| | Class 2 Device Recall Exactech |  |
| Date Initiated by Firm | August 11, 2022 |
| Create Date | September 09, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1732-2022 |
| Recall Event ID |
90279 |
| 510(K)Number | K121392 |
| Product Classification |
Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented - Product Code LZO
|
| Product | GXL acetabular liners, orthopedic hip implant component. Product Line Numbers:
a. NOVATION GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS, REF 130-32-51
b. NOVATION GXL LINER NEUTRAL, 36MM ID, GROUP 2 CUPS, REF 130-36-52
c. NOVATION GXL LINER NEUTRAL, 40MM ID, GROUP 3 CUPS, REF 130-40-53
d. NOVATION GXL LINER NEUTRAL, 40MM ID, GROUP 4 CUPS, REF 130-40-54
e. NOVATION GXL LINER NEUTRAL, 40MM ID, GROUP 5 CUPS, REF 130-40-55
f. NOVATION GXL LINER LIPPED 32MM ID, GROUP 1 CUPS, REF 132-32-51
g. NOVATION GXL LINER LIPPED 36MM ID GROUP 2, REF 132-36-52
h. NOVATION GXL LINER LIPPED 40MM ID, GROUP 3 CUPS, REF 132-40-53
i. NOVATION GXL LINER LIPPED 40MM ID, GROUP 4 CUPS, REF 132-40-54
j. NOVATION GXL LINER LIPPED 40MM ID, GROUP 5 CUPS, REF 132-40-55
k. NOVATION GXL LINER +5 LAT 40MM GROUP 3 CUPS, REF 136-40-53
l. NOVATION GXL LINER +5 LAT 40MM ID GROUP 4 CUPS, REF 136-40-54
m. NOVATION GXL LINER +5 LAT 40MM ID GROUP 5 CUPS, REF 136-40-55
n. NOVATION GXL LINER 10 DEG FACE 40MM ID, GRP 3 CUP, REF 138-40-53
o. NOVATION GXL LINER 10 DEG FACE, 40MM ID, GRP 4 CUP, REF 138-40-54
p. NOVATION GXL LINER 10 DEG FACE, 40MM ID, GRP 5 CUP, REF 138-40-55 |
| Code Information |
a. REF 130-32-51, GUDID 10885862207074, Serial Numbers: 2409984, 2409985, 2409986, 2409987, 2409988, 2409989, 2409990, 2409991, 2409993, 2409994, 2409995, 2409996, 2409997, 2409998, 2409999, 2410000, 2410001, 2410002, 2410003, 2410004, 2410005, 2410006, 2410007, 2410008, 2410009, 2410010, 2410011, 2410012, 2410013, 2410014, 2424081, 2424082, 2424083, 2424084, 2424085, 2424086, 2424087, 2424088, 2424089, 2424090, 2424091, 2424092, 2424093, 2424094, 2424096, 2424097, 2424098, 2424099, 2424100, 2433191, 2433192, 2433193, 2433194, 2433195, 2433196, 2433197, 2433198, 2433199, 2433200, 2433201, 2433202, 2433203, 2433204, 2433205, 2433206, 2433207, 2433208, 2433209, 2433210, 2433211, 2581934, 2581935, 2581936, 2581937, 2581938, 2581939, 2581940, 2581941, 2581942, 2581943, 2581944, 2581945, 2581946, 2581947, 2581948, 2581950, 2581951, 2581952, 2581953, 2581954, 2581955, 2581956, 2581957, 2581958, 2581959, 2581960, 2581961, 2581962, 2581963, 2581964, 2581965, 2581966, 2581967, 2581968, 2581969, 2581970, 2581971, 2581972, 2581973, 2581974, 2581975, 2581976, 2581977, 2581978, 2581979, 2581980, 2581981, 2581982, 2581983, 2674321, 2674322, 2674323, 2674324, 2674325, 2674326, 2674327, 2674328, 2674329, 2674330, 2674331, 2674332, 2674333, 2674334, 2674335, 2674336, 2674337, 2674338, 2674339, 2674340, 2674341, 2674342, 2674343, 2674344, 2674345, 2674346, 2674347, 2674348, 2674349, 2674350, 2674351, 2674352, 2674354, 2674355, 2674356, 2674357, 2674358, 2674359, 2674360, 2674361, 2674362, 2674363, 2674364, 2674365, 2674366, 2674367, 2674368, 2674369, 2674370, 2730574, 2730575, 2730576, 2730577, 2730578, 2730579, 2730580, 2730581, 2730582, 2730583, 2730584, 2730585, 2730586, 2730587, 2730588, 2730589, 2730590, 2730591, 2730592, 2730593, 2730594, 2730595, 2730596, 2730597, 2730598, 2730599, 2730600, 2730601, 2730602, 2730603, 2730604, 2730605, 2730606, 2730607, 2730608, 2730609, 2730610, 2730611, 2730612, 2806954, 2806955, 2806956, 2806957, 2806958, 2806959, 2806960, 2806961, 2806962, 2806963, 2806964, 2806965, 2806966, 2806967, 2806968, 2806969, 2846209, 2846210, 2846211, 2846212, 2846213, 2846214, 2846215, 2846216, 2846217, 2846218, 2846219, 2846220, 2846221, 2846222, 2846223, 2846224, 2846225, 2846226, 2846227, 2846228, 2846229, 2846230, 2846231, 2846232, 2846233, 2846234, 2846235, 2846236, 2846237, 2846238, 2846240, 2846241, 2846242, 2846243, 2846244, 2846245, 2846246, 2846247, 2870305, 2870306, 2870307, 2870308, 2870309, 2870310, 2870311, 2870312, 2870313, 2870314, 2870315, 2870316, 2870317, 2870318, 2870319, 2870320, 2870321, 2870322, 2870323, 2870324, 2870325, 2870326, 2870327, 2870328, 2870329, 2870330, 2870331, 2870332, 2870333, 2870334, 2870335, 2870336, 2870337, 2870338, 2870339, 2952437, 2952438, 2952439, 2952440, 2952441, 2952442, 2952443, 2952444, 2952445, 2952446, 2952447, 2952448, 2952449, 2952450, 2952451, 2952452, 2952453, 2952454, 2952455, 2952456, 2952457, 2952458, 2952459, 2952460, 2952462, 2952463, 2960964, 2960966, 2960967, 2960968, 2960969, 2960970, 2960971, 2960972, 2960973, 2960974, 2960975, 2960976, 2960977, 2960978, 2960979, 2960980, 2960981, 2960982, 2960983, 2960984, 2960985, 2960986, 2960987, 2960988, 3000004, 3000005, 3000006, 3000007, 3000008, 3000009, 3000010, 3000011, 3000012, 3000013, 3000014, 3000015, 3000016, 3000017, 3000018, 3000019, 3000020, 3000021, 3000022, 3000023, 3000024, 3000025, 3000026, 3000027, 3000028, 3000029, 3000030, 3543578, 3543579, 3543580, 3543581, 3543582, 3543583, 3543584, 3543585, 3543586, 3543587, 3543588, 3543589, 3543590, 3543591, 3543592, 3543593, 3543594, 3543595, 3543596, 3543597, 3543598, 3543599, 3543600, 3543601, 3543602, 3543603, 3543604, 3574998, 3574999, 3575000, 3575001, 3575002, 3575003, 3575004, 3575005, 3575006, 3575007, 3575008, 3575009, 3575010, 3575011, 3575012, 3575013, 3575014, 3575015, 3575016, 3575017, 3575018, 3575019, 3575020, 3575021, 3575022, 3575023, 3575024, 3575025, 3575026, 3575028, 3575029, 3575030, 3575031, 3575032, 3575033, 3575034, 3575035, 3604542, 3604543, 3604544, 3604546, 3604547, 3604548, 3604549, 3604550, 3604551, 3604552, 3604553, 3604554, 3604555, 3604556, 3604557, 3604558, 3604559, 3604560, 3604561, 3604562, 3604563, 3604564, 3604565, 3604566, 3604567, 3604568, 3604569, 3604570, 3604571, 3604572, 3604573, 3604574, 3644656, 3644657, 3644658, 3644659, 3644660, 3644661, 3644662, 3644663, 3644664, 3644665, 3644666, 3644667, 3644668, 3644669, 3644670, 3644671, 3644672, 3644673, 3644674, 3644675, 3644676, 3644677, 3644678, 3644679, 3644680, 3644681, 3644682, 3665320, 3665321, 3665322, 3665323, 3665324, 3665325, 3665326, 3665327, 3665328, 3665329, 3665330, 3665331, 3665332, 3665333, 3665334, 3665335, 3665336, 3665337, 3665338, 3665339, 3665340, 3665341, 3665342, 3665343, 3665344, 3665345, 3665346, 3665347, 3665348, 3665349, 3665350, 3665351, 3665352, 3665353, 3665354, 3665355, 3665356, 3665357, 3665358, 3665359, 3665360, 3665361, 3665362, 3665363, 3665364, 3665365, 3665366, 3665367, 3679522, 3679523, 3679524, 3679525, 3679526, 3679527, 3679528, 3679529, 3679530, 3679531, 3679532, 3679533, 3679534, 3679535, 3679536, 3679537, 3679538, 3679539, 3679540, 3679541, 3679542, 3679543, 3679544, 3679545, 3679546, 3679547, 3679548, 3719277, 3719278, 3719279, 3719280, 3719281, 3719282, 3719283, 3719284, 3719285, 3719286, 3719287, 3719288, 3719289, 3719290, 3719291, 3719292, 3719293, 3719294, 3719295, 3719296, 3719297, 3719298, 3719299, 3719300, 3719301, 3719302, 3719303, 3719304, 3719305, 3719306, 3719307, 3719308, 3719309, 3719310, 3719311, 3719312, 3719313, 3719314, 3719315, 3719316, 3719317, 3719318, 3719319, 3719320, 3719321, 3719322, 3719323, 3719324, 3719325, 3719326, 3720665, 3720666, 3720667, 3720668, 3720669, 3720670, 3720671, 3720672, 3720673, 3720674, 3748423, 3748424, 3748425, 3748426, 3748427, 3748428, 3748429, 3748430, 3748431, 3748432, 3748433, 3748434, 3748435, 3748436, 3748437, 3748438, 3748439, 3748440, 3748441, 3748443, 3748444, 3748445, 3748446, 3748447, 3748448, 3748449, 3748450, 3748451, 3748452, 3748453, 3748454, 3748455, 3748456, 3748457, 3748458, 3748459, 3748460, 3748461, 3748462, 3748463, 3748464, 3748465, 3748466, 3748467, 3748468, 3748469, 3748470, 3748471, 3748472, 3748581, 3748582, 3748583, 3748584, 3748585, 3748586, 3748587, 3748588, 3748589, 3748590, 3748591, 3748592, 3748593, 3748594, 3748595, 3748596, 3748597, 3748598, 3748599, 3748600, 3748601, 3748602, 3748603, 3748604, 3748605, 3748606, 3748607, 3748608, 3748609, 3748610, 3748611, 3748612, 3748613, 3748614, 3748615, 3748616, 3748617, 3748618, 3748619, 3748620, 3748621, 3748622, 3748623, 3748624, 3748625, 3748626, 3748627, 3748628, 3748629, 3748630, 3802885, 3802886, 3802887, 3802888, 3802889, 3802890, 3802891, 3802892, 3802893, 3802894, 3802895, 3802896, 3802897, 3802898, 3802899, 3802900, 3802901, 3802902, 3802903, 3802904, 3802905, 3802906, 3802907, 3802908, 3802909, 3802910, 3802911, 3802912, 3802913, 3802914, 3802915, 3802916, 3802917, 3802918, 3802919, 3802920, 3802921, 3802922, 3802923, 3802924, 3802925, 3802926, 3802927, 3802928, 3802929, 3802930, 3802931, 3802932, 3802933, 3802934, 3822551, 3822552, 3822553, 3822554, 3822555, 3822556, 3822557, 3822558, 3822559, 3822560, 3822561, 3822562, 3822563, 3822564, 3822565, 3822566, 3822567, 3822568, 3822569, 3822570, 3822571, 3822572, 3822573, 3822574, 3822575, 3822576, 3822577, 3822578, 3822579, 3822580, 3822581, 3822582, 3822583, 3822584, 3822585, 3822586, 3822587, 3822588, 3822589, 3822590, 3822591, 3822592, 3822593, 3822594, 3822595, 3822596, 3822597, 3822598, 3822599, 3822600, 3832123, 3832124, 3832125, 3832126, 3832127, 3832128, 3832129, 3832130, 3832131, 3832132, 3832133, 3832134, 3832135, 3832137, 3832138, 3832139, 3832140, 3832141, 3857365, 3857366, 3857367, 3857368, 3857369, 3857370, 3857371, 3857372, 3857373, 3857374, 3857375, 3857376, 3857377, 3857378, 3857379, 3857380, 3857381, 3857382, 3857383, 3857384, 3857385, 3857386, 3857387, 3857388, 3857389, 3857390, 3857391, 3857392, 3857393, 3857394, 3857395, 3857396, 3857397, 3857398, 3857399, 3857400, 3857401, 3857402, 3857403, 3857404, 3857405, 3857406, 3857407, 3857408, 3857409, 3857410, 3857411, 3857412, 3857413, 3857414, 3860872, 3860873, 3860874, 3860875, 3860876, 3860877, 3860878, 3860879, 3860880, 3860881, 3860882, 3860883, 3860884, 3860885, 3860886, 3860887, 3860888, 3860889, 3921900, 3921901, 3921902, 3921903, 3921904, 3921905, 3921906, 3921907, 3921908, 3921909, 3921910, 3921911, 3921912, 3921913, 3921914, 3921915, 3921916, 3921918, 3921919, 3921920, 3921921, 3921922, 3921923, 3921924, 3921925, 3921926, 3921927, 3921928, 3921929, 3921930, 3921931, 3921932, 3921933, 3921934, 3921935, 3921936, 3921937, 3921938, 3921939, 3921940, 3921941, 3921942, 3921943, 3921944, 3921945, 3921946, 3921947, 3921948, 3921949, 3921950, 3921951, 3921952, 3921953, 3921954, 3921955, 3921956, 3921957, 3921958, 3921959, 3984444, 3984445, 3984446, 3984447, 3984448, 3984449, 3984450, 3984451, 3984452, 3984453, 3984454, 3984455, 3984456, 3984457, 3984458, 3984459, 3984460, 3984461, 3986668, 3986669, 3986670, 3986671, 3986672, 3986673, 3986674, 3986675, 3986676, 3986677, 3986678, 3986679, 3986680, 3986681, 3986682, 3986683, 3986684, 3986685, 3986686, 3986687, 3986688, 3986689, 3986690, 3986691, 3986692, 3986693, 3986694, 3986695, 3986696, 3986697, 3986758, 3986759, 3986760, 3986761, 3986762, 3986763, 3986764, 3986765, 3986767, 3986768, 3986769, 3986770, 3986771, 3986772, 3986773, 3986774, 3986775, 3986776, 3986777, 3986779, 3986780, 3986782, 3986783, 3986784, 3986785, 3986786, 3986787, 3986788, 3986789, 3986790, 3986791, 3986792, 3986793, 3986794, 3986795, 3986796, 3986797, 3986798, 3986799, 3986800, 3986801, 3986802, 3986803, 3986804, 3986805, 3986806, 3986807, 3986808, 3986809, 3986810, 3986811, 3986812, 3986813, 3986814, 3986815, 3986816, 3986817, 4050066, 4050067, 4050068, 4050069, 4050070, 4050071, 4050072, 4050073, 4050074, 4050075, 4050076, 4050077, 4050078, 4050079, 4050080, 4050081, 4050082, 4050083, 4050084, 4050085, 4050086, 4050087, 4050088, 4050089, 4050090, 4050091, 4050092, 4050093, 4050094, 4050095, 4050096, 4050097, 4050098, 4050099, 4050100, 4050101, 4050102, 4050103, 4050104, 4050105, 4050106, 4050107, 4050108, 4091400, 4091401, 4091402, 4091403, 4091404, 4091405, 4091406, 4091407, 4091408, 4091409, 4091410, 4091411, 4091412, 4091413, 4091414, 4091415, 4091416, 4091417, 4091418, 4091419, 4091420, 4091421, 4091422, 4091423, 4091424, 4120672, 4120673, 4120674, 4120675, 4120676, 4120677, 4120678, 4120679, 4120680, 4120681, 4120682, 4120683, 4120684, 4120685, 4120686, 4120687, 4120688, 4120689, 4120690, 4120691, 4120692, 4120693, 4120694, 4120695, 4120696, 4120697, 4120698, 4120699, 4120700, 4120701, 4130292, 4130293, 4130294, 4130295, 4130296, 4130297, 4130298, 4130299, 4130300, 4130301, 4130302, 4130303, 4130304, 4130305, 4130306, 4130307, 4130308, 4130309, 4130310, 4130311, 4130312, 4130313, 4130314, 4130315, 4130316, 4130317, 4130318, 4130319, 4130320, 4130321, 4130322, 4130323, 4130324, 4130325, 4130326, 4130327, 4130328, 4130329, 4130330, 4130331, 4130332, 4130333, 4130334, 4130335, 4130336, 4130337, 4130338, 4130339, 4130340, 4130341, 4130342, 4130343, 4130344, 4130345, 4130346, 4154971, 4154972, 4154973, 4154974, 4154975, 4154976, 4154978, 4154979, 4154980, 4154981, 4154982, 4154983, 4154984, 4154985, 4154986, 4154987, 4154988, 4154989, 4154990, 4154991, 4154992, 4154993, 4154994, 4154995, 4154996, 4154997, 4154998, 4154999, 4155000, 4155001, 4155002, 4155003, 4155004, 4155005, 4155006, 4172941, 4172942, 4172943, 4172944, 4172945, 4172946, 4172947, 4172948, 4172949, 4172950, 4172951, 4172952, 4172953, 4172954, 4172955, 4172956, 4172957, 4172958, 4172959, 4172960, 4172961, 4172962, 4172963, 4172964, 4172965, 4172966, 4172967, 4172968, 4172969, 4172970, 4172971, 4172972, 4172973, 4172974, 4172975, 4172976, 4172977, 4172978, 4172979, 4172980, 4172981, 4172982, 4172983, 4172984, 4172985, 4172986, 4172987, 4172988, 4172989, 4172991, 4172992, 4172993, 4172994, 4172995, 4172996, 4172997, 4172998, 4172999, 4173000.4173001, 4173002, 4173003, 4173004, 4173005, 4186262, 4186263, 4186264, 4186265, 4186266, 4186267, 4186268, 4186269, 4186270, 4186271, 4186272, 4186273, 4186274, 4186275, 4186276, 4186277, 4186278, 4186279, 4186280, 4186281, 4186282, 4186283, 4186284, 4186285, 4186286, 4186287, 4186288, 4186289, 4186290, 4186291, 4186292, 4186293, 4186294, 4186295, 4186296, 4186297, 4186298, 4186299, 4186300, 4186301, 4186302, 4186303, 4186304, 4186305, 4186306, 4186307, 4186308, 4186309, 4186310, 4186311, 4186312, 4186313, 4186314, 4186315, 4186316, 4186317, 4186318, 4186319, 4186320, 4186321, 4186322, 4186323, 4186324, 4186325, 4186326, 4186327, 4186328, 4186329, 4186330, 4186331, 4186332, 4186333, 4186334, 4186335, 4186336, 4186337, 4186339, 4186340, 4186341, 4186342, 4186343, 4186344, 4186345, 4186346, 4186347, 4186348, 4186349, 4186350, 4186351, 4186352, 4186353, 4186354, 4186355, 4186356, 4186357, 4186358, 4186359, 4186360, 4186361, 4212721, 4212722, 4212723, 4212724, 4212725, 4212726, 4212727, 4212728, 4212729, 4212730, 4212731, 4212732, 4212733, 4212734, 4212735, 4212736, 4212737, 4212738, 4212739, 4212740, 4212741, 4262010, 4262011, 4262012, 4262013, 4262014, 4262015, 4262016, 4262017, 4262018, 4262019, 4262020, 4262021, 4262022, 4262023, 4262024, 4262025, 4262026, 4262027, 4262028, 4262029, 4262030, 4262031, 4262032, 4262033, 4262034, 4262035, 4262036, 4262038, 4262039, 4262090, 4262091, 4262092, 4262093, 4262094, 4262095, 4262096, 4262097, 4262098, 4262099, 4262100, 4262101, 4262102, 4262103, 4262104, 4262105, 4262106, 4262107, 4262108, 4262109, 4262110, 4262111, 4262112, 4262113, 4262114, 4262115, 4262116, 4262117, 4262118, 4262119, 4262170, 4262171, 4262172, 4262173, 4262174, 4262175, 4262176, 4262177, 4262178, 4262179, 4262180, 4262181, 4262182, 4262183, 4262184, 4262185, 4262186, 4262187, 4262188, 4262189, 4262191, 4262192, 4262193, 4262194, 4262195, 4262196, 4262197, 4262198, 4262199, 4283951, 4283952, 4283953, 4283954, 4283955, 4283956, 4283957, 4283958, 4283960, 4283961, 4283962, 4283963, 4283964, 4283965, 4283966, 4283967, 4283968, 4283969, 4283970, 4283971, 4283972, 4283973, 4283974, 4283975, 4305191, 4305192, 4325215, 4325216, 4325217, 4325218, 4325219, 4325220, 4325221, 4325222, 4325223, 4325224, 4325225, 4325226, 4325227, 4325229, 4325230, 4325231, 4325232, 4325233, 4325234, 4325235, 4325236, 4325237, 4325239, 4325240, 4325241, 4325242, 4325243, 4325244, 4325245, 4325246, 4325247, 4325248, 4325249, 4325250, 4325251, 4325252, 4325253, 4325254, 4325255, 4325256, 4325257, 4325258, 4325259, 4325260, 4325261, 4325262, 4325263, 4325264, 4325265, 4325266, 4325267, 4325268, 4325269, 4325270, 4325271, 4325272, 4325273, 4325274, 4325275, 4344271, 4344272, 4344273, 4344274, 4344275, 4344276, 4344277, 4344278, 4344279, 4344280, 4344281, 4344282, 4344283, 4344284, 4344285, 4344286, 4344288, 4344290, 4344291, 4344292, 4344293, 4344294, 4344295, 4344296, 4344297, 4344298, 4344299, 4344300, 4344301, 4344302, 4344303, 4344304, 4344305, 4344306, 4344308, 4344309, 4344310, 4344311, 4344312, 4344313, 4344315, 4344316, 4344317, 4344318, 4344319, 4344320, 4344321, 4344322, 4344324, 4344325, 4344326, 4344327, 4344328, 4344329, 4354908, 4354909, 4354910, 4354911, 4354912, 4354913, 4354914, 4354915, 4354916, 4354917, 4354918, 4354919, 4354920, 4354921, 4354922, 4354923, 4354924, 4354925, 4354926, 4354927, 4354928, 4354929, 4354930, 4354931, 4354932, 4354933, 4354934, 4354935, 4354936, 4354937, 4354938, 4354969, 4354970, 4354971, 4354972, 4354973, 4354974, 4354975, 4354976, 4354977, 4354978, 4354979, 4354980, 4354981, 4354982, 4354983, 4354984, 4354985, 4354987, 4354988, 4354989, 4354990, 4354991, 4354992, 4354993, 4354994, 4354995, 4354996, 4354997, 4354998, 4354999, 4378255, 4378256, 4378257, 4378258, 4378259, 4378260, 4378261, 4378262, 4378263, 4378264, 4378265, 4378266, 4378267, 4378268, 4378269, 4378270, 4378271, 4378272, 4378273, 4378274, 4378275, 4378276, 4378277, 4378278, 4378279, 4378280, 4378281, 4378282, 4378283, 4378284, 4378285, 4378286, 4378287, 4378288, 4378289, 4378290, 4378291, 4378292, 4378293, 4378294, 4378295, 4378296, 4409405, 4409406, 4409407, 4409408, 4409409, 4409410, 4409411, 4409412, 4409414, 4409415, 4409416, 4409417, 4409418, 4409421, 4409422, 4409423, 4409424, 4409425, 4409426, 4409427, 4409428, 4409429, 4409430, 4409431, 4409432, 4409433, 4409434, 4409435, 4409436, 4409437, 4409438, 4409439, 4409440, 4409441, 4409442, 4409443, 4409444, 4409445, 4409446, 4409447, 4409448, 4409449, 4409450, 4409451, 4409452, 4409453, 4409454, 4409455, 4409456, 4409457, 4409458, 4409459, 4409460, 4409461, 4409462, 4409463, 4409464, 4424448, 4424449, 4424450, 4424451, 4424452, 4424453, 4424454, 4424455, 4424456, 4424457, 4424458, 4424459, 4424460, 4424461, 4424462, 4424463, 4465810, 4465811, 4465812, 4465813, 4465814, 4465815, 4465816, 4465817, 4465818, 4465819, 4465820, 4465821, 4465822, 4465823, 4465824, 4465825, 4465826, 4465827, 4465828, 4465829, 4465830, 4465831, 4465832, 4465833, 4465834, 4465835, 4465836, 4465837, 4465838, 4465839, 4490836, 4490837, 4490838, 4490839, 4490840, 4490841, 4490842, 4490843, 4490844, 4490845, 4490846, 4490847, 4490848, 4490849, 4490851, 4490852, 4490853, 4490854, 4504084, 4504085, 4504086, 4504087, 4504088, 4504089, 4504090, 4504091, 4504092, 4504093, 4504094, 4504095, 4504096, 4504097, 4504099, 4504100, 4504101, 4504102, 4504103, 4504104, 4504105, 4504106, 4504107, 4504108, 4504109, 4504110, 4504111, 4504112, 4504113, 4535111, 4535112, 4535113, 4535114, 4535115, 4535116, 4535117, 4535118, 4535119, 4535120, 4535121, 4535122, 4535123, 4535124, 4535125, 4535126, 4535127, 4535128, 4535129, 4535130, 4535131, 4535132, 4535133, 4535134, 4535135, 4535136, 4535137, 4535138, 4535139, 4535140, 4535141, 4535142, 4535143, 4535144, 4535145, 4535146, 4535147, 4535148, 4535149, 4535150, 4535151, 4535152, 4535153, 4535154, 4535155, 4535156, 4535157, 4535158, 4535159, 4535160, 4535161, 4535162, 4535163, 4535164, 4535165, 4535167, 4535168, 4535169, 4535170, 4558533, 4558534, 4558535, 4558536, 4558537, 4558538, 4558539, 4558540, 4558541, 4558542, 4558543, 4558544, 4558545, 4558546, 4558547, 4558548, 4558549, 4558550, 4558551, 4558552, 4558553, 4558554, 4558555, 4558556, 4558557, 4558558, 4558559, 4558560, 4558561, 4558562, 4558683, 4558684, 4558685, 4558686, 4558687, 4558689, 4558690, 4558691, 4558692, 4558693, 4558694, 4558695, 4558696, 4558697, 4558698, 4558699, 4558700, 4558701, 4558702, 4558703, 4558704, 4558705, 4558706, 4558707, 4558708, 4558709, 4558710, 4558711, 4558712, 4578066, 4578067, 4578068, 4578069, 4578070, 4578071, 4578072, 4578073, 4578074, 4578075, 4578076, 4578077, 4578078, 4578079, 4578080, 4578081, 4578082, 4578083, 4578084, 4578085, 4578086, 4578087, 4578088, 4578089, 4578090, 4578091, 4578092, 4578093, 4578094, 4578095, 4615375, 4615376, 4615377, 4615378, 4615379, 4615380, 4615381, 4615382, 4615383, 4615384, 4615385, 4615386, 4615387, 4615388, 4615389, 4615390, 4615391, 4615392, 4615393, 4615394, 4615395, 4615396, 4615397, 4615398, 4615399, 4615400, 4615401, 4615402, 4615403, 4615404, 4625095, 4625096, 4625097, 4625098, 4625100, 4625101, 4625102, 4625104, 4625105, 4625107, 4625108, 4625109, 4625110, 4625111, 4625112, 4625113, 