• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Portrait VSM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Portrait VSMsee related information
Date Initiated by FirmSeptember 26, 2025
Date PostedNovember 07, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0456-2026
Recall Event ID 97786
510(K)NumberK233810 
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
ProductPortrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms
Code Information UDI-DI 00195278561312 Serial Nos. SWP25300030SA SWP25300031SA SWP25130712SA SWP25130713SA SWP25130714SA SWP25130715SA SWP25130716SA SWP25130717SA SWP25130718SA SWP25130719SA SWP25130720SA SWP25130721SA SWP25130722SA SWP25130723SA SWP25130724SA SWP25130725SA SWP25130726SA SWP25130727SA SWP25130728SA SWP25130729SA SWP25130730SA SWP25130731SA SWP25130732SA SWP25130733SA SWP25130734SA SWP25130735SA SWP25130736SA SWP25130737SA SWP25130738SA SWP25130739SA SWP25130740SA SWP25130741SA SWP25230019SA SWP25230020SA SWP25230021SA SWP25230022SA SWP25230026SA SWP25210083SA SWP25210084SA SWP25130751SA SWP25130752SA SWP25130753SA SWP25170019SA SWP25170020SA SWP25170021SA SWP25170022SA SWP25170023SA SWP25170024SA SWP25200033SA SWP25200034SA SWP25200035SA SWP25160027SA SWP25180121SA SWP25180122SA SWP25180123SA SWP25230034SA SWP25230035SA SWP25230036SA SWP25250032SA SWP25250033SA SWP25290061SA SWP25260114SA SWP25260115SA SWP25260116SA SWP25260117SA SWP25250046SA SWP25250047SA SWP25250048SA SWP25250050SA SWP25240013SA SWP25120633SA SWP25130746SA SWP25170056SA SWP25250034SA SWP25290065SA SWP25051286SA SWP25110533SA SWP25190088SA SWP25210044SA SWP25260155SA SWP25260156SA SWP25260157SA SWP25260158SA SWP25120623SA SWP25120627SA SWP25130766SA SWP25130843SA SWP25130844SA SWP25151063SA SWP25151071SA SWP25151087SA SWP25170027SA SWP25170028SA SWP25290048SA SWP25290053SA SWP25110399SA SWP25151065SA SWP25180130SA SWP25260120SA SWP25151070SA SWP25180126SA SWP25180127SA SWP25220010SA SWP25300001SA SWP25300002SA SWP25300003SA SWP25300004SA SWP25300005SA SWP25300006SA SWP25300007SA SWP25300008SA SWP25210078SA SWP25220050SA SWP25190090SA SWP25190091SA SWP25190092SA SWP25220063SA SWP25120626SA SWP25210082SA SWP25110389SA SWP25110390SA SWP25120616SA SWP25120617SA SWP25130744SA SWP25130745SA SWP25160028SA SWP25180128SA SWP25220023SA SWP25220024SA SWP25220025SA SWP25220026SA SWP25220027SA SWP25220028SA SWP25220029SA SWP25220030SA SWP25220031SA SWP25220032SA SWP25220034SA SWP25220035SA SWP25220036SA SWP25220037SA SWP25220038SA SWP25220039SA SWP25220040SA SWP25220041SA SWP25220042SA SWP25220043SA SWP25220044SA SWP25220046SA SWP25220047SA SWP25220048SA SWP25220067SA SWP25270075SA SWP25270076SA SWP25130749SA SWP25130750SA SWP25051387SA SWP25051388SA SWP25051389SA SWP25051390SA SWP25090172SA SWP25090173SA SWP25141031SA SWP25141032SA SWP25220018SA SWP25220019SA SWP25220020SA SWP25280039SA SWP25280040SA SWP25200030SA SWP25200031SA SWP25200032SA SWP25090190SA SWP25090191SA SWP25090192SA SWP25090193SA SWP25100013SA SWP25170048SA SWP25250021SA SWP25250022SA SWP25040958SA SWP25040959SA SWP25200017SA SWP25200018SA SWP25200019SA SWP25200020SA SWP25200021SA SWP25200022SA SWP25051391SA SWP25120622SA SWP25230059SA SWP25230060SA SWP25270073SA SWP25280001SA SWP25280002SA SWP25280003SA SWP25280004SA SWP25280005SA SWP25280006SA SWP25280007SA SWP25280008SA SWP25280009SA SWP25280013SA SWP25280014SA SWP25280015SA SWP25280016SA SWP25280017SA SWP25160001SA SWP25160002SA SWP25160003SA SWP25160004SA SWP25160005SA SWP25160006SA SWP25160007SA SWP25160008SA SWP25160009SA SWP25160010SA SWP25160011SA SWP25160012SA SWP25160013SA SWP25160014SA SWP25160015SA SWP25160016SA SWP25160017SA SWP25160019SA SWP25160020SA SWP25160021SA SWP25160022SA SWP25160023SA SWP25160024SA SWP25220049SA SWP25310005SA SWP25040956SA SWP25260102SA SWP25260118SA SWP25290055SA SWP25151080SA SWP25151081SA SWP25270071SA SWP25220065SA SWP25151102SA SWP25151103SA SWP25151104SA SWP25151105SA SWP25151106SA SWP25151107SA SWP25210036SA SWP25210037SA SWP25210038SA SWP25210039SA SWP25210040SA SWP25210041SA SWP25210042SA SWP25210043SA SWP25151088SA SWP25151089SA SWP25151090SA SWP25151091SA SWP25151092SA SWP25151093SA SWP25151094SA SWP25151095SA SWP25151096SA SWP25151097SA SWP25151098SA SWP25151099SA SWP25151100SA SWP25151101SA SWP25110529SA SWP25110530SA SWP25110531SA SWP25120632SA SWP25300062SA SWP25300063SA SWP25130669SA SWP25130670SA SWP25170057SA SWP25240001SA SWP25240002SA SWP25130682SA SWP25130683SA SWP25130684SA SWP25130685SA SWP25130686SA SWP25130687SA SWP25130688SA SWP25130689SA SWP25130690SA SWP25130691SA SWP25130692SA SWP25130693SA SWP25130694SA SWP25130695SA SWP25040965SA SWP25040966SA SWP25040967SA SWP25040968SA SWP25040969SA SWP25040970SA SWP25040971SA SWP25040972SA SWP25040973SA SWP25040974SA SWP25040975SA SWP25040976SA SWP25040977SA SWP25040978SA SWP25040979SA SWP25040980SA SWP25040981SA SWP25040982SA SWP25040983SA SWP25040984SA SWP25040985SA SWP25040986SA SWP25040987SA SWP25040988SA SWP25040989SA SWP25040990SA SWP25040991SA SWP25040992SA SWP25040993SA SWP25040994SA SWP25040995SA SWP25040996SA SWP25040997SA SWP25040998SA SWP25040999SA SWP25041000SA SWP25041001SA SWP25041002SA SWP25041003SA SWP25041004SA SWP25041005SA SWP25041006SA SWP25041007SA SWP25041008SA SWP25041009SA SWP25041010SA SWP25041011SA SWP25041012SA SWP25041013SA SWP25041014SA SWP25041015SA SWP25041016SA SWP25041017SA SWP25041018SA SWP25041019SA SWP25041020SA SWP25041021SA SWP25041022SA SWP25041023SA SWP25041024SA SWP25041025SA SWP25041026SA SWP25041027SA SWP25041028SA SWP25041029SA SWP25041030SA SWP25041031SA SWP25041032SA SWP25041033SA SWP25041034SA SWP25041035SA SWP25041036SA SWP25041037SA SWP25041038SA SWP25041039SA SWP25041040SA SWP25041041SA SWP25041042SA SWP25041043SA SWP25041044SA SWP25041045SA SWP25041046SA SWP25041047SA SWP25041048SA SWP25041049SA SWP25041050SA SWP25041051SA SWP25041052SA SWP25041053SA SWP25041054SA SWP25041055SA SWP25041056SA SWP25041057SA SWP25041058SA SWP25041059SA SWP25041060SA SWP25041061SA SWP25041062SA SWP25041063SA SWP25041064SA SWP25041065SA SWP25041066SA SWP25041067SA SWP25041068SA SWP25041069SA SWP25041070SA SWP25041071SA SWP25041072SA SWP25041073SA SWP25051074SA SWP25051075SA SWP25051076SA SWP25051077SA SWP25051078SA SWP25051079SA SWP25051080SA SWP25051081SA SWP25051082SA SWP25051083SA SWP25051084SA SWP25051085SA SWP25051086SA SWP25051087SA SWP25051088SA SWP25051089SA SWP25051090SA SWP25051091SA SWP25051092SA SWP25051093SA SWP25051094SA SWP25051095SA SWP25051096SA SWP25051097SA SWP25051098SA SWP25051099SA SWP25051100SA SWP25051101SA SWP25051102SA SWP25051103SA SWP25051104SA SWP25051105SA SWP25051106SA SWP25051107SA SWP25051108SA SWP25051109SA SWP25051110SA SWP25051111SA SWP25051112SA SWP25051113SA SWP25051114SA SWP25051115SA SWP25051116SA SWP25051117SA SWP25051118SA SWP25051119SA SWP25051120SA SWP25051121SA SWP25051122SA SWP25051123SA SWP25051124SA SWP25051125SA SWP25051126SA SWP25051127SA SWP25051128SA SWP25051129SA SWP25051130SA SWP25051131SA SWP25051132SA SWP25051133SA SWP25051134SA SWP25051135SA SWP25051136SA SWP25051137SA SWP25051138SA SWP25051139SA SWP25051140SA SWP25051141SA SWP25051142SA SWP25051143SA SWP25051144SA SWP25051145SA SWP25051146SA SWP25051147SA SWP25051148SA SWP25051149SA SWP25051150SA SWP25051151SA SWP25051152SA SWP25051153SA SWP25051154SA SWP25051155SA SWP25051156SA SWP25051158SA SWP25051159SA SWP25051160SA SWP25051161SA SWP25051162SA SWP25051182SA SWP25051183SA SWP25051188SA SWP25051189SA SWP25051190SA SWP25051206SA SWP25051208SA SWP25051209SA SWP25051210SA SWP25051219SA SWP25051220SA SWP25080006SA SWP25080007SA SWP25080008SA SWP25080009SA SWP25080010SA SWP25080011SA SWP25080012SA SWP25080015SA SWP25080016SA SWP25080017SA SWP25080018SA SWP25080019SA SWP25080020SA SWP25080021SA SWP25080022SA SWP25080023SA SWP25080033SA SWP25080034SA SWP25080035SA SWP25080049SA SWP25080056SA SWP25080057SA SWP25080059SA SWP25080060SA SWP25080061SA SWP25080062SA SWP25080063SA SWP25080065SA SWP25080105SA SWP25080106SA SWP25080107SA SWP25080108SA SWP25080109SA SWP25080110SA SWP25080111SA SWP25080112SA SWP25080113SA SWP25080114SA SWP25080115SA SWP25080116SA SWP25080117SA SWP25080118SA SWP25080119SA SWP25080120SA SWP25080121SA SWP25080122SA SWP25080123SA SWP25080124SA SWP25080125SA SWP25080126SA SWP25080127SA SWP25080128SA SWP25080129SA SWP25080130SA SWP25080131SA SWP25080132SA SWP25080133SA SWP25080134SA SWP25080135SA SWP25080136SA SWP25080137SA SWP25080138SA SWP25080139SA SWP25080140SA SWP25080141SA SWP25080142SA SWP25080143SA SWP25080144SA SWP25080145SA SWP25080146SA SWP25080147SA SWP25080148SA SWP25080149SA SWP25080150SA SWP25080152SA SWP25080153SA SWP25080154SA SWP25080155SA SWP25080156SA SWP25080157SA SWP25080158SA SWP25080159SA SWP25080160SA SWP25080161SA SWP25080162SA SWP25080163SA SWP25080164SA SWP25080165SA SWP25080166SA SWP25080167SA SWP25080168SA SWP25080169SA SWP25110381SA SWP25110382SA SWP25110383SA SWP25110384SA SWP25110385SA SWP25110386SA SWP25110387SA SWP25110388SA SWP25110401SA SWP25110402SA SWP25110403SA SWP25110404SA SWP25110405SA SWP25110406SA SWP25110407SA SWP25110408SA SWP25110409SA SWP25110410SA SWP25110411SA SWP25110413SA SWP25110414SA SWP25110415SA SWP25110416SA SWP25110417SA SWP25110418SA SWP25110419SA SWP25110420SA SWP25110421SA SWP25110422SA SWP25110423SA SWP25110424SA SWP25110425SA SWP25110426SA SWP25110427SA SWP25110428SA SWP25110429SA SWP25110430SA SWP25110431SA SWP25110432SA SWP25110433SA SWP25110434SA SWP25110435SA SWP25110436SA SWP25110437SA SWP25110438SA SWP25110439SA SWP25110440SA SWP25110441SA SWP25110442SA SWP25110443SA SWP25110444SA SWP25110445SA SWP25110446SA SWP25110447SA SWP25110448SA SWP25110450SA SWP25110451SA SWP25110452SA SWP25110453SA SWP25110454SA SWP25110455SA SWP25110456SA SWP25110457SA SWP25110458SA SWP25110459SA SWP25110460SA SWP25110461SA SWP25110462SA SWP25110463SA SWP25110464SA SWP25110465SA SWP25110466SA SWP25110493SA SWP25110494SA SWP25110495SA SWP25110496SA SWP25110497SA SWP25110498SA SWP25110499SA SWP25110500SA SWP25110501SA SWP25110502SA SWP25110504SA SWP25110505SA SWP25110506SA SWP25110507SA SWP25110508SA SWP25110509SA SWP25110510SA SWP25110511SA SWP25110512SA SWP25110513SA SWP25110514SA SWP25110515SA SWP25110516SA SWP25110517SA SWP25110518SA SWP25130651SA SWP25130652SA SWP25130653SA SWP25130654SA SWP25130655SA SWP25130656SA SWP25130657SA SWP25130658SA SWP25130659SA SWP25130660SA SWP25130661SA SWP25130662SA SWP25130663SA SWP25130664SA SWP25130665SA SWP25130666SA SWP25130667SA SWP25130668SA SWP25160029SA SWP25160030SA SWP25080014SA SWP25080024SA SWP25080025SA SWP25080026SA SWP25080027SA SWP25080028SA SWP25080029SA SWP25080030SA SWP25080031SA SWP25080032SA SWP25151051SA SWP25151052SA SWP25151053SA SWP25170040SA SWP25170041SA SWP25190001SA SWP25190002SA SWP25190003SA SWP25190004SA SWP25190005SA SWP25190006SA