| | Class 1 Device Recall GE Healthcare CARESTATION |  |
| Date Initiated by Firm | March 21, 2025 |
| Date Posted | April 29, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1624-2025 |
| Recall Event ID |
96634 |
| 510(K)Number | K151570 |
| Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
| Product | GE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000. anesthesia systems |
| Code Information |
GTIN 00840682103985, Serial Numbers: SM623390002MA, SM623390003MA, SM623020017MA, SM623480018MA, SM623480019MA, SM623480020MA, SM623480021MA, SM623110013MA, SM623060017MA, SM623060018MA, SM623060019MA, SM623060020MA, SM624440004MA, SM624440005MA, SM623320003MA, SM624050005MA, SM624050006MA, SM624130001MA, SM624200013MA, SM623440005MA, SM623440006MA, SM624140002MA, SM624480001MA, SM624480002MA, SM624510003MA, SM623110002MA, SM624110001MA, SM624380012MA, SM623360001MA, SM624100016MA, SM625050001MA, SM625050002MA, SM625050003MA, SM623050014MA, SM623320010MA, SM623320011MA, SM623320012MA, SM623360006MA, SM623360007MA, SM624010030MA, SM624040016MA, SM624040017MA, SM624040018MA, SM623020014MA, SM623020011MA, SM623020012MA, SM623020013MA, SM624260001MA, SM623030011MA, SM623030012MA, SM623220006MA, SM624500001MA, SM623480014MA, SM623020016MA, SM623290014MA, SM623360003MA, SM623360004MA, SM624100014MA, SM623060032MA, SM623030006MA, SM623030007MA, SM624200017MA, SM623310030MA, SM624200008MA, SM624200009MA, SM624200010MA, SM624200011MA, SM624200012MA, SM624040006MA, SM624040007MA, SM624110013MA, SM624110014MA, SM624110015MA, SM624370003MA, SM624400001MA, SM624400002MA, SM624400003MA, SM624400004MA, SM624510008MA, SM624510009MA, SM624510010MA, SM624510011MA, SM624510012MA, SM624510013MA, SM623480005MA, SM623480006MA, SM624120012MA, SM623060023MA, SM624160005MA, SM624160007MA, SM624180024MA, SM624260004MA, SM624460002MA, SM623310036MA, SM623310037MA, SM623310038MA, SM624300003MA, SM624300005MA, SM624300006MA, SM624340001MA, SM624340002MA, SM623100015MA, SM625030001MA, SM624320008MA, SM623300005MA, SM623300006MA, SM623300007MA, SM623300008MA, SM623300009MA, SM623300010MA, SM623300011MA, SM623080004MA, SM623080021MA, SM623080022MA, SM623160009MA, SM623160010MA, SM624280003MA, SM624280004MA, SM624140018MA, SM625020002MA, SM625020003MA, SM624090011MA, SM624090012MA, SM624090013MA, SM624050007MA, SM624230008MA, SM624230009MA, SM624230010MA, SM623480015MA, SM623060031MA, SM624060001MA, SM624320009MA, SM624320010MA, SM624320011MA, SM624320012MA, SM624320013MA, SM624320014MA, SM624230012MA, SM624310004MA, SM624260002MA, SM624260003MA, SM623340023MA, SM623340024MA, SM624200001MA, SM624200002MA, SM624060017MA, SM624300004MA, SM624300007MA, SM624310015MA, SM624310016MA, SM624310017MA, SM624310018MA, SM623170009MA, SM623170010MA, SM623140018MA, SM624390001MA, SM624390002MA, SM624390003MA, SM623320001MA, SM623260001MA, SM624080001MA, SM624090002MA, SM624090003MA, SM624090004MA, SM624090008MA, SM624090009MA, SM623200007MA, SM623200008MA, SM623200009MA, SM623200010MA, SM623200011MA, SM624080002MA, SM624080003MA, SM624080004MA, SM624080005MA, SM624080006MA, SM624080007MA, SM624080008MA, SM624080009MA, SM624080010MA, SM624080011MA, SM624080012MA, SM624080013MA, SM623180021MA, SM623180022MA, SM623180023MA, SM623180024MA, SM623070006MA, SM623200016MA, SM623200017MA, SM623500003MA, SM623500004MA, SM623200012MA, SM623320023MA, SM623320024MA, SM623140016MA, SM623250001MA, SM624100012MA, SM624100013MA, SM623020015MA, SM624130015MA, SM624210002MA, SM624380013MA, SM624160008MA, SM623390006MA, SM623390007MA, SM624150001MA, SM624070003MA, SM625040003MA, SM623480016MA, SM624250002MA, SM624250003MA, SM624330002MA, SM624330004MA, SM623110001MA, SM623300016MA, SM623390005MA, SM623430017MA, SM623430018MA, SM624100011MA, SM623420001MA, SM624140003MA, SM624230002MA, SM624230003MA, SM624230004MA, SM623190015MA, SM624150015MA, SM624160006MA, SM624190030MA, SM624200015MA, SM623110003MA, SM625020004MA, SM625020005MA, SM625020006MA, SM625020007MA, SM623170007MA, SM623170008MA, SM624410004MA, SM624410001MA, SM624340006MA, SM624340007MA, SM623380001MA, SM623380002MA, SM623380003MA, SM623380004MA, SM623200002MA, SM623200003MA, SM623200004MA, SM623200005MA, SM623200006MA, SM624190031MA, SM623260008MA, SM623260009MA, SM623260010MA, SM623260011MA, SM624400006MA, SM624400007MA, SM624220022MA, SM624370004MA, SM623370019MA, SM623370020MA, SM623210001MA, SM623210002MA, SM623060015MA, SM623230002MA, SM623040005MA, SM623120026MA, SM623040003MA, SM623040004MA, SM623040006MA, SM623050013MA, SM623160011MA, SM623160012MA, SM623110006MA, SM623370001MA, SM623430003MA, SM623430004MA, SM624400005MA, SM624160011MA, SM624160012MA, SM624