| Class 2 Device Recall | |
Date Initiated by Firm | May 26, 2005 |
Date Posted | July 16, 2005 |
Recall Status1 |
Terminated 3 on June 12, 2006 |
Recall Number | Z-1016-05 |
Recall Event ID |
32303 |
510(K)Number | K980283 |
Product Classification |
Enzyme Immunoassay, Carbamazepine - Product Code KLT
|
Product | Vitros Chemistry CRBM Slides, REF 889 2382. There are 90 slides per box. For in vitro diagnostic use. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626. |
Code Information |
Generations 18, 19, 20, 21, and 23 |
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics 100 Indigo Creek Dr Rochester NY 14626-5101
|
For Additional Information Contact | Judy M. Strzepek 908-218-8524 |
Manufacturer Reason for Recall | Falsely elevated Carbamazepine results. |
FDA Determined Cause 2 | Other |
Action | Letters dated 5/26/05 with instructions to discontinue use and discard slides. Customers also instructed to review previous results with Laboratory Medical Director to determine need for further consultation with health care provider. |
Quantity in Commerce | 5,846 boxes |
Distribution | 423 U.S. consignees nationwide, and 22 OCD/J&J foreign affiliates (Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Peru, Ecuador, India, Japan, Mexico, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain) |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = KLT
|
|
|
|