SWP25190007SA SWP25190008SA SWP25190009SA SWP25190010SA SWP25190011SA SWP25190012SA SWP25190013SA SWP25190015SA SWP25190016SA SWP25190017SA SWP25190018SA SWP25190019SA SWP25190021SA SWP25190027SA SWP25190028SA SWP25190029SA SWP25190030SA SWP25190031SA SWP25190032SA SWP25190033SA SWP25190034SA SWP25190035SA SWP25190036SA SWP25190037SA SWP25190039SA SWP25190040SA SWP25190041SA SWP25190043SA SWP25190045SA SWP25190047SA SWP25190048SA SWP25190049SA SWP25190050SA SWP25190052SA SWP25190053SA SWP25190054SA SWP25190055SA SWP25190056SA SWP25190057SA SWP25190058SA SWP25190059SA SWP25190060SA SWP25210070SA SWP25280041SA SWP25280042SA SWP25190102SA SWP25190103SA SWP25220005SA SWP25220007SA SWP25260159SA SWP25280031SA SWP25120631SA SWP25151039SA SWP25151040SA SWP25151041SA SWP25151042SA SWP25151043SA SWP25151044SA SWP25151045SA SWP25151046SA SWP25270086SA SWP25270087SA SWP25270088SA SWP25141038SA SWP25051283SA SWP25250025SA SWP25250026SA SWP25250027SA SWP25250028SA SWP25250029SA SWP25230028SA SWP25300019SA SWP25300020SA SWP25300021SA SWP25300022SA SWP25300023SA SWP25300024SA SWP25170038SA SWP25170039SA SWP25250053SA SWP25240023SA SWP25240024SA SWP25240025SA SWP25240026SA SWP25240027SA SWP25240028SA SWP25240029SA SWP25240030SA SWP25240031SA SWP25240032SA SWP25240033SA SWP25240034SA SWP25240035SA SWP25240036SA SWP25130845SA SWP25130846SA SWP25130854SA SWP25130855SA SWP25141019SA SWP25141020SA SWP25141021SA SWP25141022SA SWP25141023SA SWP25141024SA SWP25141025SA SWP25141026SA SWP25141027SA SWP25141028SA SWP25141029SA SWP25141030SA SWP25210005SA SWP25210006SA SWP25210007SA SWP25210008SA SWP25210009SA SWP25210010SA SWP25210011SA SWP25210012SA SWP25210013SA SWP25210014SA SWP25210015SA SWP25210016SA SWP25210017SA SWP25210018SA SWP25210019SA SWP25210020SA SWP25210021SA SWP25210022SA SWP25210023SA SWP25210024SA SWP25210025SA SWP25210026SA SWP25210027SA SWP25210028SA SWP25210029SA SWP25210030SA SWP25210031SA SWP25210032SA SWP25210033SA SWP25210034SA SWP25210081SA SWP25290063SA SWP25230027SA SWP25250038SA SWP25250039SA SWP25250040SA SWP25170025SA SWP25170026SA SWP25280035SA SWP25280036SA SWP25300066SA SWP25200024SA SWP25200028SA SWP25151108SA SWP25151109SA SWP25170001SA SWP25170002SA SWP25170003SA SWP25170004SA SWP25170005SA SWP25170006SA SWP25170007SA SWP25170008SA SWP25170009SA SWP25170010SA SWP25170011SA SWP25170012SA SWP25170013SA SWP25220071SA SWP25130840SA SWP25130841SA SWP25130842SA SWP25090142SA SWP25090143SA SWP25090144SA SWP25090145SA SWP25090146SA SWP25090147SA SWP25090148SA SWP25090149SA SWP25090150SA SWP25090151SA SWP25090152SA SWP25090153SA SWP25090154SA SWP25090155SA SWP25090156SA SWP25090157SA SWP25090158SA SWP25090159SA SWP25090160SA SWP25090161SA SWP25090162SA SWP25090163SA SWP25090164SA SWP25090165SA SWP25090166SA SWP25090167SA SWP25090168SA SWP25090169SA SWP25120636SA SWP25151059SA SWP25151060SA SWP25151061SA SWP25151062SA SWP25240012SA SWP25040960SA SWP25040961SA SWP25061397SA SWP25200036SA SWP25200037SA SWP25140935SA SWP25140936SA SWP25140937SA SWP25250057SA SWP25250058SA SWP25061395SA SWP25061396SA SWP25080103SA SWP25170047SA SWP25230015SA SWP25230016SA SWP25230017SA SWP25100012SA SWP25130768SA SWP25130847SA SWP25260119SA SWP25220051SA SWP25220052SA SWP25220053SA SWP25220054SA SWP25220055SA SWP25220056SA SWP25220057SA SWP25220058SA SWP25220059SA SWP25220060SA SWP25220061SA SWP25220062SA SWP25170055SA SWP25240009SA SWP25240010SA SWP25240011SA SWP25300032SA SWP25300033SA SWP25300034SA SWP25300035SA SWP25151066SA SWP25151068SA SWP25151072SA SWP25190104SA SWP25250051SA SWP25250054SA SWP25250055SA SWP25250056SA SWP25190107SA SWP25090171SA SWP25130743SA SWP25141037SA SWP25260097SA SWP25280030SA SWP25280033SA SWP25280034SA SWP25310001SA SWP25170050SA SWP25130671SA SWP25130672SA SWP25130673SA SWP25130674SA SWP25130675SA SWP25130676SA SWP25250049SA SWP25260162SA SWP25170031SA SWP25210069SA SWP25250030SA SWP25250031SA SWP25300054SA SWP25300055SA SWP25300056SA SWP25300057SA SWP25300058SA SWP25300059SA SWP25300060SA SWP25300061SA SWP25270084SA SWP25270085SA SWP25051290SA SWP25051291SA SWP25170042SA SWP25170043SA SWP25230001SA SWP25230002SA SWP25230003SA SWP25151069SA SWP25210080SA SWP25220021SA SWP25220022SA SWP25250017SA SWP25250018SA SWP25250019SA SWP25250020SA SWP25250023SA SWP25250024SA SWP25120629SA SWP25120642SA SWP25270079SA SWP25270080SA SWP25200041SA SWP25210001SA SWP25210002SA SWP25240016SA SWP25250041SA SWP25270070SA SWP25110467SA SWP25110468SA SWP25110469SA SWP25110470SA SWP25110471SA SWP25110472SA SWP25110473SA SWP25110474SA SWP25110475SA SWP25110476SA SWP25110477SA SWP25130770SA SWP25130771SA SWP25130772SA SWP25130773SA SWP25130774SA SWP25130775SA SWP25130776SA SWP25130777SA SWP25130778SA SWP25130779SA SWP25130780SA SWP25130781SA SWP25130782SA SWP25130783SA SWP25130784SA SWP25130785SA SWP25130786SA SWP25130787SA SWP25130788SA SWP25130789SA SWP25130790SA SWP25130791SA SWP25130792SA SWP25130793SA SWP25130794SA SWP25130795SA SWP25130796SA SWP25130797SA SWP25130798SA SWP25130799SA SWP25130800SA SWP25130801SA SWP25130802SA SWP25130803SA SWP25130804SA SWP25130805SA SWP25130806SA SWP25130807SA SWP25130808SA SWP25130809SA SWP25130810SA SWP25130811SA SWP25130812SA SWP25130813SA SWP25130814SA SWP25130815SA SWP25130816SA SWP25130817SA SWP25130818SA SWP25130819SA SWP25130851SA SWP25130852SA SWP25130853SA SWP25140942SA SWP25140943SA SWP25140944SA SWP25140946SA SWP25140947SA SWP25140948SA SWP25140949SA SWP25140953SA SWP25140954SA SWP25140959SA SWP25140960SA SWP25140961SA SWP25140962SA SWP25140963SA SWP25140965SA SWP25140966SA SWP25140967SA SWP25140991SA SWP25140992SA SWP25140993SA SWP25180001SA SWP25180002SA SWP25180003SA SWP25180004SA SWP25180005SA SWP25180006SA SWP25180007SA SWP25180008SA SWP25180009SA SWP25180010SA SWP25180011SA SWP25180012SA SWP25180013SA SWP25180014SA SWP25180015SA SWP25180016SA SWP25180017SA SWP25180018SA SWP25180019SA SWP25180020SA SWP25180021SA SWP25180022SA SWP25180023SA SWP25180024SA SWP25180025SA SWP25180026SA SWP25180028SA SWP25180029SA SWP25180030SA SWP25180031SA SWP25180032SA SWP25180033SA SWP25180034SA SWP25180035SA SWP25180036SA SWP25180037SA SWP25180038SA SWP25180039SA SWP25180040SA SWP25180041SA SWP25180042SA SWP25180043SA SWP25180044SA SWP25180045SA SWP25180046SA SWP25180047SA SWP25180048SA SWP25180049SA SWP25180050SA SWP25180134SA SWP25180135SA SWP25180152SA SWP25180153SA SWP25180156SA SWP25180157SA SWP25180158SA SWP25180159SA SWP25180161SA SWP25180162SA SWP25180163SA SWP25180164SA SWP25180165SA SWP25180166SA SWP25180167SA SWP25180168SA SWP25180169SA SWP25180170SA SWP25180171SA SWP25180172SA SWP25180173SA SWP25180174SA SWP25180175SA SWP25180176SA SWP25180177SA SWP25180178SA SWP25180179SA SWP25180180SA SWP25180181SA SWP25180182SA SWP25120588SA SWP25120589SA SWP25120604SA SWP25120605SA SWP25230061SA SWP25230062SA SWP25230053SA SWP25230054SA SWP25230055SA SWP25230056SA SWP25230057SA SWP25230058SA SWP25120624SA SWP25120625SA SWP25120640SA SWP25120641SA SWP25110483SA SWP25110480SA SWP25110481SA SWP25110482SA SWP25170049SA SWP25240019SA SWP25240020SA SWP25290043SA SWP25290044SA SWP25290045SA SWP25061404SA SWP25120637SA SWP25090194SA SWP25090195SA SWP25090196SA SWP25090197SA SWP25151058SA SWP25180101SA SWP25180102SA SWP25180103SA SWP25180104SA SWP25180105SA SWP25180106SA SWP25180107SA SWP25180108SA SWP25180109SA SWP25180110SA SWP25180111SA SWP25180112SA SWP25180113SA SWP25180114SA SWP25180115SA SWP25180116SA SWP25180117SA SWP25190086SA SWP25190087SA SWP25210077SA SWP25260164SA SWP25051382SA SWP25051383SA SWP25051384SA SWP25051385SA SWP25051386SA SWP25051392SA SWP25051393SA SWP25051394SA SWP25061398SA SWP25061399SA SWP25061400SA SWP25061401SA SWP25080093SA SWP25080102SA SWP25100007SA SWP25100008SA SWP25100009SA SWP25100014SA SWP25110400SA SWP25110484SA SWP25120534SA SWP25120535SA SWP25120536SA SWP25120537SA SWP25120538SA SWP25120539SA SWP25120540SA SWP25120541SA SWP25120542SA SWP25120543SA SWP25120544SA SWP25120545SA SWP25120546SA SWP25120547SA SWP25120548SA SWP25120549SA SWP25120550SA SWP25120551SA SWP25120552SA SWP25120553SA SWP25120554SA SWP25120555SA SWP25120556SA SWP25120557SA SWP25120558SA SWP25120559SA SWP25120560SA SWP25120561SA SWP25120562SA SWP25120563SA SWP25120564SA SWP25120565SA SWP25120566SA SWP25120567SA SWP25120568SA SWP25120569SA SWP25120570SA SWP25120571SA SWP25120572SA SWP25120573SA SWP25120574SA SWP25120575SA SWP25120576SA SWP25120577SA SWP25120578SA SWP25120579SA SWP25120580SA SWP25120581SA SWP25120582SA SWP25120583SA SWP25120584SA SWP25120585SA SWP25130848SA SWP25170036SA SWP25170037SA SWP25170054SA SWP25180118SA SWP25180119SA SWP25180120SA SWP25180129SA SWP25200040SA SWP25260163SA SWP25310002SA SWP25260103SA SWP25260104SA SWP25230004SA SWP25230005SA SWP25230006SA SWP25230007SA SWP25230009SA SWP25230010SA SWP25090170SA SWP25110532SA SWP25290049SA SWP25290050SA SWP25051221SA SWP25051222SA SWP25051223SA SWP25051224SA SWP25051225SA SWP25051226SA SWP25051227SA SWP25051228SA SWP25051229SA SWP25051230SA SWP25051231SA SWP25051232SA SWP25051233SA SWP25051234SA SWP25051235SA SWP25051236SA SWP25051237SA SWP25051238SA SWP25051239SA SWP25051240SA SWP25051241SA SWP25051242SA SWP25051244SA SWP25051245SA SWP25051246SA SWP25051247SA SWP25051248SA SWP25051249SA SWP25051250SA SWP25051251SA SWP25051252SA SWP25051253SA SWP25051254SA SWP25051255SA SWP25051256SA SWP25051257SA SWP25051258SA SWP25051259SA SWP25051260SA SWP25051261SA SWP25051262SA SWP25051263SA SWP25051264SA SWP25051265SA SWP25051266SA SWP25051267SA SWP25051268SA SWP25051269SA SWP25051270SA SWP25051271SA SWP25051272SA SWP25051273SA SWP25051274SA SWP25051275SA SWP25051276SA SWP25051277SA SWP25051278SA SWP25051279SA SWP25051280SA SWP25051281SA SWP25051282SA SWP25130765SA SWP25210035SA SWP25110395SA SWP25110396SA SWP25110397SA SWP25110398SA SWP25130704SA SWP25130705SA SWP25130706SA SWP25130707SA SWP25130701SA SWP25130702SA SWP25130703SA SWP25160031SA SWP25240017SA SWP25240018SA SWP25130700SA SWP25300037SA SWP25170014SA SWP25170015SA SWP25170016SA SWP25170017SA SWP25170018SA SWP25160032SA SWP25270074SA SWP25120620SA SWP25120621SA SWP25190061SA SWP25190062SA SWP25190063SA SWP25190064SA SWP25190065SA SWP25190066SA SWP25190067SA SWP25190068SA SWP25190069SA SWP25190070SA SWP25190071SA SWP25190072SA SWP25190073SA SWP25190074SA SWP25190075SA SWP25190076SA SWP25190077SA SWP25190078SA SWP25100286SA SWP25250001SA SWP25250002SA SWP25250003SA SWP25250004SA SWP25250005SA SWP25250006SA SWP25250007SA SWP25250008SA SWP25250009SA SWP25250010SA SWP25250011SA SWP25250012SA SWP25250013SA SWP25250014SA SWP25250015SA SWP25250016SA SWP25230029SA SWP25110394SA SWP25100287SA SWP25120618SA SWP25120619SA SWP25120638SA SWP25120639SA SWP25130759SA SWP25130760SA SWP25130762SA SWP25130763SA SWP25200038SA SWP25230063SA SWP25210073SA SWP25210074SA SWP25210075SA SWP25290054SA SWP25260151SA