• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Actiwatch

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Actiwatchsee related information
Date Initiated by FirmJune 27, 2007
Date PostedJanuary 17, 2008
Recall Status1 Terminated 3 on October 20, 2010
Recall NumberZ-0324-2008
Recall Event ID 38368
510(K)NumberK983533 
Product Classification Patient Physiological Monitor - Product Code M HX
ProductActiwatch 16 Part Number 198-0300-00, Actiwatch 16 AM 198-0300-AM; Actiwatch 64 Part Number 198-0301-00; Actiwatch 64 AM Part Number 198-0301-AM, Mini Mitter, a Respironics Company, Bend, Oregon
Code Information Serial Numbers: V960641, V960642, V960655, V960656, V960657, V960658, V960660, V960661, V960986, V961113, V961114, V961115, V961116, V961117, V961118, V961119, V961120, V961177, V961178, V961179, V961180, V961181, V961182, V961183, V961184, V961185, V961186, V961187, V961188, V961189, V961191, V961192, V961193, V961194, V961195, V961196, V961197, V961198, V961199, V961200, V961513, V961514, V961515, V961516, V961517, V961518, V961519, V961520, V961593, V961594, V961595, V961596, V961597, V961598, V961599, V961600, V961601, V961602, V961603, V961604, V961605, V961606, V961607, V961608, V961609, V961610, V961611, V961612, V961613, V961614, V961615, V961616, V961617, V961121, V961122, V961123, V961124, V961125, V961126, V961127, V961128, V961321, V961322, V961323, V961324, V961325, V961326, V961327, V961328, V961329, V961330, V961331, V961332, V961333, V961334, V961335, V961336, V961337, V961338, V961339, V961340,V960611, V960617, V960618, V960619, V960620, V960621, V960622, V960629, V960630, V960631, V960632, V960635, V960638, V960639, V960640, V960643, V960644, V960645, V960663, V960669, V960671, V960673, V960678, V960681, V960682, V960684, V960685, V960686, V960689, V960690, V960692, V960693, V960694, V960695, V960696, V960697, V960698, V960701, V960702, V960709, V960710, V960720, V960721, V960722, V960723, V960724, V960725, V960726, V960728, V960729, V960730, V960731, V960732, V960736, V960737, V960742, V960743, V960744, V960745, V960746, V960747, V960748, V960749, V960751, V960752, V960753, V960754, V960755, V960757, V960758, V960760, V960761, V960834, V960835, V960836, V960837, V960838, V960839, V960840, V960841, V960842, V960843, V960844, V960845, V960846, V960847, V960848, V960849, V960850, V960851, V960854, V960855, V960856, V960857, V960858, V960859, V960860, V960861, V960862, V960863, V960871, V960872, V960875, V960876, V960877, V960878, V960879, V960882, V960889, V960891, V960892, V960893, V960894, V960895, V960896, V960897, V960898, V960900, V960902, V960903, V960904, V960913, V960914, V960915, V960916, V960917, V960918, V960919, V960920, V960921, V960922, V960923, V960924, V960925, V960928, V960929, V960930, V960931, V960932, V960933, V960934, V960935, V960936, V960937, V960938, V960939, V960940, V960941, V960942, V960943, V960944, V960945, V960946, V960947, V960948, V960949, V960950, V960951, V960952, V960953, V960954, V960955, V960956, V960957, V960958, V960959, V960960, V960961, V960962, V960963, V960964, V960965, V960966, V960967, V960968, V960969, V960970, V960971, V960972, V960973, V960974, V960975, V960976, V960977, V960978, V960979, V960980, V960981, V960982, V960983, V960984, V960993, V961009, V961016, V961017, V961018, V961021, V961023, V961024, V961025, V961026, V961027, V961028, V961029, V961030, V961032, V961034, V961035, V961036, V961038, V961039, V961040, V961042, V961043, V961044, V961045, V961047, V961048, V961049, V961050, V961051, V961052, V961053, V961054, V961056, V961057, V961058, V961059, V961060, V961064, V961066, V961067, V961068, V961130, V961134, V961144, V961145, V961146, V961147, V961148, V961149, V961150, V961151, V961152, V961201, V961202, V961203, V961204, V961205, V961206, V961207, V961208, V961209, V961210, V961211, V961212, V961213, V961214, V961215, V961216, V961260, V961278, V961283, V961320, V961348, V961