| Date Initiated by Firm | May 31, 2024 |
| Date Posted | June 25, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2180-2024 |
| Recall Event ID |
94777 |
| 510(K)Number | K971452 |
| Product Classification |
Table, radiographic, non-tilting, powered - Product Code IZZ
|
| Product | AXIOM MULTIX MT - Intended Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO and TOP Systems radiographic devices
Material Number: 8395399 |
| Code Information |
UDI-DI: N/A
Serial Numbers:
1003
1032
1039
1060
1065
1071
1076
1082
1083
1085
1089
1092
1104
1170
1174
1181
1183
1194
1304
1305
1309
1310
1315
1318
1323
1325
1331
1348
1359
1382
1395
1398
1423
1443
1444
1447
1449
1452
1453
1468
1498
1500
1508
1509
1512
1513
1515
1516
1525
1531
|
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 40 Liberty Blvd Malvern PA 19355-1418
|
| For Additional Information Contact | SAME 610-219-4834 |
Manufacturer Reason for Recall | Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Siemens issued Urgent Medical Device Correction Letter (XP008/24/S) to the
Attention: Director of Diagnostic Imaging and/ or Radiology Laboratory on 5/31/24.
Letter states reason for recall, health risk, and action to take:
What steps can the user take to avoid the possible risks associated with this issue?
It is recommended to switch off the system / Touch Display when it is not in use and is unattended.
How will the issue be resolved?
To ensure continued safe operation of your system and to provide safeguards for the life of the system, the
Siemens Healthineers Service Organization will perform the replacement of the PSU of the Touch Display via UI
XP006/24/S beginning in the second quarter (June) of 2024. This replacement will be provided to you free of
charge. Based on our investigation you can continue to use your system following the recommendation to switch
off the system / Touch Display when it is not in use and is unattended.
Our service organization will contact you to arrange a date to perform this corrective action once it is available.
Please feel free to contact our Customer Care Center at 1-800-888-7436 for an earlier appointment beginning in June 2024.
Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. Please respond within 14 days of receiving this notification.
|
| Quantity in Commerce | 50 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|