• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall GE Healthcare CARESTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall GE Healthcare CARESTATIONsee related information
Date Initiated by FirmMarch 21, 2025
Date PostedApril 29, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1628-2025
Recall Event ID 96634
510(K)NumberK151570 
Product Classification Gas-machine, anesthesia - Product Code BSZ
ProductGE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000. Anesthesia system.
Code Information GTIN 00840682103947, Serial Numbers: SM724310010MA, SM723370006MA, SM723080015MA, SM724440001MA, SM723370003MA, SM723240009MA, SM723360010MA, SM723360011MA, SM723360012MA, SM723360013MA, SM723360014MA, SM723360015MA, SM723360016MA, SM723360017MA, SM724100019MA, SM724500002MA, SM724500003MA, SM724500004MA, SM724500005MA, SM723370008MA, SM724140005MA, SM723370002MA, SM723370004MA, SM723370005MA, SM723370007MA, SM723020044MA, SM723020045MA, SM723020046MA, SM723050009MA, SM723050010MA, SM723370001MA, SM725060005MA, SM724480001MA, SM724390036MA, SM724390037MA, SM724150044MA, SM724300003MA, SM724300004MA, SM724390010MA, SM725040006MA, SM725040007MA, SM725040008MA, SM724150023MA, SM724150024MA, SM723370009MA, SM724200001MA, SM724250009MA, SM725050005MA, SM724490002MA, SM723440001MA, SM723020034MA, SM723030003MA, SM723030004MA, SM723030032MA, SM724390006MA, SM724390007MA, SM724390008MA, SM723020029MA, SM723020030MA, SM724370011MA, SM724460002MA, SM723170023MA, SM723170024MA, SM723240007MA, SM723240008MA, SM723240010MA, SM723240011MA, SM723240012MA, SM723240013MA, SM725020001MA, SM724010037MA, SM724010038MA, SM724010039MA, SM724250029MA, SM724190144MA, SM724190145MA, SM724190146MA, SM724190147MA, SM724190148MA, SM724190149MA, SM723040004MA, SM723040005MA, SM723040006MA, SM723040007MA, SM723040008MA, SM724400004MA, SM724400005MA, SM724400006MA, SM724400007MA, SM724360001MA, SM724360002MA, SM723160001MA, SM724140024MA, SM724140025MA, SM724140026MA, SM724140027MA, SM724320016MA, SM723310001MA, SM723260014MA, SM724150025MA, SM724150026MA, SM724150027MA, SM723120001MA, SM724110014MA, SM723260011MA, SM724440002MA, SM724440003MA, SM723120012MA, SM723120013MA, SM723120014MA, SM723120015MA, SM723120016MA, SM723490002MA, SM723490003MA, SM723500021MA, SM723300024MA, SM723300025MA, SM723300026MA, SM723300027MA, SM723300028MA, SM723260008MA, SM723260009MA, SM723260010MA, SM724200017MA, SM724200018MA, SM724200019MA, SM723240006MA, SM724210001MA, SM723160016MA, SM724150048MA, SM724150049MA, SM723110014MA, SM723100027MA, SM723110013MA, SM724510001MA, SM724340008MA, SM724030027MA, SM724030029MA, SM724030030MA, SM723170009MA, SM723170010MA, SM724040001MA, SM724230001MA, SM724230002MA, SM724230003MA, SM724360017MA, SM724040015MA, SM724040016MA, SM724040017MA, SM724040018MA, SM724230011MA, SM724030028MA, SM724030031MA, SM724040010MA, SM724040011MA, SM724040012MA, SM724040013MA, SM724200007MA, SM724200008MA, SM724200009MA, SM723170017MA, SM723170018MA, SM723170019MA, SM723170020MA, SM723170021MA, SM723170022MA, SM724320001MA, SM724310001MA, SM724310002MA, SM724150050MA, SM724300001MA, SM723390003MA, SM723200007MA, SM723200009MA, SM723030035MA, SM723030036MA, SM724310003MA, SM723110029MA, SM724200003MA, SM724200004MA, SM724100001MA, SM724100002MA, SM723260013MA, SM724170094MA, SM724170095MA, SM723070030MA, SM723070031MA, SM723070032MA, SM723400001MA, SM723340012MA, SM723340027MA, SM723360005MA, SM723360006MA, SM723400002MA, SM723110003MA, SM723110004MA, SM723110040MA, SM724320011MA, SM724320012MA, SM724320013MA, SM724320014MA, SM724320015MA, SM723490006MA, SM723490007MA, SM723490008MA, SM723490009MA, SM723400003MA, SM723400004MA, SM723120011MA, SM724020022MA, SM724020023MA, SM724020024MA, SM724010056MA, SM724010057MA, SM724010058MA, SM724010059MA, SM724010060MA, SM724010061MA, SM723060030MA, SM723060031MA, SM724200002MA, SM723160017MA, SM723160018MA, SM723210020MA, SM723210021MA, SM723210022MA, SM723210023MA, SM723210024MA, SM723210026MA, SM723210027MA, SM723210028MA, SM723210029MA, SM723140050MA, SM723140051MA, SM723140052MA, SM723140053MA, SM723140054MA, SM723140055MA, SM724150043MA, SM723100002MA, SM723100003MA, SM723120035MA, SM723120036MA, SM724080056MA, SM723360007MA, SM723360008MA, SM723360009MA, SM724330003MA, SM723070077MA, SM723070078MA, SM723070079MA, SM723070080MA, SM723070081MA, SM723070082MA, SM723070083MA, SM723070084MA, SM723070085MA, SM723070086MA, SM723070087MA, SM723070088MA, SM724340006MA, SM724090029MA, SM723110048MA, SM723110049MA, SM724080055MA, SM724080057MA, SM724080058MA, SM724170098MA, SM724170099MA, SM724170100MA, SM723120007MA, SM724420001MA, SM723060007MA, SM723110019MA, SM723110020MA, SM723110021MA, SM723110022MA, SM723110023MA, SM723430005MA, SM723140040MA, SM723140041MA, SM723140042MA, SM723140043MA, SM723140044MA, SM725030009MA, SM725030010MA, SM725030011MA, SM725030012MA, SM723080010MA, SM723080011MA, SM723080012MA, SM724190122MA, SM724190123MA, SM724190124MA, SM724260019MA, SM724260020MA, SM724260021MA, SM724260022MA, SM723150017MA, SM723150018MA, SM723150019MA, SM723150020MA, SM723150021MA, SM723150022MA, SM723150023MA, SM723150024MA, SM723150025MA, SM723150026MA, SM723150027MA, SM723150028MA, SM723150029MA, SM723150030MA, SM723150031MA, SM723150032MA, SM723150033MA, SM723150034MA, SM723150035MA, SM723150036MA, SM723150037MA, SM723150038MA, SM723150039MA, SM723150040MA, SM723150041MA, SM723150042MA, SM723150043MA, SM723150044MA, SM724160057MA, SM724160058MA, SM723020035MA, SM723020036MA, SM723030008MA, SM723030009MA, SM723140017MA, SM723140018MA, SM723140019MA, SM723140020MA, SM723140021MA, SM724420002MA, SM724420003MA, SM723210058MA, SM723210059MA, SM723210060MA, SM723210061MA, SM723210062MA, SM723050011MA, SM723060027MA, SM725030024MA, SM725030025MA, SM723160006MA, SM723160007MA, SM723160008MA, SM723160009MA, SM723060058MA, SM723080016MA, SM724060008MA, SM723110034MA, SM723110035MA, SM723170006MA, SM723450049MA, SM723130015MA, SM723130016MA, SM723450040MA, SM723450041MA, SM723450042MA, SM723450043MA, SM723450044MA, SM723450045MA, SM723450046MA, SM723450047MA, SM723450048MA, SM723450050MA, SM723450051MA, SM723450052MA, SM723450053MA, SM723450054MA, SM723450055MA, SM724160073MA, SM724160074MA, SM724160075MA, SM724160076MA, SM724160077MA, SM724160078MA, SM724160079MA, SM724160080MA, SM724160081MA, SM724160082MA, SM724160083MA, SM724160084MA, SM724160085MA, SM724160086MA, SM724160087MA, SM724160088MA, SM724160089MA, SM724160090MA, SM724160091MA, SM724160092MA, SM724160093MA, SM724240024MA, SM724240025MA, SM724240026MA, SM724240027MA, SM724240028MA, SM724240029MA, SM724240030MA, SM723180047MA, SM724100024MA, SM723320001MA, SM724380018MA, SM723190030MA, SM723190031MA, SM723380019MA, SM723380020MA, SM724190125MA, SM724190126MA, SM724190127MA, SM724190128MA, SM724190129MA, SM724190130MA, SM724190131MA, SM724190132MA, SM724190133MA, SM724190134MA, SM724360016MA, SM724370010MA, SM723170004MA, SM723500009MA, SM724060009MA, SM724060010MA, SM724060011MA, SM724010052MA, SM723390002MA, SM724480002MA, SM724210006MA, SM724210010MA, SM724210012MA, SM724210015MA, SM724210016MA, SM724210019MA, SM724100023MA, SM724250028MA, SM724150047MA, SM724260023MA, SM724260024MA, SM724260025MA, SM724260026MA, SM724260027MA, SM724260028MA, SM724260029MA, SM724440004MA, SM724440005MA, SM724440006MA, SM724440007MA, SM725020012MA, SM724210005MA, SM724210007MA, SM724210008MA, SM724210009MA, SM724210011MA, SM724210013MA, SM724210017MA, SM724210014MA, SM724210018MA, SM723290011MA, SM723290012MA, SM724100027MA, SM724100026MA, SM724180102MA, SM723210010MA, SM723210011MA, SM723210012MA, SM723210013MA, SM723210014MA, SM723210015MA, SM723210016MA, SM723210017MA, SM723210018MA, SM723210019MA, SM723220024MA, SM723220025MA, SM723220026MA, SM723220027MA, SM723220028MA, SM724230004MA, SM724230005MA, SM724230006MA, SM724230007MA, SM724230008MA, SM724360014MA, SM724400001MA, SM724400002MA, SM724400003MA, SM724