4625115, 4625116, 4625117, 4625118, 4625119, 4625120, 4625121, 4625122, 4625123, 4625124, 4636779, 4636780, 4636781, 4636782, 4636783, 4636784, 4636785, 4636786, 4636787, 4636788, 4662720, 4662721, 4662722, 4662723, 4662724, 4662725, 4662726, 4662727, 4662728, 4662729, 4662730, 4662731, 4662732, 4662733, 4662734, 4662735, 4662736, 4662737, 4662738, 4662739, 4662740, 4662741, 4662742, 4662743, 4662745, 4662746, 4662747, 4662749, 4698957, 4698958, 4698959, 4698960, 4698961, 4698962, 4698963, 4698964, 4698965, 4698966, 4698967, 4698968, 4698969, 4698970, 4698971, 4698972, 4698973, 4698974, 4698975, 4698976, 4698977, 4698978, 4698979, 4698980, 4698981, 4698982, 4698983, 4698984, 4698985, 4698986, 4715647, 4715649, 4715650, 4715651, 4715652, 4715653, 4715654, 4715655, 4715656, 4715657, 4715658, 4715659, 4715660, 4715661, 4715662, 4715663, 4715664, 4715665, 4715666, 4715667, 4715668, 4715669, 4715671, 4715672, 4715673, 4715674, 4715675, 4715676, 4766810, 4766811, 4766812, 4766813, 4766814, 4766815, 4766816, 4766817, 4766818, 4766819, 4766820, 4766821, 4766822, 4766823, 4766824, 4766825, 4766826, 4766827, 4766828, 4766829, 4766830, 4766831, 4766832, 4766833, 4766834, 4766835, 4766836, 4766837, 4766838, 4766839, 4766844, 4766845, 4766846, 4766847, 4766848, 4766849, 4766850, 4766851, 4766852, 4766853, 4766854, 4766855, 4766856, 4766857, 4766858, 4766859, 4766860, 4766861, 4766862, 4766863, 4766864, 4766865, 4766866, 4766867, 4766868, 4766869, 4766870, 4766871, 4766872, 4766873, 4789022, 4789023, 4789024, 4789025, 4789026, 4789027, 4789028, 4789029, 4789030, 4789031, 4789032, 4789033, 4789034, 4789035, 4789036, 4789037, 4789038, 4789039, 4789040, 4789041, 4789042, 4789043, 4789044, 4789045, 4789046, 4789047, 4789048, 4789049, 4789050, 4789051, 4800798, 4800799, 4800800, 4800801, 4800802, 4800803, 4800804, 4800805, 4800806, 4800807, 4800808, 4800809, 4800810, 4800811, 4800812, 4800813, 4800814, 4800815, 4800816, 4800817, 4800818, 4800819, 4800820, 4800821, 4800822, 4800823, 4800824, 4800825, 4800826, 4800827, 4828476, 4828477, 4828478, 4828479, 4828480, 4828481, 4828482, 4828483, 4828484, 4828485, 4828486, 4828487, 4828488, 4828489, 4828490, 4828491, 4828492, 4828493, 4828494, 4828495, 4828496, 4828497, 4828498, 4828499, 4828500, 4828501, 4828502, 4828503, 4828504, 4828505, 4845500, 4845501, 4845502, 4845503, 4845504, 4845505, 4845506, 4845507, 4845508, 4845509, 4845510, 4845511, 4845512, 4845513, 4845514, 4845515, 4845516, 4845517, 4845519, 4845520, 4845521, 4845522, 4845523, 4845525, 4845526, 4845527, 4845528, 4845529, 4861307, 4861308, 4861309, 4861310, 4861311, 4861312, 4861313, 4861314, 4861315, 4861316, 4861317, 4861318, 4861319, 4861320, 4861321, 4861322, 4861323, 4861324, 4861325, 4861326, 4861327, 4861328, 4861329, 4861330, 4861331, 4861332, 4861333, 4861334, 4861335, 4861336, 4910795, 4910796, 4910797, 4910798, 4910799, 4910800, 4910801, 4910802, 4910803, 4910804, 4910805, 4910806, 4910807, 4910808, 4910809, 4910810, 4910811, 4910812, 4910813, 4910814, 4910815, 4910816, 4910817, 4910818, 4910819, 4910820, 4910821, 4910822, 4910823, 4910824, 4930586, 4930587, 4930588, 4930589, 4930590, 4930591, 4930592, 4930593, 4930594, 4930595, 4930596, 4930597, 4930598, 4930599, 4930600, 4930601, 4930602, 4930603, 4930604, 4930605, 4930606, 4930607, 4930608, 4930609, 4930610, 4930611, 4930612, 4930613, 4930614, 4930615, 4946116, 4946117, 4946118, 4946119, 4946120, 4946121, 4946122, 4946123, 4946124, 4946125, 4946126, 4946127, 4946128, 4946129, 4946130, 4946131, 4946132, 4946133, 4946134, 4946135, 4946136, 4946137, 4946138, 4946139, 4946140, 4946141, 4946142, 4946143, 4946144, 4946145, 4966364, 4966365, 4966366, 4966367, 4966368, 4966369, 4966370, 4966371, 4966372, 4966373, 4966374, 4966375, 4966376, 4966377, 4966378, 4966379, 4966380, 4966381, 4966382, 4966383, 4966384, 4966385, 4966386, 4966387, 4966388, 4966389, 4966390, 4966391, 4966392, 4966393, 4976693, 4976694, 4976695, 4976696, 4976697, 4976698, 4976699, 4976700, 4976701, 4976702, 4976703, 4976705, 4976706, 4976707, 4976708, 4976709, 4976710, 4976711, 4976712, 4976713, 4976714, 4976715, 4976716, 4976717, 4976718, 4976719, 4976720, 4976721, 4976722, 4992696, 4992699, 4992700, 4992701, 4992702, 4992703, 4992704, 4992705, 4992706, 4992707, 4992708, 4992709, 4992710, 4992711, 4992712, 4992713, 4992714, 4992715, 4992716, 4992717, 4992718, 4992719, 4992720, 4992721, 4992722, 4992723, 4992724, 4992725, 5013440, 5013441, 5013442, 5013443, 5013444, 5013445, 5013446, 5013447, 5013448, 5013449, 5013450, 5013451, 5013452, 5013453, 5013454, 5013455, 5013456, 5013457, 5013458, 5013459, 5013460, 5013461, 5013462, 5013463, 5013464, 5013465, 5013466, 5013467, 5013468, 5013469, 5046826, 5046827, 5046828, 5046829, 5046830, 5046831, 5046832, 5046833, 5046834, 5046836, 5046837, 5046838, 5046839, 5046840, 5046841, 5046842, 5046843, 5046844, 5046845, 5046846, 5046847, 5046848, 5046849, 5046850, 5046851, 5046852, 5046853, 5046854, 5046855, 5091125, 5091126, 5091127, 5091128, 5091129, 5091130, 5091131, 5091132, 5091133, 5091134, 5091135, 5091136, 5091137, 5091138, 5091139, 5091140, 5091141, 5091142, 5091143, 5091144, 5091145, 5091146, 5091147, 5091148, 5091149, 5091150, 5091151, 5091152, 5091153, 5091154, 5091155, 5091156, 5091157, 5091158, 5094734, 5094735, 5094736, 5094737, 5094738, 5094739, 5094740, 5094741, 5094742, 5094743, 5094744, 5094745, 5094746, 5094747, 5094748, 5094749, 5094750, 5094751, 5094752, 5094753, 5094754, 5094755, 5094756, 5094757, 5094758, 5094759, 5094760, 5094761, 5094762, 5094763, 5094764, 5094765, 5094766, 5094767, 5094768, 5094769, 5094770, 5094771, 5094772, 5094773, 5094774, 5094775, 5094776, 5094777, 5094778, 5094779, 5094780, 5094781, 5094782, 5094783, 5094784, 5094785, 5094786, 5094787, 5094788, 5094789, 5094790, 5094791, 5094792, 5099065, 5099066, 5099067, 5099068, 5099069, 5099070, 5099071, 5099072, 5099073, 5099074, 5099075, 5099076, 5099077, 5099078, 5099079, 5099080, 5099081, 5099082, 5099083, 5099084, 5099085, 5099086, 5099087, 5099088, 5099089, 5099090, 5099091, 5099092, 5099093, 5099094, 5140226, 5140227, 5140228, 5140229, 5140230, 5140231, 5140232, 5140233, 5140234, 5140235, 5140236, 5140237, 5140238, 5140239, 5140240, 5140241, 5140242, 5140243, 5140244, 5140245, 5140246, 5140247, 5140248, 5140249, 5140250, 5140251, 5140252, 5140253, 5140254, 5140255, 5145246, 5145247, 5145248, 5145249, 5145250, 5145251, 5145252, 5145253, 5145254, 5145255, 5145257, 5145258, 5145259, 5145260, 5145261, 5145262, 5145263, 5145264, 5145265, 5145266, 5145267, 5145268, 5145269, 5145270, 5145271, 5145272, 5145273, 5145274, 5145275, 5201354, 5201355, 5201356, 5201357, 5201358, 5201359, 5201360, 5201361, 5201362, 5201363, 5201364, 5201365, 5201366, 5201367, 5201369, 5201370, 5201371, 5201372, 5201373, 5201374, 5201375, 5201376, 5201377, 5201378, 5201379, 5201380, 5201381, 5201382, 5201383, 5228063, 5228064, 5228065, 5228066, 5228067, 5228068, 5228069, 5228070, 5228071, 5228072, 5228073, 5228074, 5228075, 5228076, 5228077, 5228078, 5228079, 5228080, 5228081, 5228082, 5228083, 5228084, 5228085, 5228086, 5228087, 5228088, 5228089, 5228090, 5228091, 5228092, 5228093, 5250989, 5250990, 5250991, 5250992, 5250993, 5250994, 5250995, 5250996, 5250997, 5250998, 5250999, 5251000, 5251001, 5251002, 5251003, 5251004, 5251005, 5251006, 5251007, 5251008, 5251009, 5251010, 5251011, 5251012, 5251013, 5251014, 5251015, 5251016, 5251017, 5251018, 5266954, 5266955, 5266956, 5266957, 5266958, 5266959, 5266960, 5266961, 5266962, 5266963, 5266964, 5266965, 5266966, 5266967, 5266968, 5266969, 5266970, 5266971, 5266972, 5266973, 5266974, 5266975, 5266976, 5266977, 5266978, 5266979, 5266980, 5266981, 5266982, 5266983, 5272282, 5272283, 5272284, 5272285, 5272286, 5272287, 5272288, 5272289, 5272290, 5272291, 5272292, 5272293, 5272294, 5272295, 5272296, 5272297, 5272298, 5272299, 5272300, 5272301, 5272302, 5272303, 5272304, 5272305, 5272306, 5272307, 5272308, 5272309, 5272310, 5272311, 5342053, 5342054, 5342055, 5342056, 5342057, 5342058, 5342059, 5342060, 5342061, 5342062, 5342063, 5342064, 5342065, 5342066, 5342068, 5342069, 5342070, 5342071, 5342072, 5342073, 5342074, 5342075, 5342076, 5342077, 5342078, 5342079, 5342080, 5342081, 5365890, 5365891, 5365892, 5365893, 5365894, 5365895, 5365896, 5365897, 5365898, 5365899, 5365900, 5365901, 5365902, 5365903, 5365904, 5365905, 5365906, 5365907, 5365908, 5365909, 5365910, 5365911, 5365912, 5365913, 5365914, 5365915, 5365916, 5365917, 5365918, 5365919, 5401662, 5401663, 5401664, 5401665, 5401666, 5401667, 5401669, 5401670, 5401671, 5401672, 5401673, 5401674, 5401675, 5401676, 5401677, 5401678, 5401679, 5401680, 5401681, 5401682, 5401683, 5401684, 5401685, 5401686, 5401687, 5401688, 5401689, 5401690, 5401691, 5416415, 5416416, 5416417, 5416418, 5416419, 5416420, 5416421, 5416422, 5416423, 5416424, 5416425, 5416426, 5416427, 5416428, 5416429, 5416430, 5416431, 5416432, 5416433, 5416435, 5416436, 5416437, 5416438, 5416439, 5416440, 5416441, 5416442, 5416443, 5416444, 5436158, 5436159, 5436160, 5436161, 5436162, 5436164, 5436165, 5436166, 5436167, 5436168, 5436169, 5436170, 5436171, 5436172, 5436173, 5436174, 5436175, 5436176, 5436177, 5436178, 5436179, 5436180, 5436181, 5436182, 5436183, 5436184, 5436185, 5436186, 5436187, 5469633, 5469634, 5469635, 5469636, 5469637, 5469639, 5469640, 5469641, 5469642, 5469643, 5469644, 5469645, 5469646, 5469648, 5469649, 5469650, 5469651, 5469652, 5469653, 5469654, 5469655, 5469656, 5469657, 5469658, 5469659, 5469661, 5469662, 5469663, 5469664, 5469665, 5469666, 5469667, 5469668, 5489564, 5489565, 5489566, 5489567, 5489568, 5489570, 5489571, 5489572, 5489573, 5489574, 5489575, 5489576, 5489577, 5489578, 5489579, 5489580, 5489581, 5489582, 5489583, 5489584, 5489585, 5489586, 5489587, 5489588, 5489589, 5489590, 5489591, 5489592, 5489593, 5489594, 5489595, 5489596, 5489597, 5489598, 5504667, 5504668, 5504669, 5504670, 5504671, 5504673, 5504674, 5504675, 5504676, 5504677, 5504678, 5504679, 5504680, 5504681, 5504682, 5504683, 5504684, 5504685, 5504686, 5504687, 5504688, 5504689, 5504690, 5504691, 5504692, 5504693, 5504694, 5504695, 5504696, 5509564, 5509565, 5509566, 5509567, 5509568, 5509569, 5509570, 5509571, 5509572, 5509573, 5509574, 5509575, 5509576, 5509577, 5509578, 5509579, 5509580, 5509581, 5509582, 5509583, 5509584, 5509585, 5509586, 5509587, 5509588, 5509589, 5509590, 5509591, 5509592, 5551448, 5551450, 5551451, 5551452, 5551453, 5551454, 5551455, 5551456, 5551457, 5551458, 5551459, 5551460, 5551461, 5551462, 5551464, 5551465, 5551466, 5551467, 5551468, 5551469, 5551470, 5551471, 5551472, 5551473, 5551475, 5551476, 5551477, 5551891, 5551892, 5551893, 5551894, 5551895, 5551896, 5551897, 5551898, 5551899, 5551900, 5551901, 5551902, 5551903, 5551904, 5551905, 5551906, 5551907, 5551908, 5551909, 5551910, 5551912, 5551913, 5551914, 5551916, 5551917, 5551918, 5551919, 5551920, 5577698, 5577699, 5577700, 5577701, 5577702, 5577703, 5577704, 5577705, 5577706, 5577707, 5577708, 5577709, 5577710, 5577711, 5577712, 5577713, 5577714, 5577715, 5577716, 5577717, 5579488, 5579490, 5579492, 5579493, 5579494, 5579495, 5579496, 5579497, 5579498, 5579499, 5579500, 5579501, 5579502, 5579503, 5579504, 5579505, 5579506, 5579507, 5579508, 5579509, 5579510, 5579511, 5579512, 5579513, 5579514, 5579515, 5579516, 5579517, 5653544, 5653545, 5653546, 5653547, 5653548, 5653549, 5653551, 5653552, 5653553, 5653554, 5653555, 5653556, 5653557, 5653558, 5653559, 5653560, 5653561, 5653562, 5653563, 5653564, 5653565, 5653566, 5653567, 5653568, 5653895, 5653897, 5653898, 5653899, 5653900, 5653901, 5653902, 5653903, 5653904, 5653905, 5653906, 5653907, 5653909, 5653910, 5653911, 5653912, 5653913, 5653914, 5653915, 5653916, 5653917, 5653918, 5653919, 5665869, 5665870, 5665871, 5665872, 5665873, 5665874, 5665875, 5665876, 5665877, 5665878, 5665879, 5665880, 5665881, 5665882, 5665883, 5665885, 5665886, 5665887, 5665888, 5665889, 5665890, 5665893, 5665894, 5665895, 5665896, 5665897, 5665898, 5686882, 5686883, 5686884, 5686885, 5686886, 5686887, 5686888, 5686889, 5686890, 5686891, 5686892, 5686893, 5686894, 5686895, 5686896, 5686897, 5686898, 5686899, 5686900, 5686901, 5768082, 5768083, 5768084, 5768085, 5768086, 5768087, 5768088, 5768089, 5768090, 5768091, 5768092, 5768093, 5768094, 5768095, 5768096, 5768097, 5768098, 5768099, 5768100, 5768102, 5768103, 5768104, 5768105, 5768106, 5768107, 5768108, 5768109, 5768110, 5768111, 5768113, 5768114, 5768115, 5768116, 5768117, 5768118, 5768119, 5768121, 5768122, 5768123, 5768124, 5768125, 5768126, 5768128, 5768129, 5768130, 5768131, 5768132, 5768133, 5768134, 5768135, 5768136, 5768137, 5768138, 5768139, 5768140, 5768141, 5768142, 5768143, 5768144, 5768145, 5768146, 5768147, 5768148, 5768149, 5768150, 5768151, 5768152, 5768153, 5768154, 5768155, 5768156, 5768157, 5768158, 5768159, 5768160, 5768161, 5768162, 5768163, 5768164, 5768166, 5768167, 5768168, 5768169, 5768170, 5768201, 5768202, 5768203, 5768204, 5768205, 5768206, 5768207, 5768208, 5768209, 5768210, 5768211, 5768212, 5768214, 5768215, 5768216, 5768217, 5768218, 5768219, 5768221, 5768222, 5768223, 5768224, 5768225, 5768226, 5768227, 5768229, 5768231, 5768232, 5768233, 5768234, 5768235, 5768236, 5768237, 5768238, 5768239, 5768240, 5768241, 5768242, 5768243, 5768244, 5768245, 5768246, 5768247, 5768248, 5768249, 5768250, 5768251, 5768252, 5768253, 5768254, 5768256, 5768257, 5768258, 5768259, 5768260, 6060293, 6060294, 6060295, 6060296, 6060297, 6060298, 6060299, 6060300, 6060301, 6060302, 6060303, 6060305, 6060306, 6060307, 6060308, 6060309, 6060310, 6060311, 6060312, 6060313, 6060314, 6060315, 6060316, 6060317, 6172817, 6172818, 6172819, 6172820, 6172821, 6172822, 6172823, 6172824, 6172825, 6172826, 6172827, 6172828, 6172829, 6172830, 6172831, 6172832, 6172833, 6172834, 6172835, 6172836, 6172837, 6172838, 6172839, 6172840, 6172841, 6172842, 6418450, 6418451, 6418452, 6418453, 6418454, 6418455, 6418457, 6418458, 6418459, 6418460, 6418461, 6418462, 6418463, 6418464, 6418466, 6418467, 6418468, 6418469, 6418470, 6418471, 6418473, 6418474, 6418475, 6418476, 6418477, 6418478, 6418479, 6421644, 6421645, 6421647, 6421648, 6421650, 6421651, 6421652, 6421655, 6421656, 6421657, 6421658, 6421659, 6421660, 6421662, 6421664, 6510680, 6510682, 6510683, 6510684, 6510685, 6510686, 6510687, 6510688, 6510689, 6510690, 6510692, 6510693, 6510694, 6510695, 6510697, 6510698, 6510699, 6510700, 6510702, 6510703, 6510704, 6510705, 6510706, 6510708, 6510709, 6510710, 6510711, 6510712, 6510713, 6510714, 6510715, 6510717, 6510718, 6510720, 6510721, 6510722, 6510723, 6510724, 6510725, 6510726, 6510727, 6510728, 6510729, 6510730, 6510731, 6510732, 6510733, 6510734, 6956916, 6956917, 6956919, 6956920, 6956922, 6956923, 6956924, 6956927, 6956929, 6956930, 6956931, 6956932, 6956934, 6956935, 6956936, 6956937, 6956938, 6956939, 6956940, 7003458, 7003461, 7003462, 7003463, 7003465, 7003467, 7003470, 7003472, 7003473, 7003475, 7003478, 7104895, 7104896, 7104897, 7104901, 7104902, 7104907, 7104909;
b. REF 130-36-52, GUDID 10885862207081, Serial Numbers: 2410443, 2410444, 2410445, 2410446, 2410447, 2410448, 2410449, 2410450, 2410451, 2410452, 2410453, 2410454, 2410455, 2410456, 2410457, 2410458, 2410459, 2410460, 2410461, 2410462, 2410463, 2410464, 2410465, 2410466, 2410467, 2410468, 2410469, 2410470, 2410471, 2410472, 2410473, 2410474, 2410475, 2410476, 2410477, 2410478, 2410479, 2410481, 2410482, 2410483, 2410484, 2410485, 2410486, 2410487, 2410488, 2410489, 2410490, 2410491, 2429090, 2429091, 2429092, 2429093, 2429094, 2429095, 2429096, 2429097, 2429098, 2429099, 2429100, 2429101, 2429102, 2429103, 2429104, 2429105, 2429106, 2429107, 2429108, 2429109, 2429110, 2429111, 2429112, 2429113, 2429114, 2429173, 2429174, 2429175, 2429176, 2429177, 2429178, 2429179, 2429180, 2429181, 2429182, 2429183, 2429184, 2429185, 2429186, 2429187, 2429188, 2429189, 2429190, 2429191, 2429192, 2429193, 2429194, 2429195, 2429197, 2441118, 2441119, 2441120, 2441121, 2441122, 2441123, 2441124, 2441125, 2441126, 2441127, 2441128, 2441129, 2441130, 2441131, 2441132, 2441133, 2441134, 2441135, 2441136, 2441137, 2601196, 2601197, 2601198, 2601199, 2601200, 2601201, 2601202, 2601203, 2601204, 2601205, 2601206, 2601207, 2601208, 2601209, 2601210, 2601211, 2601212, 2601214, 2601215, 2601216, 2601217, 2601218, 2601219, 2601220, 2601221, 2601222, 2601223, 2601224, 2601225, 2601226, 2601227, 2659348, 2659349, 2659350, 2659351, 2659352, 2659353, 2659354, 2659355, 2659356, 2659357, 2659358, 2659359, 2659360, 2659361, 2659362, 2659363, 2659364, 2659365, 2659366, 2659367, 2659368, 2659369, 2659370, 2659371, 2659372, 2659373, 2659374, 2659375, 2659376, 2659377, 2659378, 2659379, 2659380, 2659381, 2659382, 2659383, 2659384, 2659385, 2659386, 2659387, 2659388, 2659389, 2659390, 2659391, 2659392, 2659393, 2659394, 2659395, 2659396, 2659397, 2672030, 2672031, 2672032, 2672033, 2672034, 2672035, 2672036, 2672037, 2672038, 2672039, 2672040, 2672041, 2672042, 2672044, 2672045, 2672047, 2672048, 2672049, 2672050, 2672051, 2672052, 2672053, 2672054, 2672055, 2672056, 2672057, 2672058, 2672059, 2672060, 2672061, 2672062, 2672063, 2672064, 2672065, 2672066, 2672067, 2672068, 2672069, 2672070, 2672071, 2672072, 2672073, 2672074, 2672075, 2672076, 2672077, 2672078, 2672079, 2693915, 2693916, 2693917, 2693918, 2693919, 2693920, 2693921, 2693922, 2693923, 2693924, 2693925, 2693926, 2693927, 2693928, 2693929, 2693930, 2693931, 2693932, 2693933, 2693934, 2693936, 2693938, 2693939, 2693940, 2693941, 2751008, 2751009, 2751010, 2751011, 2751012, 2751013, 2751014, 2751015, 2751016, 2751017, 2751018, 2751019, 2751020, 2751021, 2751022, 2751023, 2751024, 2751025, 2751026, 2751027, 2751028, 2751029, 2751030, 2751031, 2751032, 2751033, 2782314, 2782315, 2782316, 2782317, 2782318, 2782319, 2782320, 2782321, 2782322, 2782323, 2782324, 2799544, 2799545, 2799546, 2799547, 2799548, 2799549, 2799550, 2799551, 2799552, 2799553, 2799554, 2799555, 2799556, 2799557, 2799558, 2799559, 2799560, 2799561, 2799562, 2799563, 2799564, 2799565, 2799566, 2799567, 2799568, 2799569, 2799570, 2799571, 2799572, 2799573, 2840744, 2840745, 2840746, 2840747, 2840748, 2840749, 2840750, 2840751, 2840752, 2840753, 2840754, 2840755, 2840756, 2840757, 2840758, 2840759, 2840760, 2840761, 2840762, 2840763, 2840764, 2840765, 2840766, 2840767, 2840768, 2840769, 2840770, 2840771, 2840772, 2840773, 2840774, 2840775, 2870275, 2870276, 2870277, 2870278, 2870279, 2870280, 2870281, 2870282, 2870283, 2870284, 2870285, 2870286, 2870287, 2870288, 2870289, 2870290, 2870291, 2870292, 2870293, 2870294, 2870295, 2870296, 2870297, 2870298, 2870299, 2870300, 2870301, 2870302, 2870303, 2870304, 2891773, 2891774, 2891775, 2891776, 2891777, 2891778, 2891779, 2891780, 2891781, 2891782, 2891783, 2891784, 2891785, 2891786, 2891787, 2891788, 2891789, 2891790, 2891791, 2891792, 2891793, 2891794, 2891795, 2891796, 2891797, 2891798, 2891799, 2891800, 2891801, 2891802, 2949967, 2949968, 2949969, 2949970, 2949971, 2949972, 2949973, 2949974, 2949975, 2949976, 2949977, 2949978, 2949979, 2949980, 2949981, 2949982, 2949983, 2949984, 2949985, 2949986, 2949987, 2949988, 2949989, 2949990, 2949991, 2949992, 2949993, 2952464, 2952465, 2952466, 2952467, 2952468, 2952469, 2952470, 2952471, 2952472, 2952473, 2952474, 2952475, 2952476, 2952477, 2952478, 2952479, 2952480, 2952481, 2952482, 2952483, 2952484, 2952485, 2952486, 2952487, 2952489, 2952490, 2952491, 2952492, 2952493, 2952494, 2952495, 2952496, 2952497, 2952498, 2952499, 2952500, 2952501, 2952502, 2952503, 2952504, 2952505, 2952506, 2952507, 2952508, 2952509, 2952510, 2952511, 2952512, 2952513, 2952514, 2952515, 2952516, 2952517, 2952518, 2952519, 2952520, 2952521, 2952522, 2952524, 2952525, 2952526, 2952527, 2952528, 2968373, 2968374, 2968375, 2968376, 2968377, 2968378, 2968379, 2968380, 2968381, 2968382, 2968383, 2968384, 2968385, 2968386, 2968387, 2968388, 2968389, 2968390, 2968391, 2968392, 2968393, 2968394, 2968395, 2968396, 2968397, 2968398, 2968399, 3534389, 3534390, 3534391, 3534392, 3534393, 3534394, 3534395, 3534396, 3534397, 3534398, 3534399, 3534400, 3534401, 3534402, 3534403, 3534404, 3534405, 3534406, 3534407, 3534408, 3534409, 3534410, 3534411, 3534413, 3534414, 3534415, 3550811, 3550812, 3550813, 3550814, 3550815, 3550816, 3550817, 3550818, 3550819, 3550820, 3550821, 3550822, 3550823, 3550824, 3550825, 3550826, 3550827, 3550828, 3550829, 3550830, 3550831, 3550832, 3550833, 3550834, 3550835, 3550836, 3550837, 3550838, 3550839, 3550840, 3550841, 3550842, 3550843, 3550844, 3550845, 3550846, 3550847, 3550848, 3565810, 3565811, 3565812, 3565813, 3565814, 3565815, 3565816, 3565817, 3565818, 3565819, 3565820, 3565821, 3565822, 3565823, 3565824, 3565825, 3565826, 3565827, 3565828, 3565829, 