160013MA, SM624160014MA, SM624160015MA, SM623040001MA, SM623040002MA, SM623080019MA, SM623080020MA, SM624020010MA, SM624020011MA, SM624020012MA, SM624020013MA, SM624020014MA, SM624020015MA, SM624020016MA, SM623030008MA, SM623030009MA, SM623030010MA, SM623250009MA, SM623250010MA, SM623250011MA, SM623250012MA, SM623090002MA, SM623090003MA, SM623500005MA, SM623500006MA, SM623500007MA, SM623440015MA, SM623440016MA, SM623480022MA, SM623480023MA, SM623480024MA, SM623480025MA, SM624510001MA, SM624510002MA, SM623120029MA, SM623120030MA, SM623120031MA, SM623120032MA, SM623110015MA, SM623110017MA, SM624120001MA, SM624440001MA, SM624510016MA, SM624510017MA, SM624510018MA, SM624510019MA, SM624510020MA, SM624510021MA, SM623390008MA, SM623390009MA, SM624400008MA, SM624400009MA, SM624400010MA, SM624400011MA, SM623060022MA, SM624140001MA, SM624340003MA, SM624340004MA, SM624360008MA, SM624360009MA, SM623250002MA, SM624390004MA, SM623050002MA, SM623050007MA, SM623050008MA, SM623200013MA, SM624370012MA, SM623480026MA, SM623480027MA, SM623480028MA, SM623480029MA, SM623480030MA, SM623480031MA, SM623360005MA, SM624100001MA, SM624100002MA, SM624090010MA, SM623440019MA, SM623440020MA, SM623440021MA, SM623440022MA, SM623440023MA, SM623440024MA, SM623440025MA, SM623440026MA, SM623440027MA, SM623440028MA, SM623440029MA, SM623440030MA, SM623440031MA, SM623440032MA, SM623440033MA, SM623440034MA, SM623440035MA, SM624360001MA, SM624360002MA, SM624360003MA, SM624360004MA, SM624360007MA, SM625040001MA, SM625030006MA, SM624460001MA, SM624110004MA, SM624110005MA, SM624110006MA, SM624110007MA, SM623030003MA, SM624360005MA, SM624360006MA, SM624330001MA, SM624330003MA, SM624510014MA, SM624510015MA, SM623300012MA, SM623300013MA, SM623300014MA, SM623300015MA, SM623390004MA, SM624360010MA, SM624360011MA, SM624360012MA, SM624360013MA, SM624360014MA, SM623370017MA, SM624090005MA, SM624090006MA, SM624090007MA, SM623430001MA, SM623430002MA, SM624380014MA, SM624380015MA, SM624380016MA, SM624380017MA, SM623100003MA, SM623100009MA, SM623100010MA, SM623100011MA, SM623100012MA, SM623190016MA, SM623190017MA, SM623190018MA, SM623190019MA, SM624100015MA, SM624210001MA, SM624250001MA, SM624280001MA, SM623180014MA, SM623180015MA, SM623180016MA, SM623180001MA, SM623180002MA, SM624200014MA, SM624200016MA, SM624370007MA, SM624370009MA, SM624370010MA, SM624460003MA, SM623030005MA, SM623030004MA, SM623120027MA, SM623120028MA, SM623440003MA, SM623440004MA, SM624220023MA, SM624220024MA, SM623110014MA, SM623110016MA, SM623110018MA, SM623110019MA, SM623160015MA, SM623160016MA, SM623350009MA, SM623350010MA, SM624070001MA, SM624070002MA, SM624200018MA, SM624230001MA, SM623340042MA, SM623340043MA, SM623340044MA, SM623340045MA, SM623340046MA, SM623340047MA, SM623340048MA, SM623100014MA, SM623230001MA, SM623280021MA, SM623280022MA, SM623280023MA, SM623080005MA, SM623080006MA, SM624160009MA, SM624160010MA, SM623110007MA, SM623110008MA, SM623110009MA, SM623100002MA, SM623060016MA, SM623070007MA, SM623070008MA, SM623070009MA, SM623070010MA, SM623070011MA, SM623070012MA, SM623070013MA, SM623070014MA, SM623070015MA, SM623070016MA, SM623070017MA, SM623090001MA, SM623100001MA, SM623110010MA, SM623110011MA, SM623110012MA, SM624470001MA, SM623320002MA, SM623320021MA, SM623050012MA, SM623120024MA, SM623100013MA, SM623120023MA, SM623120025MA, SM625030007MA, SM623050001MA, SM624060011MA, SM624060012MA, SM624060013MA, SM624060015MA, SM624060016MA, SM623320005MA, SM623280018MA, SM623230007MA, SM623230008MA, SM623230009MA, SM623230010MA, SM623230011MA, SM623230012MA, SM623230013MA, SM623230014MA, SM623230015MA, SM623230016MA, SM623230017MA, SM623170014MA, SM623170015MA, SM623170016MA, SM623170017MA, SM623170018MA, SM623170019MA, SM623320004MA, SM624410010MA, SM624410011MA, SM624090001MA, SM624160001MA, SM624380007MA, SM624060014MA, SM623310031MA, SM623310032MA, SM623400001MA, SM624310011MA, SM624340005MA, SM624240001MA, SM624240002MA, SM624240003MA, SM624240004MA, SM624240005MA, SM624130014MA, SM624370008MA, SM623200001MA, SM623110004MA, SM624200003WA, SM624200004WA, SM624040003WA, SM623160020WA, SM623150008WA, SM623160017WA, SM623160018WA, SM624030002WA, SM624030003WA, SM624010001WA, SM624010005WA, SM624290002WA, SM624290003WA, SM624290004WA, SM624290005WA, SM624290006WA, SM624120002WA, SM624120003WA, SM624120004WA, SM624120005WA, SM624120008WA, SM624120009WA, SM624120010WA, SM624120012WA, SM624030011WA, SM624030012WA, SM623040009WA, SM624120001WA, SM624120006WA, SM624120007WA, SM624120011WA, SM624060018WA, SM624060019WA, SM624060020WA, SM624060021WA, SM624060022WA, SM624060023WA, SM624060024WA, SM623140015WA, SM623040008WA, SM623050048WA, SM623050049WA, SM623050050WA, SM623080016WA, SM623080017WA, SM623280025WA, SM623280026WA, SM623280027WA, SM623280028WA, SM623280029WA, SM623220008WA, SM623220007WA, SM623220009WA, SM623150013WA, SM623150009WA, SM623150010WA, SM623150011WA, SM623150012WA, SM623160019WA, SM624050001WA, SM624050002WA, SM624050003WA, SM624050004WA, SM623280039WA, SM623280040WA, SM623280041WA, SM623290042WA, SM623290043WA, SM623290044WA, SM623290045WA, SM623290046WA, SM623290047WA, SM623050046WA, SM623050047WA, SM623100016WA, SM623100017WA, SM623220018WA, SM623020033WA, SM623280036WA, SM623370003WA, SM624200003MA, SM624200004MA, SM624200005MA, SM624200006MA, SM624200007MA, SM624130016MA, SM624130017MA, SM624130018MA, SM624040019MA, SM624040020MA, SM624040021MA, SM624040022MA, SM623240016MA, SM623240017MA, SM623440007MA, SM623440008MA, SM623440009MA, SM623310004WA, SM623490007WA, SM623490013WA, SM623490021WA, SM623490002WA, SM623490014WA, SM623490019WA, SM623310008WA, SM623160007WA, SM623160008WA, SM623310009WA, SM623310023WA, SM623440011WA, SM623490008WA, SM623490020WA, SM623310022WA, SM623490009WA, SM623490018WA, SM623310007WA, SM623490003WA, SM623310016WA, SM623310029WA, SM623350006WA, SM623350012WA, SM623350013WA, SM623450054WA, SM623220003WA, SM623220004WA, SM623220005WA, SM623420005WA, SM623490004WA, SM623490011WA, SM623490016WA, SM623310019WA, SM623310020WA, SM623310026WA, SM623420004WA, SM623420007WA, SM623420002WA, SM623230006WA, SM623350025WA, SM623350026WA, SM623350027WA, SM623420003WA, SM623420006WA, SM623370001WA, SM623370002WA, SM623440007WA, SM623440008WA, SM623440009WA, SM623440010WA, SM623440014WA, SM623310006WA, SM623310010WA, SM623310024WA, SM623160003WA, SM623160004WA, SM623160005WA, SM623310018WA, SM623490006WA, SM623490010WA, SM623490012WA, SM623490022WA, SM623310001WA, SM623310002WA, SM623310003WA, SM623310005WA, SM623310021WA, SM623310025WA, SM623490001WA, SM623490005WA, SM623490015WA, SM623160001WA, SM623160002WA, SM623160006WA, SM623490017WA, SM623310017WA, SM623310027WA, SM623310028WA, SM623500001MA, SM624370005MA, SM624370006MA, SM624320007MA, SM623140017MA, SM624280002MA, SM623280016MA, SM623500002MA, SM623390001MA, SM623240018MA, SM623240019MA, SM623240020MA, SM623240021MA, SM623240022MA, SM623240023MA, SM623240024MA, SM623250006MA, SM623250007MA, SM623250008MA, SM623290012MA, SM624040001MA, SM624040002MA, SM624040003MA, SM624040004MA, SM624040005MA, SM623240007MA, SM623240008MA, SM623240009MA, SM623240010MA, SM623240011MA, SM623240012MA, SM623240013MA, SM623240014MA, SM623320022MA, SM623210005MA, SM623210012MA, SM623210014MA, SM623200014MA, SM623200015MA, SM623210003MA, SM623210004MA, SM623210006MA, SM623210010MA, SM623210011MA, SM623210013MA, SM623340027MA, SM623480002MA, SM624510004MA, SM624510005MA, SM624510006MA, SM624510007MA, SM623340037MA, SM623340038MA, SM623340039MA, SM623340040MA, SM623340041MA, SM623180025MA, SM623180026MA, SM623180027MA, SM623180028MA, SM623110005MA, SM624130007MA, SM624130008MA, SM624130009MA, SM624120011MA, SM624390005MA, SM624390006MA, SM624390007MA, SM624420001MA, SM624420002MA, SM624420003MA, SM624420004MA, SM624420005MA, SM624420006MA, SM624420007MA, SM624420008MA, SM624420009MA, SM624420010MA, SM624440018MA, SM623320014MA, SM625040004MA, SM623320017MA, SM624310014MA, SM623050003MA, SM623320015MA, SM623320013MA, SM623320016MA, SM623320018MA, SM623320019MA, SM623320020MA, SM624040008MA, SM624210003MA, SM624210004MA, SM624210005MA, SM624210006MA, SM623050005MA, SM623050006MA, SM623360002MA, SM624060007MA, SM624060008MA, SM623050015WA, SM623050016WA, SM623140043WA, SM623140044WA, SM623280031WA, SM623280034WA, SM623140049WA, SM623140058WA, SM624010014WA, SM624010016WA, SM624110003WA, SM624110006WA, SM623280010WA, SM623280011WA, SM623280035WA, SM623140052WA, SM624110001WA, SM624010013WA, SM623280032WA, SM623280013WA, SM623280012WA, SM623280033WA, SM624110005WA, SM623280030WA, SM623290018WA, SM623190014MA, SM624190025MA, SM624190026MA, SM624190027MA, SM624190028MA, SM624190029MA, SM623100004MA, SM623100005MA, SM623290013MA, SM623290015MA, SM623290016MA, SM623440015WA, SM623440016WA, SM623440017WA, SM623440018WA, SM623440019WA, SM624090005WA, SM624090006WA, SM624090007WA, SM624090008WA, SM624090009WA, SM623170020MA, SM623170021MA, SM623170022MA, SM623170023MA, SM623180017MA, SM623180018MA, SM623180019MA, SM623180020MA, SM624100007MA, SM624100008MA, SM624100009MA, SM624100010MA, SM623060024MA, SM623060025MA, SM623060026MA, SM623050022MA, SM623050023MA, SM623050024MA, SM623050025MA, SM623050026MA, SM623050027MA, SM623050028MA, SM623050029MA, SM623050030MA, SM623050031MA, SM623050032MA, SM623050033MA, SM623050034MA, SM623050035MA, SM623050036MA, SM623