• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall Medline Strider Midi 3 Scooter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 3 Device Recall Medline Strider Midi 3 Scootersee related information
Date Initiated by FirmJuly 20, 2006
Date PostedSeptember 23, 2006
Recall Status1 Terminated 3 on September 02, 2009
Recall NumberZ-1522-06
Recall Event ID 36171
510(K)NumberK042763 
Product Classification scooter - Product Code INI
ProductMedline Strider Midi 3 Scooter; a three wheeled battery operated scooter; manufactured by KYMCO, Kaohsiung, Taiwan for Medline Industries, Inc., One Medline Place, Mudelein, IL 60060 USA; model MDS807500 - red and MDS807500B - blue
Code Information model MDS807500 - red and model MDS807500B - blue, scooter no., controller no.: EV10DB-*US00110*,13010613; EV10DB-*US00102*,13010615; EV10DB-*US00103*,53010802; EV10DB-*US00101*,53010803; EV10DB-*US00106*,53010806; EV10DB-*US00107*,53010807; EV10DB-*US00111*,53011072; EV10DB-*US00118*,53050871; EV10DB-*US00112*,53050872; EV10DB-*US00115*,53050873; EV10DB-*US00116*,53050874; EV10DB-*US00117*,53050875; EV10DB-*US00105*,73010706; EV10DB-*US00108*,73010710; EV10DB-*US00109*,73010711; EV10DB-*US00119*,73050684; EV10DB-*US00113*,73050685; EV10DB-*US00120*,73050686; EV10DB-*US00121*,73050688; EV10DB-*US00130*,73050805; EV10DB-*US00123*,73050806; EV10DB-*US00127*,73050807; EV10DB-*US00126*,73050808; EV10DB-*US00129*,73050809; EV10DB-*US00128*,73050810; EV10DB-*US00125*,73050812; EV10DB-*US00122*,73060000; EV10DB-*US00151*,13010484; EV10DB-*US00131*,23010651; EV10DB-*US00137*,33010608; EV10DB-*US00133*,53010863; EV10DB- *US00152*,53010928; EV10DB-*US00135*,53010929; EV10DB-*US00157*,53050630; EV10DB-*US00155*,53050631; EV10DB-*US00153*,53050632; EV10DB-*US00142*,53050865; EV10DB-*US00143*,53050866; EV10DB-*US00149*,53050867; EV10DB-*US00145*,53050868; EV10DB-*US00140*,53050869; EV10DB-*US00146*,53050870; EV10DB-*US00132*,73010716; EV10DB-*US00136*,73010765; EV10DB-*US00150*,73010824; EV10DB-*US00138*,73010825; EV10DB-*US00156*,73050441; EV10DB-*US00159*,73050443; EV10DB-*US00158*,73050444; EV10DB-*US00147*,73050680; EV10DB-*US00148*,73050681; EV10DB-*US00141*,73050682; EV10DB-*US00139*,73050683; EV10DB-*US00160*,73050442; EV10DB-*UK00150*,53010988; EV10DB-*UK00155*,53010676; EV10DB-*UK00158*,53010677; EV10DB-*UK00163*,73010884; EV10DB-*UK00173*,13010532; EV10DB-*UK00179*,53010668; EV10DB-*UK00181*,53010670; EV10DB-*UK00189*,13010530; EV10DB-*UK00192*,73010599; EV10DB-*UK00195*,53010791; EV10DB-*UK00200*,53010785; EV10DB-*UK00202*,53010789; EV10DB-*UK00203*,53010847; EV10DB-*UK00205*,73010750; EV10DB-*UK00206*,73010749; EV10DB-*UK00209*,53010787; EV10DB-*UK00220*,73010780; EV10DB-*UK00222*,53010776; EV10DB-*UK00227*,33010692; EV10DB-*UK00228*,73010691; EV10DB-*UK00230*,53010779; EV10DB-*UK00231*,73010689; EV10DB-*UK00232*,43010700; EV10DB-*UK00233*,53010788; EV10DB-*UK00236*,33010695; EV10DB-*UK00241*,13010525; EV10DB-*UK00248*