| | Class 2 Device Recall XScribe |  |
| Date Initiated by Firm | December 08, 2023 |
| Date Posted | January 10, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0707-2024 |
| Recall Event ID |
93209 |
| Product Classification |
Electrocardiograph - Product Code DPS
|
| Product | XScribe CP Cardiac Stress Testing System, version 6, with below product names and codes:
a) X-SCRIBE CPI KIT WIRELESS USB 608 MHz, 41000-030-50;
b) X-SCRIBE CPI KIT WIRELESS USB 2500 MHz, 41000-030-51;
c) X-SCRIBE CPI KIT DIRECT CONNECT USB, 41000-030-52;
d) X-SCRIBE CPI KIT QUICK CONNECT USB, 41000-030-53;
e) XSCRIBE CP KIT WIRELESS, 41000-030-60;
f) XSCRIBE CP KIT WIRED, 41000-030-62;
g) XSCRIBE CP SW KIT, 41000-030-64 |
| Code Information |
Product Code, UDI/DI Number, Serial Numbers
a) 41000-030-50, UDI/DI None, Serial Numbers: 110120039437, 110120039438, 110120039439, 110190044556, 110190044557, 110190044558, 110190044559, 110190044560, 110190044561, 110190044562, 110190044563, 110190044564, 110190044565, 110190044566, 110190044567, 110200045831, 110200045832, 110200045833, 110200045834, 110200045835, 110260052818, 110260052819, 110260052820, 110260052821, 110260052822, 110260052823, 110420061166, 110420061167, 110420061168, 110420061169, 110460064753, 110460064754, 110460064755, 110460064697, 110460064698, 110460064699, 110460064700, 110460064702, 110460064703, 110460064704, 110460064705, 110460064706, 111150077803, 111150077804, 111150077805, 111150077806, 111150077807, 111150077808, 111150077809, 111200081334, 111200081335, 111200081336, 111200081337, 111200081338, 111200081339, 111200081340, 111200081341, 111300088621, 111300088622, 111300088623, 111300088624, 111300088625, 111300088626, 111300088627, 111340091242, 111340091243, 111360092140, 111360092141, 111360092142, 111360092143, 111360092144, 111340091246, 111340091247, 111340091248, 111340091249, 111340091250, 111420097839, 111420097840, 111420097841, 111420097842, 111420097843, 111500103083, 111500103084, 111500103085, 111500103086, 111530104948, 111530104949, 111530104950, 111530104951, 112020106363, 112020106364, 112020106365, 112020106366, 112020106367, 112020106368, 112020106369, 112020106370, 112180115629, 112180115630, 112180115631, 112180115632, 112180115633, 112180115634, 112180115635, 112200117094, 112200117095, 112200117096, 112200117097, 112200117098, 112200117099, 112200117100, 112200117101, 112200117102, 112200117103, 112320128027, 112320128028, 112320128029, 112320128030, 112320128031, 112350130930, 112350130931, 112350130932, 112350130933, 112350130934, 112350130935, 112350130936, 112350130937, 112440138921, 112440138922, 112440138923, 112440138924, 112440138925, 112440138926, 112440138927, 112440138928, 112440138929, 112440138930, 112440138931, 112440138932, 113010147090, 113010147091, 113010147092, 113010147093, 113070153116, 113070153117, 113070153118, 113070153119, 113070153120, 113070153121, 113070153122, 113070153123, 113070153124, 113070153125, 113070153126, 113070153127, 113070153128, 113070153129, 113270168996, 113270168997, 113270168998, 113270168999, 113270169000, 113310172113, 113310172114, 113310172115, 113310172116, 113350174038, 113350174039, 113350174040, 113350174041, 113350174046, 113350174047, 113350174042, 113350174043, 113350174044, 113350174045, 113410177775, 113410177776, 113410177777, 113410177778, 113410177779, 113410177780, 113410177781, 113440180951, 113440180952, 113440180953, 113470184299, 113470184300, 113470184301, 113530189429, 113530189430, 113530189431, 113530189432, 113530189433, 113530189434, 114050192901, 114050192902, 114050192903, 114090196065, 114090196066, 114090196067, 113460182788, 113460182789, 113460182790, 113460182791, 113460182792, 113460182793, 114420232656, 114420232657, 114420232658, 114420232659, 114420232660, 