351, V961352, V961353, V961354, V961355, V961356, V961357, V961358, V961360, V961367, V961368, V961369, V961370, V961371, V961372, V961373, V961374, V961375, V961376, V961377, V961378, V961379, V961381, V961382, V961383, V961388, V961389, V961390, V961391, V961392, V961393, V961394, V961396, V961397, V961398, V961399, V961400, V961401, V961402, V961403, V961404, V961405, V961406, V961407, V961408, V961409, V961410, V961411, V961412, V961413, V961414, V961415, V961416, V961417, V961418, V961419, V961420, V961421, V961422, V961423, V961424, V961425, V961426, V961427, V961428, V961429, V961430, V961431, V961432, V961433, V961434, V961435, V961436, V961437, V961438, V961439, V961440, V961442, V961444, V961445, V961446, V961447, V961448, V961449, V961450, V961451, V961452, V961453, V961454, V961455, V961456, V961457, V961458, V961459, V961460, V961461, V961462, V961463, V961464, V961465, V961466, V961467, V961468, V961469, V961470, V961471, V961472, V961473, V961474, V961475, V961476, V961477, V961478, V961479, V961480, V961481, V961485, V961486, V961487, V961488, V961510, V961511, V961512, V961521, V961522, V961523, V961524, V961525, V961526, V961529, V961530, V961531, V961532, V961534, V961535, V961536, V961538, V961539, V961540, V961542, V961543, V961544, V961545, V961546, V961547, V961548, V961549, V961550, V961551, V961552, V961553, V961554, V961555, V961556, V961561, V961591, V961592, V961625, V961626, V961627, V961628, V961629, V961630, V961631, V961632, V961633, V961634, V961635, V961636, V961637, V961638, V961639, V961640, V961641, V961642, V961643, V961644, V961645, V961647, V961648, V961649, V961650, V961652, V961653, V961654, V961655, V961656, V961689, V961690, V961691, V961692, V961693, V961694, V961695, V961696, V961697, V961698, V961699, V961700, V961701, V961702, V961703, V961704, V961705, V961706, V961707, V961708, V961713, V961714, V961715, V961716, V961717, V961718, V961727, V961728, V961729, V961730, V961731, V961732, V961733, V961735, V961736, V961737, V961738, V961739, V961740, V961741, V961742, V961896, V961897, V961900, V961901, V961902, V961903, V961904, V961905, V961906, V961907, V961908, V961909, V961910, V961569, V961570, V961571, V961572, V961573, V961574, V961575, V961576, V961752, V961753, V961754, V961755, V961756, V961757, V961758, V961759, V961760, V961761, V961762, V961763, V961764, V961765, V961766, V961768, V961769, V961770, V961771, and V961772.
FEI Number 3007223238
Recalling Firm/
Manufacturer
Mini-Mitter Company, Inc.
20300 Empire Ave Ste B3
Bend OR 97701
For Additional Information ContactFrederic Viaud
724-387-7919
Manufacturer Reason
for Recall
Communication Failure: the devices have a memory chip that has a limited number of read/write cycles which can trigger set up information and/or subject information anomalies or intermittent communication annomalies. If left unaddressed, the issues can lead to permanent communication failure.
FDA Determined
Cause 2
Software design
ActionA Voluntary Medical Device Recall letter dated June 26, 2007. The recall letter will instruct the customers to 1) fill out a business reply form to confirm contact information and devices serial numbers in their possession. Then FAX this info back immediately to Stericycle or Mini-Mitter. 2) Refer to the work scheduled rework date reserved for them and noted in their recall notification letter and return the device for service at that time; and 3) Package the device appropriately, insure it and return it via UPS using the provided UPS shipping label. Product returned to Mini-Mitter will be quarantined, decontaminated, and repaired. Min-Mitter will rework all the devices returned. Each will have a new memory chip installed, be re-calibrated, retested, and have a new battery installed if required. Min-Mitter will provide a new one year warranty.
Quantity in Commerce635
DistributionWorldwide

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = M HX
-
-