| | Class 2 Device Recall da Vinci S, Si Monopolar Curved Scissors Instrument |  |
| Date Initiated by Firm | December 09, 2025 |
| Date Posted | March 30, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1696-2026 |
| Recall Event ID |
98383 |
| 510(K)Number | K050369 K180033 |
| Product Classification |
System, surgical, computer controlled instrument - Product Code NAY
|
| Product | Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument
Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument
Model/Catalog Number: 420179
Software Version: N/A
Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use.
All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient).
Component: N/A |
| Code Information |
Lot Code: 420179: Monopolar Curved Scissors (Hot Shears)
GTIN: 00886874111505
FDA Medical Device Number: D088920
Material Number: 420179-10
Batch / System Numbers:
M10101102 SH0935
M10101201 SH0935
M10101202 SH1279
M10101203 SH0059
M10101204 USG294
M10101206 USG294
M10101206 SH1378
M10101214 USG022
M10101214 SH0035
M10101214 SH0495
M10110104 SH1279
M10110111 SH1378
M10110112 USG294
M10110113 SH0075
M10110121 SG704
M10110121 SH1279
M10110124 SH0495
M10110126 USG721
M10110126 SH1391
M10110127 USG721
M10110127 SH1279
M10110127 SH2144
M10110202 USG294
M10110202 SH0495
M10110203 SH2144
M10110208 SG704
M10110210 SH1378
M10110216 SH1368
M10110216 USG294
M10110216 SH1391
M10110217 SH0495
M10110217 SH1279
M10110222 SH1279
M10110228 SH0547
M10110228 USG721
M10110228 USG022
M10110228 SH1378
M10110228 SH1279
M10110301 SH1279
M10110303 SH0495
M10110304 USG294
M10110304 SH0729
M10110304 SH0761
M10110307 USG294
M10110308 USG294
M10110308 SH0035
M10110308 SH0495
M10110308 SH1279
M10110311 USG294
M10110311 SH2144
M10110315 USG294
M10110318 USG294
M10110318 SH0935
M10110318 SH1279
M10110323 USG022
M10110324 SH0059
M10110324 USG022
M10110324 SH0495
M10110328 USG294
M10110328 SH0935
M10110328 SH1279
M10110329 USG294
M10110329 SH0935
M10110331 SH0935
M10110405 SH1378
M10110413 USG294
M10110413 SH1279
M10110414 SG704
M10110414 SH0495
M10110420 SH1279
M10110425 SH0035
M10110426 SH0059
M10110428 USG022
M10110429 SG2026
M10110429 SH0035
M10110504 USG294
M10110512 USG294
M10110512 USG022
M10110512 SH2144
M10110517 USG294
M10110517 SH0495
M10110518 SH0035
M10110524 USG294
M10110524 SH1279
M10110525 USG294
M10110526 USG294
M10110526 SH1378
M10110602 SH1279
M10110603 SH1378
M10110614 SH1279
M10110616 SH1368
M10110617 SH1182
M10110617 SH0495
M10110621 SH2144
M10110622 SH0935
M10110623 SH0729
M10110624 SH1378
M10110629 USG294
M10110630 SH1279
M10110630 SH2144
M10110701 SH1378
M10110705 SH0035
M10110711 USG294
M10110712 SH1378
M10110712 SH0495
M10110715 SH1279
M10110715 SH2144
M10110718 SH0761
M10110721 SH0935
M10110728 SH0729
M10110728 SH0761
M10110729 USG721
M10110801 SH1279
M10110802 SH0935
M10110803 USG294
M10110803 USG022
M10110803 SH2144
M10110805 SG704
M10110805 SH0495
M10110808 SH0035
M10110809 USG721
M10110810 SH1279
M10110811 USG022
M10110811 SH0729
M10110812 USG721
M10110812 SH0729
M10110815 SH0935
M10110815 SH1279
M10110816 SG704
M10110816 USG721
M10110816 SH0729
M10110819 USG294
M10110822 SG704
M10110822 USG294
M10110823 SH0729
M10110824 SH0059
M10110824 SH0729
M10110824 SH0495
M10110825 USG721
M10110826 USG721
M10110829 SH1279
M10110830 USG022
M10110902 SG704
M10110902 SH1279
M10110902 SH0761
M10110906 USG022
M10110908 USG721
M10110908 SH1378
M10110908 SH1279
M10110909 SH0495
M10110913 USG721
M10110922 SH2144
M10110923 USG294
M10110923 SH0729
M10110930 SH0935
M10110930 SH0495
M10111003 SG704
M10111004 USG294
M10111005 USG294
M10111006 SH0729
M10111007 SH1368
M10111007 SH0035
M10111007 SH2144
M10111010 SH0935
M10111011 SH0495
M10111011 SH1279
M10111013 SH1378
M10111015 USG022
M10111017 SH1015
M10111018 SH1368
M10111018 SH1279
M10111019 USG294
M10111020 SH0547
M10111020 SH1279
M10111021 USG022
M10111027 SG704
M10111027 SH1378
M10111028 SH0059
M10111028 USG294
M10111028 SH2144
M10111031 SH1015
M10111104 SH1182
M10111107 SH0059
M10111107 SH0729
M10111107 SH0761
M10111108 SH1279
M10111108 SH2144
M10111109 SH0495
M10111111 SH0935
M10111114 USG294
M10111114 SH0729
M10111114 SH1279
M10111115 SG704
M10111115 USG294
M10111115 USG022
M10111116 SH0825
M10111117 SH0729
M10111117 SH1391
M10111117 SH0761
M10111118 SH0935
M10111122 SH0495
M10111122 SH0761
M10111123 SH1390
M10111123 USG294
M10111123 SH0035
M10111124 USG294
M10111129 SH0935
M10111130 SH1378
M10111201 SH1368
M10111202 USG294
M10111202 SH0729
M10111208 USG294
M10111208 SH1279
M10111212 SH1378
M10111213 SG704
M10111213 SH1279
M10111214 USG294
M10111214 SH0729
M10111214 SH1015
M10111216 SH0495
M10111221 USG294
M10111221 SH1378
M10120104 SH1182
M10120104 USG022
M10120104 SH1279
M10120105 USG294
M10120109 USG022
M10120111 SH2144
M10120112 USG721
M10120113 SH0495
M10120117 SH0495
M10120117 SH1279
M10120127 USG294
M10120130 SG704
M10120130 SH1279
M10120131 SH0729
M10120201 SH1390
M10120201 SH0729
M10120203 SH0935
M10120208 SH0729
M10120208 SH1279
M10120208 SH2144
M10120209 SH1378
M10120213 SH0495
M10120214 SH1279
M10120216 SG704
M10120216 SH0495
M10120217 SH1182
M10120217 SH0761
M10120221 SH0729
M10120222 SH1391
M10120224 SH1390
M10120224 USG294
M10120224 SH0729
M10120228 SH1279
M10120301 SH1015
M10120305 SH1279
M10120308 SH0729
M10120309 SH0935
M10120312 USG022
M10120312 SH0935
M10120312 SH1015
M10120313 SH0825
M10120315 SH1279
M10120315 SH2144
M10120316 SH1390
M10120316 SH0729
M10120316 SH1279
M10120317 SH1276
M10120320 SG704
M10120320 SH1368
M10120320 SH1378
M10120320 SH0729
M10120321 SG2026
M10120326 SH1076
M10120327 SH0935
M10120330 SH0059
M10120330 USG294
M10120402 SH1279
M10120403 SH0729
M10120403 SH0495
M10120404 SH1279
M10120409 SH2144
M10120411 SG2026
M10120411 SH2144
M10120412 SH0935
M10120412 SH0761
M10120413 SH1276
M10120416 SH1276
M10120423 USG294
M10120424 USG294
M10120424 SH0035
M10120425 SH1015
M10120425 SH1279
M10120426 SH1390
M10120501 SG704
M10120501 USG721
M10120503 USG022
M10120503 SH0729
M10120503 SH2144
M10120503 SH1276
M10120507 SH1279
M10120515 SH1182
M10120515 SG704
M10120515 USG721
M10120516 SH2144
M10120518 SH1368
M10120521 SH0761
M10120522 USG294
M10120524 USG294
M10120524 SH0729
M10120528 USG294
M10120528 SH1279
M10120529 SH0935
M10120529 SH1276
M10120604 SH0495
M10120606 SH1378
M10120607 SH0547
M10120607 SH0825
M10120612 USG294
M10120612 SH0035
M10120612 SH1279
M10120612 SH0761
M10120614 SH1390
M10120614 SH1368
M10120615 SH0729
M10120622 SG704
M10120622 SH1378
M10120622 SH1015
M10120626 USG294
M10120626 SH1279
M10120627 SH2144
M10120629 USG022
M10120629 SH1378
M10120629 SH0935
M10120629 SH0495
M10120629 SH0761
M10120702 SH0729
M10120702 SH1015
M10120705 SH0059
M10120705 SH1368
M10120705 USG294
M10120705 SH0495
M10120705 SH0761
M10120706 SH0935
M10120706 SH1391
M10120709 SH1390
M10120711 USG294
M10120711 SH1076
M10120713 USG294
M10120713 SH0729
M10120713 SH1279
M10120713 SH2144
M10120713 SH0761
M10120719 USG294
M10120719 SH1279
M10120723 SG704
M10120726 USG721
M10120730 SH0729
M10120730 SH0035
M10120730 SH0761
M10120731 SH1385