114420232661, 114480237409, 114480237410, 114420232662, 114420232663, 114420232664, 114420232665, 114520240984, 114520240985, 114520240986, 114520240987, 114520240988, 114520240989, 115050243901, 115050243902, 115050243903, 114470236646, 114470236647, 114470236648, 115230258572, 115230258573, 115230258574, 115230258575, 115230258576, 115230258424, 115230258425, 115230258426, 115280263454, 115280263455, 115280263456, 115280263457, 115280263458, 115280263459, 115280264206, 115280264207, 115280264208, 115280264209, 115280264210, 115280264211, 115330268899, 115330268900, 115330268901, 115330268902, 115330268903, 115330268904, 115400275479, 115400275480, 115400275481, 115400275482, 115400275483, 115400275484, 115400275485, 115400275486, 115400275487, 115400275488, 116020287198, 116020287199, 116020287200, 116100335156, 116100335157, 116100335158, 116130382423, 116130382424, 116130382425, 116130382426, 116130382427, 116180422043, 116180422044, 116180422045, 116180422046, 116180422047, 116240477718, 116240477719, 116240477720, 116240477721, 116270497993, 116270497994, 116270497995, 116270497996, 116380527123, 116380527124, 116380527125, 116380527126, 116380527127, 116380527128, 116380527129, 116380527130;
b) 41000-030-51, UDI/DI None, Serial Numbers: 110120039425, 110120039426, 110120039427, 110190044568, 110190044569, 110190044570, 110200045836, 110200045837, 110200045838, 110330056675, 110330056676, 110370059009, 110370059010, 110370059011, 110390059824, 110420060941, 110420060942, 110500066898, 110500066899, 110500066900, 111060071714, 111060071715, 111060071716, 111150077462, 111150077463, 111200081342, 111200081343, 111270086404, 111270086405, 111270086406, 111270086407, 111380093553, 111380093554, 111380093555, 111380093556, 111380093557, 111470101214, 111470101215, 111500103087, 111500103088, 111500103089, 112030106627, 112030106628, 112050108028, 112050108029, 112100110764, 112100110765, 112050108026, 112050108027, 112220118711, 112220118712, 112220118713, 112270124286, 112270124287, 112270124288, 112320128032, 112320128033, 112320128034, 112410137135, 112410137136, 112410137137, 112440139667, 112440139668, 112440138933, 112440138934, 112440138935, 112440138936, 112440138937, 112440138938, 113020147730, 113020147731, 113020147732, 113020147733, 113020147734, 113020147735, 113020147736, 113020147737, 113130159216, 113130159217, 113130159218, 113130159219, 113130159220, 113130159221, 113130159222, 113130159223, 113130159224, 113130159225, 113130159226, 113130159227, 113130159228, 113130159229, 113130159230, 113450181907, 113450181908, 113480185307, 113480185308, 113450181644, 113450181645, 113450181646, 113450181647, 113450181648, 113450181649, 113450181650, 113450181651, 113450181652, 113450181653, 114380229638, 114380229639, 114440234317, 114440234325, 114470236649, 114470236650, 114480237351, 114480237352, 114480237353, 114480237354, 114480237349, 114480237350, 115060244042, 115060244043, 115100247551, 115100247552, 115100247553, 115140251901, 115140251902, 115140251903, 115140251904, 115140251905, 115140251906, 115140251907, 115140251908, 115200256104, 115200256105, 115230258565, 115230258566, 115230258567, 115230258568, 115240258971, 115240258972, 115240258973, 115240258974, 115200256100, 115200256101, 115200256102, 115200256103, 116020287201, 116020287202, 116050291262, 116050291263, 116100335159, 116100335160, 116140387942, 116140387943, 116140387944, 116180423551, 116180423552;