SWP25190093SA SWP25100015SA SWP25151084SA SWP25151085SA SWP25220012SA SWP25220013SA SWP25120634SA SWP25120635SA SWP25210067SA SWP25210068SA SWP25220045SA SWP25250036SA SWP25250037SA SWP25051288SA SWP25051289SA SWP25290060SA SWP25220069SA SWP25270099SA SWP25140863SA SWP25140864SA SWP25140865SA SWP25140866SA SWP25140867SA SWP25140868SA SWP25140869SA SWP25140870SA SWP25140871SA SWP25140872SA SWP25140873SA SWP25140874SA SWP25140875SA SWP25140876SA SWP25140877SA SWP25140878SA SWP25140879SA SWP25140880SA SWP25140881SA SWP25140882SA SWP25140883SA SWP25140884SA SWP25140885SA SWP25140886SA SWP25140887SA SWP25140888SA SWP25140889SA SWP25140890SA SWP25090198SA SWP25090199SA SWP25090200SA SWP25090201SA SWP25120591SA SWP25120592SA SWP25120593SA SWP25120594SA SWP25210088SA SWP25210089SA SWP25210090SA SWP25240014SA SWP25240015SA SWP25260150SA SWP25170029SA SWP25210057SA SWP25210058SA SWP25210059SA SWP25210060SA SWP25210064SA SWP25210065SA SWP25210066SA SWP25220014SA SWP25220015SA SWP25230018SA SWP25300009SA SWP25300010SA SWP25300011SA SWP25300012SA SWP25300013SA SWP25300014SA SWP25300015SA SWP25300016SA SWP25300017SA SWP25300018SA SWP25130757SA SWP25130758SA SWP25051381SA SWP25100016SA SWP25100017SA SWP25100018SA SWP25100019SA SWP25100020SA SWP25100021SA SWP25100022SA SWP25100024SA SWP25100025SA SWP25100026SA SWP25100027SA SWP25100028SA SWP25100029SA SWP25100030SA SWP25100031SA SWP25100032SA SWP25100033SA SWP25100034SA SWP25100035SA SWP25100036SA SWP25100037SA SWP25100038SA SWP25100039SA SWP25100040SA SWP25100041SA SWP25100042SA SWP25100043SA SWP25100044SA SWP25100045SA SWP25100046SA SWP25100047SA SWP25100048SA SWP25100049SA SWP25100050SA SWP25100051SA SWP25100052SA SWP25100053SA SWP25100054SA SWP25100055SA SWP25100056SA SWP25100057SA SWP25100058SA SWP25100059SA SWP25100060SA SWP25100061SA SWP25100062SA SWP25100063SA SWP25100064SA SWP25100065SA SWP25100066SA SWP25100067SA SWP25100068SA SWP25100069SA SWP25100070SA SWP25100071SA SWP25100072SA SWP25100073SA SWP25100074SA SWP25100075SA SWP25100076SA SWP25100077SA SWP25100078SA SWP25100079SA SWP25100080SA SWP25100081SA SWP25100082SA SWP25100083SA SWP25100084SA SWP25100085SA SWP25100086SA SWP25100087SA SWP25100088SA SWP25100089SA SWP25100090SA SWP25100091SA SWP25100092SA SWP25100093SA SWP25100094SA SWP25100095SA SWP25100096SA SWP25100097SA SWP25100098SA SWP25100099SA SWP25100100SA SWP25100101SA SWP25100102SA SWP25100103SA SWP25100104SA SWP25100105SA SWP25100106SA SWP25100107SA SWP25100108SA SWP25100109SA SWP25100110SA SWP25100111SA SWP25100112SA SWP25100113SA SWP25100114SA SWP25100115SA SWP25100116SA SWP25100117SA SWP25100118SA SWP25100119SA SWP25100120SA SWP25100121SA SWP25100122SA SWP25100123SA SWP25100124SA SWP25100125SA SWP25100126SA SWP25100127SA SWP25100128SA SWP25100129SA SWP25100130SA SWP25100131SA SWP25100132SA SWP25100133SA SWP25100134SA SWP25100135SA SWP25100136SA SWP25100137SA SWP25100138SA SWP25100139SA SWP25100140SA SWP25100141SA SWP25100142SA SWP25100143SA SWP25100144SA SWP25100145SA SWP25100146SA SWP25100147SA SWP25100148SA SWP25100149SA SWP25100150SA SWP25100151SA SWP25100152SA SWP25100153SA SWP25100154SA SWP25100155SA SWP25100156SA SWP25100157SA SWP25100158SA SWP25100159SA SWP25100160SA SWP25100161SA SWP25100162SA SWP25100163SA SWP25100164SA SWP25100165SA SWP25100166SA SWP25100167SA SWP25100168SA SWP25100169SA SWP25100170SA SWP25100171SA SWP25100172SA SWP25100174SA SWP25100175SA SWP25100176SA SWP25100177SA SWP25100178SA SWP25100179SA SWP25100180SA SWP25100181SA SWP25100182SA SWP25100183SA SWP25100184SA SWP25100185SA SWP25100186SA SWP25100187SA SWP25100188SA SWP25100189SA SWP25100190SA SWP25100191SA SWP25100192SA SWP25100193SA SWP25100194SA SWP25100195SA SWP25100196SA SWP25100197SA SWP25100198SA SWP25100199SA SWP25100200SA SWP25100201SA SWP25100202SA SWP25100203SA SWP25100204SA SWP25100205SA SWP25100206SA SWP25100207SA SWP25100208SA SWP25100209SA SWP25100210SA SWP25100211SA SWP25100212SA SWP25100213SA SWP25100214SA SWP25100215SA SWP25100216SA SWP25100217SA SWP25100218SA SWP25100219SA SWP25100220SA SWP25100221SA SWP25100222SA SWP25100223SA SWP25100224SA SWP25100225SA SWP25100226SA SWP25100227SA SWP25100228SA SWP25100229SA SWP25100230SA SWP25100231SA SWP25100232SA SWP25100233SA SWP25100234SA SWP25100235SA SWP25100236SA SWP25100237SA SWP25100238SA SWP25100239SA SWP25100240SA SWP25100241SA SWP25100242SA SWP25100243SA SWP25100244SA SWP25100245SA SWP25100246SA SWP25100247SA SWP25100248SA SWP25100249SA SWP25100250SA SWP25100251SA SWP25100252SA SWP25100253SA SWP25100254SA SWP25100255SA SWP25100256SA SWP25100257SA SWP25100258SA SWP25100259SA SWP25100260SA SWP25100261SA SWP25100262SA SWP25100263SA SWP25100264SA SWP25100265SA SWP25100266SA SWP25100267SA SWP25100268SA SWP25100269SA SWP25100270SA SWP25100271SA SWP25100272SA SWP25100273SA SWP25100274SA SWP25100275SA SWP25100276SA SWP25100277SA SWP25100278SA SWP25100279SA SWP25100280SA SWP25100281SA SWP25100282SA SWP25100283SA SWP25100284SA SWP25100285SA SWP25110293SA SWP25110294SA SWP25110295SA SWP25110296SA SWP25110297SA SWP25110298SA SWP25110299SA SWP25110300SA SWP25110301SA SWP25110302SA SWP25110303SA SWP25110304SA SWP25110305SA SWP25110306SA SWP25110307SA SWP25110308SA SWP25110309SA SWP25110310SA SWP25110311SA SWP25110312SA SWP25110313SA SWP25110314SA SWP25110315SA SWP25110316SA SWP25110317SA SWP25110318SA SWP25110319SA SWP25110320SA SWP25110321SA SWP25110322SA SWP25110323SA SWP25110324SA SWP25110325SA SWP25110326SA SWP25110327SA SWP25110328SA SWP25110329SA SWP25110330SA SWP25110331SA SWP25110332SA SWP25110333SA SWP25110334SA SWP25110335SA SWP25110336SA SWP25110337SA SWP25110338SA SWP25110339SA SWP25110340SA SWP25110341SA SWP25110342SA SWP25110343SA SWP25110344SA SWP25110345SA SWP25110346SA SWP25110347SA SWP25110348SA SWP25110349SA SWP25110350SA SWP25110351SA SWP25110352SA SWP25110353SA SWP25110354SA SWP25110355SA SWP25110356SA SWP25110358SA SWP25110359SA SWP25110360SA SWP25110361SA SWP25110362SA SWP25110363SA SWP25110364SA SWP25110365SA SWP25110366SA SWP25110367SA SWP25110368SA SWP25110369SA SWP25110370SA SWP25110371SA SWP25110372SA SWP25110373SA SWP25110485SA SWP25110486SA SWP25110487SA SWP25180124SA SWP25180125SA SWP25270100SA SWP25270101SA SWP25270102SA SWP25270103SA SWP25290052SA SWP25061405SA SWP25210071SA SWP25210072SA SWP25300036SA SWP25270082SA SWP25270083SA SWP25210079SA SWP25110488SA SWP25110489SA SWP25110490SA SWP25110491SA SWP25110492SA SWP25130696SA SWP25130697SA SWP25130698SA SWP25130699SA SWP25230064SA SWP25190098SA SWP25190099SA SWP25190100SA SWP25190101SA SWP25080170SA SWP25080171SA SWP25080172SA SWP25080173SA SWP25080174SA SWP25080175SA SWP25080176SA SWP25080177SA SWP25080178SA SWP25080179SA SWP25080180SA SWP25080181SA SWP25080182SA SWP25080183SA SWP25080184SA SWP25080185SA SWP25080186SA SWP25080187SA SWP25080188SA SWP25080189SA SWP25080190SA SWP25080191SA SWP25080192SA SWP25080193SA SWP25080194SA SWP25080195SA SWP25080196SA SWP25080197SA SWP25080198SA SWP25080199SA SWP25080200SA SWP25080201SA SWP25080202SA SWP25080203SA SWP25080204SA SWP25080205SA SWP25080206SA SWP25080207SA SWP25080208SA SWP25080209SA SWP25080210SA SWP25080211SA SWP25080212SA SWP25080213SA SWP25080214SA SWP25080215SA SWP25080216SA SWP25080217SA SWP25080218SA SWP25080219SA SWP25080220SA SWP25080221SA SWP25080222SA SWP25080223SA SWP25080224SA SWP25080225SA SWP25080226SA SWP25080227SA SWP25080228SA SWP25080229SA SWP25080230SA SWP25080231SA SWP25080232SA SWP25080233SA SWP25080234SA SWP25080235SA SWP25080236SA SWP25080237SA SWP25080238SA SWP25080239SA SWP25090001SA SWP25090002SA SWP25090003SA SWP25090004SA SWP25090005SA SWP25090006SA SWP25090007SA SWP25090008SA SWP25090009SA SWP25090010SA SWP25090011SA SWP25090012SA SWP25090013SA SWP25090014SA SWP25090015SA SWP25090016SA SWP25090017SA SWP25090018SA SWP25170051SA SWP25120595SA SWP25120596SA SWP25120597SA SWP25120598SA SWP25260160SA SWP25260161SA SWP25290057SA SWP25051287SA SWP25051322SA SWP25051323SA SWP25051324SA SWP25051325SA SWP25051326SA SWP25051327SA SWP25051328SA SWP25051329SA SWP25051330SA SWP25051331SA SWP25051332SA SWP25051333SA SWP25051334SA SWP25051335SA SWP25051336SA SWP25051337SA SWP25130748SA SWP25130756SA SWP25130761SA SWP25130767SA SWP25151064SA SWP25190089SA SWP25230030SA SWP25240008SA SWP25130742SA SWP25130754SA SWP25130755SA SWP25130820SA SWP25130821SA SWP25130822SA SWP25130823SA SWP25130824SA SWP25130825SA SWP25130826SA SWP25130827SA SWP25130828SA SWP25130829SA SWP25130830SA SWP25130831SA SWP25130832SA SWP25130833SA SWP25130834SA SWP25130835SA SWP25130836SA SWP25130837SA SWP25130838SA SWP25130839SA SWP25230023SA SWP25140919SA SWP25140920SA SWP25140921SA SWP25120644SA SWP25120645SA SWP25120646SA SWP25120647SA SWP25120648SA SWP25120649SA SWP25120650SA SWP25260109SA SWP25260110SA SWP25260111SA SWP25260112SA SWP25260113SA SWP25110503SA SWP25051340SA SWP25051341SA SWP25051342SA SWP25051345SA SWP25051350SA SWP25051352SA SWP25051353SA SWP25051356SA SWP25051357SA SWP25051358SA SWP25051359SA SWP25051360SA SWP25051361SA SWP25051362SA SWP25051363SA SWP25051364SA SWP25051365SA SWP25051366SA SWP25051369SA SWP25051371SA SWP25051372SA SWP25051373SA SWP25051374SA SWP25051375SA SWP25051376SA SWP25051377SA SWP25051378SA SWP25051379SA SWP25051380SA SWP25090175SA SWP25090176SA SWP25090177SA SWP25090178SA SWP25090179SA SWP25090180SA SWP25090181SA SWP25090182SA SWP25090183SA SWP25090184SA SWP25120586SA SWP25120587SA SWP25120590SA SWP25130747SA SWP25140906SA SWP25140907SA SWP25140934SA SWP25141033SA SWP25141034SA SWP25141035SA SWP25141036SA SWP25151047SA SWP25151048SA SWP25151049SA SWP25151050SA SWP25190105SA SWP25250052SA SWP25260148SA SWP25270081SA SWP25290058SA SWP25290059SA SWP25090185SA SWP25090186SA SWP25090187SA SWP25090188SA SWP25090189SA SWP25130677SA SWP25130678SA SWP25130679SA SWP25130680SA SWP25130681SA SWP25210045SA SWP25210046SA SWP25210047SA SWP25210048SA SWP25210049SA SWP25210050SA SWP25210051SA SWP25210076SA SWP25290051SA SWP25070004SA SWP25090112SA SWP25090113SA SWP25090114SA SWP25090115SA SWP25090116SA SWP25090117SA SWP25090118SA SWP25090119SA SWP25090120SA SWP25090121SA SWP25090122SA SWP25090123SA SWP25090124SA SWP25090125SA SWP25090126SA SWP25090127SA SWP25090128SA SWP25090129SA SWP25090130SA SWP25090131SA SWP25090132SA SWP25090133SA SWP25090134SA SWP25090135SA SWP25090136SA SWP25090137SA SWP25090138SA SWP25090139SA SWP25090140SA SWP25090141SA SWP25120628SA SWP25210052SA SWP25210053SA SWP25210054SA SWP25210055SA SWP25210056SA SWP25210061SA SWP25210062SA SWP25210063SA SWP25210091SA SWP25210092SA SWP25210093SA SWP25090019SA SWP25090020SA SWP25090021SA SWP25090022SA SWP25090023SA SWP25090024SA SWP25090025SA SWP25090026SA SWP25090027SA SWP25090028SA SWP25090029SA SWP25090030SA SWP25090031SA SWP25090032SA SWP25090033SA SWP25090034SA SWP25090035SA SWP25090036SA SWP25090037SA SWP25090038SA SWP25270105SA