• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Mindray

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Mindraysee related information
Date Initiated by FirmJuly 22, 2024
Date PostedSeptember 04, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2949-2024
Recall Event ID 95099
510(K)NumberK201957 
Product Classification Gas-machine, anesthesia - Product Code BSZ
ProductSevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number 115-066755-00
Code Information UDI-DI: 06936415975784 Serial Number: AH6D28000517, AH6D28000518
FEI Number 2221819
Recalling Firm/
Manufacturer
Mindray DS USA, Inc. dba Mindray North America
800 Macarthur Blvd
Mahwah NJ 07430-2001
For Additional Information ContactSAME
800-288-2121
Manufacturer Reason
for Recall
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
FDA Determined
Cause 2
Nonconforming Material/Component
ActionMindray issued an "URGENT Medical Device Recall" letter on July 22, 2024. Letter states reason for recall, health risk and actions to be taken: The V90 electronic vaporizer can continue to be used normally. If "Vaporizer Output Abnormal" alarm or other alarms associated with abnormal vaporizer output or high anesthetic agent alarms occur during use, take appropriate intervention per the A9 Operators manual and contact Mindray Technical Support for assistance. The intervention could be switching to another V90 electronic vaporizer, or the user can decide to switch to intravenous anesthesia. For vaporizer replacement, refer to A9 Operators manual part number H-046-017199-00, section 3.4. Vaporizer . The manual is available on the Mindray website. Mindray s Technical Support team can assist and may be reached at (877) 913-9663 (Option 1) Monday through Friday, 8:30 a.m. 5:30 p.m. ET. Please ensure that applicable clinical staff are aware of the content of this letter. Mindray will replace all affected V90 electronic vaporizers located at your facility. The V90 electronic vaporizers will be shipped directly to your facility, and a Mindray Care Team representative will contact you to schedule the installation and collect the affected V90 electronic vaporizers Contact: Director, Quality Operations and Regulatory Affairs US Agent for Shenzhen Mindray Biomedical Electronics Corp., Ltd Mindray DS USA, Inc. 201.995.8407
Quantity in Commerce2 units
DistributionUS Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = BSZ
-
-