| | Class 2 Device Recall GE Healthcare |  |
| Date Initiated by Firm | February 19, 2025 |
| Date Posted | March 28, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1486-2025 |
| Recall Event ID |
96375 |
| 510(K)Number | K042525 K063628 |
| Product Classification |
System, image processing, radiological - Product Code LLZ
|
| Product | GE Healthcare Centricity Cardiology CA1000 (CA1000), Model Numbers:
1) 2033901-001;
2) 2038437-001;
3) 2038437-009;
4) 2038437-014;
5) 2038437-015 ;
6) 2038437-0XX;
7) 2109571-007;
8) 2109571-010;
9) 2109571-011;
10) 2109571-012;
11) MANLEGACY111;
12) MANLEGACY112;
13) MANLEGACY47;
14) P00011TS. |
| Code Information |
Model/Catalog/Code # 2033901-001, Installed Product ID 14335625-Centricity CA1000-19675863, UDI UDI Not Required, System ID HCIT4082079CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 11558-1-Centricity CA1000-00099550, UDI UDI Not Required, System ID 520855CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103987-1-Centricity CA1000-00102841, UDI UDI Not Required, System ID 928537CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 9990-1-Centricity CA1000-00101135, UDI UDI not required, System ID 661327CWS01;
Model/Catalog/Code # 2038437-001, Installed Product ID 10318-1-Centricity CA1000-00105877, UDI UDI not required, System ID 818397CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10994-1-Centricity CA1000-00102387, UDI UDI Not required, System ID 909464CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 102131-1-Centricity CA1000-00105886, UDI UDI not required, System ID 818409CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4161-1-Centricity CA1000-00105894, UDI UDI Not Required, System ID 818502CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103743-1-Centricity CA1000-00099665, UDI UDI Not Required, System ID 559582CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4408-2-Centricity CA1000-00101725, UDI UDI not required, System ID 714843CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 4532-4-Centricity CA1000-00107356, UDI UDI Not Required, System ID 310673CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4009-2-Centricity CA1000-00100791, UDI UDI not required, System ID 619260CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1354-8-Centricity CA1000-00102389, UDI UDI not required, System ID 909558CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 9927-2-Centricity CA1000-00099762, UDI UDI not required, System ID 562491CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11299-9-Centricity CA1000-00109567, UDI UDI Not Required, System ID SVMCA6;
Model/Catalog/Code # 2038437-001, Installed Product ID 4446-2-Centricity CA1000-00107805, UDI (01)00840682125260(10)20P14H2038437016, System ID 323726CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103447-1-Centricity CA1000-22271916, UDI (01)00840682125260(10)20P14J2038437016, System ID CALIFORNIA22271916;
Model/Catalog/Code # 2038437-001, Installed Product ID 100256-3-Centricity CA1000-00101156, UDI (01)00840682125260(10)20P14J2038437016, System ID 661948CA1RSW;
Model/Catalog/Code # 2038437-001, Installed Product ID 11169-1-Centricity CA1000-00102586, UDI UDI not required, System ID 916225CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1223-17-Centricity CA1000-00106365, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 858554CA1SW14;
Model/Catalog/Code # 2038437-001, Installed Product ID 1845-2-Centricity CA1000-00100829, UDI UDI not required, System ID 619691CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 2195-12-Centricity CA1000-16385905, UDI UDI Not Required, System ID UCLAMC16385905;
Model/Catalog/Code # 2038437-001, Installed Product ID 9985-1-Centricity CA1000-00106974, UDI UDI not required, System ID 303695CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103533-5-Centricity CA1000-02299486, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID BOULDERCO02299486;
Model/Catalog/Code # 2038437-001, Installed Product ID 10617-1-Centricity CA1000-00107034, UDI UDI not required, System ID 303839CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 100385-1-Centricity CA1000-00106873, UDI UDI not required, System ID 303320CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10482-1-Centricity CA1000-00103250, UDI UDI not required, System ID 970764CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10841-1-Centricity CA1000-00106998, UDI UDI not required, System ID 303788CVIT5;
Model/Catalog/Code # 2038437-001, Installed Product ID 9900-1-Centricity CA1000-00106902, UDI UDI not required, System ID 303451CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 2386-2-Centricity CA1000-00116051, UDI UDI not required, System ID 203276CA1000RV1;
Model/Catalog/Code # 2038437-001, Installed Product ID 101581-1-Centricity CA1000-00095981, UDI UDI not required, System ID 202745CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4698-3-Centricity CA1000-00108747, UDI UDI not required, System ID MCTHCA10001;
Model/Catalog/Code # 2038437-001, Installed Product ID 101194-2-Centricity CA1000-00103078, UDI UDI not required, System ID 941792HCA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10298-1-Centricity CA1000-00105482, UDI UDI not required, System ID 813681CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10298-1-Centricity CA1000-00105491, UDI UDI not required, System ID 813681CA1SW19;
Model/Catalog/Code # 2038437-001, Installed Product ID 11209-1-Centricity CA1000-00095905, UDI UDI not required, System ID 0004053422A;
Model/Catalog/Code # 2038437-001, Installed Product ID 10218-4-Centricity CA1000-00108483, UDI UDI not required, System ID FWBCA1005;
Model/Catalog/Code # 2038437-001, Installed Product ID 10218-2-Centricity CA1000-00095912, UDI UDI not required, System ID FWBCA14;
Model/Catalog/Code # 2038437-001, Installed Product ID 103638-1-Centricity CA1000-00107955, UDI UDI not required, System ID 352331CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10849-3-Centricity CA1000-00107230, UDI UDI Not Required, System ID 305364CA2100;
Model/Catalog/Code # 2038437-001, Installed Product ID 103203-1-Centricity CA1000-22144957, UDI (01)00840682125260(10)20P14J2038437016, System ID BIHT22144957;
Model/Catalog/Code # 2038437-001, Installed Product ID 10458-5-Centricity CA1000-00102189, UDI UDI not required, System ID 904399HCACA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11208-2-Centricity CA1000-00103172, UDI UDI not required, System ID 954974CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 101427-1-Centricity CA1000-00104669, UDI UDI not required, System ID 727848HCA1SW7;
Model/Catalog/Code # 2038437-001, Installed Product ID 10264-1-Centricity CA1000-00095914, UDI UDI not required, System ID 850769HCA1SW11;
Model/Catalog/Code # 2038437-001, Installed Product ID 101180-1-Centricity CA1000-00104672, UDI UDI not required, System ID 727863CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103577-2-Centricity CA1000-00104941, UDI UDI not required, System ID 772581CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 104179-1-Centricity CA1000-00104625, UDI UDI Not required, System ID 727521CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103788-1-Centricity CA1000-00104637, UDI UDI not required, System ID 727521OPCA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 11639-1-Centricity CA1000-00104600, UDI UDI not required, System ID 727384CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10475-1-Centricity CA1000-00097981, UDI UDI not required, System ID 407321SWCA1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11041-1-Centricity CA1000-00103047, UDI UDI Not Required, System ID 941342CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11743-1-Centricity CA1000-00095913, UDI (01)00840682125260(10)20P14H2038437016, System ID 0004275549A;
Model/Catalog/Code # 2038437-001, Installed Product ID 11743-1-Centricity CA1000-01973685, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 850325HCA1SW5;
Model/Catalog/Code # 2038437-001, Installed Product ID 13367-1-Centricity CA1000-02041150, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID HCAE02041150;
Model/Catalog/Code # 2038437-001, Installed Product ID 100620-13-Centricity CA1000-00163680, UDI UDI not required, System ID 813885HCA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1657-7-Centricity CA1000-00105550, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID 813971CA10003;
Model/Catalog/Code # 2038437-001, Installed Product ID 10292-2-Centricity CA1000-00104678, UDI UDI Not Required, System ID 727942CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 100115-1-Centricity CA1000-00101433, UDI UDI Not Required, System ID 706651HCACA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 3901-2-Centricity CA1000-00101421, UDI UDI not required, System ID 706596CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10147-2-Centricity CA1000-22155331, UDI (01)00840682125260(10)20P14J2038437016, System ID CRH522155331;
Model/Catalog/Code # 2038437-001, Installed Product ID 9847-2-Centricity CA1000-00108204, UDI UDI not required, System ID 404501CA1SW1A;
Model/Catalog/Code # 2038437-001, Installed Product ID 10209-1-Centricity CA1000-00098680, UDI UDI Not Required, System ID 478274CA1SW100;
Model/Catalog/Code # 2038437-001, Installed Product ID 11047-6-Centricity CA1000-00101185, UDI UDI not required, System ID 678957CA1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11047-6-Centricity CA1000-00101186, UDI UDI not required, System ID 678957CA2;
Model/Catalog/Code # 2038437-001, Installed Product ID 3929-2-Centricity CA1000-00105388, UDI UDI not required, System ID 808485CA1SW2;
Model/Catalog/Code # 2038437-001, Installed Product ID 1458-6-Centricity CA1000-00107439, UDI UDI Not Required, System ID 312RUSH1CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 1566-4-Centricity CA1000-00102050, UDI UDI not required, System ID 877453CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1566-4-Centricity CA1000-00102051, UDI UDI not required, System ID 877453CA1SW2;
Model/Catalog/Code # 2038437-001, Installed Product ID 101349-2-Centricity CA1000-00109702, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID UICCA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 101349-2-Centricity CA1000-22176245, UDI (01)00840682125260(10)20P14J2038437016, System ID UIMC22176245;
Model/Catalog/Code # 2038437-001, Installed Product ID 101454-1-Centricity CA1000-00101519, UDI UDI Not Required, System ID 708202CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10691-2-Centricity CA1000-00096590, UDI UDI Not required, System ID 217258CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11367-1-Centricity CA1000-00101569, UDI UDI not required, System ID 708681CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10251-3-Centricity CA1000-00100783, UDI UDI not required, System ID 618241CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 100086-1-Centricity CA1000-00107335, UDI UDI not required, System ID 309CA1000SW4;