090031MA, SM724090032MA, SM724200005MA, SM724200006MA, SM724200010MA, SM724200011MA, SM724200012MA, SM724200013MA, SM724320002MA, SM724320003MA, SM723030007MA, SM723030010MA, SM723030011MA, SM723030012MA, SM723030013MA, SM723020025MA, SM725030023MA, SM723260004MA, SM723260005MA, SM723390006MA, SM723060009MA, SM723060010MA, SM723060011MA, SM723060012MA, SM724240012MA, SM724240013MA, SM723230004MA, SM723290025MA, SM723290026MA, SM723290027MA, SM723290028MA, SM723340006MA, SM723340008MA, SM723340010MA, SM723360004MA, SM723450022MA, SM723450023MA, SM724060004MA, SM724060005MA, SM724060022MA, SM724060023MA, SM724060024MA, SM724060026MA, SM723120002MA, SM723130008MA, SM723130009MA, SM723130010MA, SM723130011MA, SM723130013MA, SM723130014MA, SM724330001MA, SM724060027MA, SM724060028MA, SM724060025MA, SM723450028MA, SM723450029MA, SM723450030MA, SM723450031MA, SM723450032MA, SM723080017MA, SM723090011MA, SM723090012MA, SM723090013MA, SM723090014MA, SM723090015MA, SM723090016MA, SM723090017MA, SM723090018MA, SM723090019MA, SM723090021MA, SM724040008MA, SM724040009MA, SM724060030MA, SM724060031MA, SM724090033MA, SM724090034MA, SM724090035MA, SM724050021MA, SM724050022MA, SM724050023MA, SM723090020MA, SM723110005MA, SM723110047MA, SM724050008MA, SM724050009MA, SM724050010MA, SM724050011MA, SM725030020MA, SM725030021MA, SM725030022MA, SM723140045MA, SM723140046MA, SM723140047MA, SM723140048MA, SM723140049MA, SM723340021MA, SM723340024MA, SM723350017MA, SM723350018MA, SM723350020MA, SM724090036MA, SM724090037MA, SM724250010MA, SM724250011MA, SM724250012MA, SM724250013MA, SM724410007MA, SM724410008MA, SM724410009MA, SM723340020MA, SM723340022MA, SM723340023MA, SM723350014MA, SM723350015MA, SM723350016MA, SM723350019MA, SM723350021MA, SM724040006MA, SM724040007MA, SM723500001MA, SM723500002MA, SM723500003MA, SM723500004MA, SM723500005MA, SM723500006MA, SM723500007MA, SM723040022MA, SM725030005MA, SM725030006MA, SM725030007MA, SM725030008MA, SM724380001MA, SM724380002MA, SM724380003MA, SM724380004MA, SM724380005MA, SM724380006MA, SM724380007MA, SM724380008MA, SM724380009MA, SM724380010MA, SM724380011MA, SM724380012MA, SM724380013MA, SM724380014MA, SM724390009MA, SM723030002MA, SM724040002MA, SM724040003MA, SM724040004MA, SM724040005MA, SM723160002MA, SM723160003MA, SM723160004MA, SM723160005MA, SM723380018MA, SM723110050MA, SM723110051MA, SM723110052MA, SM723110053MA, SM723080013MA, SM723080014MA, SM725040009MA, SM723170011MA, SM723170012MA, SM723230002MA, SM724100052MA, SM724100053MA, SM724120002MA, SM724120003MA, SM724160070MA, SM724030025MA, SM724030026MA, SM724260018MA, SM725040010MA, SM725040011MA, SM725040012MA, SM723380010MA, SM723380011MA, SM723380012MA, SM723380013MA, SM723380014MA, SM723380015MA, SM723380016MA, SM723380017MA, SM723380026MA, SM723380027MA, SM723380028MA, SM723380029MA, SM723380030MA, SM723380031MA, SM723380032MA, SM723380033MA, SM723380034MA, SM723380035MA, SM724150022MA, SM723040002MA, SM723040003MA, SM725030013MA, SM725030014MA, SM725030015MA, SM725030016MA, SM725030017MA, SM725030018MA, SM725030019MA, SM723480002MA, SM724410001MA, SM724090030MA, SM723110024MA, SM723110025MA, SM723110026MA, SM723110027MA, SM723110028MA, SM724310012MA, SM724400008MA, SM724400009MA, SM723110012MA, SM723110036MA, SM723110037MA, SM723110038MA, SM723110039MA, SM723110041MA, SM723110042MA, SM723110043MA, SM723110044MA, SM723110045MA, SM723480008MA, SM723480009MA, SM723480010MA, SM723130002MA, SM723130003MA, SM723030037MA, SM723030038MA, SM723030039MA, SM724250026MA, SM724300008MA, SM724300009MA, SM724300010MA, SM723340025MA, SM723340026MA, SM723060059MA, SM723060060MA, SM723060061MA, SM723060062MA, SM723060028MA, SM723060029MA, SM723210063MA, SM723210064MA, SM724110007MA, SM724110008MA, SM724160054MA, SM723210044MA, SM723210045MA, SM724010042MA, SM724010041MA, SM724010043MA, SM724010044MA, SM724010045MA, SM724010046MA, SM724010047MA, SM724010048MA, SM724010049MA, SM724010050MA, SM723190032MA, SM723190033MA, SM723190034MA, SM723190035MA, SM723130001MA, SM724160061MA, SM724160062MA, SM724160063MA, SM724160064MA, SM724160065MA, SM724160066MA, SM724160067MA, SM724230034MA, SM723160020MA, SM723050012MA, SM723050013MA, SM723050014MA, SM723050015MA, SM723050016MA, SM723050017MA, SM724110040MA, SM724110041MA, SM724110042MA, SM724110043MA, SM724110044MA, SM724110045MA, SM723470001MA, SM723380002MA, SM723290020MA, SM723290021MA, SM723290022MA, SM723290023MA, SM723140037MA, SM723140038MA, SM723140039MA, SM724310011MA, SM723490015MA, SM723490016MA, SM723490017MA, SM723490018MA, SM723490019MA, SM723490020MA, SM723490021MA, SM723110007MA, SM723110008MA, SM723060057MA, SM723110033MA, SM723110063MA, SM724100003MA, SM724100004MA, SM724100005MA, SM724250027MA, SM723280066MA, SM723290016MA, SM723290017MA, SM723290018MA, SM723290019MA, SM723210001MA, SM724290031MA, SM724200015MA, SM724200016MA, SM724100025MA, SM723130004MA, SM723130005MA, SM724180103MA, SM724180104MA, SM724180105MA, SM724110056MA, SM724510002MA, SM724510003MA, SM724220002MA, SM724490001MA, SM724140007MA, SM723300014MA, SM723300015MA, SM723300016MA, SM723300017MA, SM723300018MA, SM724370003MA, SM724370004MA, SM723310017MA, SM723310018MA, SM723310019MA, SM723310020MA, SM723310021MA, SM723310022MA, SM723310023MA, SM723310024MA, SM723310025MA, SM723310026MA, SM723310027MA, SM724310004MA, SM724310005MA, SM724310006MA, SM724310007MA, SM724310008MA, SM724310009MA, SM723250006MA, SM723420001MA, SM724190135MA, SM724190136MA, SM723090040MA, SM723090041MA, SM723090042MA, SM723090043MA, SM723090044MA, SM723090045MA, SM723090046MA, SM723090047MA, SM723090048MA, SM723090049MA, SM723090050MA, SM723090051MA, SM723090052MA, SM723090053MA, SM723090054MA, SM723090055MA, SM723090056MA, SM723090057MA, SM723090058MA, SM723090059MA, SM723090060MA, SM723200010MA, SM723200011MA, SM723200012MA, SM723200013MA, SM723200014MA, SM723200015MA, SM723200016MA, SM723200017MA, SM723200018MA, SM723200019MA, SM723200020MA, SM723200021MA, SM723050003MA, SM723500008MA, SM724100022MA, SM723050001MA, SM723180052MA, SM723180053MA, SM723180054MA, SM723180055MA, SM723200008MA, SM724010036MA, SM723190027MA, SM724250001MA, SM724250002MA, SM724250003MA, SM723190025MA, SM723440003MA, SM723440004MA, SM724120019MA, SM724120020MA, SM724120021MA, SM724120022MA, SM724120023MA, SM724120024MA, SM724120025MA, SM724120026MA, SM724120027MA, SM724120028MA, SM724120029MA, SM724120030MA, SM724120031MA, SM724120032MA, SM724120033MA, SM724120034MA, SM724120035MA, SM724120036MA, SM724340001MA, SM724340002MA, SM724340003MA, SM724340004MA, SM724340005MA, SM723500010MA, SM723500011MA, SM723500012MA, SM723500013MA, SM723500014MA, SM723500015MA, SM723500016MA, SM723500017MA, SM723500018MA, SM723500019MA, SM723500020MA, SM724180113MA, SM724210022MA, SM724210023MA, SM724210024MA, SM724210025MA, SM724210026MA, SM724210027MA, SM724210028MA, SM724210029MA, SM724160068MA, SM723110002MA, SM724160069MA, SM724240001MA, SM724240002MA, SM724240003MA, SM724240004MA, SM724240005MA, SM724240006MA, SM724240007MA, SM724240008MA, SM724240009MA, SM724240010MA, SM724240011MA, SM723430001MA, SM723100026MA, SM723130012MA, SM723160019MA, SM723160021MA, SM723230003MA, SM723230005MA, SM724100049MA, SM724100050MA, SM724100051MA, SM724180106MA, SM724180107MA, SM724180108MA, SM724180109MA, SM724180110MA, SM724180111MA, SM724180112MA, SM724180114MA, SM724180115MA, SM723380021MA, SM723380022MA, SM723380023MA, SM723380024MA, SM723380025MA, SM723210069MA, SM723210070MA, SM723210071MA, SM723210072MA, SM724210003MA, SM724210004MA, SM723180012MA, SM723180013MA, SM723180014MA, SM723180015MA, SM723180016MA, SM723180017MA, SM723180018MA, SM723180019MA, SM723180020MA, SM723180021MA, SM723140056MA, SM723140057MA, SM723450033MA, SM723450034MA, SM723450035MA, SM723450036MA, SM723450037MA, SM723450038MA, SM723450039MA, SM723490004MA, SM723490005MA, SM724140006MA, SM724190138MA, SM724060007MA, SM724260032MA, SM724260030MA, SM724260031MA, SM724260