| Class 2 Device Recall Powerheart G3 |  |
Date Initiated by Firm | October 13, 2008 |
Date Posted | February 25, 2009 |
Recall Status1 |
Terminated 3 on December 08, 2011 |
Recall Number | Z-0919-2009 |
Recall Event ID |
50064 |
510(K)Number | K031987 |
Product Classification |
Automated External Defibrillator - Product Code MKJ
|
Product | Cardiac Science Powerheart Automated External Defibrillator G3 9390A |
Code Information |
Serial numbers: 4038448, 4038449, 4038450, 4038451, 4038452, 4040238, 4040239, 4040240, 4040241, 4040242, 4040243, 4040244, 4040245, 4040246, 4040247, 4040248, 4040249, 4040250, 4040251, 4040252, 4040253, 4040254, 4040255, 4040256, 4040257, 4040258, 4040259, 4040260, 4040261, 4040262, 4040263, 4040264, 4040265, 4040266, 4040267, 4040268, 4040269, 4040270, 4040271, 4040272, 4040273, 4040274, 4040275, 4040276, 4040277, 4040278, 4040279, 4040280, 4040281, 4040282, 4040283, 4040284, 4040285, 4040286, 4040287, 4040288, 4040289, 4040290, 4040291, 4040292, 4040293, 4040294, 4040295, 4040296, 4040297, 4040298, 4040299, 4040300, 4040301, 4040302, 4040303, 4040304, 4040305, 4040306, 4040307, 4040308, 4040309, 4040310, 4040311, 4040312, 4040313, 4040314, 4040315, 4040316, 4040317, 4040318, 4040319, 4040320, 4040321, 4040322, 4040323, 4040324, 4040325, 4040326, 4040327, 4040328, 4040329, 4040330, 4040331, 4040332, 4040333, 4040334, 4040335, 4040336, 4040337, 4040338, 4040339, 4040340, 4040341, 4040342, 4040343, 4040344, 4040345, 4040346, 4040347, 4040348, 4040349, 4040350, 4040351, 4040352, 4040353, 4040354, 4040355, 4040356, 4040357, 4040358, 4040359, 4040360, 4040361, 4040362, 4040363, 4040364, 4040365, 4040366, 4040367, 4040368, 4040369, 4040370, 4040371, 4040372, 4040373, 4040374, 4040375, 4040376, 4040377, 4040378, 4040379, 4040380, 4040381, 4040382, 4040383, 4040384, 4040385, 4040386, 4040387, 4040388, 4040389, 4040390, 4040391, 4040392, 4040393, 4040394, 4040395, 4040396, 4040397, 4040398, 4040399, 4040400, 4040401, 4040402, 4040403, 4040404, 4040405, 4040406, 4040407, 4040408, 4040409, 4040410, 4040411, 4040412, 4040413, 4040414, 4040415, 4040416, 4040417, 4040418, 4040419, 4040420, 4040421, 4040422, 4040423, 4040424, 4040425, 4040426, 4040427, 4040428, 4040429, 4040430, 4040431, 4040432, 4040433, 4040434, 4040435, 4040436, 4040437, 4040438, 4040439, 4040440, 4040441, 4040442, 4040443, 4040444, 4040445, 4040446, 4040447, 4040448, 4040449, 4040450, 4040451, 4040452, 4040453, 4040454, 4040455, 4040456, 4040457, 4040458, 4040459, 4040460, 4040461, 4040462, 4040463, 4040464, 4040465, 4040466, 4040467, 4040468, 4040469, 4040470, 4040471, 4040472, 4040473, 4040474, 4040475, 4040476, 4040477, 4040478, 4040479, 4040480, 4040481, 4040482, 4040483, 4040484, 4040485, 4040486, 4040487, 4040488, 4040489, 4040490, 4040491, 4040492, 4040493, 4040494, 4040495, 4040496, 4040497, 4040498, 4040499, 4040500, 4040501, 4040502, 4040503, 4040504, 4040505, 4040506, 4040507, 4040508, 4040509, 4040510, 4040511, 4040512, 4040513, 4040514, 4040515, 4040516, 4040517, 4040518, 4040519, 4040520, 4040521, 4040522, 4040523, 4040524, 4040525, 4040526, 4040527, 4040528, 4040529, 4040530, 4040531, 4040532, 4040533, 4040534, 4040535, 4040536, 