• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall DuraDiagnost

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall DuraDiagnostsee related information
Date Initiated by FirmMarch 10, 2025
Date PostedApril 11, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1554-2025
Recall Event ID 96435
510(K)NumberK141381 K201640 
Product Classification System, x-ray, stationary - Product Code KPR
ProductPhilips DXR System, DuraDiagnost Rel. 4. Stationary X-Ray System.
Code Information REF 712214; UDI: (01)00884838064751(21); Serial No. SN170009, SN170010, SN180072, SN180020 200037, SN160061, SN180075, 200089, SN180045, SN160096.
FEI Number 3016618143
Recalling Firm/
Manufacturer
Philips North America Llc
222 Jacobs St
Cambridge MA 02141-2289
For Additional Information ContactPhilips Customer Care Solutions
1-800-722-9377
Manufacturer Reason
for Recall
Potential for collimator to fall as a result of incorrect installation.
FDA Determined
Cause 2
Under Investigation by firm
ActionConsignees received an URGENT MEDICAL DEVICE CORRECTION NOTICE dated 3/7/25. The notice instructs customers of safety precautions to take during scan preparation to ensure devices are fit for use. If consignees notice that the laser from the collimator or crosshairs of the collimator's light field is not able to align with the silkscreen on the front cover of the detector, consignees are to immediately stop use and call their Philips representative. The provided notice is to be posted near affected systems and circulated to all device users. Philips will reach out to consignees to schedule an inspection of devices by field service engineers. Engineers may also replace the collimator, if necessary. Consignees with questions are to call Philips Customer Care Solutions Center at 1-800-722-9377 from Monday through Friday (8:00 AM to 8:00 PM EST).
Quantity in Commerce83 units
DistributionWorldwide - US Nationwide distribution in the states of AK, FL, KS, MS, NH, NY, OR, PA, TX, WI and the countries of Argentina, Australia, Canada, Chile, China, Germany, Hungary, India, Indonesia, Italy, Kenya, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, Slovenia, South Africa, Spain, Thailand, United Kingdom, Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = KPR
-
-