• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall 2008 Series Hemodialysis Machines

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall 2008 Series Hemodialysis Machinessee related information
Date Initiated by FirmSeptember 06, 2023
Date PostedOctober 06, 2023
Recall Status1 Open3, Classified
Recall NumberZ-0028-2024
Recall Event ID 92788
510(K)NumberK150708 K173972 K222952 
Product Classification Dialyzer, high permeability with or without sealed dialysate system - Product Code KDI
Product2008T Hemodialysis SYS, with CDX
Code Information Model No: 190713; UDI-DI: 00840861100897; Serial No. 1T0S261032, 1T0S261004, 1T0S261029, 1T0S260995, 1T0S260937, 1T0S260739, 1T0S260687, 1T0S260573, 2T0S263945, 2T0S263947, 2T0S263959, 2T0S263924, 2T0S263962, 2T0S263956, 2T0S263951, 2T0S263950, 2T0S263707, 2T0S263692, 2T0S263690, 2T0S263682, 2T0S263700, 2T0S264084, 2T0S263944, 2T0S263954, 2T0S263957, 2T0S264087, 2T0S264144, 2T0S264132, 2T0S264073, 2T0S264139, 2T0S264061, 2T0S263979, 2T0S263983, 2T0S263973, 2T0S263977, 2T0S263989, 2T0S263938, 2T0S263960, 2T0S263943, 2T0S263968, 2T0S263941, 2T0S263940, 2T0S263939, 2T0S263934, 2T0S263933, 2T0S263932, 2T0S263931, 2T0S263930, 2T0S263927, 2T0S263926, 2T0S263925, 2T0S263923, 2T0S263922, 2T0S263921, 2T0S263920, 2T0S263919, 2T0S263918, 2T0S263916, 2T0S263912, 2T0S263911, 2T0S263910, 2T0S263792, 2T0S263791, 2T0S263784, 2T0S263780, 2T0S263778, 2T0S263776, 2T0S263775, 2T0S263771, 2T0S263769, 2T0S263767, 2T0S263754, 2T0S263752, 2T0S263721, 2T0S263715, 2T0S263711, 2T0S263703, 2T0S263948, 2T0S263718, 2T0S263935, 2T0S263946, 2T0S263953, 2T0S263785, 2T0S263710, 2T0S263615, 2T0S264240, 2T0S264210, 2T0S264246, 2T0S264220, 2T0S264243, 2T0S264250, 2T0S264211, 2T0S264221, 2T0S264247, 2T0S264255, 2T0S264245, 2T0S264248, 2T0S264241, 2T0S264249, 2T0S264264, 2T0S264258, 2T0S264232, 2T0S264234, 2T0S264235, 2T0S264259, 2T0S264265, 2T0S264263, 2T0S264260, 2T0S264227, 2T0S264226, 2T0S264229, 2T0S264224, 2T0S264230, 2T0S264231, 2T0S264236, 2T0S264256, 2T0S264239, 1T0S260442, 1T0S260555, 1T0S260595, 1T0S260603, 1T0S260606, 1T0S260607, 1T0S260609, 1T0S260610, 1T0S260611, 1T0S260612, 1T0S260613, 1T0S260614, 1T0S260615, 1T0S260619, 1T0S260622, 1T0S260627, 1T0S260635, 1T0S260639, 1T0S260641, 1T0S260650, 1T0S260661, 1T0S260678, 1T0S260724, 1T0S260760, 1T0S260768, 1T0S260771, 1T0S260774, 1T0S260783, 2T0S262650, 2T0S262689, 2T0S262694, 2T0S262698, 2T0S262712, 2T0S263634, 2T0S263821, 2T0S263836, 2T0S266610, 2T0S266633, 2T0S264058, 2T0S264081, 8T0S226502, 8T0S229547, 8T0S229559, 8T0S229563, 8T0S229570, 8T0S229574, 8T0S229575, 8T0S229582, 8T0S229583, 8T0S229586, 8T0S229590, 8T0S229601, 8T0S233656, 8T0S233657, 2T0S264214, 0T0S251261, 0T0S251286, 0T0S251319, 0T0S251792, 0T0S251811, 0T0S251809, 0T0S251810, 9T0S236793, 0T0S251309, 2T0S263496, 2T0S263531, 0T0S248136, 0T0S248150, 0T0S248148, 0T0S248156, 1T0S260397, 1T0S260419, 1T0S260349, 1T0S260427, 1T0S260346, 1T0S260433, 1T0S260352, 1T0S260358, 1T0S260370, 1T0S260429, 1T0S260374, 1T0S260367, 1T0S260423, 1T0S260616, 1T0S260637, 1T0S260707, 1T0S260823, 1T0S260405, 1T0S260424, 1T0S260311.
FEI Number 3001451489
Recalling Firm/
Manufacturer
Fresenius Medical Care Holdings, Inc.
920 Winter St Bld 920
Waltham MA 02451-1521
For Additional Information ContactFresenius Medical Care Technical Service
800-227-2572
Manufacturer Reason
for Recall
Potential PCBA leaching from tubing of hemodialysis machines.
FDA Determined
Cause 2
Device Design
ActionConsignees were sent an URGENT MEDICAL DEVICE CORRECTION NOTICE via USPS, dated 09/06/2023. The notice informed consignees that there is a potential for NDL-PCBAs could leach from silicone tubing manufactured with a peroxide-based catalyst. The firm also states that it has been demonstrated that dialysate run through 2008 Series machines used during typical maintenance HD treatments for at least 36 days, cumulatively, showed levels of NDL-PCBAs that were below toxicologically relevant thresholds. Consignees are to contact FMCRTG Technical Service at 1-800-227-2572 to schedule an update to devices to receive the correct tubing or if they have any further questions. Consignees with clinical questions are to contact Global Medical Information and Education Office at 1-855-2309. Consignees are to return the provided response form via fax or mail.
Quantity in Commerce703 units
DistributionDomestic: Nationwide Distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = KDI
-
-