M10120801 SH2144
M10120802 SH0059
M10120807 SH1378
M10120808 SH1076
M10120809 SH0729
M10120809 SH1276
M10120810 USG294
M10120813 USG721
M10120815 SH1378
M10120821 SH0935
M10120822 SH1378
M10120822 SH0935
M10120823 USG022
M10120823 SH0729
M10120823 SH0761
M10120827 USG294
M10120828 SH1279
M10120903 SG704
M10120903 SH1276
M10120904 SH0495
M10120906 USG022
M10120911 SH0825
M10120919 SH1276
M10120920 SH0729
M10120920 SH0035
M10120925 SH0495
M10120925 SH1015
M10120928 SH0729
M10121001 SH0059
M10121002 SH1015
M10121004 USG294
M10121004 SH0729
M10121004 SH0495
M10121004 SH1385
M10121006 SH1391
M10121008 SH1279
M10121008 SH0761
M10121010 USG294
M10121012 USG294
M10121015 SH1378
M10121016 SH0495
M10121017 SH2144
M10121019 SH1368
M10121019 SH0825
M10121019 SH0935
M10121023 SH0729
M10121025 USG022
M10121029 SH0729
M10121029 SH1391
M10121029 SH0761
M10121031 USG294
M10121031 SH0935
M10121102 SH1279
M10121106 USG294
M10121106 USG721
M10121112 USG721
M10121113 USG689
M10121114 SG704
M10121114 USG294
M10121115 SH1390
M10121115 USG721
M10121115 SH1391
M10121121 USG294
M10121123 SH0059
M10121123 SH1390
M10121123 SH0495
M10121126 SH0935
M10121128 SH0059
M10121128 SH0729
M10121129 SH1276
M10121204 USG022
M10121205 SH2144
M10121206 SH0059
M10121215 SH1378
M10121215 SH1391
M10121215 SH0495
M10121218 USG294
M10121219 SH0547
M10130102 SH0547
M10130102 SH0059
M10130102 SH1279
M10130104 USG294
M10130108 SH1390
M10130108 USG022
M10130108 SH1279
M10130109 SH1368
M10130110 SH1368
M10130110 SH0935
M10130114 SH0729
M10130115 USG294
M10130116 SH1570
M10130116 SH0935
M10130117 SH0059
M10130117 SH0825
M10130117 SH0729
M10130117 SH0495
M10130118 USG294
M10130121 SH1390
M10130122 USG721
M10130122 SH1015
M10130123 SH1182
M10130124 SH1378
M10130124 SH0761
M10130205 USG022
M10130207 SH1391
M10130207 SH1015
M10130208 SG704
M10130211 USG689
M10130212 SH0729
M10130212 SH0761
M10130213 SH0547
M10130214 SH0729
M10130214 SH0495
M10130215 SH1182
M10130215 SH1378
M10130222 SH1390
M10130227 SH0495
M10130228 SH1622
M10130304 SH0825
M10130304 SH0495
M10130305 SH1570
M10130306 SH0761
M10130306 SH1276
M10130307 USG294
M10130308 SG2026
M10130308 SH1378
M10130308 SH1279
M10130313 USG294
M10130313 SH0935
M10130315 SH2144
M10130315 SH0761
M10130318 SH0059
M10130319 SH1279
M10130319 SH2144
M10130322 SH1279
M10130402 USG022
M10130404 USG022
M10130408 SH1378
M10130408 SH0729
M10130409 SH0059
M10130409 SH0761
M11110225 SG704
M11110323 SH1368
M11110517 SH0729
M11110517 SH1279
M11110517 SH2144
M11110525 SH0761
M11110526 SG704
M11110531 SH0495
M11110602 USG294
M11110602 SH0729
M11110623 SH1368
M11110623 SH1279
M11110625 SH0547
M11110625 USG294
M11110630 SG704
M11110630 USG294
M11110701 USG721
M11110701 SH0495
M11110701 SH0761
M11110808 SH0495
M11110808 SH0761
M11110809 USG294
M11110809 SH1391
M11110809 SH2144
M11110810 SH0761
M11110825 SH0059
M11110825 SH0761
M11110826 SH2144
M11110929 SG2026
M11110929 SH1279
M11111026 SH0729
M11111027 USG294
M11111028 SH0935
M11111028 SH0495
M11111117 USG022
M11111117 SH0729
M11111117 SH0761
M11111124 SH1015
M11111129 USG721
M11111130 USG294
M11111130 SH2144
M11120103 SH0059
M11120103 SH0729
M11120116 USG294
M11120116 SH0495
M11120126 SH0035
M11120127 SH1279
M11120130 USG294
M11120131 SH1182
M11120208 USG294
M11120209 SH0761
M11120210 SH0935
M11120213 SH0729
M11120214 SG2026
M11120330 USG294
M11120402 SH1391
M11120403 USG022
M11120404 SH1279
M11120517 USG294
M11120517 SH1279
M11120522 SH1368
M11120522 SH1276
M11120607 USG294
M11120607 USG022
M11120608 USG022
M11120717 USG294
M11120801 SH1015
M11120813 SH1368
M11120814 SH0729
M11120814 SH1276
M11120816 SH0729
M11120821 USG294
M11120906 SH1378
M11120919 USG689
M11120928 SH0935
M11120928 SH1276
M11121004 USG294
M11121004 SH1015
M11121004 SH1279
M11121010 USG294
M11121011 SH1182
M11121015 SH1378
M11121016 SH2144
M11121114 SH1557
M11121114 SH1378
M11121114 SH0935
M11121114 SH1279
M11121114 SH1276
M11121115 SH1076
M11121116 SH1279
M11121129 USG294
M11121129 SH0761
M11121130 SH0059
M11121203 USG294
M11121205 SH0035
M11121206 SH1604
M11121206 SH1385
M11121211 SG704
M11121211 SH0935
M11121211 SH2144
M11121214 SH1622
M11121218 SH0495
M11121219 SH1279
M11130112 USG294
M11130121 SH1276
M11130124 USG294
M11130124 SH0729
M11130124 SH1279
M11130207 USG721
M11130211 SH1378
M11130213 USG294
M11130214 SH2144
M11130225 SH0729
M11130226 USG721
M11130227 SH0059
M11130227 USG294
M11130228 SH0729
M11130301 SG704
M11130306 SH0059
M11130307 SH1378
M11130315 SH1378
M11130315 SH1391
M11130315 SH1015
M11130322 SH0729
M11130322 SH1276
M11130402 SH1279
M11130404 SH1015
M12110228 SH1279
M12120614 USG294
M12120614 SH1378
M12120801 SH1378
M12120816 SH1279
M12120906 SH1279
M12121011 SH1276
M12121105 SH0495
M12121207 SH1368
M12121211 SH0729
M12130109 SH0495
M12130111 SG704
M12130213 SH1279
M12130221 USG721
M12130226 SH0729
M12130226 SH0035
M12130320 USG294
M12130320 SH0935
M12130320 SH2144
M13120801 SH1390
M13120801 SH0495
M13121206 SH1378
M13130110 USG294
M13130110 SH0761
M13130111 USG721
M13130111 SH0935
M13130112 SH0035
M20111026 SH0935
M20111118 USG294
M20111206 SH0935
M20111206 SH1279
M20120116 USG294
M20120116 SH2144
M20120125 SH1391
M20120125 SH1076
M20120306 SH0495
M20120307 SH0935
M20120307 SH1279
M20120308 SH0035
M20120314 USG294
M20120314 SH0495
M20120514 USG721
M20120514 SH0825
M20120514 SH0729
M20120514 SH1276
S12110708 USG294
Material Number: 420179-12
Batch / System Numbers:
M10130601 SH1390
M10130603 SH0825
M10130603 SH0495
M10130605 USG022
M10130605 SH2144
M10130606 SH1570
M10130611 SH1390
M10130612 SH2144
M10130612 SH1276
M10130614 SH0935
M10130619 SG704
M10130620 SH1378
M10130621 SH0729
M10130622 SH1276
M10130703 SH1182
M10130703 SH0035
M10130704 SH2144
M10130710 SH1378
M10130713 SH0729
M10130713 SH1385
M10130713 SH0761
M10130716 SH0761
M10130716 SH1276
M10130717 SG2026
M10130717 SH0935
M10130717 SH1279
M10130718 SH0729
M10130722 USG294
M10130722 SH0761
M10130725 USG022
M10130725 SH1279
M10130726 SH1557
M10130726 USG294
M10130726 SH1391
M10130731 USG294
M10130731 SH0495
M10130801 SH1182
M10130802 SH0495
M10130805 SH0059
M10130805 SH0729
M10130806 USG294
M10130806 SH1391
M10130812 SH1276
M10130814 SH0035
M10130815 SH1378
M10130815 SH0729
M10130815 SH2144
M10130816 SH1570
M10130816 USG721
M10131105 USG294
M10131111 SH0935
M10131111 SH0729
M10131115 USG294
M10131115 SH1279
M10131119 USG022
M10131120 USG294
M10131120 SH1378
M10131120 SH1279
M10131121 SG704
M10131121 USG022
M10131122 SH0935
M10131122 SH0761
M10131123 SH0547
M10131123 SH1837
M10131123 SH1390
M10131128 SH0825
M10131128 SH1378
M10131202 SH0761
M10131202 SH1276
M10131204 SH1604
M10131206 SH1182
M10131206 SH0547
M10131206 SH2144
M10131207 SH0729
M10131207 SH1279
M10131207 SH1276
M10131211 USG294
M10131212 SH0059
M10131212 USG022
M10131212 SH0729
M10131212 SH1279
M10131214 SG704
M10131216 SH0729
M10131217 SH1570
M10131217 SH1557
M10131217 SH0729
M10131217 SH1279
M10131217 SH2144
M10131218 SH1368
M10140108 SH0761
M10140113 SH1378
M10140115 SH1390
M10140115 SH0935
M10140116 SH1557
M10140116 USG721
M10140116 SH1622
