| | Class 2 Device Recall LOSPA Knee System |  |
| Date Initiated by Firm | March 29, 2024 |
| Date Posted | May 31, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1934-2024 |
| Recall Event ID |
94380 |
| 510(K)Number | K110404 K130673 K160157 |
| Product Classification |
Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
|
| Product | LOSPA Tibial Insert
Model/Catalog Number: See below
Software Version: N/A
Product Description: Tibial Insert |
| Code Information |
Lot Code: LOSPA Tibial Insert (CR)
Model No: 01.10.1XX
LOSPA Tibial Insert (PS)
Model No: 01.10.2XX
LOSPA Fixed Constrained Tibial Insert
Model No: K2.IBN.4061-XXX, K2.IBN.4264-XXX, K2.IBN.4467-XXX, K2.IBN.4672-XXX, K2.IBN.4874-XXX, K2.IBN.5076-XXX, K2.IBN.5281-XXX, K2.IBN.5684-XXX
UDI/Lot Numbers: See attached Lospa recall Attachment 1
Lospa Knee System
01.10.132
Lospa Tibial Insert (CR) - Size 3 / 10mm
0108806373820953172605281070AFDM33
70AFDM33
5/28/2026
Lospa Knee System
01.10.132
Lospa Tibial Insert (CR) - Size 3 / 10mm
0108806373820953172703281070AFL130
70AFL130
3/28/2027
Lospa Knee System
01.10.134
Lospa Tibial Insert (CR) - Size 3 / 12mm
0108806373820960172605281070AFDM34
70AFDM34
5/28/2026
Lospa Knee System
01.10.134
Lospa Tibial Insert (CR) - Size 3 / 12mm
0108806373820960172712281070AFL131
70AFL131
12/28/2027
Lospa Knee System
01.10.136
Lospa Tibial Insert (CR) - Size 3 / 14mm
0108806373820977172510281070AEK548
70AEK548
10/28/2025
Lospa Knee System
01.10.136
Lospa Tibial Insert (CR) - Size 3 / 14mm
0108806373820977172601281070AEK548
70AEK548
1/28/2026
Lospa Knee System
01.10.138
Lospa Tibial Insert (CR) - Size 3 / 16mm
0108806373820984172602281070AFAR33
70AFAR33
2/28/2026
Lospa Knee System
01.10.138
Lospa Tibial Insert (CR) - Size 3 / 16mm
0108806373820984172710281070AFL133
70AFL133
10/28/2027
Lospa Knee System
01.10.138
Lospa Tibial Insert (CR) - Size 3 / 16mm
0108806373820984172712281070AFL133
70AFL133
12/28/2027
Lospa Knee System
01.10.138
Lospa Tibial Insert (CR) - Size 3 / 16mm
0108806373820984172801281070AFL133
70AFL133
1/28/2028
Lospa Knee System
01.10.13A
Lospa Tibial Insert (CR) - Size 3 / 18mm
0108806373820991172601281070AEKE31
70AEKE31
1/28/2026
Lospa Knee System
01.10.142
Lospa Tibial Insert (CR) - Size 4 / 10mm
0108806373821158172811281070AHME32
70AHME32
11/28/2028
Lospa Knee System
01.10.142
Lospa Tibial Insert (CR) - Size 4 / 10mm
0108806373821158172812281070AJA734
70AJA734
12/28/2028
Lospa Knee System
01.10.146
Lospa Tibial Insert (CR) - Size 4 / 14mm
0108806373821172172509281070AEK632
70AEK632
9/28/2025
Lospa Knee System
01.10.146
Lospa Tibial Insert (CR) - Size 4 / 14mm
0108806373821172172602281070AFAR35
70AFAR35
2/28/2026
Lospa Knee System
01.10.146
Lospa Tibial Insert (CR) - Size 4 / 14mm
0108806373821172172803281070AHBR39
70AHBR39
3/28/2028
Lospa Knee System
01.10.148
Lospa Tibial Insert (CR) - Size 4 / 16mm
0108806373821189172512281070AELH38
70AELH38
12/28/2025
Lospa Knee System
01.10.14A
Lospa Tibial Insert (CR) - Size 4 / 18mm
0108806373821196172602281070AFAR36
70AFAR36
2/28/2026
Lospa Knee System
01.10.156
Lospa Tibial Insert (CR) - Size 5 / 14mm
0108806373821271172804281060AHAA36
60AHAA36
4/28/2028
Lospa Knee System
01.10.15A
Lospa Tibial Insert (CR) - Size 5 / 18mm
0108806373821295172612281060AFL143
60AFL143
12/28/2026
Lospa Knee System
01.10.178
Lospa Tibial Insert (CR) - Size 7 / 16mm
0108806373821486172812281060AHLF32
60AHLF32
12/28/2028
Lospa Knee System
01.10.17A
Lospa Tibial Insert (CR) - Size 7 / 18mm
0108806373821493172710281060AFL148
60AFL148
10/28/2027
Lospa Knee System
01.10.17A
Lospa Tibial Insert (CR) - Size 7 / 18mm
0108806373821493172902281060AHAJ33
60AHAJ33
2/28/2029
Lospa Knee System
01.10.17C
Lospa Tibial Insert (CR) - Size 7 / 20mm
0108806373821509172804281060AHDB36
60AHDB36
4/28/2028
Lospa Knee System
01.10.184
Lospa Tibial Insert (CR) - Size 8 / 12mm
0108806373821561172807281070AHFX35
70AHFX35
7/28/2028
Lospa Knee System
01.10.194
Lospa Tibial Insert (CR) - Size 9 / 12mm
0108806373821660172602281080AFAR40
80AFAR40
2/28/2026
Lospa Knee System
01.10.196
Lospa Tibial Insert (CR) - Size 9 / 14mm
0108806373821677172605281080AFCR45
80AFCR45
5/28/2026
Lospa Knee System
01.10.196
Lospa Tibial Insert (CR) - Size 9 / 14mm
0108806373821677172902281080AJBF34
80AJBF34
2/28/2029
Lospa Knee System
01.10.198
Lospa Tibial Insert (CR) - Size 9 / 16mm
0108806373821684172902281080AJC239
80AJC239
2/28/2029
Lospa Knee System
01.10.19A
Lospa Tibial Insert (CR) - Size 9 / 18mm
0108806373821691172902281080AJC330
80AJC330
2/28/2029
Lospa Knee System
01.10.1B2
Lospa Tibial Insert (CR) - Size 11 / 10mm
0108806373821851172902281030AJC331
30AJC331
2/28/2029
Lospa Knee System
01.10.1B4
Lospa Tibial Insert (CR) - Size 11 / 12mm
0108806373821868172902281030AJC332
30AJC332
2/28/2029
Lospa Knee System
01.10.1C6
Lospa Tibial Insert (CR) - Size 12 / 14mm
0108806373821974172303281060accs32
60accs32
3/28/2023
Lospa Knee System
01.10.1C6
Lospa Tibial Insert (CR) - Size 12 / 14mm
0108806373821974172902281060AHH936
60AHH936
2/28/2029
Lospa Knee System
01.10.1C6
Lospa Tibial Insert (CR) - Size 12 / 14mm
0108806373821974172909281060AJK237
60AJK237
9/28/2029
Lospa Knee System
01.10.1C8
Lospa Tibial Insert (CR) - Size 12 / 16mm
0108806373821981172902281060AHA133
60AHA133
2/28/2029
Lospa Knee System
01.10.232
Lospa Tibial Insert (PS) - Size 3 / 10mm
0108806373822353172511281070AEK536
70AEK536
11/28/2025
Lospa Knee System
01.10.232
Lospa Tibial Insert (PS) - Size 3 / 10mm
0108806373822353172605281070AFDL30
70AFDL30
5/28/2026
Lospa Knee System
01.10.232
Lospa Tibial Insert (PS) - Size 3 / 10mm
0108806373822353172701281070AGA137
70AGA137
1/28/2027
Lospa Knee System
01.10.232
Lospa Tibial Insert (PS) - Size 3 / 10mm
0108806373822353172812281070AJA334
70AJA334
12/28/2028
Lospa Knee System
01.10.232
Lospa Tibial Insert (PS) - Size 3 / 10mm
0108806373822353172903281070AJD632
70AJD632
3/28/2029
Lospa Knee System
01.10.234
Lospa Tibial Insert (PS) - Size 3 / 12mm
0108806373822360172503281070AEB930
70AEB930
3/28/2025
Lospa Knee System
01.10.234
Lospa Tibial Insert (PS) - Size 3 / 12mm
0108806373822360172506281070AEDF31
70AEDF31
6/28/2025
Lospa Knee System
01.10.234
Lospa Tibial Insert (PS) - Size 3 / 12mm
0108806373822360172603281070AFD534
70AFD534
3/28/2026
Lospa Knee System
01.10.234
