• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Synchro SELECT Guidewire

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Synchro SELECT Guidewiresee related information
Date Initiated by FirmApril 18, 2024
Date PostedMay 31, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1940-2024
Recall Event ID 94543
510(K)NumberK202522 
Product Classification Guide, wire, catheter, neurovasculature - Product Code MOF
ProductThe Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.
Code Information Catalog/UDI-DI/Lots: SSFT215PRE/07613327508598/0000139994, 0000144321, 0000146842, 0000154408; SSFT215STR/07613327508574/0000122897, 0000123151, 0000123405, 0000124931, 0000125040, 0000125719, 0000125720, 0000125787, 0000126356, 0000129220, 0000129221, 0000130861, 0000131723, 0000132954, 0000133179, 0000137045, 0000138580, 0000138581, 0000142280, 0000142281, 0000143349, 0000143584; SSFT300STR/07613327508581/0000144375, 0000155022; SSTD215PRE/07613327508512/0000122145, 0000122866, 0000127369, 0000139986, 0000140672, 0000141825, 0000141826, 0000143802, 0000146926, 0000149561, 0000154409, 0000154410, 0000154411, 0000176472; SSTD215STR/07613327508499/0000122481, 0000122485, 0000122650, 0000123069, 0000123070, 0000123143, 0000125409, 0000127494, 0000127495, 0000128254, 0000128446, 0000128447, 0000131661, 0000131667, 0000131668, 0000133396, 0000133563, 0000133564, 0000140971, 0000141392, 0000141393, 0000142532, 0000143125, 0000143209, 0000143806, 0000143834, 0000144140, 0000148427, 0000148428, 0000148477, 0000149560, 0000154679, 0000154680, 0000154681, 0000155021, 0000155662, 0000156792, 0000164588, 0000164589, 0000164591, 0000172898, 0000172899, 0000172900, 0000172911, 0000175987, 0000176018, 0000176019; SSTD300STR/07613327508505/0000145095, 0000145694, 0000151281; SSUP215PRE/07613327508635/0000144317, 0000148425; SSUP300PRE/07613327508642/0000145692, 0000151280; SSUP215STR/07613327508611/0000128253, 0000129625, 0000137050, 0000140671, 0000141784, 0000143342, 0000144318, 0000145693, 0000145875, 0000164662, 0000165477; SSUP300STR/07613327508628/0000121919, 0000141789, 0000145691
FEI Number 3008853977
Recalling Firm/
Manufacturer
Stryker Neurovascular
47900 Bayside Pkwy
Fremont CA 94538-6515
Manufacturer Reason
for Recall
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
FDA Determined
Cause 2
Component design/selection
ActionOn 4/29/24, recall notices were mailed to Supply Chain Management/ Recall Coordinator/ Inventory Manager customers who were asked to do the following: 1) When using impacted guidewires, please do not backload the guidewire, follow the Instructions for Use (IFU) provided with the device, and do not use the guidewire if you see any visible signs of coating damage. 2) Circulate this customer notification internally to all interested/affected parties. 3) Inform firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that the firm can inform the recipients appropriately. 4) Inform firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: NVCustomerCare@stryker.com) 5) Complete and return the customer response form via email to: nvfieldactions@stryker.com. 6) Contact the firm if you would like a replacement of the guidewire with a new optional introducer accessory.
Quantity in Commerce41,074
DistributionWorldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, DE, AK, WY, MT, NM, SD, MS, RI, GU and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, JORDAN, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UK.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = MOF
-
-