4040537, 4040538, 4040539, 4040540, 4040541, 4040542, 4040543, 4040544, 4040545, 4040546, 4040547, 4040548, 4040549, 4040550, 4040551, 4040552, 4040553, 4040554, 4040555, 4040556, 4040557, 4040558, 4040559, 4040560, 4040561, 4040562, 4040563, 4040564, 4040565, 4040566, 4040567, 4040568, 4040569, 4040570, 4040571, 4040572, 4040573, 4040574, 4040575, 4040576, 4040577, 4040578, 4040579, 4040580, 4040581, 4040582, 4040583, 4040584, 4040585, 4040586, 4040587, 4040588, 4040589, 4040590, 4040591, 4040592, 4040593, 4040594, 4040595, 4040596, 4040597, 4040598, 4040599, 4040600, 4040601, 4040602, 4040603, 4040604, 4040605, 4040606, 4040607, 4040608, 4040609, 4040610, 4040611, 4040612, 4040613, 4040614, 4040615, 4040616, 4040617, 4040618, 4040619, 4040620, 4040621, 4040622, 4040623, 4040624, 4040625, 4040626, 4040627, 4040628, 4040629, 4040630, 4040631, 4040632, 4040633, 4040634, 4040635, 4040636, 4040637, 4040638, 4040639, 4040640, 4040641, 4040642, 4040643, 4040644, 4040645, 4040646, 4040647, 4040648, 4040649, 4040650, 4040651, 4040652, 4040653, 4040654, 4040655, 4040656, 4040657, 4040658, 4040659, 4040660, 4040661, 4040662, 4040663, 4040664, 4040665, 4040666, 4040667, 4040668, 4040669, 4040670, 4040671, 4040672, 4040673, 4040674, 4040675, 4040676, 4040677, 4040678, 4040679, 4040680, 4040681, 4040682, 4040683, 4040684, 4040685, 4040686, 4040687, 4040688, 4040689, 4040690, 4040691, 4040692, 4040693, 4040694, 4040695, 4040696, 4040697, 4040698, 4040699, 4040700, 4040701, 4040702, 4040703, 4040704, 4040705, 4040706, 4040707, 4040708, 4040709, 4040710, 4040711, 4040712, 4040713, 4040714, 4040715, 4040716, 4040717, 4040718, 4040719, 4040720, 4040721, 4040722, 4040723, 4040724, 4040725, 4040726, 4040727, 4040728, 4040729, 4040730, 4040731, 4040732, 4040733, 4040734, 4040735, 4040736, 4040737, 4040738, 4040739, 4040740, 4040741, 4040742, 4040743, 4040744, 4040745, 4040746, 4040747, 4040748, 4040749, 4040750, 4040751, 4040752, 4040753, 4040754, 4040755, 4040756, 4040757, 4040758, 4040759, 4040760, 4040761, 4040762, 4040763, 4040764, 4040765, 4040766, 4040767, 4040768, 4040769, 4040770, 4040771, 4040772, 4040773, 4040774, 4040775, 4040776, 4040777, 4040778, 4040779, 4040780, 4040781, 4040782, 4040783, 4040784, 4040785, 4040786, 4040787, 4040789, 4040790, 4040791, 4040792, 4040793, 4040794, 4040795, 4040796, 4040797, 4040798, 4040799, 4040800, 4040801, 4040803, 4040804, 4040805, 4040806, 4040807, 4040808, 4040809, 4040810, 4040811, 4040812, 4040813, 4040814, 4040815, 4040816, 4040817, 4040818, 4040819, 4040820, 4040821, 4040824, 4040825, 4040826, 4040827, 4040828, 4040829, 4040830, 4040831, 4040832, 4040834, 4040835, 4040837, 4040838, 4040839, 4040840, 4040841, 4040842, 4040845, 4040847, 4040849, 4040851, 4040852, 4040853, 4040854, 4040855, 4040856, 4040858, 4040860, 4040863, 4040864, 4040865, 4040866, 4040868, 4040871, 4040872, 4040873, 4040874, 4040875, 4040877, 4040880, 4040881, 4040883, 4040884, 4040886, 4040887, 4040888, 4040889, 4040893, 4040894, 4040895, 4040898, 4040900, 4040901, 4040903, 4040905, 4040907, 4040908, 4040909, 4040910, 4040911, 4040912, 4040913, 4040914, 4040916, 4040918, 4040922, 4040924, 4040925, 4040927, 4040928, 4040929, 4040935, 