Lospa Tibial Insert (PS) - Size 3 / 12mm
0108806373822360172611281070AFG431
70AFG431
11/28/2026
Lospa Knee System
01.10.234
Lospa Tibial Insert (PS) - Size 3 / 12mm
0108806373822360172710281070AGL433
70AGL433
10/28/2027
Lospa Knee System
01.10.236
Lospa Tibial Insert (PS) - Size 3 / 14mm
0108806373822377172407281070ADGA32
70ADGA32
7/28/2024
Lospa Knee System
01.10.236
Lospa Tibial Insert (PS) - Size 3 / 14mm
0108806373822377172412281070AEAC32
70AEAC32
12/28/2024
Lospa Knee System
01.10.236
Lospa Tibial Insert (PS) - Size 3 / 14mm
0108806373822377172512281070AEMM32
70AEMM32
12/28/2025
Lospa Knee System
01.10.236
Lospa Tibial Insert (PS) - Size 3 / 14mm
0108806373822377172710281070AGL434
70AGL434
10/28/2027
Lospa Knee System
01.10.236
Lospa Tibial Insert (PS) - Size 3 / 14mm
0108806373822377172812281070AHBR37
70AHBR37
12/28/2028
Lospa Knee System
01.10.238
Lospa Tibial Insert (PS) - Size 3 / 16mm
0108806373822384172601281070AFAR32
70AFAR32
1/28/2026
Lospa Knee System
01.10.238
Lospa Tibial Insert (PS) - Size 3 / 16mm
0108806373822384172711281070AGLR30
70AGLR30
11/28/2027
Lospa Knee System
01.10.23A
Lospa Tibial Insert (PS) - Size 3 / 18mm
0108806373822391172608281070AFH131
70AFH131
8/28/2026
Lospa Knee System
01.10.23C
Lospa Tibial Insert (PS) - Size 3 / 20mm
0108806373822407172807281070AFH132
70AFH132
7/28/2028
Lospa Knee System
01.10.242
Lospa Tibial Insert (PS) - Size 4 / 10mm
0108806373822551172503281070AEB933
70AEB933
3/28/2025
Lospa Knee System
01.10.244
Lospa Tibial Insert (PS) - Size 4 / 12mm
0108806373822568172603281070AFCR38
70AFCR38
3/28/2026
Lospa Knee System
01.10.246
Lospa Tibial Insert (PS) - Size 4 / 14mm
0108806373822575172503281070AEB935
70AEB935
3/28/2025
Lospa Knee System
01.10.246
Lospa Tibial Insert (PS) - Size 4 / 14mm
0108806373822575172603281070AFCR40
70AFCR40
3/28/2026
Lospa Knee System
01.10.246
Lospa Tibial Insert (PS) - Size 4 / 14mm
0108806373822575172710281070AGEY32
70AGEY32
10/28/2027
Lospa Knee System
01.10.246
Lospa Tibial Insert (PS) - Size 4 / 14mm
0108806373822575172808281070AHBR35
70AHBR35
8/28/2028
Lospa Knee System
01.10.248
Lospa Tibial Insert (PS) - Size 4 / 16mm
0108806373822582172502281070AEB936
70AEB936
2/28/2025
Lospa Knee System
01.10.248
Lospa Tibial Insert (PS) - Size 4 / 16mm
0108806373822582172703281070AGCL37
70AGCL37
3/28/2027
Lospa Knee System
01.10.248
Lospa Tibial Insert (PS) - Size 4 / 16mm
0108806373822582172710281070AGL437
70AGL437
10/28/2027
Lospa Knee System
01.10.248
Lospa Tibial Insert (PS) - Size 4 / 16mm
0108806373822582172811281070AHK731
70AHK731
11/28/2028
Lospa Knee System
01.10.248
Lospa Tibial Insert (PS) - Size 4 / 16mm
0108806373822582172902281070AJAK35
70AJAK35
2/28/2029
Lospa Knee System
01.10.248
Lospa Tibial Insert (PS) - Size 4 / 16mm
0108806373822582172903281070AJBD37
70AJBD37
3/28/2029
Lospa Knee System
01.10.248
Lospa Tibial Insert (PS) - Size 4 / 16mm
0108806373822582172907281070AJDT32
70AJDT32
7/28/2029
Lospa Knee System
01.10.24A
Lospa Tibial Insert (PS) - Size 4 / 18mm
0108806373822599172603281070AFD540
70AFD540
3/28/2026
Lospa Knee System
01.10.24A
Lospa Tibial Insert (PS) - Size 4 / 18mm
0108806373822599172701281070AFLJ32
70AFLJ32
1/28/2027
Lospa Knee System
01.10.24A
Lospa Tibial Insert (PS) - Size 4 / 18mm
0108806373822599172702281070AFLJ32
70AFLJ32
2/28/2027
Lospa Knee System
01.10.24A
Lospa Tibial Insert (PS) - Size 4 / 18mm
0108806373822599172710281070AGL438
70AGL438
10/28/2027
Lospa Knee System
01.10.24A
Lospa Tibial Insert (PS) - Size 4 / 18mm
0108806373822599172903281070AJBD38
70AJBD38
3/28/2029
Lospa Knee System
01.10.24C
Lospa Tibial Insert (PS) - Size 4 / 20mm
0108806373822605172601281070AFH135
70AFH135
1/28/2026
Lospa Knee System
01.10.24C
Lospa Tibial Insert (PS) - Size 4 / 20mm
0108806373822605172609281070AFH135
70AFH135
9/28/2026
Lospa Knee System
01.10.24C
Lospa Tibial Insert (PS) - Size 4 / 20mm
0108806373822605172807281070AHGK30
70AHGK30
7/28/2028
Lospa Knee System
01.10.24C
Lospa Tibial Insert (PS) - Size 4 / 20mm
0108806373822605172906281070AJFN30
70AJFN30
6/28/2029
Lospa Knee System
01.10.254
Lospa Tibial Insert (PS) - Size 5 / 12mm
0108806373822667172904281060AJBS31
60AJBS31
4/28/2029
Lospa Knee System
01.10.256
Lospa Tibial Insert (PS) - Size 5 / 14mm
0108806373822674172708281060AGGE37
60AGGE37
8/28/2027
Lospa Knee System
01.10.256
Lospa Tibial Insert (PS) - Size 5 / 14mm
0108806373822674172812281060AHMH30
60AHMH30
12/28/2028
Lospa Knee System
01.10.256
Lospa Tibial Insert (PS) - Size 5 / 14mm
0108806373822674172906281060AJBS32
60AJBS32
6/28/2029
Lospa Knee System
01.10.256
Lospa Tibial Insert (PS) - Size 5 / 14mm
0108806373822674172908281060AJG137
60AJG137
8/28/2029
Lospa Knee System
01.10.258
Lospa Tibial Insert (PS) - Size 5 / 16mm
0108806373822681172604281060AFDL40
60AFDL40
4/28/2026
Lospa Knee System
01.10.258
Lospa Tibial Insert (PS) - Size 5 / 16mm
0108806373822681172702281060AGB533
60AGB533
2/28/2027
Lospa Knee System
01.10.258
Lospa Tibial Insert (PS) - Size 5 / 16mm
0108806373822681172709281060AGGE38
60AGGE38
9/28/2027
Lospa Knee System
01.10.258
Lospa Tibial Insert (PS) - Size 5 / 16mm
0108806373822681172710281060AGJG36
60AGJG36
10/28/2027
Lospa Knee System
01.10.258
Lospa Tibial Insert (PS) - Size 5 / 16mm
0108806373822681172907281060AJEY35
60AJEY35
7/28/2029
Lospa Knee System
01.10.258
Lospa Tibial Insert (PS) - Size 5 / 16mm
0108806373822681172909281060AJEY35
60AJEY35
9/28/2029
Lospa Knee System
01.10.25A
Lospa Tibial Insert (PS) - Size 5 / 18mm
0108806373822698172512281060AEK535
60AEK535
12/28/2025
Lospa Knee System
01.10.25A
Lospa Tibial Insert (PS) - Size 5 / 18mm
0108806373822698172701281060AFLJ34
60AFLJ34
1/28/2027
Lospa Knee System
01.10.25A
Lospa Tibial Insert (PS) - Size 5 / 18mm
0108806373822698172710281060AGL440
60AGL440
10/28/2027
Lospa Knee System
01.10.25A
Lospa Tibial Insert (PS) - Size 5 / 18mm
0108806373822698172909281060AJKN31
60AJKN31
9/28/2029
Lospa Knee System
01.10.25C
Lospa Tibial Insert (PS) - Size 5 / 20mm
0108806373822704172808281060AHHT30
60AHHT30
8/28/2028
Lospa Knee System
01.10.264
Lospa Tibial Insert (PS) - Size 6 / 12mm
0108806373822766172701281000AGB331
00AGB331
1/28/2027
Lospa Knee System
01.10.266
Lospa Tibial Insert (PS) - Size 6 / 14mm
0108806373822773172701281000AGB332
00AGB332
1/28/2027
Lospa Knee System