Model/Catalog/Code # 2038437-001, Installed Product ID 1250-1-Centricity CA1000-00101545, UDI UDI Not Required, System ID 708361CA1000SW;
Model/Catalog/Code # 2038437-001, Installed Product ID 1237-1-Centricity CA1000-00103189, UDI UDI Not Required, System ID 956544CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 3832-1-Centricity CA1000-00105463, UDI UDI not required, System ID 812858CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 8637-1-Centricity CA1000-00107663, UDI UDI not required, System ID 317621CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 3864-3-Centricity CA1000-00105445, UDI (01)00840682125260(10)20P14J2038437016, System ID 812482CA1SW;
Model/Catalog/Code # 2038437-001, Installed Product ID 102796-1-Centricity CA1000-00104871, UDI UDI not required, System ID 765671CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10943-1-Centricity CA1000-00107645, UDI (01)00840682125260(10)20P14F2038437016, System ID 316962HCACA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103613-1-Centricity CA1000-00107629, UDI UDI not required, System ID 316858HCACA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103897-1-Centricity CA1000-00097067, UDI (01)00840682125260(10)20P14J2038437016, System ID 270465CA1SW;
Model/Catalog/Code # 2038437-001, Installed Product ID 102600-3-Centricity CA1000-00100351, UDI UDI not required, System ID 606439CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 100435-3-Centricity CA1000-00100367, UDI UDI Not Required, System ID 606759CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 10380-2-Centricity CA1000-00100361, UDI (01)00840682125260(10)20P14J2038437016, System ID 606679CA1K1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10637-1-Centricity CA1000-00107698, UDI UDI not required, System ID 318473CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4643-3-Centricity CA1000-00107915, UDI UDI Not Required, System ID 337289CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10845-1-Centricity CA1000-00098992, UDI UDI Not Required, System ID 504380CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 102718-1-Centricity CA1000-00107677, UDI UDI not required, System ID 318254CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 101572-1-Centricity CA1000-00107675, UDI UDI Not Required, System ID 318221CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 3104-3-Centricity CA1000-00100765, UDI UDI not required, System ID 617754CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103916-2-Centricity CA1000-00099172, UDI UDI Not Required, System ID 508335CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10488-1-Centricity CA1000-00099175, UDI UDI Not required, System ID 508383CA1;
Model/Catalog/Code # 2038437-001, Installed Product ID 102719-1-Centricity CA1000-00098281, UDI UDI not required, System ID 413748CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 3720-3-Centricity CA1000-00098647, UDI UDI not required, System ID 443481CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10996-4-Centricity CA1000-00096899, UDI (01)00840682125260(10)20P14J2038437016, System ID 248360CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 13399-8-Centricity CA1000-17322532, UDI (01)00840682125260(10)2.0SP14A2038437016C, System ID VANA17322532;
Model/Catalog/Code # 2038437-001, Installed Product ID 10516-1-Centricity CA1000-22189501, UDI (01)00840682125260(10)20P14J2038437016, System ID MMC1922189501;
Model/Catalog/Code # 2038437-001, Installed Product ID 13370-4-Centricity CA1000-20698703, UDI (01)00840682125260(10)20P14J2038437016, System ID ABBT20698703;
Model/Catalog/Code # 2038437-001, Installed Product ID 100796-1-Centricity CA1000-00100559, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 612638CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10159-3-Centricity CA1000-00104275, UDI UDI Not Required, System ID W128CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1651-1-Centricity CA1000-00099865, UDI UDI not required, System ID 573761CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 101536-1-Centricity CA1000-00105734, UDI UDI Not Required, System ID 816861CA1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1651-14-Centricity CA1000-00104265, UDI UDI not required, System ID W113CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11281-1-Centricity CA1000-00104246, UDI UDI not required, System ID W022CA001;
Model/Catalog/Code # 2038437-001, Installed Product ID 11445-3-Centricity CA1000-00109178, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID S011PCA1REV;
Model/Catalog/Code # 2038437-001, Installed Product ID 11445-7-Centricity CA1000-20025014, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID SHSLMHC20025014;
Model/Catalog/Code # 2038437-001, Installed Product ID 11445-2-Centricity CA1000-00104209, UDI UDI Not Required, System ID W001CA1SW10;
Model/Catalog/Code # 2038437-001, Installed Product ID 3675-3-Centricity CA1000-00099941, UDI UDI Not Required, System ID 601364CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11243-3-Centricity CA1000-00099982, UDI UDI Not Required, System ID 601883CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11243-3-Centricity CA1000-00099983, UDI UDI Not Required, System ID 601883CA1SW2;
Model/Catalog/Code # 2038437-001, Installed Product ID 10933-2-Centricity CA1000-00105927, UDI UDI not required, System ID 828262CA1000SW;
Model/Catalog/Code # 2038437-001, Installed Product ID 3725-1-Centricity CA1000-22247761, UDI (01)00840682125260(10)20P14J2038437016, System ID CTV22247761;
Model/Catalog/Code # 2038437-001, Installed Product ID 10196-1-Centricity CA1000-00100125, UDI UDI not required, System ID 603669CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4696-2-Centricity CA1000-00100128, UDI UDI Not Required, System ID 603882CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10610-1-Centricity CA1000-00100082, UDI UDI not required, System ID 603433CA1SW4;
Model/Catalog/Code # 2038437-001, Installed Product ID 3820-1-Centricity CA1000-00104711, UDI UDI not required, System ID 732294CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 104070-1-Centricity CA1000-00103474, UDI UDI not required, System ID 973956CA10001S;
Model/Catalog/Code # 2038437-001, Installed Product ID 103432-1-Centricity CA1000-00099060, UDI (01)00840682125260(10)20P14J2038437016, System ID 505848CASFT;
Model/Catalog/Code # 2038437-001, Installed Product ID 3865-2-Centricity CA1000-00099049, UDI UDI Not Required, System ID 505522CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10513-2-Centricity CA1000-00101237, UDI UDI not required, System ID 702255CA10001;
Model/Catalog/Code # 2038437-001, Installed Product ID 11239-1-Centricity CA1000-00104995, UDI UDI not required, System ID 775982CA2100;
Model/Catalog/Code # 2038437-001, Installed Product ID 4029-47-Centricity CA1000-00099399, UDI UDI not required, System ID 516562CASW7;
Model/Catalog/Code # 2038437-001, Installed Product ID 10732-1-Centricity CA1000-00100957, UDI UDI not required, System ID 631968CARR1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1196-58-Centricity CA1000-00104474, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 718920CA1SW10;
Model/Catalog/Code # 2038437-001, Installed Product ID 104038-1-Centricity CA1000-00101887, UDI UDI Not Required, System ID 718246CA1HW;
Model/Catalog/Code # 2038437-001, Installed Product ID 102250-1-Centricity CA1000-00104390, UDI UDI not required, System ID 718270CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4029-9-Centricity CA1000-00100970, UDI UDI not required, System ID 631968SSCA1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4374-3-Centricity CA1000-00106119, UDI UDI Not required, System ID 845343CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4029-8-Centricity CA1000-00099376, UDI UDI not required, System ID 516562AI10002;
Model/Catalog/Code # 2038437-001, Installed Product ID 4029-8-Centricity CA1000-00104421, UDI UDI not required, System ID 718470CASW10;
Model/Catalog/Code # 2038437-001, Installed Product ID 4029-51-Centricity CA1000-14194765, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID NSLI14194765;
Model/Catalog/Code # 2038437-001, Installed Product ID 102193-2-Centricity CA1000-00099495, UDI UDI not required, System ID 518382CA10001;
Model/Catalog/Code # 2038437-001, Installed Product ID 102193-2-Centricity CA1000-22157802, UDI (01)00840682125260(10)20P14J2038437016, System ID ELLISHOSP22157802;
Model/Catalog/Code # 2038437-001, Installed Product ID 102238-1-Centricity CA1000-00106055, UDI UDI Not Required, System ID 843668CA1SW;
Model/Catalog/Code # 2038437-001, Installed Product ID 1207-8-Centricity CA1000-00105459, UDI (01)00840682125260(10)20P14J2038437016, System ID 8125551CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10240-2-Centricity CA1000-00104793, UDI (01)00840682125260(10)20P14J2038437016, System ID 740454CA1SW3;
Model/Catalog/Code # 2038437-001, Installed Product ID 102169-1-Centricity CA1000-00099885, UDI UDI not required, System ID 580421VVCA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 103361-1-Centricity CA1000-00097922, UDI UDI not required, System ID 405ERMCCA1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1475-4-Centricity CA1000-00097935, UDI UDI not required, System ID 405STAC1000SW6;
Model/Catalog/Code # 2038437-001, Installed Product ID 1475-1-Centricity CA1000-00097938, UDI UDI Not Required, System ID 405STACA041;
Model/Catalog/Code # 2038437-001, Installed Product ID 10743-4-Centricity CA1000-00551324, UDI UDI not required, System ID 405272CA1000RV1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1467-3-Centricity CA1000-00102747, UDI UDI not required, System ID 918744CA1SW5;
Model/Catalog/Code # 2038437-001, Installed Product ID 103440-1-Centricity CA1000-00099656, UDI UDI Not Required, System ID 54174CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 1999-7-Centricity CA1000-00098806, UDI UDI not required, System ID 484488CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103653-1-Centricity CA1000-22251392, UDI (01)00840682125260(10)20P14J2038437016, System ID ASSOCIATIO22251392;
Model/Catalog/Code # 2038437-001, Installed Product ID 1999-19-Centricity CA1000-00101907, UDI UDI not required, System ID 861MCACR;
Model/Catalog/Code # 2038437-001, Installed Product ID 100254-1-Centricity CA1000-17594352, UDI (01)00840682125260(10)20P14J2038437016, System ID STMARYME17594352;
