| | Class 2 Device Recall Merit Medical |  |
| Date Initiated by Firm | June 11, 2024 |
| Date Posted | August 12, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2565-2024 |
| Recall Event ID |
94797 |
| Product Classification |
Universal drainage tray - Product Code PPD
|
| Product | Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes:
K10/K10-PI (Custom Waste Management Kit), REF:
K10T-05815
K10T-05816
K10-04722C
K10-03591KP
K10-05530
K10-04291AP
K10-04386CP
K10T-05227
K10-05120P
K10-03220AP
K10-05251AP
K10-04479P
K10-01722QP
K10-04931B
K10-04429
K10-06012P
K10-05043A
K10-06066
K10-06011AP
K10-06033AP
K10-06086P
K10-05222
K10T-00856B
K10T-01343F
K10T-03849G
K10T-05668D
K10T-05403
K10T-03958L
K10T-05139
K10-05343
K10T-04662A
K10-04800DP
K10T-05428
K10T-04278C
K10T-04942A
K10T-04216G
K10T-04110B
K10-04201GP
K10-01571C
K10T-05261
K10-04183CP
K10-00839FP
K10-00785LP
K10-00777BP
K10-04413A
K10-00592K
K10-05097A
K10-05774P
K10-05636A
K10-03327D
K10-00814L
K10-04360B
K10-03383BP
K10-01270T
K10-04621
K10-05913P
K10-05472
K10-04693D
K10-03290D
K10T-04335A
K10-04426B
K10-01700AP
K12-05018
K10-05274AP
K10-05511
K10T-04155C
K10T-03903F
K10T-04978H
K10T-03997F
K10-03364A
K10-05833P
K10-05097B
K10-05700
K10-03184
K10-05457AP
K10-05675
K10-03666B
To support the withdrawal and collection of fluids from the body. |
| Code Information |
Catalog Number/UDI-DI code/Lot Numbers:
K10-01722QP 00884450570340
H2410331
H2471342
H2480639
H2502357
H2513044
H2525139
H2543652
H2558245
H2564522
H2825008
H2831600
H2875331
H2915643
K10-03220AP 00884450174036
H2406448
H2408885
H2423250
H2428461
H2448942
H2500653
H2536816
H2567302
H2601107
H2623288
H2659637
H2665379
H2708979
H2721135
H2730471
H2760266
H2774855
H2794124
H2812819
H2829027
H2838364
H2843344
H2850151
H2868190
H2872165
H2905227
H2913401
K10-03591KP 00884450380635
H2391366
H2408970
H2441713
H2472906
H2482293
H2491990
H2496930
H2500665
H2518450
H2527822
H2551825
H2567305
H2627298
H2638378
H2645033
H2708977
H2716700
H2730479
H2751466
H2778878
H2784096
H2794135
H2808194
H2821498
H2918877
K10-04291AP 00884450176320
H2352552
H2401233
H2418439
H2431501
H2461658
H2475592
H2492022
H2511077
H2513895
H2518283
H2530047
H2538940
H2551884
H2563710
H2567356
H2641969
H2662164
H2679955
H2713607
H2728230
H2744412
H2785947
H2795948
H2809437
H2818171
H2832163
H2843600
H2866062
H2896827
H2912839
H2918929
K10-04386CP 00884450197523
H2384157
H2401236
H2423252
H2444355
H2470464
H2496947
H2522107
H2559517
H2573434
H2655272
H2659665
H2665397
H2708975
H2730487
H2744361
H2746709
H2755852
H2778887
H2834781
H2838372
H2844785
H2872168
H2882630
H2897046
H2913305
H2918921
K10-04429 00884450176795
H2418279
H2448704
H2468945
H2489657
H2505650
H2518225
H2585617
H2600807
H2642069
H2642462
H2665625
H2692001
H2703157
H2728629
H2735292
H2756121
H2776259
H2795556
H2814363
H2832784
H2845927
H2875164
H2913298
H2928657
K10-04479P 00884450177075
H2409371
H2475562
K10-04722C 00884450270431
H2429014
H2496323
H2520764
H2557465
H2589898
H2626879
H2666564
H2736535
H2777762
H2799974
H2828177
H2859884
H2864805
H2905908
K10-04931B 00884450393147
H2402920
H2469836
H2474257
H2476022
H2480793
H2494885
H2751615
H2782208
H2787882
H2800804
H2815006
H2824607
H2871355
K10-05043A 00884450202074
H2610631
H2627090
H2680755
H2728653
H2875168
H2928687
K10-05120P 00884450197202
H2399127
H2431312
H2498451
H2523383
H2527704
H2553771
H2576585
H2655265
H2669130
H2720156
H2757798
H2831587
H2837463
H2847025
H2867088
H2890107
H2915648
H2920730
K10-05222 00884450230824
H2839171
H2849535
K10-05251AP 00884450265499
H2408968
H2428471
H2448946