SWP25240003SA SWP25240004SA SWP25240005SA SWP25240006SA SWP25220064SA SWP25190106SA SWP25230024SA SWP25140996SA SWP25140997SA SWP25140998SA SWP25141003SA SWP25141004SA SWP25141007SA SWP25141008SA SWP25141010SA SWP25141011SA SWP25120630SA SWP25200001SA SWP25200002SA SWP25200003SA SWP25200004SA SWP25200005SA SWP25200006SA SWP25200007SA SWP25200008SA SWP25200009SA SWP25200010SA SWP25200011SA SWP25200012SA SWP25200013SA SWP25200014SA SWP25061403SA SWP25040962SA SWP25040963SA SWP25040964SA SWP25260105SA SWP25260106SA SWP25260107SA SWP25170046SA SWP25180132SA SWP25090039SA SWP25090040SA SWP25090041SA SWP25090042SA SWP25090043SA SWP25090044SA SWP25090045SA SWP25090046SA SWP25090047SA SWP25090048SA SWP25090049SA SWP25090050SA SWP25090051SA SWP25090052SA SWP25090053SA SWP25090054SA SWP25090055SA SWP25090056SA SWP25090057SA SWP25090058SA SWP25090059SA SWP25090060SA SWP25090061SA SWP25090062SA SWP25090063SA SWP25090064SA SWP25090065SA SWP25090066SA SWP25090067SA SWP25090068SA SWP25090069SA SWP25090070SA SWP25090071SA SWP25090072SA SWP25090073SA SWP25090074SA SWP25090075SA SWP25090076SA SWP25090077SA SWP25090078SA SWP25090079SA SWP25090080SA SWP25090081SA SWP25090082SA SWP25090083SA SWP25090084SA SWP25090085SA SWP25090086SA SWP25090087SA SWP25090088SA SWP25090089SA SWP25090090SA SWP25090091SA SWP25090092SA SWP25090093SA SWP25090094SA SWP25090095SA SWP25090096SA SWP25090097SA SWP25090098SA SWP25090099SA SWP25090100SA SWP25090101SA SWP25090102SA SWP25090103SA SWP25090104SA SWP25090105SA SWP25090106SA SWP25090107SA SWP25090108SA SWP25090109SA SWP25090110SA SWP25090111SA SWP25151057SA SWP25220066SA SWP25250059SA SWP25250060SA SWP25250061SA SWP25280032SA SWP25100288SA SWP25130764SA SWP25180131SA SWP25260096SA SWP25230025SA SWP25051284SA SWP25051285SA SWP25051292SA SWP25051293SA SWP25051294SA SWP25051295SA SWP25051296SA SWP25051297SA SWP25051298SA SWP25051299SA SWP25051300SA SWP25051301SA SWP25051302SA SWP25051303SA SWP25051304SA SWP25051305SA SWP25051306SA SWP25051307SA SWP25051308SA SWP25051309SA SWP25051310SA SWP25051311SA SWP25051312SA SWP25051313SA SWP25051314SA SWP25051315SA SWP25051316SA SWP25051317SA SWP25051318SA SWP25051319SA SWP25051320SA SWP25051321SA SWP25120599SA SWP25120600SA SWP25120601SA SWP25120602SA SWP25120603SA SWP25220068SA SWP25230065SA SWP25180051SA SWP25180052SA SWP25180053SA SWP25180054SA SWP25180055SA SWP25180056SA SWP25180057SA SWP25180058SA SWP25180059SA SWP25180060SA SWP25180061SA SWP25180062SA SWP25180063SA SWP25180064SA SWP25180065SA SWP25180066SA SWP25180067SA SWP25180068SA SWP25180069SA SWP25180070SA SWP25180071SA SWP25180072SA SWP25180073SA SWP25180074SA SWP25180075SA SWP25180076SA SWP25180077SA SWP25180078SA SWP25180079SA SWP25180080SA SWP25180081SA SWP25180082SA SWP25180083SA SWP25180084SA SWP25180085SA SWP25180086SA SWP25180087SA SWP25180088SA SWP25180089SA SWP25180090SA SWP25180091SA SWP25180092SA SWP25180093SA SWP25180094SA SWP25180096SA SWP25180097SA SWP25180098SA SWP25180099SA SWP25180100SA SWP25290183WA SWP25290184WA SWP25290185WA SWP25100015WA SWP25100016WA SWP25280001WA SWP25280002WA SWP25280003WA SWP25280004WA SWP25290176WA SWP25290177WA SWP25290178WA SWP25290179WA SWP25290180WA SWP25240001WA SWP25290175WA SWP25160051WA SWP25090117WA SWP25110027WA SWP25110028WA SWP25110029WA SWP25110030WA SWP25110031WA SWP25110032WA SWP25110033WA SWP25110034WA SWP25110035WA SWP25110036WA SWP25110037WA SWP25110038WA SWP25110039WA SWP25110040WA SWP25110041WA SWP25110042WA SWP25110043WA SWP25110044WA SWP25110045WA SWP25110046WA SWP25090118WA SWP25090119WA SWP25090120WA SWP25090121WA SWP25090122WA SWP25090123WA SWP25290023WA SWP25290181WA SWP25290182WA SWP25090130WA SWP25090131WA SWP25090132WA SWP25090133WA SWP25090134WA SWP25090135WA SWP25090136WA SWP25090137WA SWP25090138WA SWP25090139WA SWP25090140WA SWP25090141WA SWP25090142WA SWP25160044WA SWP25160045WA SWP25160046WA SWP25160047WA SWP25160048WA SWP25160043WA SWP25130110WA SWP25160052WA SWP25040062WA SWP25100017WA SWP25100018WA SWP25140374WA SWP25160137WA SWP25180001WA SWP25160138WA SWP25160139WA SWP25130121WA SWP25130122WA SWP25130123WA SWP25130124WA SWP25130125WA SWP25130126WA SWP25130127WA SWP25130128WA SWP25130129WA SWP25130130WA SWP25130131WA SWP25130132WA SWP25130133WA SWP25030146WA SWP25040005WA SWP25040027WA SWP25040028WA SWP25090143WA SWP25160090WA SWP25160091WA SWP25160092WA SWP25170011WA SWP25170056WA SWP25260001WA SWP25300016WA SWP25300017WA SWP25300018WA SWP25170032SA SWP25170033SA SWP25170034SA SWP25170035SA SWP25220001SA SWP25220002SA SWP25220003SA SWP25220004SA SWP25080092SA SWP25110374SA SWP25110375SA SWP25110376SA SWP25110378SA SWP25110379SA SWP25110380SA SWP25120643SA SWP25130849SA SWP25130850SA SWP25130856SA SWP25130857SA SWP25151086SA SWP25260121SA SWP25260122SA SWP25260123SA SWP25260124SA SWP25260142SA SWP25260143SA SWP25260144SA SWP25260145SA SWP25260146SA SWP25260147SA SWP25260154SA SWP25270069SA SWP25280029SA SWP25260135SA SWP25260136SA SWP25260137SA SWP25260138SA SWP25260139SA SWP25260140SA SWP25260141SA SWP25110412SA SWP25110392SA SWP25110393SA SWP25170044SA SWP25170045SA SWP25250035SA SWP25250043SA SWP25250044SA SWP25250045SA SWP25260153SA SWP25270068SA SWP25190079SA SWP25190080SA SWP25190081SA SWP25190082SA SWP25190083SA SWP25190084SA SWP25190085SA SWP25110478SA SWP25110479SA SWP25040957SA SWP25090202SA SWP25090203SA SWP25090204SA SWP25100289SA SWP25110391SA SWP25040946SA SWP25040947SA SWP25040948SA SWP25040949SA SWP25100010SA SWP25100011SA SWP25240021SA SWP25240022SA SWP25151055SA SWP25151056SA SWP25100001SA SWP25100002SA SWP25100003SA SWP25100004SA SWP25100005SA SWP25100006SA SWP25210085SA SWP25210086SA SWP25210087SA SWP25170053SA SWP25190094SA SWP25190095SA SWP25190096SA SWP25190097SA SWP25151082SA SWP25151083SA SWP25270089SA SWP25270090SA SWP25270091SA SWP25270093SA SWP25270094SA SWP25270095SA SWP25270096SA SWP25270097SA SWP25270098SA SWP25300025SA SWP25300026SA SWP25300027SA SWP25300028SA SWP25300029SA SWP25270055WA SWP25270056WA SWP25270057WA SWP25270058WA SWP25270059WA SWP25270060WA SWP25270061WA SWP25270062WA SWP25270063WA SWP25270064WA SWP25270065WA SWP25270066WA SWP25270067WA SWP25270068WA SWP25270069WA SWP25270070WA SWP25270071WA SWP25270072WA SWP25040006WA SWP25040007WA SWP25040029WA SWP25040030WA SWP25040031WA SWP25040032WA SWP25040033WA SWP25040034WA SWP25040035WA SWP25040036WA SWP25040037WA SWP25040038WA SWP25040039WA SWP25040040WA SWP25061402SA SWP25100020WA SWP25280005WA SWP25280006WA SWP25140375WA SWP25140376WA SWP25290010WA SWP25290011WA SWP25290012WA SWP25100019WA SWP25140417WA SWP25290024WA SWP25210010WA SWP25260002WA SWP25260003WA SWP25170012WA SWP25270045WA SWP25270046WA SWP25060043WA SWP25210001WA SWP25210002WA SWP25210003WA SWP25210004WA SWP25210005WA SWP25210006WA SWP25210007WA SWP25210008WA SWP25210009WA SWP25250002WA SWP25250003WA SWP25290013WA SWP25290035WA SWP25280007WA SWP25230002WA SWP25230003WA SWP25230004WA SWP25230005WA SWP25170001WA SWP25170002WA SWP25170003WA SWP25170004WA SWP25170005WA SWP25170006WA SWP25170007WA SWP25170008WA SWP25170009WA SWP25170010WA SWP25160053WA SWP25250069WA SWP25270047WA SWP25270048WA SWP25260004WA SWP25230001WA SWP25230185WA SWP25230186WA SWP25090144WA SWP25160002WA SWP25160003WA SWP25160004WA SWP25140416WA SWP25290025WA SWP25290026WA SWP25290027WA SWP25290028WA SWP25290029WA SWP25290030WA SWP25290031WA SWP25290032WA SWP25160093WA SWP25240002WA SWP25100022WA SWP25080013WA SWP25080014WA SWP25080015WA SWP25080016WA SWP25080017WA SWP25080018WA SWP25080019WA SWP25080020WA SWP25080021WA SWP25080022WA SWP25250065WA SWP25250066WA SWP25250067WA SWP25250068WA SWP25210030WA SWP25210031WA SWP25210032WA SWP25250004WA SWP25250005WA SWP25300031WA SWP25100021WA SWP25180002WA SWP25240003WA SWP25290033WA SWP25290034WA SWP25170013WA SWP25170014WA SWP25300019WA SWP25300020WA SWP25300021WA SWP25300022WA SWP25300023WA SWP25300024WA SWP25250006WA SWP25160001WA SWP25300029WA SWP25300030WA SWP25040008WA SWP25040009WA SWP25160094WA SWP25160095WA SWP25170015WA SWP25170016WA SWP25080002WA SWP25140401WA SWP25250008WA SWP25250009WA SWP25250010WA SWP25250011WA SWP25250012WA SWP25250013WA SWP25250014WA SWP25250015WA SWP25250016WA SWP25250017WA SWP25250018WA SWP25250019WA SWP25250020WA SWP25250021WA SWP25250022WA SWP25250023WA SWP25250024WA SWP25250025WA SWP25250026WA SWP25250027WA SWP25250028WA SWP25250029WA SWP25250030WA SWP25250031WA SWP25250032WA SWP25250033WA SWP25250034WA SWP25250035WA SWP25250036WA SWP25250037WA SWP25250038WA SWP25250039WA SWP25250040WA SWP25250041WA SWP25250042WA SWP25250043WA SWP25250044WA SWP25250045WA SWP25250046WA SWP25250047WA SWP25250048WA SWP25250049WA SWP25250050WA SWP25210053WA SWP25210054WA SWP25080023WA SWP25290036WA SWP25070228WA SWP25031784WA SWP25031785WA SWP25031786WA SWP25031787WA SWP25031788WA SWP25031789WA SWP25031790WA SWP25031791WA SWP25031792WA SWP25031793WA SWP25031794WA SWP25031795WA SWP25031796WA SWP25031797WA SWP25031798WA SWP25031799WA SWP25031800WA SWP25031801WA SWP25031802WA SWP25031803WA SWP25031804WA SWP25031805WA SWP25031806WA SWP25031807WA SWP25031808WA SWP25031809WA SWP25031810WA SWP25031811WA SWP25031812WA SWP25031813WA SWP25031814WA SWP25031815WA SWP25031816WA SWP25031817WA SWP25031818WA SWP25031819WA SWP25031820WA SWP25031821WA SWP25031822WA SWP25031823WA SWP25031824WA SWP25031825WA SWP25031826WA SWP25031827WA SWP25031828WA SWP25031829WA SWP25031830WA SWP25031831WA SWP25031832WA SWP25031833WA SWP25031834WA SWP25031835WA SWP25031836WA SWP25031837WA SWP25031838WA SWP25031839WA SWP25031840WA SWP25031841WA SWP25031842WA SWP25031843WA SWP25031844WA SWP25031845WA SWP25031846WA SWP25031847WA SWP25031848WA SWP25140377WA SWP25140378WA SWP25140379WA SWP25250007WA SWP25170057WA SWP25170058WA SWP25070102WA SWP25140380WA SWP25140381WA SWP25140382WA SWP25140383WA SWP25140384WA SWP25140385WA SWP25140386WA SWP25140387WA SWP25140388WA SWP25140389WA SWP25140390WA SWP25140391WA SWP25140392WA SWP25140393WA SWP25140394WA SWP25140395WA SWP25140396WA SWP25140397WA SWP25140398WA SWP25140399WA SWP25280008WA SWP25280009WA SWP25280010WA SWP25280011WA SWP25280012WA SWP25280013WA SWP25280014WA SWP25280015WA SWP25280016WA SWP25280017WA SWP25280018WA SWP25280019WA SWP25280020WA SWP25280021WA SWP25280022WA SWP25280023WA SWP25280024WA SWP25280025WA SWP25280026WA SWP25280027WA SWP25250051WA SWP25250052WA SWP25220001WA SWP25220002WA SWP25220003WA SWP25220004WA SWP25220005WA SWP25220021WA SWP25220022WA SWP25220023WA SWP25220024WA SWP25220025WA SWP25220026WA SWP25220027WA SWP25220028WA SWP25220029WA SWP25220030WA SWP25210018WA SWP25210019WA SWP25160005WA SWP25160006WA SWP25160007WA SWP25160008WA SWP25160009WA SWP25160010WA SWP25160011WA SWP25160012WA SWP25160013WA SWP25290037WA SWP25290038WA SWP25290039WA SWP25290040WA SWP25290041WA SWP25290042WA SWP25290043WA SWP25290044WA SWP25290045WA