,53010908; EV10DB-*UK00251*,33010620; EV10DB-*UK00252*,73010802; EV10DB-*UK00253*,53010907; EV10DB-*UK00256*,33010625; EV10DB-*UK00261*,23010532; EV10DB-*UK00265*,43010632; EV10DB-*UK00268*,73010628; EV10DB-*UK00270*,53010698; EV10DB-*UK00272*,53010688; EV10DB-*UK00276*,53010687; EV10DB-*UK00278*,23010545; EV10DB-*UK00279*,13010528; EV10DB-*UK00282*,23010533; EV10DB-*UK00286*,23010531; EV10DB-*UK00289*,73010616; EV10DB-*UK00292*,43010832; EV10DB-*UK00297*,33010838; EV10DB-*UK00298*,33010839; EV10DB-*UK00299*,43010831; EV10DB-*UK00303*,53010962; EV10DB-*UK00307*,73010859; EV10DB-*UK00309*,43010808; EV10DB-*UK00310*,43010807; EV10DB-*UK00311*,33010810; EV10DB-*UK00313*,43010811; EV10DB-*UK00315*,13010609; EV10DB-*UK00319*,43010809; EV10DB-*UK00326*,33010808; EV10DB-*UK00330*,73010858; EV10DB-*UK00341*,43010238; EV10DB-*UK00349*,53010825; EV10DB-*UK00359*,73010665; EV10DB-*UK00360*,53010738; EV10DB-*UK00361*,73010663; EV10DB-*UK00366*,73010871; EV10DB-*UK00367*,53010734; EV10DB-*UK00370*,73010776; EV10DB-*UK00376*,53010876; EV10DB-*UK00383*,43010232; EV10DB-*UK00385*,73010651; EV10DB-*UK00390*,43010602; EV10DB-*UK00398*,33010591; EV10DB-*UK00400*,33010592; EV10DB-*UK00403*,53010978; EV10DB-*UK00409*,53010722; EV10DB-*UK00411*,53010909; EV10DB-*UK00421*,23010568; EV10DB-*UK00425*,13010562; EV10DB-*UK00426*,13010563; EV10DB-*UK00429*,13010503; EV10DB-*UK00430*,73010588; EV10DB-*UK00431*,13010561; EV10DB-*UK00432*,53010914; EV10DB-*UK00433*,73010810; EV10DB-*UK00435*,53010910; EV10DB-*UK00446*,43010624; EV10DB-*UK00452*,73010754; EV10DB-*UK00461*,43010695; EV10DB-*UK00475*,53050530; EV10DB-*UK00477*,73050341; EV10DB-*UK00479*,73050344; EV10DB-*UK00487*,73050676; EV10DB-*UK00490*,43050817; EV10DB-*UK00492*,53050898; EV10DB-*UK00500*,73050732; EV10DB-*UK00501*,73050729; EV10DB-*UK00502*,73050733; EV10DB-*UK00505*,73050678; EV10DB-*UK00507*,53050915; EV10DB-*UK00510*,43050821; EV10DB-*UK00512*,53050914; EV10DB-*UK00521*,73050460; EV10DB-*UK00529*,53050677; EV10DB-*UK00560*,73050719; EV10DB-*UK00581*,53050826; EV10DB-*UK00582*,43050567; EV10DB-*UK00583*,43050806; EV10DB-*UK00595*,53050881; EV10DB-*UK00598*,73050693; EV10DB-*UK00600*,73050311; EV10DB-*UK00606*,73050696; EV10DB-*UK00613*,43050804; EV10DB-*UK00619*,73050387; EV10DB-*UK00621*,53050577; EV10DB-*UK00623*,53050576; EV10DB-*UK00627*,73050495; EV10DB-*UK00628*,53050687; EV10DB-*UK00629*,73050500; EV10DB-*UK00630*,73050497; EV10DB-*UK00636*,53050686; EV10DB-*UK00637*,73050462; EV10DB-*UK00639*,53050655; EV10DB-*UK00640*,53050618; EV10DB-*UK00643*,53050617; EV10DB-*UK00645*,73050508; EV10DB-*UK00648*,53050699; EV10DB-*UK00653*,53050654; EV10DB-*UK00655*,53050697; EV10DB-*UK00657*,53050696; EV10DB-*UK00659*,43050572; EV10DB-*UK00660*,53050653; EV10DB-*UK00661*,73050467; EV10DB-*UK00666*,73050468; EV10DB-*UK00668*,43050809; EV10DB-*UK00670*,53050906; EV10DB-*UK00671*,43050812; EV10DB-*UK00673*,73050456; EV10DB-*UK00675*,53050644; EV10DB-*UK00676*,73050455; EV10