• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall CELLDYN 3700 Hematology Analyzer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CELLDYN 3700 Hematology Analyzersee related information
Date Initiated by FirmMay 11, 2007
Date PostedAugust 29, 2007
Recall Status1 Terminated 3 on June 19, 2008
Recall NumberZ-1200-2007
Recall Event ID 38322
510(K)NumberK980614 K991605 
Product Classification Hematology Analyzer - Product Code GKZ
ProductAbbott CELL-DYN 3700 Hematology Analyzer
Code Information Model number CELL-DYN 3700 CS, List Numbers 02H30-01and 91340-02, all serial numbers distributed on or before May 3, 2007. Model number CELL-DYN 3700 SL, List Numbers 02H31-01 and 02H31-03, all serial numbers distributed on or before May 3, 2007: 02H30-01 EA CD3700 CS(EA) 1.000 20222AN 01/16/2004 0708421226 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 21423AK 04/16/2003 0705836717 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 20575AK 01/17/2001 0700657123 ZINS 1024, 02H30-01 EA CD3700 CS(EA) 1.000 20088AN 11/14/2001 0701853939 ZINS 1024, 02H31-01 EA CD3700 SL(EA) 1.000 21086AK 04/02/2002 0702540521 ZINC 1024, 02H31-01 EA CD3700 SL 1.000 20158AK 01/20/2000 30001171534T2 NC 1024, 02H30-01 EA CD3700 CS 1.000 19966AN 03/17/2000 30003160556T2 NC 1024, 02H31-01 EA CD3700 SL 1.000 20210AK 02/15/2000 30002100642T2 NC 1024, 02H30-01 EA CD3700 CS(EA) 1.000 19999AN 11/20/2000 0700475941 ZINC 1024, 02H31-01 EA CD3700 SL(EA) 1.000 20333AK 05/30/2000 0700023564 ZINS 1024, 02H31-01 EA CD3700 SL(EA) 1.000 21162AK 05/29/2002 0702793182 ZINC 1024, 02H31-01 EA CD3700 SL(EA) 1.000 20662AK 10/09/2001 0701535236 ZINC 1024, 02H31-01 EA CD3700 SL(EA) 1.000 21151K 04/26/2002 0702654355 ZINC 1024, 02H31-01 EA CD3700 SL 1.000 20215AK 02/23/2000 30002110316T2 NC 1024, 02H31-01 EA CD3700 SL(EA) 1.000 20564AK 11/29/2000 0700519556 ZINC 1024, 02H31-01 EA CD3700 SL(EA) 1.000 20564AK 11/29/2000 0700519556 ZINC 1024, 02H31-01 EA CD3700 SL(EA) 1.000 21526AK 08/22/2003 0707048787 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 21195AK 05/29/2002 0702793121 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 21193AK 05/29/2002 0702759492 ZOR 1025, 02H30-01 EA CD3700 CS 1.000 19953AN 01/21/2000 39912022083U3 SALE 1025 02H30-01 EA CD3700 CS 1.000 19966AN 03/08/2000 30002101167U2 SALE 1025 02H31-01 EA CD3700 SL 1.000 20210AK 01/27/2000 39912213420U2 SALE 1025 02H31-01 EA CD3700 SL 1.000 20215AK 01/27/2000 39912212921U2 SALE 1025 02H31-01 EA CD3700 SL 1.000 20333AK 05/18/2000 30004103499U2 SALE 1025, 02H31-01 EA CD3700 SL(EA) 1.000 20565AK 11/29/2000 0700519549 ZINS 1024, 02H30-01 EA CD3700 CS 1.000 19953AN 01/27/2000 30001270601T2 NC 1024, 02H30-01 EA CD3700 CS(EA) 1.000 19999AN 10/31/2000 0700430444 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20088AN 10/25/2001 0701722636 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20106AN 02/14/2002 0702285397 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20564AK 11/30/2000 0700531142 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20565AK 11/30/2000 0700531141 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20574AK 12/01/2000 0700419159 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20575AK 12/01/2000 0700419161 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20576AK 12/01/2000 0700419160 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20662AK 03/08/2001 0700818411 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21086AK 03/01/2002 0702067225 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21151AK 04/22/2002 0702638488 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21162AK 04/26/2002 0702536142 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 21508AK 08/18/2003 0706963179 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 22038AK 03/10/2005 0721127803 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 22039AK 03/10/2005 0721127803 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 22035AK 03/08/2005 0721102104 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 22041AK 03/10/2005 0721127061 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 20576AK 01/17/2001 0700657124 ZINS 1024, 02H31-01 EA CD3700 SL(EA) 1.000 20574AK 11/29/2001 0701921151 ZINS 1024, 02H31-01 EA CD3700 SL(EA) 1.000 22055AK 03/22/2005 0721231650 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 21836AK 08/26/2004 0710348806 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 20368AK 06/15/2000 0700067140 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20369AK 06/20/2000 0700087363 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20370AK 06/20/2000 0700087363 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20371AK 06/20/2000 0700087363 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20372AK 06/20/2000 0700087363 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20373AK 06/21/2000 0700067141 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20340AK 05/26/2000 0700022704 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20344AK 05/30/2000 0700022198 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20347AK 05/26/2000 0700022704 ZOR 1025, 02H30-01 EA CD3700 CS(EA) 1.000 20010AN 10/11/2000 0700390326 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20036AN 02/27/2001 0700702499 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20037AN 02/28/2001 0700665110 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20039AN 03/08/2001 0700764419 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20059AN 05/23/2001 0701034352 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20069AN 07/13/2001 0701274789 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20073AN 07/25/2001 0701296466 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20075AN 07/25/2001 0701296466 ZOR 1025 02H31-03 EA CD3700 SL REFURB(EA) 1.000 20117AK96 03/12/2004 0708902744 ZOR 1025 02H31-03 EA CD3700 SL REFURB(EA) 1.000 20154AK96 08/28/2006 0724669259 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20200AN 07/01/2003 