• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1821-2023
Recall Event ID 92328
PMA NumberP980016S713 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
ProductICD COBALT VR MRI DF4, Model Number DVPB3D4; Implantable Cardioverter Defibrillators
Code Information GTIN 00763000178512, Lot Serial Numbers: RSF600012S, RSF600254S, RSF600591S, RSF600655S, RSF600822S, RSF601247S, RSF602653S, RSF603908S, RSF601102S, RSF607914S, RSF605967S, RSF600633S, RSF600649S, RSF600867S, RSF600868S, RSF600869S, RSF600872S, RSF600873S, RSF600874S, RSF600875S, RSF600876S, RSF600878S, RSF600880S, RSF600883S, RSF602930S, RSF602991S, RSF602998S, RSF603002S, RSF603003S, RSF603013S, RSF603018S, RSF603019S, RSF603022S, RSF603026S, RSF603027S, RSF603030S, RSF603032S, RSF603033S, RSF603034S, RSF603035S, RSF603036S, RSF603736S, RSF603737S, RSF603741S, RSF603742S, RSF603744S, RSF603746S, RSF603747S, RSF603749S, RSF603754S, RSF603755S, RSF603756S, RSF603758S, RSF603759S, RSF603762S, RSF603763S, RSF603764S, RSF603799S, RSF603802S, RSF603804S, RSF603806S, RSF603810S, RSF603813S, RSF603818S, RSF603821S, RSF603824S, RSF602665S, RSF602703S, RSF603249S, RSF603648S, RSF603939S, RSF607644S, RSF601171S, RSF601174S, RSF602584S, RSF602743S, RSF600887S, RSF602678S, RSF608130S, RSF600767S, RSF608097S, RSF604855S, RSF606263S, RSF601206S, RSF601486S, RSF601567S, RSF601571S, RSF600924S, RSF601224S, RSF601226S, RSF602817S, RSF603405S, RSF603521S, RSF603678S, RSF602595S, RSF604129S, RSF600687S, RSF600210S, RSF602849S, RSF607243S, RSF607244S, RSF607291S, RSF604110S, RSF604146S, RSF605150S, RSF607995S, RSF600352S, RSF601546S, RSF607225S, RSF606605S, RSF600269S, RSF600940S, RSF601190S, RSF603550S, RSF604130S, RSF608014S, RSF600303S, RSF602434S, RSF602435S, RSF602449S, RSF602461S, RSF602466S, RSF607173S, RSF607174S, RSF600954S, RSF600955S, RSF602586S, RSF602587S, RSF602597S, RSF602804S, RSF603465S, RSF600982S, RSF603213S, RSF605644S, RSF606195S, RSF601058S, RSF603647S, RSF603724S, RSF604582S, RSF600219S, RSF601397S, RSF601539S, RSF602709S, RSF602757S, RSF603794S, RSF603973S, RSF604054S, RSF600443S, RSF600706S, RSF600721S, RSF600729S, RSF601434S, RSF603726S, RSF606532S, RSF603464S, RSF608114S, RSF601093S, RSF603450S, RSF607337S, RSF607359S, RSF600793S, RSF600805S, RSF600808S, RSF600864S, RSF601132S, RSF601165S, RSF601176S, RSF601540S, RSF601547S, RSF601587S, RSF602604S, RSF602874S, RSF603248S, RSF603399S, RSF603415S, RSF603428S, RSF603909S, RSF605018S, RSF605037S, RSF605025S, RSF601119S, RSF601462S, RSF607023S, RSF605918S, RSF606278S, RSF606884S, RSF600539S, RSF600523S, RSF602564S, RSF602565S, RSF602570S, RSF602571S, RSF603242S, RSF606921S, RSF601482S, RSF602165S, RSF602753S, RSF602822S, RSF602877S, RSF603585S, RSF605489S, RSF605974S, RSF607360S, RSF602220S, RSF602786S, RSF602789S, RSF604164S, RSF606324S, RSF606934S, RSF601167S, RSF601169S, RSF601172S, RSF602790S, RSF603216S, RSF603470S, RSF603549S, RSF604227S, RSF605701S, RSF605928S, RSF607222S, RSF607740S, RSF607741S, RSF607911S, RSF600260S, RSF601106S, RSF602684S, RSF603240S, RSF603468S, RSF603685S, RSF604089S, RSF604112S, RSF604121S, RSF604263S, RSF604389S, RSF604470S, RSF604700S, RSF604988S, RSF605889S, RSF600835S, RSF603897S, RSF603963S, RSF603966S, RSF605021S, RSF600263S, RSF600534S, RSF601142S, RSF601572S, RSF602208S, RSF602373S, RSF602632S, RSF602706S, RSF602848S, RSF603416S, RSF603527S, RSF603528S, RSF603662S, RSF603683S, RSF603892S, RSF604013S, RSF605570S, RSF607887S, RSF601305S, RSF600610S, RSF603260S, RSF603490S, RSF601446S, RSF605645S, RSF601534S, RSF602463S, RSF602698S, RSF603191S, RSF603510S, RSF603689S, RSF603916S, RSF603986S, RSF604266S, RSF605205S, RSF605206S, RSF605505S, RSF605670S, RSF600312S, RSF600889S, RSF604284S, RSF604471S, RSF604502S, RSF605697S, RSF605876S, RSF606313S, RSF606942S, RSF607052S, RSF607058S, RSF607714S, RSF607720S, RSF607721S, RSF603734S, RSF603870S, RSF603911S, RSF604010S, RSF604156S, RSF604289S, RSF604302S, RSF604425S, RSF604748S, RSF605054S, RSF607050S, RSF607733S, RSF607739S, RSF607763S, RSF608003S, RSF603960S, RSF605443S, RSF603543S, RSF607240S, RSF607378S, RSF607397S, RSF607406S, RSF600153S, RSF600328S, RSF600705S, RSF602707S, RSF603595S, RSF603773S, RSF600977S, RSF601471S, RSF601588S, RSF602343S, RSF602889S, RSF600162S, RSF600557S, RSF605875S, RSF606245S, RSF607055S, RSF607056S, RSF607716S, RSF601028S, RSF601033S, RSF601091S, RSF601189S, RSF601384S, RSF604197S, RSF604660S, RSF604701S, RSF604751S, RSF604905S, RSF607935S, RSF600717S, RSF603024S, RSF608070S, RSF603406S, RSF604073S, RSF600315S, RSF600420S, RSF606967S, RSF606367S, RSF606885S, RSF607025S, RSF607066S, RSF608102S, RSF608104S, RSF605699S, RSF605909S, RSF606985S, RSF607380S, RSF607697S, RSF607900S, RSF608009S, RSF600414S, RSF600671S, RSF601013S, RSF602591S, RSF602702S, RSF603130S, RSF603261S, RSF603449S, RSF603452S, RSF603492S, RSF603695S, RSF603823S, RSF603957S, RSF604087S, RSF604232S, RSF604268S, RSF604387S, RSF605559S, RSF606193S, RSF606194S, RSF606309S, RSF606534S, RSF600508S, RSF604838S, RSF606282S, RSF606513S, RSF603775S, RSF606311S, RSF607877S, RSF600708S, RSF600894S, RSF602727S, RSF602831S, RSF604604S, RSF606233S, RSF606894S, RSF607205S, RSF600815S, RSF604219S, RSF604448S, RSF607869S, RSF605419S, RSF607870S, RSF607030S, RSF607035S, RSF607934S, RSF605610S, RSF605964S, RSF606907S, RSF607073S, RSF607916S, RSF600472S, RSF600475S, RSF600696S, RSF600783S, RSF600802S, RSF600847S, RSF600911S, RSF601053S, RSF601192S, RSF601531S, RSF601560S, RSF602607S, RSF603100S, RSF603456S, RSF607366S, RSF602326S, RSF608093S, RSF603933S, RSF600241S, RSF605661S, RSF605908S, RSF600503S, RSF600886S, RSF601086S, RSF601214S, RSF601260S, RSF601279S, RSF601394S, RSF602547S, RSF602559S, RSF602654S, RSF602795S, RSF603785S, RSF601573S, RSF600801S, RSF601345S, RSF601467S, RSF603596S, RSF603649S, RSF604082S, RSF600701S, RSF605865S, RSF605866S, RSF605869S, RSF606217S, RSF606950S, RSF607044S, RSF607062S, RSF607067S, RSF601395S, RSF601469S, RSF601586S, RSF602589S, RSF602683S, RSF603214S, RSF603858S, RSF607904S, RSF606523S, RSF607928S, RSF600195S, RSF601250S, RSF601475S, RSF602578S, RSF603553S, RSF603559S, RSF603729S, RSF603832S, RSF603837S, RSF603875S, RSF603902S, RSF603982S, RSF604598S, RSF605436S, RSF600433S, RSF600615S, RSF600674S, RSF600762S, RSF600785S, RSF602744S, RSF602892S, RSF603713S, RSF603727S, RSF604237S, RSF605856S, RSF600566S, RSF601215S, RSF603602S, RSF603946S, RSF605030S, RSF603042S, RSF603587S, RSF601323S, RSF607949S, RSF607950S, RSF600552S, RSF600623S, RSF600711S, RSF600782S, RSF601088S, RSF602504S, RSF602770S, RSF605695S, RSF606528S, RSF603655S, RSF602205S, RSF600736S, RSF600865S, RSF600866S, RSF601003S, RSF601098S, RSF602177S, RSF602895S, RSF603227S, RSF603511S, RSF603544S, RSF604135S, RSF604823S, RSF605990S, RSF601100S, RSF601147S, RSF601185S, RSF601353S, RSF601354S, RSF602888S, RSF603489S, RSF604401S, RSF604968S, RSF605991S, RSF606221S, RSF607867S, RSF600360S, RSF601533S, RSF600561S, RSF600676S, RSF601549S, RSF603999S, RSF602623S, RSF600410S, RSF600937S, RSF602816S, RSF602827S, RSF603378S, RSF605945S, RSF608117S, RSF600282S, RSF600324S, RSF604759S, RSF600751S, RSF604115S, RSF607662S, RSF602592S, RSF602983S, RSF602985S, RSF603497S, RSF603519S, RSF603958S, RSF603987S, RSF603994S, RSF604186S, RSF604195S, RSF606189S, RSF606498S, RSF606535S, RSF606947S, RSF607330S, RSF602905S, RSF603104S, RSF604977S, RSF605086S, RSF607890S, RSF608062S, RSF602626S, RSF604260S, RSF602713S, RSF608041S, RSF608042S, RSF608043S, RSF608044S, RSF608045S, RSF605968S, RSF600749S, RSF600985S, RSF601212S, RSF603199S, RSF603402S, RSF603651S, RSF603652S, RSF603815S, RSF603817S, RSF603886S, RSF604217S, RSF605708S, RSF606256S, RSF600476S, RSF600477S, RSF601331S, RSF602334S, RSF602359S, RSF602629S, RSF602777S, RSF603221S, RSF603600S, RSF603735S, RSF603829S, RSF604413S, RSF604906S, RSF607227S, RSF607386S, RSF600492S, RSF600806S, RSF600975S, RSF601021S, RSF601025S, RSF601031S, RSF601265S, RSF601570S, RSF602494S, RSF602921S, RSF604154S, RSF604730S, RSF604883S, RSF604886S, RSF600154S, RSF600267S, RSF600843S, RSF601555S, RSF601574S, RSF602637S, RSF602700S, RSF602711S, RSF603054S, RSF603201S, RSF603369S, RSF603789S, RSF603918S, RSF603932S, RSF603955S, RSF603991S, RSF604098S, RSF604188S, RSF604766S, RSF600666S, RSF600949S, RSF602672S, RSF602988S, RSF603864S, RSF602619S, RSF607247S, RSF601449S, RSF603223S, RSF602423S, RSF606943S, RSF600395S, RSF600427S, RSF600504S, RSF601231S, RSF601278S, RSF601542S, RSF602802S, RSF603945S, RSF600309S, RSF600314S, RSF600560S, RSF600747S, RSF602327S, RSF602329S, RSF602411S, RSF605810S, RSF605849S, RSF605851S, RSF605852S, RSF605853S, RSF606197S, RSF606198S, RSF606200S, RSF606201S, RSF607098S, RSF607099S, RSF607101S, RSF607114S, RSF607116S, RSF607951S, RSF607952S, RSF601109S, RSF602897S, RSF603060S, RSF603062S, RSF603063S, RSF603065S, RSF603066S, RSF603069S, RSF603070S, RSF603071S, RSF603107S, RSF603639S, RSF603676S, RSF603801S, RSF603803S, RSF603807S, RSF603808S, RSF603809S, RSF603811S, RSF603812S, RSF605366S, RSF607097S, RSF607392S, RSF607393S, RSF607953S, RSF607954S, RSF607955S, RSF607956S, RSF607957S, RSF607958S, RSF607959S, RSF607960S, RSF607961S, RSF607962S, RSF607963S, RSF607964S, RSF605650S, RSF600398S, RSF600399S, RSF600401S, RSF600407S, RSF600743S, RSF600836S, RSF600838S, RSF600839S, RSF601261S, RSF601306S, RSF601314S, RSF601532S, RSF603106S, RSF603800S, RSF605678S, RSF607074S, RSF602425S, RSF602741S, RSF605956S, RSF603576S, RSF600511S, RSF600512S, RSF602372S, RSF602751S, RSF602752S, RSF602808S, RSF602932S, RSF602969S, RSF602974S, RSF603132S, RSF603362S, RSF603538S, RSF603697S, RSF603876S, RSF604012S, RSF604020S, RSF604173S, RSF604205S, RSF604344S, RSF604481S, RSF604482S, RSF604485S, RSF604500S, RSF602512S, RSF601491S, RSF606519S, RSF607271S, RSF602427S, RSF602859S, RSF602958S, RSF603675S, RSF600274S, RSF602553S, RSF607212S, RSF603430S, RSF603632S, RSF603653S, RSF608036S, RSF600142S, RSF600670S, RSF603625S, RSF601240S, RSF600532S, RSF600739S, RSF600741S, RSF604032S, RSF600967S, RSF605007S, RSF605008S, RSF606286S, RSF607865S, RSF608022S, RSF600209S, RSF600346S, RSF601075S, RSF601096S, RSF602605S, RSF602628S, RSF602780S, RSF603266S, RSF603522S, RSF603529S, RSF603491S, RSF607387S, RSF600668S, RSF602575S, RSF603512S, RSF607942S, RSF607943S, RSF607944S, RSF607945S, RSF607946S, RSF605684S, RSF605950S, RSF605962S, RSF606173S, RSF606228S, RSF606229S, RSF606230S, RSF606264S, RSF606265S, RSF606503S, RSF606504S, RSF606521S, RSF606963S, RSF607077S, RSF607266S, RSF607283S, RSF607377S, RSF600466S, RSF600605S, RSF600812S, RSF600951S, RSF601232S, RSF601235S, RSF601238S, RSF601242S, RSF601246S, RSF601262S, RSF601559S, RSF602462S, RSF602522S, RSF602524S, RSF