050037MA, SM623050038MA, SM623050039MA, SM623050040MA, SM623050041MA, SM623050042MA, SM623050043MA, SM623050044MA, SM623050045MA, SM624010031MA, SM624010032MA, SM623140045WA, SM623140046WA, SM623140047WA, SM623140048WA, SM623140050WA, SM623140051WA, SM623140053WA, SM623140054WA, SM623140055WA, SM623140056WA, SM623140057WA, SM623140059WA, SM623140060WA, SM624010011WA, SM624010012WA, SM624010015WA, SM624110002WA, SM624110004WA, SM624140012WA, SM624140013WA, SM624150004WA, SM624150005WA, SM624150006WA, SM624150007WA, SM624100025WA, SM624100026WA, SM624020035WA, SM624020036WA, SM624020037WA, SM624020038WA, SM623140002WA, SM623230003WA, SM623230004WA, SM623230005WA, SM623230018WA, SM623230019WA, SM623460055WA, SM623490023WA, SM623490024WA, SM623490025WA, SM623490028WA, SM624170017WA, SM624320004WA, SM624320005WA, SM624320006WA, SM624320007WA, SM624320008WA, SM624350001WA, SM624350002WA, SM624350003WA, SM624350004WA, SM624350005WA, SM624350006WA, SM624350007WA, SM624350008WA, SM624350009WA, SM624350010WA, SM624350011WA, SM625020001WA, SM625020002WA, SM625020003WA, SM625020004WA, SM625020005WA, SM625030001WA, SM625040002WA, SM624320019WA, SM624320020WA, SM624030005WA, SM624170006WA, SM623520001WA, SM623350031WA, SM623350032WA, SM623040012WA, SM623040013WA, SM623040014WA, SM623170004WA, SM623350037WA, SM623350038WA, SM623350039WA, SM623350040WA, SM623520002WA, SM623020025WA, SM623020030WA, SM623520005WA, SM624120016WA, SM624170008WA, SM624250004MA, SM624250005MA, SM624250007MA, SM623020027WA, SM624120018WA, SM623040011WA, SM624030004WA, SM624080022WA, SM624080023WA, SM624150010WA, SM623350034WA, SM623350035WA, SM623350036WA, SM624150012WA, SM624150011WA, SM624410008WA, SM624030006WA, SM624030007WA, SM624030008WA, SM624080027WA, SM624250006MA, SM624500002WA, SM624500003WA, SM624500004WA, SM624500005WA, SM625050004MA, SM624170005WA, SM624410006WA, SM624410007WA, SM623430005MA, SM623430006MA, SM623430007MA, SM623430008MA, SM623430009MA, SM623430010MA, SM624080025WA, SM623040015WA, SM623170001WA, SM623170002WA, SM623170003WA, SM623170005WA, SM623170006WA, SM623020028WA, SM623020029WA, SM624150009WA, SM623040010WA, SM624080024WA, SM623520003WA, SM624030009WA, SM624120015WA, SM624410009WA, SM624150008WA, SM624170004WA, SM624080026WA, SM624080028WA, SM624120014WA, SM624410005WA, SM623350033WA, SM624500001WA, SM623020026WA, SM624080029WA, SM624170007WA, SM623520004WA, SM623460036MA, SM623460037MA, SM623460038MA, SM623460039MA, SM623460040MA, SM623460041MA, SM623460042MA, SM623460043MA, SM623460044MA, SM623460045MA, SM623460046MA, SM623460047MA, SM623460048MA, SM623460049MA, SM623460050MA, SM624140004MA, SM624140005MA, SM624140006MA, SM624140007MA, SM624140008MA, SM624140009MA, SM624140010MA, SM624140011MA, SM624140012MA, SM624140013MA, SM624140014MA, SM624140015MA, SM624140016MA, SM624140017MA, SM624150007MA, SM624150008MA, SM624150009MA, SM624150010MA, SM624150011MA, SM624150012MA, SM624150013MA, SM624150014MA, SM624220006MA, SM624220007MA, SM624220008MA, SM624220009MA, SM624220010MA, SM624220011MA, SM624220012MA, SM624220013MA, SM624020001MA, SM624020002MA, SM624020003MA, SM624020004MA, SM624020005MA, SM624020006MA, SM624020007MA, SM624020008MA, SM624020009MA, SM624020021MA, SM624020022MA, SM624020023MA, SM624020024MA, SM624110020MA, SM624110021MA, SM624110022MA, SM624110023MA, SM624240006MA, SM624240007MA, SM624240008MA, SM624240009MA, SM624240010MA, SM624240011MA, SM624240012MA, SM624240013MA, SM623010001WA, SM624210004WA, SM624200002WA, SM623180010WA, SM623150006WA, SM623350002WA, SM623350003WA, SM623350004WA, SM623190021WA, SM624110012WA, SM624110013WA, SM624110014WA, SM624110015WA, SM624110016WA, SM624110017WA, SM624110018WA, SM624110019WA, SM624110020WA, SM624200001WA, SM624140003WA, SM623430024WA, SM623440005WA, SM623110020WA, SM623110021WA, SM624010007WA, SM623350029WA, SM623440030WA, SM624040006WA, SM624040007WA, SM624040008WA, SM623370004WA, SM624010008WA, SM623080012WA, SM623020024WA, SM624100027WA, SM624100030WA, SM624030016WA, SM623050018WA, SM623470068WA, SM623220001WA, SM623180006WA, SM623330001WA, SM624080019WA, SM624080016WA, SM624080017WA, SM623010010WA, SM623060029WA, SM624100024WA, SM623470062WA, SM623470063WA, SM624090001WA, SM624110007WA, SM624110008WA, SM623380001WA, SM623020031WA, SM623310011WA, SM624100031WA, SM624080033WA, SM623180007WA, SM623470064WA, SM623470065WA, SM623170013WA, SM623290051WA, SM623060034WA, SM724060037WA, SM623350028WA, SM623180029WA, SM623440032WA, SM623350020WA, SM623180008WA, SM623440033WA, SM624080001WA, SM624080002WA, SM623360043WA, SM623050017WA, SM623360042WA, SM624010009WA, SM623480004WA, SM623480005WA, SM624030015WA, SM623050021