 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmApril 08, 2024
Date PostedSeptember 06, 2024
Recall Status1 Open3, Classified
Recall NumberZ-3034-2024
Recall Event ID 94583
Product Classification Container, specimen, sterile - Product Code FMH
ProductMedline Convenience kits labeled as: 1) KIT, MIDSTREAM, W/BZK/FUNNL,3O, Pack Number DYND30212; 2) PROCESSING KIT 3 PTE0000204 , Pack Number DYNJ65728A ; 3) 30ML KIT, Pack Number DYNJ82155A ; 4) DISSECTION, Pack Number DYNJ85191; 5) FLAP HARVEST PACK , Pack Number DYNJ58575C ; 6) KIT, MIDSTREAM, W/BZK/FUNNL,3O, Pack Number DYND30212H ; 7) FLAP HARVEST PACK, Pack Number DYNJ58575C
Code Information DYND30212, Lot Number 20CBV412 ; DYND30212H , Lot Number 20CBV412 ; DYNJ58575C, Lot Number 22FMF260; DYNJ58575C, Lot Number 22EMI154; DYNJ58575C , Lot Number 22JMF580 ; DYNJ65728A , Lot Number 21IBJ821 ; DYNJ65728A , Lot Number 21IBV204 ; DYNJ65728A , Lot Number 21JBE157 ; DYNJ82155A , Lot Number 23CBD960 ; DYNJ82155A , Lot Number 23CBH024 ; DYNJ82155A , Lot Number 23DBL430 ; DYNJ82155A , Lot Number 23EBI493 ; DYNJ82155A , Lot Number 23FBO135 ; DYNJ82155A , Lot Number 23FBO795 ; DYNJ82155A , Lot Number 23GBJ181 ; DYNJ82155A , Lot Number 23HBJ559 ; DYNJ82155A , Lot Number 23IBJ041 ; DYNJ82155A , Lot Number 23IBJ042 ; DYNJ85191, Lot Number 23JBP679
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-