Class 1 Device Recall Medtronic SIGMA 100/200/300 Series IPGs

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Medtronic SIGMA 100/200/300 Series IPGssee related information
Date Initiated by FirmMay 18, 2009
Date PostedJune 11, 2009
Recall Status1 Terminated 3 on March 28, 2012
Recall NumberZ-1517-2009
Recall Event ID 52076
PMA NumberP980035S002 P890003S065 
Product Classification Implantable Pacemakere Pulse-Generator - Product Code DXY
ProductMedtronic SIGMA 100/200/300 Series IPGs, Dual Chamber atrial sensing, ventricular sensing and pacing Pacemaker, model SVDD303.
Code Information Model SVDD303, serial numbers: PJD183402H, PJD183414H, PJD183415H, PJD183422H, PJD183425H, PJD183426H, PJD183430H, PJD183437H, PJD183440H, PJD183443H, PJD183447H, PJD186130H, PJD186131H, PJD186132H, PJD186133H, PJD186134H, PJD186137H, PJD186138H, PJD186140H, PJD186141H, PJD186142H, PJD186143H, PJD186145H, PJD186147H, PJD186149H, PJD186151H, PJD186152H, PJD186155H, PJD186157H, PJD186158H, PJD186159H, PJD186160H, PJD186165H, PJD186167H, PJD186169H, PJD186172H, PJD186173H, PJD186174H, PJD186175H, PJD186176H, PJD186178H, PJD186179H, PJD186182H, PJD186185H, PJD186186H, PJD186188H, PJD186189H, PJD186192H, PJD186193H, PJD186195H, PJD186196H, PJD188418H, PJD188420H, PJD188421H, PJD188423H, PJD188424H, PJD188425H, PJD188427H, PJD188428H, PJD188429H, PJD188430H, PJD188431H, PJD188432H, PJD188433H, PJD188434H, PJD188435H, PJD188438H, PJD188441H, PJD188442H, PJD188443H, PJD188444H, PJD188445H, PJD188446H, PJD188447H, PJD188448H, PJD188449H, PJD188450H, PJD188451H, PJD188452H, PJD188453H, PJD188454H, PJD188456H, PJD188457H, PJD188458H, PJD188459H, PJD188460H, PJD188463H, PJD188464H, PJD188465H, PJD188466H, PJD188467H, PJD188469H, PJD188470H, PJD188471H, PJD188472H, PJD188473H, PJD188475H, PJD188476H, PJD188477H, PJD188478H, PJD188480H, PJD188482H, PJD188483H, PJD188484H, PJD188485H, PJD188486H, PJD188487H, PJD188488H, PJD188489H, PJD188490H, PJD188493H, PJD188495H, PJD188496H, PJD188497H, PJD188499H, PJD188500H, PJD188501H, PJD188502H, PJD188503H, PJD188504H, PJD188506H, PJD188507H, PJD188508H, PJD188509H, PJD188510H, PJD188511H, PJD188512H, PJD188513H, PJD188515H, PJD188516H, PJD188517H, PJD188519H, PJD201652H, PJD201654H, PJD201657H, PJD201658H, PJD201659H, PJD201660H, PJD201662H, PJD201665H, PJD201666H, PJD201667H, PJD201669H, PJD201670H, PJD201671H, PJD201675H, PJD201676H, PJD201677H, PJD201679H, PJD201681H, PJD201682H, PJD201687H, PJD201688H, PJD201689H, PJD201690H, PJD201694H, PJD203257H, PJD203258H, PJD203259H, PJD203261H, PJD203262H, PJD203264H, PJD203265H, PJD203266H, PJD203267H, PJD203269H, PJD203270H, PJD203271H, PJD203272H, PJD203273H, PJD203274H, PJD203275H, PJD203276H, PJD203277H, PJD203278H, PJD203279H, PJD205935H, PJD205937H, PJD205939H, PJD205940H, PJD205942H, PJD205943H, PJD205944H, PJD205945H, PJD205946H, PJD205948H, PJD205949H, PJD205953H, PJD205955H, PJD205957H, PJD205959H, PJD205962H, PJD205966H, PJD205968H, PJD205970H, PJD205973H, PJD205974H, PJD205976H, PJD205977H, PJD205983H, PJD205984H, PJD205985H, PJD205986H, PJD205990H, PJD205992H, PJD207333H, PJD207334H, PJD207341H, PJD207356H, PJD207363H, PJD207371H, PJD207375H, PJD207382H, PJD207390H, PJD207392H, PJD207396H, PJD208691H, PJD208692H, PJD208703H, PJD208706H, PJD208711H, PJD208713H, PJD208714H, PJD208715H, PJD208724H, PJD208732H, PJD208734H, PJD208736H, PJD208739H, PJD208740H, PJD208743H, PJD209427H, PJD430299H, PJD430300H, PJD430301H, PJD430302H, PJD430303H, PJD430304H, PJD430305H, PJD430306H, PJD430307H, PJD430308H, PJD430309H, PJD430310H, PJD430311H, PJD430312H, PJD430313H, PJD430314H, PJD430316H, PJD430317H, PJD430318H, PJD430319H, PJD430320H, PJD430321H, PJD430322H, PJD430323H, PJD430324H, PJD430325H, PJD430326H, PJD430327H, PJD430328H, PJD430329H, PJD430330H, PJD430331H, PJD441378H, PJD441379H, PJD441380H, PJD441382H, PJD441383H, PJD441384H, PJD441385H, PJD441386H, PJD441387H, PJD441388H, PJD441389H, PJD441390H, PJD441392H, PJD441393H, PJD441394H, PJD441396H, PJD442436H, PJD442448H, PJD445371H, PJD445372H, PJD445373H, PJD445376H, PJD445378H, PJD445806H, PJD445807H, PJD445808H, PJD445809H, PJD445810H, PJD445811H, PJD445812H, PJD445814H, PJD445815H, PJD445816H, PJD445817H, PJD445818H, PJD445819H, PJD445820H, PJD445821H, PJD445822H, PJD445824H, PJD445825H, PJD445826H, PJD445827H, PJD448138H, PJD448139H, PJD610348S, PJD610357S, PJD610358S, PJD610789S, PJD610952S, PJD610953S, PJD610962S, PJD610968S, PJD610971S, PJD611308S, PJD611322S, PJD611326S, PJD611332S, PJD611340S, PJD611349S, PJD611351S, PJD611361S, PJD611364S, PJD611367S, PJD611369S, PJD611376S, PJD611965S, PJD612711S, PJD612774S, PJD614078S, PJD614081S, PJD614086S, PJD614087S, PJD675113S, PJD675114S, PJD675115S, PJD675116S, PJD675117S, PJD675118S, PJD675119S, PJD675120S, PJD675121S, PJD675122S, PJD675123S, PJD675124S, PJD675125S, PJD675128S, PJD675130S, PJD675131S, PJD675132S, PJD675133S, PJD675134S, PJD675135S, PJD675136S, PJD675137S, PJD675138S, PJD675140S, PJD675141S, PJD675142S, PJD675143S, PJD675144S, PJD675145S, PJD675146S, PJD675147S, PJD675148S, PJD675149S, PJD675150S, PJD675151S, PJD675153S, PJD675154S, PJD675157S, PJD675158S, PJD675159S, PJD675160S, PJD675161S, PJD675162S, PJD675163S, PJD675164S, PJD675165S, PJD675166S, PJD675167S, PJD675168S, PJD675169S, PJD675170S, PJD675171S, PJD675172S, PJD675173S, PJD675174S, PJD675175S, PJD675176S, PJD675177S, PJD675178S, PJD675179S, PJD675180S, PJD675181S, PJD675182S, PJD675183S, PJD675184S, PJD675185S, PJD675187S, PJD675188S, PJD675189S, PJD675190S, PJD675191S, PJD675192S, PJD675694S, PJD675695S, PJD675696S, PJD675697S, PJD675698S, PJD675699S, PJD675701S, PJD675702S, PJD675703S, PJD675704S, PJD675705S, PJD675706S, PJD675708S, PJD675709S, PJD675710S, PJD675711S, PJD675712S, PJD675713S, PJD675714S, PJD675715S, PJD675717S, PJD675718S, PJD675719S, PJD675720S, PJD675721S, PJD675722S, PJD675723S, PJD675724S, PJD675725S, PJD675727S, PJD675728S, PJD675729S, PJD675730S, PJD675731S, PJD675732S, PJD675734S, PJD675735S, PJD675736S, PJD675739S, PJD675740S, PJD675741S, PJD675743S, PJD675744S, PJD675745S, PJD676147S, PJD676148S, PJD676149S, PJD676150S, PJD676151S, PJD676152S, PJD676153S, PJD676154S, PJD676155S, PJD676156S, PJD676157S, PJD676158S, PJD676159S, PJD676160S, PJD676161S, PJD676162S, PJD676163S, PJD676164S, PJD676165S, PJD676166S, PJD676167S, PJD676168S, PJD676169S, PJD676170S, PJD676171S, PJD676172S, PJD676173S, PJD676174S, PJD676175S, PJD676176S, PJD676177S, PJD676178S, PJD676179S, PJD676180S, PJD676181S, PJD676183S, PJD676184S, PJD676185S, PJD676186S, PJD676187S, PJD676188S, PJD676189S, PJD676190S, PJD676192S, PJD676193S, PJD676194S, PJD676195S, PJD676196S, PJD676712S, PJD676714S, PJD676715S, PJD676716S, PJD676717S, PJD676718S, PJD676719S, PJD676720S, PJD676721S, PJD676722S, PJD676723S, PJD676724S, PJD676725S, PJD676726S, PJD676727S, PJD676728S, PJD676729S, PJD676730S, PJD676731S, PJD676732S, PJD676733S, PJD676734S, PJD676735S, PJD676737S, PJD676738S, PJD676739S, PJD676740S, PJD676741S, PJD676742S, PJD676743S, PJD676744S, PJD676745S, PJD676746S, PJD676747S, PJD676748S, PJD676750S, PJD676751S, PJD676752S, PJD676753S, PJD676754S, PJD676755S, PJD676756S, PJD676758S, PJD676759S, PJD676761S, PJD676762S, PJD676763S, PJD676764S, PJD676765S, PJD676766S, PJD676767S, PJD676768S, PJD676769S, PJD676770S, PJD676771S, PJD676772S, PJD676773S, PJD676774S, PJD676775S, PJD676776S, PJD676777S, PJD676778S, PJD676779S, PJD676780S, PJD676781S, PJD676783S, PJD676784S, PJD676785S, PJD676786S, PJD676