M10140117 USG721
M10140120 USG294
M10140120 USG022
M10140123 USG294
M10140123 SH1015
M10140212 SH0547
M10140212 SH1368
M10140212 USG294
M10140212 SH0761
M10140213 SG704
M10140213 SH0729
M10140213 SH0495
M10140213 SH1279
M10140217 SH1276
M10140306 SH1570
M10140328 SH1378
M10140328 SH1391
M10140401 SH1557
M10140401 SH0729
M10140402 SH1385
M10140402 SH1279
M10140416 SH1997
M10140416 SH1391
M11130527 SH1279
M11130603 SH0059
M11130603 SH0729
M11130606 USG721
M11130606 SH1385
M11130607 SH1837
M11130607 SH1279
M11130611 SH0059
M11130613 USG721
M11130613 SH1391
M11130614 SH1570
M11130614 SH1837
M11130618 SH0059
M11130618 SH1390
M11130622 SH1390
M11130622 SH1279
M11130622 SH1276
M11130703 SH1390
M11130708 USG022
M11130709 USG294
M11130710 SH1837
M11130710 SH0935
M11130710 SH1279
M11130711 SH0495
M11130712 USG294
M11130713 SH1368
M11130713 SH0729
M11130716 SH0059
M11130716 SH0729
M11130716 SH2144
M11130718 USG689
M11130718 USG294
M11130722 SH0825
M11130723 SH1378
M11130724 USG294
M11130726 SG704
M11130726 SH1557
M11130726 SH0761
M11130730 SH1390
M11130805 SH1368
M11130805 SG2026
M11130809 USG689
M11130813 SG704
M11130813 USG294
M11130814 SH1837
M11130814 USG022
M11130815 SH0729
M11130815 SH1015
M11130816 USG294
M11130816 SH0729
M11130816 SH1279
M11131111 SH0495
M11131111 USG518
M11131202 USG294
M11131202 SH1378
M11131204 SH0935
M11131205 SH1391
M11131205 SH1279
M11131207 USG721
M11131207 SH0935
M11131207 SH0729
M11131207 SH0495
M11131207 SH2144
M11140108 USG294
M11140108 SH0495
M11140116 SH1378
M11140116 SH1279
M11140211 SH1557
M11140211 SH1378
M11140228 USG721
M11140228 SH1279
M12130531 SH0761
M12130603 SH0729
M12130608 SH0761
M12130610 SG704
M12130611 SH0059
M12130613 USG022
M12130613 SH1391
M12130615 USG518
M12130620 SH0935
M12130620 SH1385
M12130621 SH2144
M12130624 USG294
M12130624 SH1015
M12130702 SH0729
M12130704 SH0495
M12130705 SH1279
M12130708 SH1385
M12130709 SH1276
M12130710 SH1076
M12130710 SH0495
M12130712 USG721
M12130712 SH1385
M12130715 SH0547
M12130715 USG022
M12130717 SH0729
M12130717 SH1391
M12130722 SH0035
M12130726 SG704
M12130726 SH1279
M12130729 SH0729
M12130801 SH1182
M12130802 USG294
M12130802 SH2144
M12130805 SH1557
M12130807 USG294
M12130807 USG022
M12130808 SH1279
M12130809 SH1378
M12130809 SH0761
M12131123 USG689
M13130601 USG294
M13130605 SH0935
M13130605 SH1276
M13130606 SH0035
M13130608 SH0059
M13130613 SH0935
M13130613 SH0035
M13130615 SH1279
M13130617 USG294
M13130620 USG721
M13130621 SH1368
M13130621 SH1378
M13130622 SH1015
M13130624 SH1368
M13130624 SH0761
M13130626 SH0729
M13130702 USG294
M13130703 SH1279
M13130703 SH2144
M13130708 SH1279
M13130712 SH2144
M13130712 SH1276
M13130715 USG721
M13130715 SH1378
M13130715 SH0729
M13130716 USG022
M13130716 SH1076
M13130722 SH1570
M13130722 USG518
M13130724 SH1279
M13130726 USG294
M13130726 SH0825
M13130726 SH0935
M13130731 USG294
M13130731 SH0729
M13130807 SH1378
M13130808 USG294
M13131123 SH1557
M13131123 SH0729
M14130606 SH1391
M14130606 USG518
M14130607 USG294
M14130610 USG294
M14130610 SH0729
M14130613 SH0035
M14130614 SH1622
M14130617 USG294
M14130617 SH0761
M14130620 USG294
M14130620 USG721
M14130621 SH0825
M14130626 SH1570
M14130626 SH1391
M14130702 SH1279
M14130703 USG294
M14130724 SH1015
M14130724 SH1276
M14130725 SH1622
M14130730 USG294
N10130926 SH0495
N10131217 SH1015
N10131217 SH2144
N10140106 SH1015
N10140109 SH1385
N10140110 SH1378
N10140120 SH0729
N10140120 SH1279
N10140120 SH0761
N10140123 SH1378
N10140130 SH0825
N10140130 SH1378
N10140130 SH0935
N10140204 SH1837
N10140204 USG721
N10140204 SH0729
N10140207 SH1276
N10140210 SH0059
N10140210 SH1557
N10140210 USG294
N10140210 SH0729
N10140210 SH2125
N10140211 SH0935
N10140214 SG704
N10140214 USG294
N10140214 SH0495
N10140217 USG294
N10140221 SH0495
N10140224 SH0729
N10140313 USG294
N10140313 SH1378
N10140313 SH0729
N10140313 SH0495
N10140314 SH1557
N10140325 SH0761
N10140401 SH1390
N10140401 SH2186
N10140403 SH1182
N10140403 USG294
N10140408 SH2144
N10140410 SH1378
N10140410 SH0495
N10140410 SH1015
N10140414 USG294
N10140422 USG294
N10140422 SH0761
N10140423 USG294
N10140424 SH1378
N10140425 SH1390
N10140425 USG022
N10140428 SH1015
N10140428 SH1279
N10140428 SH2144
N10140501 USG294
N10140501 SH1391
N10140501 SH2125
N10140501 USG518
N10140507 SH1557
N10140507 SH0729
N10140508 SH0729
N10140512 SH1378
N10140512 SH0935
N10140512 SH1279
N10140513 USG518
N10140514 USG721
N10140519 SH1182
N10140519 USG022
N10140519 SH0495
N10140520 USG721
N10140520 SH0729
N10140523 SH1557
N10140523 SH2144
N10140527 USG721
N10140527 SH1279
N10140530 USG721
N11140204 SH0059
N11140204 USG294
N11140204 SH0729
N11140204 SH0035
N11140314 SH0547
N11140314 USG294
N11140314 USG022
N11140321 USG022
N11140423 SH1378
N11140423 SH2125
N11140428 SH0935
N11140428 SH0761
N11140501 SG704
N11140501 SH0035
N11140501 SH2125
N11140508 SH0059
N11140508 SH0761
N11140509 SH0729
N11140509 SH2144
N11140513 SH0547
N11140514 SH1378
N11140515 SH1279
N11140515 SH0761
N11140516 SH1622
N11140519 SH0547
N11140519 SH1837
N11140521 USG294
N11140526 USG689
N11140526 SH0825
N11140527 SH0729
N11140527 SH1279
N11140529 SH1378
N11140530 SH1076
N12140321 SH1182
N12140325 SH2144
N12140515 SH1378
N13140325 SH0729
N13140424 USG689
S10130603 SH1279
S11130530 SH0761
S11130603 USG294
S11130603 SH1378
Material Number: 420179-14
Batch / System Numbers:
N10140620 SH1182
N11140730 SH1182
N13140808 SH1182
N11140904 SH1182
N11150109 SH1182
N11150228 SH1182
N12140620 SH1570
N11140923 SH1570
N11141111 SH1570
N11150228 SH1570
N11140630 SG704
N10140704 SG704
N10140721 SG704
N10140729 SG704
N10140811 SG704
N13140808 SG704
N11140814 SG704
N10140829 SG704
N10140917 SG704
N10140926 SG704
N10141217 SG704
N12141217 SG704
N13150302 SG704
N11140703 SH0547
N10140909 SH0547
N13140909 SH0547
N11141213 SH0547
N10150212 SH0547
N10140617 SH0059
N12140704 SH0059
N10140718 SH0059
N11140820 SH0059
N11140916 SH0059
N10141119 SH0059
N10141125 SH0059
N11150218 SH0059
N11140729 SH1997
N12140620 SH1557
N10140701 SH1557
N10140711 SH1557
N11140724 SH1557
N10140812 SH1557
N11140822 SH1557
N10140828 SH1557
N10140909 SH1557
N10141022 SH1557
N10141111 SH1557
N13141124 SH1557
N11141209 SH1557
N11141213 SH1557
N14141223 SH1557
N10150106 SH1557
N10150115 SH1557
N10150128 SH1557
N11150213 SH1557
N11150219 SH1557
N10150303 SH1557
N10150109 SH1964
N11150218 SH1964
N11140623 SH1837
N10141010 SH1837
N12140612 SH1390
N10140701 SH1390
N10140910 SH1390
N10141206 SH1390
N10150210 SH1390
N13150302 SH1390
N12140612 USG689
N11140701 USG689
N11141001 USG689
N11140806 SH1368
N10141116 SH1368
N12141213 SH1368
N11140610 USG294
N10140620 USG294
N10140701 USG294
N11140701 USG294
N10140711 USG294
N10140718 USG294
N11140730 USG294
N11140731 USG294
N11140820 USG294
N10140829 USG294
N13141023 USG294
N14141023 USG294
N11141104 USG294
N10141201 USG294
N11141210 USG294
N12141213 USG294
N12141217 USG294
N12150121 USG294
N11150123 USG294
N12150212 USG294
N10150219 USG294
N11150228 USG294
N13141125 USG294
N12140710 USG721
N10140806 USG721
N10140814 USG721
N11140829 USG721