c) 41000-030-52, UDI/DI None, Serial Numbers: 110120039363, 110120039364, 110230049218, 110230049219, 110230049220, 110230049221, 110280053721, 110280053722, 110280053723, 110280053724, 110280053725, 110330056793, 110330056794, 110370059012, 110370059013, 110370059014, 110410060508, 110410060509, 110410060510, 110410060511, 110440063042, 110440063043, 110440063044, 110440063045, 110460064750, 110460064751, 110460064752, 110460064707, 110460064708, 110460064709, 110460064710, 110460064711, 110460064712, 110460064713, 110460064714, 111060071710, 111060071711, 111060071712, 111060071713, 111200081535, 111200081536, 111200081537, 111230083963, 111230083964, 111230083965, 111330090559, 111330090560, 111330090561, 111330090562, 111330090563, 111330090564, 111330090565, 111450099468, 111450099469, 111450099470, 111450099471, 111450099472, 111450099473, 111450099474, 111450099475, 111460100779, 111460100780, 111460100781, 111490102399, 111490102400, 111490102401, 111460100426, 111460100427, 111460100428, 112020105816, 112020105817, 112020105818, 112020105819, 112020105820, 112020105821, 112130112152, 112130112153, 112130112154, 112130112155, 112130112156, 112130112157, 112280125032, 112280125033, 112280125034, 112280125035, 112280125036, 112280125037, 112250123314, 112250123315, 112250123316, 112250123317, 113100155571, 113100155572, 113130159201, 113130159202, 113130159203, 113130159204, 113130159205, 113130159206, 113130159207, 113130159208, 113130159209, 113130159210.113130159211, 113130159212, 113130159213, 113130159214, 113130159215, 114020190211, 114020190212, 114020190213, 114020190214, 114020190215, 114020190216, 114020190217, 114180209245, 114180209246, 114020190218, 114020190219, 114180209247, 114180209248, 114180209249, 114180209250, 114180209251, 114180209252, 114250218419, 114460236532, 114460236533, 114460236534, 114460236529, 114460236530, 114460236531, 114460236526, 114460236527, 114460236528, 115260260562, 115260260563, 115260260564, 115360271118, 115360271119, 115360271120, 115440278666, 115440278667, 115440278668, 115500283264, 115500283265, 115500283266, 116050290129, 116050290130, 116050290131, 116050290132, 116110349285, 116110349286, 116110349287, 116110349288, 116110349289, 116200433074, 116200433075, 116230452680, 116230452681, 116230452682, 116280504772, 116280504773, 116280504774;
d) 41000-030-53, UDI/DI None, Serial Numbers: 110120039428, 110120039429, 110120039430, 110120039431, 110120039432, 110120039433, 110120039434, 110120039435, 110120039436;
e) 41000-030-60, UDI/DI 00817655020051, Serial Numbers: 118060611436, 118060611438, 118060611439, 118080615013, 118080615022, 118180634934, 118080615011, 118180634932, 118180634933, 118180634936, 118060611447, 118080615017, 118080615018, 118080615021, 118080615019, 118080615020, 118180634938, 118180634939, 118180634940, 118160628568, 118160628569, 118160628570, 118180634931, 118180634935, 118160628577, 118160628578, 118160628579, 118160628580, 118160628581, 118160628583, 118160628564, 118160628565, 118160628566, 118160628567, 118160628571, 118160628572, 118160628573, 118160628582, 118160628574, 118160628575, 118160628576, 118340001391, 118340001392, 118340001393, 118340001394, 118340001395, 118340001396, 118340001397, 118340001388, 118340001389, 118340001390, 119110000129, 119110000130, 119110000132, 119110000133, 119110000134, 119110000135, 119110000136, 119110000137, 119130001327, 119130001328, 119130001331, 119110000128, 119110000131, 119130001329, 119130001330, 119170000064, 119170000066, 119170000067, 119170000068, 119190000633, 119190000634, 119190000635, 119190000629, 119190000630, 119190000636, 119190000637, 119170000054, 119170000061, 119190000628, 119190000631, 119190000632, 119170000055, 119170000056, 119170000062, 119170000063, 119170000065, 119170000057, 119170000058, 119170000059, 119170000060, 119240001197, 119240001198, 119340000071, 119340000072, 119340000073, 119340000074, 119340000075, 119340000076, 119340000077, 119340000078, 119340000079, 119340000080, 119350002675, 119350002676, 119350002677, 119350002678, 119350002679, 119350002680, 119350002681, 119350002682, 119350002683, 