01.10.266
Lospa Tibial Insert (PS) - Size 6 / 14mm
0108806373822773172801281000AGB332
00AGB332
1/28/2028
Lospa Knee System
01.10.268
Lospa Tibial Insert (PS) - Size 6 / 16mm
0108806373822780172806281000AFJ137
00AFJ137
6/28/2028
Lospa Knee System
01.10.274
Lospa Tibial Insert (PS) - Size 7 / 12mm
0108806373822865172503281060AEBA31
60AEBA31
3/28/2025
Lospa Knee System
01.10.274
Lospa Tibial Insert (PS) - Size 7 / 12mm
0108806373822865172902281060AJAK43
60AJAK43
2/28/2029
Lospa Knee System
01.10.276
Lospa Tibial Insert (PS) - Size 7 / 14mm
0108806373822872172501281060AEAD32
60AEAD32
1/28/2025
Lospa Knee System
01.10.276
Lospa Tibial Insert (PS) - Size 7 / 14mm
0108806373822872172701281060AGB133
60AGB133
1/28/2027
Lospa Knee System
01.10.276
Lospa Tibial Insert (PS) - Size 7 / 14mm
0108806373822872172808281060AHBR31
60AHBR31
8/28/2028
Lospa Knee System
01.10.276
Lospa Tibial Insert (PS) - Size 7 / 14mm
0108806373822872172812281060AJA744
60AJA744
12/28/2028
Lospa Knee System
01.10.278
Lospa Tibial Insert (PS) - Size 7 / 16mm
0108806373822889172603281060AFCM34
60AFCM34
3/28/2026
Lospa Knee System
01.10.278
Lospa Tibial Insert (PS) - Size 7 / 16mm
0108806373822889172605281060AFEN43
60AFEN43
5/28/2026
Lospa Knee System
01.10.278
Lospa Tibial Insert (PS) - Size 7 / 16mm
0108806373822889172701281060AFLJ36
60AFLJ36
1/28/2027
Lospa Knee System
01.10.278
Lospa Tibial Insert (PS) - Size 7 / 16mm
0108806373822889172705281060AGDG39
60AGDG39
5/28/2027
Lospa Knee System
01.10.278
Lospa Tibial Insert (PS) - Size 7 / 16mm
0108806373822889172710281060AGKD37
60AGKD37
10/28/2027
Lospa Knee System
01.10.278
Lospa Tibial Insert (PS) - Size 7 / 16mm
0108806373822889172807281060AHBR32
60AHBR32
7/28/2028
Lospa Knee System
01.10.27A
Lospa Tibial Insert (PS) - Size 7 / 18mm
0108806373822896172512281060AEK544
60AEK544
12/28/2025
Lospa Knee System
01.10.27A
Lospa Tibial Insert (PS) - Size 7 / 18mm
0108806373822896172608281060AFH149
60AFH149
8/28/2026
Lospa Knee System
01.10.27A
Lospa Tibial Insert (PS) - Size 7 / 18mm
0108806373822896172701281060AFLJ35
60AFLJ35
1/28/2027
Lospa Knee System
01.10.27A
Lospa Tibial Insert (PS) - Size 7 / 18mm
0108806373822896172708281060AGGE41
60AGGE41
8/28/2027
Lospa Knee System
01.10.27A
Lospa Tibial Insert (PS) - Size 7 / 18mm
0108806373822896172710281060AGL431
60AGL431
10/28/2027
Lospa Knee System
01.10.27A
Lospa Tibial Insert (PS) - Size 7 / 18mm
0108806373822896172711281060AGLR43
60AGLR43
11/28/2027
Lospa Knee System
01.10.27C
Lospa Tibial Insert (PS) - Size 7 / 20mm
0108806373822902172507281060AEG932
60AEG932
7/28/2025
Lospa Knee System
01.10.27C
Lospa Tibial Insert (PS) - Size 7 / 20mm
0108806373822902172807281060AHDB30
60AHDB30
7/28/2028
Lospa Knee System
01.10.27C
Lospa Tibial Insert (PS) - Size 7 / 20mm
0108806373822902172906281060AJFN32
60AJFN32
6/28/2029
Lospa Knee System
01.10.27C
Lospa Tibial Insert (PS) - Size 7 / 20mm
0108806373822902172908281060AJFN32
60AJFN32
8/28/2029
Lospa Knee System
01.10.288
Lospa Tibial Insert (PS) - Size 8 / 16mm
0108806373822988172701281070AGB337
70AGB337
1/28/2027
Lospa Knee System
01.10.288
Lospa Tibial Insert (PS) - Size 8 / 16mm
0108806373822988172908281070AGB337
70AGB337
8/28/2029
Lospa Knee System
01.10.28A
Lospa Tibial Insert (PS) - Size 8 / 18mm
0108806373822995172701281070AGB338
70AGB338
1/28/2027
Lospa Knee System
01.10.292
Lospa Tibial Insert (PS) - Size 9 / 10mm
0108806373823053172404281080ADCH30
80ADCH30
4/28/2024
Lospa Knee System
01.10.292
Lospa Tibial Insert (PS) - Size 9 / 10mm
0108806373823053172710281080AGKD38
80AGKD38
10/28/2027
Lospa Knee System
01.10.292
Lospa Tibial Insert (PS) - Size 9 / 10mm
0108806373823053172712281080AGLS41
80AGLS41
12/28/2027
Lospa Knee System
01.10.292
Lospa Tibial Insert (PS) - Size 9 / 10mm
0108806373823053172906281080AJFK35
80AJFK35
6/28/2029
Lospa Knee System
01.10.294
Lospa Tibial Insert (PS) - Size 9 / 12mm
0108806373823060172507281080AEET35
80AEET35
7/28/2025
Lospa Knee System
01.10.294
Lospa Tibial Insert (PS) - Size 9 / 12mm
0108806373823060172701281080AGB135
80AGB135
1/28/2027
Lospa Knee System
01.10.294
Lospa Tibial Insert (PS) - Size 9 / 12mm
0108806373823060172703281080AGC933
80AGC933
3/28/2027
Lospa Knee System
01.10.294
Lospa Tibial Insert (PS) - Size 9 / 12mm
0108806373823060172810281080AHLL30
80AHLL30
10/28/2028
Lospa Knee System
01.10.296
Lospa Tibial Insert (PS) - Size 9 / 14mm
0108806373823077172405281080ADCH34
80ADCH34
5/28/2024
Lospa Knee System
01.10.296
Lospa Tibial Insert (PS) - Size 9 / 14mm
0108806373823077172504281080AEBA36
80AEBA36
4/28/2025
Lospa Knee System
01.10.296
Lospa Tibial Insert (PS) - Size 9 / 14mm
0108806373823077172512281080AEMM42
80AEMM42
12/28/2025
Lospa Knee System
01.10.296
Lospa Tibial Insert (PS) - Size 9 / 14mm
0108806373823077172604281080AFDL48
80AFDL48
4/28/2026
Lospa Knee System
01.10.296
Lospa Tibial Insert (PS) - Size 9 / 14mm
0108806373823077172703281080AGCL32
80AGCL32
3/28/2027
Lospa Knee System
01.10.296
Lospa Tibial Insert (PS) - Size 9 / 14mm
0108806373823077172810281080AHLL31
80AHLL31
10/28/2028
Lospa Knee System
01.10.298
Lospa Tibial Insert (PS) - Size 9 / 16mm
0108806373823084172403281080ADBJ30
80ADBJ30
3/28/2024
Lospa Knee System
01.10.298
Lospa Tibial Insert (PS) - Size 9 / 16mm
0108806373823084172502281080AEBA37
80AEBA37
2/28/2025
Lospa Knee System
01.10.298
Lospa Tibial Insert (PS) - Size 9 / 16mm
0108806373823084172609281080AFH230
80AFH230
9/28/2026
Lospa Knee System
01.10.298
Lospa Tibial Insert (PS) - Size 9 / 16mm
0108806373823084172711281080AGLR48
80AGLR48
11/28/2027
Lospa Knee System
01.10.298
Lospa Tibial Insert (PS) - Size 9 / 16mm
0108806373823084172712281080AGLR48
80AGLR48
12/28/2027
Lospa Knee System
01.10.298
Lospa Tibial Insert (PS) - Size 9 / 16mm
0108806373823084172811281080AHHM31
80AHHM31
11/28/2028
Lospa Knee System
01.10.298
Lospa Tibial Insert (PS) - Size 9 / 16mm
0108806373823084172902281080AJA748
80AJA748
2/28/2029
Lospa Knee System
01.10.29A
Lospa Tibial Insert (PS) - Size 9 / 18mm
0108806373823091172406281080ADCH35
80ADCH35
6/28/2024
Lospa Knee System
01.10.29A
Lospa Tibial Insert (PS) - Size 9 / 18mm
0108806373823091172603281080AFD548
80AFD548
3/28/2026
Lospa Knee System
01.10.29A
Lospa Tibial Insert (PS) - Size 9 / 18mm
0108806373823091172610281080AFH231
80AFH231
10/28/2026