SWP25290046WA SWP25290047WA SWP25290048WA SWP25290049WA SWP25290050WA SWP25290062WA SWP25290063WA SWP25290064WA SWP25290065WA SWP25290066WA SWP25290067WA SWP25290068WA SWP25290069WA SWP25290070WA SWP25290071WA SWP25290072WA SWP25290073WA SWP25290074WA SWP25290075WA SWP25290076WA SWP25290077WA SWP25290078WA SWP25290079WA SWP25290080WA SWP25290081WA SWP25290082WA SWP25290083WA SWP25290084WA SWP25290085WA SWP25290086WA SWP25290087WA SWP25290088WA SWP25290089WA SWP25290090WA SWP25290091WA SWP25290092WA SWP25290093WA SWP25290094WA SWP25290095WA SWP25290096WA SWP25290097WA SWP25290098WA SWP25290099WA SWP25290100WA SWP25290101WA SWP25290102WA SWP25290103WA SWP25290104WA SWP25290105WA SWP25290106WA SWP25290107WA SWP25290108WA SWP25290109WA SWP25290110WA SWP25290111WA SWP25290112WA SWP25290113WA SWP25290114WA SWP25290115WA SWP25290116WA SWP25290117WA SWP25290118WA SWP25290119WA SWP25290120WA SWP25290121WA SWP25290122WA SWP25290123WA SWP25290124WA SWP25290125WA SWP25290126WA SWP25290127WA SWP25290128WA SWP25290129WA SWP25290130WA SWP25290131WA SWP25290132WA SWP25290133WA SWP25290134WA SWP25290135WA SWP25290136WA SWP25290137WA SWP25290138WA SWP25290139WA SWP25290140WA SWP25290141WA SWP25290142WA SWP25290143WA SWP25290144WA SWP25290145WA SWP25290146WA SWP25290147WA SWP25290148WA SWP25290149WA SWP25290150WA SWP25290151WA SWP25290152WA SWP25290153WA SWP25290154WA SWP25290155WA SWP25290156WA SWP25290157WA SWP25290158WA SWP25290159WA SWP25290160WA SWP25290161WA SWP25290162WA SWP25290163WA SWP25290164WA SWP25290165WA SWP25290166WA SWP25290167WA SWP25290168WA SWP25290169WA SWP25290170WA SWP25290171WA SWP25290172WA SWP25290173WA SWP25290174WA SWP25240005WA SWP25240006WA SWP25270087WA SWP25270088WA SWP25220011WA SWP25170017WA SWP25170018WA SWP25170019WA SWP25170020WA SWP25170021WA SWP25300007WA SWP25300008WA SWP25160096WA SWP25070103WA SWP25160097WA SWP25160098WA SWP25160099WA SWP25160100WA SWP25160101WA SWP25160102WA SWP25030129WA SWP25030130WA SWP25120088WA SWP25120089WA SWP25120090WA SWP25160115WA SWP25160116WA SWP25160117WA SWP25160118WA SWP25160119WA SWP25160120WA SWP25160121WA SWP25160122WA SWP25160123WA SWP25160124WA SWP25160125WA SWP25160126WA SWP25160127WA SWP25160128WA SWP25160129WA SWP25160130WA SWP25160131WA SWP25160132WA SWP25160133WA SWP25160134WA SWP25160135WA SWP25160136WA SWP25090002WA SWP25090003WA SWP25090004WA SWP25090005WA SWP25090006WA SWP25120101WA SWP25120102WA SWP25200001WA SWP25200002WA SWP25270001WA SWP25270075WA SWP25290001WA SWP25210055WA SWP25210056WA SWP25210057WA SWP25210058WA SWP25220012WA SWP25220013WA SWP25220014WA SWP25220015WA SWP25220016WA SWP25220017WA SWP25220018WA SWP25220019WA SWP25220020WA SWP25290014WA SWP25290015WA SWP25290016WA SWP25290017WA SWP25290018WA SWP25290019WA SWP25290020WA SWP25290021WA SWP25290022WA SWP25140458WA SWP25210020WA SWP25210021WA SWP25210022WA SWP25250053WA SWP25250054WA SWP25250055WA SWP25250056WA SWP25250057WA SWP25250058WA SWP25250059WA SWP25250060WA SWP25250061WA SWP25300032WA SWP25130114WA SWP25130115WA SWP25030131WA SWP25030132WA SWP25030133WA SWP25030134WA SWP25030135WA SWP25030136WA SWP25030137WA SWP25030138WA SWP25030139WA SWP25030140WA SWP25170059WA SWP25170060WA SWP25250090WA SWP25290052WA SWP25290053WA SWP25290054WA SWP25290055WA SWP25290056WA SWP25290057WA SWP25290058WA SWP25290059WA SWP25290051WA SWP25170022WA SWP25100023WA SWP25100024WA SWP25100025WA SWP25040041WA SWP25040042WA SWP25040043WA SWP25070107WA SWP25070108WA SWP25070229WA SWP25080126WA SWP25080127WA SWP25080128WA SWP25080129WA SWP25080130WA SWP25100026WA SWP25140400WA SWP25210014WA SWP25210015WA SWP25210016WA SWP25210017WA SWP25210042WA SWP25250001WA SWP25110048WA SWP25110049WA SWP25160054WA SWP25170055WA SWP25170061WA SWP25210040WA SWP25210041WA SWP25270073WA SWP25300009WA SWP25060063WA SWP25150561WA SWP25150562WA SWP25150563WA SWP25150564WA SWP25150565WA SWP25150566WA SWP25150727WA SWP25150728WA SWP25260007WA SWP25270089WA SWP25270090WA SWP25070001WA SWP25230116WA SWP25230117WA SWP25230118WA SWP25230119WA SWP25230120WA SWP25230121WA SWP25230122WA SWP25230123WA SWP25230124WA SWP25230125WA SWP25230126WA SWP25230127WA SWP25230128WA SWP25230129WA SWP25230130WA SWP25230131WA SWP25230132WA SWP25230133WA SWP25230134WA SWP25230135WA SWP25230136WA SWP25230137WA SWP25230138WA SWP25230139WA SWP25230140WA SWP25230141WA SWP25230142WA SWP25230143WA SWP25230144WA SWP25230145WA SWP25230146WA SWP25230147WA SWP25230148WA SWP25230149WA SWP25230150WA SWP25230151WA SWP25230152WA SWP25230153WA SWP25230154WA SWP25230155WA SWP25230156WA SWP25230157WA SWP25230158WA SWP25230159WA SWP25230160WA SWP25230161WA SWP25230162WA SWP25230163WA SWP25230164WA SWP25230165WA SWP25230166WA SWP25230167WA SWP25230168WA SWP25230169WA SWP25230170WA SWP25230171WA SWP25230172WA SWP25230173WA SWP25230174WA SWP25230175WA SWP25230176WA SWP25230177WA SWP25230178WA SWP25230179WA SWP25230180WA SWP25230181WA SWP25230182WA SWP25230183WA SWP25230184WA SWP25160034SA SWP25160035SA SWP25160036SA SWP25070002WA SWP25070003WA SWP25070004WA SWP25070005WA SWP25070006WA SWP25070007WA SWP25070008WA SWP25070009WA SWP25070010WA SWP25070011WA SWP25070012WA SWP25070013WA SWP25070014WA SWP25070015WA SWP25070016WA SWP25070017WA SWP25070018WA SWP25070019WA SWP25070020WA SWP25070021WA SWP25070022WA SWP25070023WA SWP25070024WA SWP25070025WA SWP25070026WA SWP25070027WA SWP25070028WA SWP25070029WA SWP25070030WA SWP25070031WA SWP25070032WA SWP25070033WA SWP25070034WA SWP25070035WA SWP25070036WA SWP25070037WA SWP25070038WA SWP25070039WA SWP25070040WA SWP25070041WA SWP25070042WA SWP25070043WA SWP25070044WA SWP25070045WA SWP25070046WA SWP25070047WA SWP25070048WA SWP25070049WA SWP25070050WA SWP25070051WA SWP25070052WA SWP25070053WA SWP25070054WA SWP25070055WA SWP25070056WA SWP25070057WA SWP25070058WA SWP25070059WA SWP25070060WA SWP25070061WA SWP25070062WA SWP25070063WA SWP25070064WA SWP25070065WA SWP25070066WA SWP25070067WA SWP25070068WA SWP25070069WA SWP25070070WA SWP25070071WA SWP25070072WA SWP25070073WA SWP25070074WA SWP25070075WA SWP25070076WA SWP25070077WA SWP25070078WA SWP25070079WA SWP25070080WA SWP25070081WA SWP25070082WA SWP25070083WA SWP25070084WA SWP25070085WA SWP25070086WA SWP25070087WA SWP25070088WA SWP25070089WA SWP25070090WA SWP25070091WA SWP25070092WA SWP25070093WA SWP25070094WA SWP25070095WA SWP25070096WA SWP25070097WA SWP25070098WA SWP25070099WA SWP25070100WA SWP25070101WA SWP25070113WA SWP25070114WA SWP25070115WA SWP25070116WA SWP25070117WA SWP25070118WA SWP25070119WA SWP25070120WA SWP25070121WA SWP25070122WA SWP25070123WA SWP25070124WA SWP25070125WA SWP25070126WA SWP25070127WA SWP25070128WA SWP25070129WA SWP25070130WA SWP25070131WA SWP25070132WA SWP25070133WA SWP25070134WA SWP25070135WA SWP25070136WA SWP25070137WA SWP25070138WA SWP25070139WA SWP25070140WA SWP25070141WA SWP25070142WA SWP25070143WA SWP25070144WA SWP25070145WA SWP25070146WA SWP25070147WA SWP25070148WA SWP25070149WA SWP25070150WA SWP25070151WA SWP25070152WA SWP25070153WA SWP25070154WA SWP25070155WA SWP25070156WA SWP25070157WA SWP25070158WA SWP25070159WA SWP25070160WA SWP25070161WA SWP25070162WA SWP25070163WA SWP25070164WA SWP25070165WA SWP25070166WA SWP25070167WA SWP25070168WA SWP25070169WA SWP25070170WA SWP25070171WA SWP25070172WA SWP25070173WA SWP25070174WA SWP25070175WA SWP25070176WA SWP25070177WA SWP25070178WA SWP25070179WA SWP25070180WA SWP25070181WA SWP25070182WA SWP25070183WA SWP25070184WA SWP25070185WA SWP25070186WA SWP25070187WA SWP25070188WA SWP25070189WA SWP25070190WA SWP25070191WA SWP25070192WA SWP25070193WA SWP25070194WA SWP25070195WA SWP25070196WA SWP25070197WA SWP25070198WA SWP25070199WA SWP25070200WA SWP25070201WA SWP25070202WA SWP25070203WA SWP25070204WA SWP25070205WA SWP25070206WA SWP25070207WA SWP25070208WA SWP25070209WA SWP25070210WA SWP25070211WA SWP25070212WA SWP25080024WA SWP25080025WA SWP25080026WA SWP25080027WA SWP25080028WA SWP25080029WA SWP25080030WA SWP25080031WA SWP25080032WA SWP25080033WA SWP25080034WA SWP25080035WA SWP25080036WA SWP25080037WA SWP25080038WA SWP25080039WA SWP25080040WA SWP25080041WA SWP25080042WA SWP25080043WA SWP25080044WA SWP25080045WA SWP25080046WA SWP25080047WA SWP25080048WA SWP25080049WA SWP25080050WA SWP25080051WA SWP25080052WA SWP25080053WA SWP25080054WA SWP25080055WA SWP25080056WA SWP25080057WA SWP25080058WA SWP25080059WA SWP25080060WA SWP25080061WA SWP25080062WA SWP25080063WA SWP25080064WA SWP25080065WA SWP25080066WA SWP25080067WA SWP25080068WA SWP25080069WA SWP25080070WA SWP25080071WA SWP25080072WA SWP25080073WA SWP25080074WA SWP25080075WA SWP25080076WA SWP25080077WA SWP25080078WA SWP25080079WA SWP25080080WA SWP25080081WA SWP25080082WA SWP25080083WA SWP25080084WA SWP25080085WA SWP25080086WA SWP25080087WA SWP25080088WA SWP25080089WA SWP25080090WA SWP25080091WA SWP25080092WA SWP25080093WA SWP25080094WA SWP25080095WA SWP25080096WA SWP25080097WA SWP25080098WA SWP25080099WA SWP25080100WA SWP25080101WA SWP25080102WA SWP25080103WA SWP25080104WA SWP25080105WA SWP25080106WA SWP25080107WA SWP25080108WA SWP25080109WA SWP25080110WA SWP25080111WA SWP25080112WA SWP25080113WA SWP25080114WA SWP25080115WA SWP25080116WA SWP25080117WA SWP25080118WA SWP25080119WA SWP25080120WA SWP25080121WA SWP25080122WA SWP25080123WA SWP25080131WA SWP25080132WA SWP25080133WA SWP25080134WA SWP25080135WA SWP25080136WA SWP25080137WA SWP25080138WA SWP25080139WA SWP25080140WA SWP25080141WA SWP25080142WA SWP25080143WA SWP25080144WA SWP25080145WA SWP25080146WA SWP25080147WA SWP25080148WA SWP25080149WA SWP25080150WA SWP25080151WA SWP25080152WA SWP25080153WA SWP25080154WA SWP25080155WA SWP25080156WA SWP25080157WA SWP25080158WA SWP25080159WA SWP25080160WA SWP25080161WA SWP25080162WA SWP25080163WA SWP25080164WA SWP25080165WA SWP25080166WA SWP25080167WA SWP25080168WA SWP25080169WA SWP25080170WA SWP25080171WA SWP25080172WA SWP25080173WA SWP25080174WA SWP25080175WA SWP25080176WA SWP25080177WA SWP25080178WA SWP25080179WA SWP25080180WA SWP25080181WA SWP25080182WA SWP25080183WA SWP25080184WA SWP25080185WA SWP25080186WA SWP25080187WA SWP25080188WA SWP25080189WA SWP25080190WA SWP25080191WA SWP25080192WA SWP25080193WA SWP25080194WA SWP25080195WA SWP25080196WA SWP25080197WA SWP25080198WA SWP25080199WA SWP25080200WA SWP25080201WA SWP25080202WA SWP25080203WA SWP25080204WA SWP25080205WA SWP25080206WA SWP25080207WA SWP25080208WA SWP25080209WA SWP25080210WA SWP25080211WA SWP25080212WA SWP25080213WA SWP25080214WA SWP25080215WA SWP25080216WA SWP25080217WA SWP25080218WA SWP25080219WA SWP25080220WA SWP25080221WA SWP25080222WA SWP25080223WA SWP25080224WA SWP25080225WA SWP25080226WA SWP25080227WA SWP25080228WA SWP25080229WA SWP25080230WA SWP25090013WA SWP25090014WA SWP25090015WA SWP25090016WA SWP25090017WA SWP25090018WA SWP25090019WA SWP25090020WA SWP25090021WA SWP25090022WA SWP25090023WA SWP25090024WA SWP25090025WA