119350002684, 119380001144, 119380001145, 119380001146, 119380001147, 119380001148, 119380001149, 119430001324, 119430001325, 119430001326, 119430001327, 119430001328, 119430001329, 119430001330, 119430001331, 119430001332, 119430001333, 119470001107, 119470001108, 119470001109, 119470001110, 119470001111, 119470001112, 119470001113, 119470001114, 119470001115, 119470001116, 119500000427, 119500000428, 119500000429, 119500000430, 119500000431, 119500000432, 119500000433, 119500000434, 120080000436, 120080000437, 120080000438, 120080000439, 120080000440, 120080000441, 120080000442, 120080000443, 120080000444, 120080000445, 120080000446, 120080000447, 120080000448, 120080000449, 120080000450, 120080000451, 120080000452, 120080000453, 120080000454, 120080000455, 120260000436, 120260000437, 120260000438, 120370001869, 120370001870, 120370001872, 120260000439, 120260000440, 120370001871, 120370001873, 120370001874, 120370001875, 120370001876, 120370001877, 120370001878, 120410001420, 120410001421, 120410001422, 120410001423, 120410001424, 120410001425, 120410001426, 120410001427, 120480000382, 120480000383, 120480000384, 120480000385, 120480000386, 120480000387, 120480000388, 120480000389, 120480000390, 120480000391, 120480000392, 120480000393, 120480000394, 120480000395, 120520000178, 120520000179, 120520000180, 120520000181, 120520000182, 120520000183, 120520000184, 120520000185, 120520000186, 120520000187, 120520000188, 120520000189, 121060001022, 121060001023, 121060001024, 121060001025, 121060001026, 121060001027, 121060001028, 121060001029, 121060001030, 121060001031, 121080000918, 121080000919, 121080000920, 121080000921, 121080000922, 121080000923, 121080000924, 121130001105, 121130001106, 121130001107, 121130001108, 121130001109, 121130001110, 121130001111, 121130001112, 121130001113, 121130001114, 121130001115, 121130001116, 121130001117, 121130001118, 121130001119, 121140001520, 121140001521, 121140001522, 121140001523, 121140001524, 121170000316, 121170000317, 121170000318, 121170000319, 121170000320, 121170000321, 121170000322, 121170000323, 121170000324, 121170000325, 118050609197, 121210002553, 121210002554, 121210002555, 121210002556, 121210002557, 121210002558, 121210002559, 121210002560, 121210002561, 121210002562, 121250000651, 121250000652, 121250000653, 121250000654, 121250000655, 121250000656, 121250000657, 121250000658, 121250000659, 121250000660, 121250000661, 121250000662, 121300001879, 121300001880, 121300001881, 121300001882, 121300001883, 121300001884, 121300001885, 121300001886, 121300001887, 121300001888, 121320000722, 121320000723, 121320000724, 121320000725, 121320000726, 121320000727, 121320000728, 121320000729, 121390000580, 121390000581, 121390000582, 121390000583, 121390000584, 121390000585, 121390000586, 121390000587, 121390000588, 121390000589, 121390000590, 121390000591, 121430000457, 121430000458, 121430000459, 121430000460, 121430000461, 121430000462, 121430000463, 121430000464, 121430000465, 121430000466, 121430000467, 121430000468, 121450000127, 121450000128, 121460000618, 121490000525, 121490000526, 121490000527, 121490000528, 121490000529, 121490000530, 121490000531, 121490000532, 121490000533, 121490000534, 121490000535, 121490000536, 122130000539, 122130000540, 122130000541, 122130000542, 122130000543, 122130000544, 122130000545, 122130000546, 122190000029, 122190000030, 122190000031, 122190000032, 122190000033, 122190000034, 122200001921, 122200001922, 122200001923, 122200001924, 122200001925, 122200001926, 122200001927, 122200001928, 122200001929, 122200001930, 122200001931, 12200001930, 122240001466, 122240001467, 122240001468, 122240001469, 122240001470, 122240001471, 122230000033, 122230000036, 122230000039, 122230000040, 122230000042, 122230000044, 122230000037, 122230000038, 