4040936, 4040937, 4044789, 4044790, 4044791, 4044792, 4044793, 4044794, 4044795, 4044796, 4044797, 4044798, 4044799, 4044800, 4044802, 4044805, 4044806, 4044807, 4044808, 4044809, 4044810, 4044811, 4044812, 4044813, 4044814, 4044815, 4044816, 4044817, 4044818, 4044820, 4044821, 4044822, 4044823, 4044824, 4044826, 4044827, 4044828, 4044829, 4044830, 4044831, 4044832, 4044833, 4044834, 4044835, 4044836, 4044837, 4044838, 4044839, 4044840, 4044841, 4044842, 4044843, 4044844, 4044845, 4044846, 4044847, 4044848, 4044849, 4044850, 4044851, 4044852, 4044853, 4044855, 4044856, 4044857, 4044858, 4044859, 4044860, 4044861, 4044862, 4044863, 4044864, 4044865, 4044866, 4044867, 4044868, 4044869, 4044870, 4044871, 4044872, 4044873, 4044874, 4044875, 4044876, 4044877, 4044878, 4044879, 4044880, 4044881, 4044882, 4044883, 4044884, 4044885, 4044886, 4044887, 4044888, 4046989, 4046990, 4046991, 4046992, 4046993, 4046994, 4046995, 4046996, 4046997, 4046998, 4046999, 4047000, 4047001, 4047002, 4047003, 4047004, 4047005, 4047006, 4047007, 4047008, 4047009, 4047010, 4047011, 4047012, 4047013, 4047014, 4047015, 4047016, 4047017, 4047018, 4047019, 4047020, 4047021, 4047022, 4047023, 4047024, 4047025, 4047026, 4047027, 4047028, 4047029, 4047030, 4047031, 4047032, 4047033, 4047034, 4047035, 4047036, 4047037, 4047038, 4047039, 4047040, 4047041, 4047042, 4047043, 4047044, 4047045, 4047046, 4047047, 4047048, 4047049, 4047050, 4047051, 4047052, 4047053, 4047054, 4047055, 4047056, 4047057, 4047058, 4047059, 4047060, 4047061, 4047062, 4047063, 4047064, 4047065, 4047066, 4047067, 4047068, 4047069, 4047070, 4047071, 4047072, 4047073, 4047074, 4047075, 4047076, 4047077, 4047078, 4047079, 4047080, 4047081, 4047082, 4047083, 4047084, 4047085, 4047086, 4047087, 4047088, 4047089, 4047090, 4047091, 4047092, 4047093, 4047094, 4047095, 4047096, 4047097, 4047098, 4047099, 4047100, 4047101, 4047102, 4047103, 4047104, 4047105, 4047106, 4047107, 4047108, 4047109, 4047110, 4047111, 4047112, 4047113, 4047114, 4047115, 4047116, 4047117, 4047118, 4047119, 4047120, 4047121, 4047122, 4047123, 4047124, 4047125, 4047126, 4047127, 4047128, 4047129, 4047130, 4047131, 4047132, 4047133, 4047134, 4047135, 4047136, 4047137, 4047138, 4047139, 4047140, 4047141, 4047142, 4047143, 4047144, 4047145, 4047146, 4047147, 4047148, 4047149, 4047150, 4047151, 4047152, 4047153, 4047154, 4047155, 4047156, 4047157, 4047158, 4047159, 4047160, 4047161, 4047162, 4047163, 4047164, 4047165, 4047166, 4047167, 4047168, 4047169, 4047170, 4047171, 4047172, 4047173, 4047174, 4047175, 4047176, 4047177, 4047178, 4047179, 4047180, 4047181, 4047182, 4047183, 4047184, 4047185, 4047186, 4047187, 4047188, 4047189, 4047190, 4047191, 4047192, 4047193, 4047194, 4047195, 4047196, 4047197, 4047198, 4047199, 4047200, 4047201, 4047202, 4047203, 4047204, 4047205, 4047206, 4047207, 4047208, 4047209, 4047210, 4047211, 4047212, 4047213, 4047214, 4047215, 4047216, 4047217, 4047218, 4047219, 4047220, 4047221, 4047222, 4047223, 4047224, 4047225, 4047226, 4047227, 4047228, 4047229, 4047230, 4047231, 4047232, 4047233, 4047234, 4047235, 4047236, 4047237, 4047238, 4047239, 4047240, 4047241, 4047242, 4047243, 4047244, 4047245, 4047246, 