| | Class 2 Device Recall Zimmer Hemovac Wound Drainage Device |  |
| Date Initiated by Firm | January 29, 2008 |
| Date Posted | September 19, 2008 |
| Recall Status1 |
Terminated 3 on November 12, 2009 |
| Recall Number | Z-1498-2008 |
| Recall Event ID |
47233 |
| Product Classification |
non-powered, portable suction appartus for a single patient - Product Code GCY
|
| Product | Zimmer Hemovac Wound Drainage Device, 400 ml COMP EVAC 1/4" PVC w/trocar (4"), 6,4mm, 19Fr, PVC, 107cm, large, sterile, latex free, Rx only; Zimmer, Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio; REF 00-2590-000-10. |
| Code Information |
All lots manufactured on or before 1/18/08. Product has a 5 year expiration period. |
| FEI Number |
1000220733
|
Recalling Firm/ Manufacturer |
Zimmer Inc.
345 E Main St
Warsaw
IN
46580-2746
|
| For Additional Information Contact | 330-365-3793 |
Manufacturer Reason for Recall | The sterility of the device may be compromised due to the possible presence of slits or pinholes in the packaging. |
FDA Determined Cause 2 | Packaging process control |
| Action | Consignees were notified via recall letter dated 1/29/08 to locate and return the affected product. The recall action was expanded to additional products/lots via recall letter dated 3/17/08. Custom kit packers were instructed to notify their customers. |
| Distribution | Nationwide, Australia, Canada, Chile, China, Czech Republic, Dubai, Ecuador, France, Germany, Greece, Hungary, Italy, Japan, Korea, Poland, Romania, Russia, Saudi Arabia, Serbia & Montenegro, Singapore, Slovakia, South Africa, Spain, and Switzerland. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|