122230000041, 122230000043, 122230000034, 122260001174, 122260001175, 122260001176, 122260001177, 122260001178, 122260001179, 122280000876, 122280000877, 122280000878, 122280000879, 122280000880, 122280000881, 122340001204, 122340001205, 122340001206, 122340001207, 122390000370, 122390000371, 122390000372, 122390000373, 122390000374, 122390000375, 122420000196, 122420000197, 122420000198, 122420000199, 122420000200, 122420000203, 122420000192, 122420000193, 122420000194, 122420000195, 122420000201, 122420000202, 122450001786, 122450001787, 122450001788, 122450001789, 122450001790, 122450001791, 123100000063, 123100000069, 123100000070, 123100000071, 123100000064, 123100000065, 123100000066, 123100000067, 123100000068, 123130001066, 123130001067, 123210000710, 123210000711, 123340001553, 123340001554, 123320000551, 123320000552, 123370000077, 123370000078, 123370000079, 123370000080, 118050610039, 118180634937, 118050609198, 118180634937, 121450000127, 121450000128, 121460000618, 122230000033, 122230000034, 122230000035, 122230000036, 122230000037, 122230000038, 122230000039, 122230000040, 122230000041, 122230000042, 122230000043, 122230000044, 122500000829, 122500000830, 122500000831, 116480540886, 116480540887, 116480540885, 116480540888, 116480540889, 116480540890, 116480540891, 116500543972, 116530547195, 116530547196, 116530547197, 116530547198, 116530547199, 116530547200, 116530547201, 116530547207, 116530547208, 116530547209, 116530547202, 116530547203, 116530547204, 116530547205, 116530547206, 117050552991, 117050552992, 117140561292, 117140561293, 117140561294, 117140561295, 117140561296, 117140561297, 117140561299, 117140561300, 117140561301, 117140561302, 117190567099, 117190567100, 117140561298, 117190567101, 117190567102, 117190567103, 117190567104, 117190568321, 117190568322, 117190568323, 117210570066, 117250575592, 117250575593, 117250575594, 117250575595, 117250575596, 117250575597, 117260576261, 117260576262, 117260576263, 117250575598, 117250575599, 117250575600, 117250575601, 117310581178, 117310581179, 117350584646, 117350584647, 117350584648, 117350584649, 117350584650, 117350584642, 117350584643, 117400589264, 117400589265, 117400589266, 117400589267, 117400589268, 117400589293, 117400589294, 117400589295, 117400589296, 117400589297, 117400589298, 117400589299, 117400589269, 117400589270, 117400589271, 117400589272, 117400589273, 117400589274, 117430591637, 117430591638, 117430591639, 117430591640, 117430591980, 117430591981, 117490599622, 117490599623, 117490599624, 117490599625, 117490599627, 117490599628, 117490599630, 117490599619, 117490599620, 117490599621, 117490599626, 117490599629, 117490599631, 117490599632, 117490599633, 117510602851, 117510602852, 117510602853, 117510602854, 117520603151, 117520603152, 117520603153, 117520603154, 118030606983, 118030606984, 118030606985, 118030606986, 118030606987, 118030606988, 118030606989, 118030606990, 118030606991, 118030606992, 118050609193, 118050609194, 118050609195, 118050609196, 118050609199, 118050609200, 118050609201, 118050609202, 118060611433, 118060611429, 118060611430, 118060611431, 118060611432, 118060611434, 118060611435, 118060611437, 118060611440, 118080615016, 118080615015, 118080615014, 118080615012, 118080615010, 118080615009, 118080615008, 118080615007, 118080615006, 118080615005, 118080615004, 118080615003, 118060611448, 118060611446, 118060611445, 118060611444, 118060611443, 118060611442, 118060611441, 41000-030-60, 118180634941, 118180634942, 118340001398, 118340001399, 118340001400, 118340001401, 118340001402, 118340001403, 118340001404, 118340001405, 118340001406, 118340001407, 118340001408, 118340001409, 118340001410, 118340001411, 118340001412, 120180001027, 120180001028, 120180001029, 120180001030, 120180001031, 10081765502005100, 