787S, PJD676788S, PJD676789S, PJD676790S, PJD676791S, PJD676792S, PJD676793S, PJD676794S, PJD676795S, PJD676798S, PJD676799S, PJD676800S, PJD676801S, PJD676802S, PJD676803S, PJD676805S, PJD676806S, PJD676807S, PJD676808S, PJD676809S, PJD677182S, PJD677184S, PJD677185S, PJD677186S, PJD677187S, PJD677189S, PJD677190S, PJD677192S, PJD677193S, PJD677194S, PJD677195S, PJD677196S, PJD677197S, PJD677198S, PJD677199S, PJD677200S, PJD677201S, PJD677202S, PJD677203S, PJD677205S, PJD677206S, PJD677207S, PJD677208S, PJD677209S, PJD677210S, PJD677212S, PJD677213S, PJD677214S, PJD677215S, PJD677217S, PJD677218S, PJD677219S, PJD677220S, PJD677221S, PJD677222S, PJD677223S, PJD677224S, PJD677225S, PJD677226S, PJD677227S, PJD677228S, PJD677229S, PJD677230S, PJD677231S, PJD677232S, PJD677233S, PJD677234S, PJD677235S, PJD677236S, PJD677237S, PJD677238S, PJD677239S, PJD677240S, PJD677241S, PJD677242S, PJD677243S, PJD677244S, PJD677246S, PJD677247S, PJD677248S, PJD677249S, PJD677250S, PJD677251S, PJD677252S, PJD677253S, PJD677254S, PJD677255S, PJD677256S, PJD677257S, PJD677258S, PJD677259S, PJD677261S, PJD677471S, PJD677472S.PJD677473S, PJD677474S, PJD677475S, PJD677476S, PJD677477S, PJD677478S, PJD677479S, PJD677480S, PJD677481S, PJD677482S, PJD677483S, PJD677484S, PJD677485S, PJD677486S, PJD677487S, PJD677488S, PJD677489S, PJD677490S, PJD677492S, PJD677493S, PJD677494S, PJD677495S, PJD677496S, PJD677497S, PJD677498S, PJD677499S, PJD677945S, PJD678157S, PJD678158S, PJD678159S, PJD678160S, PJD678161S, PJD678162S, PJD678163S, PJD678164S, PJD678165S, PJD678166S, PJD678167S, PJD678168S, PJD678169S, PJD678170S, PJD678171S, PJD678172S, PJD678173S, PJD678174S, PJD678175S, PJD678176S, PJD678177S, PJD678178S, PJD678179S, PJD678180S, PJD678181S, PJD678182S, PJD678183S, PJD678184S, PJD678185S, PJD678186S, PJD678187S, PJD678188S, PJD678189S, PJD678190S, PJD678191S, PJD678192S, PJD678193S, PJD678194S, PJD678195S, PJD678196S, PJD678197S, PJD678198S, PJD678199S, PJD678200S, PJD678201S, PJD678202S, PJD678203S, PJD678204S, PJD678205S, PJD678206S, PJD678207S, PJD678208S, PJD678209S, PJD678210S, PJD678211S, PJD678212S, PJD678213S, PJD678214S, PJD678215S, PJD678216S, PJD678217S, PJD678218S, PJD678219S, PJD678220S, PJD678221S, PJD678222S, PJD678223S, PJD678224S, PJD678225S, PJD678226S, PJD678227S, PJD678228S, PJD679616S, PJD679617S, PJD679619S, PJD679622S, PJD679623S, PJD679624S, PJD679626S, PJD679628S, PJD679629S, PJD679633S, PJD679634S, PJD679636S, PJD679637S, PJD679638S, PJD679639S, PJD679640S, PJD679642S, PJD679644S, PJD679645S, PJD679647S, PJD679649S, PJD679650S, PJD679651S, PJD679652S, PJD679653S, PJD679654S, PJD679656S, PJD679657S, PJD679659S, PJD679660S, PJD679662S, PJD680212S, PJD680226S, PJD680929S, PJD680930S, PJD680933S, PJD680937S, PJD680938S, PJD680939S, PJD680942S, PJD680946S, PJD680949S, PJD680950S, PJD680952S, PJD680958S, PJD680959S, PJD680960S, PJD680962S, PJD680964S, PJD680966S, PJD680972S, PJD680975S, PJD680978S, PJD680981S, PJD680982S, PJD680984S, PJD680987S, PJD680990S, PJD680991S, PJD680993S, PJD680994S, PJD680995S, PJD680997S, PJD680998S, PJD680999S, PJD681001S, PJD681002S, PJD681003S, PJD681004S, PJD681012S, PJD681013S, PJD681016S, PJD681018S, PJD681020S, PJD681024S, PJD681025S, PJD685370S, PJD685371S, PJD685375S, PJD685377S, PJD685379S, PJD685380S, PJD685381S, PJD685382S, PJD685383S, PJD685385S, PJD685388S, PJD685399S, PJD685400S, PJD685402S, PJD685406S, PJD685408S, PJD685409S, PJD685410S, PJD686906S, PJD692195S, PJD692196S, PJD692197S, PJD692198S, PJD692199S, PJD692200S, PJD692201S, PJD692202S, PJD692203S, PJD692204S, PJD692205S, PJD692206S, PJD692207S, PJD692208S, PJD692210S, PJD692211S, PJD692212S, PJD692214S, PJD692215S, PJD692216S, PJD692217S, PJD692218S, PJD692219S, PJD692220S, PJD692221S, PJD692223S, PJD692224S, PJD692225S, PJD692226S, PJD692227S, PJD692228S, PJD692229S, PJD692230S, PJD692231S, PJD692232S, PJD692233S, PJD692234S, PJD692235S, PJD692236S, PJD692237S, PJD692238S, PJD692239S, PJD692240S, PJD692241S, PJD692242S, PJD692244S, PJD692245S, PJD692246S, PJD692247S, PJD692248S, PJD692249S, PJD692250S, PJD692252S, PJD692253S, PJD692254S, PJD692255S, PJD692256S, PJD692257S, PJD692258S, PJD692259S, PJD692260S, PJD692261S, PJD692262S, PJD692264S, PJD692265S, PJD692266S, PJD692267S, PJD692269S, PJD692270S, PJD692272S, PJD692273S, PJD692274S, PJD692275S, PJD692276S, PJD692277S, PJD692279S, PJD692282S, PJD692284S, PJD692286S, PJD692287S, PJD692288S, PJD692289S, PJD692290S, PJD692291S, PJD692293S, PJD692294S, PJD692295S, PJD692296S, PJD692297S, PJD692298S, PJD692299S, PJD692301S, PJD692302S, PJD692303S, PJD692306S, PJD692307S, PJD693032S, PJD693033S, PJD693034S, PJD693035S, PJD693036S, PJD693037S, PJD693038S, PJD693039S, PJD693040S, PJD693041S, PJD693042S, PJD693043S, PJD693044S, PJD693045S, PJD693046S, PJD693047S, PJD693049S, PJD693050S, PJD693051S, PJD693052S, PJD693053S, PJD693054S, PJD693055S, PJD693056S, PJD693057S, PJD693058S, PJD693059S, PJD693060S, PJD693061S, PJD693062S, PJD693063S, PJD693064S, PJD693065S, PJD693066S, PJD693067S, PJD693068S, PJD693069S, PJD693070S, PJD693071S, PJD693072S, PJD693073S, PJD693074S, PJD693075S, PJD693076S, PJD693077S, PJD693078S, PJD693079S, PJD693080S, PJD693081S, PJD693082S, PJD693083S, PJD693084S, PJD693085S, PJD693086S, PJD693087S, PJD693088S, PJD693089S, PJD693090S, PJD693091S, PJD693092S, PJD693094S, PJD693095S, PJD693096S, PJD693097S, PJD693098S, PJD693099S, PJD693100S, PJD693101S, PJD693102S, PJD693103S, PJD693104S, PJD693105S, PJD693106S, PJD693107S, PJD693108S, PJD693109S, PJD693110S, PJD693111S, PJD693112S, PJD693113S, PJD693114S, PJD693115S, PJD693116S, PJD693117S, PJD693118S, PJD693119S, PJD693120S, PJD693121S, PJD693692S, PJD693693S, PJD693694S, PJD693695S, PJD693696S, PJD693697S, PJD693698S, PJD693699S, PJD693700S, PJD693701S, PJD693702S, PJD693703S, PJD693704S, PJD693705S, PJD693706S, PJD693707S, PJD693708S, PJD693709S, PJD693710S, PJD693711S, PJD693712S, PJD693713S, PJD693714S, PJD693715S, PJD693716S, PJD693717S, PJD693718S, PJD693720S, PJD693721S, PJD693722S, PJD693723S, PJD693724S, PJD693725S, PJD693726S, PJD693727S, PJD693728S, PJD693729S, PJD693730S, PJD693731S, PJD693732S, PJD693733S, PJD693734S, PJD693735S, PJD693736S, PJD693737S, PJD693738S, PJD693739S, PJD693740S, PJD693741S, PJD693742S, PJD693743S, PJD693744S, PJD693745S, PJD693746S, PJD693747S, PJD693749S, PJD693750S, PJD693751S, PJD693752S, PJD693754S, PJD693755S, PJD693756S, PJD693757S, PJD693758S, PJD693759S, PJD693760S, PJD693761S, PJD693762S, PJD693763S, PJD693764S, PJD693765S, PJD693766S, PJD693767S, PJD693768S, PJD693770S, PJD693771S, PJD693772S, PJD693773S, PJD693774S, PJD693775S, PJD693776S, PJD693777S, PJD693778S, PJD693779S, PJD693780S, PJD693781S, PJD693782S, PJD693783S, PJD693784S, PJD693785S, PJD693786S, PJD693787S, PJD693788S, PJD693789S, PJD693790S, PJD693791S, PJD694177S, PJD694178S, PJD694179S, PJD694180S, PJD694181S, PJD694182S, PJD694183S, PJD694184S, PJD694185S, PJD694186S, PJD694187S, PJD694188S, PJD694189S, PJD694190S, PJD694191S, PJD694192S, PJD694193S, PJD694194S, PJD694195S, PJD694196S, PJD694197S, PJD694198S, PJD694199S, PJD694200S, PJD694201S, PJD694202S, PJD694203S, PJD694204S, PJD694205S, PJD694206S, PJD694207S, PJD694208S, PJD694209S, PJD694210S, PJD694211S, PJD694212S, PJD694213S, PJD694214S, PJD694215S, PJD694216S, PJD694217S, PJD694218S, PJD694219S, PJD694220S, PJD694221S, PJD694222S, PJD694223S, PJD694224S, PJD694225S, PJD694226S, PJD694711S, PJD694713S, PJD694715S, PJD694716S, PJD694718S, PJD694719