| Date Initiated by Firm | March 01, 2007 |
| Date Posted | September 28, 2007 |
| Recall Status1 |
Terminated 3 on May 01, 2012 |
| Recall Number | Z-0680-2007 |
| Recall Event ID |
37498 |
| 510(K)Number | K021049 K022069 K024012 |
| Product Classification |
Image-intensified fluoroscopic x-ray system - Product Code JAA
|
| Product | OEC FluoroTrak 9800 PLUS, image-intensified fluoroscopic x-ray system, GE Healthcare, Salt Lake City, UT. |
| Code Information |
Serial numbers: 89-1579-N, 89-2529-N, 82-2160-N, 8S-1932-N, 8S-2226-N. |
Recalling Firm/ Manufacturer |
GE OEC Medical Systems, Inc 384 Wright Brothers Dr Salt Lake City UT 84116-2862
|
| For Additional Information Contact | Karen Tolson 847-501-1884 |
Manufacturer Reason for Recall | During routine service, primary and/or secondary collimators may not have been properly reinstalled. |
FDA Determined Cause 2 | Other |
| Action | Customers were notified by letter on 03/01/2007. They were told to continue using the machines and that an inspection would be scheduled immediately to verify that collimators had been correctly installed. |
| Quantity in Commerce | 5 units |
| Distribution | Nationwide and Germany. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JAA
|