602526S, RSF602527S, RSF602745S, RSF602748S, RSF602749S, RSF602809S, RSF603349S, RSF603494S, RSF603619S, RSF603738S, RSF603879S, RSF603935S, RSF603938S, RSF603971S, RSF603976S, RSF604055S, RSF604153S, RSF604170S, RSF604181S, RSF604255S, RSF604440S, RSF604488S, RSF604496S, RSF604668S, RSF604669S, RSF604670S, RSF604723S, RSF604821S, RSF604826S, RSF604904S, RSF605020S, RSF605976S, RSF600413S, RSF602638S, RSF602720S, RSF602841S, RSF602869S, RSF603429S, RSF603445S, RSF603940S, RSF600558S, RSF600927S, RSF601191S, RSF601213S, RSF602873S, RSF600431S, RSF600432S, RSF601120S, RSF602844S, RSF603226S, RSF604822S, RSF600598S, RSF607735S, RSF601333S, RSF602868S, RSF600374S, RSF601241S, RSF603784S, RSF600319S, RSF602511S, RSF600377S, RSF600926S, RSF601343S, RSF602862S, RSF604216S, RSF605676S, RSF606882S, RSF607027S, RSF607772S, RSF607780S, RSF607791S, RSF607927S, RSF608001S, RSF600555S, RSF600640S, RSF600658S, RSF600685S, RSF600686S, RSF600795S, RSF601228S, RSF601264S, RSF603099S, RSF603352S, RSF603397S, RSF603424S, RSF603541S, RSF603623S, RSF603708S, RSF603856S, RSF604176S, RSF604231S, RSF604310S, RSF604952S, RSF604990S, RSF605035S, RSF607771S, RSF600393S, RSF601014S, RSF603943S, RSF601162S, RSF602364S, RSF602928S, RSF600699S, RSF601070S, RSF601139S, RSF601154S, RSF601402S, RSF602784S, RSF604245S, RSF604258S, RSF604052S, RSF605870S, RSF601448S, RSF602609S, RSF604200S, RSF601272S, RSF602961S, RSF604661S, RSF605660S, RSF605710S, RSF605925S, RSF605926S, RSF606169S, RSF600392S, RSF600424S, RSF600521S, RSF600619S, RSF600626S, RSF600637S, RSF600638S, RSF600667S, RSF600684S, RSF600773S, RSF600989S, RSF601253S, RSF601386S, RSF601558S, RSF602518S, RSF602573S, RSF602613S, RSF602733S, RSF602806S, RSF603050S, RSF603384S, RSF603407S, RSF603461S, RSF603463S, RSF603567S, RSF603637S, RSF603924S, RSF604165S, RSF604207S, RSF604285S, RSF604323S, RSF604758S, RSF601406S, RSF602716S, RSF603263S, RSF601281S, RSF602696S, RSF603869S, RSF605015S, RSF605681S, RSF605758S, RSF606175S, RSF600438S, RSF600635S, RSF600902S, RSF600993S, RSF601036S, RSF601230S, RSF601403S, RSF601429S, RSF603488S, RSF604316S, RSF604327S, RSF605046S, RSF605528S, RSF600663S, RSF600665S, RSF601087S, RSF600179S, RSF600198S, RSF600243S, RSF600332S, RSF600341S, RSF600813S, RSF600994S, RSF601039S, RSF601322S, RSF602468S, RSF602785S, RSF603179S, RSF605338S, RSF605649S, RSF605920S, RSF606053S, RSF606317S, RSF606482S, RSF607034S, RSF607038S, RSF607040S, RSF607041S, RSF607776S, RSF607778S, RSF607800S, RSF600340S, RSF600525S, RSF600925S, RSF601445S, RSF601553S, RSF602585S, RSF602612S, RSF602670S, RSF603374S, RSF603377S, RSF603389S, RSF603612S, RSF603615S, RSF603626S, RSF603719S, RSF603774S, RSF603903S, RSF604053S, RSF604238S, RSF604947S, RSF604951S, RSF604978S, RSF604999S, RSF607765S, RSF608049S, RSF608050S, RSF600354S, RSF600355S, RSF600919S, RSF601063S, RSF601177S, RSF601374S, RSF602833S, RSF603692S, RSF603972S, RSF604224S, RSF604225S, RSF604898S, RSF601286S, RSF600653S, RSF601027S, RSF602354S, RSF603383S, RSF603388S, RSF603448S, RSF603525S, RSF604787S, RSF605034S, RSF605970S, RSF606059S, RSF600295S, RSF600368S, RSF603951S, RSF604229S, RSF603867S, RSF600594S, RSF600791S, RSF601263S, RSF602692S, RSF604635S, RSF607065S, RSF606226S, RSF606270S, RSF607236S, RSF607334S, RSF607372S, RSF604250S, RSF604278S, RSF604279S, RSF607022S, RSF600161S, RSF600956S, RSF601234S, RSF605641S, RSF606181S, RSF607348S, RSF607350S, RSF601376S, RSF603469S, RSF604027S, RSF602439S, RSF605639S, RSF601104S, RSF602721S, RSF606987S, RSF600498S, RSF600770S, RSF607898S, RSF603423S, RSF603479S, RSF603548S, RSF605947S, RSF600754S, RSF600809S, RSF600942S, RSF601133S, RSF601200S, RSF603043S, RSF603849S, RSF604460S, RSF601320S, RSF603148S, RSF603218S, RSF603690S, RSF604455S, RSF602938S, RSF607880S, RSF600936S, RSF601136S, RSF601194S, RSF602181S, RSF602852S, RSF603192S, RSF603363S, RSF603591S, RSF603705S, RSF606009S, RSF606010S, RSF606019S, RSF607316S, RSF603556S, RSF605087S, RSF605088S, RSF605089S, RSF605340S, RSF605526S, RSF607317S, RSF607320S, RSF608046S, RSF600568S, RSF603090S, RSF604230S, RSF605617S, RSF600217S, RSF601082S, RSF601083S, RSF601143S, RSF602320S, RSF603193S, RSF603599S, RSF600535S, RSF603120S, RSF601000S, RSF601584S, RSF607972S, RSF607973S, RSF607976S, RSF607977S, RSF607978S, RSF607970S, RSF607971S, RSF607974S, RSF607975S, RSF607979S, RSF605655S, RSF605878S, RSF606234S, RSF606489S, RSF606491S, RSF607195S, RSF607196S, RSF607197S, RSF607198S, RSF607199S, RSF607200S, RSF607275S, RSF600147S, RSF600148S, RSF600190S, RSF600342S, RSF600347S, RSF600351S, RSF600365S, RSF600517S, RSF600590S, RSF600682S, RSF600683S, RSF600846S, RSF600860S, RSF600984S, RSF600992S, RSF601029S, RSF601166S, RSF601188S, RSF601295S, RSF601303S, RSF601310S, RSF602085S, RSF602159S, RSF602164S, RSF602179S, RSF602767S, RSF602815S, RSF602857S, RSF602870S, RSF602882S, RSF603116S, RSF603187S, RSF603663S, RSF603835S, RSF603873S, RSF604261S, RSF604262S, RSF604400S, RSF604837S, RSF605012S, RSF605262S, RSF605898S, RSF605899S, RSF605900S, RSF605901S, RSF606209S, RSF606210S, RSF606322S, RSF600311S, RSF600344S, RSF600559S, RSF600588S, RSF600600S, RSF600695S, RSF600702S, RSF600719S, RSF600800S, RSF600819S, RSF600831S, RSF600972S, RSF601182S, RSF601259S, RSF601396S, RSF601401S, RSF601489S, RSF602339S, RSF602593S, RSF602621S, RSF602627S, RSF602645S, RSF602647S, RSF602773S, RSF602774S, RSF602776S, RSF602940S, RSF603359S, RSF603375S, RSF603408S, RSF603506S, RSF603561S, RSF603745S, RSF603777S, RSF603859S, RSF603878S, RSF603905S, RSF603919S, RSF604167S, RSF604177S, RSF604208S, RSF604233S, RSF604426S, RSF604450S, RSF604477S, RSF604657S, RSF604796S, RSF605014S, RSF605038S, RSF605048S, RSF605507S, RSF605533S, RSF605534S, RSF605535S, RSF605562S, RSF605564S, RSF605518S, RSF600516S, RSF603536S, RSF604325S, RSF606968S, RSF607886S, RSF600227S, RSF600272S, RSF600457S, RSF600837S, RSF600939S, RSF601071S, RSF601112S, RSF603606S, RSF607336S, RSF600434S, RSF601390S, RSF604269S, RSF600513S, RSF601332S, RSF603173S, RSF600536S, RSF600829S, RSF601138S, RSF602403S, RSF603129S, RSF605693S, RSF605921S, RSF605922S, RSF606208S, RSF606315S, RSF607364S, RSF603681S, RSF604469S, RSF604689S, RSF604792S, RSF605027S, RSF605215S, RSF605508S, RSF605566S, RSF602923S, RSF605683S, RSF602519S, RSF603728S, RSF604265S, RSF602537S, RSF600259S, RSF600369S, RSF603779S, RSF605961S, RSF605992S, RSF600164S, RSF606237S, RSF602975S, RSF603825S, RSF603828S, RSF604138S, RSF602695S, RSF600248S, RSF600766S, RSF600172S, RSF600173S, RSF601312S, RSF603552S, RSF607717S, RSF600182S, RSF604198S, RSF603573S, RSF603644S, RSF604024S, RSF604118S, RSF605002S, RSF600562S, RSF600816S, RSF605859S, RSF606269S, RSF606331S, RSF606516S, RSF600163S, RSF600888S, RSF602430S, RSF602830S, RSF603126S, RSF603515S, RSF603547S, RSF603558S, RSF603577S, RSF603766S, RSF603881S, RSF603887S, RSF603954S, RSF604685S, RSF604696S, RSF604763S, RSF605010S, RSF605671S, RSF607045S, RSF607279S, RSF607895S, RSF602163S, RSF602793S, RSF603820S, RSF604252S, RSF604281S, RSF608055S, RSF601308S, RSF604249S, RSF600439S, RSF600714S, RSF600715S, RSF600734S, RSF600778S, RSF600856S, RSF601342S, RSF602722S, RSF603023S, RSF603581S, RSF603871S, RSF604096S, RSF604162S, RSF603588S, RSF604839S, RSF604998S, RSF607939S, RSF603364S, RSF603365S, RSF603974S, RSF604847S, RSF604901S, RSF607201S, RSF607851S, RSF608040S, RSF600885S, RSF600907S, RSF601576S, RSF602730S, RSF603234S, RSF603555S, RSF604185S, RSF604967S, RSF605239S, RSF603628S, RSF603629S, RSF600425S, RSF605647S, RSF605648S, RSF605737S, RSF605756S, RSF605913S, RSF605914S, RSF605917S, RSF606054S, RSF606055S, RSF606058S, RSF606178S, RSF606179S, RSF606204S, RSF606205S, RSF606247S, RSF606248S, RSF606249S, RSF606250S, RSF606319S, RSF606481S, RSF606485S, RSF606591S, RSF606878S, RSF606879S, RSF606883S, RSF607026S, RSF607748S, RSF607749S, RSF607752S, RSF607754S, RSF607755S, RSF607756S, RSF607760S, RSF607932S, RSF607933S, RSF608047S, RSF600330S, RSF600363S, RSF600423S, RSF600461S, RSF600491S, RSF600520S, RSF600522S, RSF600726S, RSF600727S, RSF600730S, RSF600746S, RSF600768S, RSF600798S, RSF600857S, RSF601221S, RSF601284S, RSF602513S, RSF602538S, RSF602542S, RSF603031S, RSF603144S, RSF603351S, RSF603353S, RSF603381S, RSF603495S, RSF603571S, RSF603608S, RSF603616S, RSF603967S, RSF604069S, RSF604166S, RSF604247S, RSF604318S, RSF604443S, RSF604475S, RSF604646S, RSF604672S, RSF604690S, RSF604790S, RSF604960S, RSF604963S, RSF605000S, RSF605192S, RSF605193S, RSF605219S, RSF605487S, RSF605490S, RSF605504S, RSF605506S, RSF605538S, RSF605542S, RSF605551S, RSF605553S, RSF605554S, RSF608000S, RSF607638S, RSF600300S, RSF600929S, RSF603427S, RSF603460S, RSF607893S, RSF606212S, RSF607042S, RSF607043S, RSF607297S, RSF607394S, RSF607703S, RSF600197S, RSF600912S, RSF600945S, RSF602643S, RSF602677S, RSF602851S, RSF602856S, RSF603235S, RSF603391S, RSF603560S, RSF604008S, RSF604104S, RSF604212S, RSF606185S, RSF606186S, RSF606243S, RSF606524S, RSF607202S, RSF600629S, RSF600646S, RSF601095S, RSF601111S, RSF601454S, RSF602890S, RSF603185S, RSF603583S, RSF603992S, RSF605227S, RSF605503S, RSF607728S, RSF600184S, RSF600735S, RSF601124S, RSF607937S, RSF607938S, RSF607936S, RSF605960S, RSF606253S, RSF606512S, RSF606522S, RSF607078S, RSF607228S, RSF607344S, RSF600194S, RSF600871S, RSF601458S, RSF601536S, RSF602440S, RSF602688S, RSF602824S, RSF602864S, RSF603145S, RSF603434S, RSF603907S, RSF604490S, RSF604667S, RSF600609S, RSF600614S, RSF603481S, RSF600744S, RSF604257S, RSF606254S, RSF606280S, RSF606281S, RSF606958S, RSF607230S, RSF607391S, RSF601005S, RSF602636S, RSF602686S, RSF602964S, RSF602966S, RSF602971S, RSF602978S, RSF603882S, RSF603884S, RSF603894S, RSF604199S, RSF604744S, RSF607871S, RSF601134S, RSF601090S, RSF603952S, RSF605229S, RSF603704S, RSF607874S, RSF608011S, RSF605589S, RSF605672S, RSF605673S, RSF605871S, RSF605872S, RSF606213S, RSF606215S, RSF606246S, RSF606495S, RSF606944S, RSF607060S, RSF607363S, RSF600921S, RSF601016S, RSF601051S, RSF601123S, RSF602554S, RSF602582S, RSF602614S, RSF602639S, RSF603119S, RSF603453S, RSF603487S, RSF603715S, RSF603862S, RSF604011S, RSF604253S, RSF604259S, RSF604299S, RSF604419S, RSF604833S, RSF605198S, RSF605407S, RSF605532S, RSF607061S, RSF605682S, RSF600265S, RSF600405S, RSF602556S, RSF602576S, RSF602811S, RSF603380S, RSF603393S, RSF603419S, RSF603421S, RSF603508S, RSF603621S, RSF603739S, RSF603877S, RSF603906S, RSF604276S, RSF604414S, RSF604693S, RSF605039S, RSF605235S, RSF605493S, RSF600143S, RSF600644S, RSF601220S, RSF601565S, RSF600202S, RSF600381S, RSF600480S, RSF607249S, RSF607379S, RSF607894S, RSF608090S, RSF603127S, RSF604678S, RSF603636S, RSF603997S, RSF605942S, RSF600264S, RSF601335S, RSF603442S, RSF603843S, RSF603