121040002235, 121040002236, 121040002237, 121040002238, 121040002239, 121040002240, 121040002241, 121040002242, 121040002243, 121040002244, 121050001674, 121050001675, 121050001676, 121050001677, 121050001678, 121050001679, 121050001680, 121050001681, 121050001682, 121050001683, 121080000925, 121080000926, 121080000927, 122200001932, 10081765502005100, 123390000815, 123390000816, 123390000817, 123390000818, 123400000181, 123400000182, 123400000183, 123400000184, 123420000969, 123420000970, 123420000971, 123420000972, 123430002087, 123440000214, 123440000215, 123440000216, 123440000217, 123450001456, 123450001457, 123450001458, 123450001459, 123480000269, 123480000270, 123480000271, 123480000272;
f) 41000-030-62, UDI/DI 00817655020068, Serial Numbers: 118160628584, 118160628585, 118320000033, 118330000323, 118330000325, 118320000032, 118330000324, 118330000326, 118330000320, 118330000321, 118330000322, 118450000081, 118450000082, 118330000319, 118470000102, 118470000406, 118470000407, 118470000408, 118480000446, 118480000447, 118480000448, 118480000449, 118500001826, 118500001827, 118500001834, 118500001835, 118500001828, 118500001830, 118500001829, 118500001831, 118500001832, 119190000641, 119190000642, 119190000638, 119190000639, 119190000640, 118500001825, 118500001833, 119250001407, 119320001005, 119320001006, 119320001007, 119320001002, 119320001003, 119320001004, 119350002522, 119350002523, 119350002524, 119350002525, 119430001320, 119430001321, 119430001322, 119430001323, 119430001318, 119430001319, 119470002142, 119470002144, 119470002146, 119470002143, 119470002145, 119470002147, 120040000343, 120040000344, 120050001743, 120050001744, 120050001745, 120080000456, 120080000457, 120080000458, 120080000459, 120080000460, 120080000461, 120080000462, 120080000463, 120080000464, 120080000465, 120080000466, 120080000467, 120080000468, 120080000469, 120080000470, 120080000471, 120080000472, 120080000473, 120080000474, 120330000674, 120330000676, 120050001744, 120050001745, 120330000671, 120330000672, 120330000673, 120330000675, 120330000677, 120330000678, 120330000679, 120330000680, 120050001743, 120480000397, 120480000396, 120520000680, 120520000681, 120520000682, 121060001019, 121060001020, 121060001021, 121080000928, 121080000929, 121130001120, 121130001121, 121130001122, 121170000326, 121170000327, 121170000328, 121210002563, 121210002564, 121210002565, 121250000663, 121250000664, 121250000665, 121300001889, 121300001890, 121300001891, 121340000858, 121340000859, 121340000860, 121420001162, 121420001163, 121420001164, 121420001165, 121470000236, 121470000237, 121470000238, 121510001415, 121510001416, 121510001417, 122050001914, 122050001915, 122050001916, 122080000224, 122080000225, 122080000226, 122130000547, 122130000548, 122130000549, 122240001057, 122240001061, 122240001062, 122240001058, 122240001059, 122240001060, 122240001063, 122310000395, 122310000396, 122310000397, 122320000145, 122320000146, 122320000147, 122320000881, 122320000882, 122320000883, 122380000624, 122380000625, 122380000626, 122450001792, 122450001793, 122450001794, 122480001906, 122480001909, 123020001482, 123020001483, 123070001510, 123070001513, 123070001514, 123070001515, 123100001935, 123100001936, 123070001509, 123070001511, 123070001512, 123190000747, 123190000749, 123150002638, 123190000748, 123200000231, 123150002639, 123220001435, 123220001436, 123300001463, 123300001464, 123300000098, 123300000099, 123370000075, 123370000076, 122480001907, 122480001908, 123320000988, 123300001462, 123320000989, 123320000986, 123320000987, 120050001743, 120050001744, 120050001745, 120330000674, 120330000676, 120480000396, 113400177568, 116480540898, 116530547192, 116530547193, 116530547194, 117010549589, 117010549590, 117110556514, 117110556515, 