| | Class 1 Device Recall GE |  |
| Date Initiated by Firm | December 19, 2022 |
| Date Posted | February 08, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1058-2023 |
| Recall Event ID |
91456 |
| 510(K)Number | K173816 |
| Product Classification |
System, tomography, computed, emission - Product Code KPS
|
| Product | GE Nuclear Medicine 600/800 series systems, specifically: a) 850,
b) 850 / Dismount,
c) 850 3/8 inch NM Gantry,
d) 860 / Dismount,
e) 860 3/8 inch NM Gantry,
f) 860 5/8 inch NM Gantry,
g) 860 BJ,
h) NM 830 3/8 inch BJ,
i) NM/CT 850 3/8 inch BJ,
j) NM/CT 850 3/8 inch ROW,
k) NM/CT 850 3/8 inch ROW_,
l) NM/CT 850 CT 850 3/8 inch,
m) NM/CT 850 ES 3/8 inch,
n) NM/CT 850 ES 3/8 inch ROW,
o) NM/CT 860 3/8 inch,
p) NM/CT 860 3/8 inch BJ,
q) NM/CT 860 3/8 inch ROW,
r) NM/CT 860 ADVANCE,
s) NM/CT 860 advance 3/8 inch NM Gantry,
t) NM/CT 860 BEYOND,
u) NM/CT 860 beyond 3/8 inch NM Gantry,
v) NM/CT 860 China for China,
w) NM/CT 860 CT 860 3/8 inch,
x) NM/CT 860 ES 3/8 inch,
y) NM/CT 860 ES 3/8 inch ROW,
z) NM/CT 850 3/8 inch,
aa) NM/CT 850 ES 3/8 inch ROW,
bb) NM/CT 860 3/8 inch CH,
cc) XELERIS 4 DR WORKSTATION;
Emission Computed Tomography System |
| Code Information |
a) 850: Catalog number 5376204-50-1, UDI/DI Not Applicable, Serial Numbers: NT3BG2200006BH;
b) 850 / Dismount: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3BG2100003BH;
c) 850 3/8 inch NM Gantry: Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 85EY67037, 850Z61033, 85EY67039, 85EY67048, 850Y61018, 850Y61009, 85EY67035, 85EX67001, 85EX67020, 85EY67045, 85EX67023, 85EY67032, 85EY67046, 850A61063, 850Z61047, 85EY67047, 850Z61053, 850Y61006, 850Y61010, 850X61005, 850Y61021, 85EZ67062, 85EX67019, 85EX67021, 85EA67063, 850A61069, 85EX67008, 85EY67040, 850Z61051, 850Z61038, 850A61075, 850Y61025, 850Z61030, 85EY67015, 85EX67017, 85EY67043, 850Z61042, 850Z61061, 85EY67026, 850A61065, 85EX67012, 850Y61023, 85EY67041, 850A61062, 850A61076, 850Y61012, 850A61064, 85EZ67058, 85EX67006, 850Z61043, 850Z61055, 85EY67042, 85EZ67055, 85EY67036, 85EX67024, 85EY67038, 85EX67022, 85EY67025, 85EZ67060, 850Z61058, 850Z61044, 850Z61045, 850Y61020, 850Z61032, 85EZ67056, 85EA67065, 85EY67028, 850Z61041, 85EX67002, 85EX67009, 850A61072, 850Z61036, 850A61068, 850A61071, 850Z61060, 850Z61034, 850Z61039, 850Z61035, 850Z61040, 85EY67031, 850Z61056, 850Z61057, 850Y61016, 85EX67016, 85EA67066, 85EX67004, 850Z61029, 850Z61031, 85EX67011, 85EX67010, 850Y61013, 850Y61015, 850Y61017, 850A61070, 85EX67005, 850A61073, 85EX67018, 85EY67033, 850Z61027, 85EX67013, 850Z61054, 850X61001, 85EZ67053, 85EZ67054, 85EZ67057, 850X61004, 850Z61059, 850Z61052, 850Y61014, 850Z61048, 850Y61026, 850Y61019, 85EY67044, 850Z61037, 850X61002, 85EX67003, 85EX67014, 850Y61007, 85EY67049, 85EZ67052, 850A61067, 850A61074, 85EY67050, 85EZ67051, 85EY67027;
UDI/DI (01)00840682140775(21)850B61120(11)220700, Serial Numbers: 850B61120; UDI/DI 010084068214077521850Y6101111190300, Serial Numbers: 850Y61011; UDI/DI To be provided, Serial Numbers: 850A61083, 85EA67080, 850B61127, 850B61128, 850B61129, 850B61130, 85EB67087, 850B61133, 850B61134, 850B61135, 850B61101, 85EA67070, 850B61119, 850B61102, 850B61115, 85EA67076, CHBB65669, 850B61122, 850B61132, 850B61131, 85EA67075, 85EA67077, 85EB67088, 850A61093, 85EB67086, 850B61126, 850A61089, 85EA67078, 850B61112, 850B61111, 850B61110, 850B61107, 850B61109; Catalog number To be provided, UDI/DI To be provided, Serial Numbers: 60B62111;
d) 860 / Dismount: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3CG2100006BH;
