| | Class 2 Device Recall NextSeq 550Dx Instrument |  |
| Date Initiated by Firm | April 05, 2023 |
| Date Posted | June 16, 2023 |
| Recall Status1 |
Completed |
| Recall Number | Z-1976-2023 |
| Recall Event ID |
92233 |
| Product Classification |
High throughput DNA sequence analyzer - Product Code PFF
|
| Product | NextSeq 550Dx Instrument, Catalog # 20005715, contaning Universal Copy Service software |
| Code Information |
Impacted UCS Versions: 1.4.2.927, 1.6.3.1423, 2.7.2.4320/ UDI-DI: (00)816270020125 |
| FEI Number |
3003218906
|
Recalling Firm/ Manufacturer |
Illumina, Inc. 5200 Illumina Way San Diego CA 92122-4616
|
| For Additional Information Contact | David McAlpine 347-327-1336 |
Manufacturer Reason for Recall | Cybersecurity vulnerability concerning the software used for sequencing instruments. |
FDA Determined Cause 2 | Software Manufacturing/Software Deployment |
| Action | On April 5, 2023, Illumina issued a "Urgent Medical Device Recall" Notification via XXX. Illumina ask consignees to take the following actions:
1. Download & Install Software Patch: For Each Instrument Connected to the Internet:
2. Change UCS User Configuration(s): Change the UCS User configuration to standard user
(with no administrator permissions) by following the instructions available in the customer communication.
3. Complete and return the Verification Form after carrying out all the steps in the instructions provided on your specific instrument(s) identified as affected in Table 1.
4. If you experience an adverse event due to this vulnerability with the use of any of the affected products, please report it to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. You can complete and submit the report online at
www.fda.gov/medwatch/report.htm.
|
| Quantity in Commerce | 1067 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Algeria, AU Australia, Austria, Belgium, Bulgaria, Canada, Chile, China, Cyprus, Czech Republic, Denmark, Egypt, FR France, GB Georgia, Germany, Guatemala, Hong Kong, Hungary, IT Italy, Japan, Latvia, Lithuania, Luxembourg, Macedonia, Malta, Morocco, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Qatar, Russian Fed., Saudi Arabia, SG Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Utd. Arab Emir., Vietnam, Argentina, Aerbaijan, Belgium, Bosnia-Herz., Finland, Greece, Iceland, Ireland, Israel, Japan, Nigeria, NL Norway, Romania, Serbia, Slovakia, South Africa, T¿rkiye, Ukraine. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
|
|
|