Model/Catalog/Code # 2038437-001, Installed Product ID 1581-4-Centricity CA1000-00101780, UDI UDI not required, System ID 717270CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 101079-1-Centricity CA1000-00099788, UDI (01)00840682125260(10)20P14J2038437016, System ID 570522CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 3719-1-Centricity CA1000-00104553, UDI UDI not required, System ID 724728HVCA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 4299-2-Centricity CA1000-00096454, UDI (01)00840682125260(10)20P14J2038437016, System ID 215335CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11673-4-Centricity CA1000-00096463, UDI UDI not required, System ID 215453CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 102401-1-Centricity CA1000-00101815, UDI UDI not required, System ID 717765CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 101583-1-Centricity CA1000-00099808, UDI UDI not required, System ID 570824CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10863-2-Centricity CA1000-00106092, UDI UDI Not Required, System ID 843797CHL1;
Model/Catalog/Code # 2038437-001, Installed Product ID 101518-1-Centricity CA1000-00095910, UDI (01)00840682125260(10)20P14J2038437016, System ID 0004190048A;
Model/Catalog/Code # 2038437-001, Installed Product ID 10992-2-Centricity CA1000-00106056, UDI UDI Not Required, System ID 843674CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 101512-3-Centricity CA1000-00106083, UDI UDI not required, System ID 843789CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10603-1-Centricity CA1000-00105154, UDI UDI Not Required, System ID 803329CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11305-1-Centricity CA1000-00106029, UDI UDI not required, System ID 843552CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11305-1-Centricity CA1000-00106030, UDI UDI not required, System ID 843552HCA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10109-1-Centricity CA1000-00106302, UDI UDI Not Required, System ID 850764HCA1SW6;
Model/Catalog/Code # 2038437-001, Installed Product ID 3949-4-Centricity CA1000-00101994, UDI UDI not required, System ID 865541CA1K1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10590-1-Centricity CA1000-00102019, UDI UDI not required, System ID 865693CA139;
Model/Catalog/Code # 2038437-001, Installed Product ID 3949-6-Centricity CA1000-00098555, UDI UDI not required, System ID 423586CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 100248-1-Centricity CA1000-00102862, UDI UDI not required, System ID 928758CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11011-3-Centricity CA1000-00101986, UDI UDI Not required, System ID 865481CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 11011-3-Centricity CA1000-00101988, UDI UDI not required, System ID 865481CA1SW3;
Model/Catalog/Code # 2038437-001, Installed Product ID 10433-3-Centricity CA1000-00105822, UDI UDI not required, System ID 817472CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10914-2-Centricity CA1000-00097121, UDI UDI not required, System ID 281325CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 9859-80-Centricity CA1000-20100489, UDI (01)00840682125260(10)20P14M2038437016, System ID HH1020100489;
Model/Catalog/Code # 2038437-001, Installed Product ID 100085-4-Centricity CA1000-00103005, UDI UDI not required, System ID 940384CA1SW;
Model/Catalog/Code # 2038437-001, Installed Product ID 10439-1-Centricity CA1000-00096398, UDI UDI not required, System ID 214420HCA1SW5;
Model/Catalog/Code # 2038437-001, Installed Product ID 10140-2-Centricity CA1000-00103680, UDI UDI not required, System ID BACATH50;
Model/Catalog/Code # 2038437-001, Installed Product ID 10442-1-Centricity CA1000-00103288, UDI UDI not required, System ID 972556CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 3677-2-Centricity CA1000-00104366, UDI (01)00840682125260(10)20P14J2038437016, System ID WRCAREVIEWSW;
Model/Catalog/Code # 2038437-001, Installed Product ID 10168-1-Centricity CA1000-00103895, UDI UDI not required, System ID DRCA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 104089-1-Centricity CA1000-00105850, UDI UDI not required, System ID 817639HCA1SW;
Model/Catalog/Code # 2038437-001, Installed Product ID 11227-1-Centricity CA1000-00105785, UDI UDI not required, System ID 817347CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10948-1-Centricity CA1000-00097149, UDI UDI Not required, System ID 281588CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103744-1-Centricity CA1000-00097146, UDI UDI not required, System ID 281579CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103656-1-Centricity CA1000-00105352, UDI (01)00840682125260(10)20P14J2038437016, System ID 806687CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 2163-3-Centricity CA1000-00103277, UDI UDI not required, System ID 972548HCACA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 9898-3-Centricity CA1000-00105744, UDI UDI Not Required, System ID 817225CA1SW;
Model/Catalog/Code # 2038437-001, Installed Product ID 3953-2-Centricity CA1000-00109631, UDI (01)00840682125260(10)20P14J2038437016, System ID THHBPEPCARR2;
Model/Catalog/Code # 2038437-001, Installed Product ID 10441-1-Centricity CA1000-00095903, UDI UDI not required, System ID 0004020010A;
Model/Catalog/Code # 2038437-001, Installed Product ID 4575-1-Centricity CA1000-00102111, UDI UDI not required, System ID 903885CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 3871-5-Centricity CA1000-00102105, UDI UDI Not Required, System ID 903595CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 105534-1-Centricity CA1000-00108074, UDI UDI Not Required, System ID 361788CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 102133-1-Centricity CA1000-00105849, UDI UDI not required, System ID 817599CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10811-1-Centricity CA1000-01893386, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID CAMC01893386;
Model/Catalog/Code # 2038437-001, Installed Product ID 2565-5-Centricity CA1000-00100411, UDI UDI Not Required, System ID 608775CA1;
Model/Catalog/Code # 2038437-001, Installed Product ID 102098-1-Centricity CA1000-00098301, UDI UDI not required, System ID 414447CA1;
Model/Catalog/Code # 2038437-001, Installed Product ID 106175-1-Centricity CA1000-22219327, UDI (01)00840682125260(10)20P14J2038437016, System ID MILWAUKEE22219327;
Model/Catalog/Code # 2038437-001, Installed Product ID 103367-1-Centricity CA1000-00096997, UDI UDI not required, System ID 262687CA1SW1;
Model/Catalog/Code # 2038437-001, Installed Product ID 103742-1-Centricity CA1000-00107068, UDI UDI not required, System ID 304347CA1;
Model/Catalog/Code # 2038437-001, Installed Product ID 10783-1-Centricity CA1000-00107072, UDI UDI not required, System ID 304424CASSASW;
Model/Catalog/Code # 2038437-001, Installed Product ID 080041-Centricity CA1000-00071765, UDI UDI Not required, System ID 080041ST03;
Model/Catalog/Code # 2038437-001, Installed Product ID 080041-Centricity CA1000-00059699, UDI UDI Not required, System ID 080041ST02;
Model/Catalog/Code # 2038437-001, Installed Product ID 290472-Centricity CA1000-20091941, UDI (01)00840682125260(10)20P14E2038437016, System ID HCIT4753340CA;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-02181400, UDI UDI Not Required, System ID 604875CA1000SW11;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667187, UDI UDI Not Required, System ID 604875CA1000SW01;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667188, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 604875CA1000SW02;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667189, UDI UDI Not Required, System ID 604875CA1000SW03;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667190, UDI UDI Not Required, System ID 604875CA1000SW04;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667191, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 604875CA1000SW05;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667192, UDI UDI Not Required, System ID 604875CA1000SW06;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667193, UDI UDI Not Required, System ID 604875CA1000SW07;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667194, UDI UDI Not Required, System ID 604875CA1000SW08;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667195, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 604875CA1000SW09;
Model/Catalog/Code # 2038437-001, Installed Product ID 282519-Centricity CA1000-00667196, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 604875CA1000SW10;
Model/Catalog/Code # 2038437-001, Installed Product ID 283039-Centricity CA1000-02189544, UDI (01)00840682125260(10)20P14M2038437016, System ID 613548AI1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 283039-Centricity CA1000-02189546, UDI (01)00840682125260(10)20P14M2038437016, System ID 613548CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 283007-Centricity CA1000-00828863, UDI UDI Not Required, System ID 905895CA1000A;
Model/Catalog/Code # 2038437-001, Installed Product ID 282920-Centricity CA1000-21349651, UDI (01)00840682125260(10)20P14M2038437016, System ID 416461CA1000A;
Model/Catalog/Code # 2038437-001, Installed Product ID 282920-Centricity CA1000-21349652, UDI (01)00840682125260(10)20P14M2038437016, System ID 416461CA1000B;
Model/Catalog/Code # 2038437-001, Installed Product ID 282920-Centricity CA1000-21349653, UDI (01)00840682125260(10)20P14M2038437016, System ID 416461CA1000C;
Model/Catalog/Code # 2038437-001, Installed Product ID 282913-1342026-Centricity CA1000-00604941, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID 416360CA10001sw;
Model/Catalog/Code # 2038437-001, Installed Product ID 282913-1342026-Centricity CA1000-00604942, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID 416360CA10002sw;
Model/Catalog/Code # 2038437-001, Installed Product ID 282913-1342026-Centricity CA1000-00604943, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID 416360CA10003sw;
Model/Catalog/Code # 2038437-001, Installed Product ID 282913-1342026-Centricity CA1000-00604945, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID 416360CA10004sw;
Model/Catalog/Code # 2038437-001, Installed Product ID 282913-1342026-Centricity CA1000-00604946, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID 416360CA10005sw;
Model/Catalog/Code # 2038437-001, Installed Product ID 14398615-Centricity CA1000-22180881, UDI (01)00840682125260(10)20P14J2038437016, System ID 14398615-Centricity CA1000-22180881;
Model/Catalog/Code # 2038437-001, Installed Product ID 103487-Centricity CA1000-21327072, UDI (01)00840682125260(10)20P14J2038437016, System ID 082499170038;
Model/Catalog/Code # 2038437-001, Installed Product ID 676589-Centricity CA1000-01283955, UDI UDI Not Required, System ID 082455080025;
Model/Catalog/Code # 2038437-001, Installed Product ID 676589-Centricity CA1000-21326522, UDI (01)00840682125260(10)20P14J2038437016, System ID 082499080089;