033MA, SM724500001MA, SM724480003MA, SM724480004MA, SM724480005MA, SM724480006MA, SM724480007MA, SM724480008MA, SM724480009MA, SM724480010MA, SM724480011MA, SM724480012MA, SM724380017MA, SM724430001MA, SM724360008MA, SM724360009MA, SM724360010MA, SM724360011MA, SM724360012MA, SM724360013MA, SM723210002MA, SM723210003MA, SM724260003MA, SM724260004MA, SM724260005MA, SM724260006MA, SM724260007MA, SM724260008MA, SM724260009MA, SM724260017MA, SM723340007MA, SM724050020MA, SM723190003MA, SM723190004MA, SM723190005MA, SM723190006MA, SM723040001MA, SM723060008MA, SM724120014MA, SM724120015MA, SM723210065MA, SM723210066MA, SM723210067MA, SM723210068MA, SM724230012MA, SM724230013MA, SM724230014MA, SM724230015MA, SM724230016MA, SM724230017MA, SM724230018MA, SM723020042MA, SM723020043MA, SM724220004MA, SM724220005MA, SM724220006MA, SM724220007MA, SM724220008MA, SM724220009MA, SM724220010MA, SM723210025MA, SM724110003MA, SM724110004MA, SM724290001MA, SM724290002MA, SM724260015MA, SM724260016MA, SM723110064MA, SM723110065MA, SM723110066MA, SM723110067MA, SM724190139MA, SM724190140MA, SM724190141MA, SM724190142MA, SM724190143MA, SM724310017MA, SM724310018MA, SM724310019MA, SM724310020MA, SM724310021MA, SM724310022MA, SM724310023MA, SM724310024MA, SM724310025MA, SM723110061MA, SM723180061MA, SM723180062MA, SM723180063MA, SM723180064MA, SM723180065MA, SM723180066MA, SM723020041MA, SM724110047MA, SM724110048MA, SM724110049MA, SM724110050MA, SM724110051MA, SM724110052MA, SM724110053MA, SM724110054MA, SM724370002MA, SM724320006MA, SM724410002MA, SM724410003MA, SM724410004MA, SM723020031MA, SM723020032MA, SM723020033MA, SM723230001MA, SM724160071MA, SM724160072MA, SM723290024MA, SM723340009MA, SM723340011MA, SM723450025MA, SM723450026MA, SM723450027MA, SM724210002MA, SM724220001MA, SM724220003MA, SM723380003MA, SM723380004MA, SM723380005MA, SM723380006MA, SM723380007MA, SM723380008MA, SM723380009MA, SM724260001MA, SM724040023MA, SM724040024MA, SM724040025MA, SM724040026MA, SM724040027MA, SM724040028MA, SM724110001MA, SM723070089MA, SM723070090MA, SM723070091MA, SM723070092MA, SM723070093MA, SM723070094MA, SM723070095MA, SM723070096MA, SM723070097MA, SM723070098MA, SM723070099MA, SM723070100MA, SM723070101MA, SM723070102MA, SM724190118MA, SM724190119MA, SM724190120MA, SM724190121MA, SM724280001MA, SM723430002MA, SM723110015MA, SM723110016MA, SM723110017MA, SM723110018MA, SM723210030MA, SM723020026MA, SM723480011MA, SM724100028MA, SM724100029MA, SM724100030MA, SM724100031MA, SM724100032MA, SM724100033MA, SM724460001MA, SM723260012MA, SM724300005MA, SM724300006MA, SM724300007MA, SM723120003MA, SM723120004MA, SM723120005MA, SM724200014MA, SM724250022MA, SM724250023MA, SM724250024MA, SM724250025MA, SM724310014MA, SM724310015MA, SM724310016MA, SM724320008MA, SM724320009MA, SM724320010MA, SM724380019MA, SM724380020MA, SM723450024MA, SM724330002MA, SM723050004MA, SM723050005MA, SM724140034MA, SM724140035MA, SM723390001MA, SM724190116MA, SM723150016WA, SM724050012WA, SM724050013WA, SM724050014WA, SM724050015WA, SM723310014WA, SM723290085WA, SM723290086WA, SM723290087WA, SM723290088WA, SM723400008WA, SM723400009WA, SM723400010WA, SM723070070WA, SM723070071WA, SM724220012WA, SM724220013WA, SM724120038WA, SM723420001WA, SM723420002WA, SM723420003WA, SM723420004WA, SM723460073WA, SM723180069WA, SM723180067WA, SM723280001WA, SM723280002WA, SM723280003WA, SM723280004WA, SM723280005WA, SM723280006WA, SM723280007WA, SM723280008WA, SM723280009WA, SM723280010WA, SM723280011WA, SM723280026WA, SM723280027WA, SM724330017WA, SM724330018WA, SM723280018WA, SM723520028WA, SM724320017WA, SM724320018WA, SM724320019WA, SM724220014WA, SM724220015WA, SM724010004WA, SM723310015WA, SM723310016WA, SM723280019WA, SM724010023WA, SM723280020WA, SM723090061WA, SM723090062WA, SM723060069WA, SM723060070WA, SM724170127WA, SM724100054WA, SM723470110WA, SM723470109WA, SM724290009WA, SM724290010WA, SM724120037WA, SM723140009WA, SM723140010WA, SM724080014WA, SM724080015WA, SM724080016WA, SM723100034WA, SM724010020WA, SM724290023WA, SM724290024WA, SM724140002WA, SM724010019WA, SM723400005WA, SM723400006WA, SM724290016WA, SM724290017WA, SM724290018WA, SM724290019WA, SM724290020WA, SM724140003WA, SM724220016WA, SM723190002WA, SM723060074WA, SM724330011WA, SM723060075WA, SM723220006WA, SM723220007WA, SM723220008WA, SM723220009WA, SM723360002WA, SM724290003WA, SM723080003WA, SM723330012WA, SM723330013WA, SM723430001WA, SM723430002WA, SM724270018WA, SM723220012WA, SM723330006WA, SM723330007WA, SM724250024WA, SM724200029WA, SM724250023WA, SM723140032WA, SM723140033WA, SM723140034WA, SM723150012WA, SM724080039WA, SM724150084WA, SM723100045WA, SM723350053WA, SM724010018WA, SM724320020WA, SM724290022WA, SM724320025WA, SM723210073WA, SM723210074WA, SM723210075WA, SM724270003WA, SM723350062WA, SM723050074WA, SM723160010WA, SM723140031WA, SM723030001WA, SM724080018WA, SM724080019WA, SM723270019WA, SM723270020WA, SM723150013WA, SM723150014WA, SM723150015WA, SM723260013WA, SM723210032WA, SM723180034WA, SM723090067WA, SM723090068WA, SM723100077WA, SM723020037MA, SM723020038MA, SM723070074WA, SM723070075WA, SM723070076WA, SM723460056MA, SM723310005MA, SM724290032MA, SM724290033MA, SM724290034MA, SM724170097MA, SM724190137MA, SM724280028MA, SM724280029MA, SM724030024MA, SM723120008MA, SM723030005MA, SM723030006MA, SM724130001MA, SM724130002MA, SM724130003MA, SM723240014MA, SM723240015MA, SM723240016MA, SM723240017MA, SM723240018MA, SM723240019MA, SM723240020MA, SM723240021MA, SM725030004MA, SM723090066WA, SM723160045WA, SM723440017WA, SM723440018WA, SM723440019WA, SM723120068WA, SM723120069WA, SM723120070WA, SM723120071WA, SM723120072WA, SM723420002MA, SM723420003MA, SM723420004MA, SM723420005MA, SM723420006MA, SM723280071WA, SM723280072WA, SM723280073WA, SM723420020WA, SM723490010WA, SM723490017WA, SM723420019WA, SM723420021WA, SM723420022WA, SM723420025WA, SM723420023WA, SM723420024WA, SM723490007WA, SM723490008WA, SM723490009WA, SM723490011WA, SM723490012WA, SM723490014WA, SM723490018WA, SM723490019WA, SM723490020WA, SM723490022WA, SM723490023WA, SM723110060WA, SM723220020WA, SM723420018WA, SM723490015WA, SM723490006WA, SM723490025WA, SM723490027WA, SM723490013WA, SM723490016WA, SM723490021WA, SM723490024WA, SM723490026WA, SM724390004MA, SM724390005MA, SM724440008MA, SM723490010MA, SM723490011MA, SM723490012MA, SM723490013MA, SM723490014MA, SM724150045MA, SM724150046MA, SM723380001MA, SM723300031MA, SM723300032MA, SM723300033MA, SM723140036MA, SM724080059MA, SM724080060MA, SM724080061MA, SM724080062MA, SM723100001MA, SM723020039MA, SM723020040MA, SM724040014MA, SM723180046MA, SM724090027MA, SM724260002MA, SM724410010MA, SM724410011MA, SM724310013MA, SM723260015MA, SM723110006MA, SM723480001MA, SM723200006MA, SM723210004MA, SM723210005MA, SM723210006MA, SM723210007MA, SM723210008MA, SM723210009MA, SM723190042MA, SM723190043MA, SM723190044MA, SM723190045MA, SM723190046MA, SM723190047MA, SM723330023MA, SM723330024MA, SM723330025MA, SM723330026MA, SM723330027MA, SM723330028MA, SM723330029MA, SM723330030MA, SM723330031MA, SM723330032MA, SM723330033MA, SM723330034MA, SM723330035MA, SM723330036MA, SM723330037MA, SM723330038MA, SM724250018MA, SM724250019MA, SM724040019MA, SM724040020MA, SM724040021MA, SM724040022MA, SM723060044MA, SM723060045MA, SM724140008MA, SM724140009MA, SM724140010MA, SM724140023MA, SM723180003MA, SM723180009MA, SM723180010MA, SM723180011MA, SM723110062MA, SM723120006MA, SM723190007MA, SM723190008MA, SM723190009MA, SM723190010MA, SM723190011MA, SM723190012MA, SM723190013MA, SM723190014MA, SM723190015MA, SM723190016MA, SM723180002MA, SM723180008MA, SM723180004MA, SM723180005MA, SM723180006MA, SM723180007MA, SM723260001MA, SM723010026WA, SM723440032WA, SM723440033WA, SM724110046MA, SM723040009MA, SM723040010MA, SM723040011MA, SM723040012MA, SM723040013MA, SM723040014MA, SM723040015MA, SM723040016MA, SM723040017MA, SM723040018MA, SM723040019