• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmNovember 21, 2024
Date PostedDecember 30, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0777-2025
Recall Event ID 95852
Product Classification Cardiac catheterization kit - Product Code OES
ProductMedline custom medical procedure kits labeled as: 1) CATH LAB PACK-LF, REF DYNJ0382559M; 2) CRANI, REF DYNJ56001B; 3) GENDER AFFIRMING SURGERY, REF DYNJ66252; 4) GENDER AFFIRMING SURGERY, REF DYNJ66252A; 5) MAJOR NEURO PACK-LF, REF DYNJ0578916X.
Code Information 1) REF DYNJ0382559M, UDI/DI 10889942063994 (EA) 10889942063995 (CS), Lot Numbers: 19TBC792; 2) REF DYNJ56001B, UDI/DI 10889942798278 (EA) 40889942798279 (CS), Lot Numbers: 19BDC737, 19FDA310, 19FDB230, 19JDA441; 3) REF DYNJ66252, UDI/DI 10193489324785 (EA) 40193489324786 (CS), Lot numbers: 20DBK065, 20FBN251, 20HBL176, 20HBM692, 21CBP776; 4) REF DYNJ66252A, UDI/DI 10193489887853 (EA) 40193489887854 (CS), Lot numbers: 21IBN151, 21KBO915, 21LBJ105, 21VBB714; 5) REF DYNJ0578916X, UDI/DI 10193489288384 (EA) 40193489288385 (CS), Lot numbers: 22OBL262,20CKA219, 20CKA875, 20EKB076, 20GKA543, 20GKB272, 20IKB497, 20JKB716, 21AKA500, 21BKA195, 21FMD808, 21HMD657, 21IMB172, 21JMH627, 21KMH877, 21LMC161, 22AMA533, 22DMC612, 22DMI233, 22GMB685, 22IMA041, 23AMD730, 23DMA094, 23FMF542, 23IME618, 24AMA230.
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
866-359-1704
Manufacturer Reason
for Recall
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionMedline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.
Quantity in Commerce1236 units
DistributionWorldwide distribution - US Nationwide and the countries of UAE, and Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-