Class 2 Device Recall NovaGuide 2 Traditional Ultrasound

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall NovaGuide 2 Traditional Ultrasoundsee related information
Date Initiated by FirmMay 06, 2024
Date PostedJune 27, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2217-2024
Recall Event ID 94673
510(K)NumberK213279 
Product Classification System, imaging, pulsed doppler, ultrasonic - Product Code IYN
ProductNovaGuide 2 Traditional Ultrasound , REF: NSC-TCDNG2
Code Information UDI: 00850681007405/Serial Numbers: 21092746 21122802 21122803 21092748 21092749 21122804 21123066 23063826 22074381 23063828 22033706 22064206 22074380 21092747 22113115 23063825 21122805 21082646 23063856 23104174 21123068 22103022 22064204 22064205 21123067 22074378 22074379 23033578
FEI Number 3013368530
Recalling Firm/
Manufacturer
NovaSignal Corp.
2440 S Sepulveda Blvd
Los Angeles CA 90064-1784
For Additional Information ContactNeurasignal customer service
1877-638-5251 Ext. 3
Manufacturer Reason
for Recall
Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.
FDA Determined
Cause 2
Device Design
ActionOn May 6, 2024, NeuraSignal Inc. dba NovaSignal issued a "Urgent-Medical Device Recall" notification to affected consignees. Nova Signal asked consignees to take the following actions: 1. The NeuraSignal service and customer support team will schedule a time for the onsite field correction. 2. In the timeframe leading up to the scheduled onsite field correction, the system can be used as intended. 2a. Users of the device should refrain from accessing the enclosed cable connections during routine operation. This may lead to unintentional disconnection of the chassis ground cable from the AC mains power distribution unit. It is recommended that the cart compartment remains locked while the device is commissioned for use. 3. The site should confirm that the grounding cable has not been unintentionally disconnected and maintains a secure connection to the power distribution unit 4. If necessary, resources from the Biomedical Equipment department may perform re-qualification or pre-installation qualification testing, to ensure the system is adequately grounded.
Quantity in Commerce28 systems
DistributionWorldwide - US Nationwide distribution in the states of MA, TX, NY, MN, AZ, AL, CO, CA, OH, MO, FL, TN, NJ, GA, OK, KY, WA and the country of Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.