 Class 2 Device Recall Strykersee related information
Date Initiated by FirmDecember 01, 2023
Date PostedJanuary 31, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0893-2024
Recall Event ID 93687
510(K)NumberK051948 K052917 K972967 
Product Classification Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
ProductSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3110
Code Information GTIN: 07613327033861 Lot Numbers: XEA7, 8P83, 690Y, MRM3, 67D4, 4902, YNP7, 3AYN, AYHM, W3MD, 76KN, K4P9, 792H, 43T8, MW40
FEI Number 3003070421
Recalling Firm/
Manufacturer
Howmedica Osteonics Corp.
325 Corporate Dr
Mahwah NJ 07430-2006
For Additional Information ContactLoriann Russo
201-831-5000
Manufacturer Reason
for Recall
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionStryker issued Urgent Medical Device Recall Letter (Product Field Action #: 3306872)through third party Sedgwick via UPS on 12/1/23. Letter states reason for recall, health risk and action to take: Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the affected product in Part/Lot Number Attachment: PFA 3306872 starting on page 3. 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and email to strykerortho4367@sedgwick.com 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return ALL affected product to: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 Ref. PFA 3306872; Please include Tracking information on the package. If you have any questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com
Quantity in Commerce107 units US; 176 units OUS
DistributionNationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.