| Class 2 Device Recall NxStage | |
Date Initiated by Firm | July 27, 2007 |
Date Posted | September 20, 2007 |
Recall Status1 |
Terminated 3 on September 13, 2010 |
Recall Number | Z-1228-2007 |
Recall Event ID |
38546 |
510(K)Number | K050525 K061837 |
Product Classification |
hemodialysis cartridge - Product Code KDI
|
Product | NxStage Cartridge Express, CAR-170, for hemodialysis. |
Code Information |
Lot Numbers: 7037802, 703803, 703804, 7047702, 7047703, 7047704, 7047705, 7057702, 7057703, 7057704, 7057705, 7057706, 7057707, 7057708, 7057709, 7067701, 7067702,7067703, 7067705, 7067709, 7067710, 7067711, 7067712, 7067713, 7067714, 7067715, 7077703. |
Recalling Firm/ Manufacturer |
NxStage Medical, Inc. 439 South Union Street, 5th Floor Lawrence MA 01843-2800
|
For Additional Information Contact | Same 978-687-4700 |
Manufacturer Reason for Recall | Potential for dialysate leaks from the cartridge during treatment. |
FDA Determined Cause 2 | Process change control |
Action | NxStage initiated the recall by letter on 8/24/07 advising users of the specific lots to be removed from use. Prior notifications were issued on July 27, 2007, July 30, 2007 and July 31, 2007 notifying users of the potential for leaks and instructions to detect leaks prior to use. |
Quantity in Commerce | 12,085 cases |
Distribution | Nationwide |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = KDI
|
|
|
|