| | Class 2 Device Recall OEC Elite |  |
| Date Initiated by Firm | November 21, 2024 |
| Date Posted | December 18, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0704-2025 |
| Recall Event ID |
95846 |
| 510(K)Number | K170752 K171565 K172550 K192819 |
| Product Classification |
Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
|
| Product | OEC Elite REF
Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures. |
| Code Information |
Product Name: OEC Elite
GTIN code: 00001234567895
Lot/Serial Numbers / UDI codes:
FAXXXE00254 (01)00840682121729(11)190400(21)FAXXXE00254
FBHXTX00837 (01)00840682121729(11)230300(21)FBHXTX00837
FBHXTX00912 (01)00840682121729(11)230500(21)FBHXTX00912
FBHXTX01278 (01)00840682121729(11)240800(21)FBHXTX01278
FBXXXE00731 (01)00840682121729(11)190500(21)FBXXXE00731
FBMHTX00470 (01)00840682121729(11)240900(21)FBMHTX00470
F2XXTX00801 (01)00840682121729(11)240800(21)F2XXTX00801
FBHXXX00171 (01)00840682121729(11)180600(21)FBHXXX00171
FBXXXE00715 (01)00840682121729(11)190500(21)FBXXXE00715
FBMHTX00444 (01)00840682121729(11)240600(21)FBMHTX00444
FBXXTE02974 (01)00840682121729(11)240800(21)FBXXTE02974
FBMHTX00466 (01)00840682121729(11)240900(21)FBMHTX00466
FAXXTE00672 (01)00840682121729(11)221200(21)FAXXTE00672
F9XXTX00276 (01)00840682121729(11)230900(21)F9XXTX00276
FAXXTE01160 (01)00840682121729(11)240800(21)FAXXTE01160
FBMHTX00454 (01)00840682121729(11)240800(21)FBMHTX00454
FBMHTX00455 (01)00840682121729(11)240800(21)FBMHTX00455
FBMHTX00465 (01)00840682121729(11)240900(21)FBMHTX00465
FBXXTE02975 (01)00840682121729(11)240800(21)FBXXTE02975
FBXXTE02976 (01)00840682121729(11)240800(21)FBXXTE02976
FBXXXE00312 (01)00840682121729(11)180900(21)FBXXXE00312
FAXXTE00664 (01)00840682121729(11)221100(21)FAXXTE00664
FBMHTX00450 (01)00840682121729(11)240700(21)FBMHTX00450
FAXXTE01154 (01)00840682121729(11)240700(21)FAXXTE01154
FBXXTE02994 (01)00840682121729(11)240800(21)FBXXTE02994
FAXXTE00544 (01)00840682121729(11)220600(21)FAXXTE00544
FBXXTE01055 (01)00840682121729(11)211200(21)FBXXTE01055
FBXXTE02966 (01)00840682121729(11)240800(21)FBXXTE02966
FBMHTX00467 (01)00840682121729(11)240900(21)FBMHTX00467
FAXXTE00339 (01)00840682121729(11)210600(21)FAXXTE00339
FSXXTX00282 (01)00840682121729(11)220500(21)FSXXTX00282
FAXXXE00231 (01)00840682121729(11)190300(21)FAXXXE00231
FBXXXE00653 (01)00840682121729(11)190400(21)FBXXXE00653
F2XXTX00381 (01)00840682121729(11)220100(21)F2XXTX00381
FBXXXE00008 (01)00840682121729(11)171000(21)FBXXXE00008
FBMHTX00468 (01)00840682121729(11)240900(21)FBMHTX00468
FBXXTP00008 (01)00840682121729(11)240700(21)FBXXTP00008
FSXXTX00501 (01)00840682121729(11)240400(21)FSXXTX00501
F2XXTX00802 (01)00840682121729(11)240800(21)F2XXTX00802
FAXXTE01157 (01)00840682121729(11)240700(21)FAXXTE01157
FBMHTX00447 (01)00840682121729(11)240600(21)FBMHTX00447
FBMHDX00049 (01)00195278096944(11)220100(21)FBMHDX00049
FBXXTE02312 (01)00840682121729(11)230900(21)FBXXTE02312
F9XXTX00301 (01)00840682121729(11)240100(21)F9XXTX00301
FBXXTE02965 (01)00840682121729(11)240800(21)FBXXTE02965
FBXXXE00654 (01)00840682121729(11)190400(21)FBXXXE00654
FBMHTX00452 (01)00840682121729(11)240800(21)FBMHTX00452
FBXXTE00561 (01)00840682121729(11)201100(21)FBXXTE00561
FBXXTE02963 (01)00840682121729(11)240800(21)FBXXTE02963
FBXXTE02109 (01)00840682121729(11)230600(21)FBXXTE02109
FBXXTP00006 (01)00840682121729(11)240700(21)FBXXTP00006
FBXXTP00007 (01)00840682121729(11)240700(21)FBXXTP00007
F2XXTX00021 (01)00840682121729(11)191100(21)F2XXTX00021
FBXXTE02991 (01)00840682121729(11)240800(21)FBXXTE02991
FBXXTE02992 (01)00840682121729(11)240800(21)FBXXTE02992
F2XXTX00799 (01)00840682121729(11)240800(21)F2XXTX00799
FSXXTX00514 (01)00840682121729(11)240600(21)FSXXTX00514
FBXXTE01754 (01)00840682121729(11)230100(21)FBXXTE01754
FBXXTE02914 (01)00840682121729(11)240600(21)FBXXTE02914
FBXXTE00515 (01)00840682121729(11)201000(21)FBXXTE00515