e) 860 3/8 inch NM Gantry: Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 860Y62009, 860Y62010, 860Z62027, 860Y62005, 860A62049, 860Y62019, 860Y62014, 860Y62025, 860Z62028, 860Y62024, 860Z62038, 860A62063, 860Z62042, 860Y62017, 860A62065, 860A62058, 860Z62046, 860A62057, 860Z62041, 86EA76004, 860B62118, 860Z62043, 860A62060, 860Z62026, 860Y62006, 860Y62015, 860Y62004, 860Z62047, 860Y62020, 860A62050, 860Z62030, 860Y62007, 860X62002, 860Z62032, 860Y62021, 86EA76003, 860Z62040, 860Z62037, 860Z62031, 860A62062; , UDI/DI 010084068214075121860A6207711211000, Serial Numbers: 860A62077; UDI/DI (01)00840682140751(21)860B62097(11)220300, Serial Numbers: 860B62097; UDI/DI 010084068214075121860B6209611220300, Serial Numbers: 860B62096; UDI/DI 010084068214075121860A6205911210400, Serial Numbers: 860A62059; UDI/DI 010084068214075121860A6206811210600, Serial Numbers: 860A62068; UDI/DI 010084068214075121860Y6202311191100, Serial Numbers: 860Y62023; UDI/DI 010084068214075121860Y6201311190600, Serial Numbers: 860Y62013; UDI/DI To be provided, Serial Numbers: 860A62067, 860Y62016, 860B62117, 860B62100, 860B62105, 860B62090, 860A62072, 860B62108, 860B62110, 860A62078, 860A62083, 860B62094, 860A62066, 860A62084, 860A62071, 860A62070, 860A62073, 860A62069, 860B62095, 860A62086, 860A62081, 860A62085, 860B62099, 860B62115, 860A62082, 860B62114, 860A62074, 860Z62034, 86EB76006, 860B62122, 860B62119, 860B62121, 860B62089, 86EB76007, 860A62076, 860A62079, 860B62116, 86EB76009, 860B62120, 860B62093, 860A62080, 860B62098, 860B62091, 860B62103;
f) 860 5/8 inch NM Gantry: Catalog number H3908AE, UDI/DI To be provided, Serial Numbers: 860X62001; , UDI/DI Not Applicable, Serial Numbers: 860Z62035;
g) 860 BJ: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3CG1900007BH, NT3CG1900008BH, NT3CG1900009BH;
h) NM 830 3/8 inch BJ: Catalog number 5376204-50-1, UDI/DI To be provided, Serial Numbers: NT3AG1900007BH, NT3AG2100001BH, NT3AG2100006BH;
i) NM/CT 850 3/8 inch BJ: Catalog number 5376204-50-1, UDI/DI Not Applicable, Serial Numbers: NT3BG2000003BH, NT3BG2100001BH, NT3BG2200001BH, NT3BG1900001BH, NT3BG2200002BH, NT3BG2000004BH, NT3BG2100004BH, NT3BG2200005BH; : Catalog number H2407AA, UDI/DI Not Applicable, Serial Numbers: NT3BG2000001BH; , UDI/DI To be provided, Serial Numbers: NT3BG2100002BH; : Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3BG2000002BH;
j) NM/CT 850 3/8 inch ROW: Catalog number 5376204-50-1, UDI/DI Not Applicable, Serial Numbers: NT3BG2100005BH, NT3BG1900002BH; Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 850Y61022, 850Z61046, 850Y61008; UDI/DI To be provided, Serial Numbers: 850B61100;
k) NM/CT 850 3/8 inch ROW: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 860Y62008;
l) NM/CT 850 CT 850 3/8 inch: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 850A61092, 850B61108, 850B61096, 850A61080, 850Z61050, 850B61106, 850B61097, 850A61077, 850B61118, 860B62112; Catalog number 5376204-50, UDI/DI 010084068214077521850A6109511211200, Serial Numbers: 850A61095; UDI/DI 010084068214077521850A6108811210900, Serial Numbers: 850A61088; UDI/DI 010084068214077521850B6112111220700, Serial Numbers: 850B61121; UDI/DI 010084068214077521850B6110411220400, Serial Numbers: 850B61104; UDI/DI 010084068214077521850A6108711210900, Serial Numbers: 850A61087; UDI/DI 010084068214077521850A6108211210800, Serial Numbers: 850A61082; UDI/DI 010084068214077521850B6110511220400, Serial Numbers: 850B61105; Catalog number H3907AA, UDI/DI To be provided, Serial Numbers: 850A61081; Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 850Z61049, 850A61066; UDI/DI To be provided, Serial Numbers: 850B61098, 850B61099, 850A61084, 850B61123, 850B61113, 850B61117, 850B61116, 850A61090, 850A61079, 850A61091, 850B61114, 850A61078, 850A61086; : Catalog number To be provided, UDI/DI To be provided, Serial Numbers: 850B61125, 850B61103;
m) NM/CT 850 ES 3/8 inch: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 85EA67068; Catalog number 5376204-50, UDI/DI 01008406821407752185EA6707411210800, Serial Numbers: 85EA67074, 85EA67067, 85EA67073; Catalog number 5376204-50-1, UDI/DI To be provided, Serial Numbers: NT3BG1900003BH; Catalog number H3907AA, UDI/DI To be provided, Serial Numbers: 85EA67071; Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 85EX67007; UDI/DI To be provided, Serial Numbers: 85EA67069, 85EA67072, 85EA67079, 85EB67084;
n) NM/CT 850 ES 3/8 inch ROW: Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 85EZ67059, 85EA67064, 85EY67034;
