• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Merit Medical

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Merit Medicalsee related information
Date Initiated by FirmJune 11, 2024
Date PostedAugust 12, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2560-2024
Recall Event ID 94797
Product Classification Angiography/angioplasty kit - Product Code OEQ
ProductVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K02 (Custom Syringe Kit), REF: K02-00873A, Inject fluids into, or withdraw fluids from, the body.
Code Information Catalog Number: K02-00873A UDI-DI code: 00884450156261 Lot Numbers: H2597212 H2620955 H2641148 Catalog Number: K02-01105A UDI-DI code: 00884450156124 H2486384 H2542259 H2628908 Catalog Number: K02-01481A UDI-DI code: 00884450164464 H2387157 H2418377 H2472043 H2486729 H2507912 H2517550 H2526240 H2552843 H2595093 H2628968 H2655130 H2678790 H2699921 H2712471 H2747868 H2822453 H2831126 H2846806 H2868403 H2884580 H2897149 H2916452
Recalling Firm/
Manufacturer
Merit Medical Systems, Inc.
1600 W Merit Pkwy
South Jordan UT 84095-2416
For Additional Information Contact
801-253-1600
Manufacturer Reason
for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA Determined
Cause 2
No Marketing Application
ActionOn 06/11/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTICE" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization. In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/. Customer are instructed to: 1. Immediately determine if any of the products identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution. 2. Ensure that appropriate personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them. Additional distribution details may be required by health authorities. 4. Follow the attached instructions for placing the recall labels on the products containing Jiangsu Shenli syringes, which inform the users to not use the affected syringes. If additional recall labels are needed, contact Customer Service at (801) 208-4381. Product may be returned to inventory once properly over-labeled. 5.At the point of use, the Jiangsu Shenli syringes are to be discarded. Photographs of Jiangsu Shenli syringes are enclosed to assist in identification. 6.Post this URGENT MEDICAL DEVICE RECALL NOTICE and Jiangsu Shenli Syringe photographs on or near the affected products. 7.If replacement syringes are needed, contact Customer Service at (801) 208-4381 for recommended syringe suppliers. For questions contact Merit Sales Representative or Merit Customer Service at 801-208-4381, Hours: 6 am to 6 pm MST Mon.-Fri.
Quantity in Commerce3,647 kits
DistributionU.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-