• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Angiodynamics DURAFLOW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Angiodynamics DURAFLOWsee related information
Date Initiated by FirmMarch 18, 2024
Date PostedMay 03, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1767-2024
Recall Event ID 94349
510(K)NumberK101843 
Product Classification Catheter, hemodialysis, implanted - Product Code MSD
ProductDURAMAX STACKED TIP 24CM STR. BASIC KIT, 5-UP PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103028035
Code Information UDI-DI: 25051684011836 Lot Numbers: 5807194, 5809616, 5809462
Recalling Firm/
Manufacturer
Angiodynamics, Inc.
603 Queensbury Ave
Queensbury NY 12804-7619
For Additional Information ContactSAME
518-798-1215
Manufacturer Reason
for Recall
Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
FDA Determined
Cause 2
Under Investigation by firm
ActionAngiodynamics issued Recall Notifications Packages via Federal Express to 1 Domestic Consignee on 2024-03-18 . Letter states reason for recall, health risk and action to take: IMMEDIATELY o Locate the affected product in your Inventory, cease any and all distribution segregate in a secure location pending return. o Forward a copy of this recall notification to any locations to which you have distributed affected product. 2. Complete and return the Reply Verification Tracking Form. " Promptly complete, sign, and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return), following the directions on this page and the Reply Verification Tracking Form. Email Reply Verification Tracking Form (preferred): recall@angiodynamics.com Fax Reply Verification Tracking Form: Attn: Hemodialysis Catheter Kits with Tri-Ball Tunneller Fax number 1-855-273-0519 " Call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. Package any product that is being returned in an appropriate shipping box. " Write the RMA number on the RMA label (provided on the Recall Verification Tracking Form) and affix the label to the outside of the shipping box. " Seal the box and return via your preferred carrier to: AngioDynamics, Inc. 24 Native Drive Queensbury, NY 12804 Attn: Hemodialysis Catheter Kits with Tri-Ball Tunneller
Quantity in Commerce555 units
DistributionVA
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = MSD
-
-