MA, SM723060047MA, SM723060048MA, SM723060049MA, SM723060050MA, SM723060051MA, SM723060052MA, SM723060053MA, SM723060054MA, SM723060055MA, SM723170080MA, SM723170081MA, SM723320003MA, SM723320004MA, SM723320005MA, SM724280002MA, SM724280003MA, SM724280004MA, SM724280005MA, SM724280006MA, SM724280007MA, SM724280008MA, SM724280009MA, SM724280010MA, SM724280011MA, SM724280012MA, SM724280013MA, SM724280014MA, SM724280015MA, SM724280016MA, SM724280017MA, SM724280018MA, SM724280019MA, SM724280020MA, SM724280021MA, SM724280022MA, SM724280023MA, SM724280024MA, SM724280026MA, SM724280027MA, SM724280030MA, SM724280031MA, SM724280032MA, SM724280033MA, SM724280034MA, SM724280035MA, SM724280036MA, SM724280037MA, SM724280038MA, SM724280039MA, SM724280040MA, SM724280041MA, SM723060046MA, SM723180001MA, SM723070012WA, SM723190052WA, SM723350046WA, SM723350047WA, SM723490001WA, SM723490002WA, SM723490003WA, SM723360001WA, SM723180070WA, SM723110030WA, SM724270001WA, SM723190054WA, SM724020062WA, SM724020063WA, SM723280059WA, SM723280060WA, SM723280061WA, SM724270012WA, SM723280068WA, SM723040040WA, SM723040041WA, SM723420015WA, SM723170045WA, SM723170046WA, SM723070013WA, SM723070022WA, SM723070023WA, SM723210056WA, SM724160098WA, SM724270009WA, SM723100064WA, SM723280074WA, SM723020017WA, SM723020018WA, SM724100018WA, SM724010005WA, SM723240015WA, SM723410010WA, SM723140007WA, SM724200001WA, SM724200002WA, SM724040037WA, SM723180030WA, SM724200014WA, SM724200015WA, SM723080008WA, SM723080009WA, SM724270010WA, SM723040039WA, SM723490029WA, SM724290013WA, SM724290014WA, SM724290015WA, SM724290011WA, SM724290012WA, SM723350057WA, SM723070020WA, SM724370001WA, SM724370002WA, SM724370004WA, SM724370005WA, SM724370008WA, SM724370009WA, SM724370010WA, SM725070001WA, SM725070002WA, SM725070003WA, SM725070004WA, SM725070005WA, SM725070006WA, SM725070007WA, SM725070008WA, SM725070009WA, SM725070010WA, SM725070011WA, SM725070012WA, SM725070013WA, SM725070014WA, SM725070015WA, SM725070016WA, SM725070017WA, SM725070018WA, SM725070019WA, SM725070020WA, SM725070021WA, SM725070022WA, SM723270016WA, SM724150081WA, SM724150082WA, SM723440003WA, SM723440004WA, SM724140041WA, SM723030041MA, SM723030040MA, SM723030042MA, SM723030044MA, SM723030045MA, SM723030046MA, SM723030047MA, SM723030048MA, SM723030049MA, SM723120017MA, SM723120018MA, SM723120019MA, SM723120020MA, SM723120021MA, SM724010026MA, SM723360002MA, SM723360003MA, SM724010032MA, SM724010033MA, SM724010025MA, SM724010027MA, SM724010028MA, SM724010029MA, SM724010030MA, SM724010034MA, SM724010031MA, SM723030043MA, SM724320015WA, SM723400004WA, SM724230005WA, SM724230006WA, SM724150078WA, SM724150079WA, SM724020005WA, SM724150071WA, SM724250008WA, SM724250009WA, SM723130037WA, SM723130038WA, SM724280017WA, SM724280002WA, SM723010039WA, SM723010040WA, SM723010041WA, SM723010042WA, SM724410008WA, SM724280004WA, SM724280018WA, SM724280016WA, SM723220021MA, SM723220022MA, SM723370005WA, SM723030028MA, SM724150074WA, SM723300048WA, SM723300049WA, SM724170123WA, SM724170124WA, SM723350028WA, SM724120004WA, SM723360073WA, SM723350045WA, SM723470106WA, SM724030017WA, SM724030018WA, SM724110023WA, SM724110024WA, SM723050059WA, SM724250012WA, SM724250013WA, SM723420011WA, SM724080034WA, SM724280021WA, SM724510002WA, SM723370007WA, SM724150073WA, SM724080033WA, SM723420008WA, SM723420009WA, SM723420010WA, SM724250010WA, SM724410009WA, SM724410004WA, SM724020003WA, SM724120007WA, SM723420012WA, SM724280023WA, SM723270007WA, SM723170057WA, SM723300044WA, SM723300045WA, SM723300046WA, SM723300050WA, SM724020006WA, SM724020007WA, SM724170130WA, SM724170131WA, SM723280033WA, SM723280034WA, SM723370009WA, SM724170125WA, SM724170126WA, SM723350038WA, SM723230011WA, SM723230012WA, SM723230013WA, SM723350029WA, SM723420013WA, SM723420014WA, SM723270011WA, SM723310033WA, SM723350040WA, SM723350043WA, SM723370006WA, SM723370008WA, SM723310034WA, SM723310035WA, SM723310036WA, SM723310037WA, SM723470099WA, SM724410011WA, SM724170129WA, SM723350036WA, SM723350037WA, SM723030030MA, SM723030031MA, SM723280035WA, SM724120008WA, SM724410003WA, SM723050062WA, SM723050020WA, SM724120009WA, SM724120010WA, SM724120011WA, SM724120012WA, SM724120013WA, SM724280025WA, SM723370010WA, SM723370012WA, SM723370013WA, SM723030029MA, SM724110025WA, SM724110026WA, SM724250007WA, SM723350041WA, SM725030001MA, SM725030002MA, SM725030003MA, SM723370014WA, SM723220023MA, SM725020008MA, SM725020009MA, SM725020010MA, SM725020011MA, SM724080030WA, SM723120024WA, SM723120025WA, SM725020004MA, SM725020005MA, SM725020006MA, SM725020007MA, SM724250014WA, SM724250015WA, SM724250016WA, SM724250017WA, SM724250018WA, SM723370011WA, SM723310039WA, SM723310038WA, SM723470105WA, SM723130021WA, SM723280031WA, SM723280032WA, SM723050022WA, SM723050023WA, SM723050057WA, SM723200001MA, SM723200002MA, SM723270008WA, SM723310040WA, SM723310041WA, SM723310042WA, SM724020001WA, SM724020002WA, SM723130017WA, SM723270013WA, SM723370015WA, SM723370016WA, SM724080032WA, SM724150066WA, SM724150067WA, SM724150068WA, SM724150069WA, SM724150070WA, SM724410007WA, SM724410010WA, SM724440007WA, SM724440008WA, SM724410005WA, SM723050060WA, SM723050061WA, SM723270006WA, SM724170122WA, SM724280001WA, SM723130018WA, SM723350044WA, SM723350027WA, SM724510003WA, SM724510004WA, SM724510005WA, SM725010001WA, SM725010002WA, SM725010003WA, SM725010004WA, SM725010005WA, SM725010006WA, SM725010007WA, SM725010008WA, SM725010009WA, SM725010010WA, SM723350042WA, SM723170058WA, SM723170059WA, SM723170061WA, SM723170062WA, SM723300047WA, SM723170053WA, SM723170054WA, SM723170055WA, SM723170056WA, SM723410006WA, SM723410007WA, SM723410008WA, SM723410009WA, SM723270009WA, SM723270010WA, SM723280036WA, SM723280037WA, SM723280045WA, SM724250001WA, SM724150072WA, SM724280019WA, SM724280020WA, SM724410001WA, SM724410002WA, SM723470102WA, SM723270003WA, SM723230006WA, SM723230007WA, SM723230010WA, SM723230008WA, SM723230009WA, SM723300043WA, SM724020009WA, SM723050018WA, SM723280029WA, SM723280030WA, SM724170132WA, SM724020004WA, SM724020008WA, SM723050021WA, SM723360072WA, SM723120034WA, SM723350031WA, SM723350032WA, SM723350033WA, SM723350034WA, SM723010043WA, SM723010044WA, SM723050019WA, SM723120022WA, SM723120023WA, SM723170060WA, SM724280013WA, SM724280014WA, SM724030009WA, SM724030010WA, SM724030011WA, SM724030012WA, SM724030013WA, SM724030014WA, SM724080031WA, SM724120005WA, SM724120006WA, SM723470097WA, SM723470098WA, SM723350030WA, SM723470103WA, SM723470104WA, SM724280024WA, SM723050058WA, SM724080029WA, SM723470100WA, SM723470101WA, SM724150076WA, SM724150077WA, SM724020010WA, SM723270005WA, SM724510001WA, SM724250021WA, SM724280011WA, SM723120026WA, SM723120027WA, SM723120028WA, SM723120032WA, SM723120033WA, SM723350039WA, SM723270012WA, SM723300051WA, SM723300052WA, SM723350035WA, SM724280015WA, SM724250020WA, SM723130019WA, SM723130020WA, SM723420005WA, SM724250006WA, SM724280012WA, SM724030015WA, SM724030016WA, SM724250003WA, SM724250004WA, SM724250005WA, SM724280003WA, SM724280005WA, SM724280022WA, SM724250002WA, SM724150080WA, SM724170128WA, SM723270004WA, SM723350026WA, SM723420006WA, SM723420007WA, SM724150075WA, SM723240005MA, SM723180058MA, SM724360003MA, SM724360004MA, SM724360005MA, SM724360006MA, SM724360007MA, SM724370005MA, SM724370006MA, SM724370007MA, SM724370008MA, SM724370009MA, SM724390011MA, SM724390013MA, SM724390014MA, SM724390015MA, SM724400011MA, SM723170002MA, SM723260006MA, SM723260007MA, SM723290013MA, SM723290014MA, SM723290015MA, SM724250004MA, SM724250005MA, SM724250006MA, SM724250007MA, SM724250008MA, SM724240019MA, SM724240020MA, SM724240021MA, SM724240022MA, SM724240023MA, SM724260010MA, SM724260011MA, SM724260012MA, SM724260013MA, SM724260014MA, SM723050008MA, SM723060065MA, SM723050006MA, SM723050007MA, SM724240014MA, SM724240015MA, SM724240016MA, SM724240017MA, SM724240018MA, SM723060066MA, SM723180050MA, SM723180059MA, SM723180060MA, SM723300036MA, SM723300037MA, SM723300038MA, SM723320007MA, SM723320008MA, SM723320009MA, SM723320010MA, SM723340016MA, SM723340017MA, SM723340018MA, SM723340019MA, SM723430006MA, SM723430007MA, SM723430008MA, SM723430009MA, SM723430010MA, SM724010007MA, SM724010008MA, SM724010009MA, SM724010010MA, SM724010011MA, SM724060012MA, SM724060013MA, SM724060014MA, SM724060015MA, SM724060016MA, SM724060017MA, SM724060018MA, SM724060019MA, SM724060020MA, SM724060021MA, SM724100021MA, SM724130004MA, SM724130005MA, SM724130006MA, SM724130007MA, SM723240002MA, SM723170003MA, SM723190029MA, SM723180049MA, SM723180051MA, SM723180048MA, SM723300034MA, SM723300035MA, SM723060063MA, SM723060064MA, SM723320006MA, SM723240003MA, SM723180056MA, SM723180057MA, SM723190028MA, SM723240001MA, SM723170001MA, SM723440002MA, SM723440010MA, SM723440011MA, SM724010024MA, SM724290040MA, SM724290041MA, SM723120010MA, SM724290030MA, SM724130008MA, SM724130009MA, SM724130010MA, SM724130011MA, SM724130012MA, SM724290035MA, SM724290036MA, SM723170007MA, SM723170008MA, SM723170052MA, SM723060067MA, SM723240004MA, SM723310004MA, SM724140017MA, SM724140018MA, SM724140019MA, SM724140020MA, SM724140021MA, SM724140022MA, SM724150028MA, SM724150029MA, SM724150030MA, SM724150031MA, SM724150032MA, SM724150033MA, SM724150034MA, SM724150035MA, SM724150036MA, SM724150037MA, SM724150038MA, SM724150039MA, SM723440025WA, SM723440026WA, SM723440027WA, SM723440028WA, SM723450035WA, SM723450036WA, SM723450037WA, SM723450038WA, SM723450039WA, SM723450040WA, SM723450041WA, SM723450042WA, SM723450043WA, SM723450044WA, SM723450045WA, SM723450046WA, SM723450047WA, SM723450048WA, SM724170133WA, SM724170134WA, SM724170135WA, SM724450012WA, SM724450013WA, SM724450014WA, SM724450015WA, SM724450016WA, SM724450017WA, SM724450018WA, SM724450019WA, SM724450020WA, SM724450021WA, SM724450022WA, SM724450023WA, SM724450024WA, SM723100069WA, SM723100070WA, SM723100071WA, SM723290093WA, SM723440005WA, SM723440006WA, SM723440007WA, SM723440008WA, SM723440009WA, SM723440012WA, SM723440013WA, SM723440014WA, SM723440015WA, SM723440016WA, SM724190155WA, SM724190156WA, SM724270013WA, SM724270014WA, SM724270015WA, SM724270016WA, SM724270017WA, SM724320001WA, SM724320002WA, SM724320003WA, SM724320004WA, SM724320005WA, SM724320006WA, SM724320007WA, SM724320008WA, SM724320009WA, SM724320010WA, SM724320011WA, SM724320012WA, SM724320013WA, SM724320014WA, SM724190161WA, SM723280021WA, SM723280022WA, SM723280023WA, SM723060072WA, SM724190139WA, SM724190140WA, SM724200003WA, SM724200004WA, SM724200005WA, SM724190137WA, SM723350001WA, SM723140022WA, SM723280025WA, SM723160011WA, SM723180068WA, SM723190017WA, SM723020011WA, SM723500024WA, SM724100013WA, SM723280028WA, SM724340001WA, SM724340002WA, SM724340003WA, SM723160012WA, SM724190138WA, SM723290009WA, SM723420026WA, SM723010008WA, SM723010009WA, SM723020013WA, SM723020012WA, SM723350002WA, SM724060002WA, SM723350063WA, SM723020009WA, SM723290094WA, SM723220013WA, SM723480014WA, SM723280082WA, SM723280083WA, SM723280084WA, SM724070023WA, SM723300040WA, SM723300041WA, SM723090063WA, SM723060068WA, SM724020027WA, SM724020028WA, SM723060043WA, SM724080042WA, SM723350055WA, SM723350056WA, SM724080041WA, SM723020020WA, SM723020021WA, SM723020022WA, SM723020023WA, SM723020024WA, SM723180043WA, SM723330004WA, SM723050027WA, SM723060042WA, SM723440030WA, SM723440031WA, SM723030025WA, SM723050029WA, SM724030020WA, SM724070035WA, SM724040034WA, SM724040035WA, SM723050070WA, SM723280063WA, SM723050025WA, SM723480001WA, SM723060032WA, SM723310009WA, SM723110001WA, SM723460069WA, SM723460076WA, SM723160014WA, SM723170049WA, SM723170050WA, SM723060006WA, SM723030015WA, SM724110016WA, SM724020017WA, SM724090004WA, SM723480005WA, SM723160013WA, SM723470113WA, SM724080024WA, SM723350065WA, SM723100092WA, SM724110056WA, SM724090066WA, SM724090067WA, SM724100012WA, SM723460065WA, SM723180037WA, SM723280014WA, SM723380001WA, SM723380002WA, SM723380003WA, SM724100011WA, SM723430018WA, SM723430019WA, SM724090063WA, SM723440001WA, SM724110015WA, SM723470108WA, SM723350068WA, SM723020003WA, SM723350066WA, SM723220019WA, SM723310006WA, SM724040036WA, SM723470111WA, SM723460063WA, SM723440023WA, SM723060033WA, SM723060037WA, SM723300007WA, SM723280075WA, SM724020026WA, SM723050024WA, SM723460085WA, SM723460064WA, SM723310002WA, SM723020002WA, SM723460089WA, SM724040029WA, SM723370002WA, SM723220011WA, SM723100094WA, SM723100095WA, SM723100096WA, SM723280065WA, SM723180032WA, SM723020005WA, SM723430021WA, SM723430022WA, SM723440020WA, SM723030024WA, SM723050030WA, SM723050031WA, SM723050032WA, SM723150011WA, SM723170048WA, SM723280067WA, SM723340029WA, SM724080005WA, SM724080006WA, SM723490030WA, SM723300008WA, SM723180072WA, SM723180073WA, SM724030045WA, SM723030018WA, SM723450052WA, SM723300009WA, SM723300010WA, SM724080009WA, SM724080010WA, SM723430024WA, SM723310011WA, SM723310012WA, SM724050017WA, SM723030022WA, SM723030023WA, SM724080007WA, SM724080008WA, SM723150010WA, SM723050073WA, SM723500025WA, SM724080040WA, SM723020004WA, SM723180041WA, SM723180042WA, SM723030016WA, SM723030017WA, SM724050018WA, SM724050019WA, SM723180039WA, SM723180040WA, SM724080002WA, SM724080003WA, SM724080004WA, SM723310003WA, SM724070025WA, SM724080012WA, SM724070024WA, SM724110017WA, SM724110018WA, SM723350022WA, SM723350023WA, SM723300011WA, SM723300012WA, SM724070034WA, SM723060040WA, SM723440002WA, SM723250005WA, SM723450049WA, SM723460074WA, SM723460075WA, SM723460060WA, SM723030014WA, SM724080022WA, SM723440010WA, SM723180071WA, SM723350024WA, SM723350025WA, SM724080036WA, SM723010029WA, SM723010030WA, SM723180031WA, SM724110019WA, SM724110020WA, SM724040033WA, SM723010037WA, SM723010038WA, SM724010021WA, SM723450051WA, SM723400001WA, SM723400002WA, SM723400003WA, SM723330020WA, SM723020008WA, SM723460080WA, SM723460081WA, SM723180033WA, SM723470114WA, SM723470115WA, SM723380004WA, SM723430003WA, SM723450050WA, SM723430016WA, SM723430017WA, SM723410001WA, SM723410002WA, SM723060036WA, SM723430032WA, SM723020006WA, SM723460086WA, SM723030019WA, SM723300042WA, SM723460068WA, SM723360077WA, SM723280064WA, SM723450054WA, SM724100044WA, SM723430023WA, SM723110055WA, SM723110058WA, SM723110059WA, SM723250003WA, SM723250004WA, SM723050072WA, SM723060035WA, SM723430020WA, SM723450053WA, SM723350059WA, SM723350060WA, SM724030021WA, SM724030022WA, SM724010002WA, SM724010003WA, SM723460079WA, SM723010027WA, SM723010028WA, SM723410015WA, SM724040031WA, SM723370004WA, SM723060034WA, SM723350064WA, SM723160015WA, SM723430015WA, SM723460091WA, SM723460093WA, SM723470107WA, SM723470128WA, SM723460066WA, SM723460077WA, SM723280077WA, SM723110054WA, SM723110056WA, SM723110057WA, SM724090001WA, SM724090002WA, SM724090003WA, SM723470112WA, SM723470120WA, SM724010022WA, SM723060071WA, SM723020007WA, SM723310010WA, SM723280078WA, SM723290089WA, SM723290090WA, SM723290091WA, SM723290092WA, SM723350054WA, SM723250001WA, SM723250002WA, SM723300005WA, SM723300006WA, SM724080013WA, SM723350067WA, SM723430033WA, SM723210031WA, SM723220001WA, SM723220002WA, SM723220003WA, SM723220004WA, SM723220005WA, SM723220010WA, SM723330003WA, SM723050056WA, SM723310028WA, SM723310029WA, SM723310030WA, SM723310031WA, SM724070026WA, SM724070027WA, SM724070028WA, SM724070029WA, SM724070030WA, SM724070031WA, SM724070032WA, SM724070033WA, SM723060039WA, SM723440029WA, SM723030020WA, SM723030021WA, SM723440021WA, SM723440022WA, SM723150008WA, SM723150009WA, SM723350058WA, SM723460082WA, SM723310013WA, SM723180038WA, SM723460092WA, SM723100093WA, SM723360021WA, SM723340028WA, SM723350061WA, SM723060038WA, SM724110022WA, SM723020001WA, SM724090065WA, SM723280079WA, SM723280080WA, SM723280081WA, SM724090064WA, SM723050063WA, SM723110046WA, SM723030026WA, SM723030027WA, SM724040030WA, SM724050036WA, SM723010031WA, SM723310032WA, SM723460078WA, SM724050016WA, SM723010025WA, SM723010024WA, SM723050028WA, SM723020010WA, SM724080037WA, SM723210026WA, SM724110021WA, SM723060041WA, SM723470127WA, SM723460061WA, SM723460062WA, SM723460088WA, SM723050026WA, SM723180044WA, SM723270017WA, SM723270018WA, SM723050071WA, SM723070068WA, SM723070069WA, SM724080020WA, SM724080021WA, SM724290001WA, SM724290002WA, SM724320023WA, SM724320024WA, SM723310007WA, SM723460094WA, SM724050028WA, SM724080001WA, SM723460095WA, SM724150043WA, SM724020012WA, SM724440001WA, SM724440002WA, SM723190026MA, SM723460096WA, SM724340005WA, SM724340006WA, SM724340007WA, SM724320021WA, SM724320022WA, SM723170005MA, SM723010004WA, SM723010005WA, SM723060026MA, SM723170013MA, SM723170014MA, SM723170015MA, SM723170016MA, SM724080043MA, SM724080044MA, SM724200020MA, SM724200021MA, SM724200022MA, SM724200023MA, SM724200024MA, SM724200025MA, SM724200026MA, SM724200027MA, SM724290003MA, SM724290004MA, SM724290005MA, SM724290006MA, SM724160055MA, SM724150048WA, SM724150049WA, SM723070009WA, SM723070010WA, SM723070019WA, SM723300004WA, SM723210046WA, SM723210047WA, SM723210048WA, SM723210049WA, SM723360075WA, SM723360076WA, SM723210022WA, SM723210023WA, SM723210024WA, SM723210025WA, SM723210050WA, SM723360009WA, SM723360010WA, SM723360011WA, SM723210055WA, SM724050003WA, SM724050004WA, SM724050002WA, SM723350069WA, SM724160100WA, SM724160101WA, SM724160102WA, SM724160103WA, SM723490031WA, SM724050005WA, SM724050006WA, SM724050007WA, SM724050001WA, SM723490033WA, SM723210051WA, SM723210054WA, SM723210053WA, SM723210052WA, SM723350070WA, SM723350071WA, SM723490032WA, SM724220004WA, SM724100047WA, SM724100048WA, SM724450007WA, SM724450008WA, SM724450009WA, SM724450010WA, SM724220002WA, SM724220003WA, SM724220005WA, SM724220006WA, SM724290004WA, SM724290005WA, SM724290006WA, SM724290007WA, SM724290008WA, SM723330008WA, SM723330009WA, SM723330010WA, SM723330019WA, SM724330004WA, SM724330005WA, SM723060014MA, SM723060015MA, SM723060016MA, SM724170120WA, SM724110002MA, SM723060013MA, SM724140040WA, SM723060017MA, SM723060018MA, SM724130018MA, SM724330006WA, SM724330007WA, SM724330008WA, SM724330009WA, SM724330010WA, SM724440003WA, SM724440004WA, SM723100052WA, SM724070001MA, SM724070003MA, SM724070004MA, SM724070005MA, SM724070006MA, SM724070007MA, SM724110009MA, SM724110010MA, SM724110011MA, SM724110012MA, SM724110013MA, SM724070002MA, SM724100058WA, SM724100059WA, SM724100060WA, SM724100061WA, SM724100062WA, SM723080005WA, SM723080006WA, SM724140020WA, SM724290021WA, SM723340013MA, SM723490001MA, SM724010014MA, SM723120009MA, SM724010012MA, SM724010013MA, SM724010015MA, SM724010016MA, SM724010017MA, SM723070072WA, SM723070073WA, SM723010006WA, SM723010007WA, SM723070055WA, SM724050025WA, SM724050026WA, SM724020013WA, SM724200028WA, SM723410012WA, SM723060073WA, SM724020016WA, SM723460067WA, SM723070033WA, SM723170051WA, SM723280024WA, SM723330001WA, SM723330002WA, SM724040050WA, SM724040051WA, SM723370001WA, SM723430027WA, SM723010010WA, SM723170063MA, SM723260002MA, SM723260003MA, SM724080045MA, SM724080046MA, SM724080047MA, SM724080048MA, SM724080049MA, SM724080050MA, SM724080051MA, SM724080052MA, SM724080053MA, SM724080054MA, SM724090038MA, SM724090039MA, SM724090040MA, SM724090041MA, SM724090042MA, SM724090043MA, SM724090044MA, SM724090045MA, SM724090046MA, SM724090047MA, SM724090048MA, SM724090049MA, SM724090050MA, SM724090051MA, SM724100034MA, SM724100035MA, SM724100036MA, SM724100037MA, SM724100038MA, SM724100039MA, SM724100040MA, SM724100041MA, SM724100042MA, SM724100043MA, SM724140011MA, SM724140012MA, SM724140013MA, SM724140014MA, SM724140015MA, SM724140016MA, SM724140004MA, SM724150051MA, SM724150052MA, SM724160056MA, SM723070029MA, SM724160053MA, SM723100007MA, SM723100008MA, SM723100009MA, SM723100010MA, SM723100011MA, SM723100012MA, SM723100013MA, SM723470124WA, SM724020011WA, SM723370010MA, SM724170096MA, SM724190117MA, SM724440009WA, SM724080038WA, SM724170111WA, SM724170112WA, SM724170119WA, SM724200010WA, SM723100024WA, SM723100025WA, SM723100028WA, SM723100032WA, SM723100033WA, SM723100078WA, SM723100079WA, SM723210057WA, SM723220014WA, SM723220015WA, SM723220016WA, SM723220017WA, SM723220018WA, SM723220076WA, SM723220077WA, SM724100006WA, SM724100007WA, SM724100008WA, SM724100009WA, SM724170113WA, SM724170114WA, SM724170115WA, SM724170116WA, SM724170117WA, SM724170118WA, SM724330012WA, SM724330013WA, SM724330014WA, SM724330015WA, SM724330016WA, SM724410015WA, SM725040001WA, SM725040002WA, SM725070023WA, SM723500005WA, SM724210003WA, SM723470125WA, SM723500001WA, SM723070004WA, SM724150042WA, SM724150052WA, SM724150053WA, SM723170047WA, SM723410017WA, SM723410018WA, SM723280069WA, SM723280070WA, SM723500023WA, SM723420016WA, SM723420017WA, SM723280017WA, SM723480002WA, SM724190136WA, SM723360012WA, SM723450034WA, SM723500004WA, SM723500021WA, SM724220001WA, SM724220011WA, SM724450002WA, SM724230001WA, SM724230002WA, SM723410013WA, SM723410014WA, SM723500017WA, SM723500018WA, SM723500019WA, SM723500020WA, SM723040023WA, SM723040024WA, SM723180045WA, SM723410016WA, SM723150007WA, SM723500008WA, SM723500009WA, SM724250022WA, SM724230003WA, SM723500016WA, SM723510026WA, SM723500006WA, SM723500007WA, SM723150006WA, SM723500015WA, SM724050024WA, SM724150046WA, SM724150060WA, SM723300013WA, SM724030046WA, SM723330005WA, SM723410005WA, SM723360014WA, SM723360015WA, SM723360016WA, SM723150004WA, SM723150005WA, SM723500003WA, SM723500014WA, SM724360008WA, SM724360009WA, SM723430026WA, SM723360017WA, SM723360018WA, SM723360019WA, SM723360022WA, SM723360023WA, SM723360024WA, SM723360025WA, SM723360026WA, SM723360027WA, SM724150061WA, SM723190001WA, SM723350003WA, SM723500002WA, SM723510027WA, SM724230004WA, SM724150047WA, SM723070011WA, SM723070021WA, SM723070058WA, SM723410019WA, SM723360013WA, SM723210043WA, SM723360020WA, SM723460087WA, SM723480003WA, SM723350004WA, SM723070059WA, SM723070060WA, SM724450003WA, SM724450004WA, SM724450005WA, SM724450006WA, SM723370003WA, SM723500022WA, SM723500123WA, SM723500124WA, SM723500125WA, SM724370001MA, SM723440018MA, SM724010051MA, SM723480006WA, SM723480007WA, SM723480008WA, SM723480009WA, SM723480010WA, SM723480011WA, SM723480012WA, SM723480013WA, SM724390002WA, SM723170075WA, SM723280012WA, SM723330021WA, SM723330022WA, SM723340030WA, SM723340031WA, SM723360008WA, SM723410020WA, SM723430025WA, SM723470116WA, SM723470117WA, SM723470118WA, SM723470119WA, SM724280010WA, SM724330001WA, SM724350001WA, SM723270001WA, SM723280055WA, SM724410016WA, SM724410017WA, SM724370003WA, SM723280042WA, SM723280015WA, SM723280016WA, SM723280058WA, SM723280062WA, SM724430001WA, SM723280039WA, SM723170076WA, SM723170077WA, SM723170078WA, SM723170079WA, SM723180029WA, SM723460070WA, SM723460071WA, SM724440005WA, SM724440006WA, SM723180022WA, SM723180023WA, SM723490004WA, SM723490005WA, SM723190021WA, SM723190022WA, SM723190023WA, SM723190024WA, SM723180024WA, SM723180025WA, SM723180026WA, SM723180027WA, SM723180028WA, SM724140023WA, SM724140024WA, SM724140025WA, SM724140026WA, SM724140027WA, SM724140028WA, SM724140029WA, SM724140030WA, SM724220007WA, SM724220008WA, SM724220009WA, SM724220010WA, SM724190141WA, SM724190142WA, SM724190143WA, SM724190144WA, SM724190145WA, SM724190146WA, SM724190147WA, SM724190148WA, SM723280038WA, SM723090065WA, SM723280043WA, SM723280044WA, SM723280046WA, SM723280047WA, SM723280048WA, SM723280049WA, SM723280050WA, SM723280051WA, SM723280052WA, SM723280053WA, SM723280054WA, SM723280056WA, SM723280057WA, SM723020019WA, SM723020045WA, SM723020046WA, SM723020047WA, SM723040026WA, SM723040027WA, SM723040028WA, SM723040029WA, SM723040030WA, SM723040031WA, SM723040032WA, SM723040033WA, SM723040034WA, SM723040035WA, SM723040036WA, SM723040037WA, SM723040038WA, SM724270002WA, SM723280040WA, SM723280041WA, SM723450055WA, SM723450056WA, SM723450057WA, SM723450059WA, SM723450058WA, SM723010001WA, SM723010002WA, SM723010003WA, SM723