| | Class 2 Device Recall Inion CPS |  |
| Date Initiated by Firm | April 14, 2010 |
| Date Posted | December 16, 2010 |
| Recall Status1 |
Terminated 3 on June 06, 2012 |
| Recall Number | Z-0635-2011 |
| Recall Event ID |
55861 |
| Product Classification |
Plate, bone - Product Code JEY
|
| Product | 1.5 mm Inion CPS, 4 hole plate, REF PLT-1005, sterile, Manufacturer, INION Oy, Finland. |
| Code Information |
Lot numbers of 0700000 and above. |
| FEI Number |
3005101424
|
Recalling Firm/ Manufacturer |
Stryker Craniomaxillofacial Division
750 Trade Centre Way
Ste 200
Portage
MI
49002-0482
|
| For Additional Information Contact | 877-534-2464 |
Manufacturer Reason for Recall | Inadequate Instructions for Use. The stated strength retention is ambiguous and the maximum immersion time of the plates/meshed is not defined. |
FDA Determined Cause 2 | Other |
| Action | Stryker issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 14, 2010. The notice informed users of the change in the instructions for use (IFU), and included a copy of the revised IFU. The consignees were informed that the affected product does not need to be returned.
Stryker can be contacted about this action at 877-534-2464 (option "Regulatory Affairs", reference Product Field Action #2009-445) |
| Quantity in Commerce | 8,631 all products |
| Distribution | Nationwide Distribution: Throughout the US |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|