| Class 2 Device Recall Philips M1013A IntelliVue G1 Anesthesia Gas Modules with manifolds |  |
Date Initiated by Firm | July 21, 2010 |
Date Posted | August 24, 2010 |
Recall Status1 |
Terminated 3 on March 09, 2012 |
Recall Number | Z-2263-2010 |
Recall Event ID |
56273 |
510(K)Number | K060541 |
Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
Product | Philips M1013A IntelliVue G1 Anesthesia Gas Modules with manifolds |
Code Information |
Serial Number range ARYE-0014 to ASBC-0013 |
Recalling Firm/ Manufacturer |
Philips Healthcare Inc. 3000 Minuteman Road Andover MA 01810
|
For Additional Information Contact | 978-687-1501 |
Manufacturer Reason for Recall | The manifold seal in the Philips M1657B Water Trap used on the Philips M1013A IntelliVue G1, M1019A, IntelliVue G5, and Anesthesia Gas Modules M1026B/BT may be unintentionally removed when changing the traps. A missing seal may result in leakage and may cause incorrect monitor readings of inspiratory or expiratory gas levels. |
FDA Determined Cause 2 | Device Design |
Action | Philips Healthcare issued an "Urgent Device Correction" letter dated July 14, 2010 with an addendum to the Instructions for Use including additional instructions for the replacement of the water trap for the Philips M1013A IntelliVue G1, M1019A IntelliVue G5, M1026B/BT Anesthesia Gas Modules. This addendum is being sent along with the Urgent Medical Device Correction notice to all affected customers.
Customers will be requested to incorporate the information contained in the Urgent Medical Device Correction notice and the Addendum to the Instructions for Use with the existing labeling (Instructions for Use) provided with the product. Customers can contact Philips at 1-800-722-9377 for further information.. |
Quantity in Commerce | 531 units |
Distribution | Worldwide Distribution: United States, and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Chile ,China ,Colombia, Czech Republic, Estonia, Finland, France, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Republic of Korea, Latvia, Libyan Arab Jamahiriya, Luxembourg, Malaysia, Mexico, Mozambique, Netherlands, New Zealand, Norway, Oman, Pakistan, occupied Palestinian territory, Peru, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Viet Nam. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = BSZ
|
|
|
|