N11140901 USG721
N10140903 USG721
N10141029 USG721
N11141103 USG721
N10141211 USG721
N14141223 USG721
N11150211 USG721
N10140701 USG022
N10140703 USG022
N10140806 USG022
N10140814 USG022
N11140917 USG022
N11141211 USG022
N10141213 USG022
N10141218 USG022
N11150115 USG022
N11150209 USG022
N11150211 USG022
N11141110 SH2186
N11141206 SH2186
N12141208 SH2186
N10141218 SH2186
N10140619 SH0825
N11140806 SH0825
N11140903 SH0825
N10141022 SH0825
N14141218 SH0825
N10140620 SH1378
N13140623 SH1378
N10140630 SH1378
N12140704 SH1378
N11140724 SH1378
N10140729 SH1378
N10140909 SH1378
N10140905 SH1378
N11141022 SH1378
N13141023 SH1378
N11141104 SH1378
N10141127 SH1378
N12141129 SH1378
N10141202 SH1378
N11141208 SH1378
N13141218 SH1378
N10141222 SH1378
N11150116 SH1378
N11150219 SH1378
N11150225 SH1378
N10140701 SH0935
N12140710 SH0935
N10140813 SH0935
N13140808 SH0935
N11140909 SH0935
N10141028 SH0935
N10141104 SH0935
N11141208 SH0935
N12141210 SH0935
N11150126 SH0935
N11150204 SH0935
N12150212 SH0935
N12140623 SH0729
N12140704 SH0729
N10140721 SH0729
N10140730 SH0729
N11140729 SH0729
N10140815 SH0729
N10140820 SH0729
N10140904 SH0729
N11140904 SH0729
N10140909 SH0729
N12140910 SH0729
N13141022 SH0729
N11141105 SH0729
N10141127 SH0729
N12141202 SH0729
N11150119 SH0729
N10150109 SH0729
N11150123 SH0729
N11150211 SH0729
N10150224 SH0729
N13150225 SH0729
N10140630 SH0035
N11140801 SH0035
N10141001 SH0035
N11141201 SH0035
N10150218 SH0035
N10140617 SH1391
N13140808 SH1391
N11141129 SH1391
N12150216 SH1391
N13150224 SH1391
N10140607 SH2125
N12141125 SH2125
N10140711 SH1622
N12150121 SH1622
N12140910 SH1076
N11140611 SH1604
N12141023 SH1604
N12141211 SH1604
N10140617 SH0495
N10140703 SH0495
N11140707 SH0495
N10140722 SH0495
N11140812 SH0495
N11140904 SH0495
N10140910 SH0495
N10141103 SH0495
N10141124 SH0495
N11141208 SH0495
N11141217 SH0495
N10141222 SH0495
N11150203 SH0495
N10150225 SH0495
N13150302 SH0495
N10140607 SH1015
N11140612 SH1015
N10140711 SH1015
N11140724 SH1015
N11140926 SH1015
N10141001 SH1015
N10141022 SH1015
N10141106 SH1015
N12141116 SH1015
N10150209 SH1015
N12150216 SH1015
N10150216 SH1015
N13150302 SH1015
N10140619 SH1385
N10140729 SH1385
N12140808 SH1385
N11140917 SH1385
N10141202 SH1385
N10150219 SH1385
N13150302 SH1385
N11140612 SH1279
N10140613 SH1279
N10140620 SH1279
N11140630 SH1279
N11140711 SH1279
N10140718 SH1279
N10140820 SH1279
N13140909 SH1279
N11140926 SH1279
N10141010 SH1279
N10141104 SH1279
N10141110 SH1279
N13141116 SH1279
N12141208 SH1279
N12141211 SH1279
N12141217 SH1279
N10150109 SH1279
N11150218 SH1279
N10150228 SH1279
N11150303 SH1279
N11140610 SH2144
N11140611 SH2144
N10140904 SH2144
N10140903 SH2144
N11140909 SH2144
N10141021 SH2144
N12141116 SH2144
N12141124 SH2144
N11141201 SH2144
N12141222 SH2144
N10150203 SH2144
N10150226 SH2144
N12140630 SH0761
N10140722 SH0761
N11140801 SH0761
N11140923 SH0761
N11141103 SH0761
N11141201 SH0761
N12141213 SH0761
N11141217 SH0761
N12150121 SH0761
N11150211 SH0761
N11140630 SH1276
N11140711 SH1276
N10140806 SH1276
N11140822 SH1276
N10141110 USG518
N10141116 USG518
N12141125 USG518
N10150217 USG518
Material Number: 420179-15
Batch / System Numbers:
N11150625 SH1182
N10150810 SH1182
N11150810 SH1182
N10150911 SH1182
N10150916 SH1182
N11150922 SH1182
N10151117 SH1182
N12151202 SH1182
N10160107 SH1182
N10160129 SH1182
N11160309 SH1182
N10160620 SH1182
N12160708 SH1182
N10160715 SH1182
N10161014 SH1182
N11161116 SH1182
N11170106 SH1182
N12170110 SH1182
N10170126 SH1182
N11170217 SH1182
N10170309 SH1182
N11170526 SH1182
N10170626 SH1182
N11170706 SH1182
N11170627 SH1182
N10170830 SH1182
N10171128 SH1182
N10180125 SH1182
N10180222 SH1182
N10180326 SH1182
N10180316 SH1182
N10180424 SH1182
N10150605 SH1570
N10150603 SH1570
N11150825 SH1570
N10151015 SH1570
N10151202 SH1570
N11151216 SH1570
N10160112 SH1570
N10160422 SH1570
N14160725 SH1570
N12160712 SH1570
N10161221 SH1570
N12170309 SH1570
N11170427 SH1570
N10170630 SH1570
N11170609 SH1570
N10180208 SH1570
N10180405 SH1570
N11180220 SH1570
N11150311 SG704
N10150416 SG704
N11150505 SG704
N11150609 SG704
N11150730 SG704
N10150729 SG704
N10150801 SG704
N10151005 SG704
N10151202 SG704
N12151208 SG704
N10151222 SG704
N10160119 SG704
N10160220 SG704
N13160305 SG704
N11160504 SG704
N11160609 SG704
N13160801 SG704
N11160912 SG704
N10161013 SG704
N11161102 SG704
N11170103 SG704
N11170217 SG704
N10170223 SG704
N10170421 SG704
N11170427 SG704
N10170515 SG704
N11170526 SG704
N10170626 SG704
N10170714 SG704
N10170728 SG704
N10171002 SG704
N10171115 SG704
N10171204 SG704
N10171207 SG704
N10180104 SG704
N10180125 SG704
N10180202 SG704
N11180123 SG704
N10180321 SG704
N10180306 SG704
N10180618 SG704
N13150304 SH0547
N12150511 SH0547
N11150519 SH0547
N10150624 SH0547
N10150921 SH0547
N10150923 SH0547
N11151210 SH0547
N13151208 SH0547
N10160210 SH0547
N10160414 SH0547
N10160519 SH0547
N11160614 SH0547
N11160811 SH0547
N11150312 SH0059
N11150401 SH0059
N12150319 SH0059
N10150507 SH0059
N10150519 SH0059
N10150601 SH0059
N10150602 SH0059
N10150623 SH0059
N10150731 SH0059
N10150903 SH0059
N10150929 SH0059
N11151030 SH0059
N10151209 SH0059
N12151202 SH0059
N10151221 SH0059
N11151210 SH0059
N10160718 SH0059
N12160815 SH0059
N13160823 SH0059
N10161020 SH0059
N10161104 SH0059
N10161116 SH0059
N11161117 SH0059
N10170425 SH0059
N10170426 SH0059
N10170509 SH0059
N10170610 SH0059
N10170704 SH0059
N10170717 SH0059
N10170718 SH0059
N10170720 SH0059
N10170926 SH0059
N10171109 SH0059
N10171205 SH0059
N10171215 SH0059
N10171214 SH0059
N10180202 SH0059
N10180509 SH0059
N11180511 SH0059
N11150312 SH1557
N10150330 SH1557
N10150417 SH1557
N11150429 SH1557
N10150526 SH1557
N11150528 SH1557
N10150622 SH1557
N10150803 SH1557
N10150909 SH1557
N11150825 SH1557
N10151009 SH1557
N11151012 SH1557
N12151118 SH1557
N11160116 SH1557
N10160215 SH1557
N10160225 SH1557
N10160408 SH1557
N11160504 SH1557
N10160516 SH1557
N11160606 SH1557
N10160623 SH1557
N11160614 SH1557
N11160715 SH1557
N11160808 SH1557
N13160823 SH1557
N11160930 SH1557
N12160930 SH1557
N10161128 SH1557
N11161202 SH1557
N10161219 SH1557
N10170201 SH1557
N11170120 SH1557
N10170225 SH1557
N12170404 SH1557
N11170421 SH1557
N10170509 SH1557
N11170526 SH1557
N10170606 SH1557
N10170706 SH1557
N10170718 SH1557
N10170810 SH1557
N10170815 SH1557
N10171011 SH1557
N10171023 SH1557
N10171117 SH1557
N10180123 SH1557
N10180202 SH1557
N10180216 SH1557
N10180302 SH1557
N11180413 SH1557
N10180423 SH1557
N10180306 SH1557
N10180418 SH1557
N10150427 SH1964
N11150429 SH1964
N10150622 SH1964
N11150626 SH1964
N11150728 SH1964
N10150814 SH1964
N10151014 SH1964
N11150928 SH1964
N10151102 SH1964
N10160106 SH1964
N10160303 SH1964
N10160321 SH1964
N10160505 SH1964
N10160705 SH1964
N11160809 SH1964
N10161028 SH1964
N11161107 SH1964
N11161125 SH1964
N10170309 SH1964
N10170404 SH1964
N10170519 SH1964
N10170526 SH1964
N11170620 SH1964
N10170707 SH1964
N11170906 SH1964
N10171101 SH1964
N10180119 SH1964
N10180201 SH1964
N10180209 SH1964
N10180122 SH1964
N10180322 SH1964
N10180706 SH1964
N11150622 SH1837