Model/Catalog/Code # 2038437-001, Installed Product ID 485728-Centricity CA1000-21326499, UDI (01)00840682125260(10)20P14J2038437016, System ID 082499080088;
Model/Catalog/Code # 2038437-001, Installed Product ID 445933-Centricity CA1000-24986501, UDI (01)00840682125260(10)20P14M2038437016, System ID 445933-Centricity CA1000-24986501;
Model/Catalog/Code # 2038437-001, Installed Product ID 270440-Centricity CA1000-21286717, UDI (01)00840682125260(10)20P14M2038437016, System ID 082499290077;
Model/Catalog/Code # 2038437-001, Installed Product ID 270440-Centricity CA1000-23208991, UDI (01)00840682125260(10)20P14M2038437016, System ID 082499290077;
Model/Catalog/Code # 2038437-001, Installed Product ID 270440-Centricity CA1000-23208992, UDI (01)00840682125260(10)20P14M2038437016, System ID 082499290077;
Model/Catalog/Code # 2038437-001, Installed Product ID 270440-Centricity CA1000-23208993, UDI (01)00840682125260(10)20P14M2038437016, System ID 082499290077;
Model/Catalog/Code # 2038437-001, Installed Product ID 329371-Centricity CA1000-21326494, UDI (01)00840682125260(10)20P14J2038437016, System ID 082499290055;
Model/Catalog/Code # 2038437-001, Installed Product ID 417238-Centricity CA1000-21327080, UDI (01)00840682125260(10)20P14J2038437016, System ID 082499050079;
Model/Catalog/Code # 2038437-001, Installed Product ID 14613630-Centricity CA1000-21327075, UDI (01)00840682125260(10)20P14J2038437016, System ID 082499180095;
Model/Catalog/Code # 2038437-001, Installed Product ID 322771-Centricity CA1000-21327078, UDI (01)00840682125260(10)20P14J2038437016, System ID 082499100430;
Model/Catalog/Code # 2038437-001, Installed Product ID 322771-Centricity CA1000-21327077, UDI (01)00840682125260(10)20P14J2038437016, System ID 082499100426;
Model/Catalog/Code # 2038437-001, Installed Product ID 2690882-Centricity CA1000-19676018, UDI (01)00840682125260(10)20P14H2038437016, System ID HCIT1706622CA;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02031272, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CH Annaba;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02093137, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CHU Amiens;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02093138, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CHU Amiens;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02321213, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CH Beni Messous;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02321214, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CH Beni Messous;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02321216, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CH Perigueux;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02321217, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CH Beni Messous;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02291845, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CH Gonnesse;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02291847, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CH Cayenne;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02291849, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - Bordeaux Clinique St Augustin;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-02291875, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - Strabourg Clinique de l'Orangerie;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-01458176, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - Lyon - Clinique de la Sauvegarde;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-01458195, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - CH Pau;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-01810827, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - Hopital Priv¿ Dijon Bourgogne;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-01815969, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Medireport - CHT Gaston-Bourret Noumea;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23257801, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Ch Compi¿gne;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23257806, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Ch METZ;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23257813, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Ch Marignane;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23257817, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Chu Martinique;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23258666, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Chu Nancy;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23258716, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Ch Cholet;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260790, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Ch Compi¿gne;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260791, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Ch Compi¿gne;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260792, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Ch Compi¿gne;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260793, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Ch Compi¿gne;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260794, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Ch METZ;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260845, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Chu Martinique;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260889, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Chu Nancy;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260890, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Chu Nancy;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260892, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Chu Nancy;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260893, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Chu Nancy;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260894, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Chu Nancy;
Model/Catalog/Code # 2038437-001, Installed Product ID Medireport-Centricity CA1000-23260895, UDI (01)00840682125260(10)20P14M2038437016, System ID Medireport - Chu Nancy;
Model/Catalog/Code # 2038437-001, Installed Product ID M51736-Centricity CA1000-00043875, UDI UDI Not required, System ID M5173622;
Model/Catalog/Code # 2038437-001, Installed Product ID M51736-Centricity CA1000-00043876, UDI UDI Not required, System ID M5173623;
Model/Catalog/Code # 2038437-001, Installed Product ID M51736-Centricity CA1000-00043877, UDI UDI Not required, System ID M5173625;
Model/Catalog/Code # 2038437-001, Installed Product ID D48589-Centricity CA1000-00044449, UDI UDI Not required, System ID D4858912;
Model/Catalog/Code # 2038437-001, Installed Product ID 1851954X - 553003-Centricity CA1000-00000726, UDI UDI not required, System ID 553003PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID HC2840-Centricity CA1000-00001304, UDI UDI not required, System ID HC2840PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID HC1536-Centricity CA1000-00001156, UDI UDI not required, System ID HC1536PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID 2081255X - 622223-Centricity CA1000-00000798, UDI UDI Not Required, System ID 622223PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID 120111 - UESS01-Centricity CA1000-00001918, UDI UDI Not required, System ID UESS01PC05;
Model/Catalog/Code # 2038437-001, Installed Product ID X9007643 - HC7643-Centricity CA1000-00001521, UDI UDI Not required, System ID HC7643PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID UJJF01-Centricity CA1000-00001957, UDI (01)00840682125260(10)20P14J2038437016, System ID UJJF01PC06;
Model/Catalog/Code # 2038437-001, Installed Product ID HC1217-Centricity CA1000-00001118, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID HC1217PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID X9004359 - HC4359-Centricity CA1000-00001428, UDI UDI Not required, System ID HC4359PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID X9004014 - HC4014-Centricity CA1000-00001382, UDI (01)00840682125260(10)20P14M2038437016, System ID HC4014PC02;
Model/Catalog/Code # 2038437-001, Installed Product ID CC2266-Centricity CA1000-00000905, UDI (01)00840682125260(10)20P14M2038437016, System ID CC2266PC02;
Model/Catalog/Code # 2038437-001, Installed Product ID X9000468 - HC0468-Centricity CA1000-00001043, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID HC0468PC02;
Model/Catalog/Code # 2038437-001, Installed Product ID HHAN01-Centricity CA1000-02265341, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID tobecreated;
Model/Catalog/Code # 2038437-001, Installed Product ID 362520-Centricity CA1000-02332540, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID .;
Model/Catalog/Code # 2038437-001, Installed Product ID 362520-Centricity CA1000-02314228, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID .;
Model/Catalog/Code # 2038437-001, Installed Product ID X9009025 - HC9025-Centricity CA1000-00001558, UDI UDI Not Required, System ID HC9025PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID X9003387 - HHEI01-Centricity CA1000-00001608, UDI UDI Not required, System ID HHEI01PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID UC1989-Centricity CA1000-22186141, UDI (01)00840682125260(10)20P14J2038437016, System ID tobecreated;
Model/Catalog/Code # 2038437-001, Installed Product ID HC9020-Centricity CA1000-02371112, UDI (01)00840682125260(10)20P14J2038437016, System ID tobecreated;
Model/Catalog/Code # 2038437-001, Installed Product ID HC9020-Centricity CA1000-00001550, UDI (01)00840682125260(10)20P14M2038437016, System ID HC9020PC02;
Model/Catalog/Code # 2038437-001, Installed Product ID HKOE01-Centricity CA1000-00001631, UDI UDI not required, System ID HKOE01PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID 1766262X - 536596-Centricity CA1000-00000712, UDI UDI Not required, System ID 536596PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID X9004621 - HLAH01-Centricity CA1000-00059332, UDI UDI not required, System ID HLAH01PC04;
Model/Catalog/Code # 2038437-001, Installed Product ID HL0090-Centricity CA1000-00001633, UDI (01)00840682125260(10)20P14M2038437016, System ID HL0090PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID X9002035 - UX2305-Centricity CA1000-00001977, UDI (01)00840682125260(10)20P14M2038437016, System ID UX2305PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID 606427-Centricity CA1000-20509409, UDI (01)00840682125260(10)20P14E2038437016, System ID 606427DH01;
