| | Class 2 Device Recall GE Healthcare Centricty |  |
| Date Initiated by Firm | September 14, 2023 |
| Date Posted | October 20, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0139-2024 |
| Recall Event ID |
93099 |
| 510(K)Number | K123174 |
| Product Classification |
System, image processing, radiological - Product Code LLZ
|
| Product | Centricity PACS-IW with Universal Viewer, versions 5.0 SPx with PACS-IW foundation; Radiological Image Processing System |
| Code Information |
UDI/DI none, Installed Product ID #: 100707-1-Centricity Universal Viewer-00558649, 1045-1-Centricity Universal Viewer-00558651, 1045-1-Centricity Universal Viewer-19607785, 100759-1-Centricity Universal Viewer-02360986, 9842-4-Centricity Universal Viewer-19386293, 4723-1-Centricity Universal Viewer-02303576, 104648-2-Centricity Universal Viewer-02354806, 101272-1-Centricity Universal Viewer-00558671, 100713-1-Centricity Universal Viewer-00558259, 100713-1-Centricity Universal Viewer-01982629, 4705-1-Centricity Universal Viewer-02398923, 100512-1-Centricity Universal Viewer-00558264, 3727-1-Centricity Universal Viewer-00558421, HC3062-Centricity Universal Viewer-01462397, RU9761-UW PACS-IW UP TO 10K-02336112, ZA2533-Centricity Universal Viewer-00116577, YP3150-Centricity Universal Viewer-01639018
|
| FEI Number |
3004526608
|
Recalling Firm/ Manufacturer |
GE Healthcare
500 W Monroe St
Chicago
IL
60661-3671
|
| For Additional Information Contact | GE HealthCare Service 800-437-1171 |
Manufacturer Reason for Recall | GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and Centricity Universal Viewer V6.0 with PACS-IW foundation. |
FDA Determined Cause 2 | Software design |
| Action | GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/14/2023 by letter. The notice explained the problem, risk to the patient, and provided two workaround options for the continued use of the instruments until the software update is ready for installation.
Questions: please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative |
| Quantity in Commerce | 17 units |
| Distribution | US and Austria, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, Chile, Colombia, France, Germany, India, Ireland, Italy, Kuwait, Malaysia, Netherlands, Netherlands Antilles, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Kingdom, Uruguay, Venezuela, Vietnam |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LLZ
|
|
|
|