N12150725 SH1837
N10150804 SH1837
N10150714 SH1837
N10160105 SH1837
N12160831 SH1837
N12161014 SH1837
N10161128 SH1837
N10170913 SH1837
N10171020 SH1837
N10180219 SH1837
N11180511 SH1837
N10180301 SH1822
N13150318 SH1390
N10150518 SH1390
N10150624 SH1390
N10150623 SH1390
N10151009 SH1390
N10151028 SH1390
N12151202 SH1390
N11160225 SH1390
N12160305 SH1390
N11160419 SH1390
N11160428 SH1390
N10160628 SH1390
N11160725 SH1390
N11160714 SH1390
N13160830 SH1390
N13160815 SH1390
N11161010 SH1390
N11161028 SH1390
N10161124 SH1390
N11161223 SH1390
N11170421 SH1390
N10170505 SH1390
N10170612 SH1390
N10170707 SH1390
N10171027 SH1390
N10171101 SH1390
N10180112 SH1390
N10180215 SH1390
N10180302 SH1390
N10180306 SH1390
N10150325 USG689
N10150630 USG689
N12150525 USG689
N10150729 USG689
N10150814 USG689
N10150904 USG689
N10160415 USG689
N10160512 USG689
N11160901 USG689
N10161003 USG689
N10170202 USG689
N10170209 USG689
N12170304 USG689
N11171221 USG689
N10171214 USG689
N10180326 USG689
N10180503 USG689
N10150422 SH1368
N11150708 SH1368
N10150730 SH1368
N11150922 SH1368
N10160106 SH1368
N11160213 SH1368
N10160317 SH1368
N11160418 SH1368
N10170116 SH1368
N11170517 SH1368
N11171220 SH1368
N10171214 SH1368
N10180413 SH1368
N10180504 SH1368
N11150306 USG294
N10150324 USG294
N10150326 USG294
N11150407 USG294
N13150318 USG294
N11150416 USG294
N10150424 USG294
N11150505 USG294
N11150519 USG294
N10150528 USG294
N11150529 USG294
N10150616 USG294
N12150609 USG294
N10150708 USG294
N10150727 USG294
N10150804 USG294
N10150810 USG294
N10150824 USG294
N10150902 USG294
N10150909 USG294
N10150916 USG294
N11150824 USG294
N10151005 USG294
N11151013 USG294
N12150820 USG294
N11151014 USG294
N11151030 USG294
N10151203 USG294
N12151006 USG294
N11151208 USG294
N10151216 USG294
N10160104 USG294
N12160116 USG294
N11160128 USG294
N10160210 USG294
N10160213 USG294
N10160301 USG294
N10160307 USG294
N11160312 USG294
N10160321 USG294
N11160330 USG294
N10160422 USG294
N10160505 USG294
N10160512 USG294
N10160517 USG294
N10160613 USG294
N10160623 USG294
N10160630 USG294
N10160714 USG294
N13160803 USG294
N11160809 USG294
N11160815 USG294
N11160818 USG294
N11160906 USG294
N14160815 USG294
N10160930 USG294
N14160913 USG294
N10161010 USG294
N10161021 USG294
N10161014 USG294
N11161124 USG294
N10160915 USG294
N11170103 USG294
N10170117 USG294
N11170126 USG294
N11170120 USG294
N11170309 USG294
N11170327 USG294
N10170330 USG294
N11170427 USG294
N10170428 USG294
N10170427 USG294
N10170511 USG294
N10170607 USG294
N11170606 USG294
N10170610 USG294
N10170620 USG294
N12170610 USG294
N10170707 USG294
N10170720 USG294
N10170803 USG294
N10170810 USG294
N10170901 USG294
N10170907 USG294
N10170929 USG294
N10171030 USG294
N10171128 USG294
N10171207 USG294
N10171219 USG294
N10180119 USG294
N10180130 USG294
N11180123 USG294
N10180209 USG294
N10180314 USG294
N10180326 USG294
N10180413 USG294
N10180424 USG294
N10180316 USG294
N11180423 USG294
N10180306 USG294
N10180510 USG294
N10150304 USG721
N11150312 USG721
N10150325 USG721
N10150602 USG721
N10150608 USG721
N11150622 USG721
N11150625 USG721
N12150725 USG721
N11150814 USG721
N10151019 USG721
N10150725 USG721
N11151110 USG721
N11151205 USG721
N12151202 USG721
N10151210 USG721
N11160513 USG721
N10160601 USG721
N10160606 USG721
N10160609 USG721
N10161020 USG721
N10161026 USG721
N11161007 USG721
N11161020 USG721
N10161128 USG721
N11160921 USG721
N11161214 USG721
N10170105 USG721
N11170213 USG721
N10170228 USG721
N10170224 USG721
N10170426 USG721
N10170512 USG721
N10170519 USG721
N10170620 USG721
N11170627 USG721
N10170720 USG721
N10170726 USG721
N10170810 USG721
N10170817 USG721
N10170815 USG721
N10171025 USG721
N10171026 USG721
N10171124 USG721
N10171206 USG721
N10180104 USG721
N10180131 USG721
N10180207 USG721
N10180301 USG721
N10180216 USG721
N11180103 USG721
N10180220 USG721
N10150428 USG022
N13150427 USG022
N10150603 USG022
N10150622 USG022
N10150623 USG022
N12150806 USG022
N10150923 USG022
N10151118 USG022
N13151204 USG022
N10160206 USG022
N11160310 USG022
N10160316 USG022
N10160419 USG022
N10160531 USG022
N11160615 USG022
N11160725 USG022
N10160830 USG022
N11160927 USG022
N10161028 USG022
N11161107 USG022
N10161122 USG022
N11170120 USG022
N11170403 USG022
N10170504 USG022
N10170515 USG022
N11160811 USG022
N10170626 USG022
N11170706 USG022
N10170720 USG022
N11170814 USG022
N10171017 USG022
N10171101 USG022
N11171221 USG022
N10171214 USG022
N10180504 USG022
N10180322 USG022
N10150605 SH2186
N10150624 SH2186
N10151106 SH2186
N12151202 SH2186
N11160513 SH2186
N10160705 SH2186
N12170225 SH2186
N10170404 SH2186
N10170530 SH2186
N10171023 SH2186
N10171128 SH2186
N10180426 SH2186
N12150511 SH0825
N11150820 SH0825
N11151205 SH0825
N11160120 SH0825
N10160303 SH0825
N14160725 SH0825
N10160809 SH0825
N14160815 SH0825
N11160923 SH0825
N10161031 SH0825
N10170311 SH0825
N10170526 SH0825
N10170630 SH0825
N10171215 SH0825
N10180129 SH0825
N10180328 SH0825
N10150312 SH1378
N10150318 SH1378
N10150506 SH1378
N11150525 SH1378
N10150529 SH1378
N11150529 SH1378
N10150624 SH1378
N10150623 SH1378
N10150908 SH1378
N10150915 SH1378
N10150929 SH1378
N10151016 SH1378
N10150714 SH1378
N12151006 SH1378
N12151202 SH1378
N10160212 SH1378
N10160301 SH1378
N10160308 SH1378
N12160305 SH1378
N10160328 SH1378
N10160421 SH1378
N10160512 SH1378
N11160504 SH1378
N10160518 SH1378
N11160628 SH1378
N11160714 SH1378
N10160718 SH1378
N11160823 SH1378
N13160815 SH1378
N10161013 SH1378
N10161104 SH1378
N10161215 SH1378
N11170103 SH1378
N10170208 SH1378
N10170209 SH1378
N10170213 SH1378
N11170223 SH1378
N10170316 SH1378
N11170403 SH1378
N10170421 SH1378
N11170503 SH1378
N10170606 SH1378
N10170610 SH1378
N10170713 SH1378
N10170802 SH1378
N10170814 SH1378
N10170831 SH1378
N10170926 SH1378
N10171020 SH1378
N10171027 SH1378
N10171114 SH1378
N10171129 SH1378
N10180102 SH1378
N10180113 SH1378
N10180206 SH1378
N10180213 SH1378
N10180305 SH1378
N11180125 SH1378
N11180103 SH1378
N10180328 SH1378
N10180306 SH1378
N10180418 SH1378
N10180509 SH1378
N11180511 SH1378
N10180702 SH1378
N10150306 SH0935
N10150408 SH0935
N12150511 SH0935
N11150605 SH0935
N10150622 SH0935
N12150609 SH0935
N10150923 SH0935
N11150922 SH0935
N10151014 SH0935
N10151026 SH0935
N12151006 SH0935
N10151214 SH0935
N10151216 SH0935
N10151222 SH0935
N11160310 SH0935
N12160309 SH0935
N11160606 SH0935
N10160617 SH0935
N11160714 SH0935
N10160718 SH0935
N10161003 SH0935
N11161004 SH0935
N10161020 SH0935
N11161026 SH0935
N11161124 SH0935
N10170110 SH0935
N12170208 SH0935
N10170301 SH0935
N11170410 SH0935
N10170510 SH0935
N10170606 SH0935
N10170615 SH0935
N10170623 SH0935
N10170707 SH0935
N10170901 SH0935
N11170906 SH0935
N10171026 SH0935
N10171113 SH0935
N10171208 SH0935
N10180111 SH0935
N10180130 SH0935
N10180216 SH0935
N10180326 SH0935
N10180405 SH0935
N11180413 SH0935
N10180306 SH0935
N10150326 SH0729
N11150326 SH0729
N12150311 SH0729
N10150424 SH0729
N10150514 SH0729
N10150611 SH0729
N10150619 SH0729
N10150622 SH0729
N12150722 SH0729
N10150730 SH0729
N10150731 SH0729
N10150824 SH0729
N11150911 SH0729