Model/Catalog/Code # 2038437-001, Installed Product ID CC0668-Centricity CA1000-00000902, UDI (01)00840682125260(10)20P14M2038437016, System ID CC0668PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID HC1642-Centricity CA1000-00001193, UDI (01)00840682125260(10)20P14M2038437016, System ID HC1642PC02;
Model/Catalog/Code # 2038437-001, Installed Product ID X9001819 - HC1819-Centricity CA1000-00001218, UDI (01)00840682125260(10)20P14M2038437016, System ID HC1819PC04;
Model/Catalog/Code # 2038437-001, Installed Product ID 1424036-Centricity CA1000-20685090, UDI (01)00840682125260(10)20P14E2038437016, System ID .;
Model/Catalog/Code # 2038437-001, Installed Product ID PC0052-Centricity CA1000-00111454, UDI UDI Not required, System ID PC0052PC03;
Model/Catalog/Code # 2038437-001, Installed Product ID HNUE04-Centricity CA1000-00001645, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID HNUE04PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID 2073908X - 620200-Centricity CA1000-00162674, UDI (01)00840682125260(10)20P14M2038437016, System ID 620200PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID 507902-Centricity CA1000-00000587, UDI UDI Not required, System ID 507902PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID HC0222-Centricity CA1000-00001025, UDI UDI Not Required, System ID HC0222PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID X9003884 - HC3884-Centricity CA1000-02395246, UDI (01)00840682125260(10)20P14M2038437016, System ID .;
Model/Catalog/Code # 2038437-001, Installed Product ID X9003884 - HC3884-Centricity CA1000-00555963, UDI UDI not required, System ID HC3884PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID UC4372-Centricity CA1000-00001901, UDI UDI Not required, System ID UC4372PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID X9000019 - UULM03-Centricity CA1000-00001972, UDI UDI Not required, System ID UULM03PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID HC0099-Centricity CA1000-02032905, UDI UDI Not Required, System ID HC0099PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID TP0575-Centricity CA1000-00001806, UDI UDI not required, System ID TP0575PC01;
Model/Catalog/Code # 2038437-001, Installed Product ID HU1099-Centricity CA1000-02170352, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID HU1099SW03;
Model/Catalog/Code # 2038437-001, Installed Product ID HU1099-Centricity CA1000-02170353, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID HU1099WS01;
Model/Catalog/Code # 2038437-001, Installed Product ID 254526-Centricity CA1000-24944066, UDI (01)00840682125260(10)20P14M2038437016, System ID CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID HABIB-Centricity CA1000-01821061, UDI UDI Not Required, System ID HABIB_CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID TMH-Centricity CA1000-01820948, UDI (01)00840682125260(10)20P14J2038437016, System ID 083055207483617;
Model/Catalog/Code # 2038437-001, Installed Product ID AIIMS-CN-Centricity CA1000-01892398, UDI (01)00840682125260(10)20P14M2038437016, System ID 083055003435816;
Model/Catalog/Code # 2038437-001, Installed Product ID MIOT-Centricity CA1000-01819915, UDI (01)00840682125260(10)20P14J2038437016, System ID 083055803467916;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000128482-Centricity CA1000-21337484, UDI (01)00840682125260(10)20P14E2038437016, System ID DR_SRUJAL_SHAH_CA1000_20;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000129151-Centricity CA1000-21338125, UDI (01)00840682125260(10)20P14E2038437016, System ID VH_CA1000_20;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000129616-Centricity CA1000-21335934, UDI (01)00840682125260(10)20P14M2038437016, System ID CH_CA1000_2.0;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000128152-Centricity CA1000-21337496, UDI (01)00840682125260(10)20P14M2038437016, System ID RHC_CA1000_20;
Model/Catalog/Code # 2038437-001, Installed Product ID 1296534-Centricity CA1000-21335931, UDI (01)00840682125260(10)20P14M2038437016, System ID MMM_WS1-CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 1296534-Centricity CA1000-25177693, UDI (01)00840682125260(10)20P14M2038437016, System ID MMM-WS2-CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 1296534-Centricity CA1000-25177694, UDI (01)00840682125260(10)20P14M2038437016, System ID MMM-WS3-CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000129358-Centricity CA1000-21337486, UDI (01)00840682125260(10)20P14E2038437016, System ID GCC_CA1000_20;
Model/Catalog/Code # 2038437-001, Installed Product ID 9858601-Centricity CA1000-21335936, UDI (01)00840682125260(10)20P14M2038437016, System ID DR_MOHD_CA1000_2.0;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000129144-Centricity CA1000-21337493, UDI (01)00840682125260(10)20P14M2038437016, System ID KMSC_CA1000_1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000128456-Centricity CA1000-21337494, UDI (01)00840682125260(10)20P14E2038437016, System ID MH_CA1000_20;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000129084-Centricity CA1000-21337495, UDI (01)00840682125260(10)20P14E2038437016, System ID NHCI_CA1000_20;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000129488-Centricity CA1000-21338124, UDI (01)00840682125260(10)20P14E2038437016, System ID SGRH_CA1000_20;
Model/Catalog/Code # 2038437-001, Installed Product ID 1000129439-Centricity CA1000-21337488, UDI (01)00840682125260(10)20P14J2038437016, System ID GMC_CA1000_20;
Model/Catalog/Code # 2038437-001, Installed Product ID 1401508-Centricity CA1000-24341896, UDI (01)00840682125260(10)20P14E2038437016, System ID 5512AW0032XR1_SO4293494_Line47_HCS_Bundle;
Model/Catalog/Code # 2038437-001, Installed Product ID 1360525-Centricity CA1000-24341744, UDI (01)00840682125260(10)20P14E2038437016, System ID 0835490040XVAW1_SO4017068_Line153_HCS_Bundle;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02265925, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02265926, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02265927, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02265928, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02041527, UDI (01)00840682125260(10)20P14J2038437016, System ID PTMULYA-CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02041910, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02041911, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02041912, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02041913, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02042684, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02042685, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02042686, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02042687, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02043217, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02090080, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02093423, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02093424, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02098782, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02098783, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02098784, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02098785, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02098786, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 327750-Centricity CA1000-02263361, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 81031-Centricity CA1000-02287220, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID IE1031RA01;
Model/Catalog/Code # 2038437-001, Installed Product ID IE1077-Centricity CA1000-22154168, UDI (01)00840682125260(10)20P14J2038437016, System ID 81077TRS03;
Model/Catalog/Code # 2038437-001, Installed Product ID IL1469-Centricity CA1000-00040968, UDI (01)00840682125260(10)20P14J2038437016, System ID IL1469IT03;
Model/Catalog/Code # 2038437-001, Installed Product ID A54039-Centricity CA1000-00040774, UDI (01)00840682125260(10)20P14J2038437016, System ID A5403915;
Model/Catalog/Code # 2038437-001, Installed Product ID D58105-Centricity CA1000-00040369, UDI (01)00840682125260(10)20P14J2038437016, System ID D5810526;
Model/Catalog/Code # 2038437-001, Installed Product ID A1892038-Centricity CA1000-02093304, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID A189203802;
Model/Catalog/Code # 2038437-001, Installed Product ID A54013-Centricity CA1000-00040698, UDI UDI Not Required, System ID A5401385;
Model/Catalog/Code # 2038437-001, Installed Product ID B58208-Centricity CA1000-00040397, UDI (01)00840682125260(10)20P14J2038437016, System ID B5820891;
Model/Catalog/Code # 2038437-001, Installed Product ID A51194-Centricity CA1000-00040516, UDI (01)00840682125260(10)20P14J2038437016, System ID A5119446;
Model/Catalog/Code # 2038437-001, Installed Product ID A51151-Centricity CA1000-00040787, UDI UDI Not Required, System ID A5115142;
Model/Catalog/Code # 2038437-001, Installed Product ID A51581-Centricity CA1000-00040866, UDI UDI Not required, System ID A51581111;
Model/Catalog/Code # 2038437-001, Installed Product ID A51581-Centricity CA1000-00040871, UDI UDI Not required, System ID A51581126;
Model/Catalog/Code # 2038437-001, Installed Product ID A51581-Centricity CA1000-00040874, UDI UDI Not required, System ID A51581122;
Model/Catalog/Code # 2038437-001, Installed Product ID A51581-Centricity CA1000-00040875, UDI UDI Not required, System ID A5158185;
Model/Catalog/Code # 2038437-001, Installed Product ID A57219-Centricity CA1000-00040567, UDI UDI Not required, System ID A5721908;
Model/Catalog/Code # 2038437-001, Installed Product ID A51051-Centricity CA1000-00040734, UDI (01)00840682125260(10)20P14J2038437016, System ID A51051110;