H2469092
H2475567
H2562411
H2567311
H2585822
H2627308
H2633766
H2655299
H2665378
H2675063
H2708968
H2744380
H2755857
H2774890
H2784101
H2789811
H2802599
H2829053
H2843351
H2877200
H2891663
H2928416
K10-05530 00884450284407
H2581869
K10-06011AP 00884450634516
H2752358
H2771050
H2818172
H2849571
H2866057
K10-06012P 00884450681107
H2464776
H2507506
H2507672
H2518503
H2576565
H2585848
H2641118
H2650735
H2707287
H2763420
H2900287
H2915472
K10-06033AP 00884450663707
H2780818
H2804119
H2816921
H2831647
H2837566
H2867095
H2890128
H2910148
K10-06066 00884450695524
H2744543
H2824018
H2869593
H2884336
H2912589
K10-06086P 00884450695111
H2804075
H2813661
H2831607
H2867102
K10T-05227 00884450619957
T2736090
T2842237
T2862543
T2899709
K10T-05815 00884450486061
T2143908
T2310781
T2323193
T2330789
T2340443
T2363413
T2363416
T2414967
T2427552
T2437298
T2474307
T2534990
T2534991
T2576721
T2580922S
K10T-05816 00884450486115
T2198303
T2237161
T2238504
T2259632
T2259636
T2259646
T2323194
T2330794
T2330795
T2330796
T2363411
T2363412
T2363417
T2370536
T2399814
T2417139
T2427554
T2427562
T2456678
T2469555
T2474316
T2474987
T2474987S
T2534987
T2534988
T2542105
T2553454
T2553455
T2553457
T2580725S
T2590403
T2626099
K10-00592K 00884450235140
H2393452
H2420172
H2443412
H2451992
H2470701
H2476538
H2521149
H2521296
H2544998
H2577007
H2613404
H2642940
H2697212
H2707453
H2744182
H2763384
H2792508
H2846785
H2850808
H2864775
H2911026
K10-00777BP 00884450171356
H2382837
H2412350
H2412351
H2439664
H2454880
H2469915
H2496127
H2521359
H2532904
H2549693
H2564318
H2608724
H2630383
H2671723
H2676769
H2681498
H2729338
H2744246
H2782393
H2796088
H2800051
H2823930
H2828210
H2900295
H2908480
H2928030
K10-00785LP 00884450171363
H2354646
H2359761
H2381074
H2500634
H2576482
H2592743
H2633681
H2746688
H2829020
K10-00814L 00884450375211
H2391149
H2429178
H2429179
H2474142
H2519369
H2702874
H2709084
H2729408
H2865738
H2912314
H2932289
K10-00839FP 00884450171509
H2353207
H2442209
H2474355
H2559946
H2582189
H2666829
H2703436
H2730574
H2770405
H2777617
H2795425
H2819611
H2828354
H2847065
H2853704
H2859507
H2905308
K10-01270T 00884450320792
H2406427
H2413487
H2480743
H2519372
H2585454
H2596107
H2600560
H2615014
H2705266
H2717708
H2729411
H2753894
H2767717
H2781958
H2805418
H2824729
H2846005
H2879898
H2885139
K10-01571C 00884450172704
H2437814
H2469804
H2475959
H2519374
H2537897
H2562540
H2585455
H2591201
H2607153
H2705250
H2713444
H2738386
H2749188
H2767718
H2781960
H2800719
H2828298
H2865687
H2912221
H2927507
K10-01700AP 00884450172896
H2432646
H2475553
H2513924
H2518194
H2523106
H2546141
H2627293
H2633685
H2684906
H2755834
H2760262
H2789795
H2802581
H2834774
H2850144
H2896990
K10-03184 00884450173978
H2723085
K10-03290D 00884450174173
H2398057
H2429642
H2440121
H2470702
H2476533
H2497323
H2521293
H2552981
H2569338
H2593800
H2602335
H2626906
H2657735
H2736546
H2758364
H2767991
K10-03327D 00884450358474
H2399047
H2475961
H2490868
H2500928
H2642098
H2699127
H2763063
H2772973
H2781982
H2805423
H2814903
H2866206
H2916581
K10-03364A 00884450174302
H2767140
H2787066
H2825503
K10-03383BP 00884450174470
H2413766
H2431314
H2442065
H2451601
H2480561
H2531173
H2543647
H2559858
H2678789
H2690630
H2729477
H2746472
H2780753
H2802285
H2818310
H2847026
H2870440
H2895058
K10-03666B 00884450175088 H2823758
K10-04183CP 00884450176054
H2351711
H2354650
H2432649
H2469087
H2530021
H2556100
H2576500
H2596269
H2615644
H2669945
H2675061
H2680064
H2721140
H2746703
H2755848
H2765532
H2778883
H2798349
H2838371
H2844086
H2857414
H2882625
H2897043
H2905266
H2912943
H2928402