N11150824 SH0729
N10150929 SH0729
N10151009 SH0729
N11150924 SH0729
N10151014 SH0729
N10150725 SH0729
N11151020 SH0729
N10151026 SH0729
N10151209 SH0729
N11151205 SH0729
N10160107 SH0729
N10160128 SH0729
N13160128 SH0729
N10160215 SH0729
N10160301 SH0729
N11160310 SH0729
N11160317 SH0729
N10160322 SH0729
N10160426 SH0729
N11160506 SH0729
N10160607 SH0729
N10160615 SH0729
N10160620 SH0729
N11160624 SH0729
N12160815 SH0729
N11160811 SH0729
N13160823 SH0729
N10161005 SH0729
N12161014 SH0729
N10161122 SH0729
N10170111 SH0729
N11170324 SH0729
N12170329 SH0729
N11170417 SH0729
N10170505 SH0729
N11170512 SH0729
N10170707 SH0729
N10170712 SH0729
N10170803 SH0729
N10170926 SH0729
N10171023 SH0729
N10171207 SH0729
N10171206 SH0729
N10171228 SH0729
N10180102 SH0729
N10180111 SH0729
N10180122 SH0729
N10180618 SH0729
N10150330 SH0035
N13150427 SH0035
N11150722 SH0035
N11150904 SH0035
N12151204 SH0035
N12151208 SH0035
N10160412 SH0035
N11160513 SH0035
N10160803 SH0035
N10161010 SH0035
N10161013 SH0035
N10170203 SH0035
N11170405 SH0035
N10170610 SH0035
N10170803 SH0035
N10171109 SH0035
N10180111 SH0035
N10180301 SH0035
N10150901 SH1391
N13150304 SH2125
N11150818 SH2125
N10160706 SH2125
N11160815 SH2125
N11161012 SH2125
N11161208 SH2125
N10170620 SH2125
N10170901 SH2125
N10171109 SH2125
N10180103 SH2125
N10180221 SH2125
N10180122 SH2125
N10180418 SH2125
N10150417 SH1622
N10150804 SH1622
N11160105 SH1622
N10160226 SH1622
N11160923 SH1622
N10170316 SH1622
N10171002 SH1622
N10171027 SH1622
N10180129 SH1622
N10160322 SH1076
N10160803 SH1076
N11161013 SH1076
N10170106 SH1076
N11170607 SH1076
N10171025 SH1076
N10171109 SH1076
N10171204 SH1076
N10180207 SH1076
N11180220 SH1076
N12151204 SH1604
N11160614 SH1604
N10161004 SH1604
N10170202 SH1604
N10170322 SH1604
N10170408 SH1604
N12170606 SH1604
N10171030 SH1604
N10171219 SH1604
N10180221 SH1604
N10180305 SH1604
N10180424 SH1604
N10180511 SH1604
N10180418 SH1604
N11150401 SH0495
N10150413 SH0495
N10150506 SH0495
N11150506 SH0495
N10150611 SH0495
N11150609 SH0495
N10150623 SH0495
N11150706 SH0495
N10150804 SH0495
N10150813 SH0495
N10150929 SH0495
N10151118 SH0495
N11151208 SH0495
N11160823 SH0495
N10161114 SH0495
N10170209 SH0495
N11170317 SH0495
N11170526 SH0495
N11170606 SH0495
N11170706 SH0495
N10170714 SH0495
N10180413 SH0495
N10150304 SH1015
N12150311 SH1015
N11150416 SH1015
N10150506 SH1015
N11150509 SH1015
N10150528 SH1015
N11150814 SH1015
N10150921 SH1015
N10151012 SH1015
N10151109 SH1015
N10151110 SH1015
N11151103 SH1015
N10151106 SH1015
N10160106 SH1015
N10160118 SH1015
N12160128 SH1015
N10160224 SH1015
N10160307 SH1015
N12160305 SH1015
N10160420 SH1015
N10160504 SH1015
N10160630 SH1015
N11160708 SH1015
N12160712 SH1015
N10160801 SH1015
N10160802 SH1015
N10160816 SH1015
N11160815 SH1015
N11160831 SH1015
N10161017 SH1015
N10161110 SH1015
N10161104 SH1015
N10161107 SH1015
N10161116 SH1015
N10161219 SH1015
N11170126 SH1015
N11170309 SH1015
N10170322 SH1015
N11170425 SH1015
N11170427 SH1015
N11170517 SH1015
N11170526 SH1015
N10170607 SH1015
N10170724 SH1015
N10170816 SH1015
N10170824 SH1015
N10171102 SH1015
N10171128 SH1015
N10180424 SH1015
N10180227 SH1015
N10180316 SH1015
N10180328 SH1015
N10150521 SH1385
N11150609 SH1385
N11151013 SH1385
N11151207 SH1385
N11151210 SH1385
N11160419 SH1385
N10160706 SH1385
N11160906 SH1385
N14160913 SH1385
N11161007 SH1385
N11161124 SH1385
N11170111 SH1385
N11170606 SH1385
N10170704 SH1385
N10170710 SH1385
N11170609 SH1385
N10171206 SH1385
N11171220 SH1385
N10180206 SH1385
N10180413 SH1385
N10180405 SH1385
N10180412 SH1385
N11180216 SH1385
N10150311 SH1279
N10150313 SH1279
N11150429 SH1279
N11150507 SH1279
N10150519 SH1279
N10150528 SH1279
N11150529 SH1279
N11150626 SH1279
N11150722 SH1279
N10150722 SH1279
N10150805 SH1279
N10150901 SH1279
N10150902 SH1279
N11150820 SH1279
N10150915 SH1279
N10151015 SH1279
N10151016 SH1279
N10151117 SH1279
N12151118 SH1279
N10160106 SH1279
N12160116 SH1279
N12160130 SH1279
N13160219 SH1279
N10160308 SH1279
N10160412 SH1279
N10160420 SH1279
N11160418 SH1279
N10160516 SH1279
N10160523 SH1279
N10160627 SH1279
N11160712 SH1279
N11160802 SH1279
N10160901 SH1279
N12160831 SH1279
N11161003 SH1279
N11161007 SH1279
N11160719 SH1279
N11161130 SH1279
N10170103 SH1279
N10170201 SH1279
N10170324 SH1279
N11170330 SH1279
N10170425 SH1279
N10170526 SH1279
N10170608 SH1279
N11170607 SH1279
N11170620 SH1279
N10170717 SH1279
N10170724 SH1279
N10170815 SH1279
N10170929 SH1279
N10171030 SH1279
N10171102 SH1279
N10180103 SH1279
N10180119 SH1279
N10180131 SH1279
N10180202 SH1279
N10180215 SH1279
N10180326 SH1279
N11180413 SH1279
N10180328 SH1279
N10180220 SH1279
N11150306 SH2144
N10150407 SH2144
N10150413 SH2144
N10150521 SH2144
N10150602 SH2144
N11150708 SH2144
N11150722 SH2144
N10150911 SH2144
N11150724 SH2144
N11151012 SH2144
N10151202 SH2144
N10151216 SH2144
N13160128 SH2144
N10160206 SH2144
N10160222 SH2144
N10160303 SH2144
N10160322 SH2144
N10160504 SH2144
N11160509 SH2144
N10160714 SH2144
N10160718 SH2144
N13160912 SH2144
N11160928 SH2144
N12160930 SH2144
N10161107 SH2144
N11161125 SH2144
N10170103 SH2144
N10170202 SH2144
N11170223 SH2144
N11170403 SH2144
N10170421 SH2144
N10170505 SH2144
N10170515 SH2144
N10170626 SH2144
N10170824 SH2144
N10170926 SH2144
N10171115 SH2144
N10171221 SH2144
N10180119 SH2144
N10180221 SH2144
N11180125 SH2144
N11180216 SH2144
N10180220 SH2144
N10180706 SH2144
N13150304 SH0761
N10150413 SH0761
N11150509 SH0761
N11150521 SH0761
N10150617 SH0761
N10150619 SH0761
N11150622 SH0761
N11150601 SH0761
N12150730 SH0761
N11150814 SH0761
N10150917 SH0761
N10151012 SH0761
N10151013 SH0761
N11151012 SH0761
N10151104 SH0761
N11151027 SH0761
N12151117 SH0761
N10151221 SH0761
N10160225 SH0761
N10160307 SH0761
N11160321 SH0761
N11160330 SH0761
N11160418 SH0761
N10160609 SH0761
N11160620 SH0761
N10160705 SH0761
N11160823 SH0761
N12160831 SH0761
N13160912 SH0761
N11160922 SH0761
N10161027 SH0761
N10161110 SH0761
N11161110 SH0761
N11160719 SH0761
N10170202 SH0761
N11170126 SH0761
N10170213 SH0761
N10170210 SH0761
N11170304 SH0761
N11170309 SH0761
N10170324 SH0761
N10170405 SH0761
N10170424 SH0761
N10170504 SH0761
N10170606 SH0761
N10170706 SH0761
N10170704 SH0761
N10170712 SH0761
N10170913 SH0761
N10170929 SH0761
N10171025 SH0761
N10171113 SH0761
N10171205 SH0761
N11171221 SH0761
N10180112 SH0761
N10180130 SH0761
N10180206 SH0761
N10180226 SH0761
N10180223 SH0761
N10180412 SH0761
N10180511 SH0761
N10180418 SH0761
N10180509 SH0761
N10180702 SH0761
N10150304 SH1276
N11150312 SH1276
N10151215 SH1276
N10150318 USG518
N13150427 USG518
N11160105 USG518
N11160312 USG518
N10170105 USG518
N11170103 RSH0152
N11170424 RSH0152
N11170706 RSH0152
N10171214 RSH0152
N10180122 RSH0152
Material Numbers: 420179-18
Batch / System Numbers:
N12180517 SH1182
N11180518 SH1182
N11180518 SH1570
N10180516 SH0059
N10180521 SH0059
N10180409 SH0059
N10180411 SH1557
N11180605 SH1557
N10180613 SH1557
N10180615 SH1557
N10180409 SH1557
N10171213 SH1557
N10180604 SH1964
N10180615 SH1964
N10180409 SH1964
N10180416 SH1964