Lospa Knee System
01.10.29A
Lospa Tibial Insert (PS) - Size 9 / 18mm
0108806373823091172611281080AFH231
80AFH231
11/28/2026
Lospa Knee System
01.10.29A
Lospa Tibial Insert (PS) - Size 9 / 18mm
0108806373823091172611281080AFLJ38
80AFLJ38
11/28/2026
Lospa Knee System
01.10.29A
Lospa Tibial Insert (PS) - Size 9 / 18mm
0108806373823091172707281080AFLJ38
80AFLJ38
7/28/2027
Lospa Knee System
01.10.29A
Lospa Tibial Insert (PS) - Size 9 / 18mm
0108806373823091172710281080AGL442
80AGL442
10/28/2027
Lospa Knee System
01.10.29A
Lospa Tibial Insert (PS) - Size 9 / 18mm
0108806373823091172812281080AGLR49
80AGLR49
12/28/2028
Lospa Knee System
01.10.2B2
Lospa Tibial Insert (PS) - Size 11 / 10mm
0108806373823251172610281030AFG739
30AFG739
10/28/2026
Lospa Knee System
01.10.2B2
Lospa Tibial Insert (PS) - Size 11 / 10mm
0108806373823251172711281030AGKD40
30AGKD40
11/28/2027
Lospa Knee System
01.10.2B2
Lospa Tibial Insert (PS) - Size 11 / 10mm
0108806373823251172801281030AGLS43
30AGLS43
1/28/2028
Lospa Knee System
01.10.2B4
Lospa Tibial Insert (PS) - Size 11 / 12mm
0108806373823268172502281030AEBA40
30AEBA40
2/28/2025
Lospa Knee System
01.10.2B4
Lospa Tibial Insert (PS) - Size 11 / 12mm
0108806373823268172509281030AEHF43
30AEHF43
9/28/2025
Lospa Knee System
01.10.2B4
Lospa Tibial Insert (PS) - Size 11 / 12mm
0108806373823268172711281030AGKD41
30AGKD41
11/28/2027
Lospa Knee System
01.10.2B4
Lospa Tibial Insert (PS) - Size 11 / 12mm
0108806373823268172905281030AJCK32
30AJCK32
5/28/2029
Lospa Knee System
01.10.2B6
Lospa Tibial Insert (PS) - Size 11 / 14mm
0108806373823275172411281030ADM439
30ADM439
11/28/2024
Lospa Knee System
01.10.2B6
Lospa Tibial Insert (PS) - Size 11 / 14mm
0108806373823275172501281030AEAD40
30AEAD40
1/28/2025
Lospa Knee System
01.10.2B6
Lospa Tibial Insert (PS) - Size 11 / 14mm
0108806373823275172507281030AEBA41
30AEBA41
7/28/2025
Lospa Knee System
01.10.2B6
Lospa Tibial Insert (PS) - Size 11 / 14mm
0108806373823275172609281030AFG741
30AFG741
9/28/2026
Lospa Knee System
01.10.2B6
Lospa Tibial Insert (PS) - Size 11 / 14mm
0108806373823275172610281030AFG741
30AFG741
10/28/2026
Lospa Knee System
01.10.2B6
Lospa Tibial Insert (PS) - Size 11 / 14mm
0108806373823275172704281030AGDH36
30AGDH36
4/28/2027
Lospa Knee System
01.10.2B6
Lospa Tibial Insert (PS) - Size 11 / 14mm
0108806373823275172710281030AGL443
30AGL443
10/28/2027
Lospa Knee System
01.10.2B6
Lospa Tibial Insert (PS) - Size 11 / 14mm
0108806373823275172808281030AGLS45
30AGLS45
8/28/2028
Lospa Knee System
01.10.2B6
Lospa Tibial Insert (PS) - Size 11 / 14mm
0108806373823275172901281030AJAJ31
30AJAJ31
1/28/2029
Lospa Knee System
01.10.2B8
Lospa Tibial Insert (PS) - Size 11 / 16mm
0108806373823282172501281030AEAP36
30AEAP36
1/28/2025
Lospa Knee System
01.10.2B8
Lospa Tibial Insert (PS) - Size 11 / 16mm
0108806373823282172603281030AFD631
30AFD631
3/28/2026
Lospa Knee System
01.10.2B8
Lospa Tibial Insert (PS) - Size 11 / 16mm
0108806373823282172609281030AFH232
30AFH232
9/28/2026
Lospa Knee System
01.10.2B8
Lospa Tibial Insert (PS) - Size 11 / 16mm
0108806373823282172705281030AGB835
30AGB835
5/28/2027
Lospa Knee System
01.10.2B8
Lospa Tibial Insert (PS) - Size 11 / 16mm
0108806373823282172710281030AGL444
30AGL444
10/28/2027
Lospa Knee System
01.10.2B8
Lospa Tibial Insert (PS) - Size 11 / 16mm
0108806373823282172810281030AHFM31
30AHFM31
10/28/2028
Lospa Knee System
01.10.2BA
Lospa Tibial Insert (PS) - Size 11 / 18mm
0108806373823299172406281030ADCH38
30ADCH38
6/28/2024
Lospa Knee System
01.10.2BC
Lospa Tibial Insert (PS) - Size 11 / 20mm
0108806373823305172412281020ACC543
20ACC543
12/28/2024
Lospa Knee System
01.10.2BC
Lospa Tibial Insert (PS) - Size 11 / 20mm
0108806373823305172608281030AFJ133
30AFJ133
8/28/2026
Lospa Knee System
01.10.2BC
Lospa Tibial Insert (PS) - Size 11 / 20mm
0108806373823305172609281030AFJ133
30AFJ133
9/28/2026
Lospa Knee System
01.10.2C2
Lospa Tibial Insert (PS) - Size 12 / 10mm
0108806373823350172507281060AEHF30
60AEHF30
7/28/2025
Lospa Knee System
01.10.2C2
Lospa Tibial Insert (PS) - Size 12 / 10mm
0108806373823350172909281060AJJG30
60AJJG30
9/28/2029
Lospa Knee System
01.10.2C4
Lospa Tibial Insert (PS) - Size 12 / 12mm
0108806373823367172407281060ADF236
60ADF236
7/28/2024
Lospa Knee System
01.10.2C4
Lospa Tibial Insert (PS) - Size 12 / 12mm
0108806373823367172505281060AEDG36
60AEDG36
5/28/2025
Lospa Knee System
01.10.2C4
Lospa Tibial Insert (PS) - Size 12 / 12mm
0108806373823367172602281060AFBP37
60AFBP37
2/28/2026
Lospa Knee System
01.10.2C4
Lospa Tibial Insert (PS) - Size 12 / 12mm
0108806373823367172906281060AJEY38
60AJEY38
6/28/2029
Lospa Knee System
01.10.2C6
Lospa Tibial Insert (PS) - Size 12 / 14mm
0108806373823374172407281060ADF237
60ADF237
7/28/2024
Lospa Knee System
01.10.2C6
Lospa Tibial Insert (PS) - Size 12 / 14mm
0108806373823374172505281060AEDG37
60AEDG37
5/28/2025
Lospa Knee System
01.10.2C8
Lospa Tibial Insert (PS) - Size 12 / 16mm
0108806373823381172505281060AEDG38
60AEDG38
5/28/2025
Lospa Knee System
01.10.2C8
Lospa Tibial Insert (PS) - Size 12 / 16mm
0108806373823381172508281060AEJE42
60AEJE42
8/28/2025
Lospa Knee System
01.10.2C8
Lospa Tibial Insert (PS) - Size 12 / 16mm
0108806373823381172602281060AFBP38
60AFBP38
2/28/2026
Lospa Knee System
01.10.2C8
Lospa Tibial Insert (PS) - Size 12 / 16mm
0108806373823381172604281060AFCM32
60AFCM32
4/28/2026
Lospa Knee System
01.10.2C8
Lospa Tibial Insert (PS) - Size 12 / 16mm
0108806373823381172812281060AHF135
60AHF135
12/28/2028
Lospa Knee System
01.10.2CA
Lospa Tibial Insert (PS) - Size 12 / 18mm
0108806373823398172505281060AEDG39
60AEDG39
5/28/2025
Lospa Knee System
01.10.2CA
Lospa Tibial Insert (PS) - Size 12 / 18mm
0108806373823398172611281060AFM830
60AFM830
11/28/2026
Lospa Knee System
01.10.2CA
Lospa Tibial Insert (PS) - Size 12 / 18mm
0108806373823398172906281060AJEY41
60AJEY41
6/28/2029
Lospa Knee System
01.10.2CC
Lospa Tibial Insert (PS) - Size 12 / 20mm
0108806373823404172411281060ACCR35
60ACCR35
11/28/2024
Lospa Knee System
01.10.1E4
Lospa Tibial Insert (CR) - Size 14 / 12mm
0108806373852794172502281010aebg33
10aebg33
2/28/2025
Lospa Knee System
01.10.1E4
Lospa Tibial Insert (CR) - Size 14 / 12mm
0108806373852794172505281010AEDH40
10AEDH40
5/28/2025
Lospa Knee System
01.10.1E6
Lospa Tibial Insert (CR) - Size 14 / 14mm
