| | Class 1 Device Recall Medline procedure kits |  |
| Date Initiated by Firm | April 08, 2024 |
| Date Posted | August 08, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2444-2024 |
| Recall Event ID |
94580 |
| Product Classification |
Cardiac catheterization kit - Product Code OES
|
| Product | Medline procedure kits labeled as:
1) WMC HYBRID OR PACK PART 1 , Pack Number CMPJ08068F ;
2) VASCULAR PACK , Pack Number CMPJ11455B ;
3) SALEM CATH LAB-LF , Pack Number DYNJ26938J ;
4) EH PEDIATRIC CATH , Pack Number DYNJ49920L ;
5) ARTERIOGRAM PACK, Pack Number DYNJ53018;
6) CATH PACK , Pack Number DYNJ57729C ;
7) PEDS CATH PACK, Pack Number DYNJ66856;
8) PEDS CATH PACK, Pack Number DYNJ66856A ;
9) STEWARD ROCKLEDGE CATH RRS, Pack Number DYNJ81061A ;
10) CATH LAB PACK , Pack Number DYNJ84617;
11) UA/UV CATHETERIZATION TRAY, Pack Number UVT1190A |
| Code Information |
CMPJ08068F , Lot Number 24ABN950 ;
CMPJ11455B , Lot Number 23IBD357 ;
DYNJ26938J , Lot Number 22FBD709 ;
DYNJ26938J , Lot Number 22GBH459 ;
DYNJ26938J , Lot Number 22HBR225 ;
DYNJ26938J , Lot Number 22IBJ562 ;
DYNJ26938J , Lot Number 22JBP469 ;
DYNJ26938J , Lot Number 23ABM365 ;
DYNJ26938J , Lot Number 23CBM109 ;
DYNJ26938J , Lot Number 23DBN340 ;
DYNJ26938J , Lot Number 23GBH016 ;
DYNJ26938J , Lot Number 23IBS867 ;
DYNJ53018, Lot Number 21IBX266 ;
DYNJ53018, Lot Number 22GBN555 ;
DYNJ53018, Lot Number 23ABN207 ;
DYNJ57729C , Lot Number 23HBK882 ;
DYNJ57729C , Lot Number 24ABO025 ;
DYNJ66856, Lot Number 21KBH003 ;
DYNJ66856A , Lot Number 22DBD158 ;
DYNJ66856A , Lot Number 22FBY433 ;
DYNJ66856A , Lot Number 22GBK576 ;
DYNJ66856A , Lot Number 22HBQ139 ;
DYNJ66856A , Lot Number 22IBO303 ;
DYNJ66856A , Lot Number 22JBO861 ;
DYNJ66856A , Lot Number 22JBQ914 ;
DYNJ66856A , Lot Number 23ABT952 ;
DYNJ66856A , Lot Number 23CBF157 ;
DYNJ66856A , Lot Number 23EBE461 ;
DYNJ66856A , Lot Number 23EBI477 ;
DYNJ66856A , Lot Number 23HBD973 ;
DYNJ81061A , Lot Number 23HBN396 ;
DYNJ84617, Lot Number 23FBR368 ;
DYNJ49920L , Lot Number 23KMB066 ;
UVT1190A , Lot Number 23BLA187 ;
UVT1190A , Lot Number 23DLA246 ;
UVT1190A , Lot Number 23KLA117 ;
UVT1190A , Lot Number 23LLA329
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
Manufacturer Reason for Recall | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA.
Distributors were directed to notify their customers.
The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
Link for the posting: https://www.medline.com/media/mkt/pdf/Shenli%20Syringe%20Kitting%20Recall%20info.pdf
If you have any questions, please contact the Recall Department at 866-359-1704. |
| Quantity in Commerce | 1,679,067 units in total |
| Distribution | Worldwide Distribution: US (nationwide) and OUS countries of: AE, AI, BB, CA, CL, CY, DE, GT, HN, JP, MX, MY, PA, PE, PR, SA, SG, TC, TT, VI, PR, TT, MX, BM, ES, NI, DO, GU, CR, AT, CD, DE, DF, W, and ON. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|