N11180518 SH1822
N11180605 SH1390
N12180517 USG689
N10180507 USG689
N11180524 USG294
N11180228 USG721
N11180517 USG721
N12180517 USG721
N10180507 USG721
N11180323 USG022
N11180517 USG022
N10180409 USG022
N10180529 SH2186
N11180110 SH0825
N11180525 SH0825
N11180228 SH1378
N10180430 SH1378
N11180517 SH1378
N11180524 SH1378
N10180416 SH0935
N10180523 SH0729
N10180521 SH0729
N10180525 SH0729
N10180521 SH2125
N10180619 SH2125
N11180518 SH1622
N10180516 SH1076
N10180615 SH1076
N10180605 SH0495
N10180430 SH1015
N10180409 SH1015
N10180416 SH1015
N10180507 SH1015
N11180613 SH1385
N10171123 SH1385
N12180517 SH1279
N10180521 SH1279
N10180613 SH1279
N11180524 SH1279
N11180601 SH2144
N11180613 SH2144
N11180517 SH0761
N10180619 SH0761
Material Number: 420179-22
Batch / System Numbers:
N10180926 SH1182
N10181205 SH1182
N11181102 SH1182
N10190110 SH1182
N10190131 SH1182
N10190617 SH1182
N10190415 SH1182
N14190701 SH1182
N11190812 SH1182
N14190729 SH1182
N12200302 SH1182
N10200803 SH1182
N11200810 SH1182
N11181009 SG704
N11181019 SG704
N10181024 SG704
N10181129 SG704
N10181203 SG704
N11180927 SG704
N10190603 SH0059
N11200810 SH1997
N10180920 SH1557
N10181005 SH1557
N10181015 SH1557
N10181101 SH1557
N11181210 SH1557
N11181023 SH1557
N11181030 SH1557
N11181122 SH1557
N10190102 SH1557
N10190109 SH1557
N10190124 SH1557
N10190205 SH1557
N10190315 SH1557
N10190415 SH1557
N11190708 SH1557
N12190715 SH1557
N11190722 SH1557
N12190729 SH1557
N10190729 SH1557
N10190805 SH1557
N10190812 SH1557
N11191028 SH1557
N10191209 SH1557
N11200127 SH1557
N10200221 SH1557
N10200317 SH1557
N10200406 SH1557
N10180926 SH1557
N10200218 SH1557
N10200803 SH1557
N11200817 SH1557
N10200525 SH1557
N11180921 SH1964
N11180924 SH1964
N11181010 SH1964
N10181113 SH1964
N10181102 SH1964
N11181029 SH1964
N11190114 SH1964
N10190201 SH1964
N10190306 SH1964
N12190603 SH1964
N11190701 SH1964
N10190708 SH1964
N11190708 SH1964
N10190722 SH1964
N12190722 SH1964
N10190808 SH1964
N10190902 SH1964
N13190805 SH1964
N13200127 SH1964
N10200221 SH1964
N12200127 SH1964
N10200406 SH1964
N11200810 SH1964
N11200309 SH1964
N10190206 SH1837
N11190715 SH1837
N10200218 SH1837
N10200622 SH1837
N10200525 SH1837
N10180920 SH1822
N10190121 SH1822
N11190715 SH1822
N11191028 SH1822
N11200810 SH1822
N10180918 SH1390
N11180919 SH1390
N10181217 SH1390
N10181102 SH1390
N10180927 SH1390
N11181030 SH1390
N14190701 SH1390
N10190722 SH1390
N13190805 SH1390
N11200323 SH1390
N10200309 SH1390
N11200309 SH1390
N11200817 SH1390
N10181122 USG689
N10181023 USG689
N11181030 USG689
N10190201 USG689
N11190812 USG689
N10181204 SH1368
N11191104 SH1368
N12200302 SH1368
N10180921 USG294
N11181024 USG294
N10181211 USG294
N10190610 USG294
N11190722 USG294
N10200323 USG294
N10200803 USG294
N11180920 USG721
N11180926 USG721
N10181003 USG721
N10181024 USG721
N11181211 USG721
N10190109 USG721
N10190111 USG721
N11190114 USG721
N11190327 USG721
N12190603 USG721
N10190415 USG721
N11190701 USG721
N12190708 USG721
N14190729 USG721
N10191007 USG721
N11191104 USG721
N11200127 USG721
N10200317 USG721
N11200323 USG721
N11200309 USG721
N10180919 USG022
N10181212 USG022
N10181217 USG022
N11181102 USG022
N10190104 USG022
N10190109 USG022
N10190603 USG022
N10190701 USG022
N11190617 USG022
N10190805 USG022
N10200217 USG022
N10200317 USG022
N14200127 USG022
N11200309 USG022
N10180926 SH2186
N10181017 SH2186
N10180925 SH2186
N10190225 SH2186
N10190315 SH2186
N11190327 SH2186
N10190722 SH2186
N11190805 SH2186
N11200817 SH2186
N10190201 SH0825
N11190715 SH0825
N12181023 SH1378
N10180928 SH1378
N10181102 SH1378
N11181023 SH1378
N10181207 SH1378
N11181022 SH1378
N11181101 SH1378
N10190107 SH1378
N10190111 SH1378
N10190225 SH1378
N10190306 SH1378
N11190327 SH1378
N10190617 SH1378
N10190708 SH1378
N11190709 SH1378
N12190722 SH1378
N13190805 SH1378
N11191104 SH1378
N11180918 SH0935
N11180921 SH0935
N11180925 SH0935
N10181122 SH0935
N10181114 SH0935
N10190104 SH0935
N10190124 SH0935
N10190131 SH0935
N10190214 SH2180
N12190603 SH2180
N10200221 SH2180
N12200127 SH2180
N10181005 SH0729
N10181203 SH0729
N11181101 SH0729
N10190104 SH0729
N10190315 SH0729
N12190708 SH0729
N11190715 SH0729
N11181101 SH1391
N10181211 SH1391
N11180921 SH2125
N10181218 SH2125
N10190124 SH2125
N10190312 SH2125
N10190603 SH2125
N12190708 SH2125
N10181101 SH1622
N10190114 SH1622
N10180919 SH1076
N12181023 SH1076
N11181210 SH1076
N11181029 SH1076
N10190124 SH1076
N10190701 SH1076
N12190722 SH1076
N11200810 SH1076
N11200309 SH1076
N11181008 SH1604
N10181205 SH1604
N13190708 SH1604
N11191028 SH1604
N10191209 SH1604
N12200302 SH1604
N10200317 SH1604
N10200406 SH1604
N10200601 SH1604
N12181011 SH0495
N11181010 SH0495
N10181204 SH0495
N10181212 SH0495
N10190114 SH0495
N10190314 SH0495
N11190327 SH0495
N13190708 SH0495
N11190805 SH0495
N12191014 SH0495
N10200803 SH0495
N10181113 SH1015
N11181114 SH1015
N12181023 SH1015
N10181012 SH1015
N11181101 SH1015
N10190124 SH1015
N11190124 SH1015
N10190306 SH1015
N10190315 SH1015
N10190408 SH1015
N11190722 SH1015
N11190812 SH1015
N14200127 SH1015
N10200221 SH1015
N11200323 SH1015
N10200406 SH1015
N11180925 SH1385
N11181009 SH1385
N10181015 SH1385
N10181113 SH1385
N11181114 SH1385
N10181217 SH1385
N11181011 SH1385
N10181221 SH1385
N10181218 SH1385
N10190115 SH1385
N10190205 SH1385
N11190617 SH1385
N14190701 SH1385
N11190708 SH1385
N12190708 SH1385
N14190708 SH1385
N10190729 SH1385
N13190805 SH1385
N10191007 SH1385
N10191021 SH1385
N11191104 SH1385
N11180919 SH1279
N10180921 SH1279
N10181015 SH1279
N10181203 SH1279
N10181121 SH1279
N10181012 SH1279
N10181123 SH1279
N10190102 SH1279
N11181029 SH1279
N10190312 SH1279
N10190610 SH1279
N11190701 SH1279
N14190708 SH1279
N11190709 SH1279
N10190730 SH1279
N10191021 SH1279
N10200221 SH1279
N11200217 SH1279
N12200302 SH1279
N10200803 SH1279
N10200810 SH1279
N11180920 SH2144
N10180926 SH2144
N11181024 SH2144
N11180927 SH2144
N10181102 SH2144
N11181030 SH2144
N10190115 SH2144
N10190227 SH2144
N12190603 SH2144
N12190701 SH2144
N11190715 SH2144
N10200323 SH2144
N10180919 SH0761
N11181010 SH0761
N10181017 SH0761
N10181101 SH0761
N10181114 SH0761
N10181102 SH0761
N10180927 SH0761
N10181023 SH0761
N11181031 SH0761
N11181122 SH0761
N10190116 SH0761
N10190513 SH0761
N14190708 SH0761
N10190709 SH0761
N10190805 SH0761
N11191104 SH0761
N10191209 SH0761
N11200217 SH0761
N10200317 SH0761
N10200218 SH0761
N10200601 SH0761
N10200810 SH0761
N10200622 SH0761
N10181217 USG518
N11180920 RSH0152
N11191104 RSH0152
Material Number: 420179-23
Batch / System Numbers:
N13190522 SH1182
N10191209 SH1182
N10200831 SH1182
N11201026 SH1182
N10201130 SH1182
N11201207 SH1182
N11210316 SH1182
N10210405 SH1182
N11191007 SH1997
N11191014 SH1997
N11191209 SH1997
N11201026 SH1997
N11201130 SH1997
N10210830 SH1997
N10210913 SH1997
K11220828 SH1997
K11231123 SH1997
K10240104 SH1997
K10240125 SH1997
N11191007 SH1557
N10191007 SH1557
N11191111 SH1557
N10200107 SH1557
N13200120 SH1557
N10200831 SH1557
N11201026 SH1557
N10201117 SH1557
N12210202 SH1557
N10210222 SH1557
N12210316 SH1557
N10210405 SH1557