Model/Catalog/Code # 2038437-001, Installed Product ID B51772-Centricity CA1000-00040400, UDI (01)00840682125260(10)20P14J2038437016, System ID B5177230;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130453-Centricity CA1000-01279453, UDI UDI Not Required, System ID PK1471;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130453-Centricity CA1000-01279454, UDI UDI Not Required, System ID PK1472;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130453-Centricity CA1000-01279455, UDI UDI Not Required, System ID PK1473;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130314-Centricity CA1000-02089303, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS006;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130314-Centricity CA1000-02089304, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS007;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130314-Centricity CA1000-02089305, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS008;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130314-Centricity CA1000-02089306, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS009;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130314-Centricity CA1000-01279408, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1325;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130314-Centricity CA1000-01279412, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1329;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130314-Centricity CA1000-01279414, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1330;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130087-Centricity CA1000-01984456, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1536;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130087-Centricity CA1000-01984457, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1537;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130087-Centricity CA1000-01984458, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1538;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130087-Centricity CA1000-01984459, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1539;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130087-Centricity CA1000-01984460, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1540;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130087-Centricity CA1000-01984461, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1543;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-230109-Centricity CA1000-22298029, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1552;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-230109-Centricity CA1000-22298030, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1553;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-230109-Centricity CA1000-22298031, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1554;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-230157-Centricity CA1000-20021135, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS024;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-270581-Centricity CA1000-01279510, UDI UDI Not Required, System ID PK1493;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-380108-Centricity CA1000-01279363, UDI UDI Not Required, System ID PK1277;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-400042-Centricity CA1000-01279351, UDI UDI Not Required, System ID PK1263;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-420052-Centricity CA1000-15141138, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS021;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-420052-Centricity CA1000-15141139, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS022;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-420052-Centricity CA1000-15141140, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS023;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-240130-Centricity CA1000-02183913, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS001;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-240130-Centricity CA1000-02183914, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS002;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-240130-Centricity CA1000-02183916, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS003;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-260121-Centricity CA1000-01279457, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1541;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-260121-Centricity CA1000-01279458, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1542;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-010974-Centricity CA1000-01279462, UDI UDI Not Required, System ID PK1494;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-270367-Centricity CA1000-20501595, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS026;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-270367-Centricity CA1000-20501596, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS027;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-270367-Centricity CA1000-20501597, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS028;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-040271-Centricity CA1000-01279232, UDI UDI Not Required, System ID PK0991;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-330087-Centricity CA1000-01279400, UDI UDI Not Required, System ID PK1316;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-330087-Centricity CA1000-01279446, UDI UDI Not Required, System ID PK1466;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-330087-Centricity CA1000-01279492, UDI UDI Not Required, System ID PK1467;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-330087-Centricity CA1000-21785110, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1549;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-330087-Centricity CA1000-21785111, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1550;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-320008-Centricity CA1000-02344211, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS016;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-320008-Centricity CA1000-02344212, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS017;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-320008-Centricity CA1000-02344213, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS018;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-320008-Centricity CA1000-01801162, UDI UDI Not Required, System ID JF0995;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340347-Centricity CA1000-01891095, UDI UDI Not Required, System ID PK1534;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340237-Centricity CA1000-01817039, UDI UDI Not Required, System ID PK1532;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340173-Centricity CA1000-01279451, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1489;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340173-Centricity CA1000-01279495, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1485;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340173-Centricity CA1000-01279496, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1484;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340173-Centricity CA1000-01279497, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1479;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340173-Centricity CA1000-01279502, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1488;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340173-Centricity CA1000-21785106, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1548;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340173-Centricity CA1000-23290351, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1569;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340173-Centricity CA1000-23290371, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1570;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-340499-Centricity CA1000-01279467, UDI UDI Not Required, System ID PK1498;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-380097-Centricity CA1000-14374787, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS020;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-380123-Centricity CA1000-20101321, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID PKS025;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-380009-Centricity CA1000-01279426, UDI UDI Not Required, System ID PK1339;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130440-Centricity CA1000-02040805, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1544;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130440-Centricity CA1000-02040806, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1545;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130440-Centricity CA1000-02040807, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1546;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130440-Centricity CA1000-02040808, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1547;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-130422-Centricity CA1000-01984313, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1535;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298215, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1555;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298216, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1556;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298217, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1557;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298218, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1558;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298219, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1559;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298220, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1560;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298221, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1561;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298223, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1562;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298224, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1563;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298225, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1564;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298226, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1565;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298228, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1566;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298231, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1567;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-070013-Centricity CA1000-22298233, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1568;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100014-Centricity CA1000-02343523, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS011;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100014-Centricity CA1000-02343524, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS012;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100014-Centricity CA1000-02343525, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS013;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100014-Centricity CA1000-02343526, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS014;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100014-Centricity CA1000-02343527, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS015;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100014-Centricity CA1000-01279433, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1347;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100014-Centricity CA1000-01279434, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1348;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100014-Centricity CA1000-01825457, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1529;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100088-Centricity CA1000-01767277, UDI UDI Not Required, System ID PK1527;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100088-Centricity CA1000-01767278, UDI UDI Not Required, System ID PK1528;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-100088-Centricity CA1000-22298025, UDI (01)00840682125260(10)20P14M2038437016, System ID PK1551;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-081357-Centricity CA1000-02266057, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS010;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-081007-Centricity CA1000-01279512, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1506;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-081007-Centricity CA1000-01279513, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1507;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-081007-Centricity CA1000-01279514, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1508;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-081007-Centricity CA1000-01279515, UDI (01)00840682125260(10)20P14J2038437016, System ID PK1509;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-420388-Centricity CA1000-01279511, UDI UDI Not Required, System ID PK1501;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-221185-Centricity CA1000-01279321, UDI UDI Not Required, System ID PK1133;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-220005-Centricity CA1000-02091033, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS004;