0108806373852800172606281010AFBP44
10AFBP44
6/28/2026
Lospa Knee System
01.10.1EC
Lospa Tibial Insert (CR) - Size 14 / 20mm
0108806373852831172804281010AHDB44
10AHDB44
4/28/2028
Lospa Knee System
01.10.2E2
Lospa Tibial Insert (PS) - Size 14 / 10mm
0108806373852886172510281010AEJE43
10AEJE43
10/28/2025
Lospa Knee System
01.10.2E2
Lospa Tibial Insert (PS) - Size 14 / 10mm
0108806373852886172606281010AFCS30
10AFCS30
6/28/2026
Lospa Knee System
01.10.2E2
Lospa Tibial Insert (PS) - Size 14 / 10mm
0108806373852886172710281010AGBE30
10AGBE30
10/28/2027
Lospa Knee System
01.10.2E4
Lospa Tibial Insert (PS) - Size 14 / 12mm
0108806373852893172410281010ADL931
10ADL931
10/28/2024
Lospa Knee System
01.10.2E4
Lospa Tibial Insert (PS) - Size 14 / 12mm
0108806373852893172606281010AFCS31
10AFCS31
6/28/2026
Lospa Knee System
01.10.2E4
Lospa Tibial Insert (PS) - Size 14 / 12mm
0108806373852893172710281010AGB836
10AGB836
10/28/2027
Lospa Knee System
01.10.2E6
Lospa Tibial Insert (PS) - Size 14 / 14mm
0108806373852909172411281010ADL932
10ADL932
11/28/2024
Lospa Knee System
01.10.2E8
Lospa Tibial Insert (PS) - Size 14 / 16mm
0108806373852916172507281010AEHF45
10AEHF45
7/28/2025
Lospa Knee System
01.10.2E8
Lospa Tibial Insert (PS) - Size 14 / 16mm
0108806373852916172702281010AGB837
10AGB837
2/28/2027
Lospa Knee System
01.10.2E8
Lospa Tibial Insert (PS) - Size 14 / 16mm
0108806373852916172711281010AGB837
10AGB837
11/28/2027
Lospa Knee System
01.10.2EA
Lospa Tibial Insert (PS) - Size 14 / 18mm
0108806373852923172410281010ADL934
10ADL934
10/28/2024
Lospa Knee System
01.10.2EA
Lospa Tibial Insert (PS) - Size 14 / 18mm
0108806373852923172511281010AEJE47
10AEJE47
11/28/2025
Lospa Knee System
01.10.2EA
Lospa Tibial Insert (PS) - Size 14 / 18mm
0108806373852923172702281010AGB838
10AGB838
2/28/2027
Lospa Knee System
01.10.2EC
Lospa Tibial Insert (PS) - Size 14 / 20mm
0108806373852930172804281010AHDB32
10AHDB32
4/28/2028
Lospa Knee System
01.10.9EC
LOSPA Patella Component 38mm
0108806373853692172503281000AED827
00AED827
3/28/2025
Lospa Knee System
01.10.9EC
LOSPA Patella Component 38mm
0108806373853692172603281000AFAM25
00AFAM25
3/28/2026
Lospa Modular Knee System
K2.IBN.4061-10A
TIBIAL INSERT-FIXED-CONSTRAINED-03-10T
0108806373861161172706281010AGCN30
10AGCN30
6/28/2027
Lospa Modular Knee System
K2.IBN.4061-10A
TIBIAL INSERT-FIXED-CONSTRAINED-03-10T
0108806373861161172708281010AGHQ33
10AGHQ33
8/28/2027
Lospa Modular Knee System
K2.IBN.4061-10A
TIBIAL INSERT-FIXED-CONSTRAINED-03-10T
0108806373861161172712281010AGL830
10AGL830
12/28/2027
Lospa Modular Knee System
K2.IBN.4061-10A
TIBIAL INSERT-FIXED-CONSTRAINED-03-10T
0108806373861161172802281010AHAC38
10AHAC38
2/28/2028
Lospa Modular Knee System
K2.IBN.4061-12A
TIBIAL INSERT-FIXED-CONSTRAINED-03-12T
0108806373861185172708281010AGHQ34
10AGHQ34
8/28/2027
Lospa Modular Knee System
K2.IBN.4061-12A
TIBIAL INSERT-FIXED-CONSTRAINED-03-12T
0108806373861185172712281010AGL831
10AGL831
12/28/2027
Lospa Modular Knee System
K2.IBN.4061-12A
TIBIAL INSERT-FIXED-CONSTRAINED-03-12T
0108806373861185172802281010AHAC39
10AHAC39
2/28/2028
Lospa Modular Knee System
K2.IBN.4061-14A
TIBIAL INSERT-FIXED-CONSTRAINED-03-14T
0108806373861208172706281010AGCP31
10AGCP31
6/28/2027
Lospa Modular Knee System
K2.IBN.4061-14A
TIBIAL INSERT-FIXED-CONSTRAINED-03-14T
0108806373861208172712281010AGLA30
10AGLA30
12/28/2027
Lospa Modular Knee System
K2.IBN.4061-14A
TIBIAL INSERT-FIXED-CONSTRAINED-03-14T
0108806373861208172802281010AHAC40
10AHAC40
2/28/2028
Lospa Modular Knee System
K2.IBN.4061-16A
TIBIAL INSERT-FIXED-CONSTRAINED-03-16T
0108806373861222172706281010AGCP32
10AGCP32
6/28/2027
Lospa Modular Knee System
K2.IBN.4061-16A
TIBIAL INSERT-FIXED-CONSTRAINED-03-16T
0108806373861222172708281010AGHQ36
10AGHQ36
8/28/2027
Lospa Modular Knee System
K2.IBN.4061-16A
TIBIAL INSERT-FIXED-CONSTRAINED-03-16T
0108806373861222172712281010AGLA31
10AGLA31
12/28/2027
Lospa Modular Knee System
K2.IBN.4061-16A
TIBIAL INSERT-FIXED-CONSTRAINED-03-16T
0108806373861222172801281010AHAC41
10AHAC41
1/28/2028
Lospa Modular Knee System
K2.IBN.4061-18A
TIBIAL INSERT-FIXED-CONSTRAINED-03-18T
0108806373861246172706281010AGCP33
10AGCP33
6/28/2027
Lospa Modular Knee System
K2.IBN.4061-18A
TIBIAL INSERT-FIXED-CONSTRAINED-03-18T
0108806373861246172707281010AGHQ30
10AGHQ30
7/28/2027
Lospa Modular Knee System
K2.IBN.4061-18A
TIBIAL INSERT-FIXED-CONSTRAINED-03-18T
0108806373861246172710281010AGL930
10AGL930
10/28/2027
Lospa Modular Knee System
K2.IBN.4061-18A
TIBIAL INSERT-FIXED-CONSTRAINED-03-18T
0108806373861246172711281010AGL930
10AGL930
11/28/2027
Lospa Modular Knee System
K2.IBN.4061-18A
TIBIAL INSERT-FIXED-CONSTRAINED-03-18T
0108806373861246172802281010AHAC42
10AHAC42
2/28/2028
Lospa Modular Knee System
K2.IBN.4061-20A
TIBIAL INSERT-FIXED-CONSTRAINED-03-20T
0108806373861260172706281010AGG430
10AGG430
6/28/2027
Lospa Modular Knee System
K2.IBN.4061-20A
TIBIAL INSERT-FIXED-CONSTRAINED-03-20T
0108806373861260172708281010AGJ130
10AGJ130
8/28/2027
Lospa Modular Knee System
K2.IBN.4061-20A
TIBIAL INSERT-FIXED-CONSTRAINED-03-20T
0108806373861260172712281010AGLA32
10AGLA32
12/28/2027
Lospa Modular Knee System
K2.IBN.4061-23A
TIBIAL INSERT-FIXED-CONSTRAINED-03-23T
0108806373861291172809281010AHHN30
10AHHN30
9/28/2028
Lospa Modular Knee System
K2.IBN.4061-26A
TIBIAL INSERT-FIXED-CONSTRAINED-03-26T
0108806373861321172809281010AHHN31
10AHHN31
9/28/2028
Lospa Modular Knee System
K2.IBN.4264-10A
TIBIAL INSERT-FIXED-CONSTRAINED-04-10T
0108806373861352172708281010AGHQ37
10AGHQ37
8/28/2027
Lospa Modular Knee System
K2.IBN.4264-10A
TIBIAL INSERT-FIXED-CONSTRAINED-04-10T
0108806373861352172712281010AGL832
10AGL832
12/28/2027
Lospa Modular Knee System
K2.IBN.4264-10A
TIBIAL INSERT-FIXED-CONSTRAINED-04-10T
0108806373861352172803281010AHAC44
10AHAC44
3/28/2028
Lospa Modular Knee System
K2.IBN.4264-12A
TIBIAL INSERT-FIXED-CONSTRAINED-04-12T
0108806373861376172706281010AGCP35
10AGCP35
6/28/2027
Lospa Modular Knee System
K2.IBN.4264-12A
TIBIAL INSERT-FIXED-CONSTRAINED-04-12T
0108806373861376172708281010AGHQ38
10AGHQ38
8/28/2027
Lospa Modular Knee System
K2.IBN.4264-12A
TIBIAL INSERT-FIXED-CONSTRAINED-04-12T
0108806373861376172712281010AGL833
10AGL833
12/28/2027
Lospa Modular Knee System
K2.IBN.4264-14A
TIBIAL INSERT-FIXED-CONSTRAINED-04-14T
