| | Class 2 Device Recall da Vinci S, Si Mega Needle Driver Instrument |  |
| Date Initiated by Firm | December 09, 2025 |
| Date Posted | March 30, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1698-2026 |
| Recall Event ID |
98383 |
| 510(K)Number | K050369 |
| Product Classification |
System, surgical, computer controlled instrument - Product Code NAY
|
| Product | Brand Name: da Vinci S, Si Mega Needle Driver Instrument
Product Name: da Vinci S, Si Mega Needle Driver Instrument
Model/Catalog Number: 420194
Software Version: NA
Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use.
All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient).
Component: NA |
| Code Information |
Lot Code: 420194, Mega Needle Driver
GTIN: 00886874111611
FDA Medical Device Listing Number: D088920
Material Number: 420194-08
Batch / System Numbers:
M10120201 SH1182
M20120412 SH1182
M10130121 SH1570
M10130306 SH1570
M10110602 SH0547
M10110627 SH0547
M10121129 SH0059
M10121214 SH0059
M10130429 SH0059
M10130521 SH0059
M10121105 SH1557
M10130517 SH1837
M10121109 SH1390
M10130508 SH1390
M10131004 SH1390
M10130306 USG689
M10130412 USG689
M10110302 SH1368
M10110614 SH1368
M10110620 SH1368
M10110817 SH1368
M10120705 SH1368
M10121022 SH1368
M10130429 SH1368
M10130509 SH1368
M10130722 SH1368
M10110131 USG294
M10110511 USG294
M10110616 USG294
M10110811 USG294
M10111003 USG294
M20111103 USG294
M10111206 USG294
M20120103 USG294
M10120130 USG294
M10120305 USG294
M10120425 USG294
M10120713 USG294
M10121004 USG294
M10121017 USG294
M10121022 USG294
M10121105 USG294
M10121107 USG294
M10121206 USG294
M10121207 USG294
M10130108 USG294
M10130112 USG294
M10130201 USG294
M10130213 USG294
M10130215 USG294
M10130529 USG294
M11130608 USG294
M10130718 USG294
M10130829 USG294
M10120321 SG2026
M10120605 SG2026
M10111114 SH0825
M10110330 SH0935
M10110401 SH0935
M10120123 SH0935
M10120313 SH0935
M10120404 SH0935
M10120802 SH0935
M10120919 SH0935
M10121121 SH0935
M10110302 SH0729
M10110503 SH0729
M10110701 SH0729
M10110809 SH0729
M10110817 SH0729
M10110825 SH0729
M10110913 SH0729
M10110923 SH0729
M10111116 SH0729
M20111123 SH0729
M10111201 SH0729
M10120120 SH0729
M10120130 SH0729
M10120210 SH0729
M10120217 SH0729
M10120301 SH0729
M10120412 SH0729
M10120507 SH0729
M10120525 SH0729
M20120530 SH0729
M10120703 SH0729
M10120724 SH0729
M10120803 SH0729
M10120810 SH0729
M10120827 SH0729
M10121025 SH0729
M10130102 SH0729
M10130212 SH0729
M10130213 SH0729
M10130314 SH0729
M10130418 SH0729
M10130517 SH0729
M10130603 SH0729
M10130606 SH0729
M10130617 SH0729
M10130701 SH0729
M10131014 SH0729
N10140430 SH0729
M10101206 SH0035
M10110624 SH0035
M10101206 SH1391
S10110307 SH1391
M10110302 SH1391
M10110701 SH1391
M10120404 SH1391
M10120705 SH1391
M10121010 SH1391
M10130118 SH1391
M10130125 SH1391
M10130515 SH1391
M10130626 SH1391
M10121206 SH1604
M10130308 SH1604
M10101127 SH0495
M10101217 SH0495
M10110131 SH0495
M10110512 SH0495
M10110817 SH0495
M10110901 SH0495
M10111008 SH0495
M20111221 SH0495
M10120217 SH0495
M10120224 SH0495
M10110527 SH0495
M10120727 SH0495
M10121004 SH0495
M10130102 SH0495
M10130213 SH0495
M10130301 SH0495