SWP25090026WA SWP25090027WA SWP25090028WA SWP25090029WA SWP25090030WA SWP25090031WA SWP25090032WA SWP25090033WA SWP25090034WA SWP25090035WA SWP25090036WA SWP25090037WA SWP25090038WA SWP25090039WA SWP25090040WA SWP25090041WA SWP25090042WA SWP25090043WA SWP25090044WA SWP25090045WA SWP25090046WA SWP25090047WA SWP25090048WA SWP25090049WA SWP25090050WA SWP25090051WA SWP25090052WA SWP25090053WA SWP25090054WA SWP25090055WA SWP25090056WA SWP25090057WA SWP25090058WA SWP25090059WA SWP25090060WA SWP25090061WA SWP25090062WA SWP25090063WA SWP25090064WA SWP25090065WA SWP25090066WA SWP25090067WA SWP25090068WA SWP25090069WA SWP25090070WA SWP25090071WA SWP25090072WA SWP25090073WA SWP25090074WA SWP25090075WA SWP25090076WA SWP25090077WA SWP25090078WA SWP25090079WA SWP25090080WA SWP25090081WA SWP25090082WA SWP25090083WA SWP25090084WA SWP25090085WA SWP25090086WA SWP25090087WA SWP25090088WA SWP25090089WA SWP25090090WA SWP25090091WA SWP25090092WA SWP25090093WA SWP25090094WA SWP25090095WA SWP25090096WA SWP25090097WA SWP25090098WA SWP25090099WA SWP25090100WA SWP25090101WA SWP25090102WA SWP25090103WA SWP25090104WA SWP25090105WA SWP25090106WA SWP25090107WA SWP25090108WA SWP25090109WA SWP25090110WA SWP25090111WA SWP25090112WA SWP25130134WA SWP25130135WA SWP25130136WA SWP25130137WA SWP25130138WA SWP25130139WA SWP25130140WA SWP25130141WA SWP25130142WA SWP25130143WA SWP25130144WA SWP25130145WA SWP25130146WA SWP25130147WA SWP25130148WA SWP25130149WA SWP25130150WA SWP25130151WA SWP25130152WA SWP25130153WA SWP25130154WA SWP25130155WA SWP25130156WA SWP25130157WA SWP25130158WA SWP25130159WA SWP25130160WA SWP25130161WA SWP25130162WA SWP25130163WA SWP25130164WA SWP25130165WA SWP25130166WA SWP25130167WA SWP25130168WA SWP25130169WA SWP25130170WA SWP25130171WA SWP25130172WA SWP25130173WA SWP25130174WA SWP25130175WA SWP25130176WA SWP25130177WA SWP25130178WA SWP25130179WA SWP25130180WA SWP25130181WA SWP25130182WA SWP25130183WA SWP25130184WA SWP25130185WA SWP25130186WA SWP25130187WA SWP25130188WA SWP25130189WA SWP25130190WA SWP25130191WA SWP25130192WA SWP25130193WA SWP25130194WA SWP25130195WA SWP25130196WA SWP25130197WA SWP25130198WA SWP25130199WA SWP25130200WA SWP25130201WA SWP25130202WA SWP25130203WA SWP25130204WA SWP25130205WA SWP25130206WA SWP25130207WA SWP25130208WA SWP25130209WA SWP25130210WA SWP25130211WA SWP25130212WA SWP25130213WA SWP25130214WA SWP25130215WA SWP25130216WA SWP25130217WA SWP25130218WA SWP25130219WA SWP25130220WA SWP25130221WA SWP25130222WA SWP25130223WA SWP25130224WA SWP25130225WA SWP25130226WA SWP25130227WA SWP25130228WA SWP25130229WA SWP25130230WA SWP25130231WA SWP25130232WA SWP25130233WA SWP25130234WA SWP25130235WA SWP25130236WA SWP25130237WA SWP25130238WA SWP25130239WA SWP25130240WA SWP25130241WA SWP25130242WA SWP25130243WA SWP25130244WA SWP25130245WA SWP25130246WA SWP25130247WA SWP25130248WA SWP25130249WA SWP25130250WA SWP25130251WA SWP25130252WA SWP25130253WA SWP25130254WA SWP25130255WA SWP25130256WA SWP25130257WA SWP25130258WA SWP25130259WA SWP25130260WA SWP25130261WA SWP25130262WA SWP25130263WA SWP25130264WA SWP25130265WA SWP25130266WA SWP25130267WA SWP25130268WA SWP25130269WA SWP25130270WA SWP25130271WA SWP25130272WA SWP25130273WA SWP25130274WA SWP25130275WA SWP25130276WA SWP25130277WA SWP25130278WA SWP25130279WA SWP25130280WA SWP25130281WA SWP25130282WA SWP25130283WA SWP25130284WA SWP25130285WA SWP25130286WA SWP25130287WA SWP25130288WA SWP25130289WA SWP25130290WA SWP25130291WA SWP25130292WA SWP25130293WA SWP25140294WA SWP25140295WA SWP25140296WA SWP25140297WA SWP25140298WA SWP25140299WA SWP25140300WA SWP25140301WA SWP25140302WA SWP25140303WA SWP25140304WA SWP25140305WA SWP25140306WA SWP25140307WA SWP25140308WA SWP25140309WA SWP25140310WA SWP25140311WA SWP25140312WA SWP25140313WA SWP25140314WA SWP25140315WA SWP25140316WA SWP25140317WA SWP25140318WA SWP25140319WA SWP25140320WA SWP25140321WA SWP25140322WA SWP25140323WA SWP25140324WA SWP25140325WA SWP25140326WA SWP25140327WA SWP25140328WA SWP25140329WA SWP25140330WA SWP25140331WA SWP25140332WA SWP25140333WA SWP25140334WA SWP25140335WA SWP25140336WA SWP25140337WA SWP25140338WA SWP25140339WA SWP25140340WA SWP25140341WA SWP25140342WA SWP25140343WA SWP25140344WA SWP25140345WA SWP25140346WA SWP25140347WA SWP25140348WA SWP25140349WA SWP25140350WA SWP25140351WA SWP25140352WA SWP25140353WA SWP25140354WA SWP25140355WA SWP25140356WA SWP25140357WA SWP25140358WA SWP25140359WA SWP25140360WA SWP25140361WA SWP25140362WA SWP25140363WA SWP25140364WA SWP25140365WA SWP25140366WA SWP25140367WA SWP25140368WA SWP25140369WA SWP25140370WA SWP25140371WA SWP25140372WA SWP25140373WA SWP25140402WA SWP25140403WA SWP25140404WA SWP25140405WA SWP25140406WA SWP25140407WA SWP25140408WA SWP25140409WA SWP25140410WA SWP25140411WA SWP25150461WA SWP25150462WA SWP25150463WA SWP25150464WA SWP25150465WA SWP25150466WA SWP25150467WA SWP25150468WA SWP25150469WA SWP25150470WA SWP25150471WA SWP25150472WA SWP25150473WA SWP25150474WA SWP25150475WA SWP25150476WA SWP25150477WA SWP25150478WA SWP25150479WA SWP25150480WA SWP25150481WA SWP25150482WA SWP25150483WA SWP25150484WA SWP25150485WA SWP25150486WA SWP25150487WA SWP25150488WA SWP25150489WA SWP25150490WA SWP25150491WA SWP25150492WA SWP25150493WA SWP25150494WA SWP25150495WA SWP25150496WA SWP25150497WA SWP25150498WA SWP25150499WA SWP25150500WA SWP25150501WA SWP25150502WA SWP25150503WA SWP25150504WA SWP25150505WA SWP25150506WA SWP25150507WA SWP25150508WA SWP25150509WA SWP25150510WA SWP25150511WA SWP25150512WA SWP25150513WA SWP25150514WA SWP25150515WA SWP25150516WA SWP25150517WA SWP25150518WA SWP25150519WA SWP25150520WA SWP25150521WA SWP25150522WA SWP25150523WA SWP25150524WA SWP25150525WA SWP25150526WA SWP25150527WA SWP25150528WA SWP25150529WA SWP25150530WA SWP25150531WA SWP25150532WA SWP25150533WA SWP25150534WA SWP25150535WA SWP25150536WA SWP25150537WA SWP25150538WA SWP25150539WA SWP25150540WA SWP25150541WA SWP25150542WA SWP25150543WA SWP25150544WA SWP25150545WA SWP25150546WA SWP25150547WA SWP25150548WA SWP25150549WA SWP25150550WA SWP25150551WA SWP25150552WA SWP25150553WA SWP25150554WA SWP25150555WA SWP25150556WA SWP25150557WA SWP25150558WA SWP25150559WA SWP25150560WA SWP25150567WA SWP25150568WA SWP25150569WA SWP25150570WA SWP25150571WA SWP25150572WA SWP25150573WA SWP25150574WA SWP25150575WA SWP25150576WA SWP25150577WA SWP25150578WA SWP25150579WA SWP25150580WA SWP25150581WA SWP25150582WA SWP25150583WA SWP25150584WA SWP25150585WA SWP25150586WA SWP25150587WA SWP25150588WA SWP25150589WA SWP25150590WA SWP25150591WA SWP25150592WA SWP25150593WA SWP25150594WA SWP25150595WA SWP25150596WA SWP25150597WA SWP25150598WA SWP25150599WA SWP25150600WA SWP25150601WA SWP25150602WA SWP25150603WA SWP25150604WA SWP25150605WA SWP25150606WA SWP25150607WA SWP25150608WA SWP25150609WA SWP25150610WA SWP25150611WA SWP25150612WA SWP25150613WA SWP25150614WA SWP25150615WA SWP25150616WA SWP25150617WA SWP25150618WA SWP25150619WA SWP25150620WA SWP25150621WA SWP25150622WA SWP25150623WA SWP25150624WA SWP25150625WA SWP25150626WA SWP25150627WA SWP25150628WA SWP25150629WA SWP25150630WA SWP25150631WA SWP25150632WA SWP25150633WA SWP25150634WA SWP25150635WA SWP25150636WA SWP25150637WA SWP25150638WA SWP25150639WA SWP25150640WA SWP25150641WA SWP25150642WA SWP25150643WA SWP25150644WA SWP25150645WA SWP25150646WA SWP25150647WA SWP25150648WA SWP25150649WA SWP25150650WA SWP25150651WA SWP25150652WA SWP25150653WA SWP25150654WA SWP25150655WA SWP25150656WA SWP25150657WA SWP25150658WA SWP25150659WA SWP25150660WA SWP25150661WA SWP25150662WA SWP25150663WA SWP25150664WA SWP25150665WA SWP25150666WA SWP25150667WA SWP25150668WA SWP25150669WA SWP25150670WA SWP25150671WA SWP25150672WA SWP25150673WA SWP25150674WA SWP25150675WA SWP25150676WA SWP25150677WA SWP25150678WA SWP25150679WA SWP25150680WA SWP25150681WA SWP25150682WA SWP25150683WA SWP25150684WA SWP25150685WA SWP25150686WA SWP25150687WA SWP25150688WA SWP25150689WA SWP25150690WA SWP25150691WA SWP25150692WA SWP25150693WA SWP25150694WA SWP25150695WA SWP25150696WA SWP25150697WA SWP25150698WA SWP25150699WA SWP25150700WA SWP25150701WA SWP25150702WA SWP25150703WA SWP25150704WA SWP25150705WA SWP25150706WA SWP25150707WA SWP25150708WA SWP25150709WA SWP25150710WA SWP25150711WA SWP25150712WA SWP25150713WA SWP25150714WA SWP25150715WA SWP25150716WA SWP25180003WA SWP25180004WA SWP25180005WA SWP25180006WA SWP25180007WA SWP25180008WA SWP25180009WA SWP25180010WA SWP25180011WA SWP25180012WA SWP25180013WA SWP25180014WA SWP25180015WA SWP25180016WA SWP25180017WA SWP25180018WA SWP25180019WA SWP25180020WA SWP25180021WA SWP25180022WA SWP25180023WA SWP25180024WA SWP25180025WA SWP25180026WA SWP25180027WA SWP25180028WA SWP25180029WA SWP25180030WA SWP25180031WA SWP25180032WA SWP25180033WA SWP25180034WA SWP25180035WA SWP25180036WA SWP25180037WA SWP25180038WA SWP25180039WA SWP25180040WA SWP25180041WA SWP25180042WA SWP25180043WA SWP25180044WA SWP25180045WA SWP25180046WA SWP25180047WA SWP25180048WA SWP25180049WA SWP25180050WA SWP25180051WA SWP25180052WA SWP25180064WA SWP25180065WA SWP25180066WA SWP25180067WA SWP25180068WA SWP25180069WA SWP25180070WA SWP25180071WA SWP25180072WA SWP25180073WA SWP25180074WA SWP25180075WA SWP25180076WA SWP25180077WA SWP25180078WA SWP25180079WA SWP25180080WA SWP25180081WA SWP25180082WA SWP25180083WA SWP25180084WA SWP25180085WA SWP25180086WA SWP25180087WA SWP25180088WA SWP25180089WA SWP25180090WA SWP25180091WA SWP25180092WA SWP25180093WA SWP25180094WA SWP25180095WA SWP25180096WA SWP25180097WA SWP25180098WA SWP25180099WA SWP25180100WA SWP25180101WA SWP25180102WA SWP25180103WA SWP25180104WA SWP25180105WA SWP25180106WA SWP25180107WA SWP25180108WA SWP25180109WA SWP25180110WA SWP25180111WA SWP25180112WA SWP25180113WA SWP25180114WA SWP25180115WA SWP25180116WA SWP25180117WA SWP25180118WA SWP25180119WA SWP25180120WA SWP25180121WA SWP25180122WA SWP25180123WA SWP25180124WA SWP25180125WA SWP25180126WA SWP25180127WA SWP25180128WA SWP25180129WA SWP25180130WA SWP25180131WA SWP25180132WA SWP25180133WA SWP25180134WA SWP25180135WA SWP25180136WA SWP25180137WA SWP25180138WA SWP25180139WA SWP25180140WA SWP25180141WA SWP25180142WA SWP25180143WA SWP25180144WA SWP25180145WA SWP25180146WA SWP25180147WA SWP25180148WA SWP25180149WA SWP25180150WA SWP25180151WA SWP25180152WA SWP25180153WA SWP25180154WA SWP25180155WA SWP25180156WA SWP25180157WA SWP25180158WA SWP25180159WA SWP25180160WA SWP25180161WA SWP25180162WA SWP25180163WA SWP25180164WA SWP25180165WA SWP25180166WA SWP25180167WA SWP25180168WA SWP25180169WA SWP25180170WA SWP25180171WA SWP25180172WA SWP25180173WA SWP25180174WA SWP25180175WA SWP25180176WA SWP25180177WA SWP25180178WA SWP25180179WA SWP25180180WA SWP25180181WA SWP25180182WA SWP25180183WA SWP25180184WA SWP25180185WA SWP25180186WA SWP25180187WA SWP25180188WA SWP25180189WA SWP25180190WA SWP25180191WA SWP25180192WA SWP25180193WA