0108806373861390172708281010AGHQ39
10AGHQ39
8/28/2027
Lospa Modular Knee System
K2.IBN.4264-14A
TIBIAL INSERT-FIXED-CONSTRAINED-04-14T
0108806373861390172712281010AGL834
10AGL834
12/28/2027
Lospa Modular Knee System
K2.IBN.4264-16A
TIBIAL INSERT-FIXED-CONSTRAINED-04-16T
0108806373861413172708281010AGHQ40
10AGHQ40
8/28/2027
Lospa Modular Knee System
K2.IBN.4264-16A
TIBIAL INSERT-FIXED-CONSTRAINED-04-16T
0108806373861413172712281010AGL835
10AGL835
12/28/2027
Lospa Modular Knee System
K2.IBN.4264-18A
TIBIAL INSERT-FIXED-CONSTRAINED-04-18T
0108806373861437172706281010AGCP38
10AGCP38
6/28/2027
Lospa Modular Knee System
K2.IBN.4264-18A
TIBIAL INSERT-FIXED-CONSTRAINED-04-18T
0108806373861437172708281010AGHQ41
10AGHQ41
8/28/2027
Lospa Modular Knee System
K2.IBN.4264-18A
TIBIAL INSERT-FIXED-CONSTRAINED-04-18T
0108806373861437172712281010AGL836
10AGL836
12/28/2027
Lospa Modular Knee System
K2.IBN.4264-18A
TIBIAL INSERT-FIXED-CONSTRAINED-04-18T
0108806373861437172804281010AHAD30
10AHAD30
4/28/2028
Lospa Modular Knee System
K2.IBN.4264-20A
TIBIAL INSERT-FIXED-CONSTRAINED-04-20T
0108806373861451172706281010AGCP39
10AGCP39
6/28/2027
Lospa Modular Knee System
K2.IBN.4264-20A
TIBIAL INSERT-FIXED-CONSTRAINED-04-20T
0108806373861451172708281010AGJ131
10AGJ131
8/28/2027
Lospa Modular Knee System
K2.IBN.4264-20A
TIBIAL INSERT-FIXED-CONSTRAINED-04-20T
0108806373861451172712281010AGLA33
10AGLA33
12/28/2027
Lospa Modular Knee System
K2.IBN.4264-23A
TIBIAL INSERT-FIXED-CONSTRAINED-04-23T
0108806373861482172809281010AHHN32
10AHHN32
9/28/2028
Lospa Modular Knee System
K2.IBN.4264-26A
TIBIAL INSERT-FIXED-CONSTRAINED-04-26T
0108806373861512172708281010AGJF42
10AGJF42
8/28/2027
Lospa Modular Knee System
K2.IBN.4264-26A
TIBIAL INSERT-FIXED-CONSTRAINED-04-26T
0108806373861512172809281010AHHN33
10AHHN33
9/28/2028
Lospa Modular Knee System
K2.IBN.4467-10A
TIBIAL INSERT-FIXED-CONSTRAINED-05-10T
0108806373861543172706281010AGCN32
10AGCN32
6/28/2027
Lospa Modular Knee System
K2.IBN.4467-10A
TIBIAL INSERT-FIXED-CONSTRAINED-05-10T
0108806373861543172711281010AGL837
10AGL837
11/28/2027
Lospa Modular Knee System
K2.IBN.4467-12A
TIBIAL INSERT-FIXED-CONSTRAINED-05-12T
0108806373861567172706281010AGCP40
10AGCP40
6/28/2027
Lospa Modular Knee System
K2.IBN.4467-12A
TIBIAL INSERT-FIXED-CONSTRAINED-05-12T
0108806373861567172708281010AGHQ43
10AGHQ43
8/28/2027
Lospa Modular Knee System
K2.IBN.4467-14A
TIBIAL INSERT-FIXED-CONSTRAINED-05-14T
0108806373861581172706281010AGCP41
10AGCP41
6/28/2027
Lospa Modular Knee System
K2.IBN.4467-14A
TIBIAL INSERT-FIXED-CONSTRAINED-05-14T
0108806373861581172711281010AGL839
10AGL839
11/28/2027
Lospa Modular Knee System
K2.IBN.4467-16A
TIBIAL INSERT-FIXED-CONSTRAINED-05-16T
0108806373861604172711281010AGL840
10AGL840
11/28/2027
Lospa Modular Knee System
K2.IBN.4467-18A
TIBIAL INSERT-FIXED-CONSTRAINED-05-18T
0108806373861628172706281010AGCP43
10AGCP43
6/28/2027
Lospa Modular Knee System
K2.IBN.4467-18A
TIBIAL INSERT-FIXED-CONSTRAINED-05-18T
0108806373861628172711281010AGLA34
10AGLA34
11/28/2027
Lospa Modular Knee System
K2.IBN.4467-18A
TIBIAL INSERT-FIXED-CONSTRAINED-05-18T
0108806373861628172801281010AHAD36
10AHAD36
1/28/2028
Lospa Modular Knee System
K2.IBN.4467-20A
TIBIAL INSERT-FIXED-CONSTRAINED-05-20T
0108806373861642172706281010AGCP44
10AGCP44
6/28/2027
Lospa Modular Knee System
K2.IBN.4467-20A
TIBIAL INSERT-FIXED-CONSTRAINED-05-20T
0108806373861642172709281010AGJ132
10AGJ132
9/28/2027
Lospa Modular Knee System
K2.IBN.4467-20A
TIBIAL INSERT-FIXED-CONSTRAINED-05-20T
0108806373861642172711281010AGL931
10AGL931
11/28/2027
Lospa Modular Knee System
K2.IBN.4467-23A
TIBIAL INSERT-FIXED-CONSTRAINED-05-23T
0108806373861673172809281050AHHN34
50AHHN34
9/28/2028
Lospa Modular Knee System
K2.IBN.4467-26A
TIBIAL INSERT-FIXED-CONSTRAINED-05-26T
0108806373861703172708281010AGJF45
10AGJF45
8/28/2027
Lospa Modular Knee System
K2.IBN.4467-26A
TIBIAL INSERT-FIXED-CONSTRAINED-05-26T
0108806373861703172809281050AGJF45
50AGJF45
9/28/2028
Lospa Modular Knee System
K2.IBN.4467-26A
TIBIAL INSERT-FIXED-CONSTRAINED-05-26T
0108806373861703172809281050AHHN35
50AHHN35
9/28/2028
Lospa Modular Knee System
K2.IBN.4672-10A
TIBIAL INSERT-FIXED-CONSTRAINED-07-10T
0108806373861925172711281010AGL841
10AGL841
11/28/2027
Lospa Modular Knee System
K2.IBN.4672-12A
TIBIAL INSERT-FIXED-CONSTRAINED-07-12T
0108806373861949172711281010AGLY31
10AGLY31
11/28/2027
Lospa Modular Knee System
K2.IBN.4672-12A
TIBIAL INSERT-FIXED-CONSTRAINED-07-12T
0108806373861949172804281010AHAD39
10AHAD39
4/28/2028
Lospa Modular Knee System
K2.IBN.4672-14A
TIBIAL INSERT-FIXED-CONSTRAINED-07-14T
0108806373861963172708281010AGHR37
10AGHR37
8/28/2027
Lospa Modular Knee System
K2.IBN.4672-16A
TIBIAL INSERT-FIXED-CONSTRAINED-07-16T
0108806373861987172706281010AGCP47
10AGCP47
6/28/2027
Lospa Modular Knee System
K2.IBN.4672-16A
TIBIAL INSERT-FIXED-CONSTRAINED-07-16T
0108806373861987172708281010AGHR38
10AGHR38
8/28/2027
Lospa Modular Knee System
K2.IBN.4672-16A
TIBIAL INSERT-FIXED-CONSTRAINED-07-16T
0108806373861987172712281010AGL844
10AGL844
12/28/2027
Lospa Modular Knee System
K2.IBN.4672-16A
TIBIAL INSERT-FIXED-CONSTRAINED-07-16T
0108806373861987172808281010AHAD41
10AHAD41
8/28/2028
Lospa Modular Knee System
K2.IBN.4672-18A
TIBIAL INSERT-FIXED-CONSTRAINED-07-18T
0108806373862007172706281010AGCP48
10AGCP48
6/28/2027
Lospa Modular Knee System
K2.IBN.4672-18A
TIBIAL INSERT-FIXED-CONSTRAINED-07-18T
0108806373862007172709281010AGJF30
10AGJF30
9/28/2027
Lospa Modular Knee System
K2.IBN.4672-18A
TIBIAL INSERT-FIXED-CONSTRAINED-07-18T
0108806373862007172712281010AGLA35
10AGLA35
12/28/2027
Lospa Modular Knee System
K2.IBN.4672-18A
TIBIAL INSERT-FIXED-CONSTRAINED-07-18T
0108806373862007172806281010AHAD42
10AHAD42
6/28/2028
Lospa Modular Knee System
K2.IBN.4672-18A
TIBIAL INSERT-FIXED-CONSTRAINED-07-18T
0108806373862007172901281010AHAD42
10AHAD42
1/28/2029
Lospa Modular Knee System
K2.IBN.4672-18A
TIBIAL INSERT-FIXED-CONSTRAINED-07-18T
0108806373862007172906281010AHAD42
10AHAD42
6/28/2029
Lospa Modular Knee System
K2.IBN.4672-20A
TIBIAL INSERT-FIXED-CONSTRAINED-07-20T
0108806373862021172706281010AGCP49
10AGCP49
6/28/2027
Lospa Modular Knee System
K2.IBN.4672-20A
TIBIAL INSERT-FIXED-CONSTRAINED-07-20T
0108806373862021172708281010AGJ133
10AGJ133
8/28/2027