M10130812 SH0495
M10130924 SH0495
M10111201 SH1015
M10111208 SH1015
M10120217 SH1015
M10120425 SH1015
M10120618 SH1015
M10120713 SH1015
M10120823 SH1015
M10120827 SH1015
M10121022 SH1015
M10130212 SH1015
M10130213 SH1015
M10130828 SH1015
M10120831 SH1385
M10120918 SH1385
M10110412 SH2144
M10110706 SH2144
M10110923 SH2144
M10111008 SH2144
M10111205 SH2144
M20120103 SH2144
M10120210 SH2144
M10120713 SH2144
M10121121 SH2144
M10130325 SH2144
M10130529 SH2144
M10110302 SH0761
M10121129 SH0761
M10130719 SH0761
M10131001 SH0761
M20120322 SH1276
M10120525 SH1276
M10130503 USG518
M10130821 USG518
M10130828 USG518
M10131014 USG518
Material Number: 420194-10
Batch / System Numbers:
N10170714 SH1182
N10170817 SH1182
N10171019 SH1182
M10140303 SH1570
N10140805 SH1570
N10150702 SH1570
N10160311 SH1570
N10160421 SH1570
N10161206 SH1570
N10161223 SH1570
N10180227 SG704
M10131121 SH0547
N10140609 SH0547
M10140221 SH0059
M10140307 SH0059
N10140508 SH0059
N10140827 SH0059
N10150320 SH0059
N10160519 SH0059
N10160728 SH0059
N11160907 SH0059
N11161013 SH0059
N10161129 SH0059
N10170419 SH0059
N11170407 SH0059
N10170612 SH0059
N10170705 SH0059
N10180202 SH0059
N10180323 SH0059
N10180321 SH0059
N10180412 SH0059
M10140225 SH1557
N10140605 SH1557
N10140620 SH1557
N10140827 SH1557
N12140929 SH1557
N10150211 SH1557
N10150320 SH1557
N10150528 SH1557
N10150616 SH1557
N10150721 SH1557
N10151209 SH1557
N10160307 SH1557
N10160519 SH1557
N10160509 SH1557
N10160728 SH1557
N10160907 SH1557
N10161031 SH1557
N11170111 SH1557
N11161216 SH1557
N11170427 SH1557
N10170504 SH1557
N10170605 SH1557
N10170914 SH1557
N10171016 SH1557
N10171027 SH1557
N10180118 SH1557
N10180223 SH1557
N10180321 SH1557
N10180427 SH1557
N10150604 SH1964
N10150626 SH1964
N10151207 SH1964
N10151211 SH1964
N10180131 SH1964
N10180223 SH1964
N10180404 SH1964
N10180426 SH1964
N10150814 SH1837
N10160105 SH1837
N10160728 SH1837
N11160901 SH1837
N10150717 SH1390
N10160419 SH1390
M10131211 USG689
N10160519 USG689
M10140515 SH1368
N11141113 SH1368
N10160105 SH1368
N10160718 SH1368
N10170901 SH1368
M10131102 USG294
M10140107 USG294
M10140109 USG294
M10140221 USG294
M10140408 USG294
N10140516 USG294
N10140923 USG294
N10141016 USG294
N10150219 USG294
N10150325 USG294
N10150723 USG294
N10151002 USG294
N10151013 USG294
N10160104 USG294
N10160307 USG294
N10160504 USG294
N11160616 USG294
N11160720 USG294
N10160907 USG294
N11160923 USG294
N10161118 USG294
N10161207 USG294
N10170302 USG294
N10170421 USG294
N10170621 USG294
N10170817 USG294
N10170927 USG294
N10180202 USG294
N10171214 USG294
N10180131 USG294
M10140324 SH2186
N10141127 SH2186
N10150320 SH2186
N10150626 SH2186
N10150731 SH2186
N10160901 SH2186
N10170427 SH2186
N10170509 SH2186
N10171221 SH2186
N10180321 SH2186
N11161207 SH0825
N10170511 SH0825
N10180301 SH0825
M10131109 SH1378
M10131104 SH0935
M10140225 SH0935
N10140718 SH0935
M10131109 SH0729
M10131130 SH0729
M10131202 SH0729
M10140107 SH0729
N10140611 SH0729
N10140716 SH0729
N10140717 SH0729
N10140728 SH0729
N10140826 SH0729
N11140929 SH0729
N10141118 SH0729
N10141205 SH0729
N10150219 SH0729
N10150226 SH0729
N10150421 SH0729
N10150428 SH0729
