 Class 2 Device Recall Spectral CTsee related information
Date Initiated by FirmNovember 16, 2024
Date PostedDecember 03, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0583-2025
Recall Event ID 95810
510(K)NumberK203020 
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductSpectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.
Code Information UDI-DI: (01)00884838101111. Software Version: 5.0.X. Model Number (Serial Numbers): 728333 (10266, 10261, 10264, 10219, 10265, 10197, 10182, 10175, 10192, 10179, 10196, 10180, 10188, 10190, 10181, 10183, 10178, 10186, 10185, 10199, 10184, 10198, 10193, 10187, 10189, 10195, 10191, 10194, 10018, 10023, 10027, 10062, 10072, 10078, 10079, 10080, 10084, 10088, 10101, 10104, 10112, 10117, 10113, 10116, 10032, 10049, 10133, 10140, 10152, 10153, 10166, 10170, 10171, 10173, 10172, 10174, 10016, 10015, 10017, 10019, 10022, 10021, 10020, 10025, 10024, 10033, 10031, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10043, 10042, 10014, 10044, 10046, 10048, 10045, 10047, 10051, 10050, 10053, 10055, 10052, 10056, 10058, 10057, 10059, 10060, 10061, 10065, 10066, 10068, 10067, 10070, 10071, 10063, 10069, 10074, 10075, 10077, 10076, 10081, 10082, 10087, 10089, 10085, 10086, 10091, 10094, 10093, 10095, 10096, 10090, 10099, 10098, 10102, 10100, 10103, 10097, 10105, 10106, 10107, 10108, 10109, 396001, 396003, 10110, 10111, 10114, 10115, 10122, 10118, 10120, 10121, 10123, 10119, 10124, 10125, 10126, 10127, 10129, 10128, 10131, 10130, 10132, 10135, 10136, 10134, 10137, 10139, 10138, 10142, 10141, 10143, 10144, 10145, 10146, 10147, 10149, 10150, 10151, 10155, 10156, 10159, 10148, 10160, 10161, 10162, 10158, 10164, 10163, 10165, 10167, 10168, 10169, 10013, 10177, 10176, 10210, 10218, 10217, 10206, 10202, 10200, 10211, 10207, 10204, 396004, 10208, 10201, 10205, 10203, 10212, 10213, 10225, 10214, 10220, 10215, 10222, 10227, 10247, 10209, 10226, 10230, 10228, 10232, 10231, 10229, 10221, 10233, 10234, 10235, 10237, 10238, 10223, 10239, 10241, 10242, 10236, 10245, 396005, 10248, 10249, 10243, 10244, 10250, 10259, 10258, 10157, 10262, 10246, 10255, 10254, 10240, 10251, 10256, 10224, 10253, 10257, 10260, 10263), 728340 (397060, 397061, 397022, 397023, 397020, 397024, 397021, 397025, 397027, 397031, 397026, 397033, 397032, 397030, 397028, 397029, 397015, 397016, 397004, 397007, 397005, 397002, 397006, 397008, 397010, 397003, 397011, 397013, 397009, 397012, 397014, 397017, 397019, 397018, 397036, 397037, 397038, 397034, 397041, 397040, 397039, 397044, 397046, 397043, 397045, 397047, 397048, 397049, 397053, 397050, 397054, 397055, 397056, 397058, 397059), 728344 (397057).
FEI Number 3016618143
Recalling Firm/
Manufacturer
Philips North America Llc
222 Jacobs St
Cambridge MA 02141-2289
For Additional Information ContactPhilips Customer Care Solutions Center
800-722-9377
Manufacturer Reason
for Recall
Potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the patient, and an incorrect patient ID software issue that may lead to misdiagnosis.
FDA Determined
Cause 2
Under Investigation by firm
ActionPhilips notified consignees on about 11/16/2024, via certified mail. Consignees were instructed to identify any affected units on hand, identify if those units contain the affected software version, complete and return the response form, and circulate the Urgent Field Safety Notice Letter to all users of the device. To protect the user and patient from the issue of Plan Box not Updated, remain vigilant and confirm table position during procedures. Press any of the Emergency STOP buttons to stop unintended patient table motion and follow safety instructions in Section 6 of the IFU: Working with Specialized Exams. To avoid the issue of Wrong Patient ID with Anonymous patient, the operator should verify the patient information loaded into the patient data field is correct before scanning and rename the Anonymous record with the patient s name at the time of the scan.
Quantity in Commerce301 units
DistributionWorldwide - US Nationwide distribution in the states of AR, AZ, CA, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, Israel, Italy, Japan, Jordan, Macao, Malaysia, Netherlands, New Zealand, Norway, Palestine, Panama, Poland, Portugal, Singapore, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.