Lospa Modular Knee System
K2.IBN.4672-20A
TIBIAL INSERT-FIXED-CONSTRAINED-07-20T
0108806373862021172712281010AGLA36
10AGLA36
12/28/2027
Lospa Modular Knee System
K2.IBN.4672-23A
TIBIAL INSERT-FIXED-CONSTRAINED-07-23T
0108806373862052172809281010AHHN36
10AHHN36
9/28/2028
Lospa Modular Knee System
K2.IBN.4672-26A
TIBIAL INSERT-FIXED-CONSTRAINED-07-26T
0108806373862083172809281010AHHN37
10AHHN37
9/28/2028
Lospa Modular Knee System
K2.IBN.4874-10A
TIBIAL INSERT-FIXED-CONSTRAINED-09-10T
0108806373862304172708281010AGHS30
10AGHS30
8/28/2027
Lospa Modular Knee System
K2.IBN.4874-10A
TIBIAL INSERT-FIXED-CONSTRAINED-09-10T
0108806373862304172712281010AGL845
10AGL845
12/28/2027
Lospa Modular Knee System
K2.IBN.4874-10A
TIBIAL INSERT-FIXED-CONSTRAINED-09-10T
0108806373862304172806281010AHAD44
10AHAD44
6/28/2028
Lospa Modular Knee System
K2.IBN.4874-12A
TIBIAL INSERT-FIXED-CONSTRAINED-09-12T
0108806373862328172708281010AGHS31
10AGHS31
8/28/2027
Lospa Modular Knee System
K2.IBN.4874-12A
TIBIAL INSERT-FIXED-CONSTRAINED-09-12T
0108806373862328172711281010AGLA37
10AGLA37
11/28/2027
Lospa Modular Knee System
K2.IBN.4874-14A
TIBIAL INSERT-FIXED-CONSTRAINED-09-14T
0108806373862342172706281010AGCQ31
10AGCQ31
6/28/2027
Lospa Modular Knee System
K2.IBN.4874-14A
TIBIAL INSERT-FIXED-CONSTRAINED-09-14T
0108806373862342172708281010AGHS32
10AGHS32
8/28/2027
Lospa Modular Knee System
K2.IBN.4874-14A
TIBIAL INSERT-FIXED-CONSTRAINED-09-14T
0108806373862342172711281010AGLA38
10AGLA38
11/28/2027
Lospa Modular Knee System
K2.IBN.4874-14A
TIBIAL INSERT-FIXED-CONSTRAINED-09-14T
0108806373862342172801281010AHAD46
10AHAD46
1/28/2028
Lospa Modular Knee System
K2.IBN.4874-16A
TIBIAL INSERT-FIXED-CONSTRAINED-09-16T
0108806373862366172708281010AGHS33
10AGHS33
8/28/2027
Lospa Modular Knee System
K2.IBN.4874-16A
TIBIAL INSERT-FIXED-CONSTRAINED-09-16T
0108806373862366172711281010AGLA39
10AGLA39
11/28/2027
Lospa Modular Knee System
K2.IBN.4874-18A
TIBIAL INSERT-FIXED-CONSTRAINED-09-18T
0108806373862380172707281010AGHQ31
10AGHQ31
7/28/2027
Lospa Modular Knee System
K2.IBN.4874-18A
TIBIAL INSERT-FIXED-CONSTRAINED-09-18T
0108806373862380172711281010AGLA40
10AGLA40
11/28/2027
Lospa Modular Knee System
K2.IBN.4874-18A
TIBIAL INSERT-FIXED-CONSTRAINED-09-18T
0108806373862380172901281010AHAE30
10AHAE30
1/28/2029
Lospa Modular Knee System
K2.IBN.4874-18A
TIBIAL INSERT-FIXED-CONSTRAINED-09-18T
0108806373862380172906281010AHAE30
10AHAE30
6/28/2029
Lospa Modular Knee System
K2.IBN.4874-20A
TIBIAL INSERT-FIXED-CONSTRAINED-09-20T
0108806373862403172706281010AGCQ34
10AGCQ34
6/28/2027
Lospa Modular Knee System
K2.IBN.4874-20A
TIBIAL INSERT-FIXED-CONSTRAINED-09-20T
0108806373862403172708281010AGJ134
10AGJ134
8/28/2027
Lospa Modular Knee System
K2.IBN.4874-20A
TIBIAL INSERT-FIXED-CONSTRAINED-09-20T
0108806373862403172712281010AGLA41
10AGLA41
12/28/2027
Lospa Modular Knee System
K2.IBN.4874-23A
TIBIAL INSERT-FIXED-CONSTRAINED-09-23T
0108806373862434172809281010AHHN38
10AHHN38
9/28/2028
Lospa Modular Knee System
K2.IBN.4874-26A
TIBIAL INSERT-FIXED-CONSTRAINED-09-26T
0108806373862465172809281010AHHN39
10AHHN39
9/28/2028
Lospa Modular Knee System
K2.IBN.5076-10A
TIBIAL INSERT-FIXED-CONSTRAINED-11-10T
0108806373862687172708281010AGHQ82
10AGHQ82
8/28/2027
Lospa Modular Knee System
K2.IBN.5076-10A
TIBIAL INSERT-FIXED-CONSTRAINED-11-10T
0108806373862687172710281010AGL932
10AGL932
10/28/2027
Lospa Modular Knee System
K2.IBN.5076-10A
TIBIAL INSERT-FIXED-CONSTRAINED-11-10T
0108806373862687172802281010AHAE32
10AHAE32
2/28/2028
Lospa Modular Knee System
K2.IBN.5076-12A
TIBIAL INSERT-FIXED-CONSTRAINED-11-12T
0108806373862700172706281010AGCQ35
10AGCQ35
6/28/2027
Lospa Modular Knee System
K2.IBN.5076-12A
TIBIAL INSERT-FIXED-CONSTRAINED-11-12T
0108806373862700172708281010AGHS34
10AGHS34
8/28/2027
Lospa Modular Knee System
K2.IBN.5076-12A
TIBIAL INSERT-FIXED-CONSTRAINED-11-12T
0108806373862700172711281010AGLA42
10AGLA42
11/28/2027
Lospa Modular Knee System
K2.IBN.5076-12A
TIBIAL INSERT-FIXED-CONSTRAINED-11-12T
0108806373862700172802281010AHAE33
10AHAE33
2/28/2028
Lospa Modular Knee System
K2.IBN.5076-14A
TIBIAL INSERT-FIXED-CONSTRAINED-11-14T
0108806373862724172706281010AGCQ36
10AGCQ36
6/28/2027
Lospa Modular Knee System
K2.IBN.5076-14A
TIBIAL INSERT-FIXED-CONSTRAINED-11-14T
0108806373862724172708281010AGHS35
10AGHS35
8/28/2027
Lospa Modular Knee System
K2.IBN.5076-14A
TIBIAL INSERT-FIXED-CONSTRAINED-11-14T
0108806373862724172712281010AGLA43
10AGLA43
12/28/2027
Lospa Modular Knee System
K2.IBN.5076-14A
TIBIAL INSERT-FIXED-CONSTRAINED-11-14T
0108806373862724172803281010AHAE34
10AHAE34
3/28/2028
Lospa Modular Knee System
K2.IBN.5076-16A
TIBIAL INSERT-FIXED-CONSTRAINED-11-16T
0108806373862748172706281010AGCQ37
10AGCQ37
6/28/2027
Lospa Modular Knee System
K2.IBN.5076-16A
TIBIAL INSERT-FIXED-CONSTRAINED-11-16T
0108806373862748172708281010AGHS36
10AGHS36
8/28/2027
Lospa Modular Knee System
K2.IBN.5076-16A
TIBIAL INSERT-FIXED-CONSTRAINED-11-16T
0108806373862748172711281010AGLA44
10AGLA44
11/28/2027
Lospa Modular Knee System
K2.IBN.5076-16A
TIBIAL INSERT-FIXED-CONSTRAINED-11-16T
0108806373862748172803281010AHAE35
10AHAE35
3/28/2028
Lospa Modular Knee System
K2.IBN.5076-18A
TIBIAL INSERT-FIXED-CONSTRAINED-11-18T
0108806373862762172706281010AGCQ38
10AGCQ38
6/28/2027
Lospa Modular Knee System
K2.IBN.5076-18A
TIBIAL INSERT-FIXED-CONSTRAINED-11-18T
0108806373862762172708281010AGHS37
10AGHS37
8/28/2027
Lospa Modular Knee System
K2.IBN.5076-18A
TIBIAL INSERT-FIXED-CONSTRAINED-11-18T
0108806373862762172710281010AGL933
10AGL933
10/28/2027
Lospa Modular Knee System
K2.IBN.5076-18A
TIBIAL INSERT-FIXED-CONSTRAINED-11-18T
0108806373862762172803281010AGL933
10AGL933
3/28/2028
Lospa Modular Knee System
K2.IBN.5076-18A
TIBIAL INSERT-FIXED-CONSTRAINED-11-18T
0108806373862762172803281010AHAE36
10AHAE36
3/28/2028
Lospa Modular Knee System
K2.IBN.5076-20A
TIBIAL INSERT-FIXED-CONSTRAINED-11-20T
0108806373862786172709281010AGJ135
10AGJ135
9/28/2027
Lospa Modular Knee System
K2.IBN.5076-20A
TIBIAL INSERT-FIXED-CONSTRAINED-11-20T
0108806373862786172711281010AGLA45
10AGLA45
11/28/2027
Lospa Modular Knee System
K2.IBN.5076-20A
TIBIAL INSERT-FIXED-CONSTRAINED-11-20T
0108806373862786172803281010AHAE37
10AHAE37
3/28/2028
Lospa Modular Knee System
K2.IBN.5076-23A