N10150721 SH0729
N10150804 SH0729
N10151123 SH0729
N10151203 SH0729
N10160311 SH0729
N11160419 SH0729
N11160907 SH0729
N10161019 SH0729
N10170324 SH0729
N11170410 SH0729
N10170511 SH0729
N10170621 SH0729
N10170810 SH0729
N10171207 SH0729
N10180202 SH0729
N11171214 SH0729
N11180321 SH0729
N10180307 SH0729
N10180412 SH0729
N10140722 SH0035
N10140910 SH0035
N10150519 SH0035
N10150707 SH0035
N11160711 SH0035
N10160804 SH0035
N10170207 SH0035
N10170511 SH0035
N10171207 SH0035
M10131121 SH1391
M10131217 SH1391
N10140516 SH1391
M10140206 SH2125
N10140530 SH2125
N10140611 SH2125
N10140926 SH2125
N10150105 SH2125
N10150413 SH2125
N10150504 SH2125
N10150804 SH2125
N10150918 SH2125
N10150928 SH2125
N10160114 SH2125
N10160504 SH2125
N10160519 SH2125
N10160601 SH2125
N10161019 SH2125
N10161216 SH2125
N10170404 SH2125
N10161025 SH2125
M10131104 SH2125
N10170712 SH2125
N10171103 SH2125
N11171214 SH2125
N10180223 SH2125
N10150818 SH2125
N10180323 SH2125
N10180321 SH2125
N10180427 SH2125
N10160219 SH1076
N10170427 SH1076
M10140116 SH1604
N10170926 SH1604
N10180122 SH1604
N10180111 SH1604
M10140121 SH0495
M10140303 SH0495
N10140522 SH0495
N10140808 SH0495
N11141112 SH0495
N10141127 SH0495
N11141205 SH0495
N10150226 SH0495
N10150626 SH0495
N10150811 SH0495
N10160523 SH0495
N10160822 SH0495
N11160822 SH0495
N10161118 SH0495
N10170301 SH0495
N10170927 SH0495
N10180202 SH0495
M10131119 SH1015
M11140109 SH1015
M10140131 SH1015
M10140307 SH1015
N10140425 SH1015
N10140708 SH1015
N10140904 SH1015
N11140902 SH1015
N10150204 SH1015
N10150211 SH1015
N10150216 SH1015
N10150511 SH1015
N10150610 SH1015
N10150630 SH1015
N10150707 SH1015
N10150828 SH1015
N10150916 SH1015
N10150928 SH1015
N10160212 SH1015
N10160229 SH1015
N10160307 SH1015
N10160330 SH1015
N10160504 SH1015
N10160610 SH1015
N10160616 SH1015
N10160630 SH1015
N10160706 SH1015
N11160711 SH1015
N10160822 SH1015
N11160907 SH1015
N10160901 SH1015
N11161013 SH1015
N11161118 SH1015
N10161118 SH1015
N10161223 SH1015
N10170202 SH1015
N10170302 SH1015
N10170316 SH1015
N11170324 SH1015
N10170331 SH1015
N10170504 SH1015
N10170511 SH1015
N10170515 SH1015
N10170623 SH1015
N10170710 SH1015
N10170720 SH1015
N10170901 SH1015
N10171003 SH1015
N10171103 SH1015
N10180312 SH1015
N10171221 SH1015
N10180122 SH1015
N10171226 SH1015
N10180124 SH1015
N10180227 SH1015
N10180426 SH1015
N10180412 SH1385
M10140303 SH2144
M10140326 SH2144
N10140605 SH2144
N10140805 SH2144
N10150223 SH2144
N10160519 SH2144
N10170913 SH2144
N10150312 SH0761
N10150528 SH0761
M10140206 SH1276
M10140213 SH1276
M10140327 SH1276
N10140508 SH1276
N10140925 SH1276
N10150305 SH1276
N10151021 SH1276
N10151125 SH1276
N10140902 USG518
N12140929 USG518
N10141008 USG518
M10140108 USG518
N10150320 USG518
N10161111 USG518
N10170103 USG518
Material Number: 420194-12
Batch / System Numbers:
N10190729 SH1182
N10180921 SH1557
N10190104 SH1557
N10190211 SH1557
N10190326 SH1557
N12190423 SH1557
N10190719 SH1557
N11191021 SH1557
N10191216 SH1557
N10210329 SH1557
N10210524 SH1557
N10210517 SH1557
N10190108 SH1837
N10200727 SH1837
N10201012 SH1837
N10201103 SH1837
N10210412 SH1837
N10191111 USG689
N10191125 USG689
