• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall QUADROX iR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall QUADROX iRsee related information
Date Initiated by FirmMay 18, 2023
Date PostedMay 30, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1662-2023
Recall Event ID 92006
510(K)NumberK102726 
Product Classification Oxygenator, cardiopulmonary bypass - Product Code DTZ
ProductQUADROX-iR, used during cardiopulmonary bypass Model Item No. Description BEQ-HMO 51100-USA 701067936 Small Adult with filter, with BIOLINE coating BEQ-HMO 71100-USA 701067880 Adult with filter, with BIOLINE Coating HMO 50100-USA 701067934 Small Adult without filter, with SOFTLINE Coating HMO 51100-USA 701067938 Small Adult with filter, with SOFTLINE Coating HMO 70100-USA 701067874 Adult without filter, with SOFTLINE Coating HMO 71100-USA 701067886 Adult with filter, with SOFTLINE Coating
Code Information Model Item No. UDI-DI BEQ-HMO 51100-USA 701067936 N/A BEQ-HMO 71100-USA 701067880 4058863164052 HMO 50100-USA 701067934 N/A HMO 51100-USA 701067938 N/A HMO 70100-USA 701067874 N/A HMO 71100-USA 701067886 N/A All lots
FEI Number 3008355164
Recalling Firm/
Manufacturer
Maquet Medical Systems USA
45 Barbour Pond Dr
Wayne NJ 07470-2094
For Additional Information ContactMaryanna Krivak
973-709-7483
Manufacturer Reason
for Recall
Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.
FDA Determined
Cause 2
Under Investigation by firm
ActionOn May 18, 2023, the firm notified customers of the recall via an Urgent Medical Device Removal letter. Actions to be taken by the customer: If an affected device is already in use, please use according to normal practices. Monitor the patient for signs and symptoms of inflammation, infection, sepsis, and ischemia. If detected, treat according to clinical protocols. Please remove any affected product in inventory from areas of use. Unopened and unexpired affected product is eligible for credit upon return. Product Return: Please contact Getinge Customer Service at 1-888-943-8872 option 2 between the hours of 6:00 AM and 5:00 PM Pacific Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. Distributors who have shipped affected products to customers should forward the notification to customers' attention for appropriate action. Maquet Cardiopulmonary GmbH is working with all possible urgency on the redesign and revalidation of a packaging solution, which is anticipated to be available by Q4 2023.
Quantity in Commerce0 (US)
DistributionNationwide domestic distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = DTZ
-
-