WA, SM623360049WA, SM623280017WA, SM624100028WA, SM623070004WA, SM623460060WA, SM623070002WA, SM623440012WA, SM623430021WA, SM623290007WA, SM624050005WA, SM623350007WA, SM624020002WA, SM623180033WA, SM623430022WA, SM623450034WA, SM624010010WA, SM623220002WA, SM623350021WA, SM623010009WA, SM624060002WA, SM624100033WA, SM624100032WA, SM623210007WA, SM623210008WA, SM623210009WA, SM623350030WA, SM623330002WA, SM623170011WA, SM623440004WA, SM623060013WA, SM623150007WA, SM623440006WA, SM623290049WA, SM623190006WA, SM623360001WA, SM623080003WA, SM623220016WA, SM623220017WA, SM623180030WA, SM623450053WA, SM623020032WA, SM623290052WA, SM623360048WA, SM624070001WA, SM624080035WA, SM624080036WA, SM623160014WA, SM623160015WA, SM624040005WA, SM623280002WA, SM624050008WA, SM624050009WA, SM623160023WA, SM623370005WA, SM623300001WA, SM623300002WA, SM623300003WA, SM623300004WA, SM624100034WA, SM623060028WA, SM624050006WA, SM624080020WA, SM624080021WA, SM623400001WA, SM623290050WA, SM623490027WA, SM623170012WA, SM623010011WA, SM624080018WA, SM623510001WA, SM623350005WA, SM624080034WA, SM623070003WA, SM623470069WA, SM623080007WA, SM623180003WA, SM624100029WA, SM623160027WA, SM623460057WA, SM623460058WA, SM623060030WA, SM623450045WA, SM624010017WA, SM623400002WA, SM623350022WA, SM623350023WA, SM623260013WA, SM623260014WA, SM623210023WA, SM623230020WA, SM623010006WA, SM623010007WA, SM624110011WA, SM623350024WA, SM623180013WA, SM624110010WA, SM623010008WA, SM623380005MA, SM623430016MA, SM623430019MA, SM623430020MA, SM623430021MA, SM623430022MA, SM623430023MA, SM623430024MA, SM623440013MA, SM623440014MA, SM624060005MA, SM624060006MA, SM624060009MA, SM624060010MA, SM624110024MA, SM624110025MA, SM624110026MA, SM624110027MA, SM624110028MA, SM624130004MA, SM624130005MA, SM624130006MA, SM623440029WA, SM623400006WA, SM624320017WA, SM624320018WA, SM624220001WA, SM624220002WA, SM624270019WA, SM624270020WA, SM624270021WA, SM624290001WA, SM623460059WA, SM624090003WA, SM623290053WA, SM623350017WA, SM623410001WA, SM623420001WA, SM624030013WA, SM624270023WA, SM623440031WA, SM624160002WA, SM624090004WA, SM624270001WA, SM624270002WA, SM624080014WA, SM624080015WA, SM623180032WA, SM623190013MA, SM623350041WA, SM623460061WA, SM624080013WA, SM624150002WA, SM624170003WA, SM623030001WA, SM623030002WA, SM624020005WA, SM624020006WA, SM624030014WA, SM624050011WA, SM624290022WA, SM623180004WA, SM623180005WA, SM624290021WA, SM623310015WA, SM624270024WA, SM624270025WA, SM623150001MA, SM623150002MA, SM623150003MA, SM623150004MA, SM623310033MA, SM623400003WA, SM623400004WA, SM623400005WA, SM623060021MA, SM623130001MA, SM624440001WA, SM623420010MA, SM623420028MA, SM623420030MA, SM623420009MA, SM623400004MA, SM623400003MA, SM623400002MA, SM623420025MA, SM623420026MA, SM623420027MA, SM623420035MA, SM623420036MA, SM623420037MA, SM623420038MA, SM623420039MA, SM623420020MA, SM623420021MA, SM623420012MA, SM623420013MA, SM623420022MA, SM623420023MA, SM623420024MA, SM623420014MA, SM623420015MA, SM623420008MA, SM623420029MA, SM623420018MA, SM623420019MA, SM623420031MA, SM623420032MA, SM623420033MA, SM623420034MA, SM623400021MA, SM623400022MA, SM623400023MA, SM623400024MA, SM623400012MA, SM623400015MA, SM623400016MA, SM623400017MA, SM623500002WA, SM623500003WA, SM623400005MA, SM623400006MA, SM623400013MA, SM623400014MA, SM623400011MA, SM623420016MA, SM623420017MA, SM623400007MA, SM623400008MA, SM623400009MA, SM623400010MA, SM623400018MA, SM623400019MA, SM623400020MA, SM623160024WA, SM623160025WA, SM623160026WA, SM623440020WA, SM624140001WA, SM624140002WA, SM624340002WA, SM623260005MA, SM623260003MA, SM623260004MA, SM623260006MA, SM624060004MA, SM623260002MA, SM623260007MA, SM624250001WA, SM623210024WA, SM623210025WA, SM623480001MA, SM623050019WA, SM623050020WA, SM623470070WA, SM623470071WA, SM624150002MA, SM624150003MA, SM624150004MA, SM624150005MA, SM624150006MA, SM623210015MA, SM623120018MA, SM623130003MA, SM623130004MA, SM623130005MA, SM624010021MA, SM624010022MA, SM624010023MA, SM624010024MA, SM624220001MA, SM623450046WA, SM623450047WA, SM623450048WA, SM623060014MA, SM623380006MA, SM624040023MA, SM624080030WA, SM624080031WA, SM624080032WA, SM623180031WA, SM623190039WA, SM623190034WA, SM623190035WA, SM623190036WA, SM623190037WA, SM623190038WA, SM623350008WA, SM623350011WA, SM623350016WA, SM623350018WA, SM623350019WA, SM624090010WA, SM624090011WA, SM624090012WA, SM624090013WA, SM624090014WA, SM624100021WA, SM624100022WA, SM624100023WA, SM624210001WA, SM624330002WA, SM625040001WA, SM625070001WA, SM625070002WA, SM623220010WA, SM623220011WA, SM624220003WA, SM623500001WA, SM624190001WA, SM623280024WA, SM623280014WA, SM624020003WA, SM624