3565830, 3565831, 3565832, 3565833, 3565834, 3565835, 3565836, 3565837, 3565838, 3565839, 3565840, 3565841, 3565842, 3565843, 3565844, 3565845, 3565846, 3565847, 3565848, 3565849, 3565850, 3565851, 3565852, 3565853, 3565854, 3565855, 3565856, 3565857, 3565858, 3565859, 3579099, 3579100, 3579101, 3579102, 3579103, 3579104, 3579105, 3579106, 3579107, 3579108, 3579109, 3579110, 3579111, 3579112, 3579113, 3579114, 3579115, 3579116, 3579117, 3579118, 3579119, 3579120, 3579121, 3579122, 3579123, 3579124, 3579125, 3579126, 3579127, 3579128, 3579129, 3579130, 3579131, 3579132, 3579133, 3579134, 3579135, 3579136, 3579137, 3579138, 3579139, 3579140, 3579141, 3579142, 3579143, 3579144, 3579145, 3579146, 3579147, 3579148, 3613003, 3613004, 3613005, 3613006, 3613007, 3613008, 3613009, 3613010, 3613011, 3613012, 3613013, 3613014, 3613015, 3613017, 3613019, 3613020, 3613021, 3613022, 3613023, 3613024, 3613025, 3613026, 3613027, 3613028, 3613029, 3613030, 3613031, 3613032, 3613033, 3613034, 3613035, 3613036, 3613037, 3613038, 3613039, 3613040, 3613041, 3613042, 3613043, 3613044, 3613045, 3613046, 3613047, 3613048, 3613049, 3613050, 3613051, 3613052, 3644555, 3644556, 3644557, 3644558, 3644559, 3644560, 3644561, 3644562, 3644563, 3644564, 3644565, 3644566, 3644567, 3644568, 3644569, 3644570, 3644571, 3644572, 3644573, 3644574, 3644575, 3644576, 3644577, 3644578, 3644579, 3644580, 3644581, 3644582, 3644583, 3644584, 3644585, 3644586, 3644587, 3644588, 3644589, 3644590, 3644591, 3644592, 3644593, 3644594, 3644595, 3686129, 3686130, 3686131, 3686132, 3686133, 3686134, 3686135, 3686136, 3686137, 3686138, 3686139, 3686140, 3686141, 3686142, 3686143, 3686144, 3686145, 3686147, 3686148, 3686149, 3686150, 3686151, 3686152, 3686153, 3686154, 3686155, 3686156, 3686157, 3686158, 3686159, 3686160, 3686161, 3686162, 3686163, 3686164, 3686165, 3686166, 3686167, 3686168, 3686169, 3686170, 3686171, 3686172, 3686173, 3686174, 3686175, 3686176, 3686177, 3686178, 3686179, 3686180, 3686181, 3686182, 3686183, 3686184, 3686185, 3686186, 3686187, 3686188, 3686189, 3686190, 3686191, 3686192, 3686193, 3686194, 3686195, 3686196, 3686197, 3686198, 3712242, 3712243, 3712244, 3712245, 3712246, 3712247, 3712248, 3712249, 3712250, 3712251, 3712252, 3712253, 3712254, 3712255, 3712256, 3712257, 3712258, 3712259, 3712260, 3712261, 3712262, 3712263, 3712264, 3712265, 3712266, 3712267, 3712268, 3712269, 3712270, 3712271, 3712272, 3712273, 3712274, 3712275, 3712276, 3712277, 3712278, 3712279, 3712280, 3712281, 3712282, 3712283, 3712284, 3712285, 3712286, 3712287, 3712288, 3712289, 3712290, 3712291, 3756905, 3756906, 3756907, 3756908, 3756909, 3756910, 3756911, 3756912, 3756913, 3756914, 3756915, 3756916, 3756917, 3756918, 3756919, 3756920, 3756921, 3756922, 3756923, 3756924, 3756925, 3756926, 3756927, 3756928, 3756929, 3756930, 3756931, 3756932, 3756933, 3756934, 3756935, 3756936, 3756937, 3756938, 3756939, 3756940, 3756941, 3756942, 3756943, 3756944, 3756945, 3756946, 3756947, 3756948, 3756949, 3756950, 3822489, 3822490, 3822491, 3822492, 3822493, 3822494, 3822495, 3822496, 3822497, 3822498, 3822499, 3822500, 3822501, 3822502, 3822503, 3822504, 3822505, 3822506, 3822507, 3822508, 3822509, 3822510, 3822511, 3822512, 3822513, 3822514, 3822515, 3822516, 3822517, 3822518, 3822519, 3822520, 3822521, 3822522, 3822523, 3822524, 3822525, 3822526, 3822527, 3822528, 3822529, 3822530, 3822531, 3822532, 3822533, 3822534, 3822535, 3822536, 3822537, 3822538, 3822539, 3822540, 3822541, 3822542, 3822543, 3822544, 3822545, 3822546, 3822547, 3822548, 3822549, 3822550, 3839978, 3839979, 3839980, 3839981, 3839982, 3839983, 3839984, 3839985, 3839986, 3839987, 3839988, 3839989, 3839990, 3839991, 3839992, 3839993, 3839994, 3839995, 3839996, 3839997, 3839998, 3839999, 3840000, 3840001, 3840002, 3840003, 3840004, 3840005, 3840006, 3840007, 3840008, 3840009, 3840010, 3840011, 3840012, 3840013, 3840014, 3840015, 3840016, 3840017, 3840018, 3840019, 3840020, 3840021, 3840022, 3840023, 3840024, 3840025, 3857315, 3857316, 3857317, 3857318, 3857319, 3857320, 3857321, 3857322, 3857323, 3857324, 3857325, 3857326, 3857327, 3857328, 3857329, 3857330, 3857331, 3857332, 3857333, 3857334, 3857335, 3857336, 3857337, 3857338, 3857339, 3857340, 3857341, 3857342, 3857343, 3857344, 3857345, 3857346, 3857347, 3857348, 3857349, 3857350, 3857351, 3857352, 3857353, 3857354, 3857355, 3857356, 3857357, 3857358, 3857359, 3857360, 3857361, 3857362, 3857363, 3857364, 3877011, 3877012, 3877013, 3877014, 3877015, 3877016, 3877017, 3877018, 3877019, 3877020, 3877021, 3877022, 3877023, 3877024, 3877025, 3877026, 3877027, 3877028, 3877029, 3877030, 3877031, 3877032, 3877033, 3877034, 3877035, 3877036, 3877037, 3877038, 3877039, 3877040, 3921997, 3921998, 3921999, 3922000, 3922001, 3922002, 3922003, 3922004, 3922005, 3922006, 3922007, 3922008, 3922009, 3922010, 3922011, 3922012, 3922013, 3922014, 3922015, 3922017, 3922018, 3922019, 3922020, 3922021, 3922022, 3922023, 3922024, 3922025, 3922026, 3922027, 3949938, 3949939, 3949940, 3949941, 3949942, 3949943, 3949944, 3949945, 3949946, 3949947, 3949948, 3949949, 3949950, 3949951, 3949952, 3949953, 3949954, 3949955, 3949956, 3949957, 3949958, 3949959, 3949960, 3949961, 3949962, 3949963, 3949964, 3949965, 3949966, 3949967, 3949968, 3949969, 3949970, 3949971, 3949972, 3949973, 3949974, 3949975, 3949976, 3949977, 3949978, 3949979, 3949980, 3949981, 3949982, 3949983, 3949984, 3983655, 3983656, 3983657, 3983658, 3983659, 3983660, 3983661, 3983662, 3983663, 3983664, 3983665, 3983666, 3983667, 3983668, 3983669, 3983670, 3983671, 3983672, 3983673, 3983674, 3983675, 3983676, 3983677, 3983678, 3983679, 3983680, 3983681, 3983682, 3983683, 3983684, 3983685, 3983686, 3983687, 3983688, 3983689, 3983690, 3983691, 3983692, 3983693, 3983694, 3983695, 3983696, 3983697, 3983698, 3983699, 3983700, 3983701, 3983702, 3983703, 3983704, 3983705, 3983706, 3983707, 3983708, 3983709, 3983710, 3983711, 3983712, 3983713, 3983714, 3983716, 3983717, 3983718, 3983719, 3983720, 3983721, 3983722, 3983723, 3983724, 3983725, 3983726, 3983727, 3983728, 3983729, 3983730, 3983731, 3983732, 3983734, 3983735, 3983736, 3983737, 3983738, 3983739, 3983740, 3983741, 3983742, 3983743, 3983744, 3983745, 3983746, 3983747, 3983748, 3983749, 3983750, 3983751, 3983752, 3983753, 3983754, 3983755, 3983756, 3983757, 3983758, 3983759, 3983760, 3983761, 3983762, 3983763, 3983765, 3983766, 3983767, 3983768, 3983769, 3983770, 3983771, 3983772, 3983773, 3983774, 3983775, 3983776, 3983777, 3983778, 3983779, 3983780, 3983781, 3983782, 3983783, 3983784, 4050017, 4050018, 4050019, 4050020, 4050021, 4050022, 4050023, 4050024, 4050025, 4050026, 4050027, 4050028, 4050029, 4050030, 4050031, 4050032, 4050033, 4050034, 4050035, 4050036, 4050037, 4050038, 4050039, 4050040, 4050041, 4050042, 4050043, 4050044, 4050045, 4050046, 4050047, 4050048, 4050049, 4050050, 4050051, 4050052, 4050053, 4050054, 4050055, 4050056, 4050057, 4050058, 4050059, 4050060, 4050061, 4050062, 4050063, 4050064, 4050065, 4056268, 4056269, 4056270, 4056271, 4056272, 4056273, 4056274, 4056276, 4056277, 4056278, 4056279, 4056280, 4056281, 4056282, 4056283, 4056284, 4056285, 4056286, 4056287, 4056288, 4056289, 4056290, 4056291, 4056292, 4056293, 4056294, 4056295, 4056296, 4056297, 4056298, 4091360, 4091361, 4091362, 4091363, 4091364, 4091365, 4091366, 4091367, 4091368, 4091369, 4091370, 4091371, 4091372, 4091373, 4091374, 4091375, 4091376, 4091377, 4091378, 4091379, 4091380, 4091381, 4091382, 4091384, 4091385, 4091386, 4091387, 4091388, 4091389, 4091390, 4091391, 4091392, 4091393, 4091394, 4091395, 4091396, 4091397, 4091398, 4091399, 4120788, 4120789, 4120790, 4120791, 4120792, 4120793, 4120794, 4120795, 4120796, 4120797, 4120798, 4120799, 4120800, 4120801, 4120802, 4120803, 4120804, 4120805, 4120806, 4120807, 4120808, 4120809, 4120810, 4120811, 4120812, 4120813, 4120814, 4120815, 4120816, 4120817, 4120818, 4120819, 4120820, 4120821, 4120822, 4120823, 4120824, 4120825, 4120826, 4120827, 4120828, 4120829, 4120830, 4120831, 4120832, 4120833, 4120834, 4120835, 4120836, 4130275, 4130276, 4130277, 4130278, 4130279, 4130280, 4130281, 4130282, 4130283, 4130284, 4130285, 4130286, 4130287, 4130288, 4130289, 4130290, 4130291, 4144930, 4144931, 4144932, 4144933, 4144934, 4144935, 4144936, 4144937, 4144938, 4144939, 4144940, 4144941, 4144942, 4144943, 4144944, 4144945, 4144946, 4144947, 4144948, 4144949, 4144950, 4144951, 4144952, 4144953, 4144954, 4144955, 4144956, 4144957, 4144958, 4144959, 4144960, 4144961, 4144962, 4144963, 4144964, 4144965, 4144966, 4144967, 4144968, 4144969, 4144970, 4144971, 4144972, 4144973, 4144974, 4144975, 4144976, 4144977, 4144978, 4144979, 4144980, 4144981, 4144982, 4144983, 4155007, 4155008, 4155009, 4155010, 4155011, 4155012, 4155013, 4155014, 4155015, 4155016, 4155017, 4155018, 4155019, 4155020, 4155021, 4155022, 4155023, 4155024, 4155025, 4155026, 4155027, 4155028, 4155029, 4187513, 4187514, 4187515, 4187516, 4187517, 4187518, 4187519, 4187520, 4187521, 4187522, 4187523, 4187524, 4187525, 4187526, 4187527, 4187528, 4187529, 4187530, 4187531, 4187532, 4187533, 4187534, 4187535, 4187536, 4187537, 4187538, 4187539, 4187540, 4187541, 4187542, 4187543, 4187544, 4187545, 4187546, 4187547, 4187548, 4187549, 4187550, 4187551, 4187552, 4187553, 4187554, 4187555, 4187556, 4187557, 4187558, 4187559, 4187560, 4187561, 4187562, 4187563, 4187565, 4187566, 4187567, 4187568, 4187569, 4187570, 4187571, 4187572, 4187573, 4187574, 4187575, 4187576, 4187577, 4187578, 4187579, 4187580, 4187581, 4187582, 4187583, 4187584, 4187585, 4187586, 4187587, 4187588, 4187589, 4187590, 4187591, 4187592, 4187593, 4187594, 4187595, 4187596, 4187597, 4187598, 4187599, 4187600, 4187601, 4187602, 4187603, 4187604, 4187605, 4187606, 4187607, 4187608, 4202417, 4202418, 4202419, 4202420, 4202421, 4202422, 4202423, 4202424, 4202425, 4202426, 4202427, 4202428, 4202429, 4202430, 4202431, 4202432, 4202433, 4202434, 4202435, 4202436, 4202437, 4202438, 4202439, 4202441, 4212742, 4212743, 4212744, 4212745, 4212746, 4212747, 4212748, 4212749, 4212750, 4212751, 4212752, 4212753, 4212754, 4212755, 4212756, 4212757, 4212758, 4212759, 4212760, 4212761, 4212762, 4212763, 4212764, 4212765, 4212766, 4212767, 4212768, 4233377, 4233378, 4233379, 4233380, 4233381, 4233382, 4233383, 4233384, 4233385, 4233386, 4233387, 4233388, 4233389, 4233390, 4233391, 4233392, 4233393, 4233394, 4233395, 4233396, 4233397, 4233398, 4233399, 4233400, 4233401, 4235145, 4235146, 4235147, 4235148, 4235149, 4235150, 4235151, 4235152, 4235153, 4235154, 4235155, 4235156, 4235157, 4235158, 4235159, 4235160, 4235161, 4235162, 4235163, 4235164, 4235165, 4235166, 4235167, 4235169, 4235170, 4235171, 4235172, 4235173, 4235174, 4235175, 4235176, 4235177, 4235178, 4235179, 4235180, 4235181, 4235183, 4235184, 4235185, 4235186, 4235187, 4235188, 4235189, 4235190, 4235191, 4235192, 4235193, 4235194, 4235195, 4247982, 4247983, 4247984, 4247985, 4247986, 4247988, 4247989, 4247990, 4247991, 4247992, 4247993, 4247994, 4247995, 4247996, 4247997, 4247998, 4247999, 4248000, 4248001, 4248002, 4248003, 4248005, 4248006, 4248007, 4248008, 4248009, 4248010, 4248011, 4248012, 4248013, 4248014, 4248015, 4248016, 4248017, 4248018, 4248019, 4248020, 4248021, 4248022, 4248023, 4248024, 4248025, 4248026, 4248027, 4248028, 4248029, 4248030, 4248031, 4248033, 4248034, 4248035, 4248036, 4248037, 4248038, 4248039, 4248040, 4248041, 4248042, 4248043, 4248044, 4248045, 4248046, 4248047, 4248048, 4248049, 4248050, 4248051, 4248052, 4248053, 4248054, 4248055, 4248056, 4248057, 4248058, 4248059, 4248060, 4248061, 4248062, 4248063, 4248064, 4248065, 4248066, 4248067, 4248068, 4248070, 4248071, 4262629, 4262630, 4262631, 4262632, 4262633, 4262634, 4262635, 4262636, 4262637, 4262638, 4262639, 4262640, 4262641, 4262642, 4262643, 4262644, 4262645, 4262646, 4262647, 4262648, 4262649, 4262650, 4262651, 4262652, 4262653, 4262654, 4262655, 4262656, 4262657, 4287105, 4287106, 4287107, 4287108, 4287109, 4287110, 4287111, 4287112, 4287113, 4287114, 4287115, 4287116, 4287117, 4287118, 4287119, 4287120, 4287121, 4287122, 4287123.4287124, 4287125, 4287126, 4287127, 4287128, 4287129, 4287130, 4287131, 4287132, 4287133, 4287135, 4287136, 4287137, 4287138, 4287139, 4287140, 4287141, 4287142, 4287143, 4287144, 4287145, 4287146, 4287147, 4287148, 4287149, 4287150, 4287151, 4287152, 4287153, 4287154, 4287155, 4287156, 4287157, 4287158, 4287159, 4287160, 4287161, 4287162, 4287163, 4287164, 4287165, 4287166, 4287167, 4287168, 4287169, 4287170, 4304864, 4304865, 4304866, 4304867, 4304868, 4304869, 4304870, 4304871, 4304872, 4304873, 4304874, 4304875, 4304876, 4304877, 4304878, 4304879, 4304880, 4304881, 4304882, 4304883, 4304884, 4304885, 4304886, 4304887, 4304888, 4304889, 4304890, 4304891, 4304892, 4304893, 4304894, 4304895, 4304896, 4304897, 4304898, 4304899, 4304900, 4304901, 4304902, 4304903, 4304904, 4304905, 4304906, 4304907, 4304908, 4304909, 4304910, 4304911, 4304912, 4304913, 4338641, 4338642, 4338643, 4338644, 4338645, 4338646, 4338647, 4338648, 4338649, 4338650, 4338651, 4338652, 4338653, 4338654, 4338655, 4338656, 4338657, 4338658, 4338659, 4338660, 4338661, 4338662, 4338663, 4338664, 4338665, 4338666, 4338667, 4338668, 4338669, 4338670, 4338671, 4338672, 4338673, 4338675, 4338676, 4338677, 4338678, 4338679, 4338680, 4338681, 4338682, 4338683, 4338684, 4338685, 4338686, 4338687, 4338688, 4338689, 4338690, 4338691, 4338692, 4338693, 4338694, 4338695, 4338697, 4338698, 4338699, 4338700, 4338701, 4338702, 4338703, 4338704, 4338705, 4338706, 4338707, 4338708, 4338709, 4338710, 4338711, 4338712, 4338713, 4338714, 4338715, 4338716, 4338717, 4338718, 4338719, 4338720, 4338721, 4338722, 4338723, 4338724, 4338725, 4338726, 4338727, 4338728, 4338729, 4338730, 4342053, 4342054, 4342055, 4342056, 4342057, 4342058, 4342059, 4342060, 4342061, 4342062, 4342063, 4342064, 4342065, 4342066, 4369554, 4369555, 4369556, 4369557, 4369558, 4369559, 4369560, 4369561, 4369562, 4369563, 4369564, 4369565, 4369566, 4369567, 4369568, 4369569, 4369570, 4369571, 4369573, 4369574, 4369576, 4369577, 4369578, 4369579, 4369580, 4369581, 4369582, 4369583, 4369584, 4369585, 4369586, 4369587, 4369588, 4369589, 4369590, 4369591, 4370097, 4370098, 4370099, 4370100, 4370102, 4370103, 4370104, 4370105, 4370106, 4370107, 4370108, 4370109, 4370110, 4370111, 4370112, 4370113, 4370114, 4370115, 4370116, 4370117, 4370118, 4370119, 4370120, 4370121, 4370122, 4370123, 4370124, 4370125, 4370126, 4377395, 4377396, 4377397, 4377398, 4377399, 4377400, 4377401, 4377402, 4377403, 4377404, 4377405, 4377406, 4377407, 4377408, 4377409, 4377410, 4377411, 4377412, 4377413, 4377414, 4377415, 4377416, 4377417, 4377418, 4377419, 4377420, 4377421, 4377422, 4377423, 4377424, 4377425, 4377426, 4377427, 4377428, 4377429, 4377430, 4377431, 4377432, 4377433, 4377434, 4377435, 4377436, 4377437, 4377438, 4377439, 4377440, 4377441, 4377442, 4377443, 4377444, 4377445, 4377446, 4377447, 4377448, 4377449, 4377450, 4377451, 4377452, 4377453, 4377454, 4402109, 4402110, 4402111, 4402112, 4402113, 4402114, 4402115, 4402116, 4402117, 4402118, 4402119, 4402120, 4402121, 4402122, 4402123, 4402124, 4402125, 4402127, 4402128, 4402129, 4402130, 4402131, 4402132, 4402133, 4402134, 4402135, 4402136, 4402137, 4402138, 4402139, 4402140, 4402141, 4402142, 4402143, 4402144, 4402145, 4402146, 4402147, 4402148, 4402149, 4402150, 4402151, 4402153, 4402154, 4402155, 4402156, 4402157, 4402158, 4402159, 4402160, 4402161, 4402162, 4402163, 4402164, 4402165, 4402166, 4402167, 4402168, 4402169, 4402170, 4402171, 4402172, 4402173, 4402174, 4402175, 4402176, 4402177, 4402178, 4402179, 4402180, 4402181, 4402182, 4402183, 4402184, 4402185, 4402186, 4402187, 4402188, 4402189, 4402190, 4402191, 4402192, 4402193, 4402194, 4402195, 4402196, 4402197, 4402198, 4402199, 4402200, 4402201, 4402202, 4402203, 4402204, 4402205, 4402206, 4402207, 4402208, 4402209, 4402210, 4462867, 4462868, 4462869, 4462870, 4462871, 4462872, 4462873, 4462874, 4462875, 4462876, 4462877, 4462878, 4462879, 4462880, 4462881, 4462882, 4462884, 4462885, 4462886, 4462887, 4462888, 4462889, 4462890, 4462891, 4462892, 4462893, 4462894, 4462895, 4462896, 4470739, 4470740, 4470741, 4470742, 4470743, 4470744, 4470745, 4470746, 4470747, 4470748, 4470749, 4470750, 4470751, 4470752, 4470753, 4470754, 4470755, 4470756, 4470757, 4470758, 4470759, 4470760, 4470761, 4470762, 4470763, 4470764, 4470765, 4470766, 4470767, 4470768, 4470769, 4470770, 4490915, 4490917, 4490918, 4490919, 4490920, 4490921, 4490922, 4490923, 4490924, 4490925, 4490926, 4490927, 4490928, 4490929, 4490930, 4490931, 4490932, 4490933, 4490934, 4490935, 4490936, 4490937, 4490938, 4490939, 4490940, 4490941, 4490942, 4490943, 4490944, 4491049, 4491050, 4491051, 4491052, 4491053, 4491054, 4491055, 4491056, 4491057, 4491058, 4491059, 4491060, 4491061, 4491062, 4491063, 4491064, 4491065, 4491066, 4491067, 4491068, 4491069, 4491070, 4491071, 4491072, 4491073, 4491074, 4491075, 4491076, 4491077, 4491078, 4491079, 4491080, 4491081, 4491082, 4491083, 4491084, 4491085, 4491086, 4491087, 4491088, 4491089, 4491090, 4491091, 4491092, 4491093, 4491094, 4491095, 4491096, 4491097, 4491098, 4491099, 4491100, 4491101, 4491102, 4491103, 4491104, 4491105, 4491106, 4491107, 4491108, 4491109, 4491110, 4491111, 4491112, 4491113, 4491114, 4491115, 4491116, 4491117, 4491118, 4491119, 4491120, 4491121, 4491122, 4491123, 4491124, 4491125, 4491127, 4491128, 4491129, 4491130, 4491131, 4491132, 4491133, 4491134, 4491135, 4491137, 4491138, 4534307, 4534308, 4534309, 4534310, 4534311, 4534312, 4534313, 4534314, 4534315, 4534316, 4534317, 4534318, 4534319, 4534320, 4534321, 4534322, 4534323, 4534324, 4534325, 4534326, 4534327, 4534328, 4534329, 4534330, 4534331, 4534332, 4534333, 4534334, 4534335, 4534336, 4534337, 4534338, 4534339, 4534340, 4534341, 4534342, 4534343, 4534344, 4534345, 4534346, 4534347, 4534348, 4534349, 4534350, 4534351, 4534352, 4534353, 4534354, 4534355, 4534356, 4534357, 4534358, 4534359, 4534360, 4534361, 4534362, 4534363, 4534364, 4534365, 4534366, 4547134, 4547135, 4547136, 4547137, 4547138, 4547139, 4547140, 4547141, 4547142, 4547143, 4547144, 4547145, 4547146, 4547147, 4547148, 4547149, 4547150, 4547151, 4547152, 4547153, 4547154, 4547155, 4547156, 4547157, 4547158, 4547159, 4547160, 4547161, 4547162, 4547163, 4577858, 4577859, 4577860, 4577861, 4577862, 4577863, 4577864, 4577865, 4577866, 4577868, 4577869, 4577870, 4577871, 4577872, 4577873, 4577875, 4577876, 4577877, 4577878, 4577879, 4577880, 4577882, 4577883, 4577885, 4577886, 4577887, 4577888, 4577889, 4577890, 4577891, 4577892, 4577893, 4577894, 4577895, 4577896, 4577897, 4577898, 4577899, 4577900, 4577901, 4577902, 4577903, 4577904, 4577905, 4577906, 4577907, 4577908, 4577909, 4577910, 4577911, 4577912, 4577913, 4577914, 4577915, 4577916, 4577917, 4578036, 4578037, 4578038, 4578039, 4578040, 4578041, 4578042, 4578043, 4578044, 4578045, 4578046, 4578047, 4578048, 4578049, 4578050, 4578051, 4578052, 4578053, 4578054, 4578055, 4578056, 4578057, 4578058, 4578059, 4578060, 4578061, 4578062, 4578063, 4578064, 4578065, 4625275, 4625277, 4625278, 4625279, 4625280, 4625281, 4625282, 4625283, 4625284, 4625285, 4625286, 4625287, 4625288, 4625289, 4625290, 4625291, 4625292, 4625293, 4625294, 4625295, 4625296, 4625297, 4625298, 4625299, 4625300, 4625301, 4625302, 4625303, 4625304, 4655794, 4655796, 4655797, 4655798, 4655799, 4655800, 4655801, 4655802, 4655803, 4655804, 4655805, 4655806, 4655807, 4655808, 4655809, 4655810, 4655811, 4655812, 4655813, 4655814, 4655815, 4655816, 4655817, 4655818, 4655819, 4655820, 4655821, 4655822, 4655823, 4655824, 4655825, 4655826, 4655827, 4655828, 4655829, 4655830, 4655831, 4655832, 4655833, 4655834, 4655835, 4655836, 4655837, 4655838, 4655839, 4655840, 4655841, 4655842, 4655843, 4655844, 4655845, 4655846, 4655847, 4655848, 4655849, 4655850, 4655851, 4655852, 4655853, 4683245, 4683246, 4683247, 4683248, 4683249, 4683250, 4683252, 4683253, 4683254, 4683255, 4683256, 4683257, 4683258, 4683260, 4683261, 4683262, 4683263, 4683264, 4683265, 4683266, 4683267, 4683268, 4683270, 4683271, 4683272, 4683273, 4683274, 4698927, 4698928, 4698929, 4698930, 4698931, 4698932, 4698933, 4698934, 4698935, 4698936, 4698937, 4698938, 4698939, 4698940, 4698941, 4698942, 4698943, 4698944, 4698945, 4698946, 4698947, 4698948, 4698949, 4698950, 4698951, 4698952, 4698953, 4698954, 4698955, 4698956, 4715587, 4715588, 4715589, 4715590, 4715591, 4715592, 4715593, 4715594, 4715595, 4715596, 4715597, 4715598, 4715599, 4715600, 4715601, 4715602, 4715603, 4715604, 4715605, 4715606, 4715607, 4715608, 4715609, 4715610, 4715611, 4715612, 4715613, 4715614, 4715615, 4715616, 4727497, 4727498, 4727499, 4727500, 4727501, 4727502, 4727503, 