0706596666 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20201AN 07/01/2003 0706596666 ZOR 1025 02H30-01 EA CD3700 CS(EA) 1.000 20233AN 02/17/2004 0708707057 ZOR 1025 02H31-03 EA CD3700 SL REFURB(EA) 1.000 20465AK96 09/29/2006 0724957687 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20645AK 02/27/2001 0700702498 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20646AK 02/27/2001 0700702498 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20700AK 03/29/2001 0700368389 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20703AK 03/29/2001 0700368389 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20704AK 03/29/2001 0700368389 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20705AK 03/29/2001 0700368389 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20707AK 03/30/2001 0700844563 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20708AK 03/29/2001 0700368389 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20710AK 03/30/2001 0700844563 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20711AK 03/30/2001 0700844563 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20712AK 03/30/2001 0700844563 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20716AK 03/30/2001 0700844563 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20742AK 04/24/2001 0700847325 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20869AK 08/06/2001 0701363650 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20870AK 08/07/2001 0701326067 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20871AK 08/07/2001 0701326067 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20877AK 08/14/2001 0701419772 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20881AK 08/14/2001 0701419772 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20882AK 08/14/2001 0701419772 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20883AK 08/15/2001 0701428194 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20884AK 08/15/2001 0701428194 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20885AK 08/16/2001 0701419773 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20886AK 08/16/2001 0701419773 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20887AK 08/16/2001 0701419773 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20888AK 08/16/2001 0701419773 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20889AK 08/17/2001 0701419778 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20891AK 08/21/2001 0701419774 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20892AK 08/21/2001 0701419774 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20893AK 08/21/2001 0701419774 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20894AK 08/23/2001 0701458396 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20896AK 08/28/2001 0701419775 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20897AK 08/23/2001 0701458396 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20898AK 08/23/2001 0701439537 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20902AK 08/28/2001 0701419775 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20903AK 08/29/2001 0701419776 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20904AK 08/28/2001 0701419775 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20905AK 08/28/2001 0701419775 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20907AK 08/28/2001 0701419775 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20908AK 08/31/2001 0701419777 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20909AK 08/29/2001 0701419776 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20910AK 08/29/2001 0701419776 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20911AK 08/29/2001 0701419776 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20912AK 08/29/2001 0701419776 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20913AK 08/31/2001 0701419777 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20914AK 08/31/2001 0701419777 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20915AK 08/31/2001 0701419777 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20917AK 08/31/2001 0701419777 ZOR 1025 02H31-03 EA CD3700 SL REFURB(EA) 1.000 20940AK96 03/08/2004 0708891681 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20967AK 10/19/2001 0701730439 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20970AK 10/19/2001 0701730439 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20987AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20989AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20991AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20992AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20993AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20994AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20996AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20997AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20998AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20999AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21000AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21001AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21002AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21003AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21004AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21005AK 12/03/2001 0701935471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21045AK 01/28/2002 0702203580 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21046AK 01/28/2002 0702203580 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21048AK 01/28/2002 0702203580 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21049AK 01/28/2002 0702203580 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21050AK 01/28/2002 0702203580 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21058AK 02/07/2002 0702203581 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21059AK 02/07/2002 0702203581 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21060AK 02/07/2002 