• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Centurion

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Centurionsee related information
Date Initiated by FirmApril 08, 2024
Date PostedAugust 28, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2780-2024
Recall Event ID 94586
Product Classification Circumcision tray - Product Code OHG
ProductCenturion procedure packs, containing Caina syringes, labeled as: 1) BREAST BIOPSY KIT, Pack Number DYNDH1512A; 2) CIRCUMCISION KIT, Pack Number CIT6535; 3) CIRCUMCISION KIT, Pack Number CIT7095; 4) CIRCUMCISION TRAY, Pack Number CIT6860A; 5) CIRCUMCISION TRAY, Pack Number DYNDA2271; 6) CIRCUMCISION TRAY, Pack Number CIT6720; 7) CIRCUMCISION TRAY, Pack Number CIT7020; 8) MEDICAL CIRCUMCISION TRAY, Pack Number CIT6780; 9) NICU CIRCUMCISION TRAY W/OUT CLAMP, Pack Number CIT6920
Code Information DYNDH1512A, Lot Number 2023030190; CIT6535, Lot Number 2023090580; CIT7095, Lot Number 2020121190; CIT7095, Lot Number 2022031590; CIT6860A, Lot Number 2021111580; CIT6860A, Lot Number 2022010480; CIT6860A, Lot Number 2022041880; CIT6860A, Lot Number 2022053080; CIT6860A, Lot Number 2022062080; CIT6860A, Lot Number 2022121280; CIT6860A, Lot Number 2023032080; CIT6860A, Lot Number 2023062680; CIT6860A, Lot Number 2023071780; CIT6860A, Lot Number 2023090580; CIT6860A, Lot Number 2023091180; CIT6860A, Lot Number 2023111380; CIT6860A, Lot Number 2024020580; DYNDA2271, Lot Number 2021072890; CIT6720, Lot Number 2021080390; DYNDA2271, Lot Number 2021083190; DYNDA2271, Lot Number 2021100790; DYNDA2271, Lot Number 2022010390; CIT6720, Lot Number 2021121590; DYNDA2271, Lot Number 2022042590; CIT7020, Lot Number 2022061390; CIT6720, Lot Number 2022072790; CIT6720, Lot Number 2022091690; CIT7020, Lot Number 2022110990; CIT6720, Lot Number 2022110990; CIT6720, Lot Number 2023063090; DYNDA2271, Lot Number 2023063090; CIT7020, Lot Number 2023080890; DYNDA2271, Lot Number 2023112990; CIT7020, Lot Number 2024020890; DYNDA2271, Lot Number 2024030690; CIT6780, Lot Number 2021110350; CIT6780, Lot Number 2022070750; CIT6780, Lot Number 2022072150; CIT6780, Lot Number 2023030190; CIT6780, Lot Number 2024022890; CIT6920, Lot Number 2022042590
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request that the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-