4727504, 4727505, 4727506, 4727507, 4727508, 4727509, 4727510, 4727511, 4727512, 4727513, 4727514, 4727515, 4727516, 4727517, 4727518, 4727519, 4727520, 4727521, 4727522, 4727523, 4727524, 4727525, 4727526, 4738550, 4738551, 4738552, 4738553, 4738554, 4738555, 4738556, 4738557, 4738558, 4738559, 4738560, 4738561, 4738562, 4738563, 4738564, 4738565, 4738566, 4738567, 4738568, 4738569, 4738570, 4738571, 4738572, 4738573, 4738574, 4738575, 4738576, 4738577, 4738578, 4738579, 4766979, 4766980, 4766981, 4766982, 4766983, 4766984, 4766985, 4766986, 4766987, 4766988, 4766989, 4766990, 4766991, 4766992, 4766993, 4766994, 4766995, 4766996, 4766997, 4766998, 4766999, 4767000, 4767001, 4767002, 4767003, 4767004, 4767005, 4767006, 4767007, 4767008, 4767009, 4767010, 4767011, 4767012, 4767013, 4767014, 4767015, 4767016, 4767017, 4767018, 4767019, 4767020, 4767021, 4767022, 4767023, 4767024, 4767025, 4767026, 4767027, 4767028, 4767029, 4767030, 4767031, 4767032, 4767033, 4767034, 4767035, 4767036, 4767037, 4767038, 4788839, 4788840, 4788841, 4788842, 4788843, 4788844, 4788845, 4788846, 4788847, 4788848, 4788849, 4788850, 4788851, 4788852, 4788853, 4788854, 4788855, 4788856, 4788857, 4788858, 4788859, 4788860, 4788861, 4788862, 4788863, 4788864, 4788865, 4788866, 4788867, 4788868, 4792008, 4792009, 4792010, 4792011, 4792012, 4792013, 4792014, 4792015, 4792016, 4792017, 4792018, 4792019, 4792020, 4792021, 4792022, 4792023, 4792024, 4792025, 4792026, 4792027, 4792028, 4792029, 4792031, 4792032, 4792033, 4792034, 4792035, 4792036, 4792037, 4817225, 4817226, 4817227, 4817228, 4817229, 4817230, 4817231, 4817232, 4817233, 4817234, 4817235, 4817236, 4817237, 4817238, 4817239, 4817241, 4817242, 4817243, 4817244, 4817245, 4817246, 4817247, 4817248, 4817249, 4817250, 4817251, 4817252, 4817253, 4817254, 4817255, 4817256, 4817257, 4817258, 4817259, 4817260, 4817261, 4817262, 4817263, 4817264, 4817265, 4817266, 4817267, 4817268, 4817269, 4817270, 4817271, 4817272, 4817273, 4817274, 4817275, 4817276, 4817277, 4817278, 4817279, 4817280, 4817281, 4817282, 4817283, 4817284, 4837077, 4837078, 4837079, 4837080, 4837081, 4837082, 4837083, 4837084, 4837085, 4837086, 4837087, 4837088, 4837089, 4837090, 4837091, 4837092, 4837093, 4837094, 4837095, 4837096, 4837097, 4837098, 4837099, 4837100, 4837101, 4837102, 4837103, 4837104, 4837105, 4837106, 4853905, 4853906, 4853907, 4853908, 4853909, 4853911, 4853912, 4853913, 4853914, 4853916, 4853917, 4853918, 4853919, 4853920, 4853921, 4853922, 4853923, 4853924, 4853926, 4853927, 4853928, 4853929, 4853930, 4853931, 4853932, 4853933, 4853934, 4883217, 4883218, 4883219, 4883220, 4883221, 4883222, 4883223, 4883224, 4883225, 4883226, 4883227, 4883228, 4883229, 4883230, 4883231, 4883232, 4883233, 4883234, 4883235, 4883236, 4883237, 4883238, 4883239, 4883240, 4883241, 4883242, 4883243, 4883244, 4902319, 4902320, 4902321, 4902322, 4902323, 4902324, 4902325, 4902326, 4902327, 4902328, 4902329, 4902330, 4902331, 4902332, 4902333, 4902334, 4902335, 4902337, 4902338, 4902339, 4902340, 4902341, 4902342, 4902343, 4902344, 4902345, 4902346, 4902347, 4902348, 4906276, 4906277, 4906278, 4906279, 4906280, 4906281, 4906282, 4906283, 4906284, 4906285, 4906286, 4906287, 4906288, 4906289, 4906290, 4906291, 4906292, 4906293, 4906294, 4906295, 4906296, 4906297, 4906298, 4906299, 4906300, 4906301, 4906302, 4906303, 4906304, 4906305, 4930616, 4930617, 4930618, 4930619, 4930620, 4930621, 4930622, 4930623, 4930624, 4930625, 4930626, 4930627, 4930628, 4930629, 4930630, 4930631, 4930632, 4930633, 4930634, 4930635, 4930636, 4930637, 4930638, 4930639, 4930640, 4930641, 4930642, 4930643, 4930644, 4930645, 4940991, 4940993, 4940994, 4940995, 4940996, 4940997, 4940998, 4940999, 4941000, 4941001, 4941002, 4941003, 4941004, 4941005, 4941006, 4941007, 4941008, 4941009, 4941010, 4941011, 4941012, 4941013, 4941014, 4941015, 4941016, 4941017, 4941018, 4941019, 4941020, 4950206, 4950207, 4950208, 4950209, 4950210, 4950211, 4950212, 4950213, 4950214, 4950215, 4950216, 4950217, 4950218, 4950219, 4950220, 4950221, 4950222, 4950223, 4950224, 4950225, 4950226, 4950227, 4950228, 4950229, 4950230, 4950231, 4950232, 4950233, 4950234, 4950235, 4966056, 4966057, 4966058, 4966059, 4966060, 4966061, 4966062, 4966063, 4966064, 4966065, 4966066, 4966067, 4966068, 4966069, 4966070, 4966071, 4966072, 4966073, 4966074, 4966075, 4966076, 4966077, 4966078, 4966079, 4966080, 4966081, 4966082, 4966083, 4966084, 4966085, 4974331, 4974332, 4974333, 4974334, 4974335, 4974336, 4974337, 4974338, 4974339, 4974340, 4974341, 4974342, 4974343, 4974344, 4974345, 4974346, 4974347, 4974348, 4974349, 4974350, 4974351, 4974352, 4974353, 4974354, 4974355, 4974356, 4974357, 4974358, 4974359, 4974360, 4992207, 4992208, 4992209, 4992210, 4992211, 4992212, 4992213, 4992214, 4992215, 4992216, 4992217, 4992218, 4992219, 4992220, 4992221, 4992222, 4992223, 4992224, 4992225, 4992226, 4992227, 4992228, 4992229, 4992230, 4992231, 4994840, 4994841, 4994842, 4994843, 4994844, 4994845, 4994846, 4994847, 4994848, 4994849, 4994850, 4994851, 4994852, 4994853, 4994854, 4994855, 4994856, 4994857, 4994858, 4994859, 4994860, 4994861, 4994862, 4994863, 4994864, 4994865, 4994866, 4994867, 4994868, 4994869, 5011003, 5011004, 5011005, 5011006, 5011007, 5011009, 5011010, 5011011, 5011012, 5011013, 5011014, 5011015, 5011016, 5011017, 5011018, 5011019, 5011020, 5011021, 5011022, 5011023, 5011024, 5011025, 5011026, 5011027, 5011028, 5011029, 5011030, 5011031, 5011032, 5046754, 5046755, 5046756, 5046757, 5046758, 5046759, 5046760, 5046761, 5046762, 5046763, 5046764, 5046765, 5046766, 5046767, 5046768, 5046769, 5046770, 5046771, 5046772, 5046773, 5046774, 5046775, 5046778, 5046779, 5046780, 5046781, 5046782, 5046783, 5078764, 5078765, 5078766, 5078767, 5078768, 5078769, 5078770, 5078771, 5078772, 5078773, 5078774, 5078775, 5078776, 5078777, 5078778, 5078779, 5078780, 5078781, 5078782, 5078783, 5078784, 5078785, 5078786, 5078787, 5078788, 5078789, 5078790, 5078791, 5078792, 5078793, 5090952, 5090953, 5090954, 5090955, 5090956, 5090957, 5090958, 5090959, 5090960, 5090962, 5090963, 5090964, 5090965, 5090966, 5090967, 5090968, 5090969, 5090970, 5090971, 5090972, 5090973, 5090974, 5090975, 5090977, 5090978, 5090979, 5090981, 5090982, 5090983, 5094634, 5094635, 5094636, 5094637, 5094638, 5094639, 5094640, 5094641, 5094642, 5094643, 5094644, 5094645, 5094646, 5094647, 5094648, 5094649, 5094650, 5094652, 5094653, 5094654, 5094655, 5094656, 5094657, 5094658, 5094660, 5094661, 5094662, 5094663, 5097929, 5097930, 5097931, 5097932, 5097933, 5097934, 5097935, 5097936, 5097937, 5097938, 5097939, 5097940, 5097941, 5097942, 5097943, 5097944, 5097945, 5097946, 5097947, 5097948, 5097949, 5097950, 5097951, 5097952, 5097953, 5097954, 5097955, 5097957, 5097958, 5099035, 5099036, 5099037, 5099038, 5099039, 5099040, 5099041, 5099042, 5099043, 5099044, 5099045, 5099046, 5099047, 5099048, 5099049, 5099050, 5099051, 5099052, 5099053, 5099054, 5099055, 5099056, 5099057, 5099058, 5099059, 5099060, 5099061, 5099062, 5099063, 5099064, 5119402, 5119403, 5119404, 5119405, 5119406, 5119407, 5119408, 5119409, 5119410, 5119411, 5119412, 5119413, 5119414, 5119415, 5119416, 5119417, 5119418, 5119419, 5119420, 5119421, 5119422, 5119423, 5119425, 5119426, 5119427, 5119428, 5119429, 5119430, 5119431, 5133277, 5133278, 5133279, 5133280, 5133281, 5133282, 5133283, 5133284, 5133285, 5133286, 5133287, 5133288, 5133289, 5133290, 5133291, 5133292, 5133293, 5133294, 5133295, 5133296, 5133297, 5133298, 5133299, 5133300, 5133301, 5133302, 5133303, 5133305, 5133306, 5145200, 5145201, 5145203, 5145204, 5145205, 5145206, 5145207, 5145208, 5145209, 5145210, 5145211, 5145212, 5145213, 5145214, 5145215, 5145216, 5145217, 5145218, 5145219, 5145220, 5145221, 5145222, 5145223, 5145224, 5145225, 5145226, 5145227, 5145228, 5145229, 5175461, 5175462, 5175463, 5175464, 5175465, 5175466, 5175467, 5175468, 5175469, 5175470, 5175471, 5175472, 5175473, 5175474, 5175475, 5175476, 5175477, 5175478, 5175479, 5175480, 5175481, 5175482, 5175485, 5175486, 5175487, 5175488, 5175489, 5175490, 5189639, 5189640, 5189641, 5189642, 5189643, 5189644, 5189645, 5189646, 5189647, 5189648, 5189649, 5189650, 5189651, 5189652, 5189653, 5189654, 5189655, 5189656, 5189657, 5189658, 5189659, 5189660, 5189661, 5189662, 5189663, 5189665, 5189666, 5189667, 5189668, 5203498, 5203499, 5203500, 5203501, 5203502, 5203503, 5203504, 5203505, 5203506, 5203507, 5203508, 5203509, 5203510, 5203511, 5203512, 5203513, 5203514, 5203515, 5203516, 5203517, 5203519, 5203520, 5203521, 5203522, 5203523, 5203524, 5203525, 5203526, 5203527, 5220892, 5220893, 5220894, 5220895, 5220896, 5220897, 5220898, 5220900, 5220901, 5220902, 5220903, 5220904, 5220905, 5220906, 5220907, 5220908, 5220909, 5220910, 5220911, 5220912, 5220913, 5220914, 5220915, 5220916, 5220917, 5220918, 5220919, 5220920, 5220921, 5236661, 5236662, 5236663, 5236664, 5236665, 5236666, 5236667, 5236668, 5236669, 5236670, 5236671, 5236672, 5236673, 5236674, 5236675, 5236676, 5236677, 5236678, 5236679, 5236680, 5236681, 5236682, 5236684, 5236685, 5236686, 5236687, 5236688, 5236689, 5236690, 5251351, 5251352, 5251353, 5251354, 5251355, 5251356, 5251357, 5251358, 5251359, 5251360, 5251361, 5251362, 5251363, 5251364, 5251365, 5251366, 5251367, 5251368, 5251369, 5251370, 5251371, 5251373, 5251374, 5251375, 5251377, 5251378, 5251379, 5251380, 5256116, 5256117, 5256118, 5256119, 5256120, 5256121, 5256122, 5256123, 5256124, 5256125, 5256126, 5256127, 5256128, 5256129, 5256130, 5256131, 5256132, 5256133, 5256134, 5256135, 5256136, 5256137, 5256138, 5256139, 5256140, 5256141, 5256142, 5256143, 5256144, 5256145, 5256146, 5268683, 5268684, 5268685, 5268686, 5268687, 5268688, 5268689, 5268690, 5268691, 5268692, 5268693, 5268694, 5268695, 5268696, 5268697, 5268698, 5268699, 5268700, 5268701, 5268702, 5268703, 5268704, 5268705, 5268706, 5268707, 5268708, 5268709, 5268710, 5268711, 5268712, 5276691, 5276692, 5276693, 5276694, 5276695, 5276696, 5276697, 5276698, 5276699, 5276700, 5276701, 5276702, 5276703, 5276704, 5276705, 5276706, 5276707, 5276708, 5276709, 5276710, 5276711, 5276712, 5276713, 5276714, 5276715, 5276716, 5276717, 5276718, 5276719, 5276720, 5329026, 5329027, 5329028, 5329029, 5329030, 5329031, 5329032, 5329033, 5329034, 5329035, 5329036, 5329037, 5329038, 5329039, 5329040, 5329041, 5329043, 5329044, 5329045, 5329046, 5329047, 5329048, 5329049, 5329050, 5329051, 5329052, 5329053, 5329054, 5329055, 5342023, 5342024, 5342025, 5342026, 5342027, 5342028, 5342029, 5342030, 5342031, 5342032, 5342033, 5342034, 5342035, 5342036, 5342037, 5342038, 5342039, 5342040, 5342041, 5342042, 5342043, 5342044, 5342045, 5342046, 5342047, 5342048, 5342049, 5342050, 5342051, 5342052, 5350384, 5350385, 5350386, 5350387, 5350388, 5350389, 5350390, 5350391, 5350392, 5350393, 5350394, 5350395, 5350396, 5350397, 5350398, 5350399, 5350400, 5350401, 5350402, 5350403, 5350404, 5350405, 5350406, 5350407, 5350408, 5350409, 5350410, 5350411, 5350412, 5350413, 5350414, 5350415, 5350416, 5350417, 5382172, 5382173, 5382174, 5382175, 5382176, 5382177, 5382178, 5382179, 5382180, 5382181, 5382182, 5382183, 5382184, 5382185, 5382186, 5382187, 5382188, 5382189, 5382190, 5382191, 5382192, 5382193, 5382194, 5382195, 5382196, 5382197, 5382198, 5382199, 5382200, 5382201, 5405193, 5405194, 5405195, 5405196, 5405197, 5405198, 5405199, 5405200, 5405201, 5405202, 5405203, 5405204, 5405205, 5405206, 5405207, 5405208, 5405209, 5405210, 5405211, 5405212, 5405213, 5405214, 5405216, 5405217, 5405218, 5405219, 5405220, 5405221, 5405222, 5405223, 5405224, 5405225, 5405226, 5427082, 5427083, 5427084, 5427085, 5427086, 5427087, 5427088, 5427089, 5427090, 5427091, 5427092, 5427093, 5427094, 5427095, 5427096, 5427097, 5427098, 5427099, 5427101, 5427102, 5427103, 5427104, 5427105, 5427106, 5427107, 5427108, 5427109, 5427110, 5427112, 5442699, 5442700, 5442701, 5442703, 5442704, 5442705, 5442707, 5442708, 5442709, 5442710, 5442711, 5442712, 5442713, 5442714, 5442715, 5442716, 5442717, 5442718, 5442719, 5442720, 5442721, 5442722, 5442723, 5442724, 5442725, 5442726, 5442727, 5442728, 5461834, 5461835, 5461836, 5461837, 5461838, 5461839, 5461840, 5461841, 5461842, 5461843, 5461844, 5461845, 5461846, 5461847, 5461848, 5461849, 5461850, 5461851, 5461852, 5461853, 5461854, 5461855, 5461856, 5461857, 5461858, 5461859, 5461860, 5461861, 5461862, 5461863, 5484190, 5484191, 5484192, 5484193, 5484194, 5484195, 5484196, 5484197, 5484198, 5484200, 5484201, 5484202, 5484203, 5484204, 5484205, 5484206, 5484207, 5484208, 5484209, 5484210, 5484211, 5484212, 5484213, 5484214, 5484215, 5484216, 5484217, 5484219, 5499209, 5499210, 5499211, 5499212, 5499213, 5499214, 5499215, 5499216, 5499217, 5499218, 5499219, 5499221, 5499222, 5525588, 5525589, 5525590, 5525591, 5525592, 5525593, 5525594, 5525595, 5525596, 5525597, 5525598, 5525599, 5525600, 5525601, 5525602, 5525603, 5525604, 5525605, 5525606, 5525607, 5525608, 5525609, 5525610, 5525611, 5525612, 5525613, 5525614, 5525615, 5525616, 5525617, 5526797, 5526798, 5526800, 5526801, 5526802, 5526803, 5526804, 5526805, 5526806, 5526807, 5526808, 5526809, 5526810, 5526811, 5526812, 5526813, 5526814, 5526815, 5526816, 5526817, 5526818, 5526819, 5526820, 5526821, 5526822, 5526823, 5526824, 5526825, 5526826, 5526827, 5526828, 5526829, 5565661, 5565662, 5565663, 5565664, 5565666, 5565667, 5565668, 5565669, 5565670, 5565671, 5565672, 5565673, 5565674, 5565675, 5565676, 5565677, 5565678, 5565680, 5565681, 5565682, 5565683, 5565684, 5565685, 5565686, 5565687, 5565688, 5565689, 5565690, 5565691, 5565692, 5565693, 5566838, 5566839, 5566840, 5566841, 5566842, 5566843, 5566844, 5566845, 5566846, 5566847, 5566848, 5566849, 5566850, 5566851, 5566852, 5566853, 5566854, 5566855, 5566856, 5566857, 5566858, 5566859, 5566860, 5566861, 5566862, 5566863, 5566864, 5566866, 5566867, 5566868, 5566869, 5657530, 5657531, 5657532, 5657533, 5657534, 5657536, 5657537, 5657538, 5657539, 5657540, 5657541, 5657542, 5657543, 5657544, 5657545, 5657546, 5657547, 5657548, 5657550, 5657551, 5657552, 5657553, 5657554, 5657555, 5657556, 5657557, 5657558, 5657559, 5657560, 5657561, 5657562, 5657563, 5657564, 5657565, 5657566, 5657567, 5657568, 5657570, 5657571, 5657572, 5657573, 5657574, 5657575, 5657576, 5657577, 5665680, 5665681, 5665682, 5665683, 5665685, 5665686, 5665687, 5665688, 5665689, 5665691, 5665692, 5665693, 5665694, 5665695, 5665697, 5665698, 5665699, 5665700, 5665701, 5665702, 5665703, 5665704, 5665706, 5665707, 5665708, 5665709, 5677819, 5677821, 5677822, 5677823, 5677824, 5677825, 5677826, 5677827, 5677828, 5677829, 5677831, 5677832, 5677833, 5677834, 5677835, 5677836, 5677837, 5768997, 5768998, 5768999, 5769000, 5769001, 5769002, 5769003, 5769004, 5769005, 5769006, 5769007, 5769008, 5769009, 5769010, 5769012, 5769013, 5769014, 5769015, 5769016, 5769017, 5769018, 5769019, 5769021, 5769023, 5769024, 5769025, 5769026, 5862702, 5862703, 5862704, 5862705, 5862706, 5862707, 5862708, 5862709, 5862710, 5862711, 5862712, 5862713, 5862714, 5862715, 5862716, 5862717, 5862718, 5862719, 5862720, 5862721, 5862722, 5862723, 5862724, 5862725, 5862726, 5862727, 5862728, 5862729, 5862730, 5862821, 5862822, 5862823, 5862825, 5862826, 5862827, 5862828, 5862829, 5862830, 5862831, 5862832, 5862833, 5862834, 5862835, 5862836, 5862837, 5862838, 5862839, 5862840, 5862841, 5862842, 5862843, 5862844, 5862845, 5862847, 5862848, 5862849, 5862850, 5877066, 5877067, 5877068, 5877069, 5877070, 5877071, 5877073, 5968668, 5968669, 5968670, 5968671, 5968672, 5968673, 5968674, 5968675, 5968676, 5968677, 5968678, 5968679, 5968680, 5968681, 5968682, 5968684, 5968685, 5968686, 5968687, 5968689, 5968690, 5968691, 5968692, 5978393, 5978394, 5978395, 5978396, 5978397, 5978398, 5978399, 5978401, 5978402, 5978403, 5978404, 5978405, 5978406, 5978407, 5978408, 5978409, 5978410, 5978411, 5978412, 5978413, 5978415, 5978417, 5978418, 5978419, 5978420, 5978421, 5978422, 6073039, 6073041, 6073042, 6073043, 6073044, 6073045, 6073046, 6073048, 6073049, 6073050, 6073051, 6073052, 6073053, 6073054, 6073458, 6073459, 6073460, 6073461, 6073462, 6073463, 6073464, 6073465, 6073466, 6073467, 6073468, 6073469, 6073470, 6073471, 6073472, 6073473, 6148568, 6148569, 6148570, 6148571, 6148572, 6148573, 6148574, 6148575, 6148577, 6148578, 6148579, 6148580, 6148581, 6148582, 6148583, 6148584, 6148585, 6148586, 6148587, 6148588, 6148590, 6148591, 6148592, 6148594, 6148595, 6149290, 6149291, 6149292, 6149293, 6149294, 6149295, 6149298, 6149299, 6149300, 6149301, 6149302, 6149303, 6149304, 6149305, 6149306, 6149307, 6149308, 6149309, 6149310, 6149311, 6149312, 6149313, 6149314, 6149315, 6155985, 6155986, 6155989, 6155990, 6155991, 6155992, 6155993, 6155994, 6155995, 6155996, 6155999, 6156000, 6156001, 6156002, 6156004, 6156005, 6156006, 6156007, 6156009, 6156010, 6503852, 6503853, 6503854, 6503856, 6503859, 6503860, 6503861, 6503862, 6503863, 6503864, 6503866, 6503867, 6503868, 6503869, 6503870, 6503871, 6503872, 6503873, 6503874, 6503875, 6503974, 6503975, 6503978, 6503979, 6503980, 6503981, 6503982, 6503983, 6503985, 6503986, 6503988, 6503989, 6503990, 6503991, 6503992, 6503993, 6503994, 6503996, 6503997, 6503998, 6600773, 6600774, 6600775, 6600776, 6600777, 6600778, 6600779, 6600780, 6600782, 6600784, 6600785, 6600786, 6600788, 6600789, 6600790, 6600791, 6600792, 6600794, 6600795, 6600796, 6600797, 6600798, 6600802, 6600803, 6600804, 6600805, 6600807, 6600808, 6600810, 6600811, 6600812, 6600813, 6600814, 6600815, 6600816, 6600818, 6600819, 6925578, 6925579, 6925581, 6925582, 6925583, 6925586, 6925587, 6925588, 6925590, 6925591, 6925592, 6925593, 6925594, 6925595, 6925596, 6925598, 6925599, 6925600, 6925601, 6925602, 6925603, 6925604, 7105452, 7105461, 7105467, 7105470, 7105474, 7105476;
c. REF 130-40-53, GUDID 10885862207098, Serial Numbers: 2408521, 2417851, 2417858, 2417861, 2417864, 2429148, 2429149, 2429150, 2429152, 2429153, 2429154, 2429155, 2429156, 2429158, 2429159, 2429161, 2429162, 2429163, 2429164, 2429167, 2429169, 2429171, 2429172, 2441099, 2441100, 2441101, 2441102, 2441103, 2441104, 2441105, 2441106, 2441107, 2441108, 2441109, 2441110, 2441111, 2441112, 2441114, 2441115, 2441116, 2441117, 2581984, 2581985, 2581987, 2581988, 2581990, 2581991, 2581992, 2581993, 2581994, 2581995, 2581996, 2581997, 2581998, 2581999, 2582000, 2582001, 2582002, 2582003, 2582004, 2582006, 2582007, 2582008, 2582009, 2582010, 2582011, 2582012, 2582013, 2582014, 2582015, 2582018, 2582019, 2655058, 2655059, 2655060, 2655061, 2655062, 2655063, 2655064, 2655066, 2655068, 2655069, 2655071, 2655072, 2655073, 2655074, 2659315, 2659316, 2659317, 2659318, 2659319, 2659320, 2659321, 2659322, 2659323, 2659325, 2659326, 2659327, 2659329, 2659330, 2659331, 2659333, 2659334, 2659335, 2659336, 2659337, 2659338, 2659339, 2659340, 2659341, 2659342, 2659343, 2659344, 2659345, 2659346, 2659347, 2667230, 2667231, 2667232, 2667234, 2667237, 2667239, 2667241, 2667242, 2667243, 2667244, 2667245, 2667246, 2667247, 2667249, 2667250, 2667251, 2667252, 2667253, 2667254, 2667255, 2667256, 2667258, 2667259, 2667260, 2667261, 2667262, 2667263, 2667265, 2667266, 2667267, 2667268, 2667269, 2667270, 2667273, 2667274, 2667275, 2667276, 2667279, 3680171, 3680172, 3680174, 3680175, 3680176, 3680177, 3680178, 3680179, 3680180, 3680181, 3680182, 3680183, 3680184, 3680185, 3680186, 3680187, 3680188, 3680189, 3680190, 3680191, 3680192, 3680193, 3757255, 3757256, 3757257, 3757258, 3757259, 3757260, 3757261, 3757262, 3757263, 3757264, 3757265, 3757266, 3757267, 3757268, 3757270, 3757271, 3757272, 3757273, 3757275, 3757276, 3757278, 3757279, 3757280, 3757281, 3757282, 3757283, 3860766, 3860767, 3860768, 3860769, 3860770, 3860772, 3860773, 3860774, 3860775, 3860776, 3860777, 3860778, 3860779, 3860782, 3860783, 3860784, 3860785, 3860787, 3860788, 3860789, 3860791, 3860792, 3860793, 3860794, 3860795, 3860797, 3860798, 3860799, 3860800, 3860801, 3860802, 3860803, 3860804, 3860805, 3860806, 3860807, 3860808, 3860809, 3860810, 3860811, 3860812, 3860813, 3860814, 3860815, 3860816, 3860817, 3860818, 3860820, 3860822, 3860823, 3860824, 4088305, 4088306, 4088307, 4088308, 4088309, 4088310, 4088311, 4088313, 4088314, 4088315, 4088316, 4154828, 4154829, 4154830, 4154831, 4154832, 4154833, 4154835, 4154836, 4154837, 4154838, 4154839, 4154840, 4154841, 4154842, 4155030, 4155031, 4155032, 4155033, 4155034, 4155036, 4155038, 4155039, 4155040, 4155041, 4155042, 4155043, 4155045, 4155046, 4155047, 4155048, 4155049, 4155050, 4155051, 4155052, 4155053, 4155054, 4155055, 4155056, 4155057, 4155058, 4155059, 4155061, 4155062, 4155063, 4155064, 4155065, 4155066, 4155067, 4155068, 4155069, 4191245, 4191246, 4191247, 4191248, 4191250, 4191251, 4191252, 4191253, 4191254, 4191255, 4191257, 4191258, 4191259, 4191260, 4191261, 4191262, 4191263, 4191264, 4191265, 4191266, 4191267, 4191268, 4191269, 4191270, 4191271, 4191272, 4191273, 4191274, 4191275, 4191276, 4191277, 4191278, 4216929, 4216930, 4216931, 4216932, 4216933, 4216934, 4216935, 4216936, 4216937, 4216938, 4369496, 4369497, 4369498, 4369500, 4369501, 4369502, 4369503, 4369504, 4369505, 4369508, 4369509, 4369511, 4369512, 4369513, 4369516, 4369519, 4369520, 4369521, 4369522, 4369523, 4369524, 4369525, 4369526, 4369527, 4369528, 4369529, 4369530, 4369531, 4369532, 4369533, 4369534, 4369535, 4369536, 4369537, 4369538, 4369539, 4369540, 4369541, 4369542, 4369544, 4369545, 4369546, 4369547, 4369548, 4369549, 4369551, 4369552, 4369553, 4715707, 4715708, 4715709, 4715712, 4715717, 4715721, 4715722, 4715723, 4715724, 4715725, 4715727, 4715728, 4715729, 4715730, 4715731, 4836629, 4836630, 4836631, 4836632, 4836633, 4836634, 4836635, 4836636, 4836637, 4836638, 4836639, 4836640, 4836642, 4836643, 4836644, 4836645, 4836646, 4836647, 4836649, 4836650, 4836652, 4836653, 4930561, 4930562, 4930563, 4930564, 4930565, 4930567, 4930568, 4930569, 4930571, 4930572, 4930573, 4930574, 4930575, 4930576, 4930577, 4930578, 4930579, 4930580, 4930581, 4930582, 4930584, 4930585, 5010848, 5010849, 5010850, 5010851, 5010852, 5010853, 5010854, 5010855, 5010857, 5010859, 5010861, 5010862, 5010864, 5010867, 5010868, 5010869, 5010871, 5091169, 5091170, 5091171, 5091174, 5091182, 5091183, 5091186, 5091187, 5091189, 5139783, 5139784, 5139785, 5139786, 5139787, 5139788, 5139789, 5139790, 5169123, 5169124, 5169125, 5169127, 5169128, 5169129, 5169130, 5169131, 5169133, 5169134, 5169136, 5169137, 5169138, 5169140, 5169141, 5169142, 5169143, 5169144, 5169145, 5169146, 5169147, 5169148, 5245420, 5245421, 5245422, 5245423, 5245424, 5245425, 5245426, 5245428, 5245429, 5245430, 5245431, 5245432, 5245434, 5245435, 5245436, 5245437, 5245438, 5245439, 5245440, 5245441, 5245443, 5245444, 5403288, 5403289, 5403290, 5403291, 5403292, 5403293, 5403294, 5403295, 5403296, 5403297, 5403298, 5403299, 5403300, 5403302, 5403303, 5403305, 5403307, 5403308, 5403309, 5403310, 5403311, 5507277, 5507278, 5507279, 5507280, 5507281, 5507282, 5507283, 5507284, 5507285, 5507286, 5507287, 5507288, 5507289, 5507290, 5507292, 5597469, 5597470, 5597472, 5597474, 5597475, 5597476, 5597477, 5597478, 5597480, 5597481, 5597483, 5804571, 5804572, 5804574, 5804575, 5804576, 5804578, 5804579, 5804580, 5804581, 5804583, 5804584, 5804585, 5804586, 5804587, 5804588, 5804589, 5804590, 5804591, 5804593, 5804594, 5804595, 5875514, 5875515, 5875517, 5875518, 5875521, 5875522, 5875523, 5875524, 5875526, 5875529, 5875530, 5875532, 5875534, 5875535, 5875536, 5875537, 5875538, 6279350, 6279353, 6279356, 6279362, 6279369, 6279370;