FAXXXE00049 (01)00840682121729(11)180200(21)FAXXXE00049
FBXXXE00272 (01)00840682121729(11)180900(21)FBXXXE00272
FSXXXX00004 (01)00840682121729(11)190400(21)FSXXXX00004
FBMHTX00433 (01)00840682121729(11)240400(21)FBMHTX00433
FBMHTX00446 (01)00840682121729(11)240600(21)FBMHTX00446
FBXXTE01644 (01)00840682121729(11)221200(21)FBXXTE01644
FBMHTX00464 (01)00840682121729(11)240900(21)FBMHTX00464
FBXXTE02978 (01)00840682121729(11)240800(21)FBXXTE02978
F2XXTX00798 (01)00840682121729(11)240800(21)F2XXTX00798
FBXXTE02979 (01)00840682121729(11)240800(21)FBXXTE02979
FBXXTE01932 (01)00840682121729(11)230300(21)FBXXTE01932
FBXXTE02977 (01)00840682121729(11)240800(21)FBXXTE02977
FBXXTE02993 (01)00840682121729(11)240800(21)FBXXTE02993
FBHXTX00821 (01)00840682121729(11)230200(21)FBHXTX00821
F2XXTX00800 (01)00840682121729(11)240800(21)F2XXTX00800
FSXXTX00284 (01)00840682121729(11)220500(21)FSXXTX00284
FBXXTE00575 (01)00840682121729(11)201200(21)FBXXTE00575
FBMHTX00469 (01)00840682121729(11)240900(21)FBMHTX00469
FBXXTE02323 (01)00840682121729(11)230900(21)FBXXTE02323
FSXXTX00422 (01)00840682121729(11)230300(21)FSXXTX00422
FAHXTX00538 (01)00840682121729(11)240600(21)FAHXTX00538
FBXXTE02996 (01)00840682121729(11)240800(21)FBXXTE02996
FSXXTX00418 (01)00840682121729(11)230300(21)FSXXTX00418
FBXXTE01863 (01)00840682121729(11)230300(21)FBXXTE01863
FBMHTX00453 (01)00840682121729(11)240800(21)FBMHTX00453
FBMHTX00329 (01)00840682121729(11)230200(21)FBMHTX00329
FAXXTE01158 (01)00840682121729(11)240800(21)FAXXTE01158
FBXXTE02995 (01)00840682121729(11)240800(21)FBXXTE02995
FAXXTE01159 (01)00840682121729(11)240800(21)FAXXTE01159
F9XXXX00064 (01)00840682121729(11)191200(21)F9XXXX00064
FBXXTE02980 (01)00840682121729(11)240800(21)FBXXTE02980
FBXXTE02981 (01)00840682121729(11)240800(21)FBXXTE02981
FBXXTE02982 (01)00840682121729(11)240800(21)FBXXTE02982
FBXXTE02983 (01)00840682121729(11)240800(21)FBXXTE02983
FBXXTE02984 (01)00840682121729(11)240800(21)FBXXTE02984
FBXXTE02986 (01)00840682121729(11)240800(21)FBXXTE02986
FBXXTE02987 (01)00840682121729(11)240800(21)FBXXTE02987
FBMHTX00456 (01)00840682121729(11)240800(21)FBMHTX00456
FBMHTX00449 (01)00840682121729(11)240700(21)FBMHTX00449
F9XXTX00329 (01)00840682121729(11)240800(21)F9XXTX00329
|
Recalling Firm/ Manufacturer |
GE OEC Medical Systems, Inc 384 N Wright Brothers Dr Salt Lake City UT 84116-2862
|
| For Additional Information Contact | GE HealthCare Service 800-437-1171 |
Manufacturer Reason for Recall | Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability. |
FDA Determined Cause 2 | Process control |
| Action | On 11/21/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that GE HealthCare has become aware that recently manufactured OEC Elite and OEC 3D mobile C-arms, as well as recently replaced X-ray tubes on certain OEC Elite, OEC 3D and OEC 9900/OEC 9800 mobile C-arms potentially have insufficient sealing that can result in an oil leak. This issue could result in loss of fluoroscopic imaging capability.
Customers are instructed to:
1.Continue to use their OEC Elite, OEC 3D, OEC 9900 and/or OEC9800 mobile C-arm.
2.If oil is found leaking from the X-ray tube on their OEC Elite, OEC 3D, OEC9900 and/or OEC 9800 mobile C-arm, contact the local GE HealthCareService representative.
3.Ensure all potential users in their facility are made aware of this safety notification and the recommended actions.
4.Please complete and return the attached acknowledgement form torecall.15150@gehealthcare.com.
GE HealthCare will correct all affected products at no cost to you.
A GE HealthCare representative will contact you to plan and schedule for the inspection and correction if needed.
for any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity in Commerce | 100 systems |
| Distribution | U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OXO 510(K)s with Product Code = OXO
|
|
|
|