SWP25180194WA SWP25180195WA SWP25180196WA SWP25180197WA SWP25180198WA SWP25180199WA SWP25180200WA SWP25180201WA SWP25180202WA SWP25180203WA SWP25180204WA SWP25180205WA SWP25180206WA SWP25180207WA SWP25180208WA SWP25180209WA SWP25180210WA SWP25180211WA SWP25180212WA SWP25180213WA SWP25180214WA SWP25180215WA SWP25180216WA SWP25180217WA SWP25180218WA SWP25180219WA SWP25180220WA SWP25180221WA SWP25180222WA SWP25180223WA SWP25180224WA SWP25180225WA SWP25180226WA SWP25180227WA SWP25180228WA SWP25180229WA SWP25180230WA SWP25180231WA SWP25180232WA SWP25180233WA SWP25180234WA SWP25180235WA SWP25180236WA SWP25180237WA SWP25180238WA SWP25180239WA SWP25180240WA SWP25180241WA SWP25180242WA SWP25180243WA SWP25180244WA SWP25180245WA SWP25180246WA SWP25180247WA SWP25180248WA SWP25180249WA SWP25180250WA SWP25180251WA SWP25180252WA SWP25180253WA SWP25180254WA SWP25180255WA SWP25180256WA SWP25180257WA SWP25180258WA SWP25180259WA SWP25180260WA SWP25180261WA SWP25180262WA SWP25180263WA SWP25180264WA SWP25180265WA SWP25180266WA SWP25180267WA SWP25180268WA SWP25180269WA SWP25180270WA SWP25180271WA SWP25180272WA SWP25180273WA SWP25180274WA SWP25180275WA SWP25180276WA SWP25180277WA SWP25180278WA SWP25180279WA SWP25180280WA SWP25180281WA SWP25180282WA SWP25180283WA SWP25180284WA SWP25180285WA SWP25180286WA SWP25180287WA SWP25180288WA SWP25180289WA SWP25180290WA SWP25180291WA SWP25180292WA SWP25180293WA SWP25180294WA SWP25180295WA SWP25180296WA SWP25180297WA SWP25180298WA SWP25180299WA SWP25180300WA SWP25180301WA SWP25180302WA SWP25180303WA SWP25180304WA SWP25180305WA SWP25180306WA SWP25180307WA SWP25180308WA SWP25180309WA SWP25180310WA SWP25180311WA SWP25180312WA SWP25180313WA SWP25180314WA SWP25180315WA SWP25180316WA SWP25180317WA SWP25180318WA SWP25180319WA SWP25180320WA SWP25180321WA SWP25180322WA SWP25180323WA SWP25180324WA SWP25180325WA SWP25180326WA SWP25180327WA SWP25180328WA SWP25180329WA SWP25180330WA SWP25180331WA SWP25180332WA SWP25180333WA SWP25180334WA SWP25180335WA SWP25180336WA SWP25180337WA SWP25180338WA SWP25180339WA SWP25180340WA SWP25180341WA SWP25180342WA SWP25180343WA SWP25180344WA SWP25180345WA SWP25180346WA SWP25180347WA SWP25180348WA SWP25180349WA SWP25180350WA SWP25180351WA SWP25180352WA SWP25180353WA SWP25180354WA SWP25180355WA SWP25180356WA SWP25180357WA SWP25180358WA SWP25180359WA SWP25180360WA SWP25180361WA SWP25180362WA SWP25180363WA SWP25190001WA SWP25190002WA SWP25190003WA SWP25190004WA SWP25190005WA SWP25190006WA SWP25190007WA SWP25190008WA SWP25190009WA SWP25190010WA SWP25190011WA SWP25190012WA SWP25190013WA SWP25190014WA SWP25190015WA SWP25190016WA SWP25190017WA SWP25190018WA SWP25190019WA SWP25190020WA SWP25190021WA SWP25190022WA SWP25190023WA SWP25190024WA SWP25190025WA SWP25190026WA SWP25190027WA SWP25190028WA SWP25190029WA SWP25190030WA SWP25190031WA SWP25190032WA SWP25190033WA SWP25190034WA SWP25190035WA SWP25190036WA SWP25190037WA SWP25190038WA SWP25190039WA SWP25190040WA SWP25190041WA SWP25190042WA SWP25190043WA SWP25190044WA SWP25190045WA SWP25190046WA SWP25190047WA SWP25190048WA SWP25190049WA SWP25190050WA SWP25190051WA SWP25190052WA SWP25190053WA SWP25190054WA SWP25190055WA SWP25190056WA SWP25190057WA SWP25190058WA SWP25190059WA SWP25190060WA SWP25190061WA SWP25190062WA SWP25190063WA SWP25190064WA SWP25190065WA SWP25190066WA SWP25190067WA SWP25190068WA SWP25190069WA SWP25190070WA SWP25190071WA SWP25190072WA SWP25190073WA SWP25190074WA SWP25190075WA SWP25190076WA SWP25190077WA SWP25190078WA SWP25190079WA SWP25190080WA SWP25190081WA SWP25190082WA SWP25190083WA SWP25190084WA SWP25190085WA SWP25190086WA SWP25190087WA SWP25190088WA SWP25190089WA SWP25190090WA SWP25190091WA SWP25190092WA SWP25190093WA SWP25190094WA SWP25190095WA SWP25190096WA SWP25190097WA SWP25190098WA SWP25190099WA SWP25190100WA SWP25190101WA SWP25190102WA SWP25190103WA SWP25190104WA SWP25190105WA SWP25190106WA SWP25190107WA SWP25190108WA SWP25190109WA SWP25190110WA SWP25190111WA SWP25190112WA SWP25190113WA SWP25190114WA SWP25190115WA SWP25190116WA SWP25190117WA SWP25190118WA SWP25190119WA SWP25190120WA SWP25190121WA SWP25190122WA SWP25190123WA SWP25190124WA SWP25190125WA SWP25190126WA SWP25190127WA SWP25190128WA SWP25190129WA SWP25190130WA SWP25190131WA SWP25190132WA SWP25190133WA SWP25190134WA SWP25190135WA SWP25190136WA SWP25190137WA SWP25190138WA SWP25190139WA SWP25190140WA SWP25190141WA SWP25190142WA SWP25190143WA SWP25190144WA SWP25190145WA SWP25190146WA SWP25190147WA SWP25190148WA SWP25190149WA SWP25190150WA SWP25220031WA SWP25220032WA SWP25220033WA SWP25220034WA SWP25220035WA SWP25220036WA SWP25220037WA SWP25220038WA SWP25220039WA SWP25220040WA SWP25220041WA SWP25220042WA SWP25220043WA SWP25220044WA SWP25220045WA SWP25220046WA SWP25220047WA SWP25220048WA SWP25220049WA SWP25220050WA SWP25220051WA SWP25220052WA SWP25220053WA SWP25220054WA SWP25220055WA SWP25220056WA SWP25220057WA SWP25220058WA SWP25220059WA SWP25220060WA SWP25220061WA SWP25220062WA SWP25220063WA SWP25220064WA SWP25220065WA SWP25220066WA SWP25220067WA SWP25220068WA SWP25220069WA SWP25220070WA SWP25220071WA SWP25220072WA SWP25220073WA SWP25220074WA SWP25220075WA SWP25220076WA SWP25220077WA SWP25220078WA SWP25220079WA SWP25220080WA SWP25220081WA SWP25220082WA SWP25220083WA SWP25220084WA SWP25220085WA SWP25220086WA SWP25220087WA SWP25220088WA SWP25220089WA SWP25220090WA SWP25220091WA SWP25220092WA SWP25220093WA SWP25220094WA SWP25220095WA SWP25220096WA SWP25220097WA SWP25220098WA SWP25220099WA SWP25220100WA SWP25220101WA SWP25220102WA SWP25220103WA SWP25220104WA SWP25220105WA SWP25220106WA SWP25220107WA SWP25220108WA SWP25220109WA SWP25220110WA SWP25220111WA SWP25220112WA SWP25220113WA SWP25220114WA SWP25220115WA SWP25220116WA SWP25220117WA SWP25220118WA SWP25220119WA SWP25220120WA SWP25220121WA SWP25220122WA SWP25220123WA SWP25220124WA SWP25220125WA SWP25220126WA SWP25220127WA SWP25220128WA SWP25220129WA SWP25220130WA SWP25220131WA SWP25220132WA SWP25220133WA SWP25220134WA SWP25220135WA SWP25220136WA SWP25220137WA SWP25220138WA SWP25220139WA SWP25220140WA SWP25220141WA SWP25220142WA SWP25220143WA SWP25220144WA SWP25220145WA SWP25220146WA SWP25220147WA SWP25220148WA SWP25220149WA SWP25220150WA SWP25220151WA SWP25220152WA SWP25220153WA SWP25220154WA SWP25220155WA SWP25220156WA SWP25220157WA SWP25220158WA SWP25220159WA SWP25220160WA SWP25220161WA SWP25220162WA SWP25220163WA SWP25220164WA SWP25220165WA SWP25220166WA SWP25220167WA SWP25220168WA SWP25220169WA SWP25220170WA SWP25220171WA SWP25220172WA SWP25220173WA SWP25220174WA SWP25220175WA SWP25220176WA SWP25220177WA SWP25220178WA SWP25220179WA SWP25220180WA SWP25220181WA SWP25220182WA SWP25220183WA SWP25220184WA SWP25220185WA SWP25220186WA SWP25220187WA SWP25220188WA SWP25220189WA SWP25220190WA SWP25220191WA SWP25220192WA SWP25220193WA SWP25220194WA SWP25220195WA SWP25220196WA SWP25220197WA SWP25220198WA SWP25220199WA SWP25220200WA SWP25220201WA SWP25220202WA SWP25220203WA SWP25220204WA SWP25220205WA SWP25220206WA SWP25220207WA SWP25220208WA SWP25220209WA SWP25220210WA SWP25220211WA SWP25220212WA SWP25220213WA SWP25220214WA SWP25220215WA SWP25220216WA SWP25220217WA SWP25220218WA SWP25220219WA SWP25220220WA SWP25220221WA SWP25220222WA SWP25220223WA SWP25220224WA SWP25220225WA SWP25220226WA SWP25220227WA SWP25220228WA SWP25220229WA SWP25220230WA SWP25220231WA SWP25220232WA SWP25220233WA SWP25220234WA SWP25220235WA SWP25220236WA SWP25220237WA SWP25220238WA SWP25220239WA SWP25220240WA SWP25220241WA SWP25220242WA SWP25220243WA SWP25220244WA SWP25220245WA SWP25220246WA SWP25220247WA SWP25220248WA SWP25220249WA SWP25220250WA SWP25220251WA SWP25220252WA SWP25220253WA SWP25220254WA SWP25220255WA SWP25220256WA SWP25220257WA SWP25220258WA SWP25220259WA SWP25220260WA SWP25220261WA SWP25220262WA SWP25220263WA SWP25220264WA SWP25220265WA SWP25220266WA SWP25220267WA SWP25220268WA SWP25220269WA SWP25220270WA SWP25220271WA SWP25220272WA SWP25220273WA SWP25220274WA SWP25220275WA SWP25220276WA SWP25220277WA SWP25220278WA SWP25220279WA SWP25220280WA SWP25220281WA SWP25220282WA SWP25220283WA SWP25220284WA SWP25220285WA SWP25220286WA SWP25220287WA SWP25220288WA SWP25220289WA SWP25220290WA SWP25220291WA SWP25220292WA SWP25220293WA SWP25220294WA SWP25220295WA SWP25220296WA SWP25220297WA SWP25220298WA SWP25220299WA SWP25220300WA SWP25220301WA SWP25220302WA SWP25220303WA SWP25220304WA SWP25220305WA SWP25220306WA SWP25220307WA SWP25220308WA SWP25220309WA SWP25220310WA SWP25220311WA SWP25220312WA SWP25220313WA SWP25220314WA SWP25220315WA SWP25220316WA SWP25220317WA SWP25220318WA SWP25220319WA SWP25220320WA SWP25220321WA SWP25220322WA SWP25220323WA SWP25220324WA SWP25220325WA SWP25220326WA SWP25220327WA SWP25220328WA SWP25220329WA SWP25220330WA SWP25220331WA SWP25220332WA SWP25220333WA SWP25220334WA SWP25220335WA SWP25220336WA SWP25220337WA SWP25220338WA SWP25220339WA SWP25220340WA SWP25220341WA SWP25220342WA SWP25220343WA SWP25220344WA SWP25220345WA SWP25220346WA SWP25220347WA SWP25220348WA SWP25220349WA SWP25220350WA SWP25220351WA SWP25220352WA SWP25220353WA SWP25220354WA SWP25220355WA SWP25220356WA SWP25220357WA SWP25220358WA SWP25220359WA SWP25220360WA SWP25220361WA SWP25220362WA SWP25220363WA SWP25220364WA SWP25220365WA SWP25220366WA SWP25220367WA SWP25220368WA SWP25220369WA SWP25220370WA SWP25220371WA SWP25220372WA SWP25220373WA SWP25220374WA SWP25220375WA SWP25220376WA SWP25220377WA SWP25220378WA SWP25220379WA SWP25220380WA SWP25220381WA SWP25220382WA SWP25220383WA SWP25220384WA SWP25220385WA SWP25220386WA SWP25220387WA SWP25220388WA SWP25220389WA SWP25220390WA SWP25220391WA SWP25220392WA SWP25220393WA SWP25220394WA SWP25220395WA SWP25220396WA SWP25220397WA SWP25220398WA SWP25220399WA SWP25220400WA SWP25220401WA SWP25220402WA SWP25220403WA SWP25220404WA SWP25220405WA SWP25220406WA SWP25220407WA SWP25220408WA SWP25220409WA SWP25220410WA SWP25220411WA SWP25220412WA SWP25220413WA SWP25220414WA SWP25220415WA SWP25220416WA SWP25220417WA SWP25220418WA SWP25220419WA SWP25220420WA SWP25220421WA SWP25220422WA SWP25220423WA SWP25220424WA SWP25220425WA SWP25220426WA SWP25220427WA SWP25220428WA SWP25220429WA SWP25220430WA SWP25230006WA SWP25230007WA SWP25230008WA SWP25230009WA SWP25230010WA SWP25230011WA SWP25230012WA SWP25230013WA SWP25230014WA SWP25230015WA SWP25230016WA SWP25230017WA SWP25230018WA SWP25230019WA SWP25230020WA SWP25230021WA SWP25230022WA SWP25230023WA SWP25230024WA SWP25230025WA SWP25230026WA SWP25230027WA SWP25230028WA SWP25230029WA SWP25230030WA SWP25230031WA SWP25230032WA SWP25230033WA SWP25230034WA SWP25230035WA SWP25230036WA SWP25230037WA SWP25230038WA SWP25230039WA SWP25230040WA SWP25230041WA SWP25230042WA SWP25230043WA SWP25230044WA SWP25230045WA SWP25230046WA SWP25230047WA SWP25230048WA SWP25230049WA SWP25230050WA SWP25230051WA SWP25230052WA