Class 2 Device Recall BVI CustomEyes Procedure Packs

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BVI CustomEyes Procedure Packssee related information
Date Initiated by FirmJuly 26, 2024
Date PostedAugust 28, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2894-2024
Recall Event ID 95068
Product Classification Kit, surgical instrument, disposable - Product Code KDD
ProductBVI CustomEyes Procedure Packs containing Sol-M Luer Lock and Luer Slip 1ml syringes, Part Number: 58001631;
Code Information Part Number: 58001631; UDI-DI: 30886158031780; Lot/Batch Number: 6061721
Recalling Firm/
Manufacturer
Beaver Visitec International, Inc.
500 Totten Pond Rd
10 City Point
Waltham MA 02451-1916
For Additional Information ContactAgnieszka Drzewiecka
781-906-7950
Manufacturer Reason
for Recall
Due to a voluntary recall of low dead space syringes, and luer slip tip syringes manufactured by Sol-Millennium Medical (Sol-M), Inc. These syringes manufactured by Sol-M are present in some of BVI CustomEyes Procedure Packs.
FDA Determined
Cause 2
No Marketing Application
ActionOn July 27, 2024 URGENT FIELD SAFETY NOTICE letters were sent to consignees. Actions required of you 1. Please immediately determine if any of the affected CustomEyes Procedure Packs are within your facility. In such case, quarantine them, and discontinue use until correct recall labels are applied. Please refer to Appendix 1 CustomEyes Procedure Packs List 2. If you have further distributed this product, please notify your customers by initiating a subrecall. Consider all potential users of this product in your user supply chain. You are encouraged to use a copy of this recall notification letter when contacting your customers. 3. Please follow the attached instructions for placing the recall labels on the products containing Sol-M syringes, which inform the users to not use the affected syringes. If additional recall labels are needed, contact bvi2958@sedgwick.com. Product may be returned to inventory once properly over-labeled. Refer to Appendix 2 Over labeling Instructions 4. At the point of clinical use, the Sol-M syringes are to be discarded. Photographs of Sol-M syringes are enclosed to assist in identification. Appendix 3 Images 5. Ensure that appropriate personnel within your organization are made aware of this field action 6. To maintain awareness, we suggest to post this URGENT MEDICAL DEVICE RECALL NOTICE and Sol-M syringe photographs on or near the affected products in your inventory location 7. Please fill out, scan and email the completed Business Response Form to BVI Sedgwick at bvi2958@sedgwick.com within seven (7) days BVI is committed to providing high quality products to you and apologizes for any inconvenience this field action may cause. If you have any questions, please reach out to your Sales Representative.
Quantity in Commerce84 units
DistributionUS Nationwide distribution.