180036WA, SM723020016WA, SM723020014WA, SM723020015WA, SM724370006WA, SM724370007WA, SM724330002WA, SM723180035WA, SM723170071MA, SM723170067MA, SM723060001MA, SM723170068MA, SM723030033MA, SM723030034MA, SM723170064MA, SM723170065MA, SM723170085MA, SM723060002MA, SM723170089MA, SM723170082MA, SM723020027MA, SM723020028MA, SM724090026MA, SM723170072MA, SM723170073MA, SM723170066MA, SM723170084MA, SM723170086MA, SM723170087MA, SM723170088MA, SM723170069MA, SM723170070MA, SM723170074MA, SM723170083MA, SM723060003MA, SM723060004MA, SM723060005MA, SM723340014MA, SM723340015MA, SM724010040MA, SM724340004WA, SM724010035MA, SM724160059MA, SM724160060MA, SM724190157WA, SM724190158WA, SM724390001MA, SM723360003WA, SM723360004WA, SM723360005WA, SM723360006WA, SM723360007WA, SM724280025MA, SM723360001MA, SM724120001WA, SM723500010WA, SM723500011WA, SM723500012WA, SM723500013WA, SM724060038WA, SM724060039WA, SM724060040WA, SM724060041WA, SM724060042WA, SM723080004WA, SM723170027WA, SM723170025WA, SM723170026WA, SM723170028WA, SM723170029WA, SM723170031WA, SM723170032WA, SM723170033WA, SM723170034WA, SM723170037WA, SM723170038WA, SM723170039WA, SM723170041WA, SM723170042WA, SM723170044WA, SM724190149WA, SM724190150WA, SM724190151WA, SM724190152WA, SM724190153WA, SM724190154WA, SM723170030WA, SM723170035WA, SM723170043WA, SM723170036WA, SM723170040WA, SM723100046WA, SM723070001WA, SM723330018WA, SM723060078WA, SM724200016WA, SM724200017WA, SM724160099WA, SM723100035WA, SM723100036WA, SM723190018WA, SM723190019WA, SM723190020WA, SM723190048WA, SM723190049WA, SM724150062WA, SM724150063WA, SM724150064WA, SM724150065WA, SM724200011WA, SM724160094WA, SM724160095WA, SM724160096WA, SM724160097WA, SM724270011WA, SM723080007WA, SM724080023WA, SM723440011WA, SM723280076WA, SM723140030WA, SM723140011WA, SM724150044WA, SM724150045WA, SM723190053WA, SM723080002WA, SM724200006WA, SM724200007WA, SM724200008WA, SM724200009WA, SM723200003WA, SM723200004WA, SM723200005WA, SM724220017WA, SM724220018WA, SM724140001WA, SM723140006WA, SM724160086WA, SM724160087WA, SM724160088WA, SM724060001WA, SM723290002WA, SM723290003WA, SM723140029WA, SM723140001WA, SM723140002WA, SM723140003WA, SM723140004WA, SM723140005WA, SM724510006WA, SM723080018WA, SM723080019WA, SM723080001WA, SM724150083WA, SM724190159WA, SM724190160WA, SM723040021WA, SM723190050WA, SM723190051WA, SM723300039WA, SM723470126WA, SM723280013WA, SM723440024WA, SM723480004WA, SM724380001WA, SM723140008WA, SM724060003WA, SM724140019WA, SM724150050WA, SM724150051WA, SM723410004WA, SM723410003WA, SM723400007WA, SM723290007WA, SM723290008WA, SM723290010WA, SM723040025WA, SM723290004WA, SM723290005WA, SM723290006WA, SM723050064WA, SM723050065WA, SM723050066WA, SM723050067WA, SM723050068WA, SM723050069WA, SM723410021WA, SM724360015MA, SM724010053MA, SM724010054MA, SM724010055MA, SM724020019MA, SM724020020MA, SM724020021MA, SM724020018MA, SM723430003MA, SM723470002MA, SM723470003MA, SM723470004MA, SM723470005MA, SM724110027MA, SM724110028MA, SM724110029MA, SM724110030MA, SM723320002MA, SM724320004MA, SM724320005MA, SM724110031MA, SM724110032MA, SM724110033MA, SM724110034MA, SM724110035MA, SM724110036MA, SM724110037MA, SM724110038MA, SM724110039MA, SM723440013MA, SM723440014MA, SM723440015MA, SM723440016MA, SM723440017MA, SM724100020MA, SM723130022MA, SM723130023MA, SM723130024MA, SM723130025MA, SM723130026MA, SM723130027MA, SM723130028MA, SM724390003MA, SM724390016MA, SM724390017MA, SM724390018MA, SM724390019MA, SM724390020MA, SM724390021MA, SM724390022MA, SM724390023MA, SM724390024MA, SM724390025MA, SM724510004MA, SM724510005MA, SM724510006MA, SM724510007MA, SM724510008MA, SM724510009MA, SM724510010MA, SM724510011MA, SM724510012MA, SM723060021MA, SM723130007MA, SM723300019MA, SM723300021MA, SM723300022MA, SM723300023MA, SM723440005MA, SM723440006MA, SM723440007MA, SM723440008MA, SM723440009MA, SM723460057MA, SM723460058MA, SM724060029MA, SM724250020MA, SM724250021MA, SM724290037MA, SM724290038MA, SM724290039MA, SM724290042MA, SM724290043MA, SM724290044MA, SM724380015MA, SM724380016MA, SM724390002MA, SM723210033WA, SM723210034WA, SM723210035WA, SM723210036WA, SM723210037WA, SM723210038WA, SM723210039WA, SM723210040WA, SM723210041WA, SM723210042WA, SM724080017WA, SM723130006MA, SM723060019MA, SM723060020MA, SM723300020MA, SM723060022MA, SM723060023MA, SM723060024MA, SM723060025MA, SM724450011WA, SM724120040WA, SM724120039WA, SM724040032WA, SM724160089WA, SM724160090WA, SM724160091WA, SM724160092WA, SM724160093WA, SM724270004WA, SM724270005WA, SM724270006WA, SM724270007WA, SM724270008WA, SM723070002WA, SM723070003WA, SM723070061WA, SM723070062WA, SM723070063WA, SM723070064WA, SM723070065WA, SM723070066WA, SM723070067WA, SM723300001WA, SM723300002WA, SM723300003WA, SM724050029WA, SM724050030WA, SM724050031WA, SM724050032WA, SM724050033WA, SM724050034WA, SM724050035WA, SM723250010MA, SM723250011MA, SM723250012MA, SM723250007MA, SM723250008MA, SM723250009MA, SM723460090WA, SM724090053WA, SM724090054WA, SM724090055WA, SM724090056WA, SM724090057WA, SM723020054WA, SM723150003WA, SM723100063WA, SM723100065WA, SM724030032WA, SM724030033WA, SM724030034WA, SM724030035WA, SM724030036WA, SM724030037WA, SM724030038WA, SM724030039WA, SM724030040WA, SM724030041WA, SM723070005WA, SM723070006WA, SM723070007WA, SM723070008WA, SM723300053WA, SM723300054WA, SM723300055WA, SM723300056WA, SM723300057WA, SM723300058WA, SM723100051WA, SM723100058WA, SM723100059WA, SM723100060WA, SM724120058WA, SM723070053WA, SM723070077WA, SM724320016WA, SM724090052WA, SM723410011WA, SM723410022WA, SM723050044WA, SM723050045WA, SM723050046WA, SM723050047WA, SM723050048WA, SM723050049WA, SM724200012WA, SM724200013WA, SM723060076WA, SM723090003WA, SM723090004WA, SM723090005WA, SM723090006WA, SM723100082WA, SM723100083WA, SM723090002WA, SM723420027WA, SM723420028WA, SM723420029WA, SM723350008WA, SM723350009WA, SM723350010WA, SM723350011WA, SM723350012WA, SM723350013WA, SM723050052WA, SM723050053WA, SM723050054WA, SM723050055WA, SM723050075WA, SM723050076WA, SM723050077WA, SM723070045WA, SM723070080WA, SM723090001WA, SM723100023WA, SM723100074WA, SM723100076WA, SM723220035WA, SM723220036WA, SM724210002WA, SM724210001WA, SM723110031WA, SM723110032WA, SM724070018WA, SM724070019WA, SM724070020WA, SM724070021WA, SM724070022WA, SM724280006WA, SM724280007WA, SM723010032WA, SM723010033WA, SM723010034WA, SM723010035WA, SM723010036WA, SM723090007WA, SM723090008WA, SM723090009WA, SM723090010WA, SM723090022WA, SM724150085WA, SM724170105WA, SM724170106WA, SM724170107WA, SM724170108WA, SM724170109WA, SM724170110WA, SM723410023WA, SM723070034WA, SM723100090WA, SM723100091WA, SM724120041WA, SM724140032WA, SM724140033WA, SM724020029WA, SM724020030WA, SM724020031WA, SM724100015WA, SM724100016WA, SM724100017WA, SM723100048WA, SM723100049WA, SM723100050WA, SM723100054WA, SM723100055WA, SM723100057WA, SM724120057WA, SM723070046WA, SM723070047WA, SM723070048WA, SM723050040WA, SM723050041WA, SM723050042WA, SM723050043WA, SM723220033WA, SM723330011WA, SM723330014WA, SM723330016WA, SM723330017WA, SM723350005WA, SM723350006WA, SM723350007WA, SM724080025WA, SM724080026WA, SM724080027WA, SM724080028WA, SM724100010WA, SM723050033WA, SM723050034WA, SM723050035WA, SM723050036WA, SM723050037WA, SM724100014WA, SM723020048WA, SM723020049WA, SM723020050WA, SM723020051WA, SM723020052WA, SM723020053WA, SM723020055WA, SM723020056WA, SM723020057WA, SM723020058WA, SM723050050WA, SM723050051WA, SM723090023WA, SM723090024WA, SM723090025WA, SM723090026WA, SM723090027WA, SM723090028WA, SM723090029WA, SM723090030WA, SM723090031WA, SM723090032WA, SM723090033WA, SM723090034WA, SM723090035WA, SM723090038WA, SM723090039WA, SM723100014WA, SM723100015WA, SM723100016WA, SM723100020WA, SM723100021WA, SM723100047WA, SM723100053WA, SM723100056WA, SM723100089WA, SM723070054WA, SM723070078WA, SM724020014WA, SM724020015WA, SM723050038WA, SM723310008WA, SM723