117110556516, 117130560456, 117130560457, 117130560458, 117150562750, 117150562748, 117150562749, 117180567076, 117180567077, 117190568320, 117210570063, 117210570064, 117210570065, 117260575996, 117260575997, 117260575998, 117310581180, 117310581181, 117350584629, 117350584630, 117350584631, 117350584632, 117350584633, 117400589290, 117400589291, 117400589292, 117430592542, 117430592543, 117430592544, 117490599636, 117490599637, 117490599639, 117490599634, 117490599635, 117490599638, 118070612685, 118070612687, 118070612686, 118080614439, 118080614440, 118090617067, 118090617068, 118090617071, 118090617070, 118090617069, 118160628588, 118160628587, 118160628586, 118160628589, 118160628590, 118320000034, 118320000035, 118330000327, 118330000328, 118330000329, 120040000345, 120040000346, 120040000347, 120480000398, 120480000399, 120480000400, 120480000401, 120480000402, 120480000403, 120480000404, 120480000405, 120480000406, 120480000407, 120480000408, 123400000068, 123400000069;
g) 41000-030-64, UDI/DI 00817655021423, Serial Numbers: 119490000440, 119500001335, 121350001043, 120070000730, 119490000439, 120070000732, 118380003022, 118380003023, 118470000103, 118470000104, 119500001336, 118380003022, 118380003023, 118470000103, 118470000104, 120070000731, 123150002068, 123150002069, 123150002070, 123160000556, 123150002055, 123200001669, 123150002057, 123150002058, 123200001670, 123200001671, 123150002066, 123150002067, 120070000733, 123150002056, 123150002061, 123150002062, 123150002063, 123150002064, 123150002065, 123150002059, 123150002060, 123160000552, 123160000553, 123160000554, 123160000555, 119500001335, 121350001043, 120070000730, 119490000439, 120070000732, 117320582006, 120070000734, 120070000735, 120070000736, 120070000737, 11847000104, 120070000738, 120070000739, 121350001044, 121350001045 |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
| For Additional Information Contact | Center for One Baxter 800-422-9837 |
Manufacturer Reason for Recall | Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled. |
FDA Determined Cause 2 | Software design |
| Action | Baxter issued an Urgent Medical Device Correction notice to its consignees on 12/08/2023 via USPS first class mail. The notice explained the problem, hazard involved, requested the following actions be taken until a software update becomes available:
- Clinicians should evaluate the potential impact of the SCF as described to determine if it should be enabled or disabled during stress testing and provided the method for disabling the setting.
- If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them, informing them of the requirement to evaluate the potential impact of the SCF during stress testing.
- Respond using the customer portal: https://BaxterFieldActionCustomerPortal.onprocess.com.
- If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal.
- If you purchased this product from a distributor, please note that the Baxter reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to your distributor/wholesaler according to their instructions.
For general questions regarding this communication, contact Baxter (legacy Welch Allyn and Hillrom) Customer Service at 800-535-6663, Option 1, or Option 2 for Technical Support, between the hours of 8:00 am and 8:00 pm Eastern Time, Monday through Friday. |
| Quantity in Commerce | 1675 units |
| Distribution | United States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia,
Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Morocco, Namibia, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tunisia, Turkey, United Arab Emirates, United Kingdom, Argentina, Bahamas, Bolivia, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, and Trinidad and Tobago. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|