d. REF 130-40-54, GUDID 10885862207104, Serial Numbers: 2417814, 2417816, 2417817, 2417818, 2417819, 2417820, 2417822, 2417824, 2417829, 2417830, 2417831, 2417833, 2417836, 2417838, 2417839, 2417844, 2417845, 2417847, 2417849, 2433346, 2433348, 2433349, 2433350, 2433353, 2433355, 2433356, 2433357, 2441078, 2441082, 2441084, 2441086, 2441090, 2441091, 2441092, 2441094, 2441096, 2585432, 2585433, 2585439, 2585441, 2585442, 2585443, 2585444, 2585447, 2585448, 2585450, 2585451, 2585455, 2585456, 2585457, 2585460, 2585462, 2585464, 2585465, 2585467, 2585468, 2655010, 2655014, 2655016, 2655017, 2655019, 2655021, 2655022, 2655024, 2655025, 2655031, 2655032, 2655037, 2655041, 2655046, 2655050, 2655051, 2655052, 2655053, 2655054, 2655055, 2655056, 3808181, 3808184, 3808185, 3808188, 3808189, 3808190, 3808191, 3808192, 3808193, 3808194, 3808196, 3808197, 3808198, 4091457, 4091459, 4091460, 4091461, 4091462, 4091463, 4091464, 4091465, 4091467, 4091469, 4091470, 4091473, 4091474, 4154919, 4154921, 4154925, 4154927, 4154928, 4154929, 4154931, 4154933, 4154934, 4154936, 4154938, 4191513, 4191514, 4191516, 4191520, 4191521, 4191523, 4191524, 4191526, 4191527, 4191528, 4191529, 4191531, 4191533, 4191534, 4191536, 4191538, 4191541, 4191545, 4191547, 4191552, 4191553, 4191554, 4383642, 4383643, 4383648, 4383649, 4383650, 4383651, 4383653, 4383657, 4383660, 4383661, 4383669, 4938386, 4938387, 4938388, 4938389, 4938390, 4938391, 4938392, 4938393, 4938395, 4972819, 4972820, 4972822, 4972824, 4972827, 4972828, 5005459, 5005461, 5005463, 5005464, 5034068, 5034069, 5034072, 5034074, 5034075, 5139182, 5139184, 5139185, 5139187, 5139190, 5168271, 5168273, 5168274, 5213802, 5213807, 5213808, 5284073, 5284075, 5284077, 5284079, 5486862, 5486863, 5486865, 5555251, 5555252, 5555253, 5673817, 5673820, 5673821, 5673823, 5673824, 5673826;
e. REF 130-40-55, GUDID 10885862207111, Serial Numbers: 2429144, 2429146, 2443025, 2531063, 2531067, 2531075, 2531081, 2588908, 2588916, 2588924, 2598644, 2598646, 2598647, 2598654, 2598655, 2659474, 2659479, 2659485, 3707201, 3707207, 3707209, 3707214, 4191641, 4191647, 4191649, 4191652, 4191676, 4415093, 4415102, 4415107, 4415109, 4415110, 4415117, 5213790, 5266946, 5266953, 5292866;
f. REF 132-32-51, GUDID 10885862207029, Serial Numbers: 2429133, 2429134, 2429135, 2429136, 2429137, 2429138, 2429139, 2429140, 2429141, 2429142, 2429143, 2443042, 2443043, 2443044, 2443047, 2443048, 2443049, 2443050, 2443051, 2443052, 2443054, 2443055, 2443056, 2443057, 2443058, 2443059, 2443060, 2443061, 2531155, 2531156, 2531157, 2531158, 2531159, 2531160, 2531161, 2531162, 2531163, 2531164, 2531165, 2531166, 2531167, 2531168, 2531170, 2531171, 2531172, 2531173, 2531174, 2531175, 2531176, 2531177, 2531178, 2531179, 2531180, 2531181, 2531182, 2531183, 2531184, 2531185, 2531186, 2531188, 2531189, 2531190, 2531191, 2531193, 2531194, 2531195, 2531196, 2609933, 2609934, 2609935, 2609936, 2609937, 2609939, 2609940, 2609941, 2609942, 2609943, 2609944, 2609946, 2609947, 2609948, 2609949, 2609950, 2609951, 2609952, 2609953, 2609954, 2609955, 2609956, 2609957, 2609958, 2609959, 2609960, 2609961, 2609962, 2609963, 2609964, 2609966, 2609967, 2609968, 2667280, 2667281, 2667282, 2667283, 2667284, 2667285, 2667286, 2667287, 2667288, 2667289, 2667290, 2667291, 2667292, 2667293, 2667294, 2667295, 2667296, 2667297, 2667298, 2667299, 2667300, 2667301, 2667302, 2667303, 2667304, 2667305, 2667306, 2667307, 2667308, 2667309, 2667310, 2667312, 2667313, 2667314, 2667315, 2667316, 2667317, 2667318, 2667319, 2667320, 2667322, 2667323, 2667324, 2667325, 2667326, 2667327, 2667328, 2667329, 2841246, 2841247, 2841248, 2841250, 2841251, 2841252, 2841253, 2841254, 2841255, 2841256, 2841257, 2841258, 2841259, 2841260, 2841261, 2841262, 2841263, 2841265, 2841266, 2891970, 2891971, 2891972, 2891973, 2891974, 2891975, 2891976, 2891977, 2891978, 2891979, 2891982, 2891983, 2891984, 2891985, 2891986, 2891987, 2891988, 2891989, 2891990, 2891991, 2957078, 2957079, 2957081, 2957082, 2957083, 2957084, 2957085, 2957087, 2957088, 2957089, 2957090, 2957091, 2957092, 2957093, 2957094, 2957095, 2957096, 2957098, 2957099, 2957100, 2957102, 2957103, 2957104, 2957105, 2957106, 2957107, 2957108, 2957109, 2957110, 2957111, 2957113, 2957116, 2957117, 2978511, 2978512, 2978513, 2978514, 2978515, 2978516, 2978517, 2978518, 2978519, 2978520, 2978521, 2978523, 2978525, 2978526, 2978527, 2978528, 2978530, 2978532, 2978533, 2978534, 3623993, 3623994, 3623995, 3623996, 3623997, 3623999, 3624000, 3624001, 3624002, 3624005, 3624006, 3624007, 3624008, 3624009, 3624011, 3624012, 3624013, 3624014, 3624015, 3673483, 3673484, 3673485, 3673486, 3673488, 3673489, 3673490, 3673492, 3673493, 3673494, 3673495, 3673496, 3673498, 3673499, 3673500, 3673501, 3673502, 3673503, 3673504, 3673505, 3673506, 3673507, 3673508, 3796420, 3796421, 3796422, 3796424, 3796425, 3796426, 3796427, 3796428, 3796429, 3796430, 3796431, 3796432, 3796433, 3796434, 3796435, 3796437, 3796438, 3796439, 3909580, 3909581, 3909582, 3909584, 3909585, 3909586, 3909587, 3909589, 3909590, 3909591, 3909592, 4086628, 4086629, 4086630, 4086631, 4086632, 4086633, 4086634, 4086635, 4086636, 4086638, 4086639, 4086640, 4086641, 4086642, 4086644, 4086645, 4086646, 4086648, 4086649, 4086650, 4086651, 4086652, 4086653, 4145054, 4145055, 4145056, 4145057, 4145058, 4145059, 4145060, 4145061, 4145062, 4145063, 4145064, 4145065, 4145066, 4145067, 4145069, 4145070, 4145071, 4145072, 4145073, 4145074, 4145075, 4145076, 4145077, 4145078, 4145079, 4145080, 4145082, 4145083, 4145084, 4145085, 4145086, 4145087, 4145088, 4145089, 4145090, 4145091, 4145092, 4145093, 4145094, 4145095, 4145096, 4235196, 4235197, 4235198, 4235199, 4235200, 4235201, 4235202, 4235203, 4235204, 4235205, 4235206, 4235207, 4235208, 4235209, 4235210, 4235211, 4235212, 4235213, 4235214, 4235215, 4235216, 4235217, 4235218, 4235219, 4235220, 4242725, 4242726, 4242727, 4242728, 4242729, 4242730, 4242731, 4242732, 4242733, 4242734, 4242735, 4242736, 4242737, 4242738, 4242739, 4242740, 4242741, 4242742, 4242743, 4242744, 4324733, 4324734, 4324735, 4324736, 4324737, 4324738, 4324739, 4324740, 4324741, 4324742, 4324743, 4324745, 4324746, 4324747, 4324748, 4324749, 4324750, 4324751, 4324752, 4324753, 4324754, 4324755, 4324756, 4324757, 4324758, 4324759, 4324760, 4324761, 4324762, 4349331, 4349332, 4349333, 4349334, 4349335, 4349336, 4349337, 4349338, 4349339, 4349340, 4349341, 4349342, 4349343, 4349344, 4349345, 4349346, 4349347, 4349348, 4349349, 4349350, 4349352, 4349353, 4349354, 4349355, 4349356, 4349357, 4349358, 4349359, 4349360, 4349361, 4349362, 4349363, 4349364, 4349365, 4349366, 4349367, 4349368, 4349371, 4349372, 4349373, 4349374, 4349375, 4349377, 4349379, 4349380, 4349381, 4349382, 4398371, 4398372, 4398373, 4398374, 4398375, 4398376, 4398377, 4398378, 4398379, 4398380, 4398381, 4398383, 4398384, 4398385, 4398386, 4398387, 4398388, 4398389, 4398390, 4398391, 4398392, 4398393, 4398394, 4398395, 4398396, 4398397, 4398398, 4398399, 4398400, 4453956, 4453958, 4453959, 4453960, 4453961, 4453962, 4453963, 4453964, 4453965, 4453966, 4453967, 4453968, 4453969, 4453971, 4453972, 4453973, 4453974, 4453975, 4453976, 4453977, 4453978, 4453979, 4453980, 4453981, 4453982, 4455723, 4455724, 4455725, 4455726, 4455727, 4455728, 4455729, 4455730, 4455731, 4455732, 4455733, 4455734, 4455735, 4455736, 4455737, 4455738, 4455739, 4455740, 4455741, 4455742, 4455744, 4455745, 4455746, 4455747, 4455748, 4462961, 4462962, 4462963, 4462964, 4462965, 4462967, 4462968, 4462969, 4462970, 4462971, 4462972, 4462973, 4462974, 4462975, 4462976, 4462977, 4462978, 4462979, 4462980, 4462981, 4462982, 4462983, 4462984, 4462985, 4462986, 4462987, 4462988, 4462989, 4462990, 4483599, 4483600, 4483601, 4483602, 4483603, 4483604, 4483605, 4483606, 4483607, 4483608, 4483609, 4483610, 4483611, 4483612, 4483613, 4483614, 4483615, 4483616, 4483617, 4483618, 4483619, 4483620, 4483621, 4483622, 4483623, 4483624, 4483625, 4483626, 4483627, 4483628, 4535081, 4535082, 4535083, 4535084, 4535085, 4535086, 4535087, 4535088, 4535089, 4535090, 4535091, 4535092, 4535093, 4535094, 4535095, 4535096, 4535097, 4535098, 4535099, 4535100, 4535101, 4535102, 4535103, 4535104, 4535105, 4535106, 4535107, 4535108, 4535109, 4535110, 4588848, 4588849, 4588850, 4588851, 4588852, 4588853, 4588854, 4588855, 4588856, 4588857, 4588858, 4588859, 4588860, 4588861, 4588862, 4588863, 4588864, 4588865, 4588866, 4588867, 4588868, 4588869, 4588870, 4588871, 4588872, 4588873, 4588874, 4588875, 4588876, 4588877, 4615405, 4615406, 4615407, 4615408, 4615409, 4615410, 4615411, 4615412, 4615413, 4615414, 4615415, 4615416, 4615417, 4615418, 4615419, 4615420, 4615421, 4615422, 4615423, 4615424, 4615425, 4615426, 4615427, 4615428, 4615429, 4615430, 4615431, 4615432, 4615433, 4615434, 4638456, 4638457, 4638458, 4638459, 4638460, 4638461, 4638462, 4638463, 4638464, 4638465, 4638466, 4638467, 4638468, 4638469, 4638470, 4638471, 4638472, 4638473, 4638474, 4638475, 4638476, 4638477, 4638478, 4638479, 4638480, 4638481, 4638482, 4638483, 4638484, 4638485, 4683165, 4683166, 4683167, 4683168, 4683169, 4683170, 4683171, 4683172, 4683173, 4683174, 4683175, 4683176, 4683177, 4683178, 4683179, 4683180, 4683181, 4683182, 4683183, 4683184, 4683185, 4683186, 4683187, 4683188, 4683189, 4683190, 4703965, 4703966, 4703967, 4703968, 4703969, 4703970, 4703971, 4703972, 4703973, 4703974, 4703975, 4703976, 4703977, 4703978, 4703979, 4703980, 4703982, 4703983, 4703984, 4703985, 4703986, 4703987, 4703988, 4703989, 4703990, 4703991, 4722558, 4722559, 4722560, 4722561, 4722562, 4722563, 4722564, 4722565, 4722566, 4722567, 4722568, 4722569, 4722570, 4722571, 4722572, 4722573, 4722574, 4722575, 4722576, 4722577, 4722578, 4722579, 4722580, 4722581, 4722582, 4722583, 4722584, 4722585, 4722586, 4722587, 4722660, 4722661, 4738003, 4738004, 4738005, 4738006, 4738007, 4738008, 4738010, 4738011, 4738012, 4738014, 4738015, 4738016, 4738017, 4738018, 4738019, 4738020, 4738021, 4738022, 4738023, 4738024, 4738025, 4738026, 4738027, 4738028, 4738029, 4738030, 4738031, 4738032, 4741233, 4741234, 4741235, 4741236, 4741237, 4741238, 4741239, 4743433, 4743434, 4743436, 4743437, 4743438, 4743440, 4743441, 4743442, 4743443, 4743444, 4743445, 4743446, 4743447, 4743449, 4743450, 4743451, 4743452, 4743453, 4743454, 4743455, 4743457, 4743458, 4743459, 4743460, 4743461, 4743462, 4754625, 4754626, 4754627, 4754628, 4754629, 4754630, 4770123, 4770124, 4770125, 4770126, 4770127, 4770128, 4770129, 4770130, 4770131, 4770132, 4770134, 4770135, 4770136, 4770137, 4770138, 4770139, 4770140, 4770141, 4770142, 4770143, 4770144, 4770145, 4770146, 4770147, 4770148, 4770149, 4770150, 4770151, 4770152, 4780956, 4780957, 4780958, 4780959, 4780960, 4780961, 4780962, 4780963, 4780964, 4780965, 4780966, 4780967, 4780968, 4780969, 4780970, 4780971, 4780972, 4780973, 4780974, 4780975, 4780976, 4780977, 4780978, 4780979, 4780980, 4780981, 4780982, 4780983, 4780984, 4780985, 4787941, 4787942, 4787943, 4787944, 4787945, 4787946, 4787947, 4787948, 4787949, 4787950, 4787951, 4787952, 4787953, 4787954, 4787955, 4787957, 4787958, 4787959, 4787960, 4787961, 4787962, 4787963, 4787964, 4787965, 4787966, 4787967, 4787968, 4787969, 4787970, 4816956, 4816957, 4816958, 4816959, 4816960, 4816961, 4816962, 4816963, 4816964, 4816965, 4816966, 4816967, 4816968, 4816969, 4816970, 4816971, 4816972, 4816973, 4816974, 4816975, 4816976, 4816978, 4816979, 4816980, 4816981, 4816982, 4816983, 4816984, 4816985, 4817902, 4817903, 4817904, 4817905, 4817906, 4817907, 4817908, 4817909, 4817910, 4817911, 4817912, 4817913, 4817914, 4817915, 4817916, 4817917, 4817918, 4817919, 4817920, 4817921, 4817922, 4817923, 4817924, 4817925, 4817926, 4817927, 4817928, 4817929, 4817930, 4817931, 4836599, 4836600, 4836601, 4836602, 4836603, 4836604, 4836605, 4836606, 4836607, 4836608, 4836609, 4836610, 4836611, 4836612, 4836613, 4836614, 4836615, 4836616, 4836617, 4836618, 4836619, 4836620, 4836621, 4836622, 4836623, 4836624, 4836625, 4836626, 4836627, 4836628, 4845448, 4845449, 4845450, 4845451, 4845452, 4845453, 4845454, 4845455, 4845456, 4845457, 4845458, 4845459, 4845460, 4845461, 4845462, 4845463, 4845464, 4845465, 4845466, 4845467, 4845468, 4845469, 4845470, 4845471, 4845472, 4845473, 4845474, 4845475, 4845476, 4845477, 4854312, 4854313, 4854314, 4861407, 4861408, 4861409, 4861410, 4861411, 4861412, 4861413, 4861414, 4861415, 4861416, 4861417, 4861418, 4861419, 4861420, 4861421, 4861422, 4861423, 4861424, 4861425, 4861426, 4861427, 4861428, 4861429, 4861430, 4861431, 4861432, 4861433, 4861434, 4861435, 4861436, 4878896, 4878897, 4878898, 4878899, 4878900, 4878901, 4878902, 4878903, 4878904, 4878905, 4878906, 4878907, 4878908, 4878909, 4878910, 4878911, 4878912, 4878913, 4878914, 4878915, 4878916, 4878917, 4878918, 4878919, 4878920, 4878921, 4878922, 4878923, 4878924, 4878925, 4878926, 4909073, 4909074, 4909075, 4909076, 4909077, 4909078, 4909079, 4909080, 4909081, 4909082, 4909083, 4909084, 4909085, 4909086, 4909087, 4909088, 4909089, 4909090, 4909091, 4909092, 4909093, 4909094, 4909095, 4909096, 4909097, 4909098, 4909099, 4909100, 4909101, 4909102, 4941978, 4941979, 4941980, 4941981, 4941982, 4941983, 4941984, 4941985, 4941986, 4941987, 4941988, 4941989, 4941990, 4941991, 4941992, 4941993, 4941994, 4941995, 4941996, 4941997, 4941998, 4941999, 4942000, 4942001, 4942002, 4942003, 4942004, 4942005, 4942006, 4942007, 4957743, 4957744, 4957745, 4957746, 4957747, 4957748, 4957749, 4957750, 4957751, 4957752, 4957753, 4957754, 4957755, 4957756, 4957757, 4957758, 4957759, 4957760, 4957761, 4957762, 4957763, 4957764, 4957765, 4957766, 4957767, 4957768, 4957769, 4957770, 4957771, 4957772, 4978371, 4978372, 4978373, 4978374, 4978375, 4978376, 4978377, 4978378, 4978379, 4978380, 4978381, 4978382, 4978383, 4978384, 4978385, 4978386, 4978387, 4978388, 4978389, 4978390, 4978391, 4978392, 4978393, 4978394, 4978395, 4978396, 4978397, 4978398, 4978399, 4978400, 5013284, 5013285, 5013286, 5013287, 5013288, 5013289, 5013290, 5013291, 5013292, 5013293, 5013294, 5013295, 5013296, 5013297, 5013298, 5013299, 5013300, 5013301, 5013302, 5013303, 5013304, 5013305, 5013306, 5013307, 5013308, 5013309, 5013310, 5013311, 5013312, 5013313, 5035638, 5035639, 5035640, 5035641, 5035642, 5035643, 5035644, 5035645, 5035646, 5035647, 5035648, 5035649, 5035650, 5035651, 5035652, 5035653, 5035654, 5035655, 5035656, 5035657, 5035658, 5035659, 5035660, 5035661, 5035662, 5035663, 5035664, 5035665, 5035666, 5035667, 5061801, 5061802, 5061803, 5061804, 5061805, 5061806, 5061807, 5061808, 5061809, 5061810, 5061811, 5061812, 5061813, 5061814, 5061815, 5061816, 5061817, 5061818, 5061819, 5061820, 5061821, 5061822, 5061823, 5061824, 5061825, 5061826, 5061827, 5061828, 5061829, 5061830, 5092378, 5092379, 5092380, 5092381, 5092382, 5092383, 5092384, 5092385, 5092386, 5092387, 5092388, 5092389, 5092390, 5092391, 5092392, 5092394, 5092395, 5092396, 5092397, 5092398, 5092399, 5092400, 5092401, 5092402, 5092403, 5092404, 5092405, 5092406, 5092407, 5103148, 5103149, 5103150, 5103151, 5103152, 5103153, 5103154, 5103155, 5103156, 5103157, 5103158, 5103159, 5103160, 5103161, 5103162, 5103163, 5103164, 5103165, 5103166, 5103167, 5103168, 5103169, 5103170, 5103171, 5103172, 5103173, 5103174, 5103175, 5103176, 5103177, 5115706, 5115707, 5115708, 5115709, 5115710, 5115711, 5115712, 5115713, 5115714, 5115715, 5115716, 5115717, 5115718, 5130617, 5130618, 5130619, 5130620, 5130621, 5130622, 5130623, 5130624, 5130625, 5130626, 5130627, 5130628, 5130629, 5130630, 5130631, 5130632, 5130633, 5130634, 5130635, 5130636, 5130637, 5130638, 5130639, 5130640, 5130641, 5130642, 5130643, 5130644, 5130645, 5130646, 5140760, 5140761, 5140762, 5140763, 5140764, 5140765, 5140766, 5140767, 5140768, 5140769, 5140770, 5140771, 5140772, 5140773, 5140774, 5140775, 5140776, 5140777, 5140778, 5140779, 5140780, 5140781, 5140782, 5140783, 5140784, 5140785, 5140786, 5140788, 5140789, 5140790, 5140791, 5140792, 5140793, 5140794, 5140795, 5140796, 5140797, 5140798, 5140799, 5140800, 5140801, 5140802, 5140803, 5140804, 5140805, 5140806, 5140807, 5140808, 5140809, 5140810, 5140811, 5140812, 5140813, 5140814, 5140815, 5140816, 5140817, 5140818, 5140819, 5157721, 5157722, 5157723, 5157724, 5157725, 5157726, 5157727, 5157728, 5187211, 5187212, 5187213, 5187214, 5187215, 5187216, 5187217, 5187218, 5187219, 5187220, 5187221, 5187222, 5187223, 5187224, 5187225, 5187226, 5187227, 5187228, 5187230, 5187231, 5187232, 5187233, 5187234, 5187235, 5187236, 5187237, 5187238, 5187239, 5187240, 5199503, 5199504, 5199505, 5199506, 5199507, 5199508, 5199509, 5199510, 5199511, 5199512, 5199513, 5199514, 5199515, 5199516, 5199517, 5199518, 5199519, 5199520, 5199521, 5199522, 5199523, 5214854, 5214855, 5214856, 5214857, 5214858, 5214859, 5214860, 5214861, 5214862, 5214864, 5214865, 5214866, 5214867, 5214868, 5214869, 5214870, 5214871, 5214872, 5214873, 5214874, 5214875, 5214876, 5214877, 5214878, 5214879, 5214880, 5214881, 5214882, 5214883, 5251174, 5251175, 5251176, 5251177, 5251178, 5251179, 5251181, 5251182, 5251183, 5251184, 5251185, 5251186, 5251187, 5251189, 5251190, 5251191, 5251192, 5251193, 5251194, 5251195, 5251196, 5251197, 5251198, 5251199, 5251200, 5251201, 5251202, 5251203, 5258782, 5258783, 5258784, 5258785, 5258786, 5258787, 5258788, 5258789, 5258790, 5258791, 5258792, 5258793, 5258794, 5258795, 5258796, 5258797, 5258798, 5258799, 5258800, 5258801, 5258802, 5258803, 5258804, 5258805, 5258806, 5258807, 5258808, 5258809, 5258810, 5258811, 5258812, 5258813, 5258814, 5258815, 5258816, 5258817, 5258818, 5258819, 5258820, 5258821, 5258822, 5258823, 5258824, 5301691, 5301692, 5301693, 5301694, 5301695, 5301696, 5301697, 5301698, 5301699, 5301700, 5301701, 5301702, 5301703, 5301704, 5301705, 5301706, 5301707, 5301708, 5301709, 5301710, 5301711, 5301712, 5301713, 5301714, 5301715, 5301716, 5301717, 5301718, 5301719, 5301720, 5331755, 5331756, 5331757, 5331758, 5331759, 5331760, 5331761, 5331763, 5331764, 5331765, 5331766, 5331767, 5331768, 5331770, 5331771, 5331772, 5331773, 5331774, 5331775, 5331776, 5331777, 5331778, 5331779, 5331780, 5331781, 5331782, 5331783, 5347803, 5347804, 5347805, 5347806, 5347807, 5347808, 5347809, 5347810, 5347812, 5347813, 5347814, 5347815, 5347816, 5347817, 5347818, 5347819, 5347820, 5347821, 5347822, 5347823, 5347825, 5347826, 5347827, 5347828, 5347829, 5347830, 5347831, 5347832, 5380122, 5380123, 5380124, 5380125, 5380126, 5380127, 5380128, 5380130, 5380131, 5380132, 5380133, 5380134, 5380135, 5380136, 5380137, 5380138, 5380139, 5380140, 5380141, 5380142, 5380143, 5380144, 5380145, 5380146, 5380147, 5380148, 5380149, 5380150, 5380151, 5380152, 5380153, 5380154, 5380155, 5380156, 5380157, 5380158, 5380159, 5380160, 5380161, 5404764, 5404765, 5404766, 5404768, 5404770, 5404771, 5404772, 5404773, 5404774, 5404775, 5404777, 5404778, 5404779, 5404780, 5404781, 5404782, 5404783, 5404784, 5404786, 5404787, 5404788, 5404789, 5404790, 5404791, 5404792, 5404793, 5404794, 5404795, 5425483, 5425484, 5425485, 5425486, 5425487, 5425488, 5425489, 5425490, 5425491, 5425492, 5425493, 5425494, 5425495, 5425496, 5425497, 5425498, 5425499, 5425500, 5425501, 5425502, 5425503, 5425504, 5425505, 5425506, 5425507, 5425508, 5425509, 5425510, 5425511, 5425512, 5468971, 5468972, 5468973, 5468974, 5468975, 5468976, 5468977, 5468978, 5468979, 5468980, 5468981, 5468982, 5468983, 5468984, 5468985, 5468986, 5468987, 5468988, 5468989, 5468990, 5468991, 5468992, 5468993, 5468994, 5468995, 5468996, 5468997, 5468998, 5468999, 5469000, 5491354, 5491355, 5491357, 5491358, 5491359, 5491360, 5491361, 5491362, 5491363, 5491364, 5491365, 5491366, 5491367, 5491368, 5491369, 5491370, 5491371, 5491372, 5491373, 5491374, 5491375, 5491376, 5491378, 5491379, 5491380, 5491381, 5491382, 5491383, 5509504, 5509505, 5509506, 5509507, 5509508, 5509509, 5509510, 5509511.5509512, 5509513, 5509514, 5509515, 5509516, 5509517, 5509518, 5509519, 5509520, 5509521, 5509522, 5509523, 5509524, 5509525, 5509526, 5509527, 5509528, 5509529, 5509530, 5509531, 5509532, 5509533, 5534213, 5534214, 5534215, 5534216, 5534217, 5534218, 5534219, 5534220, 5534221, 5534222, 5534223, 5534225, 5534226, 5534227, 5534228, 5534229, 5534230, 5534231, 5534232, 5534234, 5534235, 5534237, 5534238, 5534239, 5534240, 5534241, 5534242, 5534243, 5534244, 5534245, 5534246, 5554777, 5554778, 5554780, 5554781, 5554782, 5554783, 5554784, 5554785, 5554786, 5554787, 5554788, 5554789, 5554790, 5554791, 5554792, 5554793, 5554794, 5554795, 5554796, 5554797, 5554798, 5554799, 5554801, 5554802, 5554803, 5554804, 5554805, 5554806, 5554807, 5554808, 5554809, 5577551, 5577552, 5577553, 5577554, 5577555, 5577556, 5577557, 5577558, 5577559, 5577560, 5577561, 5577562, 5577563, 5577564, 5577565, 5577566, 5577567, 5577568, 5577569, 5577570, 5577571, 5577572, 5577573, 5577574, 5577575, 5577576, 5577577, 5577578, 5577579, 5577580, 5597188, 5597189, 5597190, 5597191, 5597192, 5597193, 5597194, 5597195, 5597196, 5597197, 5597198, 5597199, 5597200, 5597201, 5597202, 5597203, 5597204, 5597205, 5597206, 5597207, 5597208, 5597209, 5597210, 5597211, 5597212, 5597213, 5597214, 5597215, 5597216, 5597217, 5665820, 5665821, 5665822, 5665823, 5665824, 5665825, 5665826, 5665827, 5665828, 5665829, 5665830, 5665831, 5665832, 5665833, 5665835, 5665836, 5665837, 5665838, 5665839, 5665840, 5665841, 5665842, 5665843, 5665844, 5665845, 5665846, 5665847, 5665849, 5684722, 5684723, 5684724, 5684725, 5684726, 5684727, 5684728, 5684729, 5684730, 5684732, 5684733, 5684735, 5684736, 5684737, 5684738, 5684739, 5684740, 5684741, 5684742, 5684744, 5684745, 5684746, 5684747, 5684748, 5684749, 5684750, 5684751, 5736039, 5736040, 5736041, 5736042, 5736043, 5736044, 5736045, 5736047, 5736048, 5736049, 5736050, 5736051, 5736052, 5736053, 5736054, 5736055, 5736056, 5736057, 5736058, 5736059, 5736060, 5736061, 5736062, 5736063, 5736064, 5736065, 5736066, 5736067, 5754455, 5754456, 5754457, 5754458, 5754459, 5754460, 5754461, 5754462, 5754463, 5754464, 5754465, 5754466, 5754467, 5754468, 5754469, 5754470, 5754471, 5754472, 5754473, 5754474, 5754475, 5754476, 5754477, 5754478, 5754479, 5754480, 5754481, 5754482, 5754483, 5754484, 5792552, 5792553, 5792554, 5792555, 5792556, 5792557, 5792558, 5792559, 5792560, 5792561, 5792562, 5792563, 5792564, 5792565, 5792566, 5792567, 5792568, 5792569, 5792570, 5792571, 5792572, 5792573, 5792574, 5792575, 5792576, 5792577, 5792578, 5792579, 5792580, 5792581, 5792582, 5792583, 5792584, 5792585, 5792586, 5792587, 5792588, 5792589, 5792590, 5792591, 5792592, 5792593, 5792594, 5792595, 5792596, 5792597, 5792598, 5792599, 5792600, 5792601, 5792602, 5792603, 5792604, 5792605, 5792606, 5792607, 5792608, 5792609, 5792610, 5792611, 5833913, 5833914, 5833915, 5833916, 5833917, 5833918, 5833919, 5833920, 5837293, 5837294, 5837295, 5837296, 5837297, 5837298, 5837299, 5837300, 5837301, 5837302, 5837303, 5837304, 5837305, 5837306, 5837307, 5837308, 5837309, 5837310, 5837311, 5837312, 5837313, 5837314, 5837315, 5837316, 5837317, 5837319, 5837320, 5837321, 5837322, 5883132, 5883133, 5883134, 5883135, 5883136, 5883137, 5883138, 5883139, 5883140, 5883141, 5883142, 5883143, 5883144, 5883145, 5883146, 5883148, 5883149, 5883150, 5883151, 5883152, 5883153, 5883154, 5883155, 5883156, 5883157, 5883158, 5883159, 5883160, 5883161, 5898608, 5898609, 5898610, 5898611, 5898612, 5898613, 5898614, 5898615, 5898616, 5898617, 5898618, 5898619, 5898620, 5898621, 5898622, 5898623, 5898624, 5898625, 5931614, 5931615, 5931616, 5931617, 5931618, 5931619, 5931620, 5931621, 5931622, 5931623, 5931624, 5931625, 5931626, 5931627, 5931628, 5931629, 5931630, 5931631, 5932019, 5932020, 5932021, 5932022, 5932023, 5932024, 5932025, 5932026, 5932027, 5932029, 5932030, 5932031, 5932032, 5932033, 5932034, 5932035, 5932037, 5932038, 5932039, 5932040, 5932041, 5932042, 5932043, 5932044, 5932045, 5932046, 5932047, 5975081, 5975082, 5975083, 5975084, 5975085, 5975086, 5975087, 5975088, 5975089, 5975090, 5975091, 5975092, 5975093, 5975094, 5975095, 5975096, 5975097, 5975098, 5975099, 5975100, 5975399, 5975400, 5975401, 5975402, 5975403, 5975404, 5975405, 5975406, 5975407, 5975408, 5975409, 5975410, 5975411, 5975414, 5975416, 5975417, 5975418, 5975419, 5975421, 5975422, 5975424, 5975425, 5975426, 5975427, 5975428, 6073311, 6073312, 6073313, 6073314, 6073315, 6073317, 6073318, 6073319, 6073320, 6073321, 6073322, 6073323, 6073324, 6073325, 6073326, 6073327, 6073328, 6073329, 6073330, 6073331, 6073332, 6073333, 6073334, 6073335, 6073336, 6073337, 6073338, 6073339, 6079327, 6079329, 6079330, 6079331, 6079332, 6079333, 6079334, 6079335, 6079336, 6079337, 6079338, 6079339, 6079340, 6079341, 6079342, 6079343, 6079344, 6079345, 6079346, 6079347, 6079348, 6079349, 6079350, 6079351, 6079352, 6079353, 6079354, 6079355, 6149132, 6149133, 6149135, 6149136, 6149137, 6149139, 6149140, 6149141, 6149142, 6149144, 6149145, 6149146, 6149147, 6149148, 6149149, 6149150, 6149152, 6149153, 6149154, 6149155, 6149156, 6149157, 6149158, 6149159, 6149160, 6149161, 6149162, 6149163, 6149164, 6149165, 6149166, 6149167, 6149168, 6149170, 6149171, 6149172, 6149173, 6149174, 6149175, 6149176, 6149179, 6149180, 6149181, 6149182, 6149183, 6149184, 6149185, 6149186, 6149187, 6149188, 6149189, 6149190, 6149191, 6149192, 6149193, 6149194, 6155965, 6155966, 6155967, 6155968, 6155969, 6155970, 6155971, 6155972, 6155973, 6155974, 6155975, 6155976, 6155977, 6155978, 6155979, 6155980, 6155981, 6155982, 6155983, 6155984, 6156030, 6156031, 6156032, 6156033, 6156034, 6156035, 6156036, 6156037, 6156038, 6156039, 6156040, 6156041, 6156042, 6156043, 6156044, 6156045, 6156046, 6156047, 6156048, 6156049, 6156050, 6157082, 6157083, 6157084, 6157085, 6157086, 6157087, 6157088, 6157089, 6157090, 6157091, 6157092, 6157093, 6157094, 6157095, 6157096, 6157097, 6157098, 6157099, 6157100, 6157101, 6236520, 6236521, 6236522, 6236523, 6236525, 6236526, 6236527, 6236528, 6236529, 6236530, 6236531, 6236532, 6236533, 6236534, 6236535, 6236536, 6236537, 6236538, 6236539, 6236540, 6236541, 6236542, 6236543, 6236544, 6236545, 6236546, 6236547, 6236549, 6236550, 6236551, 6236552, 6236553, 6236554, 6236555, 6236556, 6236557, 6251315, 6251316, 6251317, 6251318, 6251320, 6251321, 6251322, 6251323, 6251324, 6251325, 6251326, 6251327, 6251328, 6251329, 6251330, 6251331, 6251332, 6251333, 6251334, 6251335, 6251336, 6252723, 6252724, 6252725, 6252726, 6252727, 6252730, 6252731, 6252732, 6252734, 6252735, 6252736, 6252737, 6252738, 6252739, 6252742, 6252743, 6252744, 6252745, 6252746, 6252747, 6252748, 6283720, 6283722, 6283723, 6283724, 6283725, 6283726, 6283727, 6283728, 6283729, 6283730, 6283731, 6283732, 6283733, 6283734, 6283735, 6283736, 6283737, 6283738, 6283739, 6283740, 6283741, 6283742, 6283743, 6283744, 6283745, 6283747, 6283748, 6283749, 6283750, 6283751, 6283752, 6283853, 6283854, 6283855, 6283856, 6283857, 6283858, 6283859, 6283860, 6283861, 6283862, 6283863, 6283864, 6283865, 6283866, 6283867, 6283868, 6283869, 6283870, 6283871, 6283872, 6283873, 6283874, 6283875, 6283876, 6283877, 6283878, 6283879, 6283880, 6283881, 6283882, 6283883, 6283884, 6297159, 6297160, 6297162, 6297163, 6297164, 6297165, 6297166, 6297167, 6297168, 6297169, 6297170, 6297171, 6297172, 6297173, 6297174, 6334522, 6334523, 6334524, 6334525, 6334527, 6334528, 6334529, 6334530, 6334531, 6334532, 6334533, 6334534, 6334535, 6334536, 6334537, 6334538, 6334539, 6334540, 6334541, 6334542, 6334543, 6334544, 6334545, 6334546, 6334547, 6334548, 6334549, 6334550, 6334551, 6334552, 6334553, 6334554, 6334555, 6334557, 6334558, 6334559, 6334560, 6334561, 6334563, 6334564, 6334565, 6334566, 6334567, 6334568, 6334569, 6334570, 6334571, 6334572, 6334573, 6334575, 6334576, 6334577, 6334578, 6334579, 6334580, 6334581, 6334582, 6334583, 6334584, 6334585, 6422429, 6422430, 6422431, 6422432, 6422434, 6422435, 6422437, 6422438, 6422439, 6422440, 6422441, 6422443, 6422444, 6422446, 6422447, 6422448, 6422449, 6422450, 6422451, 6422452, 6422453, 6422454, 6422455, 6422456, 6422457, 6422458, 6424725, 6424726, 6424727, 6424729, 6424730, 6424731, 6424732, 6424735, 6424736, 6424737, 6424738, 6424739, 6424740, 6424741, 6424742, 6424743, 6424744, 6424746, 6424747, 6424748, 6424749, 6424750, 6424751, 6424753, 6510399, 6510400, 6510402, 6510403, 6510404, 6510405, 6510406, 6510407, 6510408, 6510409, 6510410, 6510411, 6510412, 6510413, 6510414, 6510415, 6510416, 6510417, 6510418, 6510419, 6510421, 6510423, 6510424, 6510425, 6510426, 6510427, 6510428, 6510429, 6510430, 6510431, 6510432, 6510433, 6510434, 6510435, 6510436, 6510437, 6510438, 6510439, 6510442, 6510443, 6510444, 6510447, 6510448, 6510449, 6510450, 6510451, 6510452, 6510453, 6510454, 6510455, 6510456, 6510457, 6510458, 6510459, 6510460, 6510461, 6551623, 6551624, 6551625, 6551626, 6551627, 6551628, 6551631, 6551632, 6551633, 6551634, 6551635, 6551636, 6551637, 6551638, 6551639, 6551641, 6551643, 6551644, 6551646, 6551647, 6551648, 6551651, 6551653, 6551656, 6551657, 6551658, 6551659, 6551660, 6551661, 6551664, 6551665, 6551666, 6551668, 6551669, 6551670, 6551671, 6551672, 6551674, 6551675, 6551676, 6551677, 6551678, 6551679, 6551681, 6551682, 6551683, 6551684, 6551685, 6551686, 6551687, 6551688, 6563177, 6563178, 6563179, 6563180, 6563181, 6563182, 6563183, 6563184, 6563185, 6563186, 6563188, 6563189, 6563190, 6563191, 6563192, 6563193, 6563194, 6563195, 6563196, 6563197, 6563198, 6563199, 6563200, 6563201, 6563202, 6563203, 6563204, 6563206, 6563207, 6563208, 6563209, 6563211, 6563212, 6563213, 6563214, 6731683, 6731684, 6731685, 6731686, 6731688, 6731689, 6731690, 6731691, 6731692, 6731693, 6731694, 6731695, 6731696, 6731697, 6731699, 6731700, 6731701, 6731702, 6731703, 6731704, 6731705, 6731706, 6731707, 6731708, 6731710, 6731711, 6731712, 6731713, 6731714, 6731715, 6731716, 6731719, 6731720, 6731722, 6731723, 6731724, 6731725, 6731726, 6731727, 6731728, 6731729, 6731730, 6731731, 6731732, 6731733, 6735885, 6735886, 6735887, 6735888, 6735889, 6735890, 6735891, 6735892, 6735893, 6735894, 6735895, 6735896, 6735897, 6735898, 6735899, 6735900, 6735901, 6735902, 6735903, 6735904, 6735905, 6735906, 6735907, 6735909, 6735911, 6735912, 6735913, 6735915, 6735919, 6735920, 6735922, 6735923, 6735924, 6735925, 6735926, 6735927, 6735928, 6735929, 6735930, 6735931, 6735933, 6735934, 6735936, 6735937, 6735938, 6735939, 6735940, 6735941, 6735942, 6740457, 6740458, 6740459, 6740460, 6740461, 6740462, 6740463, 6740465, 6740466, 6740468, 6740469, 6740470, 6740471, 6740472, 6740474, 6740475, 6740476, 6740478, 6740479, 6740481, 6740482, 6740484, 6740485, 6740486, 6740487, 6740488, 6740489, 6740490, 6740492, 6740493, 6740494, 6740495, 6740497, 6740499, 6740500, 6740501, 6740502, 6740503, 6740504, 6740505, 6740506, 6740507, 6740508, 6740509, 6740510, 6740511, 6740513, 6740514, 6740516, 6740518, 6740519, 6740520, 6740521, 6740522, 6740523, 6740524, 6740525, 6740526, 6740527, 6740528, 6740529, 6740531, 6740532, 6740533, 6740535, 6740536, 6740537, 6740538, 6740539, 6740541, 6740542, 6740543, 6740544, 6740545, 6740546, 6740551, 6851311, 6851312, 6851313, 6851314, 6851315, 6851316, 6851317, 6851318, 6851319, 6851320, 6851321, 6851322, 6851323, 6851324, 6851325, 6851326, 6851327, 6851328, 6851329, 6851330, 6851331, 6851332, 6851333, 6851334, 6851335, 6851336, 6851338, 6851339, 6851342, 6851345, 6851346, 6851349, 6851350, 6851351, 6851352, 6851353, 6851354, 6851355, 6851356, 6851357, 6851358, 6851359, 6851360, 6851361, 6851362, 6851363, 6851364, 6851365, 6858964, 6858965, 6858966, 6858967, 6858968, 6858969, 6858970, 6858971, 6858972, 6858974, 6858975, 6858976, 6858977, 6858978, 6858979, 6858980, 6858981, 6858982, 6858983, 6858984, 6858985, 6858986, 6858987, 6858988, 6858989, 6858990, 6858991, 6858992, 6858994, 6858995, 6858996, 6858997, 6858998, 6858999, 6859000, 6859001, 6859002, 6859003, 6859004, 6859005, 6859006, 6859007, 6859008, 6859009, 6859010, 6859011, 6859012, 6859013, 6859453, 6859454, 6859457, 6859458, 6859462, 6859463, 6859464, 6859465, 6859466, 6859467, 6859468, 6859472, 6859474, 6859475, 6859476, 6859477, 6859479, 6859483, 6859484, 6859486, 6859487, 6859488, 6859491, 6859494, 6859495, 6859496, 6859497, 6859499, 6859500, 6859502, 6859503, 6859505, 6859507, 6859508, 6859509, 6859510, 6866440, 6866441, 6866442, 6866443, 6866444, 6866445, 6866446, 6866447, 6866448, 6866449, 6866450, 6866451, 6866452, 6866453, 6866454, 6866455, 6866456, 6866457, 6866458, 6866459, 6866460, 6866461, 6866462, 6866463, 6866464, 6866465, 6866467, 6866468, 6866469, 6866470, 6866472, 6866473, 6866474, 6866475, 6866476, 6866477, 6866480, 6866481, 6866482, 6866483, 6866484, 6866485, 6866486, 6866487, 6866488, 6866489, 6866490, 6866491, 6866492, 6866493, 6866494, 7003396, 7003398, 7003399, 7003401, 7003403, 7003404, 7003407, 7003408, 7003409, 7003412, 7003413, 7003414, 7003415, 7003419, 7003420, 7003421, 7003423, 7003425, 7003426, 7003427, 7003428, 7003429, 7003430, 7003431, 7003434, 7003435, 7003437, 7003438, 7003439, 7003440, 7003442, 7003445, 7003447, 7003448, 7003449, 7003451, 7003452, 7003453, 7003454, 7041101, 7041102, 7041104, 7041105, 7041110, 7041112, 7041113, 7041114, 7041115, 7041116, 7041117, 7041118, 7041119, 7041120, 7041122, 7041125, 7041126, 7041127, 7041128, 7041130, 7041131, 7041134, 7041135, 7041138, 7041140, 7041142, 7041143, 7041145, 7041147, 7041149, 7150692, 7150693, 7150697, 7150699, 7150703, 7150708, 7150709, 7150711, 7150712, 7150714, 7150717, 7150718, 7150719, 7150720, 7150721, 7150722, 7150723, 7150724, 7150727, 7150728, 7150729, 7150730, 7150732, 7150734, 7150735, 7150737, 7150738, 7150739, 7150740, 7150741, 7150742, 7150743;
g. REF 132-36-52, GUDID 10885862207036, Serial Numbers: 2408284, 2408285, 2408286, 2408287, 2408288, 2408289, 2408290, 2408291, 2408292, 2408293, 2408294, 2408295, 2408296, 2408297, 2408298, 2408299, 2408300, 2408301, 2408302, 2408303, 2408304, 2408305, 2408306, 2408307, 2408308, 2408309, 2408310, 2408311, 2408312, 2408313, 2408314, 2408315, 2408316, 2408317, 2408318, 2408319, 2408321, 2408322, 2408323, 2424190, 2424191, 2424192, 2424193, 2424194, 2424195, 2424196, 2424197, 2424198, 2424199, 2424200, 2424201, 2424202, 2424203, 2424205, 2424206, 2424207, 2424208, 2424209, 2424210, 2424211, 2424212, 2424213, 2424214, 2433358, 2433359, 2433360, 2433362, 2433363, 2433364, 2433365, 2433366, 2433367, 2433368, 2433369, 2433370, 2433371, 2433372, 2433373, 2433374, 2433375, 2433376, 2433377, 2433378, 2433379, 2433380, 2433381, 2433382, 2433384, 2433385, 2433386, 2433387, 2433388, 2433389, 2433390, 2443113, 2443114, 2443115, 2443116, 2443117, 2443118, 2443119, 2443120, 2443121, 2443122, 2443123, 2443124, 2443125, 2443126, 2443127, 2443128, 2443129, 2443130, 2443132, 2443133, 2443135, 2443136, 2443137, 2443138, 2443139, 2443140, 2443141, 2443142, 2443143, 2443144, 2443145, 2443146, 2443147, 2443148, 2443149, 2443150, 2664704, 2664705, 2664706, 2664707, 2664708, 2664709, 2664710, 2664711, 2664712, 2664713, 2664714, 2664715, 2664716, 2664717, 2664718, 2664719, 2664720, 2664721, 2664722, 2664723, 2664724, 2664725, 2664726, 2664727, 2664728, 2664729, 2664730, 2664731, 2664732, 2664733, 2664734, 2664735, 2664736, 2664737, 2664738, 2664739, 2664740, 2664741, 2664742, 2664743, 2664744, 2664745, 2664746, 2664747, 2664748, 2664749, 2664750, 2664751, 2664752, 2664753, 2667330, 2667331, 2667332, 2667333, 2667334, 2667335, 2667336, 2667337, 2667338, 2667339, 2667340, 2667341, 2667342, 2667343, 2667344, 2667345, 2667346, 2667347, 2667348, 2667349, 2667350, 2667351, 2667352, 2667353, 2667355, 2667356, 2667357, 2667358, 2667359, 2667360, 2667361, 2667362, 2667363, 2667364, 2667365, 2667366, 2667367, 2667368, 2667369, 2667370, 2667371, 2667372, 2667373, 2667374, 2667375, 2667376, 2667377, 2667378, 2667379, 2799526, 2799527, 2799528, 2799529, 2799530, 2799531, 2799532, 2799533, 2799534, 2799535, 2799536, 2799537, 2799538, 2799539, 2799540, 2799541, 2799542, 2799543, 2891992, 2891993, 2891994, 2891995, 2891996, 2891997, 2891998, 2891999, 2892000, 2892001, 2892002, 2892003, 2892004, 2892005, 2892006, 2892007, 2892008, 2892009, 2892010, 2892011, 2892012, 2892013, 2892014, 2892015, 2892016, 2892017, 2892018, 2892019, 2892020, 2892021, 2892022, 2892023, 2892024, 2892025, 2892026, 2892027, 2892028, 2892029, 2892030, 2892031, 2952529, 2952530, 2952531, 2952532, 2952533, 2952534, 2952535, 2952536, 2952537, 2952538, 2952539, 2952540, 2952541, 2952542, 2952543, 2952544, 2958888, 2958889, 2958890, 2958891, 2958892, 2958893, 2958894, 2958895, 2958896, 2958897, 2958898, 2958899, 2958900, 2958901, 2958902, 2958903, 2958904, 2958905, 2958906, 2958907, 2958908, 2958909, 2958910, 2958911, 2958912, 2958913, 2958914, 3544372, 3544373, 3544374, 3544375, 3544376, 3544377, 3544378, 3544379, 3544380, 3544381, 3544382, 3544383, 3544384, 3544385, 3544386, 3544387, 3544388, 3544389, 3544390, 3544391, 3579207, 3579208, 3579209, 3579210, 3579211, 3579212, 3579213, 3579214, 3579215, 3579216, 3579217, 3579218, 3579219, 3579220, 3579221, 3579222, 3579223, 3579225, 3579226, 3579227, 3604575, 3604576, 3604577, 3604578, 3604579, 3604580, 3604581, 3604582, 3604583, 3604584, 3604585, 3604586, 3604587, 3604588, 3604589, 3604590, 3604591, 3604592, 3604593, 3604594, 3604595, 3604596, 3604597, 3604598, 3604599, 3604600, 3604601, 3604602, 3604603, 3604604, 3604606, 3604607, 3668603, 3668604, 3668605, 3668606, 3668607, 3668608, 3668609, 3668610, 3668611, 3668612, 3668613, 3668614, 3668615, 3668617, 3668618, 3668619, 3668620, 3668621, 3668622, 3668623, 3668624, 3668625, 3668626, 3668627, 3668628, 3668629, 3757001, 3757002, 3757003, 3757004, 3757005, 3757006, 3757008, 3757009, 3757010, 3757011, 3757012, 3757013, 3757014, 3757015, 3757017, 3757018, 3757019, 3757020, 3840164, 3840165, 3840166, 3840167, 3840168, 3840169, 3840170, 3840171, 3844276, 3844277, 3844278, 3844279, 3844280, 3844281, 3844282, 3844283, 3844284, 3844285, 3844286, 3844287, 3844288, 3844289, 3844290, 3844291, 3844292, 3844293, 3844294, 3844295, 3844296, 3844297, 3844298, 3844299, 3844300, 3844301, 3844302, 3844303, 3844304, 3844305, 3844306, 3844307, 3844308, 3844309, 3844310, 3844311, 3844312, 3844313, 3844314, 3844315, 3866384, 3866385, 3866386, 3866387, 3866388, 3866389, 3866390, 3866391, 3866392, 3866393, 3866394, 3866395, 3866396, 3866397, 3866398, 3866399, 3866400, 3866401, 3866402, 3866403, 3866404, 3866405, 3866406, 3866407, 3866408, 3866409, 3866410, 3866411, 3866412, 3866413, 3866414, 3866415, 3866416, 3866417, 3866418, 3866419, 3866420, 3866422, 3866423, 3866424, 3937402, 3937403, 3937404, 3937405, 3937406, 3937407, 3937408, 3937409, 3937410, 3937411, 3937412, 3937413, 3937414, 3937415, 3937416, 3937417, 3937418, 3937419, 3937420, 3937421, 3937422, 3937423, 3937424, 