| Class 2 Device Recall 4D Integrated Treatment Console | |
Date Initiated by Firm | October 16, 2009 |
Date Posted | January 10, 2011 |
Recall Status1 |
Terminated 3 on July 20, 2011 |
Recall Number | Z-0891-2011 |
Recall Event ID |
54146 |
510(K)Number | K091132 |
Product Classification |
Accelerator, linear, medical - Product Code IYE
|
Product | 4D Integrated Treatment Console, versions: 8.0.12, 8.0.23 SP2, 8.1 SP2, 8.1.1.3, 8.1.1.4, 8.1.2, 8.1.3, 8.1.4, 8.3, 8.6, 8.6.1, 8.6.2, 8.8.0; Model Number: H51; Distributed and manufactured by: Varian Medical Systems Inc., Palo Alto, CA
Intended use: The 4D Integrated Treatment Console function is designed to assist the operator of radiation therapy device in providing accurate treatment setups for each patient by monitoring setup parameters and preventing the radiation therapy device from commencing irradiation when any parameter is out of conformance with the treatment plan. |
Code Information |
H510001 ,H510049 ,H510129 ,H510202 ,H510286 ,H510350 ,H510410 ,H510468 ,H510514 H510002 ,H510050 ,H510130 ,H510206 ,H510289 ,H510351 ,H510411 ,H510470 ,H510515 H510003 ,H510067 ,H510131 ,H510207 ,H510290 ,H510352 ,H510417 ,H510471 ,H510516 H510004 ,H510068 ,H510135 ,H510211 ,H510291 ,H510353 ,H510419 ,H510472 ,H510517 H510005 ,H510069 ,H510136 ,H510215 ,H510292 ,H510364 ,H510425 ,H510473 ,H510519 H510007 ,H510070 ,H510137 ,H510220 ,H510293 ,H510368 ,H510426 ,H510474 ,H510520 H510008 ,H510071 ,H510146 ,H510222 ,H510300 ,H510370 ,H510427 ,H510476 ,H510521 H510009 ,H510072 ,H510150 ,H510225 ,H510301 ,H510371 ,H510428 ,H510477 ,H510522 H510010 ,H510074 ,H510151 ,H510226 ,H510302 ,H510372 ,H510429 ,H510479 ,H510523 H510012 ,H510075 ,H510160 ,H510231 ,H510303 ,H510373 ,H510430 ,H510480 ,H510525 H510014 ,H510077 ,H510162 ,H510233 ,H510304 ,H510375 ,H510431 ,H510481 ,H510526 H510015 ,H510081 ,H510163 ,H510234 ,H510306 ,H510376 ,H510432 ,H510482 ,H510528 H510016 ,H510083 ,H510165 ,H510236 ,H510307 ,H510377 ,H510433 ,H510483 ,H510529 H510017 ,H510084 ,H510171 ,H510241 ,H510309 ,H510378 ,H510434 ,H510484 ,H510530 H510018 ,H510088 ,H510172 ,H510262 ,H510310 ,H510379 ,H510435 ,H510488 ,H510531 H510019 ,H510089 ,H510173 ,H510263 ,H510316 ,H510380 ,H510436 ,H510489 ,H510532 H510021 ,H510090 ,H510174 ,H510264 ,H510317 ,H510381 ,H510437 ,H510490 ,H510533 H510022 ,H510091 ,H510175 ,H510265 ,H510318 ,H510382 ,H510441 ,H510491 ,H510534 H510025 ,H510092 ,H510176 ,H510267 ,H510319 ,H510383 ,H510442 ,H510492 ,H510535 H510027 ,H510093 ,H510180 ,H510268 ,H510320 ,H510384 ,H510443 ,H510493 ,H510536 H510029 ,H510094 ,H510181 ,H510269 ,H510321 ,H510385 ,H510445 ,H510494 ,H510537 H510030 ,H510095 ,H510183 ,H510270 ,H510322 ,H510386 ,H510446 ,H510495 ,H510538 H510031 ,H510096 ,H510184 ,H510271 ,H510330 ,H510388 ,H510447 ,H510496 ,H510539 H510032 ,H510097 ,H510185 ,H510272 ,H510332 ,H510389 ,H510448 ,H510498 ,H510540 H510033 ,H510098 ,H510188 ,H510273 ,H510335 ,H510391 ,H510449 ,H510499 ,H510541 H510035 ,H510099 ,H510189 ,H510275 ,H510337 ,H510392 ,H510450 ,H510500 ,H510542 H510036 ,H510104 ,H510190 ,H510276 ,H510338 ,H510394 ,H510451 ,H510501 ,H510543 H510039 ,H510106 ,H510191 ,H510277 ,H510339 ,H510396 ,H510456 ,H510503 ,H510544 H510040 ,H510107 ,H510192 ,H510278 ,H510340 ,H510397 ,H510457 ,H510505 ,H510549 H510041 ,H510113 ,H510194 ,H510279 ,H510341 ,H510398 ,H510459 ,H510506 ,H510550 H510042 ,H510114 ,H510195 ,H510280 ,H510342 ,H510399 ,H510460 ,H510507 ,H510551 H510043 ,H510115 ,H510196 ,H510281 ,H510343 ,H510401 ,H510461 ,H510510 ,H510552 H510044 ,H510125 ,H510197 ,H510282 ,H510344 ,H510402 ,H510464 ,H510511 ,H510553 H510045 ,H510126 ,H510199 ,H510283 ,H510345 ,H510408 ,H510465 ,H510512 ,H510554 H510046 ,H510127 ,H510201 ,H510285 ,H510346 ,H510409 ,H510467 ,H510513 ,H510555 H510556 ,H510613 ,H510665 ,H510731 ,H510786 ,H510844 ,H510912 ,H510959 ,H511014 H510560 ,H510614 ,H510668 ,H510732 ,H510788 ,H510845 ,H510913 ,H510964 ,H511015 H510564 ,H510616 ,H510669 ,H510733 ,H510789 ,H510846 ,H510914 ,H510966 ,H511016 H510565 ,H510617 ,H510670 ,H510735 ,H510791 ,H510847 ,H510915 ,H510967 ,H511018 H510567 ,H510618 ,H510672 ,H510736 ,H510793 ,H510848 ,H510916 ,H510968 ,H511021 H510568 ,H510619 ,H510673 ,H510738 ,H510794 ,H510849 ,H510917 ,H510970 ,H511022 H510569 ,H510620 ,H510676 ,H510739 ,H510795 ,H510850 ,H510918 ,H510971 ,H511023 H510570 ,H510621 ,H510677 ,H510740 ,H510797 ,H510852 ,H510920 ,H510972 ,H511024 H510571 ,H510622 ,H510678 ,H510741 ,H510798 ,H510853 ,H510921 ,H510974 ,H511025 H510574 ,H510623 ,H510679 ,H510742 ,H510799 ,H510854 ,H510922 ,H510976 ,H511026 H510575 ,H510624 ,H510680 ,H510743 ,H510800 ,H510855 ,H510923 ,H510977 ,H511028 H510577 ,H510625 ,H510681 ,H510744 ,H510801 ,H510856 ,H510924 ,H510978 ,H511029 H510578 ,H510628 ,H510682 ,H510745 ,H510802 ,H510860 ,H510925 ,H510979 ,H511030 H510579 ,H510631 ,H510683 ,H510746 ,H510804 ,H510862 ,H510926 ,H510980 ,H511031 H510580 ,H510632 ,H510684 ,H510748 ,H510805 ,H510863 ,H510927 ,H510981 ,H511032 H510581 ,H510634 ,H510686 ,H510749 ,H510806 ,H510864 ,H510930 ,H510982 ,H511033 H510582 ,H510635 ,H510688 ,H510750 ,H510807 ,H510865 ,H510931 ,H510983 ,H511034 H510584 ,H510636 ,H510689 ,H510751 ,H510808 ,H510866 ,H510932 ,H510984 ,H511035 H510586 ,H510637 ,H510690 ,H510754 ,H510809 ,H510867 ,H510933 ,H510987 ,H511036 H510587 ,H510638 ,H510691 ,H510756 ,H510811 ,H510868 ,H510934 ,H510989 ,H511039 H510588 ,H510639 ,H510693 ,H510758 ,H510814 ,H510885 ,H510935 ,H510990 ,H511040 H510589 ,H510640 ,H510694 ,H510759 ,H510816 ,H510886 ,H510936 ,H510991 ,H511042 H510590 ,H510641 ,H510695 ,H510760 ,H510819 ,H510887 ,H510937 ,H510992 ,H511043 H510591 ,H510646 ,H510699 ,H510761 ,H510820 ,H510888 ,H510938 ,H510994 ,H511044 H510592 ,H510647 ,H510701 ,H510763 ,H510822 ,H510889 ,H510941 ,H510995 ,H511045 H510594 ,H510648 ,H510702 ,H510769 ,H510823 ,H510891 ,H510942 ,H510996 ,H511046 H510595 ,H510649 ,H510705 ,H510770 ,H510825 ,H510892 ,H510943 ,H510997 ,H511047 H510596 ,H510650 ,H510706 ,H510773 ,H510827 ,H510894 ,H510944 ,H510998 ,H511049 H510598 ,H510651 ,H510707 ,H510774 ,H510828 ,H510896 ,H510945 ,H510999 ,H511050 H510599 ,H510652 ,H510708 ,H510775 ,H510830 ,H510897 ,H510946 ,H511000 ,H511051 H510600 ,H510653 ,H510709 ,H510776 ,H510833 ,H510898 ,H510947 ,H511001 ,H511052 H510602 ,H510654 ,H510710 ,H510777 ,H510834 ,H510901 ,H510948 ,H511002 ,H511053 H510603 ,H510655 ,H510712 ,H510778 ,H510835 ,H510903 ,H510949 ,H511003 ,H511054 H510604 ,H510656 ,H510713 ,H510779 ,H510837 ,H510905 ,H510950 ,H511004 ,H511055 H510605 ,H510657 ,H510714 ,H510780 ,H510838 ,H510906 ,H510951 ,H511006 ,H511056 H510607 ,H510658 ,H510716 ,H510781 ,H510839 ,H510907 ,H510952 ,H511008 ,H511057 H510608 ,H510659 ,H510717 ,H510782 ,H510840 ,H510908 ,H510953 ,H511009 ,H511058 H510609 ,H510662 ,H510718 ,H510783 ,H510841 ,H510909 ,H510954 ,H511010 ,H511060 H510610 ,H510663 ,H510722 ,H510784 ,H510842 ,H510910 ,H510955 ,H511012 ,H511061 H510612 ,H510664 ,H510730 ,H510785 ,H510843 ,H510911 ,H510957 ,H511013 ,H511062 H511063 ,H511114 ,H511166 ,H511219 ,H511267 ,H511328 ,H511379 ,H511434 ,H511480 H511064 ,H511116 ,H511167 ,H511221 ,H511268 ,H511329 ,H511382 ,H511436 ,H511482 H511066 ,H511117 ,H511168 ,H511222 ,H511269 ,H511330 ,H511385 ,H511437 ,H511483 H511069 ,H511119 ,H511169 ,H511223 ,H511270 ,H511331 ,H511386 ,H511439 ,H511484 H511070 ,H511120 ,H511170 ,H511224 ,H511273 ,H511332 ,H511389 ,H511440 ,H511485 H511071 ,H511122 ,H511174 ,H511225 ,H511274 ,H511333 ,H511390 ,H511441 ,H511486 H511072 ,H511123 ,H511175 ,H511227 ,H511275 ,H511334 ,H511391 ,H511442 ,H511487 H511073 ,H511125 ,H511176 ,H511229 ,H511276 ,H511335 ,H511392 ,H511443 ,H511488 H511074 ,H511126 ,H511178 ,H511230 ,H511279 ,H511336 ,H511393 ,H511446 ,H511489 H511075 ,H511128 ,H511179 ,H511231 ,H511281 ,H511337 ,H511394 ,H511447 ,H511490 H511077 ,H511129 ,H511183 ,H511232 ,H511282 ,H511339 ,H511398 ,H511448 ,H511491 H511078 ,H511130 ,H511184 ,H511233 ,H511284 ,H511340 ,H511399 ,H511449 ,H511492 H511080 ,H511131 ,H511185 ,H511234 ,H511285 ,H511341 ,H511400 ,H511451 ,H511493 H511081 ,H511132 ,H511186 ,H511235 ,H511286 ,H511342 ,H511401 ,H511452 ,H511494 H511082 ,H511134 ,H511187 ,H511236 ,H511287 ,H511343 ,H511402 ,H511453 ,H511496 H511083 ,H511135 ,H511188 ,H511237 ,H511288 ,H511344 ,H511403 ,H511455 ,H511497 H511084 ,H511137 ,H511189 ,H511238 ,H511289 ,H511346 ,H511404 ,H511456 ,H511498 H511085 ,H511138 ,H511190 ,H511239 ,H511290 ,H511347 ,H511405 ,H511457 ,H511500 H511086 ,H511139 ,H511191 ,H511240 ,H511291 ,H511348 ,H511406 ,H511458 ,H511501 H511090 ,H511140 ,H511192 ,H511241 ,H511292 ,H511349 ,H511408 ,H511459 ,H511503 H511091 ,H511142 ,H511193 ,H511242 ,H511293 ,H511350 ,H511409 ,H511460 ,H511504 H511092 ,H511143 ,H511195 ,H511243 ,H511294 ,H511351 ,H511411 ,H511461 ,H511505 H511093 ,H511144 ,H511196 ,H511245 ,H511295 ,H511352 ,H511412 ,H511462 ,H511506 H511094 ,H511145 ,H511197 ,H511246 ,H511296 ,H511353 ,H511414 ,H511463 ,H511507 H511095 ,H511146 ,H511199 ,H511247 ,H511300 ,H511354 ,H511415 ,H511464 ,H511508 H511096 ,H511147 ,H511200 ,H511248 ,H511302 ,H511358 ,H511416 ,H511465 ,H511509 H511098 ,H511149 ,H511202 ,H511249 ,H511303 ,H511359 ,H511417 ,H511466 ,H511510 H511099 ,H511150 ,H511203 ,H511250 ,H511305 ,H511360 ,H511418 ,H511467 ,H511511 H511100 ,H511151 ,H511204 ,H511251 ,H511306 ,H511361 ,H511419 ,H511468 ,H511514 H511102 ,H511152 ,H511205 ,H511253 ,H511309 ,H511362 ,H511420 ,H511469 ,H511515 H511103 ,H511153 ,H511206 ,H511255 ,H511311 ,H511363 ,H511424 ,H511470 ,H511516 H511104 ,H511155 ,H511207 ,H511256 ,H511312 ,H511364 ,H511425 ,H511471 ,H511517 H511105 ,H511156 ,H511208 ,H511257 ,H511313 ,H511368 ,H511426 ,H511472 ,H511518 H511106 ,H511157 ,H511209 ,H511258 ,H511314 ,H511369 ,H511427 ,H511473 ,H511520 H511108 ,H511160 ,H511210 ,H511259 ,H511316 ,H511370 ,H511428 ,H511474 ,H511521 H511109 ,H511161 ,H511211 ,H511260 ,H511320 ,H511372 ,H511429 ,H511475 ,H511522 H511110 ,H511162 ,H511212 ,H511261 ,H511321 ,H511374 ,H511430 ,H511476 ,H511523 H511111 ,H511163 ,H511213 ,H511262 ,H511322 ,H511375 ,H511431 ,H511477 ,H511524 H511112 ,H511164 ,H511214 ,H511264 ,H511323 ,H511376 ,H511432 ,H511478 ,H511525 H511113 ,H511165 ,H511218 ,H511265 ,H511326 ,H511377 ,H511433 ,H511479 ,H511526 H511528 ,H511574 ,H511623 ,H511675 ,H511730 ,H511781 ,H511838 ,H511888 ,H511946 H511529 ,H511575 ,H511624 ,H511676 ,H511732 ,H511784 ,H511840 ,H511890 ,H511949 H511530 ,H511576 ,H511626 ,H511677 ,H511733 ,H511785 ,H511841 ,H511891 ,H511950 H511531 ,H511577 ,H511627 ,H511678 ,H511734 ,H511787 ,H511842 ,H511892 ,H511951 H511532 ,H511578 ,H511629 ,H511679 ,H511735 ,H511790 ,H511844 ,H511893 ,H511952 H511533 ,H511579 ,H511630 ,H511681 ,H511736 ,H511792 ,H511845 ,H511894 ,H511953 H511534 ,H511580 ,H511632 ,H511682 ,H511737 ,H511796 ,H511846 ,H511895 ,H511954 H511535 ,H511582 ,H511633 ,H511683 ,H511738 ,H511797 ,H511847 ,H511896 ,H511955 H511536 ,H511583 ,H511634 ,H511686 ,H511742 ,H511798 ,H511848 ,H511900 ,H511956 H511537 ,H511584 ,H511635 ,H511687 ,H511743 ,H511800 ,H511849 ,H511901 ,H511957 H511538 ,H511585 ,H511636 ,H511688 ,H511745 ,H511801 ,H511850 ,H511902 ,H511958 H511540 ,H511586 ,H511637 ,H511690 ,H511746 ,H511802 ,H511851 ,H511903 ,H511959 H511541 ,H511587 ,H511638 ,H511691 ,H511748 ,H511803 ,H511852 ,H511905 ,H511960 H511542 ,H511589 ,H511639 ,H511694 ,H511750 ,H511804 ,H511853 ,H511906 ,H511961 H511543 ,H511590 ,H511643 ,H511695 ,H511752 ,H511805 ,H511854 ,H511907 ,H511962 H511544 ,H511591 ,H511644 ,H511696 ,H511753 ,H511806 ,H511855 ,H511908 ,H511963 H511545 ,H511592 ,H511645 ,H511697 ,H511754 ,H511807 ,H511856 ,H511910 ,H511964 H511546 ,H511593 ,H511646 ,H511698 ,H511755 ,H511808 ,H511857 ,H511915 ,H511965 H511547 ,H511594 ,H511647 ,H511699 ,H511756 ,H511809 ,H511859 ,H511916 ,H511966 H511548 ,H511595 ,H511648 ,H511700 ,H511758 ,H511811 ,H511860 ,H511917 ,H511967 H511549 ,H511596 ,H511650 ,H511701 ,H511760 ,H511813 ,H511861 ,H511918 ,H511968 H511550 ,H511597 ,H511651 ,H511702 ,H511761 ,H511814 ,H511863 ,H511922 ,H511969 H511551 ,H511599 ,H511652 ,H511703 ,H511762 ,H511815 ,H511864 ,H511923 ,H511970 H511552 ,H511600 ,H511653 ,H511704 ,H511763 ,H511816 ,H511865 ,H511924 ,H511971 H511553 ,H511601 ,H511654 ,H511705 ,H511764 ,H511821 ,H511868 ,H511925 ,H511972 H511554 ,H511605 ,H511655 ,H511707 ,H511765 ,H511822 ,H511869 ,H511926 ,H511974 H511555 ,H511606 ,H511656 ,H511711 ,H511767 ,H511823 ,H511870 ,H511927 ,H511975 H511556 ,H511607 ,H511657 ,H511712 ,H511768 ,H511824 ,H511872 ,H511928 ,H511976 H511557 ,H511608 ,H511658 ,H511713 ,H511769 ,H511825 ,H511873 ,H511930 ,H511977 H511558 ,H511609 ,H511659 ,H511714 ,H511770 ,H511826 ,H511874 ,H511931 ,H511978 H511559 ,H511610 ,H511660 ,H511715 ,H511771 ,H511827 ,H511875 ,H511932 ,H511980 H511561 ,H511611 ,H511662 ,H511716 ,H511772 ,H511828 ,H511876 ,H511933 ,H511982 H511563 ,H511612 ,H511663 ,H511719 ,H511773 ,H511829 ,H511878 ,H511935 ,H511983 H511564 ,H511613 ,H511664 ,H511720 ,H511774 ,H511830 ,H511879 ,H511936 ,H511984 H511565 ,H511614 ,H511666 ,H511722 ,H511775 ,H511831 ,H511880 ,H511937 ,H511985 H511566 ,H511615 ,H511667 ,H511723 ,H511776 ,H511832 ,H511881 ,H511938 ,H511987 H511569 ,H511616 ,H511670 ,H511724 ,H511777 ,H511833 ,H511882 ,H511939 ,H511990 H511570 ,H511619 ,H511671 ,H511726 ,H511778 ,H511835 ,H511883 ,H511943 ,H511991 H511572 ,H511620 ,H511673 ,H511727 ,H511779 ,H511836 ,H511884 ,H511944 ,H511992 H511573 ,H511622 ,H511674 ,H511728 ,H511780 ,H511837 ,H511886 ,H511945 ,H511993 H511994 ,H512044 ,H512094 ,H512156 ,H512224 ,H512277 ,H512328 ,H512386 ,H512436 H511995 ,H512046 ,H512095 ,H512157 ,H512226 ,H512278 ,H512329 ,H512387 ,H512437 H511996 ,H512047 ,H512097 ,H512161 ,H512227 ,H512279 ,H512331 ,H512388 ,H512438 H511998 ,H512048 ,H512100 ,H512163 ,H512228 ,H512280 ,H512332 ,H512389 ,H512439 H511999 ,H512049 ,H512103 ,H512165 ,H512229 ,H512282 ,H512334 ,H512390 ,H512440 H512000 ,H512050 ,H512104 ,H512166 ,H512231 ,H512283 ,H512335 ,H512391 ,H512441 H512001 ,H512051 ,H512107 ,H512168 ,H512234 ,H512284 ,H512337 ,H512392 ,H512445 H512002 ,H512052 ,H512108 ,H512169 ,H512236 ,H512287 ,H512339 ,H512394 ,H512446 H512003 ,H512053 ,H512111 ,H512171 ,H512237 ,H512288 ,H512340 ,H512396 ,H512449 H512004 ,H512056 ,H512112 ,H512180 ,H512238 ,H512289 ,H512341 ,H512398 ,H512450 H512005 ,H512057 ,H512113 ,H512182 ,H512239 ,H512290 ,H512342 ,H512399 ,H512451 H512006 ,H512058 ,H512114 ,H512184 ,H512240 ,H512291 ,H512344 ,H512400 ,H512452 H512007 ,H512059 ,H512118 ,H512185 ,H512241 ,H512292 ,H512345 ,H512401 ,H512453 H512008 ,H512060 ,H512120 ,H512186 ,H512242 ,H512293 ,H512347 ,H512402 ,H512454 H512010 ,H512062 ,H512121 ,H512187 ,H512243 ,H512294 ,H512349 ,H512403 ,H512455 H512011 ,H512063 ,H512122 ,H512188 ,H512244 ,H512296 ,H512350 ,H512404 ,H512459 H512012 ,H512064 ,H512124 ,H512190 ,H512245 ,H512297 ,H512354 ,H512405 ,H512460 H512013 ,H512065 ,H512125 ,H512191 ,H512246 ,H512298 ,H512355 ,H512406 ,H512461 H512014 ,H512066 ,H512126 ,H512194 ,H512247 ,H512299 ,H512356 ,H512407 ,H512462 H512016 ,H512067 ,H512127 ,H512197 ,H512248 ,H512300 ,H512357 ,H512408 ,H512463 H512017 ,H512068 ,H512128 ,H512198 ,H512249 ,H512302 ,H512358 ,H512410 ,H512464 H512019 ,H512069 ,H512131 ,H512199 ,H512250 ,H512303 ,H512360 ,H512411 ,H512465 H512020 ,H512070 ,H512132 ,H512200 ,H512252 ,H512304 ,H512361 ,H512414 ,H512466 H512021 ,H512071 ,H512133 ,H512201 ,H512253 ,H512305 ,H512362 ,H512415 ,H512467 H512022 ,H512072 ,H512134 ,H512202 ,H512254 ,H512306 ,H512364 ,H512416 ,H512468 H512024 ,H512073 ,H512135 ,H512203 ,H512255 ,H512307 ,H512365 ,H512417 ,H512469 H512028 ,H512076 ,H512136 ,H512204 ,H512256 ,H512308 ,H512366 ,H512418 ,H512470 H512029 ,H512077 ,H512137 ,H512205 ,H512258 ,H512309 ,H512367 ,H512419 ,H512471 H512031 ,H512078 ,H512138 ,H512206 ,H512259 ,H512310 ,H512368 ,H512420 ,H512475 H512032 ,H512079 ,H512139 ,H512207 ,H512260 ,H512311 ,H512369 ,H512421 ,H512477 H512033 ,H512081 ,H512140 ,H512208 ,H512262 ,H512314 ,H512371 ,H512423 ,H512479 H512035 ,H512082 ,H512142 ,H512210 ,H512264 ,H512315 ,H512372 ,H512425 ,H512480 H512036 ,H512083 ,H512144 ,H512211 ,H512265 ,H512317 ,H512375 ,H512426 ,H512483 H512037 ,H512084 ,H512145 ,H512212 ,H512266 ,H512318 ,H512376 ,H512427 ,H512485 H512038 ,H512085 ,H512146 ,H512213 ,H512268 ,H512320 ,H512377 ,H512429 ,H512486 H512039 ,H512086 ,H512148 ,H512214 ,H512270 ,H512321 ,H512378 ,H512431 ,H512487 H512040 ,H512087 ,H512149 ,H512218 ,H512271 ,H512322 ,H512379 ,H512432 ,H512489 H512041 ,H512088 ,H512150 ,H512219 ,H512273 ,H512324 ,H512380 ,H512433 ,H512490 H512042 ,H512090 ,H512154 ,H512222 ,H512274 ,H512325 ,H512381 ,H512434 ,H512491 H512043 ,H512091 ,H512155 ,H512223 ,H512275 ,H512327 ,H512385 ,H512435 ,H512492 H512495 ,H512551 ,H512610 ,H512671 ,H512720 ,H512782 ,H512839 ,H512888 ,H512954 H512496 ,H512552 ,H512611 ,H512674 ,H512723 ,H512783 ,H512841 ,H512889 ,H512955 H512497 ,H512553 ,H512612 ,H512675 ,H512724 ,H512784 ,H512842 ,H512890 ,H512956 H512498 ,H512554 ,H512613 ,H512676 ,H512725 ,H512785 ,H512843 ,H512892 ,H512957 H512502 ,H512555 ,H512614 ,H512677 ,H512726 ,H512786 ,H512845 ,H512893 ,H512958 H512503 ,H512556 ,H512616 ,H512678 ,H512727 ,H512787 ,H512846 ,H512896 ,H512959 H512504 ,H512559 ,H512617 ,H512679 ,H512729 ,H512789 ,H512847 ,H512899 ,H512960 H512506 ,H512560 ,H512618 ,H512680 ,H512730 ,H512790 ,H512848 ,H512902 ,H512961 H512507 ,H512561 ,H512619 ,H512681 ,H512731 ,H512791 ,H512849 ,H512907 ,H512962 H512509 ,H512563 ,H512620 ,H512682 ,H512732 ,H512795 ,H512850 ,H512908 ,H512964 H512511 ,H512567 ,H512621 ,H512683 ,H512734 ,H512799 ,H512852 ,H512909 ,H512966 H512513 ,H512568 ,H512622 ,H512684 ,H512735 ,H512801 ,H512853 ,H512912 ,H512967 H512515 ,H512569 ,H512623 ,H512685 ,H512737 ,H512802 ,H512854 ,H512913 ,H512969 H512516 ,H512570 ,H512624 ,H512686 ,H512738 ,H512803 ,H512855 ,H512914 ,H512970 H512517 ,H512573 ,H512632 ,H512691 ,H512739 ,H512804 ,H512856 ,H512916 ,H512971 H512518 ,H512574 ,H512633 ,H512692 ,H512740 ,H512809 ,H512858 ,H512917 ,H512972 H512520 ,H512575 ,H512634 ,H512693 ,H512741 ,H512810 ,H512859 ,H512918 ,H512973 H512521 ,H512576 ,H512636 ,H512694 ,H512742 ,H512811 ,H512860 ,H512920 ,H512974 H512522 ,H512577 ,H512637 ,H512695 ,H512743 ,H512812 ,H512861 ,H512921 ,H512977 H512524 ,H512578 ,H512642 ,H512697 ,H512744 ,H512813 ,H512862 ,H512924 ,H512978 H512525 ,H512579 ,H512643 ,H512698 ,H512745 ,H512814 ,H512863 ,H512926 ,H512979 H512526 ,H512581 ,H512644 ,H512700 ,H512747 ,H512815 ,H512864 ,H512928 ,H512980 H512527 ,H512582 ,H512645 ,H512701 ,H512749 ,H512816 ,H512865 ,H512929 ,H512982 H512532 ,H512583 ,H512646 ,H512702 ,H512750 ,H512817 ,H512866 ,H512930 ,H512984 H512533 ,H512585 ,H512647 ,H512703 ,H512751 ,H512818 ,H512867 ,H512931 ,H512985 H512534 ,H512586 ,H512650 ,H512704 ,H512757 ,H512819 ,H512868 ,H512932 ,H512987 H512535 ,H512587 ,H512652 ,H512705 ,H512758 ,H512821 ,H512869 ,H512933 ,H512988 H512536 ,H512588 ,H512657 ,H512706 ,H512761 ,H512822 ,H512873 ,H512934 ,H512989 H512537 ,H512590 ,H512658 ,H512707 ,H512763 ,H512824 ,H512874 ,H512935 ,H512990 H512538 ,H512591 ,H512659 ,H512708 ,H512767 ,H512827 ,H512875 ,H512936 ,H512991 H512539 ,H512592 ,H512661 ,H512709 ,H512769 ,H512828 ,H512878 ,H512938 ,H512992 H512540 ,H512596 ,H512662 ,H512710 ,H512770 ,H512829 ,H512879 ,H512939 ,H512993 H512541 ,H512597 ,H512663 ,H512711 ,H512771 ,H512831 ,H512880 ,H512940 ,H512994 H512542 ,H512598 ,H512664 ,H512712 ,H512773 ,H512832 ,H512881 ,H512941 ,H512995 H512543 ,H512600 ,H512665 ,H512713 ,H512774 ,H512833 ,H512882 ,H512943 ,H512996 H512545 ,H512602 ,H512666 ,H512714 ,H512776 ,H512834 ,H512883 ,H512944 ,H512997 H512546 ,H512603 ,H512667 ,H512716 ,H512777 ,H512835 ,H512884 ,H512945 ,H512998 H512547 ,H512604 ,H512668 ,H512717 ,H512778 ,H512836 ,H512885 ,H512946 ,H512999 H512548 ,H512607 ,H512669 ,H512718 ,H512779 ,H512837 ,H512886 ,H512948 ,H513000 H512550 ,H512608 ,H512670 ,H512719 ,H512781 ,H512838 ,H512887 ,H512953 ,H513001 H513002 ,H513050 ,H513107 ,H513156 ,H513211 ,H513265 ,H513312 ,H513373 ,H513428 H513004 ,H513051 ,H513108 ,H513157 ,H513212 ,H513266 ,H513315 ,H513374 ,H513429 H513005 ,H513052 ,H513109 ,H513158 ,H513213 ,H513267 ,H513317 ,H513375 ,H513430 H513006 ,H513054 ,H513111 ,H513162 ,H513215 ,H513268 ,H513321 ,H513376 ,H513431 H513008 ,H513055 ,H513112 ,H513163 ,H513216 ,H513269 ,H513322 ,H513378 ,H513432 H513009 ,H513056 ,H513113 ,H513164 ,H513217 ,H513272 ,H513323 ,H513379 ,H513433 H513010 ,H513058 ,H513114 ,H513165 ,H513220 ,H513273 ,H513325 ,H513380 ,H513434 H513011 ,H513059 ,H513115 ,H513166 ,H513222 ,H513274 ,H513326 ,H513381 ,H513435 H513012 ,H513060 ,H513116 ,H513167 ,H513223 ,H513276 ,H513327 ,H513382 ,H513437 H513014 ,H513062 ,H513117 ,H513170 ,H513224 ,H513277 ,H513329 ,H513383 ,H513438 H513015 ,H513064 ,H513118 ,H513171 ,H513225 ,H513278 ,H513330 ,H513384 ,H513439 H513016 ,H513065 ,H513119 ,H513172 ,H513226 ,H513279 ,H513331 ,H513385 ,H513440 H513018 ,H513066 ,H513120 ,H513174 ,H513227 ,H513280 ,H513338 ,H513386 ,H513441 H513019 ,H513067 ,H513121 ,H513176 ,H513228 ,H513281 ,H513339 ,H513387 ,H513443 H513020 ,H513069 ,H513122 ,H513180 ,H513229 ,H513283 ,H513340 ,H513388 ,H513450 H513021 ,H513070 ,H513123 ,H513181 ,H513231 ,H513285 ,H513341 ,H513389 ,H513451 H513022 ,H513073 ,H513124 ,H513182 ,H513233 ,H513287 ,H513342 ,H513390 ,H513452 H513023 ,H513078 ,H513125 ,H513183 ,H513234 ,H513289 ,H513344 ,H513393 ,H513453 H513024 ,H513079 ,H513127 ,H513184 ,H513236 ,H513290 ,H513345 ,H513394 ,H513454 H513025 ,H513080 ,H513128 ,H513185 ,H513237 ,H513291 ,H513347 ,H513395 ,H513455 H513026 ,H513081 ,H513129 ,H513186 ,H513239 ,H513292 ,H513349 ,H513396 ,H513456 H513027 ,H513082 ,H513130 ,H513188 ,H513240 ,H513293 ,H513350 ,H513397 ,H513457 H513028 ,H513083 ,H513131 ,H513189 ,H513241 ,H513294 ,H513351 ,H513398 ,H513458 H513029 ,H513084 ,H513132 ,H513190 ,H513242 ,H513295 ,H513353 ,H513399 ,H513459 H513030 ,H513085 ,H513133 ,H513191 ,H513244 ,H513296 ,H513354 ,H513400 ,H513460 H513031 ,H513086 ,H513134 ,H513192 ,H513245 ,H513297 ,H513355 ,H513401 ,H513462 H513032 ,H513087 ,H513135 ,H513193 ,H513246 ,H513298 ,H513356 ,H513402 ,H513464 H513033 ,H513089 ,H513139 ,H513195 ,H513248 ,H513299 ,H513357 ,H513403 ,H513465 H513034 ,H513091 ,H513142 ,H513196 ,H513249 ,H513300 ,H513358 ,H513405 ,H513466 H513035 ,H513092 ,H513143 ,H513197 ,H513250 ,H513301 ,H513359 ,H513407 ,H513467 H513036 ,H513094 ,H513144 ,H513199 ,H513251 ,H513302 ,H513360 ,H513413 ,H513470 H513037 ,H513095 ,H513145 ,H513200 ,H513253 ,H513303 ,H513361 ,H513414 ,H513471 H513039 ,H513096 ,H513146 ,H513201 ,H513254 ,H513304 ,H513362 ,H513419 ,H513472 H513042 ,H513097 ,H513147 ,H513202 ,H513255 ,H513305 ,H513363 ,H513420 ,H513474 H513043 ,H513098 ,H513149 ,H513203 ,H513256 ,H513306 ,H513365 ,H513421 ,H513475 H513045 ,H513101 ,H513150 ,H513204 ,H513257 ,H513307 ,H513366 ,H513422 ,H513477 H513046 ,H513103 ,H513151 ,H513205 ,H513260 ,H513308 ,H513367 ,H513424 ,H513478 H513047 ,H513104 ,H513152 ,H513208 ,H513261 ,H513309 ,H513369 ,H513425 ,H513479 H513048 ,H513105 ,H513153 ,H513209 ,H513262 ,H513310 ,H513371 ,H513426 ,H513480 H513049 ,H513106 ,H513154 ,H513210 ,H513264 ,H513311 ,H513372 ,H513427 ,H513483 H513485 ,H513542 ,H513595 ,H513647 ,H513699 ,H513758 ,H513816 ,H513869 ,H513923 H513492 ,H513543 ,H513596 ,H513648 ,H513700 ,H513759 ,H513817 ,H513870 ,H513924 H513493 ,H513545 ,H513597 ,H513649 ,H513701 ,H513761 ,H513818 ,H513871 ,H513925 H513494 ,H513546 ,H513598 ,H513650 ,H513704 ,H513763 ,H513819 ,H513872 ,H513926 H513495 ,H513547 ,H513601 ,H513651 ,H513706 ,H513764 ,H513821 ,H513873 ,H513927 H513496 ,H513548 ,H513602 ,H513653 ,H513707 ,H513766 ,H513822 ,H513875 ,H513929 H513497 ,H513549 ,H513603 ,H513654 ,H513709 ,H513767 ,H513823 ,H513876 ,H513930 H513498 ,H513550 ,H513604 ,H513656 ,H513710 ,H513769 ,H513825 ,H513877 ,H513931 H513499 ,H513551 ,H513605 ,H513657 ,H513712 ,H513770 ,H513826 ,H513878 ,H513932 H513501 ,H513552 ,H513606 ,H513658 ,H513713 ,H513771 ,H513827 ,H513880 ,H513933 H513502 ,H513553 ,H513607 ,H513659 ,H513714 ,H513772 ,H513828 ,H513881 ,H513934 H513503 ,H513554 ,H513611 ,H513660 ,H513718 ,H513773 ,H513829 ,H513887 ,H513935 H513504 ,H513555 ,H513612 ,H513661 ,H513719 ,H513775 ,H513830 ,H513888 ,H513936 H513506 ,H513556 ,H513613 ,H513663 ,H513720 ,H513776 ,H513831 ,H513889 ,H513937 H513507 ,H513558 ,H513614 ,H513664 ,H513721 ,H513777 ,H513832 ,H513890 ,H513941 H513508 ,H513559 ,H513615 ,H513665 ,H513722 ,H513778 ,H513837 ,H513891 ,H513942 H513509 ,H513561 ,H513617 ,H513667 ,H513723 ,H513779 ,H513838 ,H513892 ,H513943 H513511 ,H513562 ,H513619 ,H513668 ,H513725 ,H513780 ,H513839 ,H513893 ,H513944 H513514 ,H513563 ,H513621 ,H513669 ,H513730 ,H513781 ,H513840 ,H513894 ,H513945 H513515 ,H513565 ,H513622 ,H513670 ,H513731 ,H513782 ,H513841 ,H513895 ,H513947 H513516 ,H513566 ,H513624 ,H513672 ,H513733 ,H513783 ,H513842 ,H513896 ,H513949 H513517 ,H513567 ,H513625 ,H513673 ,H513735 ,H513784 ,H513843 ,H513898 ,H513952 H513518 ,H513568 ,H513626 ,H513674 ,H513738 ,H513785 ,H513844 ,H513899 ,H513953 H513519 ,H513569 ,H513627 ,H513675 ,H513739 ,H513786 ,H513845 ,H513901 ,H513954 H513520 ,H513570 ,H513628 ,H513676 ,H513740 ,H513787 ,H513846 ,H513902 ,H513955 H513521 ,H513571 ,H513629 ,H513677 ,H513741 ,H513788 ,H513847 ,H513904 ,H513956 H513523 ,H513573 ,H513630 ,H513679 ,H513742 ,H513789 ,H513850 ,H513906 ,H513957 H513524 ,H513574 ,H513631 ,H513680 ,H513743 ,H513790 ,H513853 ,H513907 ,H513958 H513525 ,H513575 ,H513632 ,H513681 ,H513744 ,H513791 ,H513854 ,H513908 ,H513959 H513526 ,H513579 ,H513633 ,H513682 ,H513745 ,H513792 ,H513855 ,H513909 ,H513960 H513527 ,H513580 ,H513634 ,H513683 ,H513746 ,H513793 ,H513856 ,H513910 ,H513961 H513528 ,H513581 ,H513635 ,H513684 ,H513747 ,H513794 ,H513857 ,H513911 ,H513962 H513529 ,H513582 ,H513636 ,H513686 ,H513748 ,H513795 ,H513858 ,H513912 ,H513963 H513530 ,H513583 ,H513638 ,H513687 ,H513749 ,H513797 ,H513859 ,H513913 ,H513964 H513533 ,H513584 ,H513639 ,H513690 ,H513750 ,H513800 ,H513860 ,H513914 ,H513965 H513534 ,H513585 ,H513640 ,H513691 ,H513751 ,H513801 ,H513861 ,H513916 ,H513966 H513535 ,H513586 ,H513641 ,H513693 ,H513753 ,H513802 ,H513862 ,H513917 ,H513967 H513536 ,H513589 ,H513642 ,H513695 ,H513755 ,H513805 ,H513863 ,H513919 ,H513968 H513539 ,H513591 ,H513645 ,H513696 ,H513756 ,H513807 ,H513866 ,H513921 ,H513969 H513541 ,H513594 ,H513646 ,H513697 ,H513757 ,H513815 ,H513867 ,H513922 ,H513970 H513971 ,H514018 ,H514071 ,H514124 ,H514170 ,H514221 ,H514268 ,H514321 ,H514366 H513972 ,H514019 ,H514072 ,H514125 ,H514171 ,H514222 ,H514269 ,H514322 ,H514367 H513973 ,H514020 ,H514073 ,H514126 ,H514172 ,H514223 ,H514270 ,H514323 ,H514369 H513974 ,H514021 ,H514075 ,H514127 ,H514174 ,H514224 ,H514271 ,H514325 ,H514372 H513975 ,H514022 ,H514077 ,H514128 ,H514175 ,H514225 ,H514272 ,H514326 ,H514373 H513976 ,H514023 ,H514078 ,H514129 ,H514179 ,H514226 ,H514273 ,H514329 ,H514374 H513977 ,H514024 ,H514079 ,H514130 ,H514180 ,H514227 ,H514274 ,H514330 ,H514375 H513978 ,H514025 ,H514080 ,H514133 ,H514181 ,H514228 ,H514276 ,H514331 ,H514376 H513979 ,H514026 ,H514081 ,H514134 ,H514182 ,H514229 ,H514277 ,H514332 ,H514377 H513980 ,H514028 ,H514082 ,H514135 ,H514183 ,H514230 ,H514278 ,H514333 ,H514378 H513981 ,H514029 ,H514083 ,H514136 ,H514184 ,H514231 ,H514279 ,H514334 ,H514379 H513982 ,H514030 ,H514084 ,H514137 ,H514185 ,H514232 ,H514280 ,H514335 ,H514380 H513983 ,H514031 ,H514085 ,H514138 ,H514186 ,H514233 ,H514281 ,H514336 ,H514381 H513984 ,H514032 ,H514086 ,H514140 ,H514187 ,H514234 ,H514282 ,H514337 ,H514382 H513985 ,H514033 ,H514087 ,H514141 ,H514188 ,H514235 ,H514283 ,H514338 ,H514383 H513986 ,H514034 ,H514089 ,H514142 ,H514189 ,H514236 ,H514284 ,H514339 ,H514385 H513987 ,H514039 ,H514090 ,H514143 ,H514190 ,H514237 ,H514285 ,H514340 ,H514386 H513988 ,H514041 ,H514091 ,H514144 ,H514191 ,H514238 ,H514286 ,H514341 ,H514387 H513989 ,H514042 ,H514092 ,H514145 ,H514192 ,H514239 ,H514289 ,H514342 ,H514388 H513990 ,H514047 ,H514093 ,H514146 ,H514193 ,H514240 ,H514291 ,H514343 ,H514389 H513991 ,H514048 ,H514094 ,H514147 ,H514194 ,H514241 ,H514292 ,H514344 ,H514390 H513992 ,H514050 ,H514095 ,H514148 ,H514196 ,H514243 ,H514293 ,H514345 ,H514391 H513993 ,H514051 ,H514096 ,H514149 ,H514197 ,H514244 ,H514296 ,H514346 ,H514392 H513995 ,H514052 ,H514097 ,H514152 ,H514198 ,H514245 ,H514297 ,H514347 ,H514395 H513999 ,H514053 ,H514098 ,H514153 ,H514200 ,H514246 ,H514300 ,H514348 ,H514396 H514000 ,H514054 ,H514099 ,H514154 ,H514202 ,H514247 ,H514301 ,H514349 ,H514397 H514001 ,H514055 ,H514100 ,H514155 ,H514203 ,H514248 ,H514303 ,H514350 ,H514398 H514002 ,H514056 ,H514101 ,H514156 ,H514204 ,H514249 ,H514304 ,H514351 ,H514399 H514003 ,H514058 ,H514102 ,H514157 ,H514209 ,H514250 ,H514305 ,H514352 ,H514400 H514004 ,H514059 ,H514103 ,H514158 ,H514210 ,H514253 ,H514306 ,H514353 ,H514402
H514005 ,H514060 ,H514104 ,H514159 ,H514211 ,H514254 ,H514307 ,H514355 ,H514403 H514007 ,H514062 ,H514108 ,H514160 ,H514212 ,H514256 ,H514308 ,H514357 ,H514404 H514008 ,H514063 ,H514110 ,H514162 ,H514213 ,H514257 ,H514310 ,H514358 ,H514405 H514009 ,H514064 ,H514112 ,H514163 ,H514214 ,H514258 ,H514311 ,H514359 ,H514406 H514010 ,H514065 ,H514114 ,H514164 ,H514215 ,H514260 ,H514313 ,H514360 ,H514407 H514011 ,H514066 ,H514115 ,H514165 ,H514216 ,H514261 ,H514314 ,H514361 ,H514408 H514013 ,H514067 ,H514119 ,H514166 ,H514217 ,H514262 ,H514315 ,H514362 ,H514409 H514015 ,H514068 ,H514121 ,H514167 ,H514218 ,H514263 ,H514316 ,H514363 ,H514410 H514016 ,H514069 ,H514122 ,H514168 ,H514219 ,H514265 ,H514318 ,H514364 ,H514411 H514017 ,H514070 ,H514123 ,H514169 ,H514220 ,H514266 ,H514319 ,H514365 ,H514412 H514413 ,H514472 ,H514527 ,H514580 ,H514626 ,H514670 ,H514768 ,H515035 ,H515081 H514414 ,H514474 ,H514528 ,H514581 ,H514627 ,H514671 ,H514770 ,H515036 ,H515082 H514415 ,H514475 ,H514529 ,H514582 ,H514628 ,H514672 ,H514779 ,H515037 ,H515085 H514416 ,H514478 ,H514530 ,H514583 ,H514629 ,H514673 ,H514780 ,H515038 ,H515086 H514417 ,H514479 ,H514532 ,H514584 ,H514631 ,H514675 ,H514784 ,H515039 ,H515087 H514420 ,H514480 ,H514533 ,H514586 ,H514632 ,H514676 ,H514785 ,H515040 ,H515089 H514421 ,H514483 ,H514534 ,H514588 ,H514633 ,H514678 ,H514787 ,H515041 ,H515090 H514422 ,H514484 ,H514539 ,H514589 ,H514634 ,H514679 ,H514790 ,H515042 ,H515092 H514424 ,H514485 ,H514540 ,H514590 ,H514635 ,H514683 ,H514791 ,H515043 ,H515093 H514425 ,H514486 ,H514541 ,H514591 ,H514636 ,H514685 ,H514793 ,H515044 ,H515094 H514426 ,H514487 ,H514542 ,H514592 ,H514637 ,H514687 ,H514858 ,H515045 ,H515095 H514427 ,H514488 ,H514543 ,H514596 ,H514638 ,H514688 ,H515000 ,H515046 ,H515096 H514428 ,H514489 ,H514544 ,H514597 ,H514639 ,H514689 ,H515004 ,H515047 ,H515097 H514429 ,H514490 ,H514545 ,H514598 ,H514640 ,H514690 ,H515005 ,H515048 ,H515098 H514430 ,H514491 ,H514546 ,H514599 ,H514641 ,H514691 ,H515006 ,H515049 ,H515099 H514433 ,H514493 ,H514547 ,H514600 ,H514642 ,H514692 ,H515007 ,H515050 ,H515100 H514435 ,H514496 ,H514551 ,H514601 ,H514643 ,H514693 ,H515008 ,H515051 ,H515101 H514436 ,H514498 ,H514552 ,H514602 ,H514644 ,H514696 ,H515009 ,H515052 ,H515102 H514437 ,H514499 ,H514553 ,H514603 ,H514645 ,H514698 ,H515011 ,H515054 ,H515103 H514438 ,H514501 ,H514554 ,H514604 ,H514647 ,H514699 ,H515012 ,H515055 ,H515104 H514439 ,H514502 ,H514555 ,H514605 ,H514648 ,H514703 ,H515013 ,H515056 ,H515107 H514440 ,H514504 ,H514556 ,H514606 ,H514649 ,H514705 ,H515014 ,H515057 ,H515108 H514441 ,H514506 ,H514557 ,H514607 ,H514651 ,H514708 ,H515015 ,H515060 ,H515109 H514442 ,H514507 ,H514559 ,H514608 ,H514652 ,H514711 ,H515016 ,H515062 ,H515110 H514444 ,H514508 ,H514560 ,H514609 ,H514653 ,H514715 ,H515017 ,H515063 ,H515111 H514455 ,H514510 ,H514561 ,H514611 ,H514654 ,H514716 ,H515018 ,H515064 ,H515112 H514456 ,H514511 ,H514562 ,H514612 ,H514655 ,H514722 ,H515019 ,H515065 ,H515113 H514457 ,H514513 ,H514563 ,H514613 ,H514656 ,H514724 ,H515020 ,H515067 ,H515114 H514460 ,H514514 ,H514564 ,H514614 ,H514657 ,H514732 ,H515021 ,H515068 ,H515115 H514461 ,H514515 ,H514565 ,H514615 ,H514658 ,H514734 ,H515022 ,H515069 ,H515116 H514462 ,H514516 ,H514566 ,H514616 ,H514660 ,H514735 ,H515023 ,H515070 ,H515117 H514463 ,H514517 ,H514567 ,H514617 ,H514661 ,H514736 ,H515025 ,H515071 ,H515119 H514464 ,H514518 ,H514568 ,H514618 ,H514662 ,H514750 ,H515026 ,H515072 ,H515120 H514465 ,H514520 ,H514571 ,H514619 ,H514663 ,H514752 ,H515027 ,H515073 ,H515121 H514466 ,H514521 ,H514574 ,H514620 ,H514664 ,H514753 ,H515029 ,H515075 ,H515122 H514467 ,H514522 ,H514575 ,H514621 ,H514665 ,H514754 ,H515030 ,H515076 ,H515125 H514468 ,H514523 ,H514576 ,H514622 ,H514666 ,H514756 ,H515031 ,H515077 ,H515126 H514469 ,H514524 ,H514577 ,H514623 ,H514667 ,H514763 ,H515032 ,H515078 ,H515127 H514470 ,H514525 ,H514578 ,H514624 ,H514668 ,H514765 ,H515033 ,H515079 ,H515128 H514471 ,H514526 ,H514579 ,H514625 ,H514669 ,H514766 ,H515034 ,H515080 ,H515129, H515131 ,H515203 H515133 ,H515204 H515134 ,H515205 H515135 ,H515206 H515136 ,H515207 H515137 ,H515208 H515138 ,H515209 H515139 ,H515210 H515140 ,H515213 H515141 ,H515214 H515145 ,H515215 H515148 ,H515218 H515149 ,H515219 H515150 ,H515221 H515152 ,H515223 H515155 ,H515225 H515156 ,H515228 H515157 ,H515231 H515159 ,H515236 H515160 ,H515237 H515162 ,H515240 H515163 ,H515242 H515165 ,H515243 H515171 ,H515249 H515175 ,H515256 H515178 ,H515259 H515179 ,H515262 H515181 ,H515266 H515184 ,H515269 H515185 ,H515276 H515189 ,H515293 H515190 ,H516012 H515191 ,H517567 H515192 ,H519999 H515193, H515196, H515197, H515198, H515200, H515201 |
Recalling Firm/ Manufacturer |
Varian Medical Systems Oncology Systems 911 Hansen Way Palo Alto CA 94304-1028
|
For Additional Information Contact | Dave Hall 650-424-5892 |
Manufacturer Reason for Recall | Software anomaly may result in a design deficiency causing Plan Edit Settings to malfunction under specific circumstances, where the user selects to use a Cone Beam CT set up field. |
FDA Determined Cause 2 | Device Design |
Action | The firm, Varian Medical Systems, sent a Product Notification Letter/Field Safety Letter entitled "URGENT MEDICAL DEVICE CORRECTION" dated October 16, 2009, to all affected customers. The letter described the product, problem, actions to be taken by the customers and Varian.
The customers were instructed to use caution when creating CBCT imaging field and limit use of the 'select all' button when acquiring treatment parameters; review departmental policies and procedures regarding the acquisition of any gantry angle at the 4D ITC, and advise the appropriate personnel of this letter.
The firm, Varian, recommended that only vertical, longitudinal and lateral couch parameters are acquired when creating a CBCT field at the 4D ITC; notified all affected customers of this document, and will correct this issue in a future release of the 4D ITC software.
If owners of $D ITC, version 8.0.12 and higher, have any questions regarding this product notification letter, please contact Varian Medical Systems Help Desk at 1-888-VARIANS (888-827-4265) for USA and Canada, +41 41 749 8844 or support-emea@varian.com for Europe;
Emails for North American: support-americas@varian.com; Australia/New Zealand: support-anz@varian.com; South East Asia: seasia.apps.helpdesk@varian.com; China/Asia: china.apps.helpdesk@varian.com; Japan: Japan.Apps.Helpdesk@varian.com and Latin American: soporte.al@varian.com, and
Internets for Oncology Systems- www.myvarian.com and Varian Medical Systems- www.varian.com. |
Quantity in Commerce | 3629 |
Distribution | Worldwide distribution: USA and countries including: Australia, Asia, China, Japan, Latin American, New Zealand , North America, South East Asia. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IYE
|
|
|
|