S, PJD694732S, PJD694734S, PJD694735S, PJD694736S, PJD694737S, PJD694738S, PJD694739S, PJD694740S, PJD694741S, PJD694742S, PJD694743S, PJD694744S, PJD694745S, PJD694746S, PJD694747S, PJD694748S, PJD694749S, PJD694750S, PJD694751S, PJD694752S, PJD694753S, PJD694754S, PJD694755S, PJD694756S, PJD694757S, PJD694758S, PJD694759S, PJD694760S, PJD694761S, PJD694762S, PJD694763S, PJD694764S, PJD694765S, PJD694766S, PJD694767S, PJD694768S, PJD694769S, PJD694770S, PJD694771S, PJD694772S, PJD694773S, PJD694774S, PJD694775S, PJD694776S, PJD694778S, PJD694780S, PJD694781S, PJD694782S, PJD694783S, PJD694784S, PJD694785S, PJD694786S, PJD694787S, PJD694788S, PJD694789S, PJD694790S, PJD694791S, PJD694793S, PJD694794S, PJD694795S, PJD694796S, PJD694797S, PJD694798S, PJD694799S, PJD694800S, PJD695188S, PJD695189S, PJD695190S, PJD695191S, PJD695192S, PJD695193S, PJD695194S, PJD695195S, PJD695197S, PJD695199S, PJD695200S, PJD695201S, PJD695461S, PJD695462S, PJD695463S, PJD695467S, PJD695468S, PJD695469S, PJD695470S, PJD695471S, PJD695472S, PJD695473S, PJD695475S, PJD695476S, PJD695479S, PJD695480S, PJD695481S, PJD695482S, PJD695484S, PJD695485S, PJD695486S, PJD695488S, PJD695489S, PJD695491S, PJD695492S, PJD695493S, PJD695494S, PJD695495S, PJD695496S, PJD695497S, PJD695498S, PJD695499S, PJD695500S, PJD695501S, PJD695503S, PJD695504S, PJD695506S, PJD695507S, PJD695508S, PJD695509S, PJD695510S, PJD695511S, PJD695512S, PJD695514S, PJD695515S, PJD695516S, PJD695519S, PJD695520S, PJD695521S, PJD695522S, PJD695524S, PJD695525S, PJD695526S, PJD695527S, PJD695529S, PJD695530S, PJD695531S, PJD695535S, PJD695536S, PJD695537S, PJD695539S, PJD695540S, PJD695541S, PJD695542S, PJD695544S, PJD695545S, PJD695546S, PJD695547S, PJD695550S, PJD695552S, PJD695553S, PJD695554S, PJD695557S, PJD695558S, PJD695559S, PJD695562S, PJD695563S, PJD695565S, PJD695566S, PJD695570S, PJD695571S, PJD695572S, PJD696405S, PJD696617S, PJD696696S, PJD696701S, PJD696709S, PJD696739S, PJD696761S, PJD696788S, PJD696792S, PJD696807S, PJD696808S, PJD696809S, PJD696818S, PJD704625S, PJD704627S, PJD704629S, PJD730015S, PJD731580S, PJD731581S, PJD731582S, PJD731583S, PJD731584S, PJD731585S, PJD731586S, PJD731587S, PJD731588S, PJD731590S, PJD731591S, PJD731592S, PJD731593S, PJD731594S, PJD731595S, PJD731596S, PJD731597S, PJD731598S, PJD731599S, PJD731600S, PJD731601S, PJD731602S, PJD731603S, PJD731604S, PJD731605S, PJD731606S, PJD731608S, PJD731609S, PJD731610S, PJD731611S, PJD731612S, PJD731613S, PJD731614S, PJD731615S, PJD731616S, PJD731617S, PJD731619S, PJD731620S, PJD731621S, PJD731622S, PJD731623S, PJD731624S, PJD731625S, PJD731626S, PJD731627S, PJD731628S, PJD731629S, PJD731631S, PJD731632S, PJD731633S, PJD731637S, PJD731639S, PJD731640S, PJD731641S, PJD731642S, PJD731643S, PJD731644S, PJD731645S, PJD731646S, PJD731647S, PJD731649S, PJD731650S, PJD731651S, PJD731652S, PJD731654S, PJD731655S, PJD731656S, PJD731658S, PJD731659S, PJD731660S, PJD731662S, PJD731663S, PJD731664S, PJD731665S, PJD731666S, PJD731668S, PJD732144S, PJD732145S, PJD732146S, PJD732147S, PJD732148S, PJD732149S, PJD732150S, PJD732152S, PJD732153S, PJD732154S, PJD732155S, PJD732156S, PJD732157S, PJD732158S, PJD732161S, PJD732162S, PJD732163S, PJD732164S, PJD732165S, PJD732166S, PJD732168S, PJD732169S, PJD732170S, PJD732171S, PJD732174S, PJD732175S, PJD732176S, PJD732177S, PJD732178S, PJD732179S, PJD732180S, PJD732182S, PJD732185S, PJD732187S, PJD732190S, PJD732191S, PJD732193S, PJD732194S, PJD732195S, PJD732196S, PJD732198S, PJD732199S, PJD732201S, PJD732202S, PJD732204S, PJD732205S, PJD732206S, PJD732209S, PJD732210S, PJD732212S, PJD732213S, PJD732214S, PJD732215S, PJD732216S, PJD732218S, PJD732219S, PJD732220S, PJD732221S, PJD732273S, PJD732277S, PJD732279S, PJD732283S, PJD732284S, PJD732285S, PJD732287S, PJD732288S, PJD732293S, PJD732294S, PJD732295S, PJD732296S, PJD732297S, PJD732298S, PJD732299S, PJD732300S, PJD732301S, PJD733190S, PJD733201S, PJD733214S, PJD733218S, PJD733221S, PJD733238S, PJD733285S, PJD733291S, PJD733299S, PJD733305S, PJD733312S, PJD733321S, PJD738118S, PJD738119S, PJD738120S, PJD738121S, PJD738122S, PJD738123S, PJD738125S, PJD738126S, PJD738127S, PJD738128S, PJD738129S, PJD738130S, PJD738131S, PJD738132S, PJD738134S, PJD738135S, PJD738136S, PJD738137S, PJD738138S, PJD738139S, PJD738140S, PJD738141S, PJD738142S, PJD738143S, PJD738144S, PJD738145S, PJD738146S, PJD738147S, PJD738148S, PJD738149S, PJD738150S, PJD738151S, PJD738153S, PJD738154S, PJD738155S, PJD738157S, PJD738158S, PJD738160S, PJD738161S, PJD738162S, PJD738163S, PJD738164S, PJD738165S, PJD738167S, PJD738168S, PJD738169S, PJD738170S, PJD738171S, PJD738172S, PJD738173S, PJD738174S, PJD738175S, PJD738176S, PJD738177S, PJD738178S, PJD738179S, PJD738180S, PJD738181S, PJD738182S, PJD738183S, PJD738184S, PJD738185S, PJD738186S, PJD738187S, PJD738188S, PJD738189S, PJD738190S, PJD738191S, PJD738192S, PJD738193S, PJD738194S, PJD738195S, PJD738196S, PJD738198S, PJD738199S, PJD738200S, PJD738201S, PJD738202S, PJD738203S, PJD738204S, PJD738205S, PJD738206S, PJD738207S, PJD738209S, PJD738210S, PJD738211S, PJD738212S, PJD738213S, PJD738214S, PJD738215S, PJD738216S, PJD738217S, PJD738218S, PJD738219S, PJD738220S, PJD738221S, PJD738222S, PJD738223S, PJD738224S, PJD738225S, PJD738226S, PJD738227S, PJD738228S, PJD738231S, PJD738233S, PJD738234S, PJD738235S, PJD738236S, PJD738237S, PJD738238S, PJD738239S, PJD738240S, PJD738241S, PJD738242S, PJD738243S, PJD738244S, PJD738245S, PJD738246S, PJD738248S, PJD738249S, PJD738250S, PJD738252S, PJD738253S, PJD738254S, PJD738255S, PJD738256S, PJD738257S, PJD738258S, PJD738259S, PJD738260S, PJD738261S, PJD738262S, PJD738263S.PJD738264S, PJD738265S, PJD738266S, PJD738267S, PJD738268S, PJD738269S, PJD738270S, PJD738273S, PJD738274S, PJD738275S, PJD738276S, PJD738277S, PJD738278S, PJD738279S, PJD738280S, PJD738281S, PJD738282S, PJD738284S, PJD738286S, PJD738288S, PJD738289S, PJD738290S, PJD738291S, PJD738292S, PJD738293S, PJD738294S, PJD738295S, PJD738296S, PJD738299S, PJD738300S, PJD738304S, PJD738305S, PJD738306S, PJD738308S, PJD738309S, PJD738312S, PJD738315S, PJD738316S, PJD738318S, PJD738320S, PJD738325S, PJD738326S, PJD738328S, PJD738330S, PJD738331S, PJD738332S, PJD738333S, PJD738334S, PJD738336S, PJD738337S, PJD738338S, PJD738339S, PJD738342S, PJD738343S, PJD738344S, PJD738346S, PJD738347S, PJD738348S, PJD738349S, PJD738351S, PJD738352S, PJD738353S, PJD738356S, PJD738358S, PJD738362S, PJD738363S, PJD738366S, PJD738368S, PJD738369S, PJD738371S, PJD738374S, PJD738375S, PJD738376S, PJD738378S, PJD738380S, PJD738381S, PJD738382S, PJD738386S, PJD738388S, PJD738391S, PJD738392S, PJD738393S, PJD738395S, PJD738396S, PJD738398S, PJD738399S, PJD738402S, PJD738403S, PJD738404S, PJD738613S, PJD738614S, PJD738615S, PJD738616S, PJD738617S, PJD738618S, PJD738620S, PJD738621S, PJD738622S, PJD738623S, PJD738624S, PJD738625S, PJD738626S, PJD738627S, PJD738628S, PJD738629S, PJD738630S, PJD738631S, PJD738632S, PJD738633S, PJD738634S, PJD738635S, PJD738636S, PJD738637S, PJD738638S, PJD738639S, PJD738640S, PJD738641S, PJD738642S, PJD738643S, PJD738644S, PJD738645S, PJD738646S, PJD738647S, PJD738648S, PJD738649S, PJD738650S, PJD738651S, PJD738652S, PJD738860S, PJD738861S, PJD738862S, PJD738863S, PJD738864S, PJD738865S, PJD738866S, PJD738867S, PJD738868S, PJD738869S, PJD738870S, PJD738871S, PJD738872S, PJD738873S, PJD738874S, PJD738875S, PJD738876S, PJD738877S, PJD738878S, PJD738879S, PJD738880S, PJD738881S, PJD738882S, PJD738885S, PJD738886S