3937425, 3937426, 3937427, 3937428, 3937429, 3937430, 3937431, 3948561, 3948562, 3948564, 3948565, 3948566, 3948567, 3948568, 3948569, 3948571, 3948572, 3948573, 3948574, 3948575, 3948576, 3948577, 3948578, 3948579, 3948580, 3948581, 3948582, 3948583, 3948584, 3948585, 4007121, 4007122, 4007123, 4007124, 4007125, 4007126, 4007127, 4007128, 4007129, 4007130, 4007131, 4007132, 4007133, 4007134, 4007136, 4007137, 4007138, 4007139, 4007141, 4007142, 4007143, 4007144, 4007145, 4007146, 4007147, 4056176, 4056177, 4056178, 4056181, 4056182, 4056184, 4056185, 4056186, 4056187, 4056188, 4056189, 4056190, 4056191, 4056192, 4056193, 4056194, 4056195, 4056196, 4056197, 4056198, 4056199, 4056200, 4056201, 4056225, 4056227, 4056228, 4056229, 4056230, 4056231, 4056232, 4056233, 4056234, 4056235, 4056236, 4056237, 4056238, 4056239, 4056240, 4056241, 4056242, 4056243, 4056244, 4056245, 4056246, 4056248, 4056249, 4094543, 4094544, 4094545, 4094546, 4094547, 4094548, 4094549, 4094550, 4094551, 4094552, 4094553, 4094554, 4094555, 4094556, 4094557, 4094558, 4094559, 4094560, 4094561, 4094562, 4145012, 4145013, 4145014, 4145015, 4145016, 4145018, 4145019, 4145020, 4145021, 4145022, 4145023, 4145024, 4145025, 4145026, 4145027, 4145028, 4145029, 4145030, 4145031, 4145032, 4145033, 4145034, 4145035, 4145036, 4145037, 4145038, 4145039, 4145040, 4145041, 4145042, 4145043, 4145044, 4145045, 4145046, 4145047, 4145048, 4145049, 4145050, 4145051, 4145052, 4145053, 4159552, 4159553, 4159556, 4159557, 4159558, 4159559, 4159560, 4159562, 4159563, 4159564, 4159565, 4159567, 4159568, 4159569, 4159570, 4159571, 4159572, 4159573, 4159574, 4159575, 4159576, 4159577, 4159578, 4159579, 4159580, 4159581, 4159582, 4159583, 4159584, 4159586, 4159587, 4159588, 4159589, 4159590, 4159591, 4159592, 4159593, 4159595, 4159596, 4159597, 4159598, 4159599, 4159600, 4159602, 4159603, 4159604, 4159605, 4159606, 4159607, 4159608, 4159609, 4159610, 4159611, 4159612, 4159613, 4159614, 4159615, 4159616, 4159617, 4159618, 4159619, 4159620, 4159621, 4159622, 4159623, 4233581, 4233582, 4233583, 4233584, 4233585, 4233586, 4233587, 4233588, 4233589, 4233590, 4233591, 4233592, 4233593, 4233594, 4233595, 4233596, 4233597, 4233598, 4233599, 4233600, 4233601, 4233602, 4233603, 4233604, 4233605, 4233607, 4233608, 4233609, 4233610, 4233611, 4233612, 4233613, 4233615, 4233616, 4233618, 4233619, 4233620, 4233621, 4233622, 4233623, 4233624, 4233625, 4233626, 4233627, 4233628, 4233629, 4233630, 4233631, 4233632, 4233633, 4233634, 4233635, 4233636, 4325298, 4325299, 4325300, 4325301, 4325302, 4325303, 4325304, 4325305, 4325306, 4325307, 4325308, 4325309, 4325310, 4325311, 4325312, 4325313, 4325315, 4325316, 4325317, 4325318, 4325319, 4325320, 4325321, 4325322, 4325323, 4325324, 4325325, 4325326, 4325327, 4325328, 4325329, 4325330, 4325331, 4325332, 4325333, 4325334, 4325335, 4325336, 4325337, 4325338, 4325339, 4325340, 4325341, 4325342, 4325343, 4325344, 4325345, 4325346, 4325347, 4325348, 4325349, 4325350, 4345638, 4345639, 4345640, 4345641, 4345642, 4345643, 4345644, 4345645, 4345646, 4345647, 4345648, 4345650, 4345651, 4345652, 4345653, 4345654, 4345655, 4345656, 4345657, 4345658, 4345659, 4345660, 4345661, 4345662, 4345663, 4345664, 4345665, 4345666, 4345667, 4345668, 4345669, 4345670, 4345671, 4345672, 4345673, 4345674, 4345675, 4345676, 4345677, 4345678, 4345679, 4345680, 4345681, 4345682, 4345683, 4345684, 4345685, 4345686, 4345687, 4345688, 4345689, 4345691, 4345692, 4345693, 4345694, 4345695, 4345696, 4345697, 4345698, 4345699, 4345700, 4345701, 4345702, 4345703, 4345704, 4345705, 4345706, 4345707, 4345708, 4345709, 4345710, 4345711, 4345712, 4345713, 4345714, 4345715, 4345716, 4345717, 4345718, 4345719, 4345720, 4345721, 4345722, 4345723, 4345724, 4345725, 4345726, 4345727, 4345728, 4438625, 4438626, 4438627, 4438628, 4438629, 4438630, 4438631, 4438632, 4438633, 4438634, 4438635, 4438636, 4438637, 4438638, 4438639, 4438640, 4438641, 4438642, 4438643, 4438644, 4438645, 4438646, 4438647, 4438648, 4438649, 4438650, 4438651, 4438652, 4438653, 4438654, 4438655, 4438656, 4438657, 4438658, 4438659, 4438660, 4438661, 4438662, 4438663, 4438664, 4438665, 4438666, 4438667, 4438668, 4438669, 4438670, 4438671, 4438672, 4438673, 4458543, 4458544, 4458545, 4458546, 4458547, 4458548, 4458549, 4458550, 4458551, 4458552, 4458553, 4458554, 4458555, 4458556, 4458557, 4458558, 4458559, 4458560, 4458561, 4458562, 4458563, 4458564, 4458565, 4458566, 4458567, 4458568, 4458569, 4458570, 4458571, 4458572, 4459068, 4459069, 4459070, 4459071, 4459072, 4459073, 4459074, 4459075, 4459076, 4459077, 4459078, 4459079, 4459080, 4459081, 4459082, 4459083, 4459084, 4459085, 4459086, 4459087, 4459088, 4459089, 4490568, 4490569, 4490570, 4490571, 4490572, 4490573, 4490574, 4490575, 4490576, 4490577, 4490578, 4490579, 4490580, 4490581, 4490582, 4490583, 4490584, 4490585, 4490586, 4490587, 4490588, 4490591, 4490592, 4490593, 4490594, 4490595, 4490596, 4490597, 4490598, 4490599, 4490600, 4490601, 4490602, 4490603, 4490604, 4490605, 4490606, 4490607, 4490608, 4490609, 4490610, 4490611, 4490612, 4490613, 4490614, 4490615, 4490616, 4490617, 4490618, 4490619, 4490620, 4490621, 4490622, 4490623, 4490624, 4490625, 4490626, 4490627, 4538510, 4538511, 4538512, 4538513, 4538514, 4538515, 4538516, 4538517, 4538518, 4538519, 4538520, 4538521, 4538522, 4538523, 4538524, 4538525, 4538526, 4538527, 4538528, 4538529, 4538530, 4538531, 4538532, 4538533, 4538534, 4538535, 4558563, 4558564, 4558565, 4558566, 4558567, 4558568, 4558569, 4558570, 4558571, 4558572, 4558573, 4558574, 4558575, 4558576, 4558577, 4558578, 4558579, 4558580, 4558581, 4558582, 4558583, 4558584, 4558585, 4558586, 4558587, 4558588, 4558589, 4558590, 4558591, 4558592, 4558653, 4558654, 4558655, 4558656, 4558657, 4558658, 4558659, 4558660, 4558661, 4558662, 4558663, 4558664, 4558665, 4558666, 4558667, 4558668, 4558669, 4558670, 4558671, 4558672, 4558673, 4558674, 4558675, 4558676, 4558677, 4558678, 4558679, 4558680, 4558681, 4558682, 4572434, 4572435, 4572436, 4572437, 4572438, 4572439, 4572440, 4572441, 4572442, 4572443, 4572444, 4572445, 4572446, 4572447, 4572448, 4572449, 4572450, 4572451, 4572452, 4572453, 4572454, 4572455, 4572456, 4572457, 4572458, 4572459, 4572460, 4572461, 4572462, 4572463, 4572464, 4572465, 4608082, 4608083, 4608084, 4608085, 4608086, 4608087, 4608088, 4608089, 4608090, 4608091, 4608092, 4608093, 4608094, 4608095, 4608096, 4608097, 4608098, 4608099, 4608100, 4608101, 4608102, 4608103, 4608104, 4608105, 4608106, 4608107, 4608108, 4608109, 4608110, 4608111, 4636789, 4636790, 4636791, 4636792, 4636793, 4636794, 4636795, 4636796, 4636797, 4636798, 4636799, 4636800, 4636801, 4636802, 4636803, 4636804, 4636805, 4636806, 4636807, 4636808, 4636809, 4636811, 4636812, 4655944, 4655945, 4655946, 4655947, 4655948, 4655949, 4655950, 4655951, 4655952, 4655953, 4655954, 4655955, 4655956, 4655957, 4655958, 4655959, 4655960, 4655961, 4655962, 4655963, 4655964, 4655965, 4655966, 4655967, 4655968, 4655969, 4655970, 4655971, 4655972, 4655973, 4674838, 4674839, 4674840, 4674841, 4674842, 4674843, 4674844, 4674845, 4674846, 4674847, 4674848, 4674849, 4674850, 4674851, 4674852, 4674853, 4674854, 4674855, 4674856, 4674857, 4674858, 4674859, 4683394, 4683395, 4683396, 4683397, 4683398, 4683399, 4683400, 4683401, 4683402, 4683403, 4683404, 4683405, 4683406, 4683407, 4683408, 4683409, 4683410, 4683411, 4683412, 4683413, 4683414, 4683415, 4683416, 4683417, 4683418, 4683419, 4683420, 4683421, 4683422, 4683423, 4709638, 4709639, 4709640, 4709641, 4709642, 4709643, 4709644, 4709645, 4709646, 4709647, 4709648, 4709649, 4709650, 4709651, 4709652, 4709653, 4709654, 4709655, 4709656, 4709657, 4709658, 4709659, 4709660, 4709661, 4709662, 4709663, 4709664, 4709665, 4709666, 4709667, 4723128, 4723129, 4723130, 4723131, 4723132, 4723133, 4723134, 4723135, 4723136, 4723137, 4723138, 4723139, 4723140, 4723141, 4723142, 4723143, 4723144, 4723145, 4723146, 4723868, 4723869, 4723870, 4723871, 4723872, 4723873, 4723874, 4723875, 4723876, 4723877, 4723878, 4723879, 4723880, 4723881, 4723882, 4723883, 4723884, 4723885, 4723886, 4723887, 4723888, 4723889, 4723890, 4723891, 4723892, 4723893, 4723894, 4723895, 4723896, 4723897, 4738033, 4738034, 4738035, 4738036, 4738037, 4738038, 4738039, 4738040, 4738041, 4738042, 4738043, 4738044, 4738045, 4738046, 4738047, 4738048, 4738049, 4738050, 4738051, 4738052, 4738053, 4738054, 4738055, 4738056, 4738057, 4738058, 4738059, 4738060, 4738061, 4738062, 4750935, 4750936, 4750937, 4750938, 4750939, 4750940, 4750941, 4750942, 4750943, 4750944, 4750945, 4750946, 4750947, 4750948, 4750949, 4750951, 4750952, 4750953, 4750954, 4750955, 4750956, 4750957, 4750958, 4750959, 4750960, 4750961, 4750962, 4750963, 4750964, 4763527, 4763528, 4763529, 4763530, 4763531, 4763532, 4777153, 4777154, 4777155, 4777156, 4777157, 4777158, 4777159, 4777160, 4777161, 4777162, 4777163, 4777164, 4777165, 4777166, 4777167, 4777168, 4777169, 4777170, 4777171, 4777172, 4777173, 4777174, 4777175, 4777176, 4777177, 4777178, 4777179, 4777180, 4777181, 4777182, 4779224, 4779225, 4779226, 4779227, 4779228, 4779229, 4779230, 4779231, 4779232, 4779233, 4779234, 4779235, 4779236, 4779237, 4779238, 4779239, 4779240, 4779241, 4779242, 4779243, 4779244, 4779245, 4779246, 4779247, 4779248, 4779249, 4779250, 4779251, 4779252, 4779253, 4790117, 4790118, 4790120, 4790121, 4790122, 4790123, 4790124, 4790125, 4790126, 4790127, 4790128, 4790129, 4790130, 4790131, 4790132, 4790133, 4790134, 4790135, 4790136, 4790138, 4790139, 4790140, 4790141, 4790142, 4790143, 4790144, 4790145, 4790146, 4817028, 4817029, 4817030, 4817031, 4817032, 4817033, 4817034, 4817035, 4817036, 4817037, 4817038, 4817039, 4817040, 4817041, 4817042, 4817043, 4817044, 4817045, 4817046, 4817047, 4817048, 4817049, 4817050, 4817051, 4817052, 4817053, 4817054, 4817055, 4817056, 4817057, 4830357, 4830358, 4830359, 4830360, 4830361, 4830362, 4830363, 4830364, 4830365, 4830366, 4830367, 4830368, 4830369, 4830370, 4830371, 4830372, 4830373, 4830374, 4830375, 4830376, 4830377, 4830378, 4830379, 4830380, 4830381, 4830382, 4830383, 4830384, 4830385, 4830386, 4834936, 4834937, 4834938, 4834939, 4834940, 4834941, 4834942, 4834943, 4834944, 4834945, 4834946, 4834947, 4834948, 4834949, 4834950, 4834951, 4834952, 4834953, 4834954, 4834955, 4834956, 4834957, 4834958, 4834959, 4834960, 4834961, 4834962, 4834963, 4834964, 4834965, 4854441, 4854442, 4854443, 4854444, 4854445, 4854446, 4854447, 4854448, 4854449, 4854450, 4854451, 4854452, 4854453, 4854454, 4854455, 4854456, 4854457, 4854458, 4854459, 4854460, 4854461, 4854462, 4854463, 4854464, 4854465, 4854466, 4854467, 4854468, 4854469, 4854470, 4861357, 4861358, 4861359, 4861360, 4861361, 4861362, 4861363, 4861364, 4861365, 4861366, 4861367, 4861368, 4861369, 4861370, 4861371, 4861372, 4861373, 4861374, 4861375, 4861376, 4861377, 4861378, 4861379, 4861380, 4861381, 4861382, 4861383, 4861384, 4861385, 4861386, 4886453, 4886454, 4886455, 4886456, 4886457, 4886458, 4886459, 4886460, 4886461, 4886462, 4886463, 4886464, 4886465, 4886466, 4886467, 4886468, 4886469, 4886470, 4886471, 4886472, 4886474, 4886475, 4886476, 4886477, 4886478, 4886479, 4886480, 4886481, 4886482, 4902478, 4902479, 4902480, 4902481, 4902482, 4902483, 4902484, 4902485, 4902486, 4902487, 4902488, 4902489, 4902490, 4902491, 4902492, 4902493, 4902494, 4902495, 4902496, 4902497, 4902498, 4902499, 4902500, 4902501, 4902502, 4902503, 4902504, 4902505, 4902506, 4902507, 4940931, 4940932, 4940933, 4940934, 4940935, 4940936, 4940937, 4940938, 4940939, 4940940, 4940941, 4940942, 4940943, 4940944, 4940945, 4940946, 4940947, 4940948, 4940949, 4940950, 4940951, 4940952, 4940953, 4940954, 4940955, 4940956, 4940957, 4940958, 4940959, 4940960, 4957327, 4957328, 4957329, 4957330, 4957331, 4957332, 4957333, 4957334, 4957335, 4957336, 4957337, 4957338, 4957339, 4957340, 4957341, 4957342, 4957343, 4957344, 4957345, 4957346, 4957347, 4957348, 4957349, 4957350, 4957351, 4957352, 4957353, 4957354, 4957355, 4957356, 4977238, 4977239, 4977240, 4977241, 4977242, 4977243, 4977244, 4977245, 4977246, 4977247, 4977248, 4977249, 4977250, 4977251, 4977252, 4977253, 4977254, 4977255, 4977256, 4977257, 4977258, 4977259, 4977260, 4977261, 4977262, 4977263, 4977264, 4977265, 4977266, 4977267, 4991826, 4991827, 4991828, 4991829, 4991830, 4991831, 4991832, 4991833, 4991834, 4991835, 4991836, 4991837, 4991838, 4991839, 4991840, 4991841, 4991842, 4991843, 4991844, 4991845, 4991847, 4991848, 4991849, 4991850, 4991851, 4991852, 4991853, 4991854, 4991855, 5013196, 5013197, 5013198, 5013199, 5013200, 5013201, 5013202, 5013203, 5013205, 5013206, 5013207, 5013208, 5013209, 5013210, 5013211, 5013212, 5013213, 5013214, 5013215, 5013216, 5013217, 5013218, 5013219, 5013220, 5013221, 5013222, 5013223, 5013224, 5013225, 5013226, 5013227, 5023024, 5040776, 5040777, 5040778, 5040780, 5040781, 5040782, 5040783, 5040784, 5040785, 5040786, 5040787, 5040788, 5040789, 5040790, 5040791, 5040792, 5040794, 5040795, 5040796, 5040797, 5040798, 5040799, 5040801, 5040802, 5040803, 5040804, 5040805, 5040807, 5040808, 5040809, 5040810, 5040811, 5078728, 5078729, 5078730, 5078731, 5078732, 5078733, 5078734, 5078735, 5078736, 5078737, 5078738, 5078739, 5078740, 5078741, 5078742, 5078743, 5078744, 5078745, 5078746, 5078747, 5078748, 5078749, 5078750, 5078751, 5078752, 5078753, 5078754, 5078755, 5078756, 5078757, 5078758, 5078759, 5078760, 5078761, 5078762, 5078763, 5097969, 5097970, 5097971, 5097972, 5097973, 5097974, 5097975, 5097976, 5097977, 5097978, 5097979, 5097980, 5097981, 5097982, 5097983, 5097984, 5097985, 5097986, 5097987, 5097988, 5097989, 5097990, 5097991, 5097992, 5097993, 5097994, 5097995, 5097996, 5097997, 5097998, 5122839, 5122840, 5122841, 5122842, 5122843, 5122844, 5122845, 5122846, 5122847, 5122848, 5122849, 5122850, 5122851, 5122852, 5122853, 5122854, 5122855, 5122856, 5122857, 5122858, 5122859, 5122860, 5122861, 5122862, 5122863, 5122864, 5122865, 5122866, 5122867, 5122868, 5134235, 5134236, 5134237, 5134238, 5134239, 5134240, 5134241, 5134242, 5134243, 5134244, 5134245, 5134246, 5134247, 5134248, 5134249, 5134250, 5134251, 5134252, 5134253, 5134254, 5134255, 5134256, 5134257, 5134258, 5134259, 5134260, 5134261, 5134262, 5134263, 5134264, 5140286, 5140287, 5140288, 5140289, 5140290, 5140291, 5140292, 5140293, 5140294, 5140295, 5140296, 5140297, 5140298, 5140299, 5140300, 5140301, 5140302, 5140303, 5140304, 5140305, 5140306, 5140307, 5140308, 5140309, 5140310, 5140311, 5140312, 5140313, 5140314, 5140315, 5155307, 5155308, 5155309, 5155310, 5155311, 5155312, 5155313, 5155314, 5155315, 5155316, 5155317, 5155318, 5155319, 5155320, 5155321, 5155322, 5155323, 5155324, 5155325, 5155326, 5155327, 5155328, 5155329, 5155330, 5155331, 5155332, 5155333, 5155334, 5155335, 5169149, 5169150, 5169151, 5169152, 5169153, 5169154, 5169155, 5169156, 5169157, 5169158, 5169159, 5169160, 5169161, 5169162, 5169163, 5169164, 5169165, 5169166, 5169167, 5169168, 5169169, 5169170, 5169171, 5169172, 5169173, 5169174, 5169175, 5169176, 5169177, 5169178, 5196663, 5196664, 5196665, 5196666, 5196667, 5196668, 5196669, 5196670, 5196671, 5196672, 5196673, 5196674, 5196675, 5196676, 5196677, 5196678, 5196679, 5196680, 5196681, 5196682, 5196683, 5196684, 5196685, 5196686, 5196687, 5196688, 5196689, 5196690, 5196691, 5196692, 5196693, 5196694, 5196695, 5196696, 5214814, 5214815, 5214816, 5214817, 5214818, 5214819, 5214820, 5214821, 5214822, 5214823, 5214824, 5214825, 5214826, 5214827, 5214828, 5214829, 5214830, 5214831, 5214832, 5214833, 5214834, 5214835, 5214836, 5214837, 5214838, 5214839, 5214840, 5214841, 5214842, 5214843, 5256345, 5256346, 5256347, 5256348, 5256349, 5256350, 5256351, 5256352, 5256353, 5256354, 5256355, 5256356, 5256357, 5256358, 5256359, 5256360, 5256361, 5256362, 5256363, 5256364, 5256365, 5256366, 5256367, 5256368, 5256369, 5256370, 5256371, 5256372, 5256373, 5256374, 5256375, 5256376, 5256377, 5256378, 5256379, 5256380, 5256381, 5256382, 5256383, 5256384, 5266897, 5266898, 5266899, 5266900, 5266901, 5266902, 5266903, 5266904, 5266905, 5266906, 5266907, 5266908, 5266909, 5266910, 5266911, 5266912, 5266913, 5266914, 5266915, 5266916, 5266917, 5266918, 5266919, 5266920, 5266921, 5266922, 5266923, 5266924, 5266925, 5266926, 5266927, 5266928, 5266929, 5266930, 5266931, 5266932, 5266933, 5266934, 5266935, 5266936, 5303233, 5303234, 5303235, 5303236, 5303237, 5303238, 5303239, 5303240, 5303241, 5303242, 5303243, 5303244, 5303245, 5303246, 5303247, 5303248, 5303249, 5303250, 5303251, 5303252, 5303253, 5303254, 5303255, 5303256, 5303257, 5303258, 5303259, 5303260, 5303261, 5303262, 5303263, 5303264, 5303265, 5303266, 5303267, 5303268, 5303269, 5303270, 5347833, 5347834, 5347835, 5347836, 5347837, 5347838, 5347839, 5347840, 5347841, 5347842, 5347843, 5347844, 5347845, 5347846, 5347847, 5347848, 5347849, 5347850, 5347851, 5347852, 5347853, 5347854, 5347855, 5347856, 5347857, 5347858, 5347859, 5347860, 5347861, 5347862, 5380535, 5380536, 5380537, 5380538, 5380539, 5380540, 5380542, 5380543, 5380544, 5380545, 5380546, 5380547, 5380548, 5380549, 5380550, 5380551, 5380552, 5380553, 5380554, 5380555, 5380556, 5380557, 5380558, 5380559, 5380560, 5380561, 5380562, 5380563, 5380564, 5403209, 5403210, 5403211, 5403212, 5403213, 5403214, 5403215, 5403216, 5403217, 5403218, 5403219, 5403220, 5403221, 5403222, 5403223, 5403224, 5403225, 5403226, 5403227, 5403228, 5403229, 5403230, 5403231, 5403232, 5403233, 5403234, 5403235, 5403236, 5403237, 5427113, 5427114, 5427115, 5427116, 5427117, 5427118, 5427119, 5427120, 5427121, 5427122, 5427123, 5427124, 5427125, 5427126, 5427128, 5427129, 5427130, 5427131, 5427132, 5427133, 5427134, 5427135, 5427136, 5427137, 5427138, 5427139, 5427140, 5427141, 5427142, 5427143, 5427144, 5427145, 5427146, 5427147, 5442759, 5442760, 5442761, 5442762, 5442763, 5442764, 5442765, 5442766, 5442767, 5442768, 5442769, 5442770, 5442771, 5442772, 5442773, 5442774, 5442775, 5442776, 5442777, 5442778, 5442779, 5442780, 5442781, 5442782, 5442783, 5442784, 5442785, 5442786, 5442787, 5442788, 5485197, 5485198, 5485199, 5485200, 5485201, 5485202, 5485203, 5485204, 5485205, 5485206, 5485207, 5485208, 5485209, 5485210, 5485211, 5485212, 5485213, 5485214, 5485215, 5485216, 5485217, 5485218, 5485219, 5485220, 5485221, 5485222, 5485223, 5485224, 5485225, 5485226, 5504112, 5504113, 5504114, 5504115, 5504116, 5504117, 5504119, 5504120, 5504121, 5504122, 5504123, 5504124, 5504125, 5504126, 5504127, 5504128, 5504129, 5504130, 5504131, 5504132, 5504133, 5504134, 5504135, 5504136, 5504137, 5504138, 5504139, 5504140, 5504141, 5519108, 5519109, 5519110, 5519111, 5519112, 5519113, 5519114, 5519115, 5519116, 5519118, 5519119, 5519120, 5519121, 5519122, 5519123, 5519124, 5519125, 5519126, 5519127, 5519128, 5519129, 5519130, 5519131, 5519132, 5519133, 5519134, 5519135, 5519136, 5519137, 5540084, 5540085, 5540086, 5540087, 5540088, 5540089, 5540090, 5540091, 5540092, 5540093, 5540094, 5540095, 5540096, 5540097, 5540098, 5540099, 5540100, 5540101, 5540102, 5540103, 5540104, 5540105, 5540106, 5540107, 5540108, 5540109, 5540110, 5540111, 5540112, 5540113, 5558654, 5558655, 5558656, 5558657, 5558658, 5558659, 5558660, 5558661, 5558662, 5558663, 5558664, 5558665, 5558666, 5558667, 5558668, 5558669, 5558670, 5558671, 5558672, 5558673, 5558674, 5558675, 5558676, 5558677, 5558678, 5558679, 5558680, 5558681, 5558682, 5558683, 5558684, 5558685, 5598276, 5598277, 5598278, 5598279, 5598280, 5598282, 5598283, 5598284, 5598285, 5598286, 5598287, 5598288, 5598289, 5598290, 5598291, 5598292, 5598293, 5598294, 5598295, 5598296, 5598297, 5598298, 5598299, 5598300, 5598301, 5598302, 5598303, 5598304, 5598305, 5600880, 5600881, 5600882, 5600883, 5600884, 5600885, 5600886, 5600887, 5600888, 5600889, 5600890, 5600891, 5600892, 5600893, 5600894, 5600895, 5600896, 5600897, 5600898, 5600899, 5600900, 5600901, 5600902, 5600903, 5600904, 5600905, 5600906, 5600907, 5600908, 5600909, 5667240, 5667241, 5667242, 5667243, 5667244, 5667245, 5667246, 5667247, 5667248, 5667249, 5667250, 5667251, 5667252, 5667253, 5667254, 5667255, 5667256, 5667257, 5667258, 5667259, 5667260, 5667261, 5667262, 5667263, 5667264, 5667265, 5667266, 5667267, 5667268, 5667269, 5677801, 5677802, 5677803, 5677804, 5677805, 5677806, 5677807, 5677808, 5677809, 5677810, 5677811, 5677812, 5677813, 5677814, 5677815, 5677816, 5677817, 5677818, 5760837, 5760838, 5760839, 5760840, 5760841, 5760842, 5760843, 5760844, 5760845, 5760846, 5760847, 5760848, 5760849, 5760850, 5760851, 5760852, 5760853, 5760854, 5760855, 5760856, 5760857, 5760858, 5760859, 5760860, 5760861, 5760862, 5760863, 5760864, 5760865, 5760866, 5760917, 5760918, 5760919, 5760920, 5760921, 5760922, 5760923, 5760924, 5760925, 