K10-04201GP 00884450431429
H2381217
H2405971
H2453053
H2474162
H2482360
H2488359
H2500676
H2516291
H2517062
H2544212
H2549177
H2551826
K10-04360B 00884450176511
H2402908
H2434453
H2454179
H2475964
H2500929
H2519386
H2537904
H2550456
H2562543
H2591207
H2625331
H2658559
H2709092
H2713458
H2724386
H2738402
H2743422
H2758297
H2767724
H2772976
H2781983
H2787737
H2796458
H2800724
H2814909
H2824736
H2837110
H2857775
H2885153
H2912269
H2916582
K10-04413A 00884450176740
H2386135
H2412595
H2417532
H2470430
H2575472
H2591208
H2607155
H2615025
H2642082
H2653570
H2705208
H2814913
H2866220
K10-04426B 00884450435441
H2413490
H2448841
H2475965
H2519387
H2537905
H2566457
H2630955
H2709095
K10-04621 00884450177457
H2406989
H2413470
H2444262
H2475966
H2511862
H2519388
H2684812
H2709283
H2713461
H2824738
H2837113
H2860555
H2871428
H2893756
H2915982
K10-04693D 00884450262481
H2407705
H2428330
H2464965
H2475323
H2518222
H2551727
H2596200
H2627088
H2703135
H2747760
H2787067
H2800132
H2814364
H2869637
K10-04800DP 00884450196021
H2488218
H2520856
H2595396
H2614413
H2641388
H2655270
H2690568
H2708832
H2720157
H2764416
H2777638
H2795332
H2809399
H2824993
H2838132
H2844551
H2859356
H2867094
H2880073
H2890106
H2910624
K10-05097A 00884450197073
H2386175
H2448851
H2519426
H2575499
H2607205
H2633708
H2654764
K10-05097B 00884450756508
H2685180
H2781343
H2837271
H2849973
H2866212
H2913308
K10-05274AP 00884450659519
H2587850
H2655380
H2666828
H2720256
H2745963
H2770428
H2780848
H2789392
H2813743
H2842017
H2875374
H2899222
H2905310
H2921244
K10-05343 00884450235560
H2529935
H2555969
H2592259
H2600916
H2675783
H2685118
H2703568
H2735167
H2762172
H2781243
H2795447
H2819538
H2881378
K10-05457AP 00884450312780
H2746387
H2764528
H2813694
H2831628
H2837542
H2841993
H2844694
H2867091
H2875345
H2890127
H2899252
K10-05472 00884450270639
H2409929
H2413270
H2498456
H2559869
H2641711
K10-05511 00884450283646 H2448561
K10-05636A 00884450394014
H2391156
H2397178
H2424068
H2448843
H2490871
H2585460
H2615031
H2749195
H2767728
K10-05675 00884450400944
H2753581
H2837273
K10-05700 00884450466759
H2744209
H2753582
H2791763
H2800136
H2814373
H2828011
H2837274
H2849975
H2875169
K10-05774P 00884450555910
H2402965
H2414102
H2500788
H2564127
H2592795
H2675065
H2684903
H2721147
H2739358
H2746717
H2770024
H2843360
H2868179
H2877212
H2913283
K10-05833P 00884450552957 H2751209
K10-05913P 00884450522356
H2418662
H2442106
H2453336
H2488235
H2498463
H2528155
H2547566
H2576600
H2581874
H2636097
K10T-00856B 00884450257630
T2131617
T2194472
T2239571
T2268933
T2310883
T2329171
T2360494
T2417861
T2465749
T2479569
T2505793
T2527123
T2587256
T2596741
T2652615
T2698965
T2712047
T2786089
T2805471
T2840601
T2842270
T2862040
T2898698
K10T-01343F 00884450238516
T2700152
T2712393
T2741709
T2775305
T2805997
T2848063
T2872006
T2885311
K10T-03849G 00884450201459
T2256824
T2299853
T2322984
T2333571
T2355740
T2387061
T2407485
T2429021
T2437094
T2499082
T2521173
T2527253
T2550792
T2572078
T2610045
T2655959
T2677529
T2701730
T2727027
T2761356
T2807112
T2832505
T2869837
T2910727
K10T-03903F 00884450257593
T2587257
T2625819
T2652624
T2708973
K10T-03958L 00884450414910
T2167604
T2261967
T2305936
T2322986
T2340254
T2370031
T2404523
T2429022
T2446367
T2470234
T2505381
T2509730
T2527260
T2536900
T2609506
T2670069
T2745524
T2760426
T2775349
T2830814
T2864822
T2899047
K10T-03997F 00884450447017 T2662074
K10T-04110B 00884450320624
T2363588
T2382425
T2457932
T2479583
T2498833
T2536906
T2581366
T2625820
T2677550
T2746584
T2752959
T2801981
T2863415
K10T-04155C 00884450267554 T2643976
K10T-04216G 00884450240878
