Class 1 Device Recall Huber TrapA Safety Huber Needle

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Huber TrapA Safety Huber Needlesee related information
Date Initiated by FirmAugust 30, 2010
Date PostedSeptember 29, 2010
Recall Status1 Terminated 3 on January 31, 2013
Recall NumberZ-2573-2010
Recall Event ID 56548
510(K)NumberK891919 
Product Classification Hypodermic Single Lumen Needle - Product Code FMI
ProductMarquette Medical Huber Trap-A Safety Huber Needle 20ga x .75" Huber Trap with Needleless Y Model # MW203Y-HT Intended to be used to inject fluids into, withdraw fluids from inside the vascular space through a subcutaneously implanted port.
Code Information Lot Code:   5H041M   5H071M   7E065M   8C072M   8H042M   8I046M   9A033M   9E028M  
FEI Number 1221765
Recalling Firm/
Manufacturer
Multi-Med, Inc.
26 Victoria Court
Keene NH 03431
For Additional Information Contactsame
603-357-8733
Manufacturer Reason
for Recall
Huber needles may core and result in port leakage or emboli being flushed our of the port.
FDA Determined
Cause 2
Device Design
ActionMulti-Med issued an Urgent: Medical Device Recall notification on 8/30/10 via fax and followed by hard copy letter. Users were advised of the problem and asked to immediately examine their inventory and quarantine the affected product. They were to also notify their customers of the recall. The firm stated in the letter that they would contact customers to determine the number of product that they had and the amount they expected from the field. The firm will coordinate a plan for the return of the affected product. Questions should be directed to Sue Starkey at 603-357-8733.
Quantity in Commerce6,400 units
DistributionWorldwide Distribution -- USA and Mexico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = FMI
-
-