617S, RSF604682S, RSF602900S, RSF604864S, RSF606483S, RSF607079S, RSF607751S, RSF600678S, RSF604452S, RSF606954S, RSF602566S, RSF606274S, RSF600624S, RSF602625S, RSF600280S, RSF600913S, RSF601099S, RSF602682S, RSF603688S, RSF602371S, RSF602800S, RSF603385S, RSF603387S, RSF603590S, RSF603679S, RSF603860S, RSF604828S, RSF604972S, RSF605001S, RSF605026S, RSF605218S, RSF605220S, RSF606170S, RSF601435S, RSF607881S, RSF600859S, RSF606222S, RSF606956S, RSF600595S, RSF602724S, RSF602397S, RSF605675S, RSF603141S, RSF606492S, RSF604189S, RSF604881S, RSF605197S, RSF607303S, RSF600288S, RSF600764S, RSF601037S, RSF603707S, RSF603725S, RSF605919S, RSF607788S, RSF607789S, RSF607790S, RSF600601S, RSF600604S, RSF600606S, RSF600851S, RSF601216S, RSF604144S, RSF604145S, RSF605009S, RSF605234S, RSF605409S, RSF605552S, RSF605986S, RSF604458S, RSF604068S, RSF604076S, RSF604077S, RSF604078S, RSF605106S, RSF605107S, RSF600799S, RSF601381S, RSF602567S, RSF603039S, RSF603040S, RSF603390S, RSF603394S, RSF603431S, RSF603610S, RSF603872S, RSF603874S, RSF604126S, RSF604495S, RSF604778S, RSF604965S, RSF605216S, RSF605486S, RSF605753S, RSF605854S, RSF608026S, RSF608035S, RSF608056S, RSF605709S, RSF605963S, RSF606171S, RSF606174S, RSF606176S, RSF606255S, RSF607075S, RSF600361S, RSF600496S, RSF600950S, RSF603401S, RSF603446S, RSF603753S, RSF604097S, RSF604161S, RSF604286S, RSF604836S, RSF608092S, RSF605949S, RSF605954S, RSF606182S, RSF606183S, RSF606242S, RSF608002S, RSF600660S, RSF601243S, RSF601252S, RSF601452S, RSF601562S, RSF601583S, RSF602187S, RSF602399S, RSF602828S, RSF602832S, RSF602840S, RSF602879S, RSF605969S, RSF600738S, RSF608012S, RSF607245S, RSF607246S, RSF607021S, RSF607918S, RSF600487S, RSF606520S, RSF600448S, RSF605703S, RSF602659S, RSF600464S, RSF602525S, RSF602747S, RSF602803S, RSF600497S, RSF600796S, RSF601294S, RSF602457S, RSF602697S, RSF603594S, RSF604137S, RSF605051S, RSF601315S, RSF601316S, RSF601341S, RSF603029S, RSF603056S, RSF604707S, RSF602352S, RSF602853S, RSF603791S, RSF600997S, RSF600175S, RSF600178S, RSF600200S, RSF600388S, RSF601157S, RSF601159S, RSF601160S, RSF601161S, RSF601476S, RSF602756S, RSF603414S, RSF603706S, RSF604184S, RSF601441S, RSF601457S, RSF601363S, RSF601364S, RSF601368S, RSF600218S, RSF601126S, RSF603998S, RSF604445S, RSF607396S, RSF602875S, RSF605952S, RSF600308S, RSF600495S, RSF600553S, RSF600578S, RSF600579S, RSF600881S, RSF600916S, RSF600995S, RSF601219S, RSF601273S, RSF602594S, RSF602760S, RSF603896S, RSF604275S, RSF604454S, RSF601351S, RSF601367S, RSF600187S, RSF605147S, RSF600904S, RSF604768S, RSF605500S, RSF607413S, RSF603956S, RSF607090S, RSF607221S, RSF607407S, RSF607899S, RSF605545S, RSF608005S, RSF605642S, RSF606172S, RSF600642S, RSF601399S, RSF603044S, RSF603474S, RSF603578S, RSF605055S, RSF606511S, RSF601257S, RSF602791S, RSF605971S, RSF607070S, RSF607285S, RSF607358S, RSF600485S, RSF600784S, RSF601020S, RSF601163S, RSF601361S, RSF601460S, RSF601472S, RSF602269S, RSF602801S, RSF602929S, RSF603635S, RSF603953S, RSF607354S, RSF604641S, RSF605951S, RSF601418S, RSF603462S, RSF601195S, RSF604256S, RSF602521S, RSF600479S, RSF600551S, RSF601180S, RSF601199S, RSF601590S, RSF602189S, RSF602198S, RSF602199S, RSF603118S, RSF603554S, RSF603582S, RSF603671S, RSF604072S, RSF604123S, RSF604242S, RSF604014S, RSF603579S, RSF600893S, RSF603117S, RSF604066S, RSF600239S, RSF600541S, RSF603010S, RSF603095S, RSF603110S, RSF604322S, RSF604438S, RSF606332S, RSF600486S, RSF600519S, RSF601015S, RSF602959S, RSF603087S, RSF603959S, RSF605694S, RSF606238S, RSF601072S, RSF603370S, RSF605484S, RSF605238S, RSF603140S, RSF601197S, RSF601383S, RSF603094S, RSF603231S, RSF603233S, RSF603498S, RSF603667S, RSF607410S, RSF607411S, RSF602982S, RSF603250S, RSF603251S, RSF605686S, RSF605989S, RSF607866S, RSF600249S, RSF600285S, RSF601032S, RSF601034S, RSF601040S, RSF601041S, RSF601042S, RSF601046S, RSF601317S, RSF601412S, RSF602406S, RSF602633S, RSF602673S, RSF602675S, RSF602754S, RSF602814S, RSF602871S, RSF603121S, RSF603123S, RSF603124S, RSF603125S, RSF603128S, RSF603168S, RSF603220S, RSF603721S, RSF603795S, RSF603840S, RSF604459S, RSF600168S, RSF601385S, RSF600266S, RSF600718S, RSF600882S, RSF600803S, RSF600245S, RSF600326S, RSF600416S, RSF600584S, RSF600634S, RSF600978S, RSF601233S, RSF602487S, RSF602545S, RSF602768S, RSF603028S, RSF603531S, RSF600268S, RSF602821S, RSF604017S, RSF602726S, RSF602805S, RSF603673S, RSF603698S, RSF605272S, RSF605627S, RSF605653S, RSF606341S, RSF600316S, RSF600378S, RSF600415S, RSF600447S, RSF600556S, RSF600580S, RSF600582S, RSF600750S, RSF601304S, RSF601440S, RSF602766S, RSF602798S, RSF602907S, RSF603485S, RSF603524S, RSF603674S, RSF603768S, RSF604421S, RSF604764S, RSF605043S, RSF605491S, RSF600366S, RSF601023S, RSF601271S, RSF601371S, RSF602942S, RSF604132S, RSF604183S, RSF607290S, RSF601237S, RSF601348S, RSF603350S, RSF603780S, RSF605369S, RSF605413S, RSF602772S, RSF600772S, RSF602520S, RSF602640S, RSF602867S, RSF602919S, RSF603597S, RSF603888S, RSF603904S, RSF603989S, RSF603483S, RSF600533S, RSF604709S, RSF604713S, RSF604714S, RSF607746S, RSF600465S, RSF606890S, RSF606995S, RSF607203S, RSF600186S, RSF600199S, RSF600201S, RSF600230S, RSF600287S, RSF600293S, RSF600468S, RSF600469S, RSF601156S, RSF601578S, RSF602986S, RSF603021S, RSF603361S, RSF605217S, RSF600231S, RSF600286S, RSF600294S, RSF600317S, RSF600470S, RSF600810S, RSF600981S, RSF601004S, RSF601302S, RSF601485S, RSF602325S, RSF603459S, RSF606257S, RSF607054S, RSF600449S, RSF601248S, RSF602536S, RSF602539S, RSF602563S, RSF602642S, RSF602655S, RSF603420S, RSF603477S, RSF603503S, RSF603624S, RSF604241S, RSF604428S, RSF604706S, RSF604774S, RSF604775S, RSF604843S, RSF607210S, RSF600923S, RSF600930S, RSF601114S, RSF601115S, RSF601276S, RSF602365S, RSF602825S, RSF603410S, RSF603422S, RSF603786S, RSF603787S, RSF603790S, RSF603796S, RSF604650S, RSF604835S, RSF605029S, RSF605053S, RSF605226S, RSF607272S, RSF605883S, RSF600278S, RSF600301S, RSF600349S, RSF600362S, RSF600625S, RSF600792S, RSF600834S, RSF601010S, RSF601044S, RSF601130S, RSF601378S, RSF601455S, RSF601466S, RSF602161S, RSF602581S, RSF602583S, RSF602762S, RSF602878S, RSF603355S, RSF603836S, RSF604270S, RSF604695S, RSF605569S, RSF608085S, RSF605575S, RSF605880S, RSF606184S, RSF606926S, RSF600196S, RSF600277S, RSF601300S, RSF602548S, RSF602679S, RSF602731S, RSF602787S, RSF602866S, RSF603194S, RSF603486S, RSF603642S, RSF603797S, RSF603822S, RSF603845S, RSF603899S, RSF604074S, RSF604127S, RSF604141S, RSF604501S, RSF604769S, RSF605195S, RSF603085S, RSF603086S, RSF605033S, RSF600305S, RSF600542S, RSF600583S, RSF600586S, RSF600620S, RSF600690S, RSF601026S, RSF601218S, RSF601225S, RSF600509S, RSF600547S, RSF600565S, RSF600758S, RSF600760S, RSF600765S, RSF600776S, RSF600877S, RSF600879S, RSF600918S, RSF605646S, RSF605892S, RSF605906S, RSF605907S, RSF606190S, RSF606219S, RSF606240S, RSF606539S, RSF606988S, RSF607310S, RSF607339S, RSF607340S, RSF607355S, RSF607408S, RSF607409S, RSF608008S, RSF600236S, RSF600983S, RSF601293S, RSF601463S, RSF602624S, RSF602649S, RSF602650S, RSF602656S, RSF602687S, RSF602850S, RSF602972S, RSF602973S, RSF603025S, RSF603245S, RSF603247S, RSF603666S, RSF603730S, RSF604021S, RSF604474S, RSF604493S, RSF604644S, RSF604773S, RSF604941S, RSF604974S, RSF605222S, RSF605439S, RSF605441S, RSF605494S, RSF605498S, RSF605557S, RSF604626S, RSF604715S, RSF604716S, RSF605095S, RSF605099S, RSF605105S, RSF605108S, RSF605109S, RSF605110S, RSF605111S, RSF607115S, RSF600987S, RSF601043S, RSF601045S, RSF601047S, RSF601050S, RSF601052S, RSF601054S, RSF601055S, RSF601056S, RSF601057S, RSF601059S, RSF601066S, RSF601067S, RSF601068S, RSF601069S, RSF601074S, RSF602941S, RSF602953S, RSF602954S, RSF602957S, RSF602960S, RSF602962S, RSF604001S, RSF604004S, RSF604007S, RSF606321S, RSF608096S, RSF600574S, RSF600700S, RSF603722S, RSF606064S, RSF606065S, RSF603936S, RSF603975S, RSF605953S, RSF605957S, RSF606999S, RSF600693S, RSF600823S, RSF601389S, RSF601552S, RSF602488S, RSF602788S, RSF600226S, RSF601018S, RSF601110S, RSF601270S, RSF606517S, RSF600244S, RSF604067S, RSF607288S, RSF607346S, RSF601210S, RSF601292S, RSF601478S, RSF602704S, RSF603830S, RSF604226S, RSF605058S, RSF603038S, RSF606490S, RSF607047S, RSF607223S, RSF607269S, RSF607295S, RSF607381S, RSF603798S, RSF602740S, RSF603433S, RSF603609S, RSF604171S, RSF603059S, RSF604057S, RSF605668S, RSF607693S, RSF601236S, RSF601465S, RSF601223S, RSF603834S, RSF605983S, RSF608038S, RSF603944S, RSF604203S, RSF600654S, RSF601268S, RSF604680S, RSF600691S, RSF603514S, RSF605887S, RSF606241S, RSF606991S, RSF607241S, RSF607273S, RSF607305S, RSF607398S, RSF607999S, RSF600952S, RSF600953S, RSF602729S, RSF602742S, RSF603133S, RSF603395S, RSF605236S, RSF605438S, RSF608126S, RSF605704S, RSF605965S, RSF606232S, RSF606537S, RSF602503S, RSF602735S, RSF604094S, RSF604288S, RSF604964S, RSF602265S, RSF600167S, RSF600174S, RSF600242S, RSF600292S, RSF600587S, RSF600592S, RSF600613S, RSF603207S, RSF603478S, RSF603564S, RSF602955S, RSF606923S, RSF602569S, RSF603930S, RSF604009S, RSF604061S, RSF606008S, RSF606480S, RSF606881S, RSF607036S, RSF600529S, RSF600618S, RSF602769S, RSF603358S, RSF603922S, RSF604131S, RSF604449S, RSF604651S, RSF604652S, RSF605541S, RSF605689S, RSF600575S, RSF603254S, RSF603241S, RSF603670S, RSF604085S, RSF604752S, RSF600499S, RSF601084S, RSF601325S, RSF604155S, RSF604274S, RSF603096S, RSF603101S, RSF603366S, RSF603928S, RSF604673S, RSF604679S, RSF600379S, RSF606268S, RSF600181S, RSF603530S, RSF602492S, RSF602661S, RSF603669S, RSF600193S, RSF602846S, RSF605346S, RSF601329S, RSF602550S, RSF603640S, RSF606180S, RSF606206S, RSF606880S, RSF607031S, RSF607250S, RSF600526S, RSF600673S, RSF600970S, RSF602568S, RSF602606S, RSF602622S, RSF603584S, RSF603770S, RSF603900S, RSF603979S, RSF604046S, RSF604676S, RSF604686S, RSF604741S, RSF605023S, RSF607929S, RSF604113S, RSF603819S, RSF601481S, RSF602765S, RSF605665S, RSF605228S, RSF600372S, RSF602167S, RSF602718S, RSF602737S, RSF602823S, RSF606966S, RSF601362S, RSF602517S, RSF604026S, RSF601298S, RSF606887S, RSF600752S, RSF604149S, RSF600647S, RSF605094S, RSF603451S, RSF603714S, RSF603438S, RSF603948S, RSF605044S, RSF605045S, RSF604063S, RSF603993S, RSF605667S, RSF605905S, RSF606187S, RSF606220S, RSF606258S, RSF606279S, RSF606326S, RSF606515S, RSF606989S, RSF607209S, RSF607248S, RSF607265S, RSF607293S, RSF607347S, RSF607369S, RSF607415S, RSF607416S, RSF608007S, RSF601201S, RSF601203S, RSF601204S, RSF601319S, RSF602377S, RSF602415S, RSF602467S, RSF602681S, RSF602736S, RSF602829S, RSF603001S, RSF603020S, RSF603115S, RSF603224S, RSF603502S, RSF603694S, RSF603857S, RSF603988S, RSF604075S, RSF604239S, RSF604384S, RSF604442S, RSF604944