N10190805 SH1368
N10190930 SH1368
N10180808 USG294
N10180906 SH2186
N10190522 SH2186
N10201012 SH2186
N10190226 SH0825
N10190326 SH0825
N10210412 SH0825
N10181126 SH0729
N10190107 SH0729
N10190212 SH0729
N10180814 SH2125
N10180815 SH2125
N10181008 SH2125
N10190107 SH2125
N10190212 SH2125
N10190327 SH2125
N10190729 SH2125
N10191111 SH1076
N10180808 SH1604
N10190930 SH1604
N10190108 SH0495
N10190805 SH0495
N10190930 SH0495
N10200727 SH0495
N10180814 SH1015
N10180813 SH1015
N10181017 SH1015
N10190107 SH1015
N10190108 SH1015
N10190307 SH1015
N10190308 SH1015
N10190522 SH1015
N10190805 SH1015
N10190923 SH1015
N10190917 SH1015
N10191111 SH1015
N11191202 SH1015
N10191216 SH1015
N10191125 SH1015
N10191021 SH1015
N10200810 SH1015
N10201103 SH1015
N10210802 SH1015
N10181024 SH1385
N10191111 SH1385
N10190107 SH1279
N10190212 SH1279
N10190327 SH1279
N10210607 SH1279
Material Number: 420194-13
Batch / System Numbers:
K11220626 SH1964
K10220228 SH1837
|
| FEI Number |
3001675293
|
Recalling Firm/ Manufacturer |
Intuitive Surgical, Inc.
1266 Kifer Rd
Sunnyvale
CA
94086-5304
|
| For Additional Information Contact | Shahbaz Khan 1-408-5232443 |
Manufacturer Reason for Recall | Due to increased complaints for broken/frayed grip cables for reusable instruments. |
FDA Determined Cause 2 | Process control |
| Action | On 12/29/2025, the firm sent via email an "Urgent: Medical Device Recall" Letter informing customers that intuitive has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci S and Si reusable instruments.
Customers are instruct to:
1. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form.
2. Identify and quarantine any affected product(s). Affected product(s) listed above may be returned by sending an email with quantities and lot numbers to your regional customer service.
3. Ensure to include the FSCA number ISIFA2025-15-R in the return notes.
4. Credit will be provided based on the number of remaining lives.
5. Shared or further distributed these products with other sites, make sure appropriate staff at the site receive and understand this notification so they locate and return their affected product.
For questions or concern, contact Clinical Sales Representative or contact Intuitive Customer Service at the numbers listed below:
" North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com.
" Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com
" South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com
" Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com
" India: +1-800-103-6952 (9 am to 6 pm IT)
" Taiwan: +0800-86-8181 (9 am to 6 pm CT) |
| Quantity in Commerce | 7819 |
| Distribution | Worldwide - U.S. Nationwide distribution in the states of CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MO, MS, NC, NJ, NY, OH, PA, TX, VA and WI. The countries of Argentina, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Ecuador, France, Germany, Greece, Guadeloupe, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Pakistan, Philippines, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, South Korea, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|