T2360553
T2397204
T2412578
T2449689
T2494475
T2527503
T2550727
T2587447
T2627403
T2680829
T2713192
T2750452
T2758255
T2818275
T2862473
T2877074
T2918375
K10T-04278C 00884450318836
T2360557
T2397183
T2527527S
T2550628
T2620405
T2747270
T2761384
T2816472
T2824136
T2928269
K10T-04335A 00884450413852
T2732959
T2762843
T2921111
K10T-04662A 00884450306109
T2283344
T2306126
T2330944
T2352352
T2371538
T2404616
T2489387
T2537750
T2543102
T2570353
T2609818
T2658520
T2682252
T2740099
T2773384
T2779965
T2792493
T2816379
T2843421
T2862488
T2906192
T2918402
K10T-04942A 00884450319659
T2351570
T2366382
T2397184
T2429120
T2443500
T2465880
T2499019
T2499664
T2515748
T2521872
T2527598
T2581434
T2595447
T2606462
T2643847
T2680804
T2773330
T2779967
T2816321
T2833110
T2902199
T2906195
T2927195
K10T-04978H 00884450501245
T2626084
T2709021
T2750457
T2816449
T2848391
T2849852
T2905868
K10T-05139 00884450388495 T2604766
K10T-05261 00884450320297
T2408189
T2417985
T2446370
T2479737
T2637165
T2736369
T2781930
T2821650
T2848452
T2870228
T2884484
T2898771
T2927200
K10T-05403 00884450247235
T2485404
T2527717
T2584669
T2670915
T2754726
T2833118
T2899048
K10T-05428 00884450264195
T2339967
T2371539
T2383202
T2404674
T2427545
T2434463
T2458517
T2516379
T2527722
T2578263
T2587584
T2609819
T2644043
T2680608
T2701830
T2718892
T2750466
T2775548
T2823560
T2827728
T2848575
T2862546
T2902677
T2906263
T2921192
K10T-05668D 00884450486702
T2256883
T2310963
T2333977
T2355746
T2399998
T2404953
T2429306
T2498021
T2499043
T2511779
T2550738
T2572042
T2606490
T2643781
T2670926
T2699281
T2738156
T2774643
T2775584
T2800622
T2827649
T2842495
T2862612
T2898777
T2902696
T2918477
K12-05018 00884450193778
H2428338
H2615366
H2753586
H2800139
H2857780 |
| FEI Number |
1721504
|
Recalling Firm/ Manufacturer |
Merit Medical Systems, Inc. 1600 W Merit Pkwy South Jordan UT 84095-2416
|
| For Additional Information Contact | 801-253-1600 |
Manufacturer Reason for Recall | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. |
FDA Determined Cause 2 | No Marketing Application |
| Action | On 06/11/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTICE" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization. In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/.
Customer are instructed to:
1. Immediately determine if any of the products identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution.
2. Ensure that appropriate personnel within their organization are made aware of this field action.
3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them. Additional distribution details may be required by health authorities.
4. Follow the attached instructions for placing the recall labels on the products containing
Jiangsu Shenli syringes, which inform the users to not use the affected syringes. If additional recall labels are needed, contact Customer Service at (801) 208-4381. Product may be returned to inventory once properly over-labeled.
5.At the point of use, the Jiangsu Shenli syringes are to be discarded. Photographs of Jiangsu Shenli syringes are enclosed to assist in identification.
6.Post this URGENT MEDICAL DEVICE RECALL NOTICE and Jiangsu Shenli Syringe photographs on or near the affected products.
7.If replacement syringes are needed, contact Customer Service at (801) 208-4381 for recommended syringe suppliers.
For questions contact Merit Sales Representative or Merit Customer Service at 801-208-4381, Hours: 6 am to 6 pm MST Mon.-Fri.
|
| Quantity in Commerce | 138471 |
| Distribution | U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|