S, RSF605124S, RSF605194S, RSF605221S, RSF605237S, RSF605258S, RSF605546S, RSF605558S, RSF607353S, RSF600335S, RSF600336S, RSF600339S, RSF601340S, RSF602712S, RSF600304S, RSF600306S, RSF600307S, RSF600284S, RSF601006S, RSF603509S, RSF603732S, RSF606541S, RSF606538S, RSF606244S, RSF601297S, RSF608075S, RSF608111S, RSF600412S, RSF600418S, RSF600460S, RSF600462S, RSF601183S, RSF601347S, RSF601380S, RSF601470S, RSF601556S, RSF602669S, RSF602705S, RSF602796S, RSF605858S, RSF608025S, RSF608027S, RSF604982S, RSF601344S, RSF601451S, RSF605657S, RSF605868S, RSF605981S, RSF607341S, RSF600960S, RSF604193S, RSF604330S, RSF605004S, RSF606553S, RSF604140S, RSF602558S, RSF602561S, RSF600367S, RSF602424S, RSF605685S, RSF600188S, RSF600189S, RSF600276S, RSF600291S, RSF600482S, RSF600506S, RSF600514S, RSF600763S, RSF600944S, RSF601024S, RSF601146S, RSF601431S, RSF602410S, RSF602549S, RSF602693S, RSF602758S, RSF603137S, RSF603138S, RSF603143S, RSF603466S, RSF604264S, RSF605265S, RSF601283S, RSF601393S, RSF602662S, RSF601442S, RSF600010S, RSF600290S, RSF602648S, RSF600256S, RSF605561S, RSF603076S, RSF603084S, RSF607864S, RSF603767S, RSF600895S, RSF601208S, RSF603396S, RSF601209S, RSF600932S, RSF600933S, RSF602644S, RSF604385S, RSF604959S, RSF605881S, RSF603175S, RSF605502S, RSF603392S, RSF600897S, RSF602598S, RSF602599S, RSF602810S, RSF603493S, RSF603723S, RSF607017S, RSF606218S, RSF606314S, RSF606500S, RSF607069S, RSF600228S, RSF600232S, RSF600546S, RSF600631S, RSF601077S, RSF601285S, RSF602635S, RSF603771S, RSF607018S, RSF600481S, RSF603255S, RSF603580S, RSF603638S, RSF603535S, RSF604647S, RSF604671S, RSF607351S, RSF600842S, RSF602779S, RSF603605S, RSF607009S, RSF606916S, RSF607046S, RSF600761S, RSF601064S, RSF601324S, RSF601330S, RSF601359S, RSF601468S, RSF601582S, RSF602414S, RSF602611S, RSF602794S, RSF602891S, RSF604850S, RSF605855S, RSF607234S, RSF606225S, RSF601145S, RSF605666S, RSF606266S, RSF606267S, RSF606525S, RSF607107S, RSF607108S, RSF607109S, RSF608051S, RSF600253S, RSF600343S, RSF600611S, RSF600612S, RSF600628S, RSF600630S, RSF600775S, RSF600958S, RSF601019S, RSF601097S, RSF601158S, RSF602367S, RSF602398S, RSF602602S, RSF602843S, RSF602845S, RSF603147S, RSF603443S, RSF603454S, RSF603593S, RSF603641S, RSF603898S, RSF604151S, RSF600430S, RSF602671S, RSF606236S, RSF601443S, RSF602764S, RSF607892S, RSF605977S, RSF601530S, RSF604271S, RSF604770S, RSF603781S, RSF600390S, RSF600576S, RSF604845S, RSF605006S, RSF602699S, RSF602984S, RSF603371S, RSF602217S, RSF602995S, RSF603052S, RSF603476S, RSF602349S, RSF603776S, RSF600898S, RSF601121S, RSF601357S, RSF602819S, RSF603146S, RSF603210S, RSF604023S, RSF604202S, RSF604383S, RSF602608S, RSF603507S, RSF600501S, RSF605902S, RSF606316S, RSF606997S, RSF600961S, RSF601244S, RSF601373S, RSF602557S, RSF603480S, RSF603592S, RSF604684S, RSF606260S, RSF606277S, RSF601258S, RSF604235S, RSF605485S, RSF605556S, RSF607926S, RSF605904S, RSF605910S, RSF605911S, RSF605912S, RSF606223S, RSF606262S, RSF606340S, RSF606343S, RSF606346S, RSF606347S, RSF606479S, RSF606895S, RSF606897S, RSF607815S, RSF600771S, RSF600965S, RSF601022S, RSF601229S, RSF602555S, RSF602560S, RSF602580S, RSF602775S, RSF604034S, RSF604150S, RSF604240S, RSF604267S, RSF604339S, RSF604424S, RSF604649S, RSF604989S, RSF605049S, RSF605544S, RSF606896S, RSF607924S, RSF605151S, RSF602416S, RSF600596S, RSF600948S, RSF607718S, RSF607901S, RSF600900S, RSF601009S, RSF601017S, RSF602926S, RSF604319S, RSF607376S, RSF601365S, RSF601561S, RSF601564S, RSF601575S, RSF601585S, RSF603660S, RSF603551S, RSF605958S, RSF608078S, RSF605980S, RSF605189S, RSF605549S, RSF605658S, RSF606285S, RSF607071S, RSF601375S, RSF604789S, RSF600334S, RSF600494S, RSF600712S, RSF600974S, RSF602887S, RSF603177S, RSF604251S, RSF600458S, RSF600484S, RSF600608S, RSF600639S, RSF600651S, RSF600710S, RSF601211S, RSF600318S, RSF607857S, RSF607859S, RSF607860S, RSF607861S, RSF607862S, RSF607863S, RSF602660S, RSF602715S, RSF603475S, RSF602807S, RSF605581S, RSF605955S, RSF600146S, RSF600204S, RSF600313S, RSF600406S, RSF600444S, RSF600505S, RSF600554S, RSF600661S, RSF602529S, RSF603376S, RSF603537S, RSF604431S, RSF604777S, RSF604831S, RSF604892S, RSF606052S, RSF607024S, RSF600206S, RSF603731S, RSF604273S, RSF604900S, RSF606312S, RSF606494S, RSF606940S, RSF600411S, RSF600648S, RSF600689S, RSF600720S, RSF601366S, RSF604287S, RSF606335S, RSF606888S, RSF607208S, RSF607375S, RSF607903S, RSF608004S, RSF608006S, RSF600854S, RSF601474S, RSF602719S, RSF603622S, RSF603769S, RSF603891S, RSF604125S, RSF604128S, RSF604645S, RSF604666S, RSF604677S, RSF600675S, RSF605679S, RSF606239S, RSF600732S, RSF607064S, RSF600489S, RSF600707S, RSF601313S, RSF601076S, RSF605032S, RSF605948S, RSF606630S, RSF607270S, RSF600910S, RSF601131S, RSF601301S, RSF601309S, RSF602261S, RSF602701S, RSF602893S, RSF603703S, RSF603838S, RSF603865S, RSF603970S, RSF604064S, RSF604148S, RSF604201S, RSF604697S, RSF601350S, RSF600849S, RSF603710S, RSF603814S, RSF604236S, RSF604070S, RSF601007S, RSF606935S, RSF601358S, RSF600725S, RSF600728S, RSF600890S, RSF600931S, RSF600963S, RSF600966S, RSF601038S, RSF601073S, RSF601144S, RSF601179S, RSF603047S, RSF604698S, RSF605924S, RSF607915S, RSF600387S, RSF602694S, RSF607405S, RSF607996S, RSF600149S, RSF601150S, RSF601529S, RSF602353S, RSF602401S, RSF602834S, RSF602854S, RSF602855S, RSF603176S, RSF603259S, RSF603357S, RSF603400S, RSF603413S, RSF603458S, RSF603645S, RSF603682S, RSF603942S, RSF604108S, RSF604116S, RSF600845S, RSF600220S, RSF600222S, RSF600261S, RSF600325S, RSF600327S, RSF600357S, RSF600375S, RSF600389S, RSF600429S, RSF600757S, RSF600759S, RSF600781S, RSF600826S, RSF600855S, RSF601581S, RSF604429S, RSF605903S, RSF606261S, RSF601181S, RSF603964S, RSF605978S, RSF606092S, RSF603516S, RSF603613S, RSF603921S, RSF603990S, RSF604147S, RSF604169S, RSF604456S, RSF604910S, RSF604976S, RSF605016S, RSF605529S, RSF606506S, RSF603403S, RSF603778S, RSF603926S, RSF600980S, RSF600337S, RSF600436S, RSF600445S, RSF600959S, RSF601479S, RSF605691S, RSF602479S, RSF607708S, RSF600537S, RSF602392S, RSF600567S, RSF602881S, RSF605005S, RSF603618S, RSF600899S, RSF600191S, RSF600192S, RSF601245S, RSF601352S, RSF601356S, RSF601444S, RSF602221S, RSF602876S, RSF602967S, RSF603650S, RSF604059S, RSF602924S, RSF602927S, RSF603067S, RSF603068S, RSF600115S, RSF600165S, RSF600212S, RSF600213S, RSF600225S, RSF600333S, RSF600348S, RSF600483S, RSF600657S, RSF600709S, RSF600731S, RSF601080S, RSF601081S, RSF602710S, RSF603467S, RSF603523S, RSF603684S, RSF603557S, RSF604048S, RSF606889S, RSF607737S, RSF607738S, RSF602872S, RSF604476S, RSF603788S, RSF608016S, RSF601202S, RSF601405S, RSF601408S, RSF601409S, RSF602913S, RSF602916S, RSF602939S, RSF605112S, RSF605113S, RSF605114S, RSF606011S, RSF606012S, RSF606569S, RSF606578S, RSF606588S, RSF606979S, RSF606984S, RSF607117S, RSF607118S, RSF607119S, RSF607120S, RSF607121S, RSF600891S, RSF600901S, RSF601296S, RSF601321S, RSF601404S, RSF608112S, RSF603075S, RSF600473S, RSF600723S, RSF601307S, RSF602552S, RSF603686S, RSF607088S, RSF607089S, RSF606020S, RSF606024S, RSF606025S, RSF606028S, RSF606029S, RSF606031S, RSF606032S, RSF606034S, RSF606036S, RSF604618S, RSF604620S, RSF604621S, RSF604632S, RSF604633S, RSF606023S, RSF606026S, RSF606027S, RSF606033S, RSF606035S, RSF606037S, RSF606550S, RSF603244S, RSF606021S, RSF606628S, RSF606629S, RSF606631S, RSF600903S, RSF603382S, RSF603418S, RSF604480S, RSF606231S, RSF606632S, RSF606637S, RSF606638S, RSF606642S, RSF606643S, RSF606644S, RSF606275S, RSF603601S, RSF603672S, RSF603711S, RSF607729S, RSF606040S, RSF606041S, RSF606043S, RSF606044S, RSF606045S, RSF606046S, RSF606047S, RSF606048S, RSF606049S, RSF606050S, RSF606646S, RSF606648S, RSF606650S, RSF606651S, RSF606653S, RSF606658S, RSF606659S, RSF606690S, RSF600935S, RSF601266S, RSF602914S, RSF602918S, RSF603568S, RSF603665S, RSF603693S, RSF603696S, RSF603699S, RSF603700S, RSF603701S, RSF604634S, RSF604724S, RSF604726S, RSF604728S, RSF604993S, RSF604995S, RSF604996S, RSF604997S, RSF605074S, RSF605076S, RSF605077S, RSF605079S, RSF605096S, RSF605121S, RSF605122S, RSF605123S, RSF605199S, RSF605200S, RSF605201S, RSF605202S, RSF605203S, RSF605302S, RSF605303S, RSF605304S, RSF605305S, RSF605308S, RSF605321S, RSF605322S, RSF605324S, RSF605335S, RSF605336S, RSF605339S, RSF606042S, RSF606652S, RSF606662S, RSF607122S, RSF607123S, RSF607124S, RSF607125S, RSF607126S, RSF607127S, RSF607128S, RSF607129S, RSF607130S, RSF607131S, RSF607132S, RSF607133S, RSF607134S, RSF605893S, RSF600827S, RSF602732S, RSF600884S, RSF605024S, RSF603947S, RSF603977S, RSF604228S, RSF604234S, RSF602646S, RSF605988S, RSF605664S, RSF605246S, RSF606518S, RSF600599S, RSF600627S, RSF600664S, RSF600698S, RSF607274S, RSF607356S, RSF607902S, RSF600650S, RSF602863S, RSF608079S, RSF608077S, RSF608086S, RSF602541S, RSF605075S, RSF608123S, RSF605342S, RSF605343S, RSF606199S, RSF605700S, RSF600205S, RSF600452S, RSF600545S, RSF604111S, RSF604117S, RSF602631S, RSF602717S, RSF602861S, RSF603614S, RSF601277S, RSF603687S, RSF603985S, RSF606038S, RSF606039S, RSF606705S, RSF606707S, RSF606708S, RSF606709S, RSF606704S, RSF603782S, RSF600152S, RSF606329S, RSF607238S, RSF600383S, RSF600435S, RSF604018S, RSF606211S, RSF606892S, RSF607257S, RSF607698S, RSF601085S, RSF603190S, RSF603533S, RSF605036S, RSF601060S, RSF601089S, RSF602357S, RSF603105S, RSF601360S, RSF601487S, RSF600694S, RSF603360S, RSF605698S, RSF605867S, RSF607259S, RSF601338S, RSF602437S, RSF602667S, RSF603049S, RSF603051S, RSF603170S, RSF603540S, RSF605104S, RSF606291S, RSF606487S, RSF607068S, RSF601414S, RSF602987S, RSF602990S, RSF605420S, RSF603058S, RSF607072S, RSF607343S, RSF606308S, RSF601388S, RSF600722S, RSF601548S, RSF603646S, RSF600170S, RSF600442S, RSF603057S, RSF603417S, RSF607357S, RSF606994S, RSF603793S, RSF600171S, RSF605861S, RSF600680S, RSF601164S, RSF601198S, RSF602201S, RSF602267S, RSF602562S, RSF602723S, RSF603532S, RSF606993S, RSF600150S, RSF605056S, RSF600299S, RSF600527S, RSF602755S, RSF603607S, RSF605437S, RSF600622S, RSF601249S, RSF601269S, RSF600833S, RSF600991S, RSF601327S, RSF602708S, RSF603131S, RSF603598S, RSF603604S, RSF603889S, RSF603920S, RSF604746S, RSF608048S, RSF607016S, RSF602395S, RSF601566S, RSF603586S, RSF603885S, RSF604446S, RSF604806S, RSF602826S, RSF605656S, RSF606337S, RSF606339S, RSF606365S, RSF606891S, RSF606990S, RSF606992S, RSF607217S, RSF607292S, RSF607400S, RSF607401S, RSF607404S, RSF600298S, RSF600703S, RSF600820S, RSF603437S, RSF603501S, RSF603633S, RSF604889S, RSF605204S, RSF605442S, RSF607365S, RSF607403S, RSF600704S, RSF600957S, RSF602652S, RSF603473S, RSF600998S, RSF602610S, RSF603868S, RSF603965S, RSF604143S, RSF604209S, RSF604304S, RSF606952S, RSF601079S, RSF604468S, RSF608105S, RSF602246S, RSF603792S, RSF601239S, RSF605884S, RSF605885S, RSF605886S, RSF606192S, RSF606251S, RSF606252S, RSF606955S, RSF600221S, RSF600233S, RSF600235S, RSF600652S, RSF601255S, RSF602535S, RSF602544S, RSF602813S, RSF603472S, RSF603484S, RSF603717S, RSF603855S, RSF604158S, RSF604971S, RSF605041S, RSF607019S, RSF600384S, RSF601141S, RSF602317S, RSF600373S, RSF603880S, RSF600797S, RSF601137S, RSF601149S, RSF601557S, RSF602534S, RSF602641S, RSF603149S, RSF603151S, RSF604050S, RSF604940S, RSF606227S, RSF607012S, RSF600409S, RSF600440S, RSF600493S, RSF600788S, RSF600807S, RSF604174S, RSF604479S, RSF604486S, RSF602838S, RSF600279S, RSF600310S, RSF603386S, RSF603471S, RSF605013S, RSF600356S, RSF605696S, RSF605052S, RSF600964S, RSF600234S, RSF608013S, RSF600255S, RSF600257S, RSF600353S, RSF600777S, RSF603883S, RSF602347S, RSF602792S, RSF602860S, RSF602617S, RSF601267S, RSF600176S, RSF607218S, RSF607412S, RSF603630S, RSF604159S, RSF605223S, RSF600659S, RSF605659S, RSF605894S, RSF605896S, RSF606207S, RSF606276S, RSF607000S, RSF607251S, RSF600250S, RSF600252S, RSF601116S, RSF602968S, RSF602970S, RSF602980S, RSF602981S, RSF603716S, RSF604160S, RSF604213S, RSF605191S, RSF605225S, RSF605560S, RSF607984S, RSF603271S, RSF603668S, RSF605985S, RSF601122S, RSF601339S, RSF602836S, RSF602413S, RSF600135S, RSF601437S, RSF606327S, RSF600616S, RSF600672S, RSF601256S, RSF601337S, RSF602842S, RSF604409S, RSF604497S, RSF604648S, RSF605019S, RSF600321S, RSF600404S, RSF600821S, RSF602680S, RSF602880S, RSF603439S, RSF603457S, RSF603534S, RSF603572S, RSF603634S, RSF603661S, RSF603890S, RSF604062S, RSF604472S, RSF604733S, RSF604930S, RSF605674S, RSF600273S, RSF600928S, RSF601092S, RSF604206S, RSF605680S, RSF607014S, RSF607940S, RSF600543S, RSF600571S, RSF600572S, RSF600573S, RSF602385S, RSF602386S, RSF603440S, RSF603441S, RSF603570S, RSF603589S, RSF603831S, RSF603910S, RSF604157S, RSF604168S, RSF604172S, RSF604187S, RSF604272S, RSF604507S, RSF605643S, RSF605662S, RSF606325S, RSF600169S, RSF600779S, RSF601140S, RSF601355S, RSF602393S, RSF602407S, RSF602472S, RSF602630S, RSF602666S, RSF604956S, RSF604961S, RSF604962S, RSF605547S, RSF605923S, RSF600385S, RSF602433S, RSF603225S, RSF603482S, RSF602837S, RSF600215S, RSF600455S, RSF601002S, RSF601049S, RSF601101S, RSF601107S, RSF600183S, RSF600905S, RSF602168S, RSF602169S, RSF602173S, RSF605984S, RSF607294S, RSF600569S, RSF600597S, RSF600863S, RSF606188S, RSF607726S, RSF602896S, RSF605654S, RSF600502S, RSF600538S, RSF600540S, RSF600548S, RSF600549S, RSF600563S, RSF601473S, RSF601537S, RSF603863S, RSF604003S, RSF604022S, RSF604492S, RSF605196S, RSF606191S, RSF600296S, RSF603861S, RSF603752S, RSF603901S, RSF602432S, RSF602884S, RSF605973S, RSF602497S, RSF603112S, RSF601427S, RSF608110S, RSF605651S, RSF605652S, RSF605702S, RSF605927S, RSF605929S, RSF605930S, RSF605931S, RSF605936S, RSF606259S, RSF606946S, RSF607048S, RSF607049S, RSF607224S, RSF607286S, RSF607287S, RSF607296S, RSF607321S, RSF607327S, RSF607368S, RSF607382S, RSF607742S, RSF607743S, RSF607745S, RSF607910S, RSF607912S, RSF607913S, RSF600155S, RSF600156S, RSF600157S, RSF600185S, RSF600211S, RSF600240S, RSF600283S, RSF600370S, RSF600382S, RSF600450S, RSF600478S, RSF600716S, RSF600756S, RSF600850S, RSF600938S, RSF600962S, RSF600971S, RSF601001S, RSF601061S, RSF601094S, RSF601117S, RSF601170S, RSF601299S, RSF601382S, RSF601391S, RSF602499S, RSF602725S, RSF603136S, RSF603182S, RSF603186S, RSF603189S, RSF603196S, RSF603404S, RSF603412S, RSF603677S, RSF603968S, RSF603969S, RSF604025S, RSF604028S, RSF604124S, RSF604152S, RSF604196S, RSF604585S, RSF604643S, RSF604688S, RSF605565S, RSF608107S, RSF608108S, RSF603574S, RSF600946S, RSF602600S, RSF606202S, RSF606203S, RSF607093S, RSF607094S, RSF605687S, RSF605857S, RSF607095S, RSF607096S, RSF608071S, RSF600403S, RSF600607S, RSF600697S, RSF601328S, RSF602404S, RSF602441S, RSF602781S, RSF602835S, RSF603426S, RSF603702S, RSF603917S, RSF604019S, RSF604291S, RSF604489S, RSF604970S, RSF605022S, RSF605028S, RSF605047S, RSF605057S, RSF605059S, RSF605344S, RSF605347S, RSF605747S, RSF606296S, RSF601410S, RSF601416S, RSF601417S, RSF601419S, RSF602934S, RSF602935S, RSF603092S, RSF603097S, RSF603098S, RSF603102S, RSF603163S, RSF603165S, RSF603166S, RSF603851S, RSF603852S, RSF604030S, RSF604031S, RSF604033S, RSF604036S, RSF604042S, RSF604178S, RSF604179S, RSF604180S, RSF604182S, RSF604594S, RSF604602S, RSF604606S, RSF604608S, RSF604610S, RSF604613S, RSF604614S, RSF604615S, RSF604616S, RSF605100S, RSF605101S, RSF605102S, RSF605103S, RSF605115S, RSF605116S, RSF605117S, RSF605125S, RSF605692S, RSF605741S, RSF606013S, RSF606014S, RSF606015S, RSF606051S, RSF606289S, RSF606294S, RSF606295S, RSF606298S, RSF606299S, RSF606300S, RSF606301S, RSF606303S, RSF606304S, RSF606305S, RSF606306S, RSF606307S, RSF606728S, RSF606729S, RSF606731S, RSF606732S, RSF606733S, RSF606734S, RSF606735S, RSF606736S, RSF606737S, RSF606738S, RSF606775S, RSF605690S, RSF605720S, RSF600251S, RSF601411S, RSF604390S, RSF604391S, RSF604394S, RSF604398S, RSF604399S, RSF604220S, RSF604221S, RSF604578S, RSF604588S, RSF604659S, RSF606821S, RSF606849S, RSF606850S, RSF606852S, RSF606858S, RSF606860S, RSF606861S, RSF606875S, RSF607252S, RSF607253S, RSF607254S, RSF607255S, RSF607383S, RSF607384S, RSF607385S, RSF600968S, RSF604473S, RSF605677S, RSF607909S, RSF601535S, RSF603949S, RSF600456S, RSF600818S, RSF603152S, RSF603569S, RSF605993S, RSF605994S, RSF605995S, RSF605996S, RSF605997S, RSF606620S, RSF606621S, RSF606710S, RSF606711S, RSF606712S, RSF606714S, RSF606715S, RSF606716S, RSF606717S, RSF607082S, RSF607083S, RSF607084S, RSF607085S, RSF607086S, RSF607155S, RSF607169S, RSF601153S, RSF602186S, RSF602651S, RSF602763S, RSF602778S, RSF602933S, RSF602963S, RSF602965S, RSF604637S, RSF604639S, RSF604814S, RSF604942S, RSF604954S, RSF604958S, RSF605188S, RSF605190S, RSF605230S, RSF605231S, RSF605232S, RSF605233S, RSF605519S, RSF605521S, RSF605522S, RSF607153S, RSF607154S, RSF607156S, RSF607157S, RSF607158S, RSF607159S, RSF607160S, RSF607161S, RSF607162S, RSF607163S, RSF607164S, RSF607165S, RSF607166S, RSF607168S, RSF607170S, RSF607171S, RSF607172S, RSF607875S, RSF600177S, RSF600237S, RSF600238S, RSF601372S, RSF601480S, RSF601580S, RSF602797S, RSF602799S, RSF605092S, RSF605093S, RSF605270S, RSF605271S, RSF605288S, RSF605291S, RSF605295S, RSF605296S, RSF605300S, RSF605301S, RSF601105S, RSF600774S, RSF601065S, RSF601346S, RSF605688S, RSF600271S, RSF601113S, RSF601413S, RSF601415S, RSF601420S, RSF601421S, RSF601424S, RSF601428S, RSF602674S, RSF603114S, RSF603181S, RSF603183S, RSF603184S, RSF603432S, RSF603435S, RSF603436S, RSF604107S, RSF604109S, RSF604114S, RSF604119S, RSF604120S, RSF604122S, RSF604214S, RSF604215S, RSF604433S, RSF604436S, RSF604437S, RSF604584S, RSF604591S, RSF604592S, RSF604593S, RSF605084S, RSF605085S, RSF605097S, RSF605408S, RSF605411S, RSF605414S, RSF605415S, RSF605416S, RSF605417S, RSF606544S, RSF606551S, RSF606623S, RSF606624S, RSF606625S, RSF606626S, RSF606627S, RSF606633S, RSF606873S, RSF606876S, RSF607080S, RSF607081S, RSF607087S, RSF607140S, RSF607141S, RSF607142S, RSF607143S, RSF607144S, RSF607145S, RSF607146S, RSF607147S, RSF607148S, RSF607149S, RSF607150S, RSF607151S, RSF607152S, RSF605941S, RSF600208S, RSF600400S, RSF600528S, RSF600544S, RSF600585S, RSF600713S, RSF602350S, RSF603520S, RSF606589S, RSF606590S, RSF606022S, RSF606557S, RSF606558S, RSF606584S, RSF606599S, RSF606612S, RSF601488S, RSF602989S, RSF602992S, RSF602996S, RSF603000S, RSF603274S, RSF603372S, RSF604191S, RSF604192S, RSF604194S, RSF605098S, RSF606568S, RSF606596S, RSF606613S, RSF606614S, RSF606615S, RSF607103S, RSF607104S, RSF607105S, RSF607106S, RSF600371S, RSF603153S, RSF605705S, RSF605706S, RSF606540S, RSF606579S, RSF606581S, RSF606582S, RSF606616S, RSF606739S, RSF606742S, RSF600786S, RSF604466S, RSF604573S, RSF604720S, RSF605523S, RSF605524S, RSF606740S, RSF606750S, RSF606863S, RSF606970S, RSF606972S, RSF607175S, RSF607176S, RSF607177S, RSF607178S, RSF607179S, RSF607180S, RSF607181S, RSF607182S, RSF607183S, RSF607185S, RSF607186S, RSF607187S, RSF607189S, RSF607191S, RSF607193S, RSF601186S, RSF602739S, RSF603565S, RSF604204S, RSF604388S, RSF604718S, RSF600358S, RSF600247S, RSF603733S, RSF606016S, RSF606017S, RSF606018S, RSF606073S, RSF606074S, RSF606271S, RSF606272S, RSF606273S, RSF602685S, RSF604037S, RSF605031S, RSF605081S, RSF605082S, RSF605119S, RSF605120S, RSF605372S, RSF605515S, RSF605516S, RSF605517S, RSF607110S, RSF607111S, RSF607112S, RSF607113S, RSF607980S, RSF607981S, RSF607982S, RSF607983S, RSF602400S, RSF608017S, RSF600345S, RSF603518S, RSF603643S, RSF603229S, RSF605509S, RSF605510S, RSF605511S, RSF605514S, RSF606694S, RSF606695S, RSF606696S, RSF606697S, RSF606698S, RSF606699S, RSF606700S, RSF606701S, RSF606702S, RSF606703S, RSF607965S, RSF607966S, RSF607967S, RSF607969S, RSF605972S, RSF606602S, RSF603045S, RSF603046S, RSF606068S, RSF606071S, RSF604655S, RSF604656S, RSF604658S, RSF604662S, RSF604711S, RSF604717S, RSF604749S, RSF604762S, RSF604776S, RSF604786S, RSF604788S, RSF604793S, RSF604795S, RSF605421S, RSF605422S, RSF605423S, RSF605424S, RSF605425S, RSF605426S, RSF605427S, RSF605428S, RSF605429S, RSF605430S, RSF605431S, RSF605432S, RSF605433S, RSF605434S, RSF605435S, RSF606067S, RSF606069S, RSF606070S, RSF606072S, RSF606601S, RSF606603S, RSF606606S, RSF606608S, RSF606610S, RSF606611S, RSF605975S, RSF606066S, RSF605512S, RSF605707S, RSF605959S, RSF605966S, RSF606056S, RSF606057S, RSF606060S, RSF606061S, RSF606062S, RSF606063S, RSF606594S, RSF606595S, RSF606617S, RSF606618S, RSF601108S, RSF602603S, RSF602690S, RSF602761S, RSF602922S, RSF603664S, RSF604627S, RSF604630S, RSF604664S, RSF604865S, RSF604868S, RSF605090S, RSF605091S, RSF605118S, RSF605350S, RSF605351S, RSF605352S, RSF605364S, RSF605525S, RSF607091S, RSF607092S, RSF607102S, RSF607388S, RSF607389S, RSF607947S, RSF607948S, RSF607135S, RSF607136S, RSF607137S, RSF607138S, RSF607139S, RSF601287S, RSF601288S, RSF601289S, RSF601290S, RSF601291S, RSF601398S, RSF601400S, RSF601568S, RSF602213S, RSF602215S, RSF602277S, RSF602285S, RSF602292S, RSF602293S, RSF602294S, RSF602297S, RSF602299S, RSF602588S, RSF602590S, RSF602596S, RSF602615S, RSF602618S, RSF602620S, RSF602901S, RSF602903S, RSF602904S, RSF602910S, RSF602381S, RSF602383S, RSF605979S, RSF600158S, RSF600742S, RSF602759S, RSF603950S, RSF605943S, RSF600917S, RSF602714S, RSF603276S, RSF602728S, RSF602412S, RSF606196S, RSF601447S, RSF606224S, RSF600841S, RSF602356S, RSF602734S, RSF603542S, RSF604933S, RSF605982S, RSF600419S, RSF600848S, RSF601432S, RSF601550S, RSF602212S, RSF602847S, RSF603517S, RSF603545S, RSF603842S, RSF603995S, RSF604277S, RSF604293S, RSF604863S, RSF607242S, RSF607390S, RSF607395S, RSF600817S, RSF602931S, RSF606585S, RSF606587S, RSF606592S, RSF606593S, RSF600589S, RSF600621S, RSF601128S, RSF601184S, RSF601438S, RSF601483S, RSF603505S, RSF603603S, RSF601336S, RSF605998S, RSF605999S, RSF606003S, RSF606006S, RSF603654S, RSF603656S, RSF603657S, RSF603658S, RSF603934S, RSF603978S, RSF604133S, RSF604136S, RSF604139S, RSF604280S, RSF604282S, RSF604283S, RSF604290S, RSF604292S, RSF604326S, RSF604329S, RSF604352S, RSF604358S, RSF604359S, RSF604370S, RSF604381S, RSF604382S, RSF604463S, RSF604464S, RSF604595S, RSF604600S, RSF604601S, RSF604609S, RSF604611S, RSF604612S, RSF604619S, RSF604622S, RSF605060S, RSF605061S, RSF605062S, RSF605063S, RSF605064S, RSF605065S, RSF605066S, RSF605067S, RSF605068S, RSF605069S, RSF605152S, RSF605153S, RSF605166S, RSF605177S, RSF605187S, RSF605207S, RSF605208S, RSF605209S, RSF605210S, RSF605211S, RSF605212S, RSF605213S, RSF605214S, RSF605251S, RSF605252S, RSF605253S, RSF605374S, RSF605375S, RSF605376S, RSF605378S, RSF605380S, RSF605382S, RSF605384S, RSF605385S, RSF605386S, RSF605388S, RSF605389S, RSF605390S, RSF605391S, RSF605392S, RSF605393S, RSF605394S, RSF605397S, RSF605398S, RSF605399S, RSF605400S, RSF605401S, RSF605402S, RSF605403S, RSF605404S, RSF605405S, RSF606000S, RSF606001S, RSF606002S, RSF606004S, RSF606005S, RSF606007S, RSF606718S, RSF606719S, RSF606720S, RSF606721S, RSF606723S, RSF606724S, RSF606725S, RSF606726S, RSF606727S; GTIN 