, PJD738887S, PJD738888S, PJD738889S, PJD738890S, PJD738892S, PJD738893S, PJD738894S, PJD738895S, PJD738896S, PJD738898S, PJD738900S, PJD739168S, PJD739169S, PJD739170S, PJD739171S, PJD739172S, PJD739173S, PJD739174S, PJD739175S, PJD739176S, PJD739177S, PJD739178S, PJD739179S, PJD739180S, PJD739181S, PJD739182S, PJD739183S, PJD739184S, PJD739185S, PJD739186S, PJD739187S, PJD739188S, PJD739189S, PJD739190S, PJD739191S, PJD739192S, PJD739193S, PJD739196S, PJD739197S, PJD739198S, PJD739199S, PJD739201S, PJD739202S, PJD739203S, PJD739204S, PJD739205S, PJD739206S, PJD739208S, PJD739209S, PJD739211S, PJD739212S, PJD739213S, PJD739214S, PJD739215S, PJD739216S, PJD739217S, PJD739218S, PJD739220S, PJD739221S, PJD739222S, PJD739223S, PJD739224S, PJD739225S, PJD739226S, PJD739227S, PJD739228S, PJD739230S, PJD739232S, PJD739233S, PJD739234S, PJD739235S, PJD739236S, PJD739239S, PJD739240S, PJD739242S, PJD739243S, PJD739244S, PJD739246S, PJD739250S, PJD739251S, PJD739252S, PJD739253S, PJD739254S, PJD739256S, PJD739257S, PJD739258S, PJD739259S, PJD739260S, PJD739261S, PJD739262S, PJD739263S, PJD739264S, PJD739265S, PJD739267S, PJD739268S, PJD739269S, PJD739270S, PJD739272S, PJD739273S, PJD739274S, PJD739275S, PJD739276S, PJD739277S, PJD739279S, PJD739280S, PJD739281S, PJD739283S, PJD739284S, PJD739286S, PJD739288S, PJD739289S, PJD739290S, PJD739293S, PJD739294S, PJD739296S, PJD739297S, PJD739298S, PJD739301S, PJD739302S, PJD739305S, PJD739308S, PJD739310S, PJD739311S, PJD739312S, PJD739313S, PJD739314S, PJD739316S, PJD739320S, PJD739321S, PJD739323S, PJD739324S, PJD739325S, PJD739326S, PJD739647S, PJD739648S, PJD739650S, PJD739651S, PJD739652S, PJD739654S, PJD739655S, PJD739656S, PJD739657S, PJD739658S, PJD739661S, PJD739662S, PJD739665S, PJD739666S, PJD739667S, PJD739668S, PJD739669S, PJD739670S, PJD739671S, PJD739672S, PJD739673S, PJD739674S, PJD739675S, PJD739676S, PJD739677S, PJD739678S, PJD739679S, PJD739680S, PJD739681S, PJD739682S, PJD739683S, PJD739684S, PJD739685S, PJD739686S, PJD739687S, PJD739688S, PJD739689S, PJD739691S, PJD739692S, PJD739693S, PJD739694S, PJD739695S, PJD739696S, PJD739698S, PJD739699S, PJD739700S, PJD739701S, PJD739702S, PJD739703S, PJD739704S, PJD739705S, PJD739706S, PJD739707S, PJD739708S, PJD739709S, PJD739710S, PJD739712S, PJD739713S, PJD739714S, PJD739715S, PJD739716S, PJD739717S, PJD739718S, PJD739720S, PJD739721S, PJD739722S, PJD739723S, PJD739724S, PJD739725S, PJD739726S, PJD739727S, PJD739728S, PJD739729S, PJD739730S, PJD739732S, PJD739733S, PJD739734S, PJD739735S, PJD739736S, PJD739738S, PJD739739S, PJD739740S, PJD739741S, PJD739743S, PJD739744S, PJD740545S, PJD740546S, PJD740547S, PJD740548S, PJD740549S, PJD740550S, PJD740552S, PJD740553S, PJD740554S, PJD740555S, PJD740556S, PJD740557S, PJD740558S, PJD740559S, PJD740560S, PJD740561S, PJD740562S, PJD740563S, PJD740564S, PJD740565S, PJD740566S, PJD740567S, PJD740568S, PJD740569S, PJD740570S, PJD740571S, PJD740572S, PJD740573S, PJD740574S, PJD740575S, PJD740576S, PJD740577S, PJD740579S, PJD740580S, PJD740581S, PJD740582S, PJD740583S, PJD740584S, PJD740585S, PJD740586S, PJD740587S, PJD740588S, PJD740589S, PJD740590S, PJD740592S, PJD740593S, PJD740594S, PJD740595S, PJD740596S, PJD740597S, PJD740598S, PJD740599S, PJD740600S, PJD740601S, PJD740602S, PJD740603S, PJD740604S, PJD740605S, PJD740606S, PJD740607S, PJD740608S, PJD740609S, PJD740611S, PJD740612S, PJD740613S, PJD740614S, PJD740615S, PJD740616S, PJD740617S, PJD740618S, PJD740619S, PJD740620S, PJD740621S, PJD740622S, PJD740623S, PJD740624S, PJD740625S, PJD740626S, PJD740627S, PJD740628S, PJD740629S, PJD740630S, PJD740631S, PJD740632S, PJD740633S, PJD740634S, PJD740635S, PJD740636S, PJD740637S, PJD740638S, PJD740639S, PJD740640S, PJD740641S, PJD740