4047247, 4047248, 4047249, 4047250, 4047251, 4047252, 4047253, 4047254, 4047255, 4047256, 4047257, 4047258, 4047259, 4047260, 4047261, 4047262, 4047263, 4047264, 4047265, 4047266, 4047267, 4047268, 4047269, 4047270, 4047271, 4047272, 4047273, 4047274, 4047275, 4047276, 4047277, 4047278, 4047279, 4047280, 4047281, 4047282, 4047283, 4047284, 4047285, 4047286, 4047287, 4047288, 4047289, 4047290, 4047291, 4047292, 4047293, 4047294, 4047295, 4047296, 4047297, 4047298, 4047299, 4047300, 4047301, 4047302, 4047303, 4047304, 4047305, 4047306, 4047307, 4047308, 4047309, 4047310, 4047311, 4047312, 4047313, 4047314, 4047315, 4047316, 4047317, 4047318, 4047319, 4047320, 4047321, 4047322, 4047323, 4047324, 4047325, 4047326, 4047327, 4047328, 4047329, 4047330, 4047331, 4047332, 4047333, 4047334, 4047335, 4047336, 4047337, 4047338, 4047339, 4047340, 4047341, 4047342, 4047343, 4047344, 4047345, 4047346, 4047347, 4047348, 4047349, 4047350, 4047351, 4047352, 4047353, 4047354, 4047355, 4047356, 4047357, 4047358, 4047359, 4047360, 4047361, 4047362, 4047363, 4047364, 4047365, 4047366, 4047367, 4047368, 4047369, 4047370, 4047371, 4047372, 4047373, 4047374, 4047375, 4047376, 4047377, 4047378, 4047379, 4047380, 4047381, 4047382, 4047383, 4047384, 4047385, 4047386, 4047387, 4047388, 4047389, 4047390, 4047391, 4047392, 4047394, 4047395, 4047396, 4047397, 4047398, 4047399, 4047400, 4047401, 4047403, 4047404, 4047405, 4047406, 4047407, 4047408, 4047409, 4047410, 4047411, 4047412, 4047413, 4047414, 4047415, 4047416, 4047417, 4047418, 4047419, 4047420, 4047421, 4047422, 4047423, 4047424, 4047425, 4047426, 4047427, 4047428, 4047429, 4047430, 4047431, 4047432, 4047433, 4047434, 4047436, 4047437, 4047438, 4047439, 4047440, 4047441, 4047442, 4047443, 4047444, 4047445, 4047446, 4047447, 4047448, 4047449, 4047450, 4047451, 4047452, 4047455, 4047456, 4047457, 4047458, 4047459, 4047460, 4047461, 4047462, 4047463, 4047464, 4047465, 4047466, 4047467, 4047468, 4047469, 4047471, 4047472, 4047473, 4047474, 4047475, 4047476, 4047477, 4047478, 4047479, 4047480, 4047481, 4047482, 4047483, 4047484, 4047485, 4047486, 4047487, 4047488, 4104689, 4104690, 4104691, 4104692, 4104693, 4104694, 4104695, 4104696, 4104697, 4104698, 4104699, 4104700, 4104701, 4104702, 4104703, 4104704, 4104705, 4104706, 4104707, 4104708, 4104709, 4104710, 4104711, 4104712, 4104713, 4104714, 4104715, 4104716, 4104717, 4104718, 4104719, 4104720, 4104721, 4104722, 4104723, 4104724, 4104725, 4104726, 4104727, 4104728, 4104729, 4104730, 4104731, 4104732, 4104733, 4104734, 4104735, 4104736, 4104737, 4104738, 4104739, 4104740, 4104741, 4104742, 4104743, 4104744, 4104745, 4104746, 4104747, 4104748, 4104749, 4104750, 4104751, 4104752, 4104753, 4104754, 4104755, 4104756, 4104757, 4104758, 4104759, 4104760, 4104761, 4104762, 4104763, 4104764, 4104765, 4104766, 4104767, 4104768, 4104769, 4104770, 4104771, 4104772, 4104773, 4104774, 4104775, 4104776, 4104777, 4104778, 4104779, 4104780, 4104781, 4104782, 4104783, 4104784, 4104785, 4104786, 4104787, 4104788, 4104789, 4104790, 4104791, 4104792, 4104793, 4104794, 4104795, 4104796, 4104797, 4104798, 4104799, 4104800, 4104801, 4104802, 4104803, 4104804, 4104805, 4104806, 4104807, 