00763000178529. Lot Serial Numbers: RSF603289S, RSF604311S, RSF603296S, RSF604513S, RSF604636S, RSF604708S, RSF604731S, RSF604732S, RSF607535S, RSF606986S, RSF606930S, RSF604800S, RSF606355S, RSF606357S, RSF606360S, RSF606752S, RSF608362S, RSF608572S, RSF608592S, RSF608334S, RSF608571S, RSF608317S, RSF607840S, RSF607602S, RSF608610S, RSF608595S, RSF608358S, RSF606542S, RSF608300S, RSF608109S, RSF605448S, RSF606389S, RSF606505S, RSF607428S, RSF607849S, RSF608298S, RSF600035S, RSF602270S, RSF602272S, RSF602298S, RSF602438S, RSF602442S, RSF602945S, RSF602946S, RSF603298S, RSF605280S, RSF605462S, RSF605582S, RSF606376S, RSF607220S, RSF607618S, RSF608324S, RSF608355S, RSF608577S, RSF608602S, RSF600028S, RSF600031S, RSF600081S, RSF600093S, RSF600117S, RSF600033S, RSF606388S, RSF606449S, RSF606450S, RSF602252S, RSF603291S, RSF605259S, RSF605825S, RSF604809S, RSF605882S, RSF604740S, RSF604742S, RSF605254S, RSF605261S, RSF605864S, RSF606798S, RSF606868S, RSF606117S, RSF606118S, RSF606807S, RSF606810S, RSF607856S, RSF607925S, RSF608213S, RSF608216S, RSF608259S, RSF603306S, RSF604631S, RSF604805S, RSF604908S, RSF606382S, RSF607010S, RSF607994S, RSF600023S, RSF601884S, RSF602063S, RSF602384S, RSF603304S, RSF604342S, RSF604526S, RSF605353S, RSF605354S, RSF605780S, RSF605793S, RSF605808S, RSF605888S, RSF606387S, RSF606649S, RSF606839S, RSF607281S, RSF607619S, RSF607775S, RSF602180S, RSF602182S, RSF605149S, RSF601867S, RSF603312S, RSF604638S, RSF605132S, RSF601423S, RSF605769S, RSF607839S, RSF605711S, RSF607184S, RSF607579S, RSF602200S, RSF605577S, RSF605584S, RSF605828S, RSF605831S, RSF606089S, RSF606575S, RSF607319S, RSF603281S, RSF603915S, RSF606838S, RSF608149S, RSF602358S, RSF603293S, RSF603294S, RSF604084S, RSF604355S, RSF604451S, RSF604540S, RSF604547S, RSF605306S, RSF605313S, RSF605613S, RSF605743S, RSF606094S, RSF606647S, RSF603893S, RSF605139S, RSF605141S, RSF601865S, RSF601874S, RSF601877S, RSF601882S, RSF602064S, RSF602065S, RSF602066S, RSF602204S, RSF602206S, RSF602209S, RSF603563S, RSF604301S, RSF605142S, RSF605154S, RSF605155S, RSF605776S, RSF605777S, RSF605778S, RSF607580S, RSF607581S, RSF607582S, RSF602211S, RSF603288S, RSF603290S, RSF604936S, RSF605725S, RSF606459S, RSF603080S, RSF603252S, RSF603286S, RSF604386S, RSF604551S, RSF604808S, RSF605160S, RSF605164S, RSF605779S, RSF602154S, RSF600045S, RSF602004S, RSF602183S, RSF602351S, RSF604410S, RSF606448S, RSF607032S, RSF607586S, RSF607587S, RSF607757S, RSF608356S, RSF600072S, RSF606488S, RSF606814S, RSF606815S, RSF606820S, RSF606983S, RSF607188S, RSF607616S, RSF607838S, RSF600103S, RSF604491S, RSF605131S, RSF605341S, RSF600024S, RSF602007S, RSF602203S, RSF602237S, RSF604016S, RSF605325S, RSF605453S, RSF605822S, RSF606082S, RSF607323S, RSF608342S, RSF608332S, RSF604047S, RSF605848S, RSF606076S, RSF606135S, RSF606779S, RSF607445S, RSF607707S, RSF602005S, RSF602006S, RSF602102S, RSF602104S, RSF602196S, RSF602248S, RSF602471S, RSF602530S, RSF602532S, RSF603169S, RSF603339S, RSF604351S, RSF604403S, RSF604535S, RSF604694S, RSF604816S, RSF604878S, RSF604894S, RSF604899S, RSF605180S, RSF605583S, RSF605608S, RSF605823S, RSF606088S, RSF606378S, RSF606819S, RSF606851S, RSF606853S, RSF607816S, RSF608279S, RSF608316S, RSF602312S, RSF603208S, RSF603209S, RSF603302S, RSF603850S, RSF603853S, RSF603895S, RSF606456S, RSF606457S, RSF604313S, RSF603282S, RSF603847S, RSF603848S, RSF602121S, RSF602122S, RSF602185S, RSF602222S, RSF602184S, RSF603301S, RSF602146S, RSF602480S, RSF603205S, RSF603236S, RSF603246S, RSF603913S, RSF604043S, RSF604309S, RSF604361S, RSF604517S, RSF604564S, RSF604565S, RSF604734S, RSF604810S, RSF604897S, RSF604950S, RSF605314S, RSF605622S, RSF605760S, RSF606116S, RSF606543S, RSF606548S, RSF606586S, RSF606657S, RSF602067S, RSF602068S, RSF602096S, RSF602097S, RSF602243S, RSF602244S, RSF602266S, RSF602271S, RSF602290S, RSF602300S, RSF602301S, RSF602306S, RSF602307S, RSF602319S, RSF602491S, RSF602495S, RSF602496S, RSF603315S, RSF604314S, RSF604315S, RSF604362S, RSF604363S, RSF604529S, RSF604530S, RSF604567S, RSF604568S, RSF604710S, RSF604991S, RSF604992S, RSF602029S, RSF602030S, RSF602069S, RSF602175S, RSF602340S, RSF602506S, RSF602507S, RSF602533S, RSF603342S, RSF604045S, RSF604441S, RSF604447S, RSF604478S, RSF604521S, RSF604727S, RSF604896S, RSF605294S, RSF605358S, RSF605609S, RSF607311S, RSF604368S, RSF604558S, RSF604975S, RSF605567S, RSF605739S, RSF605827S, RSF607313S, RSF607442S, RSF607558S, RSF607841S, RSF608315S, RSF600026S, RSF604350S, RSF600111S, RSF600113S, RSF600126S, RSF602012S, RSF602060S, RSF602436S, RSF602948S, RSF605356S, RSF606803S, RSF606866S, RSF606919S, RSF607467S, RSF607482S, RSF602256S, RSF602171S, RSF602949S, RSF604374S, RSF604392S, RSF604825S, RSF600036S, RSF600038S, RSF600097S, RSF600121S, RSF602089S, RSF602255S, RSF602260S, RSF602445S, RSF602531S, RSF603016S, RSF603055S, RSF603061S, RSF604811S, RSF606402S, RSF606805S, RSF606836S, RSF602387S, RSF604909S, RSF604928S, RSF604955S, RSF605768S, RSF608352S, RSF606136S, RSF602041S, RSF602075S, RSF602076S, RSF602077S, RSF602078S, RSF602105S, RSF602149S, RSF602151S, RSF603195S, RSF603203S, RSF603267S, RSF603295S, RSF603297S, RSF603310S, RSF603321S, RSF603816S, RSF603912S, RSF603980S, RSF604005S, RSF604083S, RSF604086S, RSF604101S, RSF604335S, RSF604444S, RSF604484S, RSF604534S, RSF604586S, RSF604703S, RSF604704S, RSF604784S, RSF604785S, RSF602091S, RSF602092S, RSF602125S, RSF602126S, RSF602191S, RSF602227S, RSF602251S, RSF602475S, RSF603243S, RSF603311S, RSF603317S, RSF603318S, RSF604373S, RSF604559S, RSF604563S, RSF604739S, RSF604852S, RSF604860S, RSF604983S, RSF605323S, RSF605454S, RSF606126S, RSF606405S, RSF606816S, RSF607037S, RSF607312S, RSF607453S, RSF607573S, RSF607848S, RSF607993S, RSF608291S, RSF602348S, RSF604295S, RSF605862S, RSF602148S, RSF603212S, RSF603347S, RSF604349S, RSF604453S, RSF604522S, RSF604541S, RSF604875S, RSF605444S, RSF605770S, RSF605802S, RSF606130S, RSF606496S, RSF607455S, RSF607803S, RSF607991S, RSF608325S, RSF600077S, RSF602045S, RSF603180S, RSF602014S, RSF602235S, RSF603232S, RSF605787S, RSF606797S, RSF607531S, RSF607534S, RSF600076S, RSF603079S, RSF605591S, RSF606794S, RSF602010S, RSF602021S, RSF603316S, RSF604623S, RSF602074S, RSF601426S, RSF601450S, RSF602053S, RSF604353S, RSF604395S, RSF604867S, RSF604882S, RSF604937S, RSF605184S, RSF605289S, RSF605317S, RSF605572S, RSF606108S, RSF606129S, RSF606549S, RSF607002S, RSF607569S, RSF602136S, RSF602138S, RSF602139S, RSF602229S, RSF602242S, RSF602540S, RSF603202S, RSF603204S, RSF603237S, RSF603805S, RSF604035S, RSF604321S, RSF604331S, RSF604408S, RSF604435S, RSF604531S, RSF604536S, RSF604587S, RSF604607S, RSF604799S, RSF604918S, RSF604929S, RSF605242S, RSF605249S, RSF605318S, RSF605326S, RSF605458S, RSF605578S, RSF605773S, RSF606098S, RSF606381S, RSF606476S, RSF606898S, RSF602114S, RSF604360S, RSF606106S, RSF606407S, RSF606645S, RSF607432S, RSF607544S, RSF607818S, RSF604532S, RSF602107S, RSF602108S, RSF602219S, RSF602305S, RSF603253S, RSF604919S, RSF604927S, RSF605539S, RSF605587S, RSF605588S, RSF606545S, RSF606576S, RSF607229S, RSF607328S, RSF607474S, RSF602017S, RSF602084S, RSF602101S, RSF602284S, RSF603082S, RSF604550S, RSF604932S, RSF605176S, RSF605396S, RSF605576S, RSF606081S, RSF606115S, RSF606656S, RSF607318S, RSF607487S, RSF607559S, RSF607607S, RSF607612S, RSF602308S, RSF604877S, RSF602086S, RSF602087S, RSF602336S, RSF605171S, RSF605476S, RSF605513S, RSF606124S, RSF607449S, RSF602483S, RSF604354S, RSF605612S, RSF602254S, RSF602257S, RSF602258S, RSF602259S, RSF602474S, RSF602498S, RSF602508S, RSF602523S, RSF603925S, RSF603927S, RSF604312S, RSF604514S, RSF604537S, RSF604874S, RSF604916S, RSF604917S, RSF604920S, RSF604921S, RSF605181S, RSF605310S, RSF605381S, RSF605455S, RSF605606S, RSF605775S, RSF605803S, RSF606128S, RSF607430S, RSF606499S, RSF607434S, RSF602027S, RSF602031S, RSF602033S, RSF602034S, RSF602057S, RSF602127S, RSF602133S, RSF602224S, RSF602302S, RSF603162S, RSF603307S, RSF603341S, RSF604039S, RSF604338S, RSF604343S, RSF604380S, RSF604397S, RSF604402S, RSF604420S, RSF604527S, RSF605167S, RSF605601S, RSF605602S, RSF606679S, RSF606855S, RSF606856S, RSF607309S, RSF607444S, RSF607456S, RSF607480S, RSF607481S, RSF602129S, RSF602226S, RSF602249S, RSF602250S, RSF602346S, RSF602528S, RSF603164S, RSF603303S, RSF604345S, RSF605311S, RSF602026S, RSF602072S, RSF602073S, RSF602162S, RSF602283S, RSF604925S, RSF605520S, RSF606377S, RSF607677S, RSF608154S, RSF604525S, RSF602009S, RSF602079S, RSF602080S, RSF602082S, RSF602144S, RSF602178S, RSF602223S, RSF602263S, RSF602276S, RSF602303S, RSF602333S, RSF602338S, RSF602514S, RSF602516S, RSF603171S, RSF603279S, RSF603280S, RSF603308S, RSF603981S, RSF604006S, RSF604015S, RSF604296S, RSF604297S, RSF604306S, RSF604324S, RSF604328S, RSF604367S, RSF604406S, RSF604407S, RSF604465S, RSF604509S, RSF604523S, RSF604524S, RSF604539S, RSF604589S, RSF604617S, RSF604745S, RSF604771S, RSF604779S, RSF604857S, RSF604858S, RSF604869S, RSF604895S, RSF604938S, RSF604948S, RSF604953S, RSF604969S, RSF604984S, RSF605287S, RSF605395S, RSF605456S, RSF605457S, RSF605495S, RSF605599S, RSF605614S, RSF605621S, RSF605731S, RSF605767S, RSF606104S, RSF606125S, RSF606134S, RSF606383S, RSF606468S, RSF606547S, RSF606655S, RSF606872S, RSF607282S, RSF607435S, RSF607454S, RSF607555S, RSF607678S, RSF607709S, RSF607710S, RSF602128S, RSF602225S, RSF602460S, RSF603292S, RSF603344S, RSF604508S, RSF605729S, RSF603923S, RSF605480S, RSF605483S, RSF605771S, RSF605935S, RSF606122S, RSF606123S, RSF606501S, RSF606507S, RSF606508S, RSF606509S, RSF606813S, RSF607831S, RSF607194S, RSF607204S, RSF607479S, RSF607503S, RSF607575S, RSF607576S, RSF608170S, RSF608182S, RSF608190S, RSF608603S, RSF608611S, RSF608612S, RSF608623S, RSF608310S, RSF608311S, RSF602051S, RSF602052S, RSF602095S, RSF602112S, RSF602124S, RSF602197S, RSF604754S, RSF604907S, RSF604922S, RSF604924S, RSF605345S, RSF605492S, RSF605785S, RSF606080S, RSF606105S, RSF606546S, RSF606639S, RSF606741S, RSF606802S, RSF607260S, RSF607485S, RSF607556S, RSF608301S, RSF607812S, RSF607813S, RSF602291S, RSF602295S, RSF602943S, RSF603984S, RSF604913S, RSF606091S, RSF606114S, RSF607298S, RSF607299S, RSF607300S, RSF607301S, RSF607547S, RSF607548S, RSF607549S, RSF607550S, RSF608624S, RSF604505S, RSF600106S, RSF601871S, RSF602172S, RSF602304S, RSF604580S, RSF604893S, RSF605806S, RSF606397S, RSF606458S, RSF607264S, RSF607722S, RSF601349S, RSF602083S, RSF602232S, RSF602239S, RSF602289S, RSF602309S, RSF602509S, RSF604356S, RSF604729S, RSF604851S, RSF604931S, RSF605460S, RSF605531S, RSF605579S, RSF605580S, RSF605597S, RSF605786S, RSF606084S, RSF606085S, RSF606127S, RSF606400S, RSF606473S, RSF606641S, RSF606865S, RSF606904S, RSF607262S, RSF607450S, RSF607451S, RSF607517S, RSF607521S, RSF607608S, RSF607799S, RSF608343S, RSF602238S, RSF603272S, RSF602192S, RSF602193S, RSF602194S, RSF602313S, RSF602470S, RSF603074S, RSF604538S, RSF604699S, RSF604702S, RSF605240S, RSF605309S, RSF606095S, RSF606096S, RSF606399S, RSF606597S, RSF606654S, RSF607219S, RSF607512S, RSF607622S, RSF608282S, RSF603159S, RSF604038S, RSF605130S, RSF602240S, RSF602241S, RSF603270S, RSF604518S, RSF605255S, RSF602247S, RSF602450S, RSF604640S, RSF604844S, RSF600041S, RSF600069S, RSF600073S, RSF602025S, RSF603340S, RSF605447S, RSF607496S, RSF600080S, RSF602044S, RSF602109S, RSF602110S, RSF602166S, RSF602210S, RSF602231S, RSF602280S, RSF602281S, RSF602426S, RSF602477S, RSF602478S, RSF602481S, RSF602484S, RSF602944S, RSF603314S, RSF604515S, RSF604544S, RSF604552S, RSF604599S, RSF605766S, RSF602311S, RSF604546S, RSF605620S, RSF602019S, RSF605174S, RSF606398S, RSF602088S, RSF602090S, RSF602465S, RSF603197S, RSF603200S, RSF603313S, RSF603322S, RSF604336S, RSF604467S, RSF604577S, RSF604605S, RSF604842S, RSF604866S, RSF604879S, RSF605297S, RSF605446S, RSF605830S, RSF606474S, RSF606484S, RSF606791S, RSF600027S, RSF600119S, RSF602268S, RSF602458S, RSF602899S, RSF603015S, RSF607578S, RSF600032S, RSF602431S, RSF602443S, RSF604801S, RSF605134S, RSF605256S, RSF606119S, RSF606527S, RSF607784S, RSF606451S, RSF607514S, RSF604738S, RSF606471S, RSF607306S, RSF607490S, RSF608359S, RSF607706S, RSF606452S, RSF600098S, RSF602120S, RSF603343S, RSF604859S, RSF604903S, RSF604985S, RSF605383S, RSF605475S, RSF605740S, RSF605824S, RSF606854S, RSF607513S, RSF607990S, RSF608283S, RSF602062S, RSF601544S, RSF603319S, RSF602020S, RSF602278S, RSF606533S, RSF601461S, RSF602035S, RSF602055S, RSF602056S, RSF602140S, RSF603269S, RSF604079S, RSF604457S, RSF604583S, RSF604887S, RSF605185S, RSF605387S, RSF605832S, RSF606109S, RSF607278S, RSF600055S, RSF600104S, RSF600105S, RSF602111S, RSF602152S, RSF603174S, RSF604303S, RSF604337S, RSF604553S, RSF604566S, RSF604934S, RSF605600S, RSF605755S, RSF605759S, RSF606077S, RSF606107S, RSF606493S, RSF606640S, RSF607433S, RSF607545S, RSF607750S, RSF603172S, RSF604000S, RSF604002S, RSF604783S, RSF605440S, RSF605470S, RSF605586S, RSF605805S, RSF606801S, RSF607322S, RSF604512S, RSF605257S, RSF607335S, RSF602022S, RSF602024S, RSF602059S, RSF602134S, RSF602135S, RSF602236S, RSF602316S, RSF603073S, RSF603158S, RSF603299S, RSF603305S, RSF604365S, RSF604396S, RSF604434S, RSF604548S, RSF604549S, RSF604872S, RSF605312S, RSF605571S, RSF605833S, RSF606446S, RSF607606S, RSF604794S, RSF600050S, RSF600051S, RSF601459S, RSF601873S, RSF602011S, RSF602048S, RSF602093S, RSF602142S, RSF602190S, RSF602228S, RSF602473S, RSF602485S, RSF603167S, RSF603275S, RSF603323S, RSF603937S, RSF604081S, RSF604506S, RSF604519S, RSF604712S, RSF604873S, RSF605178S, RSF605277S, RSF605283S, RSF605496S, RSF605607S, RSF605727S, RSF605789S, RSF606131S, RSF606447S, RSF606660S, RSF606857S, RSF606900S, RSF607438S, RSF607475S, RSF607819S, RSF608165S, RSF608295S, RSF602321S, RSF604946S, RSF605452S, RSF606793S, RSF607215S, RSF607712S, RSF607785S, RSF602322S, RSF602330S, RSF602459S, RSF602543S, RSF603345S, RSF603929S, RSF603931S, RSF604366S, RSF604511S, RSF605250S, RSF605285S, RSF605574S, RSF605826S, RSF605860S, RSF606093S, RSF606598S, RSF607314S, RSF607315S, RSF607620S, RSF607820S, RSF608159S, RSF608330S, RSF608339S, RSF604498S, RSF604757S, RSF604797S, RSF604798S, RSF604911S, RSF604912S, RSF605127S, RSF605128S, RSF605471S, RSF605472S, RSF605473S, RSF605537S, RSF605540S, RSF605563S, RSF605807S, RSF606090S, RSF606369S, RSF606370S, RSF606403S, RSF606404S, RSF606453S, RSF606454S, RSF606455S, RSF607524S, RSF607525S, RSF607782S, RSF608615S, RSF608616S, RSF607013S, RSF608197S, RSF608202S, RSF608272S, RSF608273S, RSF608274S, RSF608288S, RSF608289S, RSF608290S, RSF608335S, RSF608336S, RSF608360S, RSF608361S, RSF608617S, RSF606846S, RSF604624S, RSF600070S, RSF600074S, RSF600102S, RSF601425S, RSF602071S, RSF602098S, RSF602264S, RSF603257S, RSF603348S, RSF604357S, RSF604885S, RSF605593S, RSF605603S, RSF606075S, RSF600013S, RSF600016S, RSF600017S, RSF600018S, RSF600019S, RSF600058S, RSF600059S, RSF600060S, RSF600061S, RSF600062S, RSF600064S, RSF600065S, RSF600066S, RSF602036S, RSF602037S, RSF602038S, RSF602039S, RSF602040S, RSF602331S, RSF602332S, RSF602335S, RSF603258S, RSF603264S, RSF604780S, RSF604781S, RSF604782S, RSF604791S, RSF604812S, RSF604813S, RSF604815S, RSF604817S, RSF604818S, RSF604819S, RSF606142S, RSF606143S, RSF606144S, RSF606145S, RSF606216S, RSF606465S, RSF606556S, RSF606619S, RSF606804S, RSF606808S, RSF606809S, RSF606840S, RSF606841S, RSF606842S, RSF606843S, RSF606844S, RSF606874S, RSF606893S, RSF607526S, RSF607527S, RSF607528S, RSF607529S, RSF607530S, RSF608597S, RSF608598S, RSF608599S, RSF608600S, RSF608601S, RSF600128S, RSF602049S, RSF602050S, RSF602061S, RSF602131S, RSF602143S, RSF602314S, RSF602315S, RSF602344S, RSF600048S, RSF602116S, RSF602117S, RSF602118S, RSF602119S, RSF602195S, RSF602202S, RSF602287S, RSF602456S, RSF603283S, RSF603284S, RSF604300S, RSF604653S, RSF604820S, RSF605137S, RSF605157S, RSF605158S, RSF605315S, RSF605477S, RSF605616S, RSF605788S, RSF606146S, RSF606147S, RSF606391S, RSF606392S, RSF606567S, RSF606634S, RSF606835S, RSF607206S, RSF607216S, RSF607505S, RSF607511S, RSF603285S, RSF605129S, RSF605371S, RSF605794S, RSF605940S, RSF605944S, RSF606141S, RSF606502S, RSF606799S, RSF607190S, RSF607508S, RSF602047S, RSF603346S, RSF604423S, RSF604935S, RSF606393S, RSF606848S, RSF607842S, RSF602328S, RSF604572S, RSF604642S, RSF604722S, RSF604725S, RSF604824S, RSF604849S, RSF605373S, RSF605377S, RSF605449S, RSF605450S, RSF606120S, RSF606121S, RSF606386S, RSF606461S, RSF606570S, RSF606902S, RSF606918S, RSF606951S, RSF607466S, RSF607585S, RSF607597S, RSF604333S, RSF604334S, RSF604422S, RSF604556S, RSF604557S, RSF604629S, RSF604683S, RSF604803S, RSF604804S, RSF604834S, RSF604994S, RSF605361S, RSF605362S, RSF605367S, RSF605592S, RSF605784S, RSF605795S, RSF605939S, RSF606078S, RSF606103S, RSF606139S, RSF606554S, RSF606555S, RSF606564S, RSF606982S, RSF607277S, RSF607463S, RSF607484S, RSF607510S, RSF607623S, RSF600101S, RSF603941S, RSF602453S, RSF602454S, RSF603230S, RSF604308S, RSF600084S, RSF604298S, RSF604504S, RSF602013S, RSF602310S, RSF604902S, RSF605165S, RSF606380S, RSF600078S, RSF603300S, RSF604528S, RSF602130S, RSF604923S, RSF605801S, RSF606112S, RSF602489S, RSF605568S, RSF605469S, RSF602153S, RSF602158S, RSF603206S, RSF603309S, RSF604393S, RSF604516S, RSF604880S, RSF605445S, RSF605863S, RSF606531S, RSF606806S, RSF602054S, RSF602100S, RSF604943S, RSF605598S, RSF605834S, RSF602947S, RSF603239S, RSF604427S, RSF604462S, RSF604561S, RSF604719S, RSF604721S, RSF604765S, RSF604802S, RSF604827S, RSF604829S, RSF604830S, RSF604840S, RSF604914S, RSF605136S, RSF605156S, RSF605368S, RSF605451S, RSF605468S, RSF605478S, RSF605543S, RSF605604S, RSF605605S, RSF605615S, RSF605618S, RSF605792S, RSF605934S, RSF606079S, RSF606460S, RSF606529S, RSF606571S, RSF607207S, RSF607470S, RSF607486S, RSF607495S, RSF600079S, RSF600085S, RSF600092S, RSF605590S, RSF605712S, RSF606795S, RSF606796S, RSF607532S, RSF600082S, RSF600107S, RSF600124S, RSF602081S, RSF606290S, RSF606351S, RSF606353S, RSF604533S, RSF604560S, RSF604562S, RSF604747S, RSF604760S, RSF605819S, RSF605829S, RSF605850S, RSF605874S, RSF605877S, RSF605890S, RSF606333S, RSF608247S, RSF608557S, RSF608564S, RSF608568S, RSF606334S, RSF606336S, RSF606338S, RSF606345S, RSF606348S, RSF606366S, RSF607747S, RSF607759S, RSF607792S, RSF606342S, RSF606358S, RSF606359S, RSF600052S, RSF603325S, RSF603844S, RSF603846S, RSF604320S, RSF604432S, RSF605800S, RSF605816S, RSF606100S, RSF606466S, RSF607459S, RSF607811S, RSF608023S, RSF604372S, RSF607469S, RSF600037S, RSF600112S, RSF600118S, RSF602099S, RSF602296S, RSF600039S, RSF600100S, RSF600108S, RSF600110S, RSF602103S, RSF602233S, RSF602323S, RSF602324S, RSF603037S, RSF604340S, RSF604369S, RSF604371S, RSF604418S, RSF604761S, RSF605264S, RSF605332S, RSF607783S, RSF605809S, RSF605811S, RSF605812S, RSF605813S, RSF605814S, RSF602230S, RSF602446S, RSF602502S, RSF606530S, RSF607779S, RSF608275S, RSF608276S, RSF602950S, RSF603072S, RSF604411S, RSF605135S, RSF605143S, RSF605263S, RSF605330S, RSF606113S, RSF607557S, RSF606086S, RSF607504S, RSF605169S, RSF605623S, RSF600134S, RSF600136S, RSF600138S, RSF600145S, RSF602137S, RSF602141S, RSF602145S, RSF602147S, RSF602150S, RSF602157S, RSF602361S, RSF602366S, RSF602374S, RSF602388S, RSF602389S, RSF602390S, RSF602391S, RSF602394S, RSF602418S, RSF602419S, RSF602422S, RSF602994S, RSF602997S, RSF603009S, RSF603083S, RSF604095S, RSF604346S, RSF605159S, RSF605172S, RSF605637S, RSF606428S, RSF606436S, RSF606439S, RSF606818S, RSF606917S, RSF606920S, RSF606924S, RSF606929S, RSF606937S, RSF606938S, RSF600131S, RSF600132S, RSF601187S, RSF601453S, RSF602058S, RSF602113S, RSF602123S, RSF602376S, RSF602378S, RSF602379S, RSF602409S, RSF602482S, RSF602956S, RSF602977S, RSF602999S, RSF603007S, RSF603008S, RSF603014S, RSF605138S, RSF605148S, RSF605161S, RSF605624S, RSF605630S, RSF605631S, RSF605632S, RSF605633S, RSF606445S, RSF606906S, RSF606910S, RSF606911S, RSF