 Class 1 Device Recall DatexOhmedasee related information
Date Initiated by FirmDecember 29, 2023
Date PostedJanuary 30, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0775-2024
Recall Event ID 93715
Product Classification Compressor, air, portable - Product Code BTI
ProductEVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002
Code Information UDI/DI ********, Serial Numbers: CBCS01269, CBCT02328, CBCT03050, CBCT02306, CBCT02301, CBCT02303, CBCT02302, CBCT02088, CBCT02309, CBCT02304, CBCU00717, CBCU00559
FEI Number 2112667
Recalling Firm/
Manufacturer
Datex-Ohmeda, Inc.
3030 Ohmeda Dr
Madison WI 53718-6704
For Additional Information ContactGE HealthCare Service
800-437-1171
Manufacturer Reason
for Recall
GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstrm Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.
FDA Determined
Cause 2
Component design/selection
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to consignees on 12/29/2023. The notice explained the safety issue with the device and requested the following actions be taken: GE HealthCare recommends that the EVair and EVair 03 (Jun-Air) compressors are not used to supply air to ventilators for neonatal and infant patients (0-2 years of age). Since the elevated levels of formaldehyde are observed when the compressors are used at higher room temperatures, GE HealthCare is also lowering the maximum room air temperature for operation of the compressors from 40oC (104oF) to 30oC (86oF). Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Retain this document and the Addendum to the user instructions for your records. Please keep a copy of the Addendum available near the areas where the compressors are used. GE sent an updated communication to consignees on 10/11/2024 stating the following: EVair Compressors: The recommended actions indicated in the previous letter for EVair Compressors no longer apply, as such: 1. The EVair Compressors can be used to supply air to CARESCAPE R860 ventilators for all patient populations, including neonatal and infant patients (0-2 years of age). 2. The EVair Compressor can be used in a room with maximum air temperature of 40oC (104F). EVair 03 (Jun-Air) Compressors: For the EVair 03 (Jun-Air) compressors (at End of Service Support): If you choose to continue to use these, the instructions in the 12/29/2023 letter continue to apply: 1. GE HealthCare recommends that the EVair 03 (Jun-Air) compressors are not used to supply air to ventilators for neonatal and infant patients (0-2 years of age). 2. GE HealthCare recommends that these compressors are used at a maximum room air temperature of 30oC (86oF). Contact for questions/concerns: GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce12 units
DistributionWorldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-