0702203581 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21061AK 02/07/2002 0702203581 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21062AK 02/08/2002 0702270892 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21063AK 02/11/2002 0702277975 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21064AK 02/08/2002 0702270892 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21065AK 02/13/2002 0702282548 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21067AK 02/11/2002 0702277975 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21068AK 02/13/2002 0702282548 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21069AK 02/15/2002 0702299928 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21070AK 02/19/2002 0702311964 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21076AK 02/26/2002 0702357048 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21082AK 02/28/2002 0702357049 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21083AK 02/28/2002 0702357049 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21084AK 02/28/2002 0702357049 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21089AK 03/07/2002 0702409793 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21090AK 03/07/2002 0702409793 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21091AK 03/08/2002 0702409156 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21092AK 03/08/2002 0702409156 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21093AK 03/07/2002 0702409793 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21094AK 03/07/2002 0702409793 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21095AK 03/08/2002 0702409156 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21096AK 03/12/2002 0702409789 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21097AK 03/12/2002 0702409789 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21098AK 03/13/2002 0702409790 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21099AK 03/12/2002 0702409789 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21100AK 03/14/2002 0702409791 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21101AK 03/13/2002 0702409790 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21102AK 03/13/2002 0702409790 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21103AK 03/19/2002 0702409792 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21104AK 03/14/2002 0702409791 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21105AK 03/19/2002 0702409792 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21107AK 03/19/2002 0702409792 ZOR 1025 02H31-03 EA CD3700 SL REFURB(EA) 1.000 21127AK96 02/27/2004 0708817941 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21163AK 04/25/2002 0702452514 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21180AK 05/09/2002 0702726859 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21181AK 05/10/2002 0702726862 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21182AK 05/09/2002 0702726859 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21183AK 05/10/2002 0702726862 ZOR 1025 02H31-03 EA CD3700 SL REFURB(EA) 1.000 21244AK96 08/28/2006 0724669259 ZOR 1025 02H31-03 EA CD3700 SL REFURB(EA) 1.000 21245AK96 08/17/2006 0724616503 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21473AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21474AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21475AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21476AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21478AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21479AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21480AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21481AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21482AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21483AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21485AK 07/01/2003 0706596666 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21590AK 11/20/2003 0707872598 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21592AK 11/20/2003 0707872598 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21624AK 01/21/2004 0708229861 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21625AK 01/21/2004 0708229861 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21627AK 01/26/2004 0708229862 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21628AK 01/26/2004 0708229862 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21629AK 01/26/2004 0708230220 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21630AK 01/27/2004 0708520940 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21639AK 02/05/2004 0708585282 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21640AK 02/05/2004 0708585282 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21643AK 02/10/2004 0708585283 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21644AK 02/12/2004 0708585284 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21645AK 02/10/2004 0708585283 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21647AK 02/17/2004 0708688571 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21650AK 02/23/2004 0708707082 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21651AK 02/24/2004 0708707083 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21652AK 02/23/2004 0708707082 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21654AK 03/04/2004 0708707085 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21655AK 03/04/2004 0708707085 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21656AK 03/04/2004 0708707086 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21657AK 03/04/2004 0708707086 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21658AK 03/04/2004 0708707089 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21659AK 03/04/2004 0708707089 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21660AK 03/05/2004 0708788543 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21662AK 03/05/2004 0708707090 