| | Class 2 Device Recall Medline |  |
| Date Initiated by Firm | November 21, 2024 |
| Date Posted | December 30, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0777-2025 |
| Recall Event ID |
95852 |
| Product Classification |
Cardiac catheterization kit - Product Code OES
|
| Product | Medline custom medical procedure kits labeled as:
1) CATH LAB PACK-LF, REF DYNJ0382559M;
2) CRANI, REF DYNJ56001B;
3) GENDER AFFIRMING SURGERY, REF DYNJ66252;
4) GENDER AFFIRMING SURGERY, REF DYNJ66252A;
5) MAJOR NEURO PACK-LF, REF DYNJ0578916X. |
| Code Information |
1) REF DYNJ0382559M, UDI/DI 10889942063994 (EA) 10889942063995 (CS), Lot Numbers: 19TBC792;
2) REF DYNJ56001B, UDI/DI 10889942798278 (EA) 40889942798279 (CS), Lot Numbers: 19BDC737, 19FDA310, 19FDB230, 19JDA441;
3) REF DYNJ66252, UDI/DI 10193489324785 (EA) 40193489324786 (CS), Lot numbers: 20DBK065, 20FBN251, 20HBL176, 20HBM692, 21CBP776;
4) REF DYNJ66252A, UDI/DI 10193489887853 (EA) 40193489887854 (CS), Lot numbers: 21IBN151, 21KBO915, 21LBJ105, 21VBB714;
5) REF DYNJ0578916X, UDI/DI 10193489288384 (EA) 40193489288385 (CS), Lot numbers: 22OBL262,20CKA219, 20CKA875, 20EKB076, 20GKA543, 20GKB272, 20IKB497, 20JKB716, 21AKA500, 21BKA195, 21FMD808, 21HMD657, 21IMB172, 21JMH627, 21KMH877, 21LMC161, 22AMA533, 22DMC612, 22DMI233, 22GMB685, 22IMA041, 23AMD730, 23DMA094, 23FMF542, 23IME618, 24AMA230.
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Haley Barclay 866-359-1704 |
Manufacturer Reason for Recall | Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Medline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit. |
| Quantity in Commerce | 1236 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of UAE, and Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|