606936S, RSF602500S, RSF606372S, RSF606514S, RSF605479S, RSF602898S, RSF604853S, RSF602447S, RSF606467S, RSF602188S, RSF602448S, RSF605182S, RSF605247S, RSF605481S, RSF605482S, RSF606371S, RSF604862S, RSF602493S, RSF604884S, RSF602894S, RSF602245S, RSF605183S, RSF600139S, RSF600141S, RSF600144S, RSF602115S, RSF602155S, RSF602363S, RSF602368S, RSF602370S, RSF602382S, RSF602405S, RSF602408S, RSF602417S, RSF603006S, RSF605173S, RSF606914S, RSF606928S, RSF600688S, RSF602018S, RSF602028S, RSF602032S, RSF602420S, RSF602421S, RSF605162S, RSF605625S, RSF605626S, RSF605628S, RSF605629S, RSF605635S, RSF605640S, RSF606925S, RSF606927S, RSF605634S, RSF605638S, RSF605636S, RSF606833S, RSF607541S, RSF605293S, RSF607592S, RSF605140S, RSF605879S, RSF606235S, RSF606283S, RSF606287S, RSF606288S, RSF606349S, RSF606350S, RSF606352S, RSF606354S, RSF603287S, RSF600129S, RSF600137S, RSF603328S, RSF605839S, RSF605840S, RSF605841S, RSF605842S, RSF606786S, RSF606787S, RSF606789S, RSF606790S, RSF606959S, RSF606960S, RSF606961S, RSF606962S, RSF608514S, RSF608515S, RSF608516S, RSF608517S, RSF608518S, RSF608519S, RSF608520S, RSF608521S, RSF608522S, RSF608523S, RSF604317S, RSF605290S, RSF600120S, RSF602337S, RSF602362S, RSF604377S, RSF604378S, RSF604412S, RSF604415S, RSF604416S, RSF604417S, RSF604510S, RSF604555S, RSF604569S, RSF604570S, RSF604571S, RSF604692S, RSF604750S, RSF604753S, RSF604756S, RSF604832S, RSF604846S, RSF604888S, RSF604945S, RSF605144S, RSF605145S, RSF605146S, RSF605245S, RSF605279S, RSF605316S, RSF605320S, RSF605331S, RSF605334S, RSF605337S, RSF605355S, RSF605365S, RSF605465S, RSF605499S, RSF605501S, RSF605595S, RSF605596S, RSF605783S, RSF605797S, RSF605798S, RSF605818S, RSF605820S, RSF605895S, RSF605915S, RSF605932S, RSF605937S, RSF605938S, RSF606101S, RSF606102S, RSF606138S, RSF606463S, RSF606561S, RSF606562S, RSF606635S, RSF606636S, RSF606788S, RSF606823S, RSF606824S, RSF606825S, RSF606826S, RSF606827S, RSF606847S, RSF606941S, RSF607001S, RSF607005S, RSF607213S, RSF607214S, RSF607425S, RSF607426S, RSF607460S, RSF607461S, RSF607462S, RSF607509S, RSF607553S, RSF607563S, RSF607762S, RSF607770S, RSF607773S, RSF607774S, RSF607823S, RSF607829S, RSF607850S, RSF608287S, RSF600057S, RSF600086S, RSF601407S, RSF601543S, RSF602016S, RSF602106S, RSF602156S, RSF602218S, RSF602274S, RSF602288S, RSF602318S, RSF602451S, RSF602469S, RSF603004S, RSF603188S, RSF604044S, RSF604307S, RSF604379S, RSF604494S, RSF604545S, RSF604743S, RSF604876S, RSF604939S, RSF604949S, RSF605163S, RSF605175S, RSF605278S, RSF605459S, RSF605619S, RSF605733S, RSF606087S, RSF607267S, RSF607324S, RSF607452S, RSF607611S, RSF607699S, RSF607855S, RSF608134S, RSF606363S, RSF606364S, RSF605244S, RSF600034S, RSF600043S, RSF600044S, RSF601868S, RSF602015S, RSF602132S, RSF602214S, RSF602342S, RSF602455S, RSF603161S, RSF603983S, RSF604348S, RSF604687S, RSF604981S, RSF605370S, RSF605379S, RSF605796S, RSF605873S, RSF606097S, RSF606572S, RSF606573S, RSF606111S, RSF600020S, RSF600021S, RSF600022S, RSF600025S, RSF600029S, RSF600030S, RSF600067S, RSF600068S, RSF602003S, RSF602008S, RSF602042S, RSF602043S, RSF602046S, RSF602094S, RSF602170S, RSF602279S, RSF602286S, RSF603160S, RSF603211S, RSF603277S, RSF604294S, RSF605843S, RSF605845S, RSF605846S, RSF605847S, RSF603335S, RSF606953S, RSF606957S, RSF607686S, RSF607687S, RSF607689S, RSF607690S, RSF607691S, RSF607692S, RSF607694S, RSF607695S, RSF607696S, RSF605754S, RSF605757S, RSF605761S, RSF605762S, RSF605763S, RSF606148S, RSF606149S, RSF606150S, RSF606151S, RSF606152S, RSF606153S, RSF606154S, RSF606155S, RSF606156S, RSF606157S, RSF606158S, RSF606159S, RSF606160S, RSF606161S, RSF606162S, RSF606163S, RSF606164S, RSF606165S, RSF606166S, RSF606167S, RSF606168S, RSF606412S, RSF606413S, RSF606414S, RSF606415S, RSF606416S, RSF606417S, RSF606418S, RSF606419S, RSF606420S, RSF606421S, RSF606422S, RSF606423S, RSF606424S, RSF606425S, RSF606426S, RSF606427S, RSF606429S, RSF606430S, RSF606431S, RSF606432S, RSF606433S, RSF606434S, RSF606435S, RSF605764S, RSF600109S, RSF600114S, RSF600122S, RSF606410S, RSF603336S, RSF603337S, RSF604058S, RSF603329S, RSF603330S, RSF603331S, RSF603332S, RSF603333S, RSF603334S, RSF604051S, RSF604065S, RSF605548S, RSF606774S, RSF606778S, RSF606780S, RSF606783S, RSF606784S, RSF606785S, RSF608224S, RSF607660S, RSF608222S, RSF605714S, RSF605715S, RSF605716S, RSF605717S, RSF605718S, RSF605719S, RSF605721S, RSF605722S, RSF605723S, RSF605724S, RSF605728S, RSF605730S, RSF605732S, RSF605734S, RSF605735S, RSF605736S, RSF605738S, RSF605742S, RSF605744S, RSF605745S, RSF605746S, RSF605748S, RSF605749S, RSF605750S, RSF605751S, RSF605752S, RSF605836S, RSF605837S, RSF605838S, RSF606411S, RSF606437S, RSF606438S, RSF606440S, RSF606441S, RSF606442S, RSF606443S, RSF606444S, RSF606713S, RSF606722S, RSF606730S, RSF606743S, RSF606747S, RSF606749S, RSF606964S, RSF606965S, RSF606969S, RSF606971S, RSF606973S, RSF606974S, RSF606975S, RSF606976S, RSF606977S, RSF606978S, RSF606980S, RSF606981S, RSF607261S, RSF607329S, RSF607414S, RSF607518S, RSF607519S, RSF607520S, RSF607621S, RSF607624S, RSF607625S, RSF607626S, RSF607627S, RSF607628S, RSF607629S, RSF607636S, RSF607647S, RSF607649S, RSF607650S, RSF607651S, RSF607658S, RSF607659S, RSF607663S, RSF607664S, RSF607665S, RSF607666S, RSF607667S, RSF607669S, RSF607670S, RSF607671S, RSF607672S, RSF607673S, RSF607674S, RSF607679S, RSF607680S, RSF607681S, RSF607682S, RSF607683S, RSF607684S, RSF607685S, RSF607845S, RSF607846S, RSF608226S, RSF608235S, RSF608238S, RSF608242S, RSF608468S, RSF608469S, RSF608470S, RSF608471S, RSF608472S, RSF608474S, RSF608487S, RSF608488S, RSF608490S, RSF608491S, RSF603338S, RSF604736S, RSF604691S, RSF604735S, RSF606083S, RSF606099S, RSF606385S, RSF607570S, RSF600008S, RSF600009S, RSF600011S, RSF600014S, RSF600099S, RSF602282S, RSF603215S, RSF603217S, RSF603219S, RSF603228S, RSF607985S, RSF607986S, RSF607987S, RSF607988S, RSF607989S, RSF605594S, RSF607675S, RSF600071S, RSF602273S, RSF603324S, RSF605817S, RSF606137S, RSF606464S, RSF606566S, RSF606607S, RSF607591S, RSF600042S, RSF605168S, RSF603326S, RSF603327S, RSF604332S, RSF604341S, RSF604364S, RSF605281S, RSF601205S, RSF603320S, RSF603839S, RSF603841S, RSF604305S, RSF604404S, RSF604405S, RSF604848S, RSF605170S, RSF605298S, RSF605299S, RSF605357S, RSF605360S, RSF605466S, RSF605815S, RSF606462S, RSF606792S, RSF606859S, RSF606864S, RSF607551S, RSF604040S, RSF605282S, RSF605286S, RSF605319S, RSF605464S, RSF605497S, RSF605782S, RSF606140S, RSF606552S, RSF606559S, RSF606560S, RSF606604S, RSF606828S, RSF606862S, RSF607506S, RSF607538S, RSF607572S, RSF607598S, RSF607613S, RSF607768S, RSF607769S, RSF607795S, RSF607796S, RSF607797S, RSF602070S, RSF603198S, RSF603914S, RSF604503S, RSF604915S, RSF605284S, RSF605363S, RSF605467S, RSF605791S, RSF606563S, RSF606565S, RSF606831S, RSF606832S, RSF607767S, RSF605133S, RSF605179S, RSF605328S, RSF605329S, RSF605573S, RSF605765S, RSF605772S, RSF605774S, RSF605781S, RSF605821S, RSF605891S, RSF606110S, RSF606132S, RSF606133S, RSF606292S, RSF606297S, RSF606302S, RSF606320S, RSF606323S, RSF606328S, RSF606356S, RSF606362S, RSF606368S, RSF606373S, RSF606379S, RSF606406S, RSF606475S, RSF606536S, RSF606577S, RSF606580S, RSF606583S, RSF606661S, RSF606773S, RSF606776S, RSF606829S, RSF606869S, RSF606870S, RSF606871S, RSF606905S, RSF606908S, RSF606912S, RSF606915S, RSF606931S, RSF606932S, RSF606933S, RSF606939S, RSF607003S, RSF607006S, RSF607007S, RSF607039S, RSF607192S, RSF607211S, RSF607226S, RSF607276S, RSF607280S, RSF607424S, RSF607427S, RSF607429S, RSF607437S, RSF607440S, RSF607441S, RSF607448S, RSF607464S, RSF607483S, RSF607488S, RSF607489S, RSF607507S, RSF607539S, RSF607554S, RSF607562S, RSF607566S, RSF607567S, RSF607568S, RSF607583S, RSF607584S, RSF607594S, RSF607601S, RSF607610S, RSF607615S, RSF607661S, RSF607676S, RSF607701S, RSF607715S, RSF607744S, RSF607761S, RSF607764S, RSF607786S, RSF607787S, RSF607793S, RSF607794S, RSF607802S, RSF607814S, RSF607817S, RSF607830S, RSF607837S, RSF607852S, RSF607854S, RSF608030S, RSF608103S, RSF608148S, RSF608157S, RSF608162S, RSF608163S, RSF608172S, RSF608210S, RSF608218S, RSF608220S, RSF608246S, RSF608249S, RSF608255S, RSF608258S, RSF608277S, RSF608278S, RSF608280S, RSF608281S, RSF608284S, RSF608286S, RSF608292S, RSF608293S, RSF608294S, RSF608299S, RSF608302S, RSF608318S, RSF608320S, RSF608321S, RSF608326S, RSF608327S, RSF608328S, RSF608329S, RSF608331S, RSF608333S, RSF608337S, RSF608338S, RSF608341S, RSF608344S, RSF608348S, RSF608349S, RSF608350S, RSF608351S, RSF608353S, RSF608354S, RSF608357S, RSF608364S, RSF608365S, RSF608366S, RSF608367S, RSF608368S, RSF608372S, RSF608475S, RSF608499S, RSF608500S, RSF608501S, RSF608502S, RSF608503S, RSF608504S, RSF608507S, RSF608508S, RSF608510S, RSF608511S, RSF608512S, RSF608513S, RSF608524S, RSF608525S, RSF608559S, RSF608573S, RSF608574S, RSF608576S, RSF608579S, RSF608580S, RSF608581S, RSF608582S, RSF608583S, RSF608584S, RSF608586S, RSF608588S, RSF608589S, RSF608593S, RSF608594S, RSF608604S, RSF608605S, RSF608606S, RSF608607S, RSF608608S, RSF608613S, RSF608614S, RSF608618S, RSF608619S, RSF608621S, RSF608625S, RSF608631S, RSF608632S, RSF608634S, RSF608637S, RSF608639S, RSF608640S, RSF608641S, RSF608644S, RSF608645S, RSF608643S, RSF608397S, RSF608398S, RSF608399S, RSF608400S, RSF608401S, RSF608403S, RSF608404S, RSF608435S, RSF608436S, RSF608437S, RSF608447S, RSF608448S, RSF608449S, RSF608450S, RSF608451S, RSF608452S, RSF608453S, RSF608454S, RSF608455S, RSF608456S, RSF608457S, RSF608458S, RSF608459S, RSF608460S, RSF608461S, RSF608462S, RSF608463S, RSF608464S, RSF608465S, RSF608466S, RSF608467S, RSF608476S, RSF608477S, RSF608478S, RSF608479S, RSF608480S, RSF608481S, RSF608482S, RSF608483S, RSF608484S, RSF608485S, RSF608486S, RSF608489S, RSF608492S, RSF608493S, RSF608494S, RSF608495S, RSF608496S, RSF608497S, RSF608498S, RSF608526S, RSF608527S, RSF608528S, RSF608529S, RSF608530S, RSF608531S, RSF608532S, RSF608533S, RSF608534S, RSF608535S, RSF608536S, RSF608537S, RSF608538S, RSF608539S, RSF608540S, RSF608541S, RSF608542S, RSF608543S, RSF608544S, RSF608546S, RSF608547S, RSF608548S, RSF608549S, RSF608550S, RSF608551S, RSF608552S, RSF608553S, RSF608554S, RSF608555S, RSF608556S, RSF608558S, RSF608560S, RSF608561S, RSF608562S, RSF608563S, RSF608565S, RSF608567S, RSF608569S, RSF608570S, RSF608649S, RSF604554S, RSF604575S, RSF604870S, RSF604891S, RSF605083S, RSF606361S, RSF606408S, RSF606510S, RSF606830S, RSF607420S, RSF607523S, RSF607609S, RSF608297S, RSF608305S, RSF608314S, RSF608340S, RSF608587S, RSF608620S
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce7048 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = LWS
PMAs with Product Code = LWS
-
-