4104808, 4104809, 4104810, 4104811, 4104812, 4104813, 4104814, 4104815, 4104816, 4104817, 4104818, 4104819, 4104821, 4104822, 4104823, 4104824, 4104825, 4104826, 4104827, 4104828, 4104829, 4104830, 4104831, 4104832, 4104833, 4104834, 4104835, 4104836, 4104837, 4104838, 4104839, 4104840, 4104841, 4104842, 4104843, 4104844, 4104845, 4104846, 4104847, 4104848, 4104849, 4104850, 4104851, 4104852, 4104853, 4104854, 4104855, 4104856, 4104857, 4104858, 4104859, 4104860, 4104861, 4104862, 4104863, 4104864, 4104865, 4104866, 4104867, 4104868, 4104869, 4104870, 4104871, 4104872, 4104873, 4104874, 4104875, 4104876, 4104877, 4104878, 4104879, 4104880, 4104881, 4104882, 4104883, 4104884, 4104885, 4104886, 4104887, 4104888, 4106005, 4106006, 4106007, 4106008, 4106009, 4106010, 4106011, 4106012, 4106013, 4106014, 4106015, 4106016, 4106017, 4106018, 4106019, 4106020, 4106021, 4106022, 4106023, 4106024, 4106025, 4106026, 4106027, 4106028, 4106029, 4106030, 4106031, 4106032, 4106033, 4106034, 4106035, 4106036, 4106037, 4106038, 4106039, 4106040, 4106041, 4106042, 4106043, 4106044, 4106045, 4106046, 4106047, 4106048, 4106049, 4106050, 4106051, 4106052, 4106053, 4106054, 4106055, 4106056, 4106057, 4106058, 4106059, 4106060, 4106061, 4106062, 4106063, 4106064, 4106065, 4106066, 4106067, 4106068, 4106069, 4106070, 4106071, 4106072, 4106073, 4106074, 4106075, 4106077, 4106078, 4106079, 4106080, 4106081, 4106082, 4106083, 4106084, 4106085, 4106086, 4106087, 4106088, 4106089, 4106090, 4106091, 4106092, 4106093, 4106094, 4106095, 4106096, 4106097, 4106098, 4106099, 4106100, 4106101, 4106102, 4106103, 4106104, 4106105, 4106106, 4106107, 4106108, 4106109, 4106110, 4106111, 4106112, 4106113, 4106114, 4106115, 4106116, 4106117, 4106118, 4106119, 4106120, 4106121, 4106122, 4106123, 4106124, 4106125, 4106126, 4106127, 4106128, 4106129, 4106130, 4106131, 4106132, 4106133, 4106134, 4106135, 4106136, 4106137, 4106138, 4106139, 4106140, 4106141, 4106142, 4106143, 4106144, 4106145, 4106146, 4106147, 4106148, 4106149, 4106150, 4106151, 4106152, 4106153, 4106154, 4106155, 4106156, 4106157, 4106158, 4106159, 4106160, 4106161, 4106162, 4106163, 4106164, 4106165, 4106166, 4106167, 4106168, 4106169, 4106170, 4106171, 4106172, 4106173, 4106174, 4106175, 4106176, 4106177, 4106178, 4106179, 4106180, 4106181, 4106182, 4106183, 4106184, 4106185, 4106186, 4106187, 4106188, 4106189, 4106190, 4106191, 4106192, 4106193, 4106194, 4106195, 4106196, 4106197, 4106198, 4106199, 4106200, 4106201, 4106202, 4106203, 4106204, 4106205, 4106206, 4106207, 4106208, 4106209, 4106210, 4106211, 4106212, 4106213, 4106214, 4106215, 4106216, 4106217, 4106218, 4106219, 4106220, 4106221, 4106222, 4106223, 4106224, 4106225, 4106226, 4106227, 4106228, 4106229, 4106230, 4106231, 4106232, 4106233, 4106234, 4106235, 4106236, 4106237, 4106238, 4106239, 4106240, 4106241, 4106242, 4106243, 4106245, 4106246, 4106247, 4106248, 4106249, 4106250, 4106251, 4106252, 4106253, 4106254, 4106255, 4106256, 4106257, 4106258, 4106259, 4106260, 4106261, 4106262, 4106263, 4106264, 4106265, 4106266, 4106268, 4106269, 4106270, 4106271, 4106272, 4106273, 4106274, 4106275, 4106276, 4106277, 4106278, 4106279, 4106280, 4106281, 