642S, PJD740643S, PJD740974S, PJD740975S, PJD740976S, PJD740977S, PJD740978S, PJD740979S, PJD740981S, PJD740984S, PJD740985S, PJD740986S, PJD740987S, PJD740988S, PJD740989S, PJD740990S, PJD740991S, PJD740992S, PJD740993S, PJD740994S, PJD740995S, PJD740996S, PJD740997S, PJD740998S, PJD740999S, PJD741000S, PJD741001S, PJD741002S, PJD741003S, PJD741004S, PJD741005S, PJD741006S, PJD741007S, PJD741008S, PJD741010S, PJD741011S, PJD741012S, PJD741013S, PJD741015S, PJD741016S, PJD741017S, PJD741018S, PJD741019S, PJD741020S, PJD741021S, PJD741022S, PJD741023S, PJD741025S, PJD741026S, PJD741027S, PJD741029S, PJD741030S, PJD741031S, PJD741032S, PJD741033S, PJD741034S, PJD741035S, PJD741036S, PJD741037S, PJD741038S, PJD741039S, PJD741040S, PJD741041S, PJD741042S, PJD741043S, PJD741044S, PJD741045S, PJD741046S, PJD741048S, PJD741049S, PJD741050S, PJD741051S, PJD741052S, PJD741053S, PJD741054S, PJD741055S, PJD741056S, PJD741057S, PJD741058S, PJD741059S, PJD741060S, PJD741062S, PJD741063S, PJD741065S, PJD741066S, PJD741067S, PJD741068S, PJD741071S, PJD741073S, PJD741074S, PJD741075S, PJD741076S, PJD741077S, PJD741078S, PJD741079S, PJD741080S, PJD741081S, PJD741082S, PJD741083S, PJD741246S, PJD741247S, PJD741251S, PJD741261S, PJD741273S, PJD741281S, PJD741283S, PJD741284S, PJD741285S, PJD741287S, PJD741293S, PJD741299S, PJD741301S, PJD741303S, PJD741305S, PJD741307S, PJD741309S, PJD741660S, PJD741672S, and PJD744349S.   
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc. Cardiac Rhythm Disease Managment
8200 Coral Sea St. N.E.
Saint Paul MN 55112
For Additional Information Contact
763-514-4000
Manufacturer Reason
for Recall
One or more bond wire pairs will lift or separate from the bonding terminals on the device hybrid. This may present clinically as loss of rate response, premature battery depletion, loss of telemetry, or no output.
FDA Determined
Cause 2
Device Design
ActionMedtronic Sales representative hand delivered an Important Patient Safety Information letter, dated May 2009, to physicians beginning 05/18/09. The letter describes the issues, provides a predicted failure rate for the 3 populations of devices, and provides Patient Management Recommendations. The letter recommended that physicians should advise their patients to seek medical attention immediately if they experience symptoms (e.g., fainting or lightheadedness). It was also recommended that physicians should consider device replacement for patients who are both pacemaker dependent and who have been implanted with a device in the affected subsets. Medtronic will offer supplemental device warranty if the device is not already at elective replacement time. The last recommendation was that physicians should continue routine follow up in accordance with standard practice for those patients who are not pacemaker dependent. The letter also provides Physician and Patient Support. Letters will be mailed to patients with registered devices beginning 05/27/09. Questions should be directed to a local Medtronic Representative or Medtronic Technical Services at 800-505-4636.
Quantity in Commerce2247
DistributionWorldwide Distribution -- USA, including Washington D.C., Guam, Puerto Rico, and Virgin Islands; Albania, Algeria, Argentina, Armenia, Aruba, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latin America, Latvia, Lebanon, Luxembourg, Martinique, Mauritius, Mexico, Morocco, Nepal, Netherlands, Netherlands Antilles, Norway, Oman, Pakistan, Palestinian Territory Occupied, Peru, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Senegal, Slovakia (Slovak Republic), Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Vatican City State (Holy See), and Venezuela.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.