DB-*UK00678*,33050776; EV10DB-*UK00679*,53050890; EV10DB-*UK00680*,43050811; EV10DB-*UK00681*,43050813; EV10DB-*UK00683*,73050720; EV10DB-*UK00685*,53050908; EV10DB-*UK00687*,53050929; EV10DB-*UK00689*,53050579; EV10DB-*UK00691*,73050394; EV10DB-*UK00693*,53050580; EV10DB-*UK00695*,73050324; EV10DB-*UK00696*,73050392; EV10DB-*UK00698*,43050457; EV10DB-*UK00699*,53050924; EV10DB-*UK00700*,73050737; EV10DB-*UK00701*,53050926; EV10DB-*UK00702*,73050741; EV10DB-*UK00703*,53050925; EV10DB-*UK00706*,73050742; EV10DB-*UK00709*,53050646; EV10DB-*UK00712*,73050740; EV10DB-*UK00713*,53050907; EV10DB-*UK00717*,73050722; EV10DB-*UK00721*,73050381; EV10DB-*UK00723*,53050571; EV10DB-*UK00725*,43050837; EV10DB-*UK00726*,33050793; EV10DB-*UK00727*,53050922; EV10DB-*UK00730*,33050794; EV10DB-*UK00731*,33050431; EV10DB-*UK00733*,73050383; EV10DB-*UK00736*,53050611; EV10DB-*UK00737*,33050792; EV10DB-*UK00739*,73050382; EV10DB-*UK00740*,73050420; EV10DB-*UK00741*,73050323; EV10DB-*UK00742*,53050509; EV10DB-*UK00743*,73050422; EV10DB-*UK00746*,73050325; EV10DB-*UK00747*,53050508; EV10DB-*UK00748*,73050659; EV10DB-*UK00750*,53050511; EV10DB-*UK00753*,73050321; EV10DB-*UK00755*,53050612; EV10DB-*UK00756*,73050421; EV10DB-*UK00758*,73050419; EV10DB-*UK00760*,73050347; EV10DB-*UK00761*,73050346; EV10DB-*UK00762*,53050534; EV10DB-*UK00763*,73050658; EV10DB-*UK00766*,43050857; EV10DB-*UK00767*,43050432; EV10DB-*UK00768*,53050935; EV10DB-*UK00769*,43050434; EV10DB-*UK00770*,43050430; EV10DB-*UK00771*,73050652; EV10DB-*UK00773*,73050744; EV10DB-*UK00777*,53050844; EV10DB-*UK00778*,53050934; EV10DB-*UK00781*,73050384; EV10DB-*UK00782*,53050568; EV10DB-*UK00785*,53050569; EV10DB-*UK00786*,73050747; EV10DB-*UK00788*,53050536; EV10DB-*UK00789*,53050613; EV10DB-*UK00791*,73050348; EV10DB-*UK00792*,43050854; EV10DB-*UK00795*,53050837; EV10DB-*UK00797*,53050610; EV10DB-*UK00800*,73050423; EV10DB-*UK00805*,73050385; EV10DB-*UK00808*,43050433; EV10DB-*UK00809*,33050430; EV10DB-*UK00811*,53050931; EV10DB-*UK00812*,53050535; EV10DB-*UK00816*,43050856; EV10DB-*UK00817*,73050738; EV10DB-*UK00818*,53050538; EV10DB-*UK00819*,33050429;
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries Inc
1 Medline Pl
Mundelein IL 60060-4486
For Additional Information Contact
800-289-9793
Manufacturer Reason
for Recall
The P&G Solo 60A electronic control unit on the Midi Scooters may overheat.
FDA Determined
Cause 2
Other
ActionMedline sent 'Voluntary Field Correction' letters dated 7/20/06 to the direct accounts on the same date, informing them that the Strider Midi 3 and Midi 4 scooters have an older version of the electronic control unit that may overheat, which Medline will replace with a newer version of the controller. Any questions were directed to 800-289-9793, press 1, then press 3 and ask to transfer to 4674 (8:00 AM to 5:00 PM CST).
Quantity in Commerce269 scooters
DistributionNationwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = INI
-
-