4106282, 4106283, 4106284, 4106285, 4106286, 4106287, 4106288, 4106289, 4106290, 4106291, 4106292, 4106293, 4106294, 4106295, 4106296, 4106297, 4106298, 4106299, 4106300, 4106301, 4106302, 4106303, 4106304, 4110311, 4110312, 4110313, 4110314, 4110315, 4110316, 4110317, 4110318, 4110319, 4110320, 4110496, 4110497, 4110498, 4110499, 4110500, 4110501, 4110502, 4110503, 4110504, 4110505, 4110506, 4110507, 4110508, 4110509, 4110510, 4110511, 4110512, 4110513, 4110514, 4110515, 4110516.4110517, 4110518, 4110519, 4110520, 4110521, 4110522, 4110523, 4110524, 4110525, 4110526, 4110527, 4110528, 4110529, 4110530, 4110531, 4110532, 4110533, 4110534, 4110535, 4110536, 4110537, 4110538, 4110539, 4110540, 4110541, 4110542, 4110543, 4110544, 4110545, 4110546, 4110547, 4110548, 4110549, 4110550, 4110551, 4110552, 4110553, 4110554, 4110555, 4110556, 4110557, 4110558, 4110559, 4110560, 4110561, 4110562, 4110563, 4110564, 4110565, 4110566, 4110567, 4110568, 4110569, 4110570, 4110571, 4110572, 4110573, 4110574, 4110575, 4110576, 4110577, 4110578, 4110579, 4110580, 4110581, 4110582, 4110583, 4110584, 4110585, 4110586, 4110587, 4110588, 4110589, 4110590, 4110591, 4110592, 4110593, 4110594, 4110595, 4110596, 4110597, 4110598, 4110599, 4110600, 4110601, 4110602, 4110603, 4110604, 4110605, 4110606, 4110607, 4110608, 4110609, 4110620, 4110622, 4110623, 4110624, 4110625, 4110626, 4110627, 4110628, 4110629, 4110630, 4110631, 4110636, 4110642, 4110643, 4110644, 4110645, 4110646, 4110647, 4110648, 4110649, 4110650, 4110651, 4110652, 4110653, 4110654, 4110655, 4110656, 4110657, 4110658, 4110659, 4110660, 4110661, 4110662, 4110663, 4110664, 4110665, 4110666, 4110667, 4110668, 4110669, 4110670, 4110671, 4110672, 4110673, 4110674, 4110675, 4110676, 4110677, 4110678, 4110679, 4110680, 4110681, 4110682, 4110683, 4110684, 4110685, 4110686, 4110687, 4110688, 4110689, 4110690, 4110691, 4110692, 4110693, 4110694, 4110695, 4110696, 4110697, 4110698, 4110699, 4110700, 4110701, 4110702, 4110703, 4110704, 4110705, 4110706, 4110707, 4110708, 4110709, 4110710, 4110711, 4110712, 4110713, 4110714, 4110715, 4110716, 4110717, 4110718, 4110719, 4110720, 4110721, 4110722, 4110723, 4110724, 4110725, 4110726, 4110727, 4110728, 4110729, 4110730, 4110731, 4110732, 4110733, 4110734, 4110735, 4110736, 4110737, 4110738, 4110739, 4110740, 4110741, 4110742, 4110743, 4110744, 4110745, 4110746, 4110747, 4110748, 4110749, 4110750, 4110751, 4110752, 4110753, 4110754, 4110755, 4110756, 4110757, 4110758, 4110759, 4110760, 4110761, 4110762, 4110763, 4110764, 4110765, 4110766, 4110767, 4110768, 4110769, 4110770, 4110771, 4110773, 4110774, 4110775, 4110776, 4110777, 4110778, 4110779, 4110780, 4110781, 4110782, 4110783, 4110784, 4110785, 4110786, 4110787, 4110788, 4110789, 4110790, 4110791, 4110792, 4110793, 4110794, 4110795, 4110796, 4110797, 4110798, 4110799, 4110800, 4110801, 4110802, 4110803, 4110804, 4110805, 4110806, 4110807, 4110808, 4110809, 4110810, 4110811, 4110812, 4110813, 4110902, 4110903, 4110910, 4111085, 4111086, 4111087, 4111088, 4111089, 4111090, 4111091, 4111092, 4111093, 4111095, 4111096, 4111097, 4111098, 4111099, 4111100, 4111101, 4111102, 4111105, 4111106, 4111107, 4111108, 