 Class 2 Device Recall LOGIQ P9 series with software version R4.5.7see related information
Date Initiated by FirmSeptember 18, 2025
Date PostedOctober 15, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0157-2026
Recall Event ID 97726
510(K)NumberK214039 
Product Classification System, imaging, pulsed doppler, ultrasonic - Product Code IYN
ProductLOGIQ P9 R4.5 ultrasound system with system software version R4.5.7., Model Number 5877533
Code Information Model 5877533 UDI-DI: 00195278464637, Serial Numbers: LP9490416 LP9490175 LP9490393 LP9490399 LP9490376 LP9490463 LP9443017 LP9442472 LP9443231 LP9442094 LP9443291 LP9443230 LP9440403 LP9442483 LP9440972 LP9440166 LP9440237 LP9440933 LP9440348 LP9442196 LP9440631 LP9440926 LP9441099 LP9441225 LP9441097 LP9440929 LP9441030 LP9440921 LP9441035 LP9440507 LP9440920 LP9440459 LP9440460 LP9441332 LP9442656 LP9443022 LP9440986
Recalling Firm/
Manufacturer
GE Medical Systems, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information ContactGE HealthCare Service
800-437-1171
Manufacturer Reason
for Recall
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
FDA Determined
Cause 2
Software design
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 9/18/25 was sent to customers. Actions to be taken by Customer/User For LOGIQ P9 and LOGIQ P10 series ultrasound systems with system software version R4.5.7, discontinue using the UGAP feature (please see Appendix for product identification). You may continue using other imaging features. After the correction has been implemented, please destroy the installation media for software R4.5.7 at your site. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.74084@gehealthcare.com. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce37 units
DistributionUS Nationwide. Global Distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = IYN
-
-