050007WA, SM624030001WA, SM624020039WA, SM624360002WA, SM624250013MA, SM625030003MA, SM625030004MA, SM625030005MA, SM623120021MA, SM623120022MA, SM623280019MA, SM623280020MA, SM623280024MA, SM623280025MA, SM623280026MA, SM623280027MA, SM623280028MA, SM624050001MA, SM624050002MA, SM623180009WA, SM623370006WA, SM624200005WA, SM623020023WA, SM623080001WA, SM623080002WA, SM624320016WA, SM624320014WA, SM624320015WA, SM623270015WA, SM623450035WA, SM623450036WA, SM623140042WA, SM623270001WA, SM623270002WA, SM623270004WA, SM623280003WA, SM623280009WA, SM623450050WA, SM623450051WA, SM623450052WA, SM624270011WA, SM624270008WA, SM624140022WA, SM624140033WA, SM623190001WA, SM624050015WA, SM624270014WA, SM623010012WA, SM623420014WA, SM623070037WA, SM623080008WA, SM623080011WA, SM623430006WA, SM624140036WA, SM623080014WA, SM623080015WA, SM624320001WA, SM624320002WA, SM624320003WA, SM623270014WA, SM624370001WA, SM623020002WA, SM623270016WA, SM624050012WA, SM623410004WA, SM623430007WA, SM623270006WA, SM623080013WA, SM624140019WA, SM623070040WA, SM623270003WA, SM624270007WA, SM624140026WA, SM624140018WA, SM623020001WA, SM623070036WA, SM623070035WA, SM624060008WA, SM624060012WA, SM624060013WA, SM624060014WA, SM624060015WA, SM624060016WA, SM624060017WA, SM624150016WA, SM624140024WA, SM624140037WA, SM623420008WA, SM624140016WA, SM624090002WA, SM623280008WA, SM624060005WA, SM623290048WA, SM624440002WA, SM624440003WA, SM624440004WA, SM624440005WA, SM624440006WA, SM623010014WA, SM623010015WA, SM623070039WA, SM623270005WA, SM623270007WA, SM623270020WA, SM623420010WA, SM623420012WA, SM623420015WA, SM623430010WA, SM623430011WA, SM623430012WA, SM623450049WA, SM624140020WA, SM624270017WA, SM624290018WA, SM624290019WA, SM624370002WA, SM624370003WA, SM624370004WA, SM624060011WA, SM624330001WA, SM624140004WA, SM624140005WA, SM624140006WA, SM624140007WA, SM624140008WA, SM624140009WA, SM624140010WA, SM624140011WA, SM623440002WA, SM623440003WA, SM623270009WA, SM623270010WA, SM623270011WA, SM624060007WA, SM624170009WA, SM624170011WA, SM624140025WA, SM624320009WA, SM624320010WA, SM623270012WA, SM624140015WA, SM624140023WA, SM623070038WA, SM624270005WA, SM624270006WA, SM623430005WA, SM623430008WA, SM624340001WA, SM624220004WA, SM623180011WA, SM624280001WA, SM624280002WA, SM623280007WA, SM623430002WA, SM624060009WA, SM623020003WA, SM624060010WA, SM623190009WA, SM623190011WA, SM623190012WA, SM623070041WA, SM623280005WA, SM624170010WA, SM623270008WA, SM623270017WA, SM623420011WA, SM623430009WA, SM624140017WA, SM624140021WA, SM624140027WA, SM624140028WA, SM624140034WA, SM624140035WA, SM624150001WA, SM624150014WA, SM624150017WA, SM624270009WA, SM624270010WA, SM624270012WA, SM624270013WA, SM624270015WA, SM624270018WA, SM624290008WA, SM624290012WA, SM624290013WA, SM624290014WA, SM624290015WA, SM624290016WA, SM624290017WA, SM624170015WA, SM624170016WA, SM623430001WA, SM623430003WA, SM623410006WA, SM624150013WA, SM623190010WA, SM623010016WA, SM624050013WA, SM624050014WA, SM624050016WA, SM624060003WA, SM624060004WA, SM624060006WA, SM623010005WA, SM624290011WA, SM623080009WA, SM624410001WA, SM624410002WA, SM623080010WA, SM624270016WA, SM624170012WA, SM623020006WA, SM623020007WA, SM624150015WA, SM624290020WA, SM623020018WA, SM623020019WA, SM623020020WA, SM623020021WA, SM623020022WA, SM623450037WA, SM623450038WA, SM623450039WA, SM623450040WA, SM623450041WA, SM623450042WA, SM623450043WA, SM623450044WA, SM623290008WA, SM623290009WA, SM623010003WA, SM623010004WA, SM623270013WA, SM623410003WA, SM623190008WA, SM623420009WA, SM624140032WA, SM623430004WA, SM623010002WA, SM623410002WA, SM623420013WA, SM623410005WA, SM623280006WA, SM624040001WA, SM624040002WA, SM623020004WA, SM623020005WA, SM623010013WA, SM624290007WA, SM624290009WA, SM624290010WA, SM623180012WA, SM623290010WA, SM624320011WA, SM624320012WA, SM624320013WA, SM624140030WA, SM623280004WA, SM623270018WA, SM623270019WA, SM623010017WA, SM624140029WA, SM624140031WA, SM624170013WA, SM624170014WA, SM623060007WA, SM623060008WA, SM623060052WA, SM623060053WA, SM623060054WA, SM623270021WA, SM623270022WA, SM624140014WA, SM623280001WA, SM624020040WA, SM623440013WA, SM623480006WA, SM623190007WA, SM624270022WA, SM623140003WA, SM623140007WA, SM623140008WA, SM623140009WA, SM623430023WA, SM623230021WA, SM623230022WA, SM623290011WA, SM624030010WA, SM623190020WA, SM623280015WA, SM623050051WA, SM623080018WA, SM623040007WA, SM623280023WA, SM623350001WA, SM624270004WA, SM623470072WA, SM623480001WA, SM623480003WA, SM624360001WA, SM624490001WA, SM623480002WA, SM623190002WA, SM623190003WA, SM623190004WA, SM623190005WA, SM623280037WA, SM623280038WA, SM624040004WA, SM624210002WA, SM624210003