4111109, 4111110, 4111123, 4111124, 4111125, 4111126, 4111127, 4111128, 4111129, 4111130, 4111131, 4111132, 4111133, 4111134, 4111405, 4111412, 4111413, 4111414, 4111415, 4111416, 4111417, 4111418, 4111419, 4111420, 4111429, 4111430, 4111431, 4111432, 4111433, 4111434, 4111435, 4111436, 4111437, 4111438, 4111439, 4111440, 4111441, 4111442, 4111443, 4111444, 4111445, 4111446, 4111447, 4111448, 4111449, 4111450, 4111451, 4111452, 4111453, 4111454, 4111630, 4111632, 4111633, 4111634, 4111635, 4111636, 4111637, 4111638, 4111639, 4111640, 4111641, 4111642, 4111643, 4111644, 4111645, 4111646, 4111647, 4111648, 4111649, 4111650, 4111651, 4111652, 4111653, 4111654, 4111655, 4111656, 4111657, 4111658, 4111659, 4111660, 4111661, 4111662, 4111663, 4111664, 4111665, 4111666, 4111667, 4111668, 4111669, 4111670, 4111671, 4111672, 4111673, 4111674, 4111675, 4111676, 4111677, 4111678, 4111679, 4111680, 4111681, 4111746, 4111747, 4111748, 4111749, 4111750, 4111751, 4111752, 4111753, 4111754, 4111755, 4111756, 4111757, 4111758, 4111759, 4111760, 4111761, 4111762, 4111763, 4111764, 4111765, 4111766, 4111767, 4111770, 4111771, 4111772, 4111773, 4111774, 4111775, 4111776, 4111777, 4111778, 4111779, 4111780, 4111781, 4111782, 4111783, 4111784, 4111785, 4111786, 4111787, 4111788, 4111789, 4111790, 4111791, 4111792, 4111793, 4111794, 4111795, 4111796, 4111797, 4111798, 4111799, 4111800, 4111801, 4111802, 4111803, 4111804, 4111805, 4111806, 4111825, 4111826, 4111827, 4111828, 4111829, 4111830, 4111831, 4111832, 4111833, 4111834, 4111835, 4111836, 4111837, 4111838, 4111839, 4111840, 4111841, 4111842, 4111843, 4111852, 4111862, 4111863, 4111864, 4111865, 4111866, 4111867, 4111868, 4111869, 4111870, 4111871, 4111872, 4111873, 4111874, 4111875, 4111876, 4111877, 4111878, 4111879, 4111880, 4111881, 4111882, 4111883, 4111884, 4111885, 4111886, 4111887, 4111888, 4111889, 4111890, 4111891, 4111892, 4111893, 4111894, 4111895, 4111896, 4111897, 4111898, 4111899, 4111900, 4111901, 4111902, 4111903, 4111904, 4111905, 4111906, 4111907, 4111908, 4111909, 4111910, 4111911, 4111912, 4111913, 4111914, 4111915, 4111916, 4111917, 4111918, 4111919, 4111920, 4111921, 4111922, 4111923, 4111924, 4111925, 4111926, 4111927, 4111928, 4111929, 4111930, 4111931, 4111932, 4111933, 4111934, 4111935, 4111936, 4111937, 4111938, 4111939, 4111940, 4111941, 4111942, 4111943, 4111944, 4111945, 4111946, 4111947, 4111948, 4111949, 4111950, 4111951, 4111952, 4111953, 4111954, 4111955, 4111956, 4111957, 4111958, 4111959, 4111960, 4111961, 4111962, 4111963, 4111964, 4111965, 4111966, 4111967, 4111968, 4111969, 4111970, 4111971, 4111972, 4111973, 4111974, 4111975, 4111976, 4111978, 4111979, 4111980, 4111981, 4111982, 4111983, 4111984, 4111985, 4111986, 4111987, 4111988, 4111989, 4111990, 4111991, 4111992, 4111993, 4111995, 4111996, 4111997, 4111998, 4112083, 4112084, 4112085, 4112086, 4112087, 4112088, 4112089, 4112100, 4112102, 4112103, 4112104, 4112105, 4112106, 4112107, 4112109, 4112110, 4112111, 4112113, 4112114, 4112115, 4112116, 4112117, 4112118, 4112119, 4112121, 4112122, 4112129, 4112130, 4112131, 4112132, 4112133, 4112134, 4112136, 4112137, 4112138, 4112142, 4112144, 4112145, 4112146, 4112147, 