 Class 2 Device Recall Centurionsee related information
Date Initiated by FirmApril 08, 2024
Date PostedAugust 28, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2779-2024
Recall Event ID 94586
Product Classification Labor and Delivery Kit - Product Code MLS
ProductCenturion procedure packs, containing Caina syringes, labeled as: 1) ARTHROGRAM TRAY, Pack Number MNS11965A; 2) CIRCUMCISION TRAY, Pack Number CIT6720; 3) CIRCUMCISION TRAY, Pack Number CIT6720A
Code Information MNS11965A, Lot Number 2023040690; CIT6720, Lot Number 2022122390; CIT6720A, Lot Number 2023110890
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request that the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-