TIBIAL INSERT-FIXED-CONSTRAINED-11-23T
0108806373862816172809281010AHHN40
10AHHN40
9/28/2028
Lospa Modular Knee System
K2.IBN.5076-26A
TIBIAL INSERT-FIXED-CONSTRAINED-11-26T
0108806373862847172809281010AHHN41
10AHHN41
9/28/2028
Lospa Modular Knee System
K2.IBN.5281-10A
TIBIAL INSERT-FIXED-CONSTRAINED-12-10T
0108806373862878172711281010AGLP30
10AGLP30
11/28/2027
Lospa Modular Knee System
K2.IBN.5281-10A
TIBIAL INSERT-FIXED-CONSTRAINED-12-10T
0108806373862878172806281010AHAE38
10AHAE38
6/28/2028
Lospa Modular Knee System
K2.IBN.5281-12A
TIBIAL INSERT-FIXED-CONSTRAINED-12-12T
0108806373862892172711281010AGLP31
10AGLP31
11/28/2027
Lospa Modular Knee System
K2.IBN.5281-12A
TIBIAL INSERT-FIXED-CONSTRAINED-12-12T
0108806373862892172806281010AHAE39
10AHAE39
6/28/2028
Lospa Modular Knee System
K2.IBN.5281-14A
TIBIAL INSERT-FIXED-CONSTRAINED-12-14T
0108806373862915172711281010AGLP32
10AGLP32
11/28/2027
Lospa Modular Knee System
K2.IBN.5281-14A
TIBIAL INSERT-FIXED-CONSTRAINED-12-14T
0108806373862915172806281010AHAE40
10AHAE40
6/28/2028
Lospa Modular Knee System
K2.IBN.5281-16A
TIBIAL INSERT-FIXED-CONSTRAINED-12-16T
0108806373862939172711281010AGLP33
10AGLP33
11/28/2027
Lospa Modular Knee System
K2.IBN.5281-16A
TIBIAL INSERT-FIXED-CONSTRAINED-12-16T
0108806373862939172806281010AHAE41
10AHAE41
6/28/2028
Lospa Modular Knee System
K2.IBN.5281-18A
TIBIAL INSERT-FIXED-CONSTRAINED-12-18T
0108806373862953172711281010AGLP34
10AGLP34
11/28/2027
Lospa Modular Knee System
K2.IBN.5281-20A
TIBIAL INSERT-FIXED-CONSTRAINED-12-20T
0108806373862977172711281010AGLP35
10AGLP35
11/28/2027
Lospa Modular Knee System
K2.IBN.5281-20A
TIBIAL INSERT-FIXED-CONSTRAINED-12-20T
0108806373862977172806281010AHAE43
10AHAE43
6/28/2028
Lospa Modular Knee System
K2.IBN.5281-23A
TIBIAL INSERT-FIXED-CONSTRAINED-12-23T
0108806373863004172809281010AHHN42
10AHHN42
9/28/2028
Lospa Modular Knee System
K2.IBN.5281-26A
TIBIAL INSERT-FIXED-CONSTRAINED-12-26T
0108806373863035172809281010AHHN43
10AHHN43
9/28/2028
Lospa Modular Knee System
K2.IBN.5684-10A
TIBIAL INSERT-FIXED-CONSTRAINED-14-10T
0108806373863257172711281010AGLP36
10AGLP36
11/28/2027
Lospa Modular Knee System
K2.IBN.5684-12A
TIBIAL INSERT-FIXED-CONSTRAINED-14-12T
0108806373863271172711281010AGLP37
10AGLP37
11/28/2027
Lospa Modular Knee System
K2.IBN.5684-14A
TIBIAL INSERT-FIXED-CONSTRAINED-14-14T
0108806373863295172711281010AGLP38
10AGLP38
11/28/2027
Lospa Modular Knee System
K2.IBN.5684-16A
TIBIAL INSERT-FIXED-CONSTRAINED-14-16T
0108806373863318172711281010AGLP39
10AGLP39
11/28/2027
Lospa Modular Knee System
K2.IBN.5684-18A
TIBIAL INSERT-FIXED-CONSTRAINED-14-18T
0108806373863332172711281010AGLP40
10AGLP40
11/28/2027
Lospa Modular Knee System
K2.IBN.5684-20A
TIBIAL INSERT-FIXED-CONSTRAINED-14-20T
0108806373863356172711281010AGLP41
10AGLP41
11/28/2027
Lospa Modular Knee System
K2.IBN.5684-23A
TIBIAL INSERT-FIXED-CONSTRAINED-14-23T
0108806373863387172809281010AHHN44
10AHHN44
9/28/2028
Lospa Modular Knee System
K2.IBN.5684-26A
TIBIAL INSERT-FIXED-CONSTRAINED-14-26T
0108806373863417172809281010AHHN45
10AHHN45
9/28/2028
Lospa Knee System
01.10.1G2
Lospa Tibial Insert (CR) - Size 16 / 10mm
0108806373868580172706281010AFDF35
10AFDF35
6/28/2027
Lospa Knee System
01.10.1G4
Lospa Tibial Insert (CR) - Size 16 / 12mm
0108806373868597172706281010AFDF36
10AFDF36
6/28/2027
Lospa Knee System
01.10.1G6
Lospa Tibial Insert (CR) - Size 16 / 14mm
0108806373868603172706281010AFDF37
10AFDF37
6/28/2027
Lospa Knee System
01.10.1G8
Lospa Tibial Insert (CR) - Size 16 / 16mm
0108806373868610172706281010AFDF38
10AFDF38
6/28/2027
Lospa Knee System
01.10.1GA
Lospa Tibial Insert (CR) - Size 16 / 18mm
0108806373868627172706281010AFDF39
10AFDF39
6/28/2027
Lospa Knee System
01.10.2G2
Lospa Tibial Insert (PS) - Size 16 / 10mm
0108806373868689172706281010AFDF30
10AFDF30
6/28/2027
Lospa Knee System
01.10.2G4
Lospa Tibial Insert (PS) - Size 16 / 12mm
0108806373868696172706281010AFDF31
10AFDF31
6/28/2027
Lospa Knee System
01.10.2G6
Lospa Tibial Insert (PS) - Size 16 / 14mm
0108806373868702172706281010AFDF32
10AFDF32
6/28/2027
Lospa Knee System
01.10.2G8
Lospa Tibial Insert (PS) - Size 16 / 16mm
0108806373868719172706281010AFDF33
10AFDF33
6/28/2027
Lospa Knee System
01.10.2GA
Lospa Tibial Insert (PS) - Size 16 / 18mm
0108806373868726172706281010AFDF34
10AFDF34
6/28/2027 |
Recalling Firm/ Manufacturer |
CORENTEC CO., LTD 12 Yeongsanhong1-Gil; Seobuk Cheonan Korea (the Republic of)
|
| For Additional Information Contact | Michael Son 949-290-9885 |
Manufacturer Reason for Recall | Due to unsupported 10 year expiration date. |
FDA Determined Cause 2 | Labeling Change Control |
| Action | On 03/29/2024, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that on 7/1/2020 the firm changed the labeling of their polyethylene (PE) and patella inserts from a 5 year shelf life to a 10 year shelf life. This change was made without adequate documentation evidence to support the increase in shelf life.
Customers are instructed to:
1. Inform individuals within your organization who need to be aware of this device recall.
2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility.
3. Quarantine and discontinue use of the recalled devices.
4. Maintain awareness of this notice internally until all required actions have been completed within your facility.
5. Inform Corentec if any of the subject devices have been distributed to other organizations.
a. Please provide Corentec contact details so that Corentec can inform the recipients appropriately.
6. Please complete the attached customer response form (acknowledgement form). It is possible that you no longer have any physical inventory of the impacted product on site. Completing this form will allow us to update our records and will also negate the need for us to send any further communications on this matter. Therefore, please complete the customer response form even if you no longer have any of the subject devices in your physical inventory.
7. Return the completed form to Corentec.
For any questions or assistance contact Bobby Pham at 310-488-2886 or via email Bobby.Pham@corentec-US.com. |
| Quantity in Commerce | 397 |
| Distribution | US Nationwide distribution in the states of AL, AZ, CA, MI, PA, TX. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JWH
|
|
|
|