070052WA, SM724020025WA, SM723430029WA, SM723430030WA, SM723430031WA, SM723100097WA, SM723220029WA, SM723220032WA, SM724010031WA, SM723490028WA, SM724050027WA, SM724330003WA, SM723430028WA, SM723050039WA, SM723070039WA, SM723070040WA, SM723070049WA, SM723070050WA, SM723350048WA, SM723350049WA, SM723350050WA, SM723350051WA, SM723350052WA, SM723100029WA, SM723100030WA, SM723100031WA, SM723100098WA, SM723100100WA, SM723460072WA, SM724280008WA, SM724280009WA, SM723100087WA, SM723220030WA, SM723220034WA, SM723010011WA, SM723010012WA, SM723010013WA, SM723010014WA, SM723010015WA, SM723010016WA, SM724120042WA, SM724120043WA, SM724120044WA, SM724120045WA, SM724120046WA, SM724120047WA, SM724120048WA, SM723010021WA, SM724150054WA, SM724150055WA, SM724150056WA, SM724150057WA, SM724150058WA, SM724150059WA, SM723290001WA, SM723100061WA, SM723070051WA, SM724140021WA, SM724140022WA, SM724030001WA, SM724030002WA, SM724030003WA, SM724030004WA, SM724030005WA, SM724030006WA, SM724030007WA, SM724030008WA, SM723070056WA, SM723070057WA, SM723100017WA, SM723100018WA, SM723100019WA, SM723100022WA, SM723330015WA, SM724100055WA, SM723100066WA, SM723100067WA, SM723100068WA, SM723100072WA, SM723100073WA, SM723100075WA, SM723100080WA, SM723100081WA, SM723470121WA, SM723470122WA, SM723470123WA, SM724100045WA, SM724100046WA, SM724200018WA, SM724200019WA, SM724200020WA, SM724200021WA, SM724200022WA, SM724200023WA, SM724200024WA, SM724200025WA, SM724200026WA, SM724200027WA, SM723010022WA, SM723010023WA, SM723270002WA, SM724090007WA, SM724090009WA, SM724090010WA, SM724090011WA, SM724090012WA, SM723010017WA, SM723010018WA, SM723010019WA, SM723010020WA, SM723100084WA, SM723100085WA, SM723100086WA, SM723100088WA, SM723290095WA, SM724030042WA, SM724030043WA, SM724030044WA, SM724030047WA, SM724030048WA, SM724030049WA, SM723100062WA, SM723150002WA, SM723070035WA, SM723070036WA, SM723070037WA, SM723070038WA, SM723430004WA, SM723430005WA, SM723430006WA, SM723430007WA, SM723430008WA, SM723430009WA, SM723430010WA, SM723430011WA, SM723430012WA, SM723430013WA, SM723430014WA, SM723220031WA, SM724010029WA, SM724010030WA, SM724090005WA, SM723070041WA, SM723070042WA, SM723070043WA, SM723070044WA, SM723100099WA, SM723150001WA, SM724140031WA, SM724090006WA, SM724090008WA, SM724090013WA, SM724090014WA, SM724090015WA, SM724090016WA, SM724090017WA, SM724090018WA, SM724090019WA, SM724090020WA, SM724030019WA, SM723070079WA, SM723050002MA, SM723360074WA, SM723110009MA, SM723110010MA, SM723110011MA, SM723440012MA, SM724240031MA, SM724300002MA, SM724340007MA, SM724110055MA, SM724130013MA, SM724130014MA, SM724130015MA, SM724130016MA, SM724130017MA, SM723100004MA, SM723100005MA, SM723100006MA, SM723340001MA, SM723340002MA, SM723340003MA, SM723340004MA, SM723340005MA, SM723440019MA, SM723440020MA, SM723440021MA, SM723460059MA, SM723460060MA, SM723460061MA, SM723460062MA, SM723460083WA, SM723460084WA, SM723470006MA, SM723470007MA, SM723470008MA, SM723470009MA, SM723470010MA, SM723470011MA, SM723470012MA, SM723470013MA, SM723470014MA, SM723470015MA, SM723470016MA, SM723470017MA, SM723470018MA, SM723470019MA, SM723470020MA, SM723470021MA, SM723470022MA, SM723470023MA, SM723480003MA, SM723480004MA, SM723480005MA, SM723480006MA, SM723480007MA, SM724030023MA, SM724090021MA, SM724090022MA, SM724090023MA, SM724090024MA, SM724090025MA, SM724290007MA, SM724290008MA, SM724290009MA, SM724290010MA, SM724290011MA, SM724290012MA, SM724290013MA, SM724290014MA, SM724290015MA, SM724290016MA, SM724290017MA, SM724290018MA, SM724290019MA, SM724290020MA, SM724290021MA, SM724290022MA, SM724290023MA, SM724290024MA, SM724290025MA, SM724290026MA, SM724290027MA, SM724290028MA, SM724290029MA, SM721060060MA, SM721060061MA, SM721090010MA, SM719180017MA, SM723170011MA, SM723170012MA, SM718380034MA, SM718380035MA, SM718380036MA, SM718380038MA, SM718380039MA, SM718380040MA, SM718380043MA, SM718380044MA, SM718380045MA, SM718380046MA, SM718380047MA, SM718380048MA, SM718380049MA, SM718380050MA, SM719250008MA, SM724260018MA, SM719160012MA, SM724310012MA, SM724400008MA, SM724400009MA, SM724250026MA, SM722050066MA, SM722050067MA, SM722050068MA, SM722050069MA, SM722050070MA, SM722050071MA, SM722050072MA, SM722050073MA, SM722050074MA, SM722050075MA, SM722050076MA, SM722050077MA, SM722050079MA, SM718130075WA, SM718130076WA, SM718130077WA, SM718130078WA, SM718130079WA, SM718130081WA, SM718130082WA, SM718490023WA, SM719120022WA, SM719120023WA, SM719120024WA, SM719120025WA, SM719120026WA, SM719120027WA, SM719120028WA, SM719120029WA, SM719120030WA, SM719120031WA, SM719120032WA, SM719120033WA, SM719120034WA, SM719120038WA, SM719120039WA, SM719120040WA, SM719120041WA, SM719120042WA, SM719120043WA, SM719120044WA, SM719120045WA, SM719120046WA, SM719120047WA, SM719120048WA, SM719120049WA, SM719120050WA, SM719130009WA, SM719130015WA, SM719130016WA, SM719130017WA, SM719130018WA, SM722090017WA, SM722090018WA, SM722090019WA, SM722090020WA, SM722090021WA, SM722090022WA, SM722090023WA, SM722090024WA, SM722090025WA, SM722090026WA, SM722090027WA, SM722090028WA, SM722090029WA, SM722090030WA, SM722090031WA, SM722090032WA, SM722090033WA, SM719100015WA, SM719100016WA, SM719100017WA, SM719100018WA, SM719100019WA, SM719100020WA, SM719100021WA, SM719100022WA, SM719100026WA, SM719100027WA, SM719100028WA, SM719100029WA, SM719490001WA, SM719490002WA, SM719490003WA, SM719490004WA, SM719490005WA, SM719490006WA, SM719490007WA, SM719490008WA, SM719490009WA, SM719490011WA, SM719490012WA, SM720140106WA, SM720140107WA, SM720140108WA, SM720140109WA, SM720140110WA, SM720140111WA, SM720140112WA, SM720150001WA, SM720150002WA, SM720150003WA, SM720150004WA, SM720150005WA.
FEI Number 3003513129
Recalling Firm/
Manufacturer
GE Medical Systems China Co., Ltd.
Dev. Zone
National Hi-Tech; No. 19 Changjiang Road; Xin
Wuxi China
For Additional Information ContactGE HealthCare Service
800-437-1171
Manufacturer Reason
for Recall
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
FDA Determined
Cause 2
Device Design
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice on 03/21/2025. The notice explained the safety issue with the device, potential risk, and requested the following actions: The anesthesia system may continue to be used following these instructions: Perform the Ventilation Screening Test specified in Appendix 1 for each affected Carestation system. If the Carestation system passes the Ventilation Screening Test, you can continue to use the device. If the Carestation system fails the Ventilation Screening Test and must be used prior to the system being corrected by GE HealthCare, follow the instructions below: 1. Use the device in accordance with the instructions in the URM with these changes: i. Use only Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) mode to mechanically ventilate a patient. ii. Do not use Volume Control Ventilation (VCV) mode to mechanically ventilate a patient. NOTE: Manual mode of the anesthesia system can be used to provide manual ventilation or allow spontaneous ventilation of the patient. 2. Users are not to use VCV mode on the device until the device has been corrected by GE HealthCare. Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions and retain this document for your record. Questions: contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. Notice sent on 04/21/25 with updated instructions. Change: Attach a 2 Liter reservoir bag (original) to Attach a reservoir bag (updated)in step 5 of the Ventilation Screening Test in Appendix 1 of the letter. Notice sent on 5/18/26 added additional affected devices for GE HealthCare Urgent Medical Device Correction notification, Field Replaceable Unit (FRU) parts: DRIVER GAS CHECK VALVE ASSY, M1807901-S; Valve Assembly, 3.5cm H2O Drive Gas Check Valve BCG, 2071125-001-S; A180 Ventilator Engine Assembly, 2071117-001-S.
Quantity in Commerce4036 units
DistributionWorldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = BSZ
-
-