WA, SM624220005WA, SM624010025MA, SM624320001MA, SM623500011MA, SM623340005MA, SM623340006MA, SM623340015MA, SM623340021MA, SM623340022MA, SM624170020MA, SM623120016MA, SM623120017MA, SM623320007MA, SM624020017MA, SM624060003MA, SM623060001MA, SM624280006MA, SM624020018MA, SM624190032MA, SM624190033MA, SM624190034MA, SM624050003MA, SM623500012MA, SM624020019MA, SM624040014MA, SM624040015MA, SM624320005MA, SM624320006MA, SM623130008MA, SM623340018MA, SM623130014MA, SM623500010MA, SM623480008MA, SM623370021MA, SM623370022MA, SM624310006MA, SM623240015MA, SM623240005MA, SM623060005MA, SM624010026MA, SM623340003MA, SM623340004MA, SM624120004MA, SM624020020MA, SM624170021MA, SM624290007MA, SM624290013MA, SM624290014MA, SM624290015MA, SM624290017MA, SM623060002MA, SM624290002MA, SM624040013MA, SM624290001MA, SM623020010MA, SM624110010MA, SM624290003MA, SM624120009MA, SM623130012MA, SM624320003MA, SM624120017MA, SM623020008MA, SM623130009MA, SM623130010MA, SM623130011MA, SM623130020MA, SM623130021MA, SM623240003MA, SM623240004MA, SM623250005MA, SM623330003MA, SM623340001MA, SM623340002MA, SM623340007MA, SM623340008MA, SM623340009MA, SM623340010MA, SM623340014MA, SM623340016MA, SM623340019MA, SM623340020MA, SM623340025MA, SM623370008MA, SM623370018MA, SM623430014MA, SM623430015MA, SM623440001MA, SM623440002MA, SM623440010MA, SM623480009MA, SM623490001MA, SM623500013MA, SM624010018MA, SM624010019MA, SM624040009MA, SM624040010MA, SM624050004MA, SM624110018MA, SM624110019MA, SM624120003MA, SM624120008MA, SM624120010MA, SM624160018MA, SM624290016MA, SM624310001MA, SM624380010MA, SM624060002MA, SM623100007MA, SM624120005MA, SM624290009MA, SM623100006MA, SM624280005MA, SM624120006MA, SM624290004MA, SM624310005MA, SM624310007MA, SM624290006MA, SM624300001MA, SM624300002MA, SM623130018MA, SM623130019MA, SM624120019MA, SM624160003MA, SM623060006MA, SM623130025MA, SM623130026MA, SM623130027MA, SM623130028MA, SM623130029MA, SM623130030MA, SM623130031MA, SM623130032MA, SM623130033MA, SM623130034MA, SM623130035MA, SM623130036MA, SM623130037MA, SM623130038MA, SM623130039MA, SM623130040MA, SM623130041MA, SM623340017MA, SM623130016MA, SM623340013MA, SM623060004MA, SM624110008MA, SM624110009MA, SM624120007MA, SM624310002MA, SM624310003MA, SM623250003MA, SM623340011MA, SM623340012MA, SM624290008MA, SM623130006MA, SM623130024MA, SM624250012MA, SM624120018MA, SM623430011MA, SM623430012MA, SM623430013MA, SM623340026MA, SM624120016MA, SM623480003MA, SM623370007MA, SM623060003MA, SM623240006MA, SM623480004MA, SM624380008MA, SM624380009MA, SM623240002MA, SM623250004MA, SM624010020MA, SM623500009MA, SM624320004MA, SM624250008MA, SM624250009MA, SM624250010MA, SM624250011MA, SM623120015MA, SM623420002MA, SM623420003MA, SM623420004MA, SM623420005MA, SM623420006MA, SM623420007MA, SM624040011MA, SM624040012MA, SM623050009MA, SM623050010MA, SM623120001MA, SM623120002MA, SM623120003MA, SM623120004MA, SM623120005MA, SM623120006MA, SM623120007MA, SM623120008MA, SM623120009MA, SM623120010MA, SM623120011MA, SM623120012MA, SM623120013MA, SM623120014MA, SM623370009MA, SM623370011MA, SM623370006MA, SM623130022MA, SM623130023MA, SM624110016MA, SM624110017MA, SM623130015MA, SM624010027MA, SM624010028MA, SM624010029MA, SM624010033MA, SM624010034MA, SM624110002MA, SM624110003MA, SM624290011MA, SM624290012MA, SM624290005MA, SM624160017MA, SM624310008MA, SM624120013MA, SM624160016MA, SM623100008MA, SM623310034MA, SM623310035MA, SM623340032MA, SM623340033MA, SM623340034MA, SM623340035MA, SM623340036MA, SM623130007MA, SM623130013MA, SM623120019MA, SM623120020MA, SM623130002MA, SM624310013MA, SM623340028MA, SM623340029MA, SM623340030MA, SM623340031MA, SM623370004MA, SM623370005MA, SM623440011MA, SM623440012MA, SM624360015MA, SM623370010MA, SM623370012MA, SM623370013MA, SM623370014MA, SM623370015MA, SM623370016MA, SM623020009MA, SM624120002MA, SM624290010MA, SM624320002MA, SM623480007MA, SM624160004MA, SM623240001MA, SM623430013WA, SM623060011WA, SM623160022WA, SM623290001WA, SM623290002WA, SM623290003WA, SM623350014WA, SM623350015WA, SM623430014WA, SM623430015WA, SM623430016WA, SM623430017WA, SM623430018WA, SM623430019WA, SM623430020WA, SM624420001WA, SM624420002WA, SM624420003WA, SM623200018WA, SM623200021WA, SM623200022WA, SM623470066WA, SM623470067WA, SM624080003WA, SM624080004WA, SM624080005WA, SM624080006WA, SM624080007WA, SM624080008WA, SM624080009WA, SM624080010WA, SM624080011WA, SM624080012WA, SM623060012WA, SM623200019WA, SM623200020WA, SM623060009WA, SM623060010WA, SM623290004WA, SM623290005WA, SM623290006WA, SM623160021WA, SM623330027WA, SM623330028WA, SM623330029WA, SM623330030WA, SM623070018WA, SM623070019WA, SM