• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Freestyle Navigator Continuous Glucose Monitor System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Freestyle Navigator Continuous Glucose Monitor Systemsee related information
Date Initiated by FirmApril 14, 2009
Date PostedApril 30, 2010
Recall Status1 Terminated 3 on March 31, 2011
Recall NumberZ-1451-2010
Recall Event ID 52635
PMA NumberP050020 
Product Classification Continuous glucose monitor - Product Code MDS
ProductFreestyle Navigator Continuous Glucose Monitor System, ABB NAV Dutch System Kit CE RE, AND ABB SYSTEM KIT, MMOL, EMEA AND ABB09 Dutch Kit New LCD. Abbot Item Number 70966-01, distributed by Abbott Lab. Medical device indicated for continually recording interstitial fluid glucose levels in people ages 18+ with diabetes mellitus for improving diabetes management. The device is indicated to be used in hospital and home settings to aid health care.
Code Information Abbott Item # 70966-01 with multiple lot numbers, kits serial numbers, and Tx serial numbers. Lot numbers are these: 01M0028 or 01M0168 or 01K 2768 or 01K 2898 or 01K 3238 or 01K3328 or 01K3478 or 01M0028 or 01M0298 or or 01M1438 or 01M1638 or 01M2808. Kit Serial numbers beginning with BAAH 186 or BAAH 220 or BAAH 223 or BAAH 243 or BAAH 261 or BAAH 290 or BAAH 299 or BAAH 341and followed by a 5 digit number. TX Serial number beginning with AAAH 008 or AAAH 90 or AAAH 139 or AAAH 158 or AAAH 187 or AAAH195 or AAAH 212 or AAAH 221 or AAAH 228 or AAAH 239 or AAAH 268 or AAAH 296 or AAAH 354 and followed by a 5 digit number.
FEI Number 3002803444
Recalling Firm/
Manufacturer
Abbott Diabetes Care, Inc.
1360 South Loop Road
Alameda CA 94502-7000
For Additional Information Contact
510-749-5400
Manufacturer Reason
for Recall
Potential failure or inaccurate readings--the device's plastic housing near the battery compartment may crack, and allow moisture to enter.
FDA Determined
Cause 2
Process control
ActionNotification letters were issued to FreeStyle Navigator customers via certified mail and notification was posted on the Abbott Diabetes Care's Website. Customers were notified of the potential issue and actions to take: 1) if moisture is observed in the FreeStyle Navigator transmitter now or at any time, discontinue use of the transmitter and call Abbott Diabetes Care Customer Care at 1-800-418-6427 and 2) if a loss of connection is encountered or if your results are not reliable or are inconsistent with how you feel, consult your User's Guide for troubleshooting guidelines. If the issue is not resolved, call Abbott Diabetes Care Customer Care at 1-800-418-6421. For outside-of-US accounts, the firm's regional offices will distribute notices to identified customers in the appropriate language.
Quantity in Commerce5,449 of total units
DistributionWorldwide Distribution -- United States (OH, FL, NJ, CA, MS, MN, MA, and NY), France, Australia, the Netherlands, Sweden, the UK, Israel, Ireland and Germany.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA DatabasePMAs with Product Code = MDS
-
-