5760926, 5760927, 5760928, 5760929, 5760930, 5760931, 5760932, 5760933, 5760934, 5760936, 5760937, 5760938, 5760939, 5760940, 5760941, 5760942, 5760943, 5760944, 5760945, 5760946, 5760947, 5760948, 5760949, 5760950, 5760951, 5760952, 5760953, 5760954, 5760955, 5760956, 5760957, 5760958, 5760959, 5760960, 5760961, 5760962, 5760963, 5760964, 5760965, 5760966, 5760967, 5760968, 5760969, 5760970, 5760971, 5760972, 5760973, 5760974, 5760975, 5760976, 5762138, 5762139, 5762140.5762141, 5762142, 5762143, 5762144, 5762145, 5762146, 5762147, 5762148, 5762149, 5762150, 5762151, 5762152, 5762153, 5762154, 5762156, 5762157, 5762158, 5762217, 5762218, 5762219, 5762220, 5762221, 5762222, 5762223, 5762224, 5762225, 5762226, 5762227, 5762228, 5762229, 5762230, 5762231, 5762232, 5762233, 5762234, 5762235, 5762236, 5762237, 5762238, 5762239, 5762240, 5762241, 5762242, 5762243, 5762244, 5762245, 5762246, 5876956, 5876957, 5876958, 5876959, 5876960, 5876961, 5876962, 5876963, 5876964, 5876965, 5876966, 5876967, 5876968, 5876969, 5876970, 5876971, 5876972, 5876973, 5876974, 5876975, 5876976, 5876977, 5876978, 5876979, 5876980, 5876981, 5876982, 5933549, 5933550, 5933551, 5933552, 5933553, 5933554, 5933555, 5933556, 5933557, 5933558, 5933559, 5933560, 5933561, 5933562, 5933563, 5933564, 5933565, 5933566, 5933567, 5933568, 5933569, 5933570, 5933571, 5933572, 5933573, 5933574, 5933575, 6072170, 6072171, 6072172, 6072173, 6072174, 6072175, 6072176, 6072177, 6072178, 6072179, 6072180, 6072181, 6072182, 6072183, 6072184, 6072185, 6072186, 6072187, 6072188, 6072189, 6072190, 6072191, 6072192, 6072193, 6072194, 6072195, 6072196, 6072197, 6072198, 6072199, 6072200, 6072201, 6072202, 6072203, 6072204, 6072205, 6072206, 6072207, 6072208, 6072209, 6072210, 6072211, 6072212, 6072213, 6072214, 6072215, 6072216, 6072217, 6072218, 6072219, 6072220, 6072221, 6072222, 6072223, 6072224, 6072225, 6072226, 6072227, 6072228, 6072230, 6072231, 6072232, 6072234, 6072235, 6072237, 6072238, 6072239, 6072240, 6072241, 6072242, 6072243, 6149316, 6149317, 6149318, 6149319, 6149320, 6149321, 6149322, 6149323, 6149324, 6149325, 6149326, 6149327, 6149328, 6149329, 6149330, 6149331, 6149332, 6149334, 6149335, 6149336, 6149337, 6149338, 6149339, 6149340, 6149341, 6149342, 6149343, 6149344, 6149345, 6149346, 6149347, 6149348, 6153633, 6153634, 6153635, 6153636, 6153637, 6153638, 6153639, 6153640, 6153641, 6153642, 6153643, 6153644, 6153645, 6153647, 6153648, 6153649, 6153650, 6153651, 6153652, 6153653, 6153654, 6153655, 6153656, 6153657, 6153658, 6153659, 6153660, 6153661, 6153662, 6153663, 6153664, 6153665, 6153666, 6153667, 6153668, 6153670, 6153671, 6153672, 6153673, 6153674, 6153675, 6153676, 6153677, 6153678, 6153679, 6153680, 6153681, 6153682, 6153683, 6153684, 6153685, 6153686, 6153687, 6153688, 6153689, 6153690, 6179202, 6179203, 6179204, 6179206, 6179207, 6179208, 6179209, 6179210, 6179211, 6179213, 6179214, 6179216, 6179217, 6179218, 6179219, 6179220, 6179221, 6179222, 6179223, 6179224, 6179226, 6179227, 6179228, 6194528, 6194529, 6194530, 6194531, 6194532, 6194533, 6194534, 6194535, 6194536, 6194537, 6194538, 6194539, 6194540, 6194541, 6194542, 6194543, 6194544, 6194545, 6194546, 6194547, 6194548, 6194549, 6194550, 6194551, 6194552, 6194553, 6194554, 6194555, 6194556, 6194557, 6194558, 6194559, 6236487, 6236488, 6236489, 6236490, 6236491, 6236492, 6236493, 6236494, 6236495, 6236496, 6236497, 6236498, 6236499, 6236500, 6236501, 6236502, 6236503, 6236504, 6236505, 6236506, 6236507, 6236508, 6236509, 6236510, 6236511, 6236512, 6236513, 6236514, 6236515, 6236516, 6236517, 6236518, 6236519, 6251287, 6251288, 6251289, 6251290, 6251291, 6251292, 6251293, 6251294, 6251295, 6251296, 6251297, 6251298, 6251299, 6251300, 6251301, 6251302, 6251303, 6251304, 6251305, 6251306, 6251307, 6251308, 6251309, 6251310, 6251311, 6251312, 6251313, 6251314, 6252774, 6252775, 6252776, 6252777, 6252778, 6252779, 6252780, 6252781, 6252783, 6252784, 6252785, 6252786, 6252787, 6252788, 6252789, 6252790, 6252791, 6252792, 6340723, 6340725, 6340726, 6340727, 6340728, 6340729, 6340730, 6340731, 6340732, 6340733, 6340734, 6340735, 6340736, 6340737, 6340738, 6340741, 6340742, 6340743, 6340744, 6340745, 6340746, 6340747, 6340748, 6340750, 6340751, 6340752, 6340753, 6340754, 6340755, 6340756, 6340757, 6340758, 6340759, 6340760, 6340761, 6340762, 6340763, 6340764, 6340765, 6340766, 6340767, 6340768, 6340769, 6340770, 6340771, 6340772, 6340773, 6340775, 6340776, 6340778, 6340779, 6340780, 6340781, 6340782, 6340783, 6340784, 6340786, 6340787, 6340788, 6340789, 6340790, 6340792, 6340793, 6340794, 6340796, 6340797, 6340798, 6340799, 6340800, 6340801, 6340802, 6340803, 6340804, 6340805, 6340808, 6340809, 6340810, 6340814, 6340815, 6340925, 6340926, 6340927, 6340928, 6340929, 6340930, 6340931, 6340932, 6340933, 6340934, 6340935, 6340936, 6340937, 6340938, 6340939, 6340940, 6340941, 6340943, 6340944, 6340945, 6340946, 6340947, 6340948, 6340949, 6340950, 6340951, 6340952, 6340953, 6515674, 6515675, 6515676, 6515677, 6515678, 6515679, 6515680, 6515681, 6515682, 6515683, 6515684, 6515686, 6515687, 6515688, 6515689, 6515690, 6515691, 6515692, 6515693, 6515694, 6515695, 6515696, 6515697, 6515698, 6515699, 6515700, 6515701, 6515702, 6515704, 6515705, 6634177, 6634178, 6634179, 6634180, 6634181, 6634182, 6634183, 6634184, 6634185, 6634186, 6634189, 6634190, 6634191, 6634192, 6634193, 6634194, 6634195, 6634196, 6634197, 6634198, 6634199, 6634200, 6634201, 6634202, 6634203, 6634204, 6634205, 6634206, 6634207, 6634208, 6634209, 6634210, 6634211, 6634212, 6634213, 6634215, 6634217, 6634218, 6634219, 6634220, 6634221, 6634222, 6634223, 6634226, 6634227, 6634229, 6634231, 6634232, 6634234, 6634235, 6634237, 6634238, 6634240, 6658249, 6658251, 6658252, 6658253, 6658254, 6658255, 6658256, 6658258, 6658259, 6658260, 6658262, 6658263, 6658264, 6658267, 6658269, 6658271, 6658272, 6658273, 6658274, 6658275, 6658277, 6658278, 6658279, 6658280, 6658281, 6658282, 6658283, 6658285, 6658286, 6658287, 6658288, 6658289, 6658290, 6658291, 6658292, 6658293, 6658294, 6658295, 6658296, 6658297, 6658298, 6658299, 6658300, 6658301, 6658303, 6658304, 6658305, 6658306, 6658307, 6658308, 6658309, 6792047, 6792048, 6792049, 6792050, 6792051, 6792052, 6792054, 6792055, 6792056, 6792057, 6792058, 6792059, 6792060, 6792061, 6792062, 6792063, 6792064, 6792065, 6792066, 6792067, 6792068, 6792070, 6792071, 6792072, 6792075, 6792076, 6792077, 6792078, 6792079, 6792080, 6792081, 6792082, 6792083, 6792085, 6792086, 6792087, 6792088, 6792089, 6792090, 6792091, 6792092, 6792093, 6792094, 6792095, 6792096, 6792097, 6792098, 6792099, 6792100, 6792101, 6792102, 6792103, 6792105, 6792106, 6797379, 6797380, 6797381, 6797382, 6797383, 6797384, 6797385, 6797386, 6797387, 6797388, 6797389, 6797390, 6797391, 6797392, 6797393, 6797394, 6797395, 6797397, 6797398, 6797399, 6797400, 6797401, 6797402, 6877129, 6877130, 6877131, 6877132, 6877133, 6877134, 6877135, 6877136, 6877137, 6877138, 6877139, 6877140, 6877141, 6877143, 6877144, 6877145, 6877146, 6877147, 6877148, 6877149, 6877150, 6877151, 6877152, 6877153, 6877154, 6877155, 6877156, 6877159, 6877160, 6877161, 6877162, 6877163, 6877164, 6877165, 6877166, 6877167, 6877169, 6877170, 6877171, 6877172, 6877173, 6877174, 6877175, 6877176, 6903408, 6903409, 6903410, 6903411, 6903412, 6903413, 6903414, 6903416, 6903417, 6903418, 6903419, 6903421, 6903422, 6903423, 6903424, 6903425, 6903426, 6903427, 6903428, 6903429, 6903430, 6903431, 6903432, 6903433, 6903434, 6903437, 6903438, 6903439, 6903440, 6903441, 6903442, 6903443, 6903444, 6903445, 6903446, 6903447, 6903449, 6903450, 6903451, 6903452, 6903483, 6903484, 6903485, 6903486, 6903487, 6903488, 6903490, 6903491, 6903492, 6903493, 6903494, 6903495, 6903497, 6903498, 6903499, 6903500, 6903501, 6903503, 6903504, 6903505, 6903507, 6903508, 6903509, 6903510, 6903511, 6903513, 6903514, 6903515, 6903516, 6903517, 6903519, 6903520, 6903521, 6903524, 6903525, 6903526, 6903527, 6919518, 6919519, 6919520, 6919521, 6919523, 6919524, 6919527, 6919528, 6919529, 6919530, 6919531, 6919534, 6919535, 6919536, 6991661, 6991662, 6991663, 6991664, 6991665, 6991666, 6991667, 6991668, 6991669, 6991670, 6991671, 6991672, 6991673, 6991676, 6991677, 6991678, 6991679, 6991681, 6991683, 6991684, 6991686, 6991687, 6991689, 6991691, 6991692, 6991693, 7036227, 7036228, 7036229, 7036230, 7036231, 7036232, 7036234, 7036235, 7036237, 7036238, 7036239, 7036240, 7036241, 7036242, 7036243, 7036244, 7036245, 7036246, 7036247, 7036248, 7036249, 7036250, 7036251, 7036252, 7036254, 7036255, 7036256, 7036257, 7036259, 7036260, 7036261, 7036262, 7036263, 7036265, 7036266, 7036267, 7036269, 7036271, 7036274, 7036275, 7105067, 7105070, 7105071, 7105073, 7105075, 7105076, 7105077, 7105078, 7105079, 7105083, 7105084, 7105085, 7105086, 7105087, 7105088, 7105090, 7105091, 7105092, 7105093, 7143511, 7143513, 7143514, 7143518, 7143523, 7143524, 7143525, 7143528, 7143534, 7143536, 7143537, 7143539, 7143540, 7143541, 7143542, 7143544, 7143545, 7143547, 7143548, 7143549, 7143551, 7143552, 7143553, 7143555, 7143558, 7143561, 7143562, 7143563;
h. REF 132-40-53, GUDID 10885862207043, Serial Numbers: 2408232, 2408233, 2408234, 2408235, 2408236, 2408237, 2408238, 2408240, 2408243, 2408244, 2408245, 2408246, 2408247, 2408248, 2408249, 2408250, 2408252, 2408253, 2408254, 2408255, 2408256, 2408257, 2408258, 2408260, 2408261, 2408262, 2408263, 2408264, 2408265, 2408266, 2408267, 2408268, 2408269, 2408270, 2408271, 2443156, 2443157, 2443159, 2443160, 2443163, 2443164, 2443165, 2443166, 2443167, 2443168, 2443169, 2443170, 2443171, 2443172, 2443173, 2443174, 2443175, 2443176, 2443177, 2581848, 2581849, 2581850, 2581851, 2581852, 2581853, 2581854, 2581855, 2581856, 2581857, 2581858, 2581859, 2581860, 2581861, 2581862, 2581863, 2581864, 2581866, 2581867, 2586430, 2586431, 2586433, 2586434, 2586435, 2586436, 2586437, 2586438, 2586440, 2586442, 2586443, 2586444, 2586445, 2586446, 2586447, 2586448, 2586449, 2586450, 2586451, 2586452, 2659398, 2659400, 2659401, 2659402, 2659403, 2659404, 2659405, 2659406, 2659407, 2659408, 2659410, 2659411, 2659412, 2659413, 2659414, 2659415, 2659417, 2659419, 2659420, 2659422, 2659423, 2659426, 2659427, 2659428, 2659429, 2659432, 2659433, 2659434, 2659436, 2659437, 2659438, 2659440, 2659442, 2659444, 2664656, 2664658, 2664660, 2664661, 2664662, 2664663, 2664664, 2664665, 2664666, 2664667, 2664668, 2664669, 2664670, 2664671, 2664672, 2664673, 2664674, 2664675, 2664677, 2664678, 2664680, 2664681, 2664682, 2664683, 2664684, 2664685, 2664686, 2664687, 2664688, 2664689, 2664690, 2664691, 2664692, 2664693, 2664695, 2664696, 2664697, 2664698, 2664700, 2664702, 2664703, 2667200, 2667201, 2667202, 3000463, 3000464, 3000465, 3000467, 3000468, 3000470, 3000472, 3000473, 3000474, 3000475, 3000476, 3000477, 3000478, 3000479, 3000480, 3000481, 3000482, 3000483, 3000484, 3000486, 3000487, 3000488, 3000489, 3000490, 3640624, 3640625, 3727280, 3727281, 3727283, 3727284, 3791783, 3791784, 3791785, 3791787, 3791788, 3791789, 3791790, 3791791, 3791792, 3808324, 3808325, 3808326, 3808327, 3808328, 3808329, 3808330, 3808331, 3808332, 3808333, 3808334, 3808335, 3808336, 3808337, 3808338, 3808339, 3808340, 3808341, 3808342, 3808343, 4201567, 4201568, 4201569, 4201570, 4201571, 4201572, 4201574, 4201575, 4201577, 4201578, 4201579, 4210568, 4210569, 4210570, 4210572, 4210573, 4210574, 4210575, 4210576, 4210577, 4210578, 4210579, 4210580, 4210581, 4210582, 4210583, 4210584, 4210585, 4210586, 4210587, 4210588, 4210589, 4210591, 4210593, 4210594, 4210595, 4210596, 4210597, 4210598, 4210599, 4210600, 4210601, 4210602, 4210603, 4210604, 4210605, 4210606, 4210607, 4210609, 4210610, 4210614, 4210615, 4210617, 4210618, 4210619, 4210620, 4210621, 4210622, 4210623, 4210624, 4254468, 4254469, 4254470, 4254471, 4254472, 4254475, 4254476, 4254478, 4254481, 4254482, 4254483, 4254485, 4254486, 4254487, 4369434, 4369435, 4369436, 4369437, 4369438, 4369439, 4369441, 4369443, 4369445, 4369446, 4369447, 4369448, 4369449, 4369450, 4369451, 4369452, 4369453, 4369454, 4369455, 4369457, 4369459, 4369460, 4369461, 4369462, 4369463, 4369464, 4369465, 4369466, 4369467, 4369468, 4369469, 4369470, 4369471, 4369472, 4369474, 4369476, 4369477, 4369478, 4369480, 4369481, 4369482, 4369485, 4369486, 4369488, 4369489, 4369490, 4369491, 4369492, 4383566, 4383567, 4383568, 4383569, 4383570, 4383572, 4383573, 4383574, 4383575, 4383576, 4383577, 4383578, 4383579, 4383580, 4383582, 4383584, 4383585, 4383586, 4383587, 4383588, 4383589, 4383590, 4383591, 4383592, 4926116, 4926117, 4926118, 4926119, 4926120, 4926121, 4926122, 4926123, 4926124, 4926125, 4926126, 4926127, 4926128, 4926129, 4926130, 4926131, 4926132, 4926133, 4926134, 4926135, 4926136, 4926137, 4926139, 4926140, 4926141, 4926142, 4926143, 4926144, 4926145, 5013314, 5013317, 5013320, 5013321, 5013323, 5013325, 5013326, 5013327, 5013328, 5013329, 5013330, 5013331, 5013332, 5013333, 5013334, 5013335, 5013336, 5013337, 5013338, 5013339, 5013340, 5013341, 5013342, 5013343, 5117568, 5117569, 5117570, 5117571, 5117572, 5117573, 5117574, 5117575, 5117576, 5117577, 5117578, 5117579, 5117580, 5117581, 5117582, 5117583, 5117586, 5117587, 5117588, 5117589, 5117590, 5117591, 5117594, 5151313, 5151314, 5151316, 5151318, 5151319, 5151320, 5151321, 5151322, 5151323, 5151324, 5151325, 5151326, 5151327, 5194535, 5194536, 5194537, 5194538, 5194539, 5194540, 5194541, 5194542, 5194543, 5194544, 5194545, 5194548, 5194549, 5194550, 5194552, 5194553, 5194555, 5194556, 5194557, 5268717, 5268719, 5268720, 5268721, 5268722, 5268723, 5268724, 5268725, 5268727, 5268728, 5268729, 5268730, 5268731, 5268732, 5268733, 5268734, 5268735, 5268736, 5268737, 5268739, 5268740, 5268743, 5268744, 5268745, 5268746, 5419807, 5419810, 5419811, 5419812, 5419815, 5419816, 5419817, 5419818, 5419821, 5419824, 5419827, 5419828, 5419829, 5419831, 5419832, 5419833, 5419836;
i. REF 132-40-54, GUDID 10885862207050, Serial Numbers: 2449597, 2449598, 2475678, 2475681, 2475682, 2475684, 2536990, 2536992, 2536994, 2536995, 2536997, 2536998, 2536999, 2537000, 2537001, 2537004, 2537006, 2537007, 2537008, 2537009, 2537010, 2537015, 2537016, 2537018, 2537019, 2537020, 2537021, 2537022, 2537024, 2537026, 2537027, 2537028, 2537029, 2537030, 2537036, 2537037, 2537038, 2537039, 2585501, 2585502, 2585503, 2585504, 2585505, 2585506, 2585508, 2585510, 2585511, 2585512, 2585513, 2585516, 2585518, 2585519, 2585520, 2585523, 2585524, 2585525, 2585526, 2585529, 2585530, 2585531, 2585534, 2585537, 2667733, 2667734, 2667737, 2667738, 2667740, 2667741, 2667743, 2667744, 2667746, 2667748, 2667752, 2667755, 2667757, 2667758, 2667760, 2667761, 2667762, 2667764, 2667765, 3719398, 3719399, 3719402, 3719404, 3719405, 3791813, 3791814, 3791815, 3791817, 3791818, 3791819, 3791820, 3791822, 3791823, 3791827, 3791828, 3791830, 3791832, 4161411, 4161413, 4161417, 4161418, 4161419, 4161421, 4161425, 4161427, 4161428, 4161429, 4161432, 4161433, 4161434, 4161435, 4161436, 4161440, 4161442, 4209955, 4209956, 4262787, 4262788, 4262795, 4262796, 4262798, 4262801, 4262803, 4262804, 4262807, 4262813, 4325367, 4325368, 4325374, 4325375, 4325376, 4325383, 4325384, 4325385, 4325387, 4325389, 4325391, 4325393, 4325394, 4325395, 4383672, 4383682, 4383685, 4383687, 4383692, 4383695, 4383696, 4383698, 4383701, 4383702, 4383705, 4383710, 4383714, 4383716, 4383717, 4383727, 4383729, 5139124, 5139126, 5139129, 5139130, 5139132, 5139133, 5139134, 5139136, 5168293, 5168294, 5284184, 5284186, 5284188, 5284196;
j. REF 132-40-55, GUDID 10885862207067, Serial Numbers: 2406944, 2406954, 2406960, 2408204, 2408207, 2408211, 2408213, 2408215, 2408217, 2408218, 2408226, 2408228, 2443151, 2598685, 2673737, 2673743, 2673746, 2673748, 3792277, 4209974, 4209982, 4235095, 4235101, 4235103, 4235104, 4235106, 4260071, 5150307, 5150310, 5272974;
k. REF 136-40-53, GUDID 10885862220301, Serial Numbers: 2601258, 2601259, 2601260, 2601262, 2601263, 2601268, 2601269, 2601270, 2601271, 2601272, 2601273, 2601274, 2601275, 2601276, 2601277, 2601278, 2601280, 2601281, 2601282, 2601283, 2601284, 2601285, 2601286, 2601288, 2601289, 2687676, 2687677, 2687679, 2687680, 2687682, 2687683, 2687684, 2687685, 2687686, 2687687, 2687688, 2687690, 2687691, 2687692, 2687693, 2687696, 2687697, 2687698, 2687700, 2687701, 3751038, 3751039, 3751040, 3751042, 3751044, 3751046, 3751047, 3751049, 3751053, 3751056, 3751057, 3751060, 3751062, 3751063, 3751064, 3751065, 3751066, 3751067.3751068, 4713723, 4713724, 4713727, 4713728, 4713731, 4725388, 4725390, 4725391, 4725392, 4764116, 4781029, 4781030, 4781031, 4781032, 4781033, 4781034, 4781035, 4817935, 4817940, 4946298, 4946299, 4946301, 4946303, 4946304, 5139738, 5139739, 5139740, 5139741, 5139743, 5139744, 5139745, 5139746, 5139747, 5139748, 5159613, 5159614, 5175504, 5213720, 5213723, 5305157, 5305158, 5305159, 5305160, 5305161, 5365956, 5365957, 5365958, 5365959, 5365960, 5365961, 5365962, 5365963, 5365964, 5365966, 5365967, 5365968, 5603823, 5603825, 5603826, 5603827, 5603828, 5603830, 6109679, 6109696;
l. REF 136-40-54, GUDID 10885862220318, Serial Numbers: 2598603, 2598604, 2598605, 2598606, 2598607, 2598609, 2598611, 2598613, 2598614, 2598616, 2598617, 2598618, 2598621, 2598622, 2598623, 2598624, 2598630, 2697701, 2697706, 2697708, 3650027, 3727302, 3727304, 3727306, 3751075, 3751076, 3751080, 3751083, 3751086, 3751088, 3751091, 3751093, 3751094, 3751102, 3751110, 4837163, 5139777, 5139782, 5151331, 5316514, 5316515;
m. REF 136-40-55, GUDID 10885862220325, Serial Numbers: 2594398, 2594405, 2594410, 2594414, 2594421, 2594425;
n. REF 138-40-53, GUDID 10885862208804, Serial Numbers: 2594428, 2594429, 2594431, 2594433, 2594435, 2594436, 2594437, 2594438, 2594440, 2594442, 2594443, 2594445, 2594446, 2594449, 2594450, 2594451, 2594453, 2594454, 2594455, 2594456, 3613070, 3613072, 3650014, 3762124, 3762125, 3762126, 3762127, 3762130, 3762131, 3762132, 3762133, 4127111, 4127112, 4127113, 4127114, 4127115, 4127116, 4127118, 4127120, 4127121, 4127122, 4127123, 4127124, 4127125, 4127128, 4127129, 4127130, 4760640, 4760641, 4760642, 4760643, 4760648, 4760650, 4760652, 4817479, 4817480, 4817481, 4817482, 4817483, 4916831, 4916832, 4916833, 4916834, 4916836, 4974530, 4974531, 4974533, 4974534, 4974535, 4974536, 5005479, 5005480, 5139749, 5139750, 5139751, 5139754, 5139755, 5139756, 5139757, 5139760, 5139762, 5139764, 5139765, 5139766, 5139767, 5174564, 5174565, 5174567, 5174568, 5245408, 5245409, 5245410, 5245412, 5245414, 5245415, 5245416, 5245418, 5323803;
o. REF 138-40-54, GUDID 10885862208811, Serial Numbers: 2598544, 2598552, 2598554, 2598559, 2598563, 2750917, 3735555, 3750994, 5334791;
p. REF 138-40-55, GUDID 10885862208828, Serial Numbers: 2598583, 2598601, 5139018.
|
| FEI Number |
1038671
|
Recalling Firm/ Manufacturer |
Exactech, Inc.
2320 Nw 66th Ct
Gainesville
FL
32653-1630
|
| For Additional Information Contact | 888-912-0403 |
Manufacturer Reason for Recall | Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear. |
FDA Determined Cause 2 | Process control |
| Action | The firm sent an "URGENT MEDICAL DEVICE RECALL" DEAR HEALTHCARE PROFESSIONAL COMMUNICATION on 08/11/2022 by email. The purpose of the current letter is to update surgeons regarding the implications of the packaging non-conformity on performance of the Connexion GXL and other conventional UHMWPE acetabular liners. In addition, this letter provides updates to patient management and follow-up recommendations since issuance of the July 2021 DHCP letter. In addition to providing surgeons with lists of all their affected GXL and/or conventional polyethylene patients since 2004, Exactech is providing surgeons with two draft letters directed at patients who have been implanted with Exactech GXL hip devices.
The Customers were instructed to take the following actions:
CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required.
IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customers inventory.
EXTEND THE DESCRIPTION OF THE ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession.
COMPLETE AND RETURN the attached Recall Inventory Response Form to
Exactech via email at recalls@exac.com.
Reach out to Customer Service at cust.svc@exac.com to request a Returned Goods
(RG) Number for ALL affected product requested to be returned.
WE ARE REQUIRING 100% EFFECTIVENESS FOR THIS RECALL.
If you have any questions, call (888) 912-0403. |
| Quantity in Commerce | 17929 units |
| Distribution | Worldwide distribution: US (nationwide) and OUS (foreign) countries of: ARGENTINA, BRAZIL, CHILE, CHINA, COLOMBIA, ECUADOR, GREECE, GUATEMALA, INDIA, LEBANON
PORTUGAL,PUERTO RICO,SPAIN
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LZO
|
|
|
|