SWP25230053WA SWP25230054WA SWP25230055WA SWP25230056WA SWP25230057WA SWP25230058WA SWP25230059WA SWP25230060WA SWP25230061WA SWP25230062WA SWP25230063WA SWP25230064WA SWP25230065WA SWP25230066WA SWP25230067WA SWP25230068WA SWP25230069WA SWP25230070WA SWP25230071WA SWP25230072WA SWP25230073WA SWP25230074WA SWP25230075WA SWP25230076WA SWP25230077WA SWP25230078WA SWP25230079WA SWP25230080WA SWP25230081WA SWP25230082WA SWP25230083WA SWP25230084WA SWP25230085WA SWP25230086WA SWP25230087WA SWP25230088WA SWP25230089WA SWP25230090WA SWP25230091WA SWP25230092WA SWP25230093WA SWP25230094WA SWP25230095WA SWP25230096WA SWP25230097WA SWP25230098WA SWP25230099WA SWP25230100WA SWP25230101WA SWP25230102WA SWP25230103WA SWP25230104WA SWP25230105WA SWP25090007WA SWP25090008WA SWP25090009WA SWP25090010WA SWP25090011WA SWP25090012WA SWP25040001WA SWP25040002WA SWP25040003WA SWP25040004WA SWP25110047WA SWP25080124WA SWP25110061WA SWP25110062WA SWP25110063WA SWP25110065WA SWP25110066WA SWP25110067WA SWP25110071WA SWP25110073WA SWP25110076WA SWP25040045WA SWP25040046WA SWP25040047WA SWP25040048WA SWP25040049WA SWP25040050WA SWP25040051WA SWP25040052WA SWP25040053WA SWP25040054WA SWP25040055WA SWP25040056WA SWP25040057WA SWP25040058WA SWP25040059WA SWP25040060WA SWP25040061WA SWP25100001WA SWP25100002WA SWP25100003WA SWP25100004WA SWP25110060WA SWP25110064WA SWP25110068WA SWP25110069WA SWP25110070WA SWP25110072WA SWP25110074WA SWP25110075WA SWP25110077WA SWP25140418WA SWP25140419WA SWP25140420WA SWP25140421WA SWP25140422WA SWP25140423WA SWP25140424WA SWP25140425WA SWP25140426WA SWP25140427WA SWP25140428WA SWP25140429WA SWP25140430WA SWP25140431WA SWP25140432WA SWP25140433WA SWP25140434WA SWP25140435WA SWP25140436WA SWP25140437WA SWP25140438WA SWP25140439WA SWP25140440WA SWP25140441WA SWP25140442WA SWP25140443WA SWP25140444WA SWP25140445WA SWP25140446WA SWP25140447WA SWP25140448WA SWP25140449WA SWP25140450WA SWP25140451WA SWP25140452WA SWP25140453WA SWP25140454WA SWP25140455WA SWP25140456WA SWP25140457WA SWP25180053WA SWP25180056WA SWP25180057WA SWP25210033WA SWP25210034WA SWP25210035WA SWP25230106WA SWP25230107WA SWP25230108WA SWP25230109WA SWP25230110WA SWP25230111WA SWP25230112WA SWP25250062WA SWP25270002WA SWP25270003WA SWP25270004WA SWP25270005WA SWP25270006WA SWP25270007WA SWP25270008WA SWP25270009WA SWP25270010WA SWP25270011WA SWP25270012WA SWP25270013WA SWP25270014WA SWP25270015WA SWP25270016WA SWP25270017WA SWP25270018WA SWP25270019WA SWP25270020WA SWP25270021WA SWP25270076WA SWP25270077WA SWP25270078WA SWP25270079WA SWP25270080WA SWP25270081WA SWP25270082WA SWP25270083WA SWP25270084WA SWP25270085WA SWP25270086WA SWP25031782WA SWP25070104WA SWP25070105WA SWP25070106WA SWP25120103WA SWP25120104WA SWP25120105WA SWP25120106WA SWP25120107WA SWP25180054WA SWP25180055WA SWP25080003WA SWP25080004WA SWP25080005WA SWP25080006WA SWP25080007WA SWP25160055WA SWP25160056WA SWP25160057WA SWP25160058WA SWP25160059WA SWP25270022WA SWP25270023WA SWP25270024WA SWP25270025WA SWP25270026WA SWP25270027WA SWP25270028WA SWP25270029WA SWP25270030WA SWP25270031WA SWP25270032WA SWP25270033WA SWP25270034WA SWP25270035WA SWP25270036WA SWP25270037WA SWP25270038WA SWP25270039WA SWP25270040WA SWP25270041WA SWP25270042WA SWP25270043WA SWP25270049WA SWP25270050WA SWP25270051WA SWP25030141WA SWP25030142WA SWP25030143WA SWP25030144WA SWP25030145WA SWP25220006WA SWP25220007WA SWP25220008WA SWP25220009WA SWP25170062WA SWP25170063WA SWP25210051WA SWP25210052WA SWP25090128WA SWP25220431WA SWP25090124WA SWP25090125WA SWP25090126WA SWP25090127WA SWP25230114WA SWP25230115WA SWP25250091WA SWP25250092WA SWP25160049WA SWP25070230WA SWP25160103WA SWP25250093WA SWP25210043WA SWP25210044WA SWP25210045WA SWP25210046WA SWP25210047WA SWP25210048WA SWP25210049WA SWP25210050WA SWP25080008WA SWP25160050WA SWP25230113WA SWP25280028WA SWP25160060WA SWP25160061WA SWP25070216WA SWP25070217WA SWP25070218WA SWP25070219WA SWP25070220WA SWP25070221WA SWP25070222WA SWP25070223WA SWP25070224WA SWP25070225WA SWP25070226WA SWP25070227WA SWP25070231WA SWP25070232WA SWP25070233WA SWP25070234WA SWP25070235WA SWP25070236WA SWP25070237WA SWP25080009WA SWP25110050WA SWP25110051WA SWP25110052WA SWP25110053WA SWP25110054WA SWP25110055WA SWP25110056WA SWP25110057WA SWP25110058WA SWP25110059WA SWP25110078WA SWP25110079WA SWP25110080WA SWP25110081WA SWP25110082WA SWP25110083WA SWP25110084WA SWP25110085WA SWP25110086WA SWP25110087WA SWP25250070WA SWP25250071WA SWP25250072WA SWP25250073WA SWP25250074WA SWP25250075WA SWP25250076WA SWP25250077WA SWP25250078WA SWP25250079WA SWP25250080WA SWP25250081WA SWP25250082WA SWP25250083WA SWP25250084WA SWP25250085WA SWP25250086WA SWP25250087WA SWP25250088WA SWP25250089WA SWP25210004SA SWP25290068SA SWP25290069SA SWP25290070SA SWP25290071SA SWP25260005WA SWP25120108WA SWP25120109WA SWP25130116WA SWP25130117WA SWP25130118WA SWP25130119WA SWP25130120WA SWP25140412WA SWP25140413WA SWP25140414WA SWP25160014WA SWP25160015WA SWP25160016WA SWP25160017WA SWP25160018WA SWP25160019WA SWP25160020WA SWP25160021WA SWP25160022WA SWP25160023WA SWP25160024WA SWP25160025WA SWP25160026WA SWP25160027WA SWP25160028WA SWP25160029WA SWP25160030WA SWP25160031WA SWP25160032WA SWP25160033WA SWP25160034WA SWP25160035WA SWP25160036WA SWP25160037WA SWP25160038WA SWP25160039WA SWP25160040WA SWP25160041WA SWP25160042WA SWP25270111WA SWP25270112WA SWP25270113WA SWP25270114WA SWP25270115WA SWP25270116WA SWP25270117WA SWP25270118WA SWP25270119WA SWP25270120WA SWP25290004WA SWP25290005WA SWP25290006WA SWP25290007WA SWP25290008WA SWP25270091WA SWP25270092WA SWP25270093WA SWP25270094WA SWP25270095WA SWP25270096WA SWP25270097WA SWP25270098WA SWP25270099WA SWP25270100WA SWP25270101WA SWP25270102WA SWP25270103WA SWP25270104WA SWP25270105WA SWP25270106WA SWP25270107WA SWP25270108WA SWP25270109WA SWP25270110WA SWP25120606SA SWP25120607SA SWP25120608SA SWP25120609SA SWP25120610SA SWP25120611SA SWP25120612SA SWP25120613SA SWP25120614SA SWP25120615SA SWP25280029WA SWP25280030WA SWP25280031WA SWP25280032WA SWP25280033WA SWP25280034WA SWP25280035WA SWP25280036WA SWP25280037WA SWP25280038WA SWP25280039WA SWP25280040WA SWP25280041WA SWP25280042WA SWP25120091WA SWP25120092WA SWP25120093WA SWP25120094WA SWP25120095WA SWP25120096WA SWP25120097WA SWP25120098WA SWP25120099WA SWP25120100WA SWP25160076WA SWP25160077WA SWP25160078WA SWP25160079WA SWP25160080WA SWP25160104WA SWP25160105WA SWP25160106WA SWP25160107WA SWP25160108WA SWP25160109WA SWP25160110WA SWP25160111WA SWP25160112WA SWP25040015WA SWP25040016WA SWP25040017WA SWP25040018WA SWP25040019WA SWP25040020WA SWP25040021WA SWP25040022WA SWP25040023WA SWP25040024WA SWP25170051WA SWP25170052WA SWP25160083WA SWP25160084WA SWP25170036WA SWP25170037WA SWP25170038WA SWP25170039WA SWP25080012WA SWP25270054WA SWP25210026WA SWP25300025WA SWP25300026WA SWP25300027WA SWP25160066WA SWP25170041WA SWP25170042WA SWP25170043WA SWP25170044WA SWP25170045WA SWP25170023WA SWP25170024WA SWP25170025WA SWP25170026WA SWP25170027WA SWP25170028WA SWP25170029WA SWP25170030WA SWP25170031WA SWP25170032WA SWP25170035WA SWP25270052WA SWP25160065WA SWP25040044WA SWP25160085WA SWP25250064WA SWP25160087WA SWP25160088WA SWP25160089WA SWP25040026WA SWP25080125WA SWP25210025WA SWP25170049WA SWP25170050WA SWP25040011WA SWP25040012WA SWP25040013WA SWP25040014WA SWP25040010WA SWP25160081WA SWP25160113WA SWP25180058WA SWP25240007WA SWP25300014WA SWP25160074WA SWP25160075WA SWP25140415WA SWP25170047WA SWP25170048WA SWP25300015WA SWP25300028WA SWP25090129WA SWP25270074WA SWP25270053WA SWP25160067WA SWP25160068WA SWP25260006WA SWP25210023WA SWP25170046WA SWP25210061WA SWP25170040WA SWP25160069WA SWP25160070WA SWP25160071WA SWP25060044WA SWP25060045WA SWP25060046WA SWP25060047WA SWP25060048WA SWP25060049WA SWP25060050WA SWP25060051WA SWP25060052WA SWP25060053WA SWP25060054WA SWP25060055WA SWP25060056WA SWP25060057WA SWP25060058WA SWP25060059WA SWP25060060WA SWP25060061WA SWP25060062WA SWP25060064WA SWP25060065WA SWP25060066WA SWP25060067WA SWP25060068WA SWP25060069WA SWP25060070WA SWP25060071WA SWP25060072WA SWP25060073WA SWP25060074WA SWP25060075WA SWP25060076WA SWP25100005WA SWP25100006WA SWP25100007WA SWP25100008WA SWP25100009WA SWP25100010WA SWP25100011WA SWP25100012WA SWP25100013WA SWP25100014WA SWP25170033WA SWP25170034WA SWP25180059WA SWP25180060WA SWP25180061WA SWP25180062WA SWP25180063WA SWP25290009WA SWP25040025WA SWP25080010WA SWP25080011WA SWP25090113WA SWP25090114WA SWP25090115WA SWP25090116WA SWP25160114WA SWP25200003WA SWP25220010WA SWP25090145WA SWP25090146WA SWP25090147WA SWP25090148WA SWP25090149WA SWP25090150WA SWP25090151WA SWP25090152WA SWP25090153WA SWP25090154WA SWP25090155WA SWP25090156WA SWP25090157WA SWP25090158WA SWP25090159WA SWP25090160WA SWP25090161WA SWP25090162WA SWP25090163WA SWP25090164WA SWP25090165WA SWP25090166WA SWP25090167WA SWP25090168WA SWP25090169WA SWP25090170WA SWP25090171WA SWP25090172WA SWP25090173WA SWP25090174WA SWP25090175WA SWP25090176WA SWP25160072WA SWP25160073WA SWP25210024WA SWP25070109WA SWP25070110WA SWP25070111WA SWP25070112WA SWP25170053WA SWP25170054WA SWP25290060WA SWP25160082WA SWP25160062WA SWP25160063WA SWP25160064WA SWP25290061WA SWP25160086WA SWP25300033WA (Updated 3/12/26 - additional serial nos. SWP24140086WA SWP24140087WA)
FEI Number 3002806783
Recalling Firm/
Manufacturer
GE Healthcare Finland Oy
Kuortaneenkatu 2
Helsinki Finland
For Additional Information ContactGE HealthCare Service
800-4371171 Ext. 2436
Manufacturer Reason
for Recall
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
FDA Determined
Cause 2
Component change control
ActionOn September 26, 2025, the recalling firm began notifying affected customers. A second letter was issued on October 3, 2025 which addressed a correction to Affected Product Details information. The customer letter advised users to use a monitor with a battery that is not within the affected battery population. However, pending correction by GE HealthCare, users may continue to use a monitor with affected batteries by following instructions included in the letter. These instructions include having mains power connected whenever possible, confirming appropriate battery status prior to use, and being aware that the monitor may still shut down during patient transport. Users are advised to be prepared to provide alternate methods of assessing the clinical status of the patient, bring a mains power cord during patient transport, and bring a spare battery. GE HealthCare will correct all affected products at no cost to the customer. A GE HealthCare representative will contact each customer to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce5990
DistributionWorldwide - US Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = MHX
-
-