Model/Catalog/Code # 2038437-001, Installed Product ID JP-220005-Centricity CA1000-02189852, UDI (01)00840682125260(10)20P14J2038437016, System ID PKS005;
Model/Catalog/Code # 2038437-001, Installed Product ID 851291-Centricity CA1000-02267862, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 0850550249;
Model/Catalog/Code # 2038437-001, Installed Product ID 851746-Centricity CA1000-18990528, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 0850550152;
Model/Catalog/Code # 2038437-001, Installed Product ID 851523-Centricity CA1000-02090679, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 851523-Centricity CA1000-02091038, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 851523-Centricity CA1000-02092831, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 851523-Centricity CA1000-02157452, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 851523-Centricity CA1000-02172339, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 851523-Centricity CA1000-02172656, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID Not Applicable;
Model/Catalog/Code # 2038437-001, Installed Product ID 851118-Centricity CA1000-02299639, UDI UDI Not Required, System ID 0850550208;
Model/Catalog/Code # 2038437-001, Installed Product ID 851118-Centricity CA1000-02401007, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 0850550208;
Model/Catalog/Code # 2038437-001, Installed Product ID 851118-Centricity CA1000-02401008, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 0850550208;
Model/Catalog/Code # 2038437-001, Installed Product ID 1221080-Centricity CA1000-20856225, UDI (01)00840682125260(10)20P14E2038437016, System ID 0850550140;
Model/Catalog/Code # 2038437-001, Installed Product ID 851468-INHAUH-CVIT-Centricity CA1000-01461264, UDI (01)00840682125260(10)2.0SP14C2038437016C, System ID 0850550316;
Model/Catalog/Code # 2038437-001, Installed Product ID 852004-Centricity CA1000-01285955, UDI UDI Not Required, System ID 0850550240;
Model/Catalog/Code # 2038437-001, Installed Product ID 59219_ERPerror-Centricity CA1000-21350998, UDI (01)00840682125260(10)20P14E2038437016, System ID KW1155PC06;
Model/Catalog/Code # 2038437-001, Installed Product ID 1296729-Centricity CA1000-24341732, UDI (01)00840682125260(10)20P14E2038437016, System ID 1553IT0032_SO4560667_Line24_HCS_Bundle;
Model/Catalog/Code # 2038437-001, Installed Product ID 1269390-Centricity CA1000-24341707, UDI (01)00840682125260(10)20P14E2038437016, System ID 1553AW0146XR1_SO4448676_Line90_HCS_Bundle;
Model/Catalog/Code # 2038437-001, Installed Product ID 2438975-Centricity CA1000-24341619, UDI (01)00840682125260(10)20P14E2038437016, System ID 1553AW0147XR1_SO4448676_Line41_HCS_Bundle;
Model/Catalog/Code # 2038437-001, Installed Product ID 12117608-Centricity CA1000-19649610, UDI UDI Not Required, System ID HCIT4794572CA;
Model/Catalog/Code # 2038437-001, Installed Product ID 2611279-Centricity CA1000-02154726, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID HCIT4538700CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 274800-Centricity CA1000-02287955, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID OECFBXXXE0039AW-2;
Model/Catalog/Code # 2038437-001, Installed Product ID _329368-Centricity CA1000-21752348, UDI (01)00840682125260(10)20P14J2038437016, System ID HCIT5044634CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID 600042-Centricity CA1000-22267349, UDI (01)00840682125260(10)20P14J2038437016, System ID .;
Model/Catalog/Code # 2038437-001, Installed Product ID 600042-Centricity CA1000-22267353, UDI (01)00840682125260(10)20P14J2038437016, System ID To be provided;
Model/Catalog/Code # 2038437-001, Installed Product ID QA1015-Centricity CA1000-21350997, UDI (01)00840682125260(10)20P14E2038437016, System ID QA1015PP06;
Model/Catalog/Code # 2038437-001, Installed Product ID 605007-Centricity CA1000-16385126, UDI (01)00840682125260(10)20P14J2038437016, System ID SA2485VA01;
Model/Catalog/Code # 2038437-001, Installed Product ID SA2191-Centricity CA1000-20879436, UDI (01)00840682125260(10)20P14J2038437016, System ID SA2191_CA1000_SO#5286069;
Model/Catalog/Code # 2038437-001, Installed Product ID SA1008-Centricity CA1000-20879437, UDI (01)00840682125260(10)20P14E2038437016, System ID SA1008_CA1000;
Model/Catalog/Code # 2038437-001, Installed Product ID GEC218-Centricity CA1000-00044870, UDI (01)00840682125260(10)20P14M2038437016, System ID GEC21843;
Model/Catalog/Code # 2038437-001, Installed Product ID 225030-Centricity CA1000-00000241, UDI (01)00840682125260(10)20P14M2038437016, System ID 225030PC04;
Model/Catalog/Code # 2038437-001, Installed Product ID 0853551073-Centricity CA1000-19678120, UDI (01)00840682125260(10)20P14M2038437016, System ID 0853551073;
Model/Catalog/Code # 2038437-001, Installed Product ID 107127-Centricity CA1000-22300468, UDI (01)00840682125260(10)20P14M2038437016, System ID 0853551140;
Model/Catalog/Code # 2038437-001, Installed Product ID 5378586-Centricity CA1000-24342071, UDI (01)00840682125260(10)20P14K2038437016, System ID 1573IT0002_SO4714137_Line41_HCS_Bundle;
Model/Catalog/Code # 2038437-001, Installed Product ID RXQ50-Centricity CA1000-01977874, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID 00638VAS03; 00638UPS01; 00638CRS04;
Model/Catalog/Code # 2038437-001, Installed Product ID 60044-Centricity CA1000-00002025, UDI (01)00840682125260(10)20P14J2038437016, System ID 60044TRS01;
Model/Catalog/Code # 2038437-001, Installed Product ID 60490-Centricity CA1000-00003387, UDI UDI not required, System ID 60490TRS03;
Model/Catalog/Code # 2038437-001, Installed Product ID 60150-Centricity CA1000-00003512, UDI (01)00840682125260(10)20P14J2038437016, System ID 60150TRS01;
Model/Catalog/Code # 2038437-001, Installed Product ID 61065-Centricity CA1000-00003505, UDI (01)00840682125260(10)20P14J2038437016, System ID 61065TRS01;
Model/Catalog/Code # 2038437-001, Installed Product ID 80115-C-PACS EE Backend-00005734, UDI UDI Not required, System ID 60115CRS01;
Model/Catalog/Code # 2038437-001, Installed Product ID 60199-Centricity CA1000-00056266, UDI UDI Not Required, System ID 60199TRS05;
Model/Catalog/Code # 2038437-001, Installed Product ID RD301-Centricity CA1000-00006091, UDI UDI Not required, System ID 60234TRS01;
Model/Catalog/Code # 2038437-009, Installed Product ID To be Provided, UDI UDI Not Required, System ID 397713PC01;
Model/Catalog/Code # 2038437-009, Installed Product ID Centricity Cardiology CA1000 , UDI UDI Not Required, System ID LDD0460;
Model/Catalog/Code # 2038437-014, Installed Product ID -Centricity CA1000-19638476, UDI UDI Not Required, System ID HCIT4936689CA1000;
Model/Catalog/Code # 2038437-015 , Installed Product ID 8133-Centricity CA1000-02176382, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID HCITCA1000B;
Model/Catalog/Code # 2038437-015 , Installed Product ID 8133-Centricity CA1000-02176383, UDI (01)00840682125260(10)2.0SPA142038437016B, System ID HCITCA1000C;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 233490II2, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000230779II1, UDI UDI Not Required, System ID 00000230779II1;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000206705II6, UDI UDI Not Required, System ID 989907INNOVA;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000204679II5, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000204680II3, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 240467II1, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 242861II3, UDI UDI Not Required, System ID HSSP1RM1D656;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000216476II2, UDI UDI Not Required, System ID 970542XR656;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 243162II5, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000222692II6, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000222699II1, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 169825II7, UDI UDI Not Required, System ID 0910555032CA;
Model/Catalog/Code # 2038437-0XX, Installed Product ID Centricity Cardiology CA1000 , UDI UDI Not Required, System ID 080041ST02;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000192785II4, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000195401II5, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000195825II5, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000195827II1, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 00000195828II9, UDI UDI Not Required, System ID To be provided;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID IS1001PC01;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055601001512;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055241260315;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055240006012;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055820007313;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055860021313;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055281120412;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055200018312;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055207483617;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055600027812;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055040018812;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055041288114;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055000041312;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055803033114;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055804781615;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055601008113;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 83055600007012;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 118564114, UDI UDI Not Required, System ID 835160016;
