Class 2 Device Recall Plum A3 Triple Channel Infusion Pumps

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Plum A3 Triple Channel Infusion Pumpssee related information
Date Initiated by FirmFebruary 14, 2011
Date PostedMarch 14, 2011
Recall Status1 Terminated 3 on November 08, 2016
Recall NumberZ-1645-2011
Recall Event ID 57724
510(K)NumberK031185 K042081 K052052 
Product Classification Pump, infusion - Product Code FRN
ProductPlum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software
Code Information a) list 12348 - serial numbers 0013740156 through 0013741025; b) list 12618 - serial numbers 0014040001 through 0018627599;  c) list 20678 - serial numbers 0015541001 through 0019616461 
Recalling Firm/
Manufacturer
Hospira Inc.
275 N Field Dr
Lake Forest IL 60045-2579
For Additional Information ContactMs. Ileana Quinones
224-212-2000
Manufacturer Reason
for Recall
Hospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe firm, Hospira, sent "URGENT DEVICE RECALL" letters dated February 14, 2011 to their customers on the same date. The letter described the product, problem and action to be taken by the customers. The Customer/Clinicians were instructed to weigh the risk associated with continued use of the device versus removal from service. If they elect to continue the devices, they were provided with instructions to perform alarm tests prior to each clinical use (during their pump cleaning process). If the alarm is not audible, they were instructed to discontinue use of the pump and contact Hospira Global Product Safety and Complaints at 1-800-441-4100. The customers were requested to contact Hospira Advanced Knowledge Center at 1-800-241-4002 for technical assistance. The accounts were also requested to complete the attached Reply Form and return it to Hospira via fax at 1-888-216-7330.
Quantity in Commerce23,020 pumps
DistributionWorldwide distribution: United States including the U.S. Virgin Islands and Puerto Rico, and countries including: Argentina, Australia, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Uruguay, United Arab Emirates and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.