N11210419 SH1557
N10210503 SH1557
N10210614 SH1557
N10210726 SH1557
N10210628 SH1557
N11210927 SH1557
N10211011 SH1557
K10220104 SH1557
N10191014 SH1964
N11200107 SH1964
N10200420 SH1964
N12200420 SH1964
N11200727 SH1964
N11200921 SH1964
N10201123 SH1964
N10201214 SH1964
N12201214 SH1964
N12210202 SH1964
N12210316 SH1964
N11210316 SH1964
N10210405 SH1964
N11210419 SH1964
N10210510 SH1964
N11210503 SH1964
N10210503 SH1964
N10210614 SH1964
N10210726 SH1964
N10210628 SH1964
N10210830 SH1964
N10210913 SH1964
N10211011 SH1964
N11210927 SH1964
K10220208 SH1964
K11220425 SH1964
K10220605 SH1964
K10220807 SH1964
K10220918 SH1964
N10191209 SH1837
N10201123 SH1837
K10220403 SH1837
K10240208 SH1837
N10200114 SH1822
N10201214 SH1822
N10210215 SH1822
N10210222 SH1822
N11190610 SH1390
N10190827 SH1390
N10200420 SH1390
N11200727 SH1390
N11201207 SH1390
N12210316 SH1390
N11210503 SH1390
N10210614 SH1390
N10210726 SH1390
N11210927 SH1390
N10210920 SH1390
K10220131 SH1390
N10211206 SH1390
N11210913 SH1390
K11220522 SH1390
K10220605 SH1390
K10220807 SH1390
K11220828 SH1390
K10230209 SH1390
N10190909 USG689
N10191014 USG689
N13200210 USG689
K10220605 USG689
K10220807 USG689
N11191028 SH1368
N10200824 SH1368
K10240104 SH1368
K10240125 SH1368
N11191007 USG294
N11200120 USG294
N11200210 USG294
N10200420 USG294
N11200921 USG294
N10201026 USG294
N10201123 USG294
N11201207 USG294
N10210112 USG294
N10210118 USG294
N10210222 USG294
N10210405 USG294
N10210412 USG294
N11210503 USG294
N10210802 USG294
N10210830 USG294
K10220131 USG294
K10220208 USG294
K10220403 USG294
K11220425 USG294
K11220522 USG294
K10220605 USG294
K10220918 USG294
K11220828 USG294
K10230209 USG294
K10240104 USG294
K10240125 USG294
N10190826 USG721
N10191007 USG721
N10190827 USG721
N10190917 USG721
N11200107 USG721
N10200120 USG721
N11200120 USG721
N10200703 USG721
N10200330 USG721
N10200831 USG721
N12200831 USG721
N10201130 USG721
N11210621 USG721
N10191014 USG022
N10200120 USG022
N10200420 USG022
N12200420 USG022
N11200921 USG022
N11201207 USG022
N10210111 USG022
N10210118 USG022
N10210215 USG022
N10210316 USG022
N10210412 USG022
N11210621 USG022
N10210405 USG022
N10210726 USG022
N11210823 USG022
N11210927 USG022
N11210913 USG022
N10210712 USG022
K10220104 USG022
K10220208 USG022
K11220425 USG022
K11220522 USG022
K10220605 USG022
K10220807 USG022
K10220918 USG022
N11210316 USG022
K11220828 USG022
K10230209 USG022
N10190826 SH2186
N11191209 SH2186
N12200420 SH2186
N10200914 SH2186
N10201026 SH2186
N10200420 SH2186
N12201214 SH2186
N10210118 SH2186
N11210419 SH2186
N10210503 SH2186
N12210524 SH2186
N10210726 SH2186
N10210913 SH2186
N10191209 SH0825
N11201026 SH0825
N10200810 SH0825
N10210726 SH0825
N10191104 SH1378
N11200113 SH1378
N11191209 SH1378
N13200120 SH1378
N10200703 SH1378
N10200330 SH1378
N11201214 SH1378
N10210112 SH1378
N10210118 SH1378
N12210202 SH1378
N12200831 SH1378
N13210222 SH1378
N11210316 SH1378
N10210405 SH1378
N11210412 SH1378
N10210503 SH1378
N10210830 SH1378
N10200420 SH0729
N11200727 SH0729
N11200921 SH0729
N11201214 SH0729
N11210215 SH0729
N13200120 SH0035
N10191209 SH1391
N10201019 SH1391
N11200921 SH1391
N10190827 SH2125
N13190522 SH2125
N10190909 SH1622
N11191014 SH1076
N10191209 SH1076
N10210112 SH1076
N10190909 SH1076
N10200914 SH1076
N11210412 SH1076
K10220918 SH1076
N10190909 SH1604
N10200703 SH1604
N10200907 SH1604
N11200120 SH0495
N10200810 SH0495
N11191104 SH1015
N11200107 SH1015
N13200120 SH1015
N10200120 SH1015
N13200210 SH1015
N11200713 SH1015
N10200831 SH1015
N11201026 SH1015
K10220124 SH1015
N10190909 SH1385
N14190522 SH1385
N10190827 SH1385
N11200107 SH1385
N10200204 SH1385
N10190827 SH1279
N10200114 SH1279
N11200120 SH1279
N11200210 SH1279
N12200420 SH1279
N10200831 SH1279
N10200420 SH1279
N10210111 SH1279
N11210111 SH1279
N10210614 SH1279
N10210726 SH1279
N10210802 SH1279
N10210913 SH1279
N10211011 SH1279
N10211206 SH1279
K10220104 SH1279
K10220208 SH1279
K10220403 SH1279
K11220425 SH1279
K11220522 SH1279
K10220605 SH1279
K10220807 SH1279
N10200107 SH2144
N11200107 SH2144
N10200713 SH2144
N10210111 SH2144
N10210118 SH2144
N10210412 SH2144
N10210628 SH2144
N10210802 SH2144
N10210830 SH2144
K11220425 SH2144
K10230209 SH2144
K11231123 SH2144
N11190819 SH0761
N11191028 SH0761
N11200107 SH0761
N12200120 SH0761
N10200114 SH0761
N11200210 SH0761
N11201214 SH0761
N12210202 SH0761
N10201130 SH0761
N12200831 SH0761
N10210405 SH0761
N10210503 SH0761
N10210614 SH0761
N10210726 SH0761
N10210830 SH0761
N10210628 SH0761
N10210913 SH0761
N11210913 SH0761
N10211206 SH0761
K10220208 SH0761
K11220425 SH0761
N10210222 SH0761
K11220522 SH0761
K10220605 SH0761
K10220807 SH0761
K11220828 SH0761
K10230209 SH0761
N10211011 SH0761
N12200420 RSH0152
N10210510 RSH0152
K10230209 RSH0152
K10220208 SH1513
K10220605 SH1513
K10220807 SH1513
K10220918 SH1513
K10230209 SH1513
K10240125 SH1513
|
| FEI Number |
3001675293
|
Recalling Firm/ Manufacturer |
Intuitive Surgical, Inc.
1266 Kifer Rd
Sunnyvale
CA
94086-5304
|
| For Additional Information Contact | Shahbaz Khan 1-408-5232443 |
Manufacturer Reason for Recall | Due to increased complaints for broken/frayed grip cables for reusable instruments. |
FDA Determined Cause 2 | Process control |
| Action | On 12/29/2025, the firm sent via email an "Urgent: Medical Device Recall" Letter informing customers that intuitive has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci S and Si reusable instruments.
Customers are instruct to:
1. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form.
2. Identify and quarantine any affected product(s). Affected product(s) listed above may be returned by sending an email with quantities and lot numbers to your regional customer service.
3. Ensure to include the FSCA number ISIFA2025-15-R in the return notes.
4. Credit will be provided based on the number of remaining lives.
5. Shared or further distributed these products with other sites, make sure appropriate staff at the site receive and understand this notification so they locate and return their affected product.
For questions or concern, contact Clinical Sales Representative or contact Intuitive Customer Service at the numbers listed below:
" North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com.
" Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com
" South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com
" Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com
" India: +1-800-103-6952 (9 am to 6 pm IT)
" Taiwan: +0800-86-8181 (9 am to 6 pm CT) |
| Quantity in Commerce | 86904 |
| Distribution | Worldwide - U.S. Nationwide distribution in the states of CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MO, MS, NC, NJ, NY, OH, PA, TX, VA and WI. The countries of Argentina, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Ecuador, France, Germany, Greece, Guadeloupe, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Pakistan, Philippines, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, South Korea, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = NAY
|
|
|
|