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21663AK 03/05/2004 0708707090 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21711AK 05/05/2004 0708707092 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21712AK 05/05/2004 0708707092 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21713AK 05/06/2004 0708707095 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21714AK 05/06/2004 0708707095 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21715AK 05/06/2004 0708707098 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21716AK 05/06/2004 0708707098 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21839AK 08/26/2004 0710361988 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21841AK 08/26/2004 0710361988 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 21843AK 08/26/2004 0710361988 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22022AK 02/10/2005 0720995106 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22111AK 04/13/2005 0721376532 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22113AK 06/02/2005 0721693890 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22114AK 06/02/2005 0721693890 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22118AK 06/02/2005 0721693890 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22119AK 06/02/2005 0721705699 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22138AK 06/02/2005 0721705699 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22154AK 06/02/2005 0721705699 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22169AK 06/02/2005 0721705699 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22171AK 06/02/2005 0721705699 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22184AK 07/27/2005 0722063784 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22186AK 08/01/2005 0722086543 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22187AK 08/01/2005 0722086543 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22200AK 08/01/2005 0722086543 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22201AK 08/01/2005 0722086543 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22215AK 07/27/2005 0722063784 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22217AK 07/27/2005 0722063784 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22218AK 07/27/2005 0722063784 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22220AK 09/09/2005 0722368703 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22221AK 09/09/2005 0722368703 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22233AK 09/08/2005 0722353178 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22260AK 10/12/2005 0722601656 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22261AK 10/12/2005 0722601656 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22287AK 11/10/2005 0722807036 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22303AK 02/02/2006 0723378074 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22318AK 12/20/2005 0723075608 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22341AK 01/18/2006 0723279431 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22357AK 03/03/2006 0723573497 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22385AK 02/02/2006 0723378074 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22404AK 03/24/2006 0723666112 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22486AK 06/08/2006 0724084813 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22518AK 06/27/2006 0724306909 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22579AK 08/14/2006 0724403267 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22581AK 08/14/2006 0724403267 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22582AK 08/14/2006 0724403267 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22584AK 08/14/2006 0724381933 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22586AK 08/16/2006 0724600552 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22606AK 08/31/2006 0724658143 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22610AK 09/11/2006 0724831647 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22618AK 09/11/2006 0724794142 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22645AK 10/09/2006 0725013284 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 20407AK 07/18/2000 0700159975 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20409AK 07/18/2000 0700159975 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20414AK 07/24/2000 0700159976 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20415AK 07/24/2000 0700159976 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20417AK 07/24/2000 0700159976 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20418AK 07/24/2000 0700159976 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20419AK 07/25/2000 0700179062 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20435AK 08/07/2000 0700214053 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20438AK 08/09/2000 0700214054 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20474AK 09/11/2000 0700242470 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20478AK 09/21/2000 0700242471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20480AK 09/21/2000 0700242471 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20484AK 09/21/2000 0700242472 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20487AK 09/21/2000 0700242472 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20516AK 10/11/2000 0700308524 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20518AK 10/13/2000 0700242473 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20527AK 10/19/2000 0700242474 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20528AK 10/19/2000 0700242474 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20531AK 10/24/2000 0700242475 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20533AK 10/24/2000 0700242475 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20555AK 11/14/2000 0700242476 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20556AK 