Model/Catalog/Code # 2038437-0XX, Installed Product ID 184161II8, UDI UDI Not Required, System ID 850550152;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 1553IT0032;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 853550927;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID 853550930;
Model/Catalog/Code # 2038437-0XX, Installed Product ID To be provided , UDI UDI Not Required, System ID UA4230PC01;
Model/Catalog/Code # 2109571-007, Installed Product ID BioTechVenez-Centricity CA1000-19737484, UDI (01)00840682125260(10)20P14E2038437016, System ID Not Applicable;
Model/Catalog/Code # 2109571-010, Installed Product ID 1596868_Lakemac-Centricity CA1000-01283281, UDI (01)00840682125260(10)20P14J2038437016, System ID 0910558008;
Model/Catalog/Code # 2109571-011, Installed Product ID 10955-5-Centricity CA1000-23251234, UDI (01)00840682125260(10)20P14K2038437016, System ID WMMC223251234;
Model/Catalog/Code # 2109571-012, Installed Product ID 801303_Flinderspriv-Centricity CA1000-01975975, UDI (01)00840682125260(10)20P14M2038437016, System ID 0910555032;
Model/Catalog/Code # MANLEGACY111, Installed Product ID 1596868_Materbris-Centricity CA1000-01283295, UDI UDI Not Required, System ID 0910554032;
Model/Catalog/Code # MANLEGACY111, Installed Product ID CVIT_KW1011-Centricity CA1000-00165354, UDI UDI not required, System ID KW1011SV01;
Model/Catalog/Code # MANLEGACY111, Installed Product ID CVIT_AlShifa-Centricity CA1000-00165355, UDI UDI not required, System ID Not Applicable;
Model/Catalog/Code # MANLEGACY111, Installed Product ID RU3623-Centricity CA1000-21331795, UDI UDI Not Required, System ID RU3623ID03;
Model/Catalog/Code # MANLEGACY112, Installed Product ID 4029-47-Centricity CA1000-01804717, UDI UDI Not Required, System ID NSLI01804717;
Model/Catalog/Code # MANLEGACY112, Installed Product ID 4361-2-Centricity CA1000-01444473, UDI UDI Not Required, System ID JCBLAIR01444473;
Model/Catalog/Code # MANLEGACY112, Installed Product ID 1497-27-Centricity CA1000-01752513, UDI UDI Not Required, System ID 01752513UPMC;
Model/Catalog/Code # MANLEGACY112, Installed Product ID 10800-3-Centricity CA1000-01460396, UDI UDI Not Required, System ID SMHORI01460396;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-02179434, UDI UDI Not Required, System ID Medireport - CH Bayonne M4164356;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458141, UDI UDI Not Required, System ID Medireport - CH Le Mans;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458166, UDI UDI Not Required, System ID Medireport - CH Belfort;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458168, UDI UDI Not Required, System ID Medireport - CCML;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458170, UDI UDI Not Required, System ID Medireport - CH Pontoise;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458173, UDI UDI Not Required, System ID Medireport - CH St Philippe de l'Homme Lille;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458180, UDI UDI Not Required, System ID Medireport - CH Auxerre;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458184, UDI UDI Not Required, System ID Medireport - CH Blida;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458186, UDI UDI Not Required, System ID Medireport - CH Argenteuil;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458188, UDI UDI Not Required, System ID Medireport - CH Fort de France;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458190, UDI UDI Not Required, System ID Medireport - CH Ottignies;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458192, UDI UDI Not Required, System ID Medireport - CH Sud Francilien (CHSF);
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458410, UDI UDI Not Required, System ID Medireport - CH Foch Suresnes;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458411, UDI UDI Not Required, System ID Medireport - CH Saintonges;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458412, UDI UDI Not Required, System ID Medireport - St Niklaas;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458413, UDI UDI Not Required, System ID Medireport - CH Colmar;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458415, UDI UDI Not Required, System ID Medireport - CH Laval;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458418, UDI UDI Not Required, System ID Medireport - CH Nevers;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458420, UDI UDI Not Required, System ID Medireport - Strasbourg Clinique de l'Orangerie;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458424, UDI UDI Not Required, System ID Medireport - Paris APHP La Piti¿;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458425, UDI UDI Not Required, System ID Medireport - CHU Rouen;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458427, UDI UDI Not Required, System ID Medireport - CH Bastia;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458429, UDI UDI Not Required, System ID Medireport - CH Pessac;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458432, UDI UDI Not Required, System ID Medireport - CH Troyes;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458433, UDI UDI Not Required, System ID Medireport - CH Saran;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01458438, UDI UDI Not Required, System ID Medireport - Paris CH St Joseph;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01445704, UDI UDI Not Required, System ID CVX Divers;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01700461, UDI UDI Not Required, System ID Medireport - CH Metz CLaude Bernard;
Model/Catalog/Code # MANLEGACY112, Installed Product ID Medireport-Centricity CA1000-01807133, UDI UDI Not Required, System ID Medireport - Clinique Axium Aix-en-Provence;
Model/Catalog/Code # MANLEGACY112, Installed Product ID HC4357-Centricity CA1000-01594559, UDI UDI Not Required, System ID HC4357PC02;
Model/Catalog/Code # MANLEGACY112, Installed Product ID 59219_ERPerror-Centricity CA1000-01457357, UDI UDI Not Required, System ID Not Applicable;
Model/Catalog/Code # MANLEGACY112, Installed Product ID 529020-Centricity CA1000-01797226, UDI UDI Not Required, System ID tobecreated;
Model/Catalog/Code # MANLEGACY47, Installed Product ID 102708-1-Centricity CA1000-01273541, UDI UDI Not Required, System ID 574364CA1000;
Model/Catalog/Code # MANLEGACY47, Installed Product ID 13399-10-Centricity CA1000-01291477, UDI UDI Not Required, System ID 317554CA1SW1;
Model/Catalog/Code # MANLEGACY47, Installed Product ID 4362-1-Centricity CA1000-02205323, UDI UDI Not Required, System ID JOHNTMA02205323;
Model/Catalog/Code # MANLEGACY47, Installed Product ID 4152-1-Centricity CA1000-02277731, UDI UDI Not Required, System ID 610988CA1SW1;
Model/Catalog/Code # MANLEGACY47, Installed Product ID 106348-1-Centricity CA1000-20080247, UDI UDI Not Required, System ID 361888CRS;
Model/Catalog/Code # MANLEGACY47, Installed Product ID 9859-76-Centricity CA1000-01301215, UDI UDI Not Required, System ID 214618CA1000SW;
Model/Catalog/Code # MANLEGACY47, Installed Product ID 283066-Centricity CA1000-01363410, UDI UDI Not Required, System ID 905848TCA10001;
Model/Catalog/Code # MANLEGACY47, Installed Product ID Medireport-Centricity CA1000-01458436, UDI UDI Not Required, System ID Medireport - CHR Metz;
Model/Catalog/Code # MANLEGACY47, Installed Product ID PC1695-Centricity CA1000-00001704, UDI UDI not required, System ID PC1695PC01;
Model/Catalog/Code # MANLEGACY47, Installed Product ID HC5284-Centricity CA1000-00001482, UDI UDI not required, System ID HC5284PC02;
Model/Catalog/Code # MANLEGACY47, Installed Product ID HC1995-Centricity CA1000-00001242, UDI UDI not required, System ID HC1995PC01;
Model/Catalog/Code # MANLEGACY47, Installed Product ID A53751-Centricity CA1000-20644429, UDI UDI Not Required, System ID A5375116;
Model/Catalog/Code # MANLEGACY47, Installed Product ID JP-260927-Centricity CA1000-01279148, UDI UDI Not Required, System ID PK0332;
Model/Catalog/Code # MANLEGACY47, Installed Product ID JP-350009-Centricity CA1000-24814093, UDI UDI Not Required, System ID PK1574;
Model/Catalog/Code # MANLEGACY47, Installed Product ID JP-100014-Centricity CA1000-24436397, UDI UDI Not Required, System ID PK1571;
Model/Catalog/Code # MANLEGACY47, Installed Product ID JP-100014-Centricity CA1000-24436400, UDI UDI Not Required, System ID PK1572;
Model/Catalog/Code # MANLEGACY47, Installed Product ID JP-100014-Centricity CA1000-24436403, UDI UDI Not Required, System ID PK1573;
Model/Catalog/Code # MANLEGACY47, Installed Product ID CVIT_KW1018-Centricity CA1000-00165347, UDI UDI Not required, System ID KW1018PC02;
Model/Catalog/Code # MANLEGACY47, Installed Product ID CVIT_KW1001-Centricity CA1000-00165348, UDI UDI Not required, System ID KW1001PC03;
Model/Catalog/Code # MANLEGACY47, Installed Product ID CVIT_KW1098-Centricity CA1000-00165351, UDI UDI Not required, System ID Not Applicable;
Model/Catalog/Code # P00011TS, Installed Product ID 362520-Centricity CA1000-01805902, UDI UDI Not Required, System ID 362520PC01. |
| FEI Number |
3004526608
|
Recalling Firm/ Manufacturer |
GE Healthcare
500 W Monroe St
Chicago
IL
60661-3671
|
| For Additional Information Contact | GE Healthcare 800-437-1171 |
Manufacturer Reason for Recall | There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data. |
FDA Determined Cause 2 | Software design |
| Action | GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 by letter delivered using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken:
"You can continue to use your device.
Please ensure all potential users in your facility are made aware of this safety notification.
Please ensure you have implemented the actions below:
1: Ensure your hospital network is secure
and
2: Outside of the hospital network implement additional safeguards such as a VPN.
Please retain this document for your records.
Please complete and return the attached acknowledgement form to recall.85480@gehealthcare.com."
"If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative."
Per status report 05/23/2025, the firm send a follow-up notice dated 05/13/2025 providing an installation manual addendum that provides instructions for changing passwords. |
| Quantity in Commerce | 633 units |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LLZ 510(K)s with Product Code = LLZ
|
|
|
|