o) NM/CT 860 3/8 inch: Catalog number H3908AD, UDI/DI To be provided, Serial Numbers: 00000860Z62033;
p) NM/CT 860 3/8 inch BJ: Catalog number H2408AA, UDI/DI To be provided, Serial Numbers: NT3CG2000001BH; Catalog number 5376204-60-1, UDI/DI Not Applicable, Serial Numbers: NT3CG2100001BH, NT3CG1900010BH, NT3CG2100003BH, NT3CG1900005BH, NT3CG1900003BH, NT3CG2100004BH, NT3CG1900006BH, NT3CG2100005BH, NT3CG2000004BH, NT3CG1900002BH, NT3CG2100002BH;
q) NM/CT 860 3/8 inch ROW: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 00000860Y62018; Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 860A62056, 860Y62011, 860A62048, 860Z62039, 860Z62044, 860A62052, 860A62055, 860Y62012, 860Y62022, 860A62054; Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3CG2000003BH; UDI/DI To be provided, Serial Numbers: 860B62102;
r) NM/CT 860 ADVANCE: Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 860Z62045, 860A62064, 860A62061; UDI/DI To be provided, Serial Numbers: 860A62075; Catalog number H3908BB , UDI/DI Not Applicable, Serial Numbers: 8BBB30002, 8BBB30001;
s) NM/CT 860 advance 3/8 inch NM Gantry: Catalog number H3908BB, UDI/DI To be provided, Serial Numbers: 8BBB30003, 8BBB30004;
t) NM/CT 860 BEYOND: Catalog number H3908BC, UDI/DI Not Applicable, Serial Numbers: 8BCB40001;
u) NM/CT 860 beyond 3/8 inch NM Gantry: Catalog number H3908BC, UDI/DI To be provided, Serial Numbers: 8BCB40002;
v) NM/CT 860 China for China: Catalog number 5376204-60-1, UDI/DI Not Applicable, Serial Numbers: NT3CG2200004BH, NT3CG2200003BH, NT3CG2200002BH; , UDI/DI To be provided, Serial Numbers: NT3CG2200001BH;
w) NM/CT 860 CT 860 3/8 inch: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 860B62109, 860Z62029, 860B62101, 860B62113; Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 860Z62036, 86EZ76001, 860A62053; UDI/DI To be provided, Serial Numbers: 860B62092, 860A62088;
x) NM/CT 860 ES 3/8 inch: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 86EB76005;
y) NM/CT 860 ES 3/8 inch ROW: Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 86EA76002;
z) NM/CT 850 3/8 inch: Catalog number H3907AA, UDI/DI Not Applicable, Serial Numbers: 850Y61024;
aa) NM/CT 850 ES 3/8 inch ROW: Catalog number H3907AH, UDI/DI Not Applicable, Serial Numbers: 85EY67030;
bb) NM/CT 860 3/8 inch CH: Catalog number H2408AA, UDI/DI Not Applicable, Serial Numbers: NT3DG1900007BH;
cc) XELERIS 4 DR WORKSTATION: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 850B61124, 850A61094
|
| FEI Number |
3004571294
|
Recalling Firm/ Manufacturer |
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING 4, Hayozma St Tirat Carmel Israel
|
| For Additional Information Contact | GE HealthCare Service 800-4371171 |
Manufacturer Reason for Recall | A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury. |
FDA Determined Cause 2 | Process control |
| Action | GE Healthcare issue an URGENT MEDICAL DEVICE CORRECTION on 12/19/2022 by mail. The notice explained the issue and the risk, and requested the following action by taken:
.GE Healthcare asks that you STOP USE of your Nuclear Medicine system, if your
system is listed in the Affected Product Details and Appendix of this letter, until an
inspection of your system is completed by GE Healthcare and you have been
provided with a letter indicating the system can be returned to clinical use.
.GE Healthcare will correct all affected products at no cost to you. GE Healthcare
will inspect your system to ensure that the appropriate mitigation is correctly
implemented to address this potential life-threatening concern. A GE Healthcare
representative will contact you to arrange for this correction.
.If you have any questions or concerns regarding this notification, please contact
GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
|
| Quantity in Commerce | 380 units |
| Distribution | Worldwide distribution. US nationwide including Puerto Rico, Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Ethiopia, Finland, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uzbekistan, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KPS
|
|
|
|