4112148, 4112149, 4112151, 4112153, 4112154, 4112156, 4112157, 4112158, 4112161, 4112163, 4112164, 4112165, 4112167, 4112168, 4112170, 4112171, 4112245, 4112246, 4112247, 4112250, 4112251, 4112252, 4112253, 4112255, 4112256, 4112257, 4112258, 4112259, 4112260, 4112261, 4112263, 4112264, 4112265, 4112266, 4112267, 4112268, 4112269, 4112270, 4112271, 4112272, 4112273, 4112275, 4112276, 4112277, 4112278, 4112279, 4112281, 4112282, 4112284, 4112285, 4112286, 4112287, 4112288, 4112289, 4112292, 4112293, 4112294, 4112295, 4112296, 4112299, 4112304, 4112308, 4112309, 4112310, 4112315, 4112318, 4112320, 4112321, 4112322, 4112324, 4112325, 4112326, 4112327, 4112328, 4112329, 4112330, 4112331, 4112332, 4112333, 4112334, 4112335, 4112336, 4112337, 4112339, 4112341, 4112342, 4112343, 4112344, 4112345, 4112346, 4112347, 4112349, 4112350, 4112351, 4112352, 4112353, 4112355, 4112356, 4112357, 4112358, 4112359, 4112360, 4112361, 4112362, 4112363, 4112364, 4112366, 4112368, 4112369, 4112370, 4112371, 4112372, 4112375, 4112377, 4114063, 4114064, 4114392, 4114393, 4114394, 4114395, 4114396, 4114397, 4114398, 4114399, 4114400, 4114401, 4114404, 4114405, 4114406, 4114408, 4114409, 4114410, 4114411, 4114412, 4114413, 4114414, 4114415, 4114416, 4114417, 4114418, and 4114419. |
Recalling Firm/ Manufacturer |
Cardiac Science Corporation 3303 Monte Villa Pkwy Bothell WA 98021-8969
|
For Additional Information Contact | Bev Magrane 425-402-2000 |
Manufacturer Reason for Recall | Potential for AED to not deliver defibrillation therapy. |
FDA Determined Cause 2 | Software design |
Action | On October 13, 2008 Cardiac Science began sending via Fed-Ex letters and software upgrade kits. Mailing of letters and software upgrade kits is anticipated to be completed by November 21, 2008. The letter instructs consignees to install the software update, assure that the AED information test file generated by the update software is returned to the firm, and complete the postage paid response card and return it to the firm. If the affected devices are no longer in the possession of the consignee, the consignee is to contact the firm and forward the update kit and letter to the person who has the AED in their possession. A separate letter is being sent to distributors. The letter advised to install the software upgrades for any units in stock or to forward the software upgrade along with the end user letter enclosed and that the AED information text file generated by the update software must be returned to the firm. The firm will visit the locations of individual national account customers and install the update. OEM partners will be notified and provided the information and update kits to provide to their customers. |
Quantity in Commerce | 2,572 devices |
Distribution | Worldwide Distribution: USA, Puerto Rico, Virgin Islands, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czechoslovakia, Denmark, Dominican Republic, France, Germany, Greece, Guam, Hong Kong, Hungary, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Kuwait, Libya, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, UK, and Ukraine. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = MKJ
|
|
|
|