623070020WA, SM623070021WA, SM623070022WA, SM623070023WA, SM623070024WA, SM623290019WA, SM623290020WA, SM623290021WA, SM623290022WA, SM623290023WA, SM624050010WA, SM624120014MA, SM624120015MA, SM624110011MA, SM624110012MA, SM623060033MA, SM623360044WA, SM623360045WA, SM623360046WA, SM623360047WA, SM624320021WA, SM624320022WA, SM624320023WA, SM624320024WA, SM624320025WA, SM624320026WA, SM624230001WA, SM624230002WA, SM624230003WA, SM624230004WA, SM624230005WA, SM624230006WA, SM624230007WA, SM623440001WA, SM623050004MA, SM623130017MA, SM623310012WA, SM623310013WA, SM623310014WA, SM623320008MA, SM623320009MA, SM623320025MA, SM623320026MA, SM623370002MA, SM623370003MA, SM623440017MA, SM623440018MA, SM623440021WA, SM623440022WA, SM623440023WA, SM623440024WA, SM623440025WA, SM623440026WA, SM623440027WA, SM623440028WA, SM623460051MA, SM623460052MA, SM623460053MA, SM623460056WA, SM623470001MA, SM623470002MA, SM623470003MA, SM623470004MA, SM623470005MA, SM623470006MA, SM623480010MA, SM623480011MA, SM623480012MA, SM623480013MA, SM623500008MA, SM624060026WA, SM624100004MA, SM624100005MA, SM624100006MA, SM624130010MA, SM624130011MA, SM624130012MA, SM624130013MA, SM624160001WA, SM624060025WA, SM624100003MA, SM624450001WA, SM624450002WA, SM624450003WA, SM624060027WA, SM624120020MA, SM624120021MA, SM624120022MA, SM624120023MA, SM624120024MA, SM624120025MA, SM624130002MA, SM624130003MA, SM624410005MA, SM624410006MA, SM624410007MA, SM624410008MA, SM624410009MA, SM624440006MA, SM624440007MA, SM624440008MA, SM624440009MA, SM624440010MA, SM624440011MA, SM624440012MA, SM624440013MA, SM624440014MA, SM624440015MA, SM624440016MA, SM624440017MA, SM619050021MA, SM619060029MA, SM619400017MA, SM619400018MA, SM619400019MA, SM619400020MA, SM619400021MA, SM619400030MA, SM619400032MA, SM619400033MA, SM619400034MA, SM619400054MA, SM619400055MA, SM619460006MA, SM620120048MA, SM620120049MA, SM620120050MA, SM620120051MA, SM620120052MA, SM620120054MA, SM620120056MA, SM620120057MA, SM620120058MA, SM620120059MA, SM620120060MA, SM620120061MA, SM620180017MA, SM620180018MA, SM620180019MA, SM620180020MA, SM620180021MA, SM620180022MA, SM620180023MA, SM620180024MA, SM620180025MA, SM620180026MA, SM620180027MA, SM620180028MA, SM620180029MA, SM620180030MA, SM620180031MA, SM620180032MA, SM620180034MA, SM620180035MA, SM621030011MA. |
| FEI Number |
3003513129
|
Recalling Firm/ Manufacturer |
GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No. 19 Changjiang Road; Xin Wuxi China
|
| For Additional Information Contact | GE HealthCare Service 800-437-1171 |
Manufacturer Reason for Recall | Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation. |
FDA Determined Cause 2 | Device Design |
| Action | GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice on 03/21/2025. The notice explained the safety issue with the device, potential risk, and requested the following actions:
The anesthesia system may continue to be used following these instructions:
Perform the Ventilation Screening Test specified in Appendix 1 for each affected Carestation system.
If the Carestation system passes the Ventilation Screening Test, you can continue to use the device.
If the Carestation system fails the Ventilation Screening Test and must be used prior to the system being corrected by GE HealthCare, follow the instructions below:
1. Use the device in accordance with the instructions in the URM with these changes:
i. Use only Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) mode to mechanically ventilate a patient.
ii. Do not use Volume Control Ventilation (VCV) mode to mechanically ventilate a patient.
NOTE: Manual mode of the anesthesia system can be used to provide manual ventilation or allow spontaneous ventilation of the patient.
2. Users are not to use VCV mode on the device until the device has been corrected by GE HealthCare.
Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions and retain this document for your record.
Questions: contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Notice sent on 04/21/25 with updated instructions. Change: Attach a 2 Liter reservoir bag (original) to Attach a reservoir bag (updated)in step 5 of the Ventilation Screening Test in Appendix 1 of the letter.
Notice sent on 5/18/26 added additional affected devices for GE HealthCare Urgent Medical Device Correction notification, Field Replaceable Unit (FRU) parts: DRIVER GAS CHECK VALVE ASSY, M1807901-S; Valve Assembly, 3.5cm H2O Drive Gas Check Valve BCG, 2071125-001-S; A180 Ventilator Engine Assembly, 2071117-001-S. |
| Quantity in Commerce | 2298 units |
| Distribution | Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = BSZ
|
|
|
|