11/15/2000 0700488515 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20607AK 01/22/2001 0700242477 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20609AK 01/23/2001 0700668678 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20619AK 02/06/2001 0700311274 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20620AK 02/06/2001 0700242478 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20621AK 02/06/2001 0700311274 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20623AK 02/06/2001 0700242478 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20642AK 02/27/2001 0700399168 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20653AK 02/28/2001 0700311275 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20684AK 03/20/2001 0700311276 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20687AK 03/20/2001 0700764417 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20688AK 03/20/2001 0700311276 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20691AK 03/22/2001 0700311277 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20692AK 03/22/2001 0700311277 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20698AK 03/27/2001 0700311278 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 20702AK 03/27/2001 0700311278 ZOR 1025, 02H31-01 EA CD3700 SL(EA) 1.000 20366AK 06/15/2000 0700067139 ZOR 1025, 02H30-01 EA CD3700 CS 1.000 19954AN 01/19/2000 30001052430U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20201AK 01/19/2000 30001052492U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20202AK 01/19/2000 30001052475U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20222AK 02/07/2000 30001272996U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20227AK 02/10/2000 30001272997U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20228AK 02/10/2000 30001272998U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20230AK 02/15/2000 30002100624U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20232AK 02/15/2000 30002100624U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20237AK 02/18/2000 30001273000U3 SALE 1025, 02H31-01 EA CD3700 SL 1.000 20239AK 02/18/2000 30001272999U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20242AK 02/28/2000 30001273003U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20243AK 02/22/2000 30001273001U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20246AK 02/23/2000 30001273002U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20280AK 03/30/2000 30003070642U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20285AK 03/31/2000 30003070647U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20286AK 03/31/2000 30003070652U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20296AK 04/13/2000 30003070646U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20298AK 04/13/2000 30003070659U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20306AK 04/21/2000 30004112193U3 SALE 1025, 02H31-01 EA CD3700 SL(EA) 1.000 22669AK 01/24/2007 0725624482 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22711AK 01/24/2007 0725624482 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22725AK 01/24/2007 0725624482 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22746AK 03/08/2007 0725853549 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22747AK 03/19/2007 0725947869 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22748AK 03/08/2007 0725853549 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22750AK 03/19/2007 0725947869 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22758AK 03/19/2007 0725947869 ZOR 1025, 02H31-01 EA CD3700 SL 1.000 20239AK 02/18/2000 30001272999U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20242AK 02/28/2000 30001273003U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20243AK 02/22/2000 30001273001U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20246AK 02/23/2000 30001273002U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20280AK 03/30/2000 30003070642U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20285AK 03/31/2000 30003070647U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20286AK 03/31/2000 30003070652U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20296AK 04/13/2000 30003070646U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20298AK 04/13/2000 30003070659U3 SALE 1025 02H31-01 EA CD3700 SL 1.000 20306AK 04/21/2000 30004112193U3 SALE 1025, 02H31-01 EA CD3700 SL(EA) 1.000 22669AK 01/24/2007 0725624482 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22711AK 01/24/2007 0725624482 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22725AK 01/24/2007 0725624482 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22746AK 03/08/2007 0725853549 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22747AK 03/19/2007 0725947869 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22748AK 03/08/2007 0725853549 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22750AK 03/19/2007 0725947869 ZOR 1025 02H31-01 EA CD3700 SL(EA) 1.000 22758AK 03/19/2007 0725947869 ZOR 1025 
FEI Number 2919069
Recalling Firm/
Manufacturer
Abbott Laboratories
5440 Patrick Henry Dr
Santa Clara CA 95054-1113
For Additional Information Contact
408-982-4800
Manufacturer Reason
for Recall
The product failed electrostatic discharge testing, which could result in product failure.
FDA Determined
Cause 2
Other
ActionMandatory Technical Service Bulletins (TSB) are being issued to Field Service to upgrade all instruments in the field. All customers will be contacted by Field Service to schedule this mandatory upgrade. The upgrades are expected to take 12 months.
Quantity in Commerce3,429 units, all models
DistributionWorldwide, including USA, Puerto Rico, Canada, Mexico, Argentina, Australia, Bahamas, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Ecuador, Germany, Guatemala, Hong Kong, Jamaica, Japan, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan, Thailand, Uruguay, and Venezuela.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = GKZ
-
-