• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ISOLOC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ISOLOCsee related information
Date Initiated by FirmMay 31, 2005
Date PostedAugust 08, 2007
Recall Status1 Terminated 3 on April 19, 2011
Recall NumberZ-1147-2007
Recall Event ID 38358
510(K)NumberK904908 
Product Classification radiation treatment software - Product Code IYE
ProductISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software.
Code Information ISOLOC Version 4.2 Part Number NT-NW-425-520.
Recalling Firm/
Manufacturer
Northwest Medical Physics Equipment Inc
1910 120th Pl SE, Suite 101
Everett WA 98208
For Additional Information ContactAnderew D. Jones
425-338-1961
Manufacturer Reason
for Recall
When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display values ten times greater than the correct values.
FDA Determined
Cause 2
Software design
ActionSoftware to fix the described problems was written and customers contacted by e-mail on 05/31/2005, that a software patch, titled